Search Results
Found 5 results
510(k) Data Aggregation
(86 days)
The NuVasive CoRoent Thoracolumbar System is indicated for intervertebral body fusion of the spine in skeletally mature patients. The System is designed for use with autogenous and/or allogeneic bone graft comprised of cancellous and/or corticocancellous bone graft to facilitate fusion and supplemental internal spinal fixation systems that are cleared by the FDA for use in the thoracolumbar spine. The devices are to be used in patients who have had at least six months of non-operative treatment.
The Nu Vasive CoRoent Thoracolumbar System (XL platform) implants are intended for use in interbody fusions in the thoracic spine, from T1 to T12, and at the thoracolumbar junction (T12-L1), and the CoRoent Thoracolumbar System (XL and L platforms) implants are intended for use in the lumbar spine, from L1 to S1, for the treatment of symptomatic disc degeneration (DDD) or degenerative spondylolisthesis at one or two adjacent levels, including thoracic disc herniation (with myelopathy and/or radiculopathy with or without axial pain). DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. The NuVasive CoRoent Thoracolumbar System (XL and L platforms) can be used as an adjunct to fusion in patients diagnosed with multilevel degenerative scoliosis and sagittal deformity.
The NuVasive CoRoent XL Interfixated System implants are indicated for intervertebral body fusion of the spine in skeletally mature patients. The System is designed for use with autogenous and/or allogeneic bone graft comprised of cancellous and/or corticocancellous bone graft to facilitate fusion and supplemental internal spinal fixation system (e.g., pedicle screw system) cleared by the FDA for use in the lumbar spine in addition to the integrated screws. The devices are to be used in patients who have had at least six months of non-operative treatment.
The NuVasive CoRoent XL Interfixated System implants are interbody fusions in the lumbar spine, from L2 to L5, following discectomy in the treatment of symptomatic degenerative disc disease (DDD) or degenerative spondylolisthesis at one or two adjacent levels. DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. The Nu Vasive CoRoent XL Interfixated System implants can also be used as an adjunct to fusion in patients diagnosed with multilevel degenerative scoliosis and sagittal deformity.
The Brigade System is indicated for spinal fusion procedures in skeletally mature patients. The Brigade Standalone System (lordotic angles of 8° and 12°) is a standalone system. The Brigade Hyperlorio System (lordotic angles of 15° to 30°) must be used with supplemental internal spinal fixation systems (e.g., posterior pedicle screw and rod system) cleared by the FDA for use in the lumbar spine in addition to the integrated screws. The System is designed for use with autogenous and/or allogeneic bone graft to facilitate fusion. The devices are to be used in patients who have had at least six months of non-operative treatment.
The Brigade System is intended for use in interbody fusions in the lumbar spine from L2 to S1, following discectorny in the treatment of symptomatic degenerative disc disease (DDD) or degenerative spondylolisthesis at one or two adjacent levels. DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. The Brigade System platform implants can also be used as an adjunct to fusion in patients diagnosed with multilevel degenerative scoliosis and sagittal deformity. However, when used in these patients at multiple levels and for patients with degenerative spondylolisthesis, the Brigade System platform must be used with a supplemental internal spinal fixation system (e.g., pedicle screw system) cleared by the FDA for use in the lumbar spine in addition to the integrated screws.
The Brigade Lateral System is indicated for spinal fusion procedures in skeletally mature patients. The Brigade Lateral System must be used with supplemental internal spinal fixation systems (e.g., posterior pedicle screw and rod system) that are cleared by the FDA for use in the lumbar spine in addition to the integrated screws. The System is designed for use with autogenous and/or allogenic bone graft comprised of cancellous and/or corticocancellous bone graft to facilitate fusion. The devices are to be used in patients who have had at least six months of non-operative treatment. The Brigade Lateral System is intended for use in interbody fusions in the lumbar spine from L2 to S1, following discectomy in the treatment of symptomatic degenerative disc disease (DDD) or degenerative spondylolisthesis at one or two adjacent levels. DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. The Brigade Lateral System implants can adjunct to fusion in patients diagnosed with multilevel degenerative scoliosis and sagittal deformity.
The BASE Interfixated Titanium System is indicated for spinal fusion procedures in skeletally mature patients. The BASE Interfixated Titanium System 10° - 20° lordotic cages may be used as a standalone system. The BASE Interfixated Titanium System 25° - 30° lordotic cages must be used with supplemental internal spinal fixation systems (i.e., posterior pedicle screw and rod system) that are cleared by the FDA for use in the System is designed for use with autogenous and/or allogeneic bone graft comprised of cancellous and/or corticocancellous bone graft to facilitate fusion. The devices are to be used in patients who have had at least six months of non-operative treatment. The BASE Interfixated Titanium System is interbody fusions in the lumbar spine from L2 to S1, following discectorny in the treatment of symptomatic degenerative disc disease (DDD) or degenerative spondylolisthesis at one or two adjacent levels. DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. The BASE Interfixated System implants can also be used as an adjunct to fusion in patients diagnosed with multilevel degenerative scoliosis and sagittal deformity; however, when used in these patients at multiple levels and for patients with degenerative spondylolisthesis, the BASE Interfixated Titanium System must be used with a supplemental internal spinal fixation system (e.g., pedicle screw system) cleared by the FDA for use in the lumbar spine in addition to the integrated screws.
The Coalesce Thoracolumbar Interbody Fusion System is indicated for intervertebral body fusion of the spine in skeletally mature patients. The System is designed for use with autogenous and/or allogeneic bone graft comprised of cancellous and/or corticocancellous bone graft to facilitate fusion and supplemental internal spinal fixation systems cleared by the FDA for use in the thoracolumbar spine. The devices are to be used in patients who have had at least six months of nonoperative treatment.
The Coalesce Thoracolumbar Interbody Fusion System is interbody fusions in the thoracic spine from T1 to T12 and at the thoracolumbar junction (T12-L1), and is intended for use in the lumbar spine, from L1 to S1, for the treatment of symptomatic disc degeneration (DDD) or degenerative spondylolisthesis at one or two adjacent levels, including thoracic disc herniation (with myelopathy and/or radiculopathy with or without axial pain). DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. The Coalesce Thoracolumbar Interbody Fusion System can be used as an adjunct to fusion in patients diagnosed with multilevel degenerative scoliosis and sagittal deformity.
The Cohere Thoracolumbar Interbody System is indicated for intervertebral body fusion of the spine in skeletally mature patients. The System is designed for use with autogenous and/or allogeneic bone graft comprised of cancellous and/or corticocancellous bone graft to facilitate fusion and supplemental internal spinal fixation systems cleared by the FDA for use in the thoracolumbar spine. The devices are to be used in patients who have had at least six months of non-operative treatment.
The Cohere Thoracolumbar Interbody System is interbody fusions in the thoracic spine, from T1 to T12, and at the thoracolumbar junction (T12-L1), and in the lumbar spine, from L1 to S1, for the treatment of symptomatic disc degeneration (DDD) or degenerative spondylolisthesis at one or two adjacent levels, including thoracic disc herniation (with myelopathy and/or radiculopathy with or without axial pain). DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. The Cohere Thoracolumbar Interbody System can be used as an adjunct to fusion in patients diagnosed with multilevel degenerative scoliosis and sagittal deformity.
The NuVasive Modulus XLIF Interbody System is indicated for intervertebral body fusion of the spine in skeletally mature patients. The System is designed for use with autogenous and/or allogeneic bone graft comprised of cancellous and/or corticocancellous bone graft to facilitate fusion. When used with or without Modulus XLIF internal fixation, the system is intended for use with supplemental spinal fixation system cleared by the FDA for use in the thoracolumbar spine. The devices are to be used in patients who have had at least six months of non-operative treatment. The NuVasive Modulus XLIF Interbody System is intended for use in interbody fusions in the thoracolumbar spine from T1 to T12 and at the thoracolumbar junction (T12-L1), and for use in the lumbar spine from L1 to S1, for the treatment of symptomatic disc degeneration (DDD) or degenerative spondylolisthesis at one or two adjacent levels, including thoracic disc herniation (with myelopathy and/or radiculopathy with or without axial pain). DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. The NuVasive Modulus XLIF Interbody System can also be used as an adjunct to fusion in patients diagnosed with multilevel degenerative scoliosis and sagittal deformity.
The NuVasive Modulus TLIF Interbody System is indicated for intervertebral body fusion of the spine in skeletally mature patients. The System is designed for use with autogenous and/or allogeneic bone graft comprised of cancellous and/or corticocancellous bone graft to facilitate fusion and supplemental internal spinal fixation systems cleared by the FDA for use in the thoracolumbar spine. The devices are to be used in patients who have had at least six months of nonoperative treatment.
The NuVasive Modulus TLIF Interbody System is intended for use in interbody fusions in the thoracolumbar spine from T1 to T12 and at the thoracolumbar junction (T12-L1), and for use in the lumbar spine from L1 to S1, for the treatment of symptomatic disc degeneration (DDD) or degenerative spondylolisthesis at one or two adjacent levels, including thoracic disc herniation (with myelopathy and/or radiculopathy with or without axial pain). DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. The NuVasive Modulus TLIF Interbody System can also be used as an adjunct to fusion in patients diagnosed with multilevel degenerative scoliosis and sagittal deformity.
The Nu Vasive 3DP Interfixated ALIF System is indicated for spinal fusion procedures in skeletally mature patients. The 3DP Interfixated ALIF System 10°-20° lordotic cages may be used as a standalone system. The 3DP Interfixated ALIF System 25°-30° lordotic cages must be used with supplemental internal spinal fixation systems (e.g., posterior pedicle screw and rod system) that are cleared by the FDA for use in the lumbar spine. The 3DP Interfixated ALIF System used with less than 3 bolts must be used with supplemental internal spinal fixation systems (e.g., posterior pedicle screw and rod system) that are cleared by the FDA for use in the lumbar spine. The System is designed for use with autogenous and/or allogeneic bone graft comprised of cancellous and/or corticocancellous bone graft to facilitate fusion. The devices are to be used in patients who have had at least six months of non-operative treatment.
The Nu Vasive 3DP Interfixated ALIF System is intended for use in interbody fusions in the lumbar spine from L2 to S1, following discectorny in the treatment of symptomatic degenerative disc disease (DDD) or degenerative spondylolisthesis at one or two adjacent levels. DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. The 3DP Interfixated ALIF System implants can also be used as an adjunct to fusion in patients diagnosed with multilevel degenerative scoliosis and sagittal deformity; however, when used in these patients at multiple levels and for patients with degenerative spondylolisthesis, the 3DP Interfixated ALIF System must be used with a supplemental internal spinal fixation system (e.g., pedicle screw system) cleared by the FDA for use in the lumbar spine in addition to the integrated screws.
The NuVasive Thoracolumbar Interbody Systems are intervertebral body fusion devices, interfixated and non-interfixated. Implants are manufactured of either PEEK-Optima® LT-1 (Polyether-ether-ketone) conforming to ASTM F2026 or Titanium alloy (Ti6Al4V ELI) conforming to ASTM F136/ISO 5832-3. PEEK implants include radiographic markers made of Titanium (Ti) conforming to ASTM F136/ISO 5832-3 or ASTM F1472, or Tantalum (Ta) conforming to ASTM F560 or ISO 13782. Interfixated implants include Titanium alloy (Ti6Al4V ELI) screws conforming to ASTM F136/ISO 5832-3, and in addition, CoRoent Thoracolumbar Interfixated implants include canted coil locking mechanism of Nickel-Cobalt-Chromium-Molybdenum alloy (MP35N) conforming to ASTM F562.
The provided text describes a 510(k) premarket notification for various intervertebral body fusion devices. This type of submission focuses on demonstrating substantial equivalence to legally marketed predicate devices, rather than proving a device meets specific acceptance criteria through a clinical study with detailed performance metrics. Therefore, much of the requested information regarding acceptance criteria, study design, expert involvement, and ground truth establishment is not typically part of a 510(k) summary for these types of Class II spinal implants.
However, I can extract and infer some information based on the document's content.
Here's a breakdown of the requested information:
1. A table of acceptance criteria and the reported device performance:
The document does not provide a table of explicit acceptance criteria or quantitative performance metrics in the way one would see for an AI/CADe device or a new technology requiring de novo clearance. For spinal implants cleared via 510(k), "acceptance criteria" are generally tied to demonstrating substantial equivalence in terms of:
- Intended Use: The device performs the same function for the same condition.
- Technological Characteristics: The materials, design, and operating principles are similar or equivalent, and any differences do not raise new questions of safety or effectiveness.
- Performance: The device performs as safely and effectively as the predicate, often demonstrated through non-clinical (e.g., mechanical, biocompatibility) testing.
The "performance" reported here is a qualitative assessment of substantial equivalence.
Acceptance Criteria (Implied for 510(k) Equivalence) | Reported Device Performance |
---|---|
Intended Use Equivalence: Device addresses the same conditions and patient population as predicates. | "The NuVasive® Thoracolumbar Interbody Systems are indicated for intervertebral body fusion of the spine in skeletally mature patients... for the treatment of symptomatic disc degeneration (DDD) or degenerative spondylolisthesis at one or two adjacent levels, including thoracic disc herniation... can be used as an adjunct to fusion in patients diagnosed with multilevel degenerative scoliosis and sagittal deformity." (Multiple specific indications for each system, mirroring predicate uses). |
Technological Characteristics Equivalence: Materials, design, and function are substantially similar to predicates, or differences don't raise new safety/effectiveness concerns. | "The subject device was shown to be substantially equivalent and have the same technological characteristics to its predicate devices through comparison in areas including design, intended use, material composition, function, and range of sizes." |
Implants are manufactured from PEEK-Optima® LT-1 or Titanium alloy (Ti6Al4V ELI). Interfixated implants include Titanium alloy screws. (Similar to predicate materials). |
| Safety and Effectiveness Equivalence: Demonstrated through established clinical use and/or non-clinical testing. | "Based on the clinical data, it was determined that the NuVasive Thoracolumbar Interbody Systems used in the treatment of thoracolumbar sagittal deformities has a safety and effectiveness profile similar to the predicate device."
"No new NuVasive Thoracolumbar Interbody Systems implant designs are being introduced... minor design modifications were made... We include the confirmatory mechanical testing that shows that the minor design modifications do not create a new worst case that would require new or additional testing."
"Therefore, no performance testing was performed for this 510(k) submission and the worst case devices included with the subject system were tested and cleared in predicate 510(k) submissions." |
2. Sample size used for the test set and the data provenance:
- Sample Size for Test Set: Not applicable in the context of this 510(k) submission. This submission primarily relies on demonstrating equivalence to existing cleared devices, particularly for a broadened indication (multilevel degenerative scoliosis and sagittal deformity). The document states, "A clinical literature analysis of thoracolumbar sagittal deformities treated with the subject device was performed." This refers to a literature review, not a new clinical study with a prospectively defined test set of patients for device performance evaluation.
- Data Provenance: The nature of the "clinical literature analysis" means the data provenance would be from published studies, which could originate from various countries and be retrospective or prospective depending on the individual studies included in the analysis. The document does not specify.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
- Not applicable. As no new clinical study was conducted with a defined "test set" requiring ground truth establishment by experts for specific outcomes, this information is not provided. The determination of "substantial equivalence" is made by the FDA based on the provided technical and comparative information and literature review.
4. Adjudication method for the test set:
- Not applicable. There was no specific test set requiring adjudication in the context of this 510(k) submission.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
- Not applicable. This is a submission for spinal implants, not an AI or imaging diagnostic device.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:
- Not applicable. This is a physical implant device, not an algorithm.
7. The type of ground truth used:
- For this 510(k) submission, the "ground truth" for the new indication (multilevel degenerative scoliosis and sagittal deformity) is established through clinical literature analysis. This means the safety and effectiveness profile of using these types of devices for this condition, as reported in existing peer-reviewed literature, serves as the basis for enabling expanded indications. The FDA relies on the predicate devices' established safety and effectiveness and the literature supporting the similar safety and effectiveness for the expanded use.
8. The sample size for the training set:
- Not applicable. No "training set" in the machine learning sense was used, as this is a physical medical device. The "training" for the device's design and use comes from existing medical knowledge, biomechanical principles, and the history of predicate devices.
9. How the ground truth for the training set was established:
- Not applicable. As above, no machine learning training set was used. The basis for the device's design and function comes from established clinical practice, biomechanical data from prior predicate devices, and manufacturing standards for medical implants.
Ask a specific question about this device
(97 days)
The ALIF System is indicated for use in skeletally mature patients with Degenerative Disc Disease (DDD) at one or two contiguous levels from L2 to S1.
DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by patient history and radiographic studies. These patients should be sketally mature and have had six months of non-operative treatment prior to treatment with the devices. These DDD patients may also have up to Grade I spondylolisthesis or retrolisthesis at the involved level(s).
The ALIF System must be used with supplemental internal spinal fixation systems (i.e. posterior pedicle screw and rod system) that are cleared by the FDA for use in the lumbar spine.
For use with autogenous and/or allogeneic bone graft comprised of cancellous and/or corticocancellous bone graft to facilitate fusion.
The Axis Spine Technologies ALIF is an inter-fixated interbody system consisting of a modular interbody spacer manufactured from Ti6Al4V. The system is designed to be assembled in vivo. The spacer comprises of two endplates and a central core. Each component is available in a variety of shapes and sizes to allow the assembled device to suit the individual pathology and anatomical conditions of the patient.
The Axis Spine Technologies ALIF System intervertebral fusion device is designed to address lumbar pathologies utilizing placement through an anterior approach. The device's hollow core or graft aperture allows for packing of autogenous and/or allogeneic bone graft to help promote a solid fusion. A rough surface on the device endplates serves to grip the adjacent vertebrae to resist migration and expulsion of the device. The subject device is indicated for use with supplemental internal spinal fixation (i.e., posterior pedicle screw and rodsystem).
The Axis Spine Technologies ALIF implant is composed of:
- one (1) inferior endplate manufactured from titanium alloy (Ti-6AI-4V ELI) conforming to ASTM F136 and ISO 5832-3.
- one (1) inferior endplate manufactured from titanium alloy (Ti-6AI-4V ELI) ● conforming to ASTM F136 and ISO 5832-3.
- one (1) core manufactured from titanium alloy (Ti-6Al-4V ELI) conforming to . ASTM F136 and ISO 5832-3.
- three (3) bone screws manufactured from titanium alloy (Ti-6Al-4V ELI) ● conforming to ASTM F136 and ISO 5832-3.
- one (1) cover plate manufactured from PEEK-Optima HA Enhanced. ●
This document is a 510(k) summary for the Axis Spine Technologies ALIF system, an intervertebral body fusion device. It details the device's characteristics, intended use, and comparative testing against predicate devices to demonstrate substantial equivalence.
Here's the breakdown of the acceptance criteria and the study that proves the device meets the acceptance criteria, based on the provided PDF:
Acceptance Criteria and Reported Device Performance
The concept of "acceptance criteria" for performance metrics like sensitivity, specificity, accuracy, etc., and "reported device performance" are typical for diagnostic or AI-driven devices where a quantitative measure of performance against a ground truth is established.
However, this document describes a spinal implant (intervertebral body fusion device). For such devices, acceptance criteria primarily revolve around mechanical performance, biocompatibility, and substantial equivalence to existing devices, rather than diagnostic accuracy. The "study" proving acceptance criteria is primarily non-clinical (bench) testing and comparison to predicate devices.
Therefore, the table format for typical medical device AI/diagnostic performance metrics (sensitivity, specificity, etc.) is not directly applicable here. Instead, the acceptance criteria are based on biocompatibility and mechanical performance tests to ensure the device is safe and effective and performs comparably to or better than predicate devices.
Table of Acceptance Criteria (for a spinal implant) and Reported Device Performance:
Acceptance Criterion Type | Specific Test/Standard | Acceptance Goal (Implicit) | Reported Device Performance (Summary) |
---|---|---|---|
Mechanical Performance | Static Compression per ASTM F2077 | Meets or exceeds predicate device performance; within design specifications | The Axis Spine Technologies ALIF meets or exceeds the performance of the predicate device. |
Dynamic Compression per ASTM F2077 | Meets or exceeds predicate device performance; within design specifications | The Axis Spine Technologies ALIF meets or exceeds the performance of the predicate device. | |
Static Compression Shear per ASTM F2077 | Meets or exceeds predicate device performance; within design specifications | The Axis Spine Technologies ALIF meets or exceeds the performance of the predicate device. | |
Dynamic Compression Shear per ASTM F2077 | Meets or exceeds predicate device performance; within design specifications | The Axis Spine Technologies ALIF meets or exceeds the performance of the predicate device. | |
Subsidence per ASTM F2267 | Resistance to subsidence comparable to or better than predicate device | The Axis Spine Technologies ALIF meets or exceeds the performance of the predicate device. | |
Expulsion and Axial Pushout of Retention Device Testing | Secure fixation and resistance to expulsion | The Axis Spine Technologies ALIF meets or exceeds the performance of the predicate device. | |
Material/Biocompatibility | Ti6Al4V ELI conforming to ASTM F136 and ISO 5832-3; PEEK-Optima HA Enhanced | Biocompatible and safe for implantation; materials meet established standards | Materials used conform to specified ASTM and ISO standards for implantable devices. |
The study that proves the device meets the acceptance criteria:
The study proving the device meets the acceptance criteria is primarily a non-clinical (bench-top) performance testing program and a comparison against legally marketed predicate devices.
Study Details:
-
Sample sizes used for the test set and the data provenance:
- Test Set (Non-Clinical): For mechanical testing, the "sample size" refers to the number of physical devices or components tested. While not explicitly stated with a specific number (e.g., n=X devices), the tests were conducted per relevant ASTM standards (ASTM F2077, ASTM F2267). These standards specify the number of samples required for each test (e.g., 6 samples for static compression, 3 for dynamic, etc.).
- Data Provenance: The device is manufactured by Axis Spine Technologies Ltd, located in the United Kingdom. The testing would have been conducted in a laboratory setting, typically in the country of manufacture or a qualified testing facility. The data is retrospective in the sense that it's generated from manufactured devices for the purpose of regulatory submission, not from a patient population study.
-
Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
- For an implantable medical device undergoing non-clinical mechanical testing, "ground truth" is established by engineering standards and specifications (e.g., ASTM standards) and the objective physical measurements of the devices' performance.
- There isn't a concept of a "human expert" establishing ground truth in the way a radiologist establishes ground truth for an AI diagnostic algorithm. The "experts" involved would be engineers and technicians conducting the tests in accordance with validated protocols and interpreting the results against the specified standards and predicate device performance. Their qualifications would be in biomedical engineering, mechanical engineering, or materials science.
-
Adjudication method for the test set:
- Not applicable in the context of non-clinical mechanical testing. Adjudication is relevant for subjective assessments (like image interpretation) where multiple readers' opinions need to be reconciled. Here, the results are quantitative physical measurements.
-
If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
- No. An MRMC study is a clinical study typically used to evaluate the performance of diagnostic devices or AI algorithms when used by human readers. This submission is for an implantable spinal fusion device.
-
If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:
- No. This is not an AI/algorithm-driven device. It is a physical implant.
-
The type of ground truth used:
- For this device, the "ground truth" is defined by established engineering performance standards (ASTM) and the demonstrated performance characteristics of the legally marketed predicate devices. The goal is to show comparable or superior mechanical performance and material properties.
-
The sample size for the training set:
- Not applicable. This device is a physical implant, not an AI algorithm that requires a training set.
-
How the ground truth for the training set was established:
- Not applicable. See point 7.
Ask a specific question about this device
(223 days)
The EL CAPITAN Anterior Lumbar Interbody System is indicated for intervertebral body fusion procedures in skeletally mature patients with degenerative disc disease (DDD) of the lumbar spine at one or two contiguous levels from L1-L2 to L5-S1. DDD is defined as discogenic pain with degeneration of the disc confirmed by history and radiographic studies. These DDD patients may also have up to Grade I spondylolisthesis or retrolisthesis at the involved level(s). EL CAPITAN System implants are to be used with autogenous bone graft and supplemental fixation. Patients should have at least six (6) months of non-operative treatment prior to treatment with an intervertebral cage.
The EL CAPITAN Spacer and Plate assembly are an integrated interbody fusion device intended for stand-alone use when used with all titanium alloy screws. When used with anchors, the assembly is intended for use with additional supplemental fixation that has been cleared by the FDA for use in the lumbar spine.
Hyperlordotic interbody devices (>20° lordosis) must be used with supplemental fixation (e.g. posterior fixation) that has been cleared by the FDA for use in the Lumbar spine.
The EL CAPITAN Anterior Lumbar Interbody devices are implants developed for the substitution of the classical autogenous bone graft blocks. The cages assist to avoid complications related to the bone graft donation site (chronic pain, hematoma, infection, bone removal from the donor site making it impossible to remove bone again, quality of the iliac bone, accessing a healthy donor site that may become an unhealthy site, hernias by the incision). The Spacers are a 2-piece modular design which allows for interchangeable plate and spacer components. The plate and spacer components contain interlocking features in addition to a spring-loaded latch mechanism which allows for intraoperative assembly prior to implantation. The spacer components are available in a range of footprints and heights, and the plates are offered in multiple fixation types and sizes to suit the individual pathology and anatomical conditions of the patient. The implants have a hollow center to allow placement of autogenous bone graft. The superior and inferior surfaces are open to promote contact of the bone graft with the vertebral end plates, allowing bone growth (arthrodesis)
This document is a 510(k) summary for the EL CAPITAN Anterior Lumbar Interbody System, a medical device for spinal fusion. This device is not an AI/ML-enabled device, and therefore, the requested information about acceptance criteria and studies demonstrating AI performance (such as sample sizes for test/training sets, expert ground truth establishment, MRMC studies, etc.) is not applicable.
The document explicitly states: "No clinical studies were performed." This indicates that the device's substantial equivalence to predicate devices was based on non-clinical tests (mechanical and material properties), rather than clinical performance data. Therefore, there are no acceptance criteria related to AI/ML performance metrics, nor any studies proving such performance.
The non-clinical tests conducted establish the device's mechanical safety and performance, demonstrating equivalence to predicate devices. The acceptance criteria for these non-clinical tests would be the established ASTM standards mentioned (e.g., ASTM F2077, ASTM F1717, ASTM F2267).
To summarize why the specific points of your request cannot be answered from this document:
- Acceptance Criteria & Reported Device Performance (Table): Not applicable for AI/ML performance. The device's performance is shown through compliance with mechanical testing standards.
- Sample sizes & Data Provenance (Test/Training Sets): Not applicable as there is no AI/ML component. Mechanical tests have their own sample size requirements per ASTM standards.
- Number of experts & Qualifications: Not applicable as there is no AI/ML component and no clinical studies were performed requiring expert reading.
- Adjudication Method: Not applicable.
- MRMC Comparative Effectiveness Study: Not applicable.
- Standalone Performance: Not applicable.
- Type of ground truth: Not applicable for AI/ML. The "ground truth" for this device's safety and efficacy is derived from its mechanical properties, material biocompatibility, and demonstrated equivalence to previously cleared predicate devices through non-clinical testing.
- Sample size for training set & How ground truth established: Not applicable.
The document focuses on demonstrating substantial equivalence to legally marketed predicate devices based on:
- Intended Use
- Design
- Materials Used (PEEK-OPTIMA LT120HA, Tantalum, Titanium Alloy, Nitinol)
- Mechanical Safety and Performance (Non-clinical tests: Static/Dynamic Compression, Static/Dynamic Compression Shear, Subsidence, Expulsion per ASTM standards).
Therefore, this document does not contain the information required to answer your prompt, as it pertains to a traditional medical device, not an AI/ML-enabled one.
Ask a specific question about this device
(92 days)
The Sovereign™ Spinal System is indicated for use with autogenous bone and/or allograft bone graft comprised of cancellous and/or corticocancellous bone graft, and/or demineralized allograft bone marrow aspirate in patients with degenerative disc disease (DDD) at one or two contiguous levels from L2 to S1. DDD is defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies. Additionally, the Sovereign™ Spinal System is indicated for use in patients diagnosed with multilevel degenerative scoliosis and sagittal deformity conditions as an adjunct to fusion. These patients should be skeletally mature and have had 6 months of non-operative treatment. When used in patients as an adjunct to fusion in patients diagnosed with multilevel degenerative scoliosis and sagittal deformity conditions, additional supplemental fixation) must be used. These implants may be implanted via a variety of open or minimally invasive approaches include anterior and oblique.
The Sovereign™ Spinal System may be used as a stand-alone device or in conjunction with supplemental fixation. When used as a stand-alone device, the Sovereign™ Spinal System is intended to be used with 3 titanium alloy fixed or variable angle screws. The accompanying cover plate MUST be used anytime the device is used with any number of variable angle screws. If the physician chooses to use less than 3 or none of the provided screws, additional supplemental fixation in the lumbar spine must be used to augment stability. Implants with lordosis angles greater than 18° are intended to be used with supplemental fixation (e.g. facet screws or posterior fixation).
The SOVEREIGN™ Spinal System is an intervertebral body fusion device with internal screw fixation. The subject devices in this submission add additional polyetheretherketone (PEEK) implant lordosis options with corresponding trials and PTC implant options with a Commercially Pure Titanium (CP Ti) coating, to the existing SOVEREIGN™ Spinal System.
The provided text describes a 510(k) summary for the Medtronic SOVEREIGN™ Spinal System, which is an intervertebral body fusion device. The submission focuses on adding new polyetheretherketone (PEEK) implant lordosis options and PTC implant options with a Commercially Pure Titanium (CP Ti) coating to the existing system. The document primarily discusses substantial equivalence to predicate devices based on mechanical testing and MRI compatibility. It does not describe an AI medical device or a study with clinical endpoints involving human readers or ground truth established by experts/pathology.
Therefore, many of the requested details about acceptance criteria, study design, expert involvement, and AI performance metrics are not applicable to the information provided in the given text.
Based on the available information:
1. A table of acceptance criteria and the reported device performance
The acceptance criteria for the SOVEREIGN™ Spinal System are based on mechanical testing standards. The reported device performance indicates that the subject implants met these criteria.
Acceptance Criteria (Standards) | Reported Device Performance |
---|---|
Mechanical Testing: | |
ASTM F2077, Test Methods For Intervertebral Body Fusion Devices | Subject implants met pre-determined acceptance criteria. |
ASTM F2267, Standard Test Method for Measuring Load Induced Subsidence of Intervertebral Body Fusion Device Under Static Axial Compression | Subject implants met pre-determined acceptance criteria. |
ASTM Draft Standard F-04.25.02.02, Static Push-out Test Method for Intervertebral Body Fusion Devices | Subject implants met pre-determined acceptance criteria. |
Specific Tests Performed: | |
Static Compression | Subject implants met pre-determined acceptance criteria. |
Compression Fatigue | Subject implants met pre-determined acceptance criteria. |
Static Compression Shear | Subject implants met pre-determined acceptance criteria. |
Compression Shear Fatigue | Subject implants met pre-determined acceptance criteria. |
Subsidence | Subject implants met pre-determined acceptance criteria. |
Expulsion | Subject implants met pre-determined acceptance criteria. |
MRI Testing: | |
Guidance for Industry and FDA Staff – Establishing Safety and Compatibility of Passive Implants in the Magnetic Resonance (MR) Environment | Subject devices were evaluated and can be classified as MR-Conditional. |
2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)
This information is not applicable as the document describes mechanical and MRI compatibility testing of a physical medical device (intervertebral body fusion device), not a software or AI device with a test set of data.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)
This information is not applicable as the document describes mechanical and MRI compatibility testing of a physical medical device. Ground truth by experts is not established for this type of testing.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
This information is not applicable as the document describes mechanical and MRI compatibility testing of a physical medical device. Adjudication methods are not used for this type of testing.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
This information is not applicable. The document does not describe an AI medical device or any study involving human readers.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
This information is not applicable. The document does not describe an AI medical device.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)
For the mechanical testing, the "ground truth" is defined by the performance requirements set forth in the referenced ASTM standards. For MRI testing, it's about meeting safety and compatibility guidelines outlined by the FDA for the MR environment. These are objective engineering and regulatory standards, not clinical ground truth like pathology or expert consensus.
8. The sample size for the training set
This information is not applicable as the document describes mechanical and MRI compatibility testing of a physical medical device, not an AI model requiring a training set.
9. How the ground truth for the training set was established
This information is not applicable as the document describes mechanical and MRI compatibility testing of a physical medical device, not an AI model requiring a training set.
Ask a specific question about this device
(90 days)
The IRIX-A™ Lumbar Integrated Fusion System is a stand-alone intervertebral body fusion device intended for use in patients with degenerative disc disease (DDD) at one (1) or two (2) contiguous levels of the lumbosacral spine (L2-S1 inclusive). DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. These patients should be skeletally mature and have had at least six (6) months of non-operative treatment. In addition, these patients may have had a previous non-fusion spinal surgery at the involved level(s) and may have had up to a Grade I spondylolisthesis or retrolisthesis at the involved level(s). The IRIX-A™ System is intended to be used with autograft and/or allograft comprised of cancellous bone graft material, and is to be used with three titanium alloy screws included as part of the system.
Hyperlordotic implants >20° are intended for use with supplemental fixation (e.g. facet screws or posterior fixation).
The IRIX-A™ Lumbar Integrated Fusion System is a stand-alone intervertebral fusion device to restore biomechanical height and act as an aid in fusion of the lumbar spine in anterior discectomy procedures. The device is generally boxed shaped with teeth on the superior and inferior faces of the device. The IRIX-A™ implant is manufactured from both titanium alloy (Ti6AI4V) in accordance with ASTM F136 and Invibio PEEK Optima LT1 in accordance with ASTM F2026, or from Ti6Al4V titanium alloy alone. The device will be supplied with the option of having the superior and inferior surfaces of the device plasma coated with medical-grade commercially pure titanium (CP Ti) per ASTM F1580.
The device is secured in location through the use of bone screws, also manufactured from titanium alloy (Ti6Al4V) per ASTM F136. The devices are provided in various sizes, and screws are offered in multiple lengths to adjust for variations in patient anatomy. The spacers of the IRIX-A™ System are supplied clean and STERILE. The screws and accompanying instruments of the IRIX™-A System are provided clean and non-sterile.
The system does not contain software/firmware or electrical equipment.
The provided document is a 510(k) Pre-market Notification from the FDA for a medical device called the IRIX-A™ Lumbar Integrated Fusion System. This submission focuses on establishing substantial equivalence to existing predicate devices, rather than a clinical study evaluating the device's diagnostic performance against specific acceptance criteria for AI/software-based devices. Therefore, much of the requested information regarding acceptance criteria for device performance in a clinical context, sample sizes for test and training sets, expert qualifications, and ground truth establishment, is not applicable or not present in this type of regulatory document.
However, I can extract information related to the performance data provided within this 510(k) summary for non-clinical testing.
Here's a breakdown of the available information:
1. Table of Acceptance Criteria and Reported Device Performance:
Acceptance Criteria (from testing standards) | Reported Device Performance |
---|---|
ASTM F2077: Static and dynamic compression testing (specific values not provided in this document) | Meets or exceeds the performance of the predicate device. |
ASTM F2267: Subsidence (specific values not provided in this document) | Meets or exceeds the performance of the predicate device. |
Expulsion as recommended by FDA (specific methodology/values not provided in this document) | Meets or exceeds the performance of the predicate device. |
Note: The document states the device "meets or exceeds the performance of the predicate device" for these tests, implying that the acceptance criteria are based on the performance of the predicate device as established through the same ASTM standards and FDA recommendations.
2. Sample size used for the test set and the data provenance:
- Test set sample size: Not applicable. This document describes non-clinical mechanical testing, not a clinical study involving a test set of data.
- Data provenance: Not applicable. The testing is described as "Nonclinical testing" and performed on the device itself.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
- Not applicable. Ground truth for clinical data is not relevant to this non-clinical mechanical testing.
4. Adjudication method for the test set:
- Not applicable.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
- Not applicable. This is not an AI/software-based diagnostic device, nor a comparative effectiveness study involving human readers.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:
- Not applicable. This is not an algorithm-only device.
7. The type of ground truth used:
- For the non-clinical performance data: The "ground truth" or reference for performance is the established performance characteristics of the predicate device under the specified ASTM standards and FDA recommendations for static and dynamic compression, subsidence, and expulsion.
8. The sample size for the training set:
- Not applicable. This product does not involve machine learning or a "training set."
9. How the ground truth for the training set was established:
- Not applicable.
Summary based on the provided document:
The IRIX-A™ Lumbar Integrated Fusion System is a physical medical device (intervertebral body fusion device), not an AI/software-based diagnostic tool. The regulatory submission (510(k)) aims to demonstrate substantial equivalence to legally marketed predicate devices. This is achieved through non-clinical performance testing (mechanical/physical tests) rather than clinical studies involving patient data and diagnostic performance metrics like sensitivity, specificity, or reader studies.
The performance data supplied focuses on:
- ASTM F2077: Static and dynamic compression testing
- ASTM F2267: Subsidence
- Expulsion testing as recommended by FDA
The conclusion is that the subject device "meets or exceeds the performance of the predicate device" for these tests, indicating that its mechanical integrity and function are comparable or superior, and it "does not introduce any new risks." The predicate device itself acts as the benchmark for establishing acceptable performance under these tests.
Ask a specific question about this device
Page 1 of 1