K Number
K170876
Date Cleared
2017-08-23

(152 days)

Product Code
Regulation Number
870.1025
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Passport 17m and Passport 12m patient monitors are intended for monitoring, displaying, alarming, and transferring of multiple physiological parameters including ECG (3-lead or 12-lead selectable, arrhythmia detection. ST segment analysis, oT analysis, and heart rate (HR)), respiration rate (Resp), temperature (Temp), pulse oxygen saturation (SpO2), pulse rate (PR), non-invasive blood pressure (NIBP), invasive blood pressure(IBP), pulmonary artery wedge pressure (PAWP), cardiac output (C.O.), continuous cardiac output (CCO), mixed/central venous oxygen saturation (Sv02/Scv02), carbon dioxide (C02), Oxygen (02), anesthetic gas (AG), impedance cardiograph (ICC), bispectral index (BIS), respiration mechanics (RM), and neuromuscular transmission monitoring (NMT). The equipment also provides an interpretation of resting 12-lead ECG.

All the parameters can be monitored on single adult, pediatric, and neonatal patients with the exception of the following:
• The arrhythmia detection, ST Segment analysis of Mortara algorithm, BIS, RM, CCO, SvO2/Scv02, PAWP monitoring and NMT monitoring are intended for adult and pediatric patients only;

  • · ST Segment analysis of Mindray algorithm is intended for adult patients only:
  • · C.O. monitoring is restricted to adult patients only;
    • ICG monitoring is only for use on adult patients who meet the following requirements: height: 122 to 229cm, weight: 30 to 155kg.

The monitor is to be used in healthcare facilities by clinical professionals or under their guidance. It should only be used by persons who have received adequate training in its use. It is not intended for helicopter transport, hospital ambulance, or home use.

Device Description

The subject Passport Series Patient Monitors includes two monitors:

  • . Passport 12m Patient Monitor
  • Passport 17m Patient Monitor ●

The Passport 12m and 17m Patient monitors provide patient monitoring capabilities based on the user selected modules that are plugged into the main monitor.

AI/ML Overview

The provided text describes a 510(k) premarket notification for the Passport Series Patient Monitors (Passport 12m and Passport 17m). The information mainly focuses on establishing substantial equivalence to predicate devices and does not detail specific acceptance criteria or a dedicated study proving performance against these criteria for the entire device.

However, the document does contain a "Device Comparison Table" that includes performance specifications for the Neuromuscular Transmission (NMT) module, which is a new feature for the subject device compared to the primary predicate, but present in a secondary predicate. This section implies a comparison of performance against the secondary predicate for the NMT module.

Based on the provided text, here's an attempt to extract the requested information, primarily focusing on the NMT module due to the lack of detailed performance study information for the overall patient monitor device:


1. Table of Acceptance Criteria and the Reported Device Performance

For the NMT module, the "acceptance criteria" can be inferred from the performance specifications of the secondary predicate device (Philips IntelliVue NMT module) and the "reported device performance" refers to the subject Mindray NMT module's specifications.

Feature / ParameterAcceptance Criteria (Secondary Predicate Device: Philips NMT - K161531)Reported Device Performance (Subject Mindray NMT)Compliance / Comparison
Stimulation Output
Current pulse type100, 200, or 300μs; monophasic rectangle pulse100, 200 or 300μs; monophasic rectangle pulse; Pulse width Accuracy: ±10%Same
Current Range5 to 60 mA in increments of 5 mA0 - 60 mA in increments of 5 mAActual output current of both sides are 5 to 60 mA. Same.
Current Accuracy± 5% or ± 2 mA, whichever is greater± 5% or ± 2 mA, whichever is greaterSame
Max. Skin Resistance / Impedance3 kOhm3kohmSame
Max. Output Voltage300 V300 VSame
Stimulation ModesSingle Twitch (Twitch); Train-Of-Four (TOF); Post-Tetanic Count (PTC); Double-Burst Stimulation (DBS)Single Twitch (ST); Train-Of-Four (TOF); Post-Tetanic Count (PTC); Double-Burst Stimulation (DBS)Same
Performance Specifications
Single Twitch Stimulation Mode:
Twitch / ST-Ratio0 to 200%0 to 200%Same
Measurement interval (ST)Manual, or 1s, 12s, 30sManual, 1s, 10s, 20sDifferent Measurement interval
Train-Of-Four Stimulation Mode:
TOF Count0 to 40 to 4Same
TOF Ratio5 to 150%5 to 160%Wider TOF-Ratio Range
Measurement interval (TOF)Manual, or 12s, 30s, 1min, 5min, 10min, 15min, 30min, 60minManual, 12s, 15s, 20s, 30s, 1min, 5min, 15min, 30min, 60minDifferent Measurement interval
Post-Tetanic-Count Stimulation Mode:
PTC0 to 200 to 20Same
Measurement interval (PTC)ManualManualSame
Double-Burst Stimulation Mode:
Measurement interval (DBS)ManualManual, 15s, 20s, 30s, 1min, 5min, 15min, 30min, 60minMore Measurement intervals

Study Proving Device Meets Acceptance Criteria:

The document states: "To establish the substantial equivalence of the Passport Series Patient Monitors (Passport 12m and Passport 17m), Mindray conducted functional and system level testing on the subject devices. The testing provided an evaluation of the performance of the device relevant to each of the modifications to the subject devices since their previous clearance. The functional and system level testing showed that the devices continue to meet specifications and the performance of the device is equivalent to the predicate."

For the NMT module specifically, the comparison table indicates that while some measurement intervals and the TOF-Ratio range differ, the underlying principles and most specifications are either "Same" or provide a "Wider TOF-Ratio Range" or "More Measurement intervals," implying that these variations are acceptable or represent an improvement.

2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

The document does not provide details on the sample size used for the test set for performance validation for the NMT module or other parameters. It only mentions "functional and system level testing." The data provenance (country of origin, retrospective/prospective) is also not specified.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

This information is not provided in the document.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

This information is not provided in the document.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

This document describes a patient monitor, not an AI-assisted diagnostic tool. Therefore, an MRMC comparative effectiveness study involving human readers and AI assistance is not applicable and not mentioned.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

The document describes the performance of a medical device (patient monitor) and its modules, including the NMT module. The performance described in the comparison table are standalone specifications for the device's functionality. There is no explicit mention of an "algorithm only" study in the context of the device's overall operation, as it is a physical monitoring device. The NMT module's performance characteristics (e.g., current ranges, accuracies, measurement modes) are inherent to its design and function.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

For the NMT module's performance specifications, the "ground truth" implicitly would be established through calibrated laboratory equipment and standard measurement techniques to verify the accuracy of stimulation outputs, ranges, and timing. For other physiological parameters, similar direct measurement against known standards would apply. The document does not explicitly state the methodology for establishing ground truth for any of the parameters.

8. The sample size for the training set

This is not applicable as the document describes a physical patient monitor with integrated measurement capabilities, not an AI/machine learning model that requires a "training set."

9. How the ground truth for the training set was established

This is not applicable for the reasons stated above.


Summary of Limitations Based on Document:

The provided 510(k) summary focuses on demonstrating "substantial equivalence" of the Passport Series Patient Monitors to existing predicate devices, particularly regarding modifications and the addition of the NMT module. It provides a technical comparison of specifications rather than detailed clinical study data with explicit acceptance criteria and performance against those for all parameters. Information on sample sizes, data provenance, expert involvement for ground truth, and specific study methodologies (beyond "functional and system level testing") is not present in this document.

{0}------------------------------------------------

Image /page/0/Picture/2 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo consists of a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" written around it. Inside the circle is a symbol of three stylized human profiles facing to the right, stacked on top of each other.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

August 23, 2017

Shenzhen Mindray Bio-medical Electronics Co., LTD. Yanhong Bai Manager Regulatory Affairs Mindrav Building. Keii 12th Road South Hi-tech Industrial Park, Nanshan Shenzhen, 518057, P.R. China

Re: K170876

Trade/Device Name: Passport Series Patient Monitors (Passport 12m, Passport 17m) Regulation Number: 21 CFR 870.1025 Regulation Name: Arrhythmia detector and alarm Regulatory Class: Class II Product Code: MHX, DSI, MLD, DRT, DXN, DSK, FLL, DQA, DPZ, CCK, CBQ, CBS, CBR, CCL, DSB, DXG, OLW, DSJ, KOI, GXY Dated: July 14, 2017 Received: July 17, 2017

Dear Yanhong Bai:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply

{1}------------------------------------------------

with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely,

Jessica E. Paulsen -S

for

Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K170876

Device Name

Passport Series Patient Monitors (Passport 12m and Passport 17m)

Indications for Use (Describe)

The Passport 17m and Passport 12m patient monitors are intended for monitoring, displaying, alarming, and transferring of multiple physiological parameters including ECG (3-lead or 12-lead selectable, arrhythmia detection. ST segment analysis, oT analysis, and heart rate (HR)), respiration rate (Resp), temperature (Temp), pulse oxygen saturation (SpO2), pulse rate (PR), non-invasive blood pressure (NIBP), invasive blood pressure(IBP), pulmonary artery wedge pressure (PAWP), cardiac output (C.O.), continuous cardiac output (CCO), mixed/central venous oxygen saturation (Sv02/Scv02), carbon dioxide (C02), Oxygen (02), anesthetic gas (AG), impedance cardiograph (ICC), bispectral index (BIS), respiration mechanics (RM), and neuromuscular transmission monitoring (NMT). The equipment also provides an interpretation of resting 12-lead ECG.

All the parameters can be monitored on single adult, pediatric, and neonatal patients with the exception of the following: • The arrhythmia detection, ST Segment analysis of Mortara algorithm, BIS, RM, CCO, SvO2/ScvO2, PAWP monitoring and NMT monitoring are intended for adult and pediatric patients only;

  • · ST Segment analysis of Mindray algorithm is intended for adult patients only:
  • · C.O. monitoring is restricted to adult patients only;

• ICG monitoring is only for use on adult patients who meet the following requirements: height: 122 to 229cm, weight: 30 to 155kg.

The monitor is to be used in healthcare facilities by clinical professionals or under their guidance. It should only be used by persons who have received adequate training in its use. It is not intended for helicopter transport, hospital ambulance, or home use.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D) ×
Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(K) SUMMARY

In accordance with 21 CFR 807.87(h) and (21 CFR 807.92) the 510(k) Summary for the Passport Series Patient Monitors is provided below.

Device Common Name:Patient Monitor
Device Proprietary Name:Passport Series Patient Monitors (Passport 12mand Passport 17m)
Submitter:Shenzhen Mindray Bio-medical Electronics Co., LTD.Mindray Building, Keji 12th Road South,Hi-tech Industrial Park, Nanshan,Shenzhen, 518057, P. R. ChinaTel: +86 755 8188 8998Fax: +86 755 2658 2680
Contact:Ms. Yanhong BaiManager Regulatory AffairsShenzhen Mindray Bio-medical Electronics Co., LTD.Mindray Building, Keji 12th Road South,High-tech Industrial Park, Nanshan, Shenzhen518057, P.R. ChinaTel: +86 755 81885635Fax: +86 755 26582680E-mail: baiyanhong@mindray.com
Date Prepared:March 17, 2017
Classification Regulation:21 CFR 870.1025, Class II, Arrhythmia detector and alarm(including ST-segment measurement and alarm)
Panel:Cardiovascular

{4}------------------------------------------------

ProductCodeRegulationNumberPanelRegulation descriptionDevice Common Name
Primary
MHX21 CFR870.1025CardiovascularArrhythmia detector and alarm (including ST-segment measurement and alarm)monitor, physiological, patient(with arrhythmia detection or alarms)
Secondary
ProductCodeRegulationNumberPanelRegulation descriptionDevice Common Name
DSI21 CFR870.1025CardiovascularArrhythmia detector and alarm (including ST-segment measurement and alarm)detector and alarm, arrhythmia
MLD21 CFR870.1025CardiovascularArrhythmia detector and alarm (including ST-segment measurement and alarm)monitor, st segment with alarm
DRT21 CFR870.2300CardiovascularCardiac Monitor (including cardiotachometer and rate alarm)monitor, cardiac (incl. cardiotachometer & rate alarm)
DXN21 CFR870.1130CardiovascularNoninvasive blood pressure measurement systemsystem, measurement, blood-pressure, non-invasive
DSK21 CFR870.1110CardiovascularBlood pressure computercomputer, blood-pressure
FLL21 CFR880.2910CardiovascularClinical electronic thermometerthermometer, electronic, clinical
DQA21 CFR870.2700CardiovascularOximeteroximeter
DPZ21 CFR870.2710CardiovascularEar oximeteroximeter, ear
CCK21 CFR868.1400AnesthesiologyCarbon dioxide gas analyzeranalyzer, gas, carbon-dioxide, gaseous-phase
CBQ21 CFR868.1500AnesthesiologyEnflurane gas analyzeranalyzer, gas, enflurane, gaseous-phase (anesthetic concentration)
CBS21CFRAnesthesiologyHalothane gas analyzeranalyzer, gas, halothane, gaseous-phase (anesthetic conc.)
CBR21CFRAnesthesiologyNitrous oxide gas analyzeranalyzer, gas, nitrous-oxide, gaseous phase (anesthetic conc.)
CCL21CFRAnesthesiologyOxygen gas analyzeranalyzer, gas, oxygen, gaseous-phase
DSB21CFRCardiovascularImpedance plethysmographplethysmograph, impedance
DXG21CFRCardiovascularSingle-function, preprogrammed diagnostic computercomputer, diagnostic, pre-programmed, single-function
OLW21CFRNeurologyElectroencephalographindex-generating electroencephalograph software
DSJ21CFRCardiovascularBlood pressure alarmalarm, blood-pressure
KOI21CFRAnesthesiologyElectrical peripheral nerve stimulatorstimulator, nerve, peripheral, electric
GXY21CFRNeurologyCutaneous electrode.electrode, cutaneous
868.1620
868.1700
868.1720
870.2770
870.1435
882.1400
870.1100
868.2775
870.1320

Classification Regulation, Classification Name and Product Codes:

{5}------------------------------------------------

Primary Predicate Device: K152902 - Passport Series Patient Monitors (Passport 12m, Passport 17m, and T1); Shenzhen Mindray Bio-Medical Electronics Co., Ltd

Secondary Predicate Device:

K161531 - IntelliVue Patient Monitors (Philips Medizin Systeme Boeblingen GmbH)

Indications for Use:

The Passport 17m and Passport 12m patient monitors are intended for monitoring, displaying, reviewing, alarming, and transferring of multiple physiological parameters including ECG (3-lead, 5-lead or 12-lead selectable, arrhythmia detection, ST segment analysis, QT analysis, and heart rate (HR)), respiration rate (Resp), temperature (Temp), pulse oxygen saturation (SpO2), pulse rate (PR), non-invasive blood pressure (NIBP), invasive blood pressure(IBP), pulmonary artery wedge pressure (PAWP), cardiac output (C.O.), continuous cardiac output (CCO), mixed/central venous oxygen saturation (SvO2/ScvO2), carbon dioxide (CO2), Oxygen (O2), anesthetic gas (AG), impedance cardiograph (ICG), bispectral index (BIS), and

{6}------------------------------------------------

respiration mechanics (RM), neuromuscular transmission monitoring(NMT). The equipment also provides an interpretation of resting 12-lead ECG.

All the parameters can be monitored on single adult, pediatric, and neonatal patients with the exception of the following:

  • The arrhythmia detection, ST Segment analysis of Mortara algorithm, BIS, RM, CCO, SvO2/ScvO2, PAWP monitoring and NMT are intended for adult and pediatric patients only:
  • ST Segment analysis of Mindray algorithm is intended for adult patients only;
  • C.O. monitoring is restricted to adult patients only;
  • ICG monitoring is only for use on adult patients who meet the following requirements: ■ height: 122 to 229cm, weight: 30 to 155kg.

The monitor is to be used in healthcare facilities by clinical professionals or under their guidance. It should only be used by persons who have received adequate training in its use. It is not intended for helicopter transport. hospital ambulance, or home use.

Device Description:

The subject Passport Series Patient Monitors includes two monitors:

  • . Passport 12m Patient Monitor
  • Passport 17m Patient Monitor ●

The Passport 12m and 17m Patient monitors provide patient monitoring capabilities based on the user selected modules that are plugged into the main monitor.

Performance Data:

  • To establish the substantial equivalence of the Passport Series Patient Monitors (Passport 12m and Passport 17m), Mindray conducted functional and system level testing on the subject devices. The testing provided an evaluation of the performance of the device relevant to each of the modifications to the subject devices since their previous clearance. The functional and system level testing showed that the devices continue to meet specifications and the performance of the device is equivalent to the predicate.
  • In addition, Mindray has conducted testing to ensure the subject devices meet relevant consensus standards.
    • IEC 60601-1:2005+A1:2012 Medical electrical equipment -- Part 1: General . requirements for basic safety and essential performance
    • . IEC 60601-1-2:2007 Medical electrical equipment – Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances - Requirements and tests
    • IEC 60601-2-10:2012 Medical electrical equipment Part 2-10: Particular . requirements for the basic safety and essential performance of nerve and muscle stimulators
    • IEC 60601-2-49:2011 Medical electrical equipment Part 2-49: Particular ●

{7}------------------------------------------------

requirements for the basic safety and essential performance of multifunction patient monitoring equipment

  • . ISO 80601-2-55:2011 Medical electrical equipment -- Part 2-55: Particular requirements for the basic safety and essential performance of respiratory gas monitors

Substantial Equivalence:

Comparison of Indications - Both the predicate device and the subject Passport 12m and Passport 17m devices are multiparameter patient monitors intended to be used in healthcare facilities under the direction of clinical professionals. The indications for use of the subject devices (Passport 12m and Passport 17m) have been modified to include the new module of NMT. Although this feature is not present in the predicate device, it is present in the cleared Philips (K161531), and thus does not constitute a new intended use for a multi-parameter monitor. In conclusion, the minor changes to the indications for use do not change the fundamental intended use of the Passport 12m and Passport 17m as multiparameter monitors.

Comparison of Technological Characteristics - The table below compares the key technological feature of the subject devices to the primary predicate device (Passport Series Patient Monitors K152902). The features in grey are the features that have been modified since their previous clearances and that are the subject of this 510(k).

Predicate Device (K152902)Subject Devices
FeaturePassport 17mPassport 12mPassport 17mPassport 12m
Integrated display and touchscreen17" 1280*1024 pixels12" 800*600 pixelsSameSame
Secondary displayIndependent control and displayDisplay is linked to integrated displaySameSame
Additional display featuresThe minitrends diagram, OxyCRG diagram, other monitor view, and calculation can be viewed when using an external LCD screenSameSame
Predicate Device (K152902)Subject Devices
FeaturePassport 17mPassport 12mPassport 17mPassport 12m
WirelessmoduleThe ASUS module isused for connecting to anetwork wirelessly,constructing a monitoringnetwork with a centralmonitoring system(CMS).The ASUS and Silexmodules are used forconnecting to a networkwirelessly, constructing amonitoring network with acentral monitoring system(CMS).SameLaird(2.4G/5G)wirelessmodule isadded.
Module rackIndependent of the patient monitor, provides 8 standardmodule slots to extend the measurement capabilities ofthe systemSame
Power supplyTwo rechargeable Lithium-ionbattery or AC power supplyOne rechargeableLithium-ion batteryor AC power supplySameSame
BatteryChargeable Lithium-Ion, 11.1 VDC, 4500 mAh, 350 gSame
Externalmemory cardCompact FlashSame
Data RecorderThe thermal recorder records patient information,measurement numerics, up to three waveforms, etc.Same
SpeakerGive alarm tones (45 to 85 dB), key tones, QRS tones;support PITCH TONE and multi-level tone modulationSame
Supports T1 asa moduleSupportedSame
ECG3-lead , 5-lead or 12-lead selectable, arrhythmiadetection, ST segment analysis, QT analysis, heart rate(HR), an interpretation of resting 12-lead ECG, J-pointAuto detection, Dual Channel Pace detection andAdjustable QRS Detection ThresholdSame
ArrhythmiaAnalysis(MindrayAlgorithm)Asystole, VFib/VTac, Vtac, Vent. Brady, Extreme Tachy,Extreme Brady, PVC, Couplet, Bigeminy, Trigeminy, Ron T, Run PVCs, PVCs, Tachy, Brady, Missed Beats,Vent. Rhythm, PNP, PNC, Multif. PVC, Nonsus. Vtac,Pause, Irr. Rhythm, AFibSame
ArrhythmiaAnalysis(MortaraAlgorithm)Asystole, Vfib, Vtac, Vent. Rhythm, Couplet, Run PVCs,Bigeminy, Trigeminy, R on T, Multif. PVC, Irr. Rhythm,Tachy, Brady, Missed Beats, PNP, PNC, PVCsSame
Predicate Device (K152902)Subject Devices
FeaturePassport 17mPassport 12mPassport 17mPassport 12m
Respiration rate(Resp)Measurement range:Adult: 0 to 120 rpm; Pediatric, neonate: 0 to 150 rpm.Accuracy:7 to 150 rpm: ±2 rpm or ±2%, whichever is greater;0 to 6 rpm: Not specified.Same
Temperature(Temp)Measurement range: 0 to 50°C (32 to 122°F)Accuracy: ±0.1°C or ±0.2°F (without probe)Same
Pulse oxygensaturation(SpO2)Mindray SpO2 ModuleMeasurement range: 0 to 100%Accuracy:70 to 100%: ±2% (adult/pediatric mode)70 to 100%: ±3% (neonate mode)0% to 69%: Not specified.Masimo SpO2 ModuleMeasurement range: 1 to 100%Accuracy:70 to 100%: ±2% (measured without motion inadult/pediatric mode)70 to 100%: ±3% (measured without motion in neonatemode)70 to 100%: ±3% (measured with motion)1% to 69%: Not specified.Nellcor SpO2 ModuleMeasurement range: 0 to 100%Accuracy:70 to 100%: ±2% (adult/pediatric)70 to 100%: ±3% (neonate)0% to 69%: Not specifiedSame
Predicate Device (K152902)Subject Devices
FeaturePassport 17mPassport 12mPassport 17mPassport 12m
Pulse rate (PR)PR from Mindray SpO2 ModuleMeasurement range: 20 to 254 bpmAccuracy: ±3 bpmPR from Masimo SpO2 ModuleMeasurement range: 25 to 240 bpmAccuracy: ±3 bpm (measured without motion)±5 bpm (measured with motion)PR from Nellcor SpO2 ModuleMeasurement range: 20 to 300 bpmAccuracy:20 to 250 bpm: ±3 bpm251 to 300 bpm, not specifiedPR from IBP ModuleMeasurement range: 25 to 350 bpmAccuracy:±1 bpm or ±1%, whichever is greater"Same
Non-invasive blood pressure (NIBP)Measurement range:AdultNeonateSystolic: 40 to 270Diastolic: 10 to 210Mean: 20 to 230Accuracy:Max mean error: ±5 mmHgMax standard deviation: 8 mmHgPediatricNeonateSystolic: 40 to 200Diastolic: 10 to 150Mean: 20 to 165SamePediatricNeonateSystolic: 40 to 135Diastolic: 10 to 100Mean: 20 to 110
Invasive blood pressure (IBP)Measurement range: -50 to 300 mmHgAccuracy: ±2% or ±1 mmHg, whichever is greater (without sensor)Same
Pulse Pressure Variation (PPV)Supported feature of IBPSame
Cardiac output (C.O.)The temperature change is displayed as a curve in the C.O. split screen, and the monitor calculates the C.O. value from this curve. The monitor is capable of storing 6 measurements.Same
Predicate Device (K152902)Subject Devices
FeaturePassport 17mPassport 12mPassport 17mPassport 12m
Continuouscardiac output(CCO)CCO/SvO2 interface module is used to interfacewith Edwards Vigilance II monitor / VigileoMonitor which measures continuous cardiacoutput (CCO), mixed venous oxygen saturation(SvO2) and central venous oxygen saturation(ScvO2).Same
Central venousoxygensaturation(ScvO2)ScvO2 module is used to measure central venous oxygensaturation (ScvO2).Same
Carbon dioxide(CO2)Sidestream CO2 Module:Measurement range: 0 to 99 mmHgAccuracy:0 to 40 mmHg: ±2 mmHg41 to 76 mmHg: ±5% of the reading77 to 99 mmHg: ±10% of the readingMicrostream CO2 Module:Measurement range: 0 to 99 mmHgAccuracy:0 to 38 mmHg: ±2 mmHg39 to 99 mmHg: ±5% of the reading ±5% of thereading+0.08% of (the reading-38)Mainstream CO2 Module:Measurement range: 0 to 150 mmHgAccuracy:0 to 40 mmHg: ±2 mmHg41 to 70 mmHg: ±5% of the reading71 to 100 mmHg: ±8% of the reading101 to 150 mmHg: ±10% of the readingSame
Anesthetic gas(AG)The AG module analyzes gas samples from the patientand calculates CO2, O2, N2O and AA waves and relatednumerics, airway respiratory rate, and MAC (minimumalveolar concentration).Supports the 3-slot AG module.Added support for the2-slot AG module
Predicate Device (K152902)Subject Devices
FeaturePassport 17mPassport 12mPassport17mPassport12m
Impedancecardiograph(ICG)Measurement range:SV: 5 to 250 mlHR: 44 to 2mC.O. 1.4 to 15 L/minAccuracy:SV: Not specified.HR: ±2 bpmC.O. Not specified.Same
Bispectral index(BIS)Measured parameters: EEG, BIS, BIS L, BIS RSame
Respirationmechanics(RM)Measurement range: Adult: 0 to 120 rpm; Pediatric: 0 to150 rpm.Accuracy: 7 to 150 rpm:±2 rpm or ±2%, whichever isgreater;0 to 6 rpm: Not specified."Same
NMTNot supportedAdded

Device Comparison Table (Compare with Passport 12m/17m (K152902))

{8}------------------------------------------------

{9}------------------------------------------------

{10}------------------------------------------------

{11}------------------------------------------------

{12}------------------------------------------------

{13}------------------------------------------------

FeatureSecondary Predicate DeviceNMT (K152902)Subject Mindray NMT in this510(k)ComparisonSystem
CompositionModule and accessory cableModule and accessory cableSame
Intended useThe IntelliVue NMTModule is intended to beused as an objectiveneuromuscular transmissionmonitor, usingaccelerometry formeasuring the musclecontraction following anelectrical stimulation of aperipheral nerve. The NMTModule is intended to beused with adult andpediatric patients.The neuromusculartransmission (NMT) moduleis intended to be used as anobjective neuromusculartransmission monitor, usingaccelerometry for measuringthe muscle contractionfollowing an electricalstimulation of a peripheralnerve. The NMT module isintended to be used with adultand pediatric patients.Same
Principle ofmeasurementNeuro MuscularTransmission (NMT) andits measurement enables theevaluation of musclerelaxation of patients underNeuromuscular Block bymeasuring the strength ofmuscle reaction afterelectrically stimulating thededicated motor nerve.The NMT Monitorelectrodes are placed on thepatient's skin over the ulnarnerve, a controllable currentsource delivers stimulationpulses to two skin surfaceelectrodes for the nervestimulation, the muscleresponse is measured withan acceleration sensor.Neuromuscular Transmission(NMT) module evaluatesmuscle relaxation of patientsunder neuromuscular block bymeasuring the strength ofmuscle reaction afterelectrically stimulating thededicated motor nerve.The electrodes are placed onthe patient's skin over ulnarnerve, a controllable currentsource delivers stimulationpulses to two skin surfaceelectrodes for the nervestimulation, and the muscleresponse is measured with anacceleration sensor.Same
FeatureSecondary Predicate DeviceNMT (K152902)Subject Mindray NMT in this510(k)ComparisonSystem
StimulationOutputCurrent pulse: 100, 200, or300μs; monophasicrectangle pulseCurrent Pulse: 100, 200 or300μs; monophasic rectanglepulse;Pulse width Accuracy: ±10%Same
Current Range: 5 to 60 mAin increments of 5 mACurrent Range:0 - 60 mA inincrements of 5 mAthe actual outputcurrent of thetwo sides areboth 5 to 60 mASame
Current Accuracy: ± 5% or± 2 mA, whichever isgreaterCurrent Accuracy: ± 5% or ±2 mA, whichever is greaterSame
Max. Skin Resistance: 3kOhmMax. Skin Impedance: 3kohmSame
Max. Output Voltage:300 VMax. output voltage: 300 VSame
StimulationModesSingle Twitch (Twitch); Train-Of-Four (TOF); Post-Tetanic Count (PTC); Double-Burst Stimulation (DBS)Single Twitch (ST); Train-Of-Four (TOF); Post-Tetanic Count (PTC); Double-Burst Stimulation (DBS)Same
PerformanceSpecificationsSingle Twitch StimulationMode:Twitch: 0 to 200%Measurement interval:Manual, or 1s, 12s, 30sSingle Twitch StimulationMode:ST-Ratio: 0 to 200%Measurement interval:Manual, 1s, 10s, 20sDifferentMeasurement interval
FeatureSecondary Predicate DeviceNMT (K152902)Subject Mindray NMT in this510(k)ComparisonSystem
Train-Of-Four StimulationMode:TOF Count (TOFcnt): 0 to4TOF Ratio (TOFrat): 5 to150%-Measurement interval:Manual, or 12s, 30s, 1min,5min, 10min, 15min,30min, 60minTrain-Of-Four StimulationMode:TOF-Count: 0 to 4TOF-Ratio: 5 to 160%Measurement interval:Manual, 12s, 15s, 20s, 30s,1min, 5min, 15min, 30min,60minWiderTOF-RatioRange; DifferentMeasurementinterval
Post-Tetanic-CountStimulation Mode:PTC:0 to 20Measurement interval:ManualPost-Tetanic-CountStimulation Mode:PTC: 0 to 20Measurement interval:ManualSame
Double-Burst StimulationMode:Measurement interval:ManualDouble-Burst StimulationMode:Manual, 15s, 20s, 30s, 1min,5min, 15min, 30min, 60minMoreMeasurementintervals

Device Comparison Table (Compare with Philips NMT module (K161531))

{14}------------------------------------------------

{15}------------------------------------------------

Substantial Equivalence Conclusion:

Based on the detailed comparison of specifications for each of the modifications to the primary predicate devices, Passport 12m and Passport 17m (K152902), the secondary predicate devices, IntelliVue Patient Monitors(K161531) , the performance testing results and conformance with applicable standards show that the Passport Series Patient Monitors (Passport 12m and Passport 17m) can be found substantially equivalent to the predicate devices.

§ 870.1025 Arrhythmia detector and alarm (including ST-segment measurement and alarm).

(a)
Identification. The arrhythmia detector and alarm device monitors an electrocardiogram and is designed to produce a visible or audible signal or alarm when atrial or ventricular arrhythmia, such as premature contraction or ventricular fibrillation, occurs.(b)
Classification. Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Arrhythmia Detector and Alarm” will serve as the special control. See § 870.1 for the availability of this guidance document.