(232 days)
The noninvasive INVOS 5100C is intended for use as an adjunct monitor of regional hemoglobin oxygen saturation of blood in the brain or in other tissue beneath the sensor. It is intended for use in individuals greater than 2.5 kg at risk for reduced-flow or no-flow ischemic states. It is also intended for use as an adjunct trend monitor of regional hemoglobin oxygen saturation of blood in the brain or in other tissue beneath the sensor in any individual. The clinical value of trend data has not been demonstrated in disease states. The INVOS System should not be used as the sole basis for diagnosis or therapy.
The INVOS 5100C is a 2 wavelength, diffuse reflectance spectroscopy system employing near infrared light to estimate the percentage of hemoglobin saturated with oxygen in tissue underneath the sensor. An adhesive sensor containing a light source and 2 photodiodes is applied to the skin over the tissue of interest and the returning light is analyzed for oxyhemoglobin and deoxyhemoglobin light absorption. Absorption signals from the photodiode closer to the light source are subtracted from those from the farther photodiode where the returning photons penetrate more deeply in the tissue. This suppresses absorption events originating in the outer layers of tissue that are common to both photodiodes, including the effects of skin pigmentation and subcutaneous tissues. This method of “spatial resolution” also allows estimation of scattering to improve measurement accuracy.
The provided text is a 510(k) Premarket Notification for the SOMANETICS INVOS 5100C Cerebral/Somatic Oximeter System. The document focuses on establishing substantial equivalence to a predicate device (Somanetics INVOS 5100C Cerebral/Somatic Oximeter System, K080769) for revised indications for use and labeling.
Based on the provided text, the specific details regarding acceptance criteria and a study proving those criteria are not explicitly present. The document states: "Extensive literature references and clinical studies are submitted demonstrating the substantial equivalence of the device for its stated indication." However, it does not provide the details of these studies, including acceptance criteria, sample sizes, ground truth establishment, or performance metrics.
Therefore, I cannot fully answer all the requested questions with the provided input. Below is an outline of what can and cannot be extracted:
1. A table of acceptance criteria and the reported device performance:
- Acceptance Criteria: Not explicitly stated in the provided text.
- Reported Device Performance: Not explicitly stated in terms of specific metrics against acceptance criteria. The document only mentions "extensive literature references and clinical studies... demonstrating the substantial equivalence."
2. Sample sized used for the test set and the data provenance:
- Sample Size for Test Set: Not explicitly stated.
- Data Provenance (country of origin, retrospective/prospective): Not explicitly stated.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
- Not explicitly stated.
4. Adjudication method (e.g., 2+1, 3+1, none) for the test set:
- Not explicitly stated.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
- Not applicable/Not stated. The device is an oximeter, not an AI-powered diagnostic imaging tool that would typically involve human reader studies in this context. The document refers to "regional hemoglobin oxygen saturation of blood," which is a physiological measurement.
6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:
- The INVOS 5100C is an oximeter system, which inherently functions as a standalone measurement device. Its performance relates to its accuracy in measuring regional hemoglobin oxygen saturation. No "algorithm only" performance separate from the device itself is detailed in terms of a study.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):
- Not explicitly stated. For an oximeter, the "ground truth" would typically involve a reference standard for blood oxygen saturation measurements (e.g., co-oximetry of arterial or venous blood samples) for accuracy studies.
8. The sample size for the training set:
- Not explicitly stated. The document refers to "extensive literature references and clinical studies" but does not detail their methodology or if a "training set" in the machine learning sense was involved.
9. How the ground truth for the training set was established:
- Not explicitly stated.
Conclusion Drawn from the Testing (as stated in the document):
"The conclusion drawn is that the revised indications for use and labeling are substantially equivalent to the predicate device and do not raise new questions of safety and effectiveness."
In summary, the provided text does not contain the detailed information necessary to answer most of your questions about specific acceptance criteria and study methodologies, as it is a 510(k) summary focused on substantial equivalence rather than a detailed clinical study report.
§ 870.2700 Oximeter.
(a)
Identification. An oximeter is a device used to transmit radiation at a known wavelength(s) through blood and to measure the blood oxygen saturation based on the amount of reflected or scattered radiation. It may be used alone or in conjunction with a fiberoptic oximeter catheter.(b)
Classification. Class II (performance standards).