(146 days)
No
The summary describes a physical spinal implant made of PEEK with a titanium coating. There is no mention of software, algorithms, or any computational processing that would suggest the use of AI or ML. The testing described is focused on the mechanical and material properties of the implant.
Yes
The device is an implantable system intended for interbody fusion with bone graft in patients with Degenerative Disc Disease, indicating a direct therapeutic purpose to treat a medical condition.
No.
The device is an implantable medical device (spinal system) used for interbody fusion, not for diagnosing conditions. Its intended use is therapeutic, involving surgical implantation to provide support during lumbar interbody fusion surgeries and to assist in spinal deformity correction.
No
The device description clearly states the system consists of PEEK cages and implants, which are physical hardware components. The performance studies also detail testing of these physical implants and their coatings.
Based on the provided text, this device is not an IVD (In Vitro Diagnostic).
Here's why:
- Intended Use: The intended use clearly describes a surgical implant for spinal fusion, specifically for treating Degenerative Disc Disease and assisting in spinal deformity correction. This is a therapeutic intervention performed directly on the patient's body.
- Device Description: The device is described as PEEK cages with a titanium coating, designed to be inserted between vertebral bodies. This is a physical implant.
- Lack of IVD Characteristics: An IVD is a medical device used to perform tests on samples taken from the human body (like blood, urine, tissue) to provide information for diagnosis, monitoring, or screening. The provided text does not mention any testing of samples or diagnostic purposes.
The device is a surgical implant used for structural support and facilitating bone fusion within the spine.
N/A
Intended Use / Indications for Use
CAPSTONE PTC™ Spinal System:
The CAPSTONE PTC™ Spinal System is indicated for interbody fusion with autogenous bone graft in patients with Degenerative Disc Disease (DDD) at one or two levels from L2 to S1. These DDD patients may also have up to Grade 1 Spondylolisthesis or retrolisthesis at the involved levels. DDD is defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies. These patients should be skeletally mature and have had six months of non-operative treatment.
Additionally, the CAPSTONE PTC™ Spinal System is indicated to assist in the setting of spinal deformity as a supplement to pedicle screw fixation in patients diagnosed with degenerative scoliosis.
These implants may be implanted via an open or a minimally invasive posterior approach. Alternatively, these implants may also be implanted via an anterior and/or transforaminal approach. These implants are to be used with autogenous bone graft. These devices are intended to be used with supplemental fixation instrumentation, which has been cleared by the Food and Drug Administration (FDA) for use in the lumbar spine.
CLYDESDALE PTC™ Spinal System:
CLYDESDALE PTC™ Spinal System is designed to be used with autogenous bone graft to facilitate interbody fusion and is intended for use with supplemental fixation systems cleared for use in the lumbar spine. The CLYDESDALE PTC™ Spinal System is used for patients diagnosed with Degenerative Disc Disease (DDD) at one or two contiguous levels from L2 to S1. These DDD patients may also have up to Grade 1 spondylolisthesis or retrolisthesis at the involved levels. DDD is defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies. These patients should be skeletally mature and have had six months of nonoperative treatment. These implants may be implanted via a minimally invasive lateral approach.
Product codes (comma separated list FDA assigned to the subject device)
MAX
Device Description
The CAPSTONE PTCTM and CLYDESDALE PTC™ Spinal Systems consist of PEEK cages of various widths and heights, which can be inserted between two lumbar or lumbosacral vertebral bodies to give support and correction during lumbar interbody fusion surgeries. The hollow geometry of the implants allows them to be packed with autogenous bone graft.
Mentions image processing
Not Found
Mentions AI, DNN, or ML
Not Found
Input Imaging Modality
Not Found
Anatomical Site
L2 to S1 (lumbar or lumbosacral vertebral bodies)
Indicated Patient Age Range
Skeletally mature
Intended User / Care Setting
Not Found
Description of the training set, sample size, data source, and annotation protocol
Not Found
Description of the test set, sample size, data source, and annotation protocol
Not Found
Summary of Performance Studies (study type, sample size, AUC, MRMC, standalone performance, key results)
Non-Clinical Testing:
- Testing of the plasma-sprayed coating was performed according to FDA's Guidance for FDA Reviewers/Staff, "Guidance for Industry on the Testing of Metallic Plasma Sprayed Coatings on Orthopedic Implants to Support Reconsideration of Postmarket Surveillance Requirements," issued February 2, 2000.
- Coating Microstructure: ASTM F1854, Standard test method for stereological evaluation of porous coatings on medical implants
- Shear Fatigue Testing: ASTM F1160, Standard Test Method for Shear and Bending Fatigue . Testing of Calcium Phosphate and Metallic Medical and Composite Calcium Phosphate/ Metallic Coatings
- Static Shear Testing: ASTM F1044, Standard test method for shear testing of calcium . phosphate coatings and metallic coatings
- Tensile Testing: ASTM F1147, Standard test method for tension testing of calcium . phosphate & metallic coatings
- Abrasion Testing: ASTM F1978. Standard Test Method for Measuring Abrasion Resistance . of Metallic Thermal Spray Coatings by Using the Taber Abraser
- Mechanical testing was conducted in accordance with ASTM F2077: Test Methods for Intervertebral Body Fusion Devices, including static compression, dynamic compression, static compression shear, and dynamic compression shear fatigue testing.
- Subsidence testing was performed in accordance with ASTM F2267: Standard Test Method for Measuring Load Induced Subsidence of the Intervertebral Body Fusion Device under Static Axial Compression.
- Static push-out (expulsion) testing was also performed in accordance with ASTM Draft Standard F04.25.02.02, Static Push-out Test Method for Intervertebral Body Fusion Devices.
- Wear particulate testing was performed in accordance with ASTM F1877, Standard Practice for Characterization of Particles.
- Animal testing was performed using canines and CP Ti coated coupons.
Key Results: Based on the risk analysis, test results, and additional supporting documentation provided in this pre-market notification, Medtronic believes the subject CAPSTONE PTC™ and CLYDESDALE PTC™ Spinal Systems are substantially equivalent to the predicates, the CAPSTONE® and CLYDESDALE® Spinal Systems.
Key Metrics (Sensitivity, Specificity, PPV, NPV, etc.)
Not Found
Predicate Device(s): If the device was cleared using the 510(k) pathway, identify the Predicate Device(s) K/DEN number used to claim substantial equivalence and list them here in a comma separated list exactly as they appear in the text. List the primary predicate first in the list.
K073291, K123027, K1130528, K122037
Reference Device(s): Identify the Reference Device(s) K/DEN number and list them here in a comma separated list exactly as they appear in the text.
Predetermined Change Control Plan (PCCP) - All Relevant Information for the subject device only (e.g. presence / absence, what scope was granted / cleared under the PCCP, any restrictions, etc).
Not Found
§ 888.3080 Intervertebral body fusion device.
(a)
Identification. An intervertebral body fusion device is an implanted single or multiple component spinal device made from a variety of materials, including titanium and polymers. The device is inserted into the intervertebral body space of the cervical or lumbosacral spine, and is intended for intervertebral body fusion.(b)
Classification. (1) Class II (special controls) for intervertebral body fusion devices that contain bone grafting material. The special control is the FDA guidance document entitled “Class II Special Controls Guidance Document: Intervertebral Body Fusion Device.” See § 888.1(e) for the availability of this guidance document.(2) Class III (premarket approval) for intervertebral body fusion devices that include any therapeutic biologic (e.g., bone morphogenic protein). Intervertebral body fusion devices that contain any therapeutic biologic require premarket approval.
(c)
Date premarket approval application (PMA) or notice of product development protocol (PDP) is required. Devices described in paragraph (b)(2) of this section shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.
0
MAR 1 3 2014
510(k) Summary
MEDTRONIC Sofamor Danek CAPSTONE PTCTM and CLYDESDALE PTC™ Spinal Systems March 2014
I. | Company: | Medtronic Sofamor Danek, USA Inc. |
---|---|---|
1800 Pyramid Place | ||
Memphis, Tennessee 38132 | ||
(901) 396-3133 | ||
II. | Contact: | Julie Bassett |
Regulatory Affairs Program Manager | ||
Telephone: (901) 399-3248 | ||
Fax: (901) 346-9738 | ||
III. | Proposed Proprietary Trade Name: | CAPSTONE PTCT™ Spinal System and CLYDESDALE PTCT™ Spinal System |
IV. | Classification Names: | Intervertebral Body Fusion Device (21 CFR 888.3080) |
Class: | II | |
Product Code: | MAX |
V. Description :
The CAPSTONE PTCTM and CLYDESDALE PTC™ Spinal Systems consist of PEEK cages of various widths and heights, which can be inserted between two lumbar or lumbosacral vertebral bodies to give support and correction during lumbar interbody fusion surgeries. The hollow geometry of the implants allows them to be packed with autogenous bone graft.
1
The CAPSTONE PTC™ and CLYDESDALE PTC™ Spinal Systems will be available in all the same sizes as the predicate systems.
VI. Indications for Use:
CAPSTONE PTCTM Spinal System:
The CAPSTONE PTC™ Spinal System is indicated for interbody fusion with autogenous bone graft in patients with Degenerative Disc Disease (DDD) at one or two levels from L2 to S1. These DDD patients may also have up to Grade 1 Spondylolisthesis or retrolisthesis at the involved levels. DDD is defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies. These patients should be skeletally mature and have had six months of non-operative treatment.
Additionally, the CAPSTONE PTC™ Spinal System is indicated to assist in the setting of spinal deformity as a supplement to pedicle screw fixation in patients diagnosed with degenerative scoliosis.
These implants may be implanted via an open or a minimally invasive posterior approach. Alternatively, these implants may also be implanted via an anterior and/or transforaminal approach. These implants are to be used with autogenous bone graft. These devices are intended to be used with supplemental fixation instrumentation, which has been cleared by the Food and Drug Administration (FDA) for use in the lumbar spine.
CLYDESDALE PTC™ Spinal System:
CLYDESDALE PTC™ Spinal System is designed to be used with autogenous bone graft to facilitate interbody fusion and is intended for use with supplemental fixation systems cleared for use in the lumbar spine. The CLYDESDALE PTC™ Spinal System is used for patients diagnosed with Degenerative Disc Disease (DDD) at one or two contiguous levels from L2 to
2
S1. These DDD patients may also have up to Grade 1 spondylolisthesis or retrolisthesis at the involved levels. DDD is defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies. These patients should be skeletally mature and have had six months of nonoperative treatment. These implants may be implanted via a minimally invasive lateral approach.
Summary of the Technological Characteristics: VII.
The CAPSTONE PTC™ and CLYDESDALE PTC™ Spinal Systems have the same fundamental technology as the predicate CAPSTONE® and CLYDESDALE® Spinal Systems. The CAPSTONE PTC™ and CLYDESDALE PTC™ System implants (subject devices) and the CAPSTONE® and CLYDESDALE® Spinal System implants (predicate devices), are both made from PEEK material with tantalum markers. The only difference in the subject and predicate devices is the subject devices also have a commercially pure titanium coating. In addition, the predicate and subject devices are both convex, bullet-nosed interbody devices designed to contain graft material and facilitate a fusion between two vertebral bodies.
VIII. Identification of Legally Marketed Devices:
The CAPSTONE PTC™ Spinal System has the same indications for use and fundamental scientific technology as the predicate, CAPSTONE® Spinal System (K073291, SE 4/24/2008 and K123027, SE 07/25/2013). The CLYDESDALE PTCTM Spinal System has the same indications for use and fundamental scientific technology as the predicate, CLYDESDALE® Spinal System (K1130528, SE 12/20/2011 and K122037, SE 3/22/2013). An additional predicate, Spinal Element's Lucent Ti-Bond Spinal System (K110632, SE 5/23/2012), is also being used to demonstrate that the commercially pure titanium coating is not new and currently exists on other legally marketed devices.
3
IX. Discussion of Non-Clinical Testing:
- Testing of the plasma-sprayed coating was performed according to FDA's Guidance for FDA Reviewers/Staff, "Guidance for Industry on the Testing of Metallic Plasma Sprayed Coatings on Orthopedic Implants to Support Reconsideration of Postmarket Surveillance Requirements," issued February 2, 2000. The testing was performed according to the applicable ASTM standards listed below:
Coating Microstructure:
- . ASTM F1854, Standard test method for stereological evaluation of porous coatings on medical implants
Shear Fatigue Testing:
- ASTM F1160, Standard Test Method for Shear and Bending Fatigue . Testing of Calcium Phosphate and Metallic Medical and Composite Calcium Phosphate/ Metallic Coatings
Static Shear Testing:
- ASTM F1044, Standard test method for shear testing of calcium . phosphate coatings and metallic coatings
Tensile Testing:
- ASTM F1147, Standard test method for tension testing of calcium . phosphate & metallic coatings
Abrasion Testing:
- ASTM F1978. Standard Test Method for Measuring Abrasion Resistance . of Metallic Thermal Spray Coatings by Using the Taber Abraser
In order to demonstrate substantial equivalence to the predicate devices, mechanical testing was conducted. Static compression, dynamic compression, static compression shear, and dynamic compression shear fatigue testing were performed in accordance with ASTM F2077: Test Methods for Intervertebral
4
Body Fusion Devices. Subsidence testing was performed in accordance with ASTM F2267: Standard Test Method for Measuring Load Induced Subsidence of the Intervertebral Body Fusion Device under Static Axial Compression. Static push-out (expulsion) testing was also performed in accordance with ASTM Draft Standard F04.25.02.02, Static Push-out Test Method for Intervertebral Body Fusion Devices. In addition, wear particulate testing was performed in accordance with ASTM F1877, Standard Practice for Characterization of Particles.
Animal testing was also performed using canines and CP Ti coated coupons.
Conclusion: XI.
Based on the risk analysis, test results, and additional supporting documentation provided in this pre-market notification, Medtronic believes the subject CAPSTONE PTC™ and CLYDESDALE PTC™ Spinal Systems are substantially equivalent to the predicates, the CAPSTONE® and CLYDESDALE® Spinal Systems.
5
DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
March 13, 2014
Medtronic Sofamor Danek USA, Incorporated Ms. Julie Bassett Principal Regulatory Affairs Specialist 1800 Pyramid Place Memphis, Tennessee 38132
Re: K133205
Trade/Device Name: CAPSTONE PTC™ Spinal System and CLYDESDALE PTC™ Spinal System Regulation Number: 21 CFR 888.3080 Regulation Name: Intervertebral body fusion device Regulatory Class: Class II Product Code: MAX Dated: February 17, 2014 Received: February 18, 2014
Dear Ms. Bassett:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
6
Page 2 - Ms. Julie Bassett
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Lori A. Wiqqins
for
Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
7
Form Approved: OMB No. 0910-0120
Expiration Date: January 31, 2017
See PRA Statement below.
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
Indications for Use
510(k) Number (if known) K133205
Device Name
CAPSTONE PTC™ Spinal System
Indications for Use (Describe)
The CAPSTONE PTC™ Spinal System is indicated for interbody fusion with autogenous bone graft in patients with Degenerative Disc Disease (DDD) at one or two levels from L2 to S1. These DDD patients may also have up to Grade 1 Spondylolisthesis or retrolisthesis at the involved levels. DDD is defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies. These patients should be skeletally mature and have had six months of non-operative treatment.
Additionally, the CAPSTONE PTC™ Spinal System is indicated to assist in the setting of spinal deformity as a supplement to pedicle screw fixation in patients diagnosed with degenerative scollosis.
These implants may be implanted via an open or a minimally invasive posterior approach. Alternatively, these implants may also be implanted via an anterior and/or transforaminal approach. These implants are to be used with autogenous bone graft. These devices are intended to be used with supplemental fixation instrumentation, which has been cleared by the Food and Drug Administration (FDA) for use in the lumbar spine.
Type of Use (Select one or both, as applicable)
2 Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
PLEASE DO NOT WRITE BELOW THIS LINE – CONTINUE ON A SEPARATE PAGE IF NEEDED.
. . . . FOR FDA USE ONLY . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . : 评论: .. Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
David Hwang Rh Division of Orthopedie Devices
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
8
Expiration Date: January 31, 2017 See PRA Statement below.
Form Approved: OMB No. 0910-0120
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
Indications for Use
510(k) Number (if known) K133205
Device Name
CLYDESDALE PTC™ Spinal System
Indications for Use (Describe)
CLYDESDALE PTC™ Spinal System is designed to be used with autogenous bone graft to facilitate interbody fusion and are intended for use with supplemental fixation systems cleared for use in the lumbar spine. The CL YDESDALE PTC™ Spinal System is used for patients diagnosed with Degenerative Disc Disease (DDD) at one or two contiguous levels from L2 to S1. These DDD patients may also have up to Grade I spondylolisthesis or retrolisthesis at the involved levels. DDD is defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies. These patients should be skeletally mature and have had six months of non-operative treatment. These implants may be implanted via a minimally invasive lateral approach.
Type of Use (Select one or both, as applicable)
2 Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
FOR FDA USE ONLY ・・・。 ...
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
David Hwang Ph.D.
Division of Orthopedie Devices
This section applies only to requirements of the Paperwork Reduction Act of 1995.
*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW."
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."