AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The NuVasive CoRoent Thoracolumbar System is indicated for intervertebral body fusion of the spine in skeletally mature patients. The System is designed for use with autogenous and/or allogeneic bone graft comprised of cancellous and/or corticocancellous bone graft to facilitate fusion and supplemental internal spinal fixation systems that are cleared by the FDA for use in the thoracolumbar spine. The devices are to be used in patients who have had at least six months of non-operative treatment.

The Nu Vasive CoRoent Thoracolumbar System (XL platform) implants are intended for use in interbody fusions in the thoracic spine, from T1 to T12, and at the thoracolumbar junction (T12-L1), and the CoRoent Thoracolumbar System (XL and L platforms) implants are intended for use in the lumbar spine, from L1 to S1, for the treatment of symptomatic disc degeneration (DDD) or degenerative spondylolisthesis at one or two adjacent levels, including thoracic disc herniation (with myelopathy and/or radiculopathy with or without axial pain). DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. The NuVasive CoRoent Thoracolumbar System (XL and L platforms) can be used as an adjunct to fusion in patients diagnosed with multilevel degenerative scoliosis and sagittal deformity.

The NuVasive CoRoent XL Interfixated System implants are indicated for intervertebral body fusion of the spine in skeletally mature patients. The System is designed for use with autogenous and/or allogeneic bone graft comprised of cancellous and/or corticocancellous bone graft to facilitate fusion and supplemental internal spinal fixation system (e.g., pedicle screw system) cleared by the FDA for use in the lumbar spine in addition to the integrated screws. The devices are to be used in patients who have had at least six months of non-operative treatment.

The NuVasive CoRoent XL Interfixated System implants are interbody fusions in the lumbar spine, from L2 to L5, following discectomy in the treatment of symptomatic degenerative disc disease (DDD) or degenerative spondylolisthesis at one or two adjacent levels. DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. The Nu Vasive CoRoent XL Interfixated System implants can also be used as an adjunct to fusion in patients diagnosed with multilevel degenerative scoliosis and sagittal deformity.

The Brigade System is indicated for spinal fusion procedures in skeletally mature patients. The Brigade Standalone System (lordotic angles of 8° and 12°) is a standalone system. The Brigade Hyperlorio System (lordotic angles of 15° to 30°) must be used with supplemental internal spinal fixation systems (e.g., posterior pedicle screw and rod system) cleared by the FDA for use in the lumbar spine in addition to the integrated screws. The System is designed for use with autogenous and/or allogeneic bone graft to facilitate fusion. The devices are to be used in patients who have had at least six months of non-operative treatment.

The Brigade System is intended for use in interbody fusions in the lumbar spine from L2 to S1, following discectorny in the treatment of symptomatic degenerative disc disease (DDD) or degenerative spondylolisthesis at one or two adjacent levels. DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. The Brigade System platform implants can also be used as an adjunct to fusion in patients diagnosed with multilevel degenerative scoliosis and sagittal deformity. However, when used in these patients at multiple levels and for patients with degenerative spondylolisthesis, the Brigade System platform must be used with a supplemental internal spinal fixation system (e.g., pedicle screw system) cleared by the FDA for use in the lumbar spine in addition to the integrated screws.

The Brigade Lateral System is indicated for spinal fusion procedures in skeletally mature patients. The Brigade Lateral System must be used with supplemental internal spinal fixation systems (e.g., posterior pedicle screw and rod system) that are cleared by the FDA for use in the lumbar spine in addition to the integrated screws. The System is designed for use with autogenous and/or allogenic bone graft comprised of cancellous and/or corticocancellous bone graft to facilitate fusion. The devices are to be used in patients who have had at least six months of non-operative treatment. The Brigade Lateral System is intended for use in interbody fusions in the lumbar spine from L2 to S1, following discectomy in the treatment of symptomatic degenerative disc disease (DDD) or degenerative spondylolisthesis at one or two adjacent levels. DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. The Brigade Lateral System implants can adjunct to fusion in patients diagnosed with multilevel degenerative scoliosis and sagittal deformity.

The BASE Interfixated Titanium System is indicated for spinal fusion procedures in skeletally mature patients. The BASE Interfixated Titanium System 10° - 20° lordotic cages may be used as a standalone system. The BASE Interfixated Titanium System 25° - 30° lordotic cages must be used with supplemental internal spinal fixation systems (i.e., posterior pedicle screw and rod system) that are cleared by the FDA for use in the System is designed for use with autogenous and/or allogeneic bone graft comprised of cancellous and/or corticocancellous bone graft to facilitate fusion. The devices are to be used in patients who have had at least six months of non-operative treatment. The BASE Interfixated Titanium System is interbody fusions in the lumbar spine from L2 to S1, following discectorny in the treatment of symptomatic degenerative disc disease (DDD) or degenerative spondylolisthesis at one or two adjacent levels. DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. The BASE Interfixated System implants can also be used as an adjunct to fusion in patients diagnosed with multilevel degenerative scoliosis and sagittal deformity; however, when used in these patients at multiple levels and for patients with degenerative spondylolisthesis, the BASE Interfixated Titanium System must be used with a supplemental internal spinal fixation system (e.g., pedicle screw system) cleared by the FDA for use in the lumbar spine in addition to the integrated screws.

The Coalesce Thoracolumbar Interbody Fusion System is indicated for intervertebral body fusion of the spine in skeletally mature patients. The System is designed for use with autogenous and/or allogeneic bone graft comprised of cancellous and/or corticocancellous bone graft to facilitate fusion and supplemental internal spinal fixation systems cleared by the FDA for use in the thoracolumbar spine. The devices are to be used in patients who have had at least six months of nonoperative treatment.

The Coalesce Thoracolumbar Interbody Fusion System is interbody fusions in the thoracic spine from T1 to T12 and at the thoracolumbar junction (T12-L1), and is intended for use in the lumbar spine, from L1 to S1, for the treatment of symptomatic disc degeneration (DDD) or degenerative spondylolisthesis at one or two adjacent levels, including thoracic disc herniation (with myelopathy and/or radiculopathy with or without axial pain). DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. The Coalesce Thoracolumbar Interbody Fusion System can be used as an adjunct to fusion in patients diagnosed with multilevel degenerative scoliosis and sagittal deformity.

The Cohere Thoracolumbar Interbody System is indicated for intervertebral body fusion of the spine in skeletally mature patients. The System is designed for use with autogenous and/or allogeneic bone graft comprised of cancellous and/or corticocancellous bone graft to facilitate fusion and supplemental internal spinal fixation systems cleared by the FDA for use in the thoracolumbar spine. The devices are to be used in patients who have had at least six months of non-operative treatment.

The Cohere Thoracolumbar Interbody System is interbody fusions in the thoracic spine, from T1 to T12, and at the thoracolumbar junction (T12-L1), and in the lumbar spine, from L1 to S1, for the treatment of symptomatic disc degeneration (DDD) or degenerative spondylolisthesis at one or two adjacent levels, including thoracic disc herniation (with myelopathy and/or radiculopathy with or without axial pain). DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. The Cohere Thoracolumbar Interbody System can be used as an adjunct to fusion in patients diagnosed with multilevel degenerative scoliosis and sagittal deformity.

The NuVasive Modulus XLIF Interbody System is indicated for intervertebral body fusion of the spine in skeletally mature patients. The System is designed for use with autogenous and/or allogeneic bone graft comprised of cancellous and/or corticocancellous bone graft to facilitate fusion. When used with or without Modulus XLIF internal fixation, the system is intended for use with supplemental spinal fixation system cleared by the FDA for use in the thoracolumbar spine. The devices are to be used in patients who have had at least six months of non-operative treatment. The NuVasive Modulus XLIF Interbody System is intended for use in interbody fusions in the thoracolumbar spine from T1 to T12 and at the thoracolumbar junction (T12-L1), and for use in the lumbar spine from L1 to S1, for the treatment of symptomatic disc degeneration (DDD) or degenerative spondylolisthesis at one or two adjacent levels, including thoracic disc herniation (with myelopathy and/or radiculopathy with or without axial pain). DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. The NuVasive Modulus XLIF Interbody System can also be used as an adjunct to fusion in patients diagnosed with multilevel degenerative scoliosis and sagittal deformity.

The NuVasive Modulus TLIF Interbody System is indicated for intervertebral body fusion of the spine in skeletally mature patients. The System is designed for use with autogenous and/or allogeneic bone graft comprised of cancellous and/or corticocancellous bone graft to facilitate fusion and supplemental internal spinal fixation systems cleared by the FDA for use in the thoracolumbar spine. The devices are to be used in patients who have had at least six months of nonoperative treatment.

The NuVasive Modulus TLIF Interbody System is intended for use in interbody fusions in the thoracolumbar spine from T1 to T12 and at the thoracolumbar junction (T12-L1), and for use in the lumbar spine from L1 to S1, for the treatment of symptomatic disc degeneration (DDD) or degenerative spondylolisthesis at one or two adjacent levels, including thoracic disc herniation (with myelopathy and/or radiculopathy with or without axial pain). DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. The NuVasive Modulus TLIF Interbody System can also be used as an adjunct to fusion in patients diagnosed with multilevel degenerative scoliosis and sagittal deformity.

The Nu Vasive 3DP Interfixated ALIF System is indicated for spinal fusion procedures in skeletally mature patients. The 3DP Interfixated ALIF System 10°-20° lordotic cages may be used as a standalone system. The 3DP Interfixated ALIF System 25°-30° lordotic cages must be used with supplemental internal spinal fixation systems (e.g., posterior pedicle screw and rod system) that are cleared by the FDA for use in the lumbar spine. The 3DP Interfixated ALIF System used with less than 3 bolts must be used with supplemental internal spinal fixation systems (e.g., posterior pedicle screw and rod system) that are cleared by the FDA for use in the lumbar spine. The System is designed for use with autogenous and/or allogeneic bone graft comprised of cancellous and/or corticocancellous bone graft to facilitate fusion. The devices are to be used in patients who have had at least six months of non-operative treatment.

The Nu Vasive 3DP Interfixated ALIF System is intended for use in interbody fusions in the lumbar spine from L2 to S1, following discectorny in the treatment of symptomatic degenerative disc disease (DDD) or degenerative spondylolisthesis at one or two adjacent levels. DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. The 3DP Interfixated ALIF System implants can also be used as an adjunct to fusion in patients diagnosed with multilevel degenerative scoliosis and sagittal deformity; however, when used in these patients at multiple levels and for patients with degenerative spondylolisthesis, the 3DP Interfixated ALIF System must be used with a supplemental internal spinal fixation system (e.g., pedicle screw system) cleared by the FDA for use in the lumbar spine in addition to the integrated screws.

Device Description

The NuVasive Thoracolumbar Interbody Systems are intervertebral body fusion devices, interfixated and non-interfixated. Implants are manufactured of either PEEK-Optima® LT-1 (Polyether-ether-ketone) conforming to ASTM F2026 or Titanium alloy (Ti6Al4V ELI) conforming to ASTM F136/ISO 5832-3. PEEK implants include radiographic markers made of Titanium (Ti) conforming to ASTM F136/ISO 5832-3 or ASTM F1472, or Tantalum (Ta) conforming to ASTM F560 or ISO 13782. Interfixated implants include Titanium alloy (Ti6Al4V ELI) screws conforming to ASTM F136/ISO 5832-3, and in addition, CoRoent Thoracolumbar Interfixated implants include canted coil locking mechanism of Nickel-Cobalt-Chromium-Molybdenum alloy (MP35N) conforming to ASTM F562.

AI/ML Overview

The provided text describes a 510(k) premarket notification for various intervertebral body fusion devices. This type of submission focuses on demonstrating substantial equivalence to legally marketed predicate devices, rather than proving a device meets specific acceptance criteria through a clinical study with detailed performance metrics. Therefore, much of the requested information regarding acceptance criteria, study design, expert involvement, and ground truth establishment is not typically part of a 510(k) summary for these types of Class II spinal implants.

However, I can extract and infer some information based on the document's content.

Here's a breakdown of the requested information:

1. A table of acceptance criteria and the reported device performance:

The document does not provide a table of explicit acceptance criteria or quantitative performance metrics in the way one would see for an AI/CADe device or a new technology requiring de novo clearance. For spinal implants cleared via 510(k), "acceptance criteria" are generally tied to demonstrating substantial equivalence in terms of:

  • Intended Use: The device performs the same function for the same condition.
  • Technological Characteristics: The materials, design, and operating principles are similar or equivalent, and any differences do not raise new questions of safety or effectiveness.
  • Performance: The device performs as safely and effectively as the predicate, often demonstrated through non-clinical (e.g., mechanical, biocompatibility) testing.

The "performance" reported here is a qualitative assessment of substantial equivalence.

Acceptance Criteria (Implied for 510(k) Equivalence)Reported Device Performance
Intended Use Equivalence: Device addresses the same conditions and patient population as predicates."The NuVasive® Thoracolumbar Interbody Systems are indicated for intervertebral body fusion of the spine in skeletally mature patients... for the treatment of symptomatic disc degeneration (DDD) or degenerative spondylolisthesis at one or two adjacent levels, including thoracic disc herniation... can be used as an adjunct to fusion in patients diagnosed with multilevel degenerative scoliosis and sagittal deformity." (Multiple specific indications for each system, mirroring predicate uses).
Technological Characteristics Equivalence: Materials, design, and function are substantially similar to predicates, or differences don't raise new safety/effectiveness concerns."The subject device was shown to be substantially equivalent and have the same technological characteristics to its predicate devices through comparison in areas including design, intended use, material composition, function, and range of sizes."Implants are manufactured from PEEK-Optima® LT-1 or Titanium alloy (Ti6Al4V ELI). Interfixated implants include Titanium alloy screws. (Similar to predicate materials).
Safety and Effectiveness Equivalence: Demonstrated through established clinical use and/or non-clinical testing."Based on the clinical data, it was determined that the NuVasive Thoracolumbar Interbody Systems used in the treatment of thoracolumbar sagittal deformities has a safety and effectiveness profile similar to the predicate device.""No new NuVasive Thoracolumbar Interbody Systems implant designs are being introduced... minor design modifications were made... We include the confirmatory mechanical testing that shows that the minor design modifications do not create a new worst case that would require new or additional testing.""Therefore, no performance testing was performed for this 510(k) submission and the worst case devices included with the subject system were tested and cleared in predicate 510(k) submissions."

2. Sample size used for the test set and the data provenance:

  • Sample Size for Test Set: Not applicable in the context of this 510(k) submission. This submission primarily relies on demonstrating equivalence to existing cleared devices, particularly for a broadened indication (multilevel degenerative scoliosis and sagittal deformity). The document states, "A clinical literature analysis of thoracolumbar sagittal deformities treated with the subject device was performed." This refers to a literature review, not a new clinical study with a prospectively defined test set of patients for device performance evaluation.
  • Data Provenance: The nature of the "clinical literature analysis" means the data provenance would be from published studies, which could originate from various countries and be retrospective or prospective depending on the individual studies included in the analysis. The document does not specify.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

  • Not applicable. As no new clinical study was conducted with a defined "test set" requiring ground truth establishment by experts for specific outcomes, this information is not provided. The determination of "substantial equivalence" is made by the FDA based on the provided technical and comparative information and literature review.

4. Adjudication method for the test set:

  • Not applicable. There was no specific test set requiring adjudication in the context of this 510(k) submission.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

  • Not applicable. This is a submission for spinal implants, not an AI or imaging diagnostic device.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

  • Not applicable. This is a physical implant device, not an algorithm.

7. The type of ground truth used:

  • For this 510(k) submission, the "ground truth" for the new indication (multilevel degenerative scoliosis and sagittal deformity) is established through clinical literature analysis. This means the safety and effectiveness profile of using these types of devices for this condition, as reported in existing peer-reviewed literature, serves as the basis for enabling expanded indications. The FDA relies on the predicate devices' established safety and effectiveness and the literature supporting the similar safety and effectiveness for the expanded use.

8. The sample size for the training set:

  • Not applicable. No "training set" in the machine learning sense was used, as this is a physical medical device. The "training" for the device's design and use comes from existing medical knowledge, biomechanical principles, and the history of predicate devices.

9. How the ground truth for the training set was established:

  • Not applicable. As above, no machine learning training set was used. The basis for the device's design and function comes from established clinical practice, biomechanical data from prior predicate devices, and manufacturing standards for medical implants.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. Underneath the square are the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.

September 25, 2020

NuVasive, Incorporated Ms. Daniela Mahan, Esq., RAC Senior Specialist, Regulatory Affairs 7475 Lusk Blvd San Diego, California 92121

Re: K201820

Trade/Device Name: NuVasive® Thoracolumbar Interbody Systems: CoRoent Thoracolumbar System, CoRoent XL Interfixated System, Brigade Standalone System and Brigade Hyperlordotic System, Brigade Lateral System, BASE Interfixated Titanium System, Coalesce Thoracolumbar Interbody Fusion System, Cohere Thoracolumbar Interbody System, Modulus XLIF Interbody System, Modulus TLIF Interbody System, 3DP Interfixated ALIF System Regulation Number: 21 CFR 888.3080

Regulation Name: Intervertebral Body Fusion Device Regulatory Class: Class II Product Code: MAX, OVD, PHM Dated: June 26, 2020 Received: July 1, 2020

Dear Ms. Mahan:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of

{1}------------------------------------------------

Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Brent L. Showalter, Ph.D. Assistant Director DHT6B: Division of Spinal Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

510(k) Number (if known) K201820

Device Name CoRoent Thoracolumbar System

Indications for Use (Describe)

The NuVasive CoRoent Thoracolumbar System is indicated for intervertebral body fusion of the spine in skeletally mature patients. The System is designed for use with autogenous and/or allogeneic bone graft comprised of cancellous and/or corticocancellous bone graft to facilitate fusion and supplemental internal spinal fixation systems that are cleared by the FDA for use in the thoracolumbar spine. The devices are to be used in patients who have had at least six months of non-operative treatment.

The Nu Vasive CoRoent Thoracolumbar System (XL platform) implants are intended for use in interbody fusions in the thoracic spine, from T1 to T12, and at the thoracolumbar junction (T12-L1), and the CoRoent Thoracolumbar System (XL and L platforms) implants are intended for use in the lumbar spine, from L1 to S1, for the treatment of symptomatic disc degeneration (DDD) or degenerative spondylolisthesis at one or two adjacent levels, including thoracic disc herniation (with myelopathy and/or radiculopathy with or without axial pain). DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. The NuVasive CoRoent Thoracolumbar System (XL and L platforms) can be used as an adjunct to fusion in patients diagnosed with multilevel degenerative scoliosis and sagittal deformity.

Type of Use (Select one or both, as applicable)

☒ Prescription Use (Part 21 CFR 801 Subpart D)☐ Over-The-Counter Use (21 CFR 801 Subpart C)
-----------------------------------------------------------------------------------------------------------------------------------------------------------------------

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

{3}------------------------------------------------

510(k) Number (if known) K201820

Device Name CoRoent XL Interfixated System

Indications for Use (Describe)

The NuVasive CoRoent XL Interfixated System implants are indicated for intervertebral body fusion of the spine in skeletally mature patients. The System is designed for use with autogenous and/or allogeneic bone graft comprised of cancellous and/or corticocancellous bone graft to facilitate fusion and supplemental internal spinal fixation system (e.g., pedicle screw system) cleared by the FDA for use in the lumbar spine in addition to the integrated screws. The devices are to be used in patients who have had at least six months of non-operative treatment.

The NuVasive CoRoent XL Interfixated System implants are interbody fusions in the lumbar spine, from L2 to L5, following discectomy in the treatment of symptomatic degenerative disc disease (DDD) or degenerative spondylolisthesis at one or two adjacent levels. DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. The Nu Vasive CoRoent XL Interfixated System implants can also be used as an adjunct to fusion in patients diagnosed with multilevel degenerative scoliosis and sagittal deformity.

X Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

{4}------------------------------------------------

510(k) Number (if known) K201820

Device Name

Brigade Standalone System and Brigade Hyperlordotic System

Indications for Use (Describe)

The Brigade System is indicated for spinal fusion procedures in skeletally mature patients. The Brigade Standalone System (lordotic angles of 8° and 12°) is a standalone system. The Brigade Hyperlorio System (lordotic angles of 15° to 30°) must be used with supplemental internal spinal fixation systems (e.g., posterior pedicle screw and rod system) cleared by the FDA for use in the lumbar spine in addition to the integrated screws. The System is designed for use with autogenous and/or allogeneic bone graft to facilitate fusion. The devices are to be used in patients who have had at least six months of non-operative treatment.

The Brigade System is intended for use in interbody fusions in the lumbar spine from L2 to S1, following discectorny in the treatment of symptomatic degenerative disc disease (DDD) or degenerative spondylolisthesis at one or two adjacent levels. DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. The Brigade System platform implants can also be used as an adjunct to fusion in patients diagnosed with multilevel degenerative scoliosis and sagittal deformity. However, when used in these patients at multiple levels and for patients with degenerative spondylolisthesis, the Brigade System platform must be used with a supplemental internal spinal fixation system (e.g., pedicle screw system) cleared by the FDA for use in the lumbar spine in addition to the integrated screws.

Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

{5}------------------------------------------------

510(k) Number (if known) K201820

Device Name Brigade Lateral System

Indications for Use (Describe)

The Brigade Lateral System is indicated for spinal fusion procedures in skeletally mature patients. The Brigade Lateral System must be used with supplemental internal spinal fixation systems (e.g., posterior pedicle screw and rod system) that are cleared by the FDA for use in the lumbar spine in addition to the integrated screws. The System is designed for use with autogenous and/or allogenic bone graft comprised of cancellous and/or corticocancellous bone graft to facilitate fusion. The devices are to be used in patients who have had at least six months of non-operative treatment. The Brigade Lateral System is intended for use in interbody fusions in the lumbar spine from L2 to S1, following discectomy in the treatment of symptomatic degenerative disc disease (DDD) or degenerative spondylolisthesis at one or two adjacent levels. DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. The Brigade Lateral System implants can adjunct to fusion in patients diagnosed with multilevel degenerative scoliosis and sagittal deformity.

Type of Use (Select one or both, as applicable)
-------------------------------------------------

X Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

{6}------------------------------------------------

510(k) Number (if known) K201820

Device Name BASE Interfixated Titanium System

Indications for Use (Describe)

The BASE Interfixated Titanium System is indicated for spinal fusion procedures in skeletally mature patients. The BASE Interfixated Titanium System 10° - 20° lordotic cages may be used as a standalone system. The BASE Interfixated Titanium System 25° - 30° lordotic cages must be used with supplemental internal spinal fixation systems (i.e., posterior pedicle screw and rod system) that are cleared by the FDA for use in the System is designed for use with autogenous and/or allogeneic bone graft comprised of cancellous and/or corticocancellous bone graft to facilitate fusion. The devices are to be used in patients who have had at least six months of non-operative treatment. The BASE Interfixated Titanium System is interbody fusions in the lumbar spine from L2 to S1, following discectorny in the treatment of symptomatic degenerative disc disease (DDD) or degenerative spondylolisthesis at one or two adjacent levels. DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. The BASE Interfixated System implants can also be used as an adjunct to fusion in patients diagnosed with multilevel degenerative scoliosis and sagittal deformity; however, when used in these patients at multiple levels and for patients with degenerative spondylolisthesis, the BASE Interfixated Titanium System must be used with a supplemental internal spinal fixation system (e.g., pedicle screw system) cleared by the FDA for use in the lumbar spine in addition to the integrated screws.

Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

{7}------------------------------------------------

510(k) Number (if known) K201820

Device Name

Coalesce Thoracolumbar Interbody Fusion System

Indications for Use (Describe)

The Coalesce Thoracolumbar Interbody Fusion System is indicated for intervertebral body fusion of the spine in skeletally mature patients. The System is designed for use with autogenous and/or allogeneic bone graft comprised of cancellous and/or corticocancellous bone graft to facilitate fusion and supplemental internal spinal fixation systems cleared by the FDA for use in the thoracolumbar spine. The devices are to be used in patients who have had at least six months of nonoperative treatment.

The Coalesce Thoracolumbar Interbody Fusion System is interbody fusions in the thoracic spine from T1 to T12 and at the thoracolumbar junction (T12-L1), and is intended for use in the lumbar spine, from L1 to S1, for the treatment of symptomatic disc degeneration (DDD) or degenerative spondylolisthesis at one or two adjacent levels, including thoracic disc herniation (with myelopathy and/or radiculopathy with or without axial pain). DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. The Coalesce Thoracolumbar Interbody Fusion System can be used as an adjunct to fusion in patients diagnosed with multilevel degenerative scoliosis and sagittal deformity.

Type of Use (Select one or both, as applicable)

☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

{8}------------------------------------------------

510(k) Number (if known) K201820

Device Name

Cohere Thoracolumbar Interbody System

Indications for Use (Describe)

The Cohere Thoracolumbar Interbody System is indicated for intervertebral body fusion of the spine in skeletally mature patients. The System is designed for use with autogenous and/or allogeneic bone graft comprised of cancellous and/or corticocancellous bone graft to facilitate fusion and supplemental internal spinal fixation systems cleared by the FDA for use in the thoracolumbar spine. The devices are to be used in patients who have had at least six months of non-operative treatment.

The Cohere Thoracolumbar Interbody System is interbody fusions in the thoracic spine, from T1 to T12, and at the thoracolumbar junction (T12-L1), and in the lumbar spine, from L1 to S1, for the treatment of symptomatic disc degeneration (DDD) or degenerative spondylolisthesis at one or two adjacent levels, including thoracic disc herniation (with myelopathy and/or radiculopathy with or without axial pain). DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. The Cohere Thoracolumbar Interbody System can be used as an adjunct to fusion in patients diagnosed with multilevel degenerative scoliosis and sagittal deformity.

Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

{9}------------------------------------------------

510(k) Number (if known) K201820

Device Name Modulus XLIF Interbody System

Indications for Use (Describe)

The NuVasive Modulus XLIF Interbody System is indicated for intervertebral body fusion of the spine in skeletally mature patients. The System is designed for use with autogenous and/or allogeneic bone graft comprised of cancellous and/or corticocancellous bone graft to facilitate fusion. When used with or without Modulus XLIF internal fixation, the system is intended for use with supplemental spinal fixation system cleared by the FDA for use in the thoracolumbar spine. The devices are to be used in patients who have had at least six months of non-operative treatment. The NuVasive Modulus XLIF Interbody System is intended for use in interbody fusions in the thoracolumbar spine from T1 to T12 and at the thoracolumbar junction (T12-L1), and for use in the lumbar spine from L1 to S1, for the treatment of symptomatic disc degeneration (DDD) or degenerative spondylolisthesis at one or two adjacent levels, including thoracic disc herniation (with myelopathy and/or radiculopathy with or without axial pain). DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. The NuVasive Modulus XLIF Interbody System can also be used as an adjunct to fusion in patients diagnosed with multilevel degenerative scoliosis and sagittal deformity.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

{10}------------------------------------------------

510(k) Number (if known) K201820

Device Name Modulus TLIF Interbody System

Indications for Use (Describe)

The NuVasive Modulus TLIF Interbody System is indicated for intervertebral body fusion of the spine in skeletally mature patients. The System is designed for use with autogenous and/or allogeneic bone graft comprised of cancellous and/or corticocancellous bone graft to facilitate fusion and supplemental internal spinal fixation systems cleared by the FDA for use in the thoracolumbar spine. The devices are to be used in patients who have had at least six months of nonoperative treatment.

The NuVasive Modulus TLIF Interbody System is intended for use in interbody fusions in the thoracolumbar spine from T1 to T12 and at the thoracolumbar junction (T12-L1), and for use in the lumbar spine from L1 to S1, for the treatment of symptomatic disc degeneration (DDD) or degenerative spondylolisthesis at one or two adjacent levels, including thoracic disc herniation (with myelopathy and/or radiculopathy with or without axial pain). DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. The NuVasive Modulus TLIF Interbody System can also be used as an adjunct to fusion in patients diagnosed with multilevel degenerative scoliosis and sagittal deformity.

Type of Use (Select one or both, as applicable)
× Prescription Use (Part 21 CFR 801 Subpart D)Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

{11}------------------------------------------------

510(k) Number (if known) K201820

Device Name 3DP Interfixated ALIF System

Indications for Use (Describe)

The Nu Vasive 3DP Interfixated ALIF System is indicated for spinal fusion procedures in skeletally mature patients. The 3DP Interfixated ALIF System 10°-20° lordotic cages may be used as a standalone system. The 3DP Interfixated ALIF System 25°-30° lordotic cages must be used with supplemental internal spinal fixation systems (e.g., posterior pedicle screw and rod system) that are cleared by the FDA for use in the lumbar spine. The System is designed for use with autogenous and/or allogeneic bone graft comprised of cancellous bone graft to facilitate fusion. The devices are to be used in patients who have had at least six months of non-operative treatment.

The Nu Vasive 3DP Interfixated ALIF System is intended for use in interbody fusions in the lumbar spine from L2 to S1, following discectorny in the treatment of symptomatic degenerative disc disease (DDD) or degenerative spondylolisthesis at one or two adjacent levels. DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. The 3DP Interfixated ALIF System implants can also be used as an adjunct to fusion in patients diagnosed with multilevel degenerative scoliosis and sagittal deformity; however, when used in these patients at multiple levels and for patients with degenerative spondylolisthesis, the 3DP Interfixated ALIF System must be used with a supplemental internal spinal fixation system (e.g., pedicle screw system) cleared by the FDA for use in the lumbar spine in addition to the integrated screws.

Type of Use (Select one or both, as applicable)

☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

{12}------------------------------------------------

Image /page/12/Picture/0 description: The image shows the logo for NuVasive. The logo consists of a stylized leaf shape in purple and gray, followed by the word "NUVASIVE" in gray sans-serif font. The leaf shape is split into two parts, with the top part in purple and the bottom part in gray.

510(k) Summary

In accordance with Title 21 of the Code of Federal Regulations, Part 807, and in particular 21 CFR §807.92, the following summary of information is provided:

Submitted by: A.

Ms. Daniela Mahan, Esq., RAC Sr. Specialist, Regulatory Affairs NuVasive, Incorporated 7475 Lusk Blvd. San Diego, California 92121 (858) 458-2285 Date Prepared: June 26, 2020

B. Device Name

Trade or Proprietary Name:NuVasive® Thoracolumbar Interbody Systems:
CoRoent Thoracolumbar System
CoRoent XL Interfixated System
Brigade Standalone System and Brigade Hyperlordotic System
Brigade Lateral System
BASE Interfixated Titanium System
Coalesce Thoracolumbar Interbody Fusion System
Cohere Thoracolumbar Interbody System
Modulus XLIF Interbody System
Modulus TLIF Interbody System
3DP Interfixated ALIF System
Common or Usual Name:Intervertebral Body Fusion Device
Classification Name:Intervertebral Body Fusion Device
Device Class:Class II
Classification:21 CFR 888.3080
Product Code:MAX, PHM, OVD

C. Predicate Devices

The subject NuVasive Thoracolumbar Interbody Systems is substantially equivalent to the following devices:

Primary Predicate Device:

Product Name510(k)NumberDate of FDA-Clearance
NuVasive CoRoent Thoracolumbar SystemK1709626/26/2017

{13}------------------------------------------------

Additional Predicate Devices:

Product Name510(k)NumberDate of FDA-Clearance
CoRoent Thoracolumbar SystemK1534194/13/2016
NuVasive Lumbar Interbody ImplantsK1612308/25/2016
Brigade SystemK1612308/25/2016
Brigade Lateral SystemK1813868/24/2018
BASE Interfixated Titanium SystemK1705924/26/2017
Vertera Spine Coalesce Thoracolumbar InterbodyFusion SystemK17315312/6/2017
NuVasive Cohere Thoracolumbar Interbody SystemK18186010/5/2018
NuVasive Cohere Thoracolumbar Interbody SystemK1935413/30/2020
Modulus TLIF Interbody SystemK17234110/26/2017
Modulus XLIF Interbody SystemK19276010/18/2019
3DP Interfixated ALIF SystemK1935933/25/2020

D. Device Description

The NuVasive Thoracolumbar Interbody Systems are intervertebral body fusion devices, interfixated and non-interfixated. Implants are manufactured of either PEEK-Optima® LT-1 (Polyether-ether-ketone) conforming to ASTM F2026 or Titanium alloy (Ti6Al4V ELI) conforming to ASTM F136/ISO 5832-3. PEEK implants include radiographic markers made of Titanium (Ti) conforming to ASTM F136/ISO 5832-3 or ASTM F1472, or Tantalum (Ta) conforming to ASTM F560 or ISO 13782. Interfixated implants include Titanium alloy (Ti6Al4V ELI) screws conforming to ASTM F136/ISO 5832-3, and in addition, CoRoent Thoracolumbar Interfixated implants include canted coil locking mechanism of Nickel-Cobalt-Chromium-Molybdenum alloy (MP35N) conforming to ASTM F562.

CoRoent Thoracolumbar System

The subject NuVasive CoRoent Thoracolumbar System are interbodies manufactured from PEEK-Optima® LT-1 conforming to ASTM F2026 or titanium alloy (Ti-6A1-V) conforming to ASTM F136. The PEEK devices contain titanium alloy radiographic markers conforming to ASTM F136 or ASTM F1472 or tantalum markers conforming to ASTM 560 or ISO 13782. The device's hollow core or graft aperture allows for packing of autograft to help promote a solid fusion. Small spikes or teeth on each end of the device serve to grip the adjacent vertebrae to resist migration and expulsion of the device.

The implants are available in a variety sizes and lordotic angles to suit the individual pathology and anatomical conditions of the patient. The device is intended to be used with supplemental internal spinal fixation systems that are cleared by the FDA for use in the lumbar spine.

{14}------------------------------------------------

CoRoent XL Interfixated System and Brigade Hyperlordotic System

The subject CoRoent Thoracolumbar System (Interfixated) and Brigade Hyperlordotic System are interbodies manufactured from PEEK-Optima LT-1 conforming to ASTM F2026 PEEK-Optima® LT-1, with radiographic markers manufactured from titanium alloy (Ti-6A1-4V) conforming to ASTM F136, and CoRoent Thoracolumbar Interfixated implants include canted coil locking mechanism of Nickel-Cobalt-Chromium-Molybdenum alloy (MP35N) conforming to ASTM F562. The devices all contain integrated screws made of titanium alloy, and include a hollow core or graft aperture which allows for packing of autograft to help promote a solid fusion. The subject implants contain small spikes or teeth on each end of the device, which serve to grip the adjacent vertebrae to resist migration and expulsion.

The implants are available in a variety sizes and lordotic angles to suit the individual pathology and anatomical conditions of the patient. In addition to the integrated screws, the CoRoent XL-F System and Brigade Hyperlordotic System devices are intended to be used with supplemental internal spinal fixation systems that are cleared by the FDA for use in the lumbar spine.

Coalesce Thoracolumbar Interbody Fusion System

The Vertera Spine Coalesce Thoracolumbar Interbody Fusion System comprises of sterile, single use implant grade polyetheretherketone (PEEK) devices, available in varied footprints and heights, designed for supplemental stabilization of the thoracolumbar spinal column in thoracolumbar intervertebral body fusion procedures.

Each device within the Coalesce System is comprised of a continuous body of PEEK formed into the final product shape with a porous architecture on select faces of the implant. The porous architecture is derived directly from the implant body and is not a sintered or otherwise additive coating. In addition to PEEK, the device assembly may contain two or more tantalum markers, depending on footprint, to enable visibility under xray in vivo.

Cohere Thoracolumbar Interbody System

The NuVasive Cohere Thoracolumbar Interbody System comprises of sterile, single use implant grade polyetheretherketone (PEEK) devices, available in varied footprints and heights, designed for supplemental stabilization of the thoracolumbar spinal column in thoracolumbar intervertebral body fusion procedures.

Each device within the Cohere Thoracolumbar Interbody System is comprised of a continuous body of PEEK formed into the final product shape with a porous architecture on select faces of the implant. The porous architecture is derived directly from the implant body and is not a sintered or otherwise additive coating. In addition to PEEK, the device assembly may contain two or more radiolucent markers, depending on footprint, to enable visibility under x-ray in vivo.

{15}------------------------------------------------

Brigade Standalone System

The NuVasive Brigade Hyperlordotic System is an interbody system manufactured from PEEK and titanium alloy conforming to industry recognized standards. The NuVasive Brigade Hyperlordotic System is available in a variety of different shapes and sizes to suit the individual pathology and anatomical conditions of the patient. The Brigade Hyperlordotic intervertebral fusion device is a device composed of a PEEK interbody implant containing radiographic titanium alloy markers, and-four (4) titanium alloy bone screws. The subject device components are made from Polyetheretherketone (PEEK OPTIMA LTI) conforming to ASTM F-2026 and titanium alloy (Ti-6A1-4V ELI) conforming to ASTM F136 and 150 5832-3.

Brigade Lateral System

The NuVasive® Brigade® Lateral System is an interfixated interbody system manufactured from PEEK and titanium alloy conforming to industry recognized standards. The NuVasive Brigade Lateral System is available in a variety of different shapes and sizes to suit the individual pathology and anatomical conditions of the patient. The Brigade Lateral System intervertebral fusion device is composed of a PEEK interbody implant containing radiographic titanium alloy markers, and two (2) titanium alloy bone screws. The subject device components are made from Polyetheretherketone (PEEK-OPTIMAL LT1) conforming to ASTM F2026 and titanium alloy (Ti-6A-4V ELI) conforming to ASTM F136 and ISO 5832-3 or (Ti-6A-4V) conforming to ASTM F1472.

Base Interfixated Titanium System

The NuVasive BASE Interfixated Titanium System is an interfixated interbody system manufactured from titanium alloy (Ti-6Al-4V ELI) conforming to ASTM F136 and ISO 5832-3. The NuVasive BASE Interfixated Titanium System is available in a variety of different shapes and sizes to suit the individual pathology and anatomical conditions of the patient. The BASE Interfixated Titanium System consists of a titanium alloy interbody and three (3) titanium alloy bone screws. The BASE Interfixated Titanium System 10° - 20° lordotic cages may be used as a standalone system. The BASE Interfixated Titanium System 25° - 30° lordotic cages must be used with supplemental internal spinal fixation systems (i.e. posterior pedicle screw and rod system) that are cleared by the FDA for use in the lumbar spine.

Modulus XLIF

The subject NuVasive Modulus XLIF Interbody System are interbody implants manufactured from titanium alloy (Ti-6Al-4V ELI) powder conforming to ASTM F3001. The solid and porous structures are simultaneously built using a powder bed fusion method. The hollow core, or graft aperture, allows for packing of graft to aid in the promotion of a solid fusion. Similarly, the macroporous internal lattice structure provides additional space for graft packing. The microporous, textured surface on the superior and inferior ends of the device serves to grip the adjacent vertebrae to resist migration and expulsion of the

{16}------------------------------------------------

Image /page/16/Picture/1 description: The image shows the Nuvasive logo. The logo consists of a purple and gray abstract leaf-like shape on the left, followed by the word "NUVASIVE" in gray, sans-serif font. The logo is simple and modern, with a clean design.

device. The device is intended to be used with supplemental internal spinal fixation systems that are cleared by the FDA for use in the thoracolumbar spine.

The implants are available in a variety sizes and lordotic angles to suit the individual pathology and anatomical conditions of the patient. The device is intended to be used with supplemental spinal fixation systems that are cleared by the FDA for use in the lumbar spine.

Modulus TLIF

The subject NuVasive Modulus TLIF Interbody System are interbody implants manufactured from titanium alloy (Ti-6Al-4V ELI) powder conforming to ASTM F3001. The solid and porous structures are simultaneously built using a powder bed fusion method. The hollow core, or graft aperture, allows for packing of graft to aid in the promotion of a solid fusion. Similarly, the macroporous internal lattice structure provides additional space for graft packing. The microporous, textured surface on the superior and inferior ends of the device serves to grip the adjacent vertebrae to resist migration and expulsion of the device. The device is intended to be used with supplemental internal spinal fixation systems that are cleared by the FDA for use in the thoracolumbar spine.

The implants are available in a variety sizes and lordotic angles to suit the individual pathology and anatomical conditions of the patient. The device is intended to be used with supplemental spinal fixation systems that are cleared by the FDA for use in the thoracolumbar spine.

3DP Interfixated ALIF System

The subject NuVasive 3DP Interfixated System are interbody implants manufactured from titanium alloy (Ti-6A1-4V ELI) powder conforming to ASTM F3001. The solid and porous structures are simultaneously built using a powder bed fusion method. The hollow core, or graft aperture, allows for packing of graft to aid in the promotion of a solid fusion. Similarly, the macroporous internal lattice structure provides additional space for graft packing. The microporous, textured surface on the superior and inferior ends of the device serves to grip the adjacent vertebrae to resist migration of the device. The implants are available in a variety sizes and lordotic angles to suit the individual pathology and anatomical conditions of the patient. The 3DP Interfixated ALIF System 10°-20° lordotic cages may be used as a standalone system. The 3DP Interfixated ALIF System 25° - 30° lordotic cages must be used with supplemental internal spinal fixation systems (e.g. posterior pedicle screw and rod system) that are cleared by the FDA for use in the lumbar spine.

E. Indications for Use

1) CoRoent Thoracolumbar System:

The NuVasive CoRoent Thoracolumbar System is indicated for intervertebral body fusion of the spine in skeletally mature patients. The System is designed for use with autogenous and/or allogeneic bone graft comprised of cancellous and/or corticocancellous bone graft to

{17}------------------------------------------------

Image /page/17/Picture/1 description: The image shows the Nuvasive logo. The logo consists of a stylized leaf shape in purple and gray, followed by the word "NUVASIVE" in gray. The leaf shape is on the left side of the logo, and the word "NUVASIVE" is on the right side.

facilitate fusion and supplemental internal spinal fixation systems that are cleared by the FDA for use in the thoracolumbar spine. The devices are to be used in patients who have had at least six months of non-operative treatment.

The NuVasive CoRoent Thoracolumbar System (XL platform) implants are intended for use in interbody fusions in the thoracic spine, from T1 to T12, and at the thoracolumbar junction (T12-L1), and the CoRoent Thoracolumbar System (XL and L platforms) implants are intended for use in the lumbar spine, from L1 to S1, for the treatment of symptomatic disc degeneration (DDD) or degenerative spondylolisthesis at one or two adjacent levels, including thoracic disc herniation (with myelopathy and/or radiculopathy with or without axial pain). DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. The NuVasive CoRoent Thoracolumbar System (XL and L platforms) can be used as an adjunct to fusion in patients diagnosed with multilevel degenerative scoliosis and sagittal deformity.

2) CoRoent XL Interfixated System:

The NuVasive CoRoent XL Interfixated System implants are indicated for intervertebral body fusion of the spine in skeletally mature patients. The System is designed for use with autogenous and/or allogeneic bone graft comprised of cancellous and/or corticocancellous bone graft to facilitate fusion and supplemental internal spinal fixation system (e.g., pedicle screw system) cleared by the FDA for use in the lumbar spine in addition to the integrated screws. The devices are to be used in patients who have had at least six months of nonoperative treatment.

The NuVasive CoRoent XL Interfixated System implants are intended for use in interbody fusions in the lumbar spine, from L2 to L5, following discectomy in the treatment of symptomatic degenerative disc disease (DDD) or degenerative spondylolisthesis at one or two adjacent levels. DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. The NuVasive CoRoent XL Interfixated System implants can also be used as an adjunct to fusion in patients diagnosed with multilevel degenerative scoliosis and sagittal deformity.

3) Brigade Standalone System and Brigade Hyperlordotic System

The Brigade System is indicated for spinal fusion procedures in skeletally mature patients. The Brigade Standalone System (lordotic angles of 8° and 12°) is a standalone system. The Brigade Hyperlordotic System (lordotic angles of 15° to 30°) must be used with supplemental internal spinal fixation systems (e.g., posterior pedicle screw and rod system) cleared by the FDA for use in the lumbar spine in addition to the integrated screws. The System is designed for use with autogenous and/or allogeneic bone graft to facilitate fusion. The devices are to be used in patients who have had at least six months of nonoperative treatment.

The Brigade System is intended for use in interbody fusions in the lumbar spine from L2 to S1, following discectomy in the treatment of symptomatic degenerative disc disease (DDD) or degenerative spondylolisthesis at one or two adjacent levels. DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. The Brigade System platform implants can also be used as an adjunct to fusion in patients diagnosed with multilevel degenerative scoliosis and sagittal deformity. However, when used in these patients at multiple levels and for patients with

{18}------------------------------------------------

Image /page/18/Picture/1 description: The image shows the Nuvasive logo. The logo consists of a stylized leaf shape in purple and gray on the left, followed by the word "NUVASIVE" in gray. The leaf shape is split into two parts, with the top part in purple and the bottom part in gray. The word "NUVASIVE" is in a sans-serif font.

degenerative spondylolisthesis, the Brigade System must be used with a supplemental internal spinal fixation system (e.g., pedicle screw system) cleared by the FDA for use in the lumbar spine in addition to the integrated screws.

4) Brigade Lateral System

The Brigade Lateral System is indicated for spinal fusion procedures in skeletally mature patients. The Brigade Lateral System must be used with supplemental internal spinal fixation systems (e.g., posterior pedicle screw and rod system) that are cleared by the FDA for use in the lumbar spine in addition to the integrated screws. The System is designed for use with autogenous and/or allogenic bone graft comprised of cancellous and/or corticocancellous bone graft to facilitate fusion. The devices are to be used in patients who have had at least six months of non-operative treatment.

The Brigade Lateral System is intended for use in interbody fusions in the lumbar spine from L2 to S1, following discectomy in the treatment of symptomatic degenerative disc disease (DDD) or degenerative spondylolisthesis at one or two adjacent levels. DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. The Brigade Lateral System implants can also be used as an adjunct to fusion in patients diagnosed with multilevel degenerative scoliosis and sagittal deformity.

5) BASE Titanium Interfixated System

The BASE Interfixated Titanium System is indicated for spinal fusion procedures in skeletally mature patients. The BASE Interfixated Titanium System 10° - 20° lordotic cages may be used as a standalone system. The BASE Interfixated Titanium System 25° -30° lordotic cages must be used with supplemental internal spinal fixation systems (i.e., posterior pedicle screw and rod system) that are cleared by the FDA for use in the lumbar spine. The System is designed for use with autogenous and/or allogeneic bone graft comprised of cancellous and/or corticocancellous bone graft to facilitate fusion. The devices are to be used in patients who have had at least six months of non-operative treatment.

The BASE Interfixated Titanium System is intended for use in interbody fusions in the lumbar spine from L2 to S1, following discectomy in the treatment of symptomatic degenerative disc disease (DDD) or degenerative spondylolisthesis at one or two adjacent levels. DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. The BASE Interfixated System implants can also be used as an adjunct to fusion in patients diagnosed with multilevel degenerative scoliosis and sagittal deformity; however, when used in these patients at multiple levels and for patients with degenerative spondylolisthesis, the BASE Interfixated Titanium System must be used with a supplemental internal spinal fixation system (e.g., pedicle screw system) cleared by the FDA for use in the lumbar spine in addition to the integrated screws.

6) Coalesce Thoracolumbar Interbody Fusion System

The Vertera Spine Coalesce Thoracolumbar Interbody Fusion System is indicated for intervertebral body fusion of the spine in skeletally mature patients. The System is designed for use with autogenous and/or allogeneic bone graft comprised of cancellous and/or corticocancellous bone graft to facilitate fusion and supplemental internal spinal

{19}------------------------------------------------

Image /page/19/Picture/1 description: The image shows the Nuvasive logo. The logo consists of a stylized leaf shape in purple and gray, followed by the word "NUVASIVE" in gray. The leaf shape is divided into two parts, with the top part in purple and the bottom part in gray. The word "NUVASIVE" is in a sans-serif font, and there is a trademark symbol after the word.

fixation systems cleared by the FDA for use in the thoracolumbar spine. The devices are to be used in patients who have had at least six months of non-operative treatment.

The Vertera Spine Coalesce Thoracolumbar Interbody Fusion System is intended for use in interbody fusions in the thoracic spine from T1 to T12 and at the thoracolumbar junction (T12-L1), and is intended for use in the lumbar spine, from L1 to S1, for the treatment of symptomatic disc degeneration (DDD) or degenerative spondylolisthesis at one or two adjacent levels, including thoracic disc herniation (with myelopathy and/or radiculopathy with or without axial pain). DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. The Vertera Spine Coalesce Thoracolumbar Interbody Fusion System can be used as an adjunct to fusion in patients diagnosed with multilevel degenerative scoliosis and sagittal deformity.

7) Cohere Thoracolumbar Interbody Systems

The NuVasive Cohere Thoracolumbar Interbody System is indicated for intervertebral body fusion of the spine in skeletally mature patients. The System is designed for use with autogenous and/or allogeneic bone graft comprised of cancellous and/or corticocancellous bone graft to facilitate fusion and supplemental internal spinal fixation systems cleared by the FDA for use in the thoracolumbar spine. The devices are to be used in patients who have had at least six months of non-operative treatment.

The NuVasive Cohere Thoracolumbar Interbody System is intended for use in interbody fusions in the thoracic spine, from T1 to T12, and at the thoracolumbar junction (T12-L1), and in the lumbar spine, from L1 to S1, for the treatment of symptomatic disc degeneration (DDD) or degenerative spondylolisthesis at one or two adjacent levels, including thoracic disc herniation (with myelopathy and/or radiculopathy with or without axial pain). DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. The NuVasive Cohere Thoracolumbar Interbody System can be used as an adjunct to fusion in patients diagnosed with multilevel degenerative scoliosis and sagittal deformity.

8) Modulus XLIF Interbody System

The NuVasive Modulus XLIF Interbody System is indicated for intervertebral body fusion of the spine in skeletally mature patients. The System is designed for use with autogenous and/or allogeneic bone graft comprised of cancellous and/or corticocancellous bone graft to facilitate fusion. When used with or without Modulus XLIF internal fixation, the system is intended for use with supplemental spinal fixation system cleared by the FDA for use in the thoracolumbar spine. The devices are to be used in patients who have had at least six months of non-operative treatment.

The NuVasive Modulus XLIF Interbody System is intended for use in interbody fusions in the thoracolumbar spine from T1 to T12 and at the thoracolumbar junction (T12-L1), and for use in the lumbar spine from L1 to S1, for the treatment of symptomatic disc degeneration (DDD) or degenerative spondylolisthesis at one or two adjacent levels, including thoracic disc herniation (with myelopathy and/or radiculopathy with or without axial pain). DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. The NuVasive Modulus XLIF Interbody System can also be used as an adjunct to fusion in patients diagnosed with multilevel degenerative scoliosis and sagittal deformity.

{20}------------------------------------------------

Image /page/20/Picture/1 description: The image shows the logo for Nuvasive. The logo consists of a purple and gray abstract shape on the left, followed by the word "NUVASIVE" in gray. The logo is simple and modern.

9) Modulus TLIF Interbody System

The NuVasive Modulus TLIF Interbody System is indicated for intervertebral body fusion of the spine in skeletally mature patients. The System is designed for use with autogenous and/or allogeneic bone graft comprised of cancellous and/or corticocancellous bone graft to facilitate fusion and supplemental internal spinal fixation systems cleared by the FDA for use in the thoracolumbar spine. The devices are to be used in patients who have had at least six months of non-operative treatment.

The NuVasive Modulus TLIF Interbody System is intended for use in interbody fusions in the thoracolumbar spine from T1 to T12 and at the thoracolumbar junction (T12-L1), and for use in the lumbar spine from L1 to S1, for the treatment of symptomatic disc degeneration (DDD) or degenerative spondylolisthesis at one or two adjacent levels, including thoracic disc herniation (with myelopathy and/or radiculopathy with or without axial pain). DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. The NuVasive Modulus TLIF Interbody System can also be used as an adjunct to fusion in patients diagnosed with multilevel degenerative scoliosis and sagittal deformity.

10) 3DP Interfixated ALIF System

The NuVasive 3DP Interfixated ALIF System is indicated for spinal fusion procedures in skeletally mature patients. The 3DP Interfixated ALIF System 10°-20° lordotic cages may be used as a standalone system. The 3DP Interfixated ALIF System 25°-30° lordotic cages must be used with supplemental internal spinal fixation systems (e.g., posterior pedicle screw and rod system) that are cleared by the FDA for use in the lumbar spine. The 3DP Interfixated ALIF System used with less than 3 bolts must be used with supplemental internal spinal fixation systems (e.g., posterior pedicle screw and rod system) that are cleared by the FDA for use in the lumbar spine. The System is designed for use with autogenous and/or allogeneic bone graft comprised of cancellous and/or corticocancellous bone graft to facilitate fusion. The devices are to be used in patients who have had at least six months of non-operative treatment.

The NuVasive 3DP Interfixated ALIF System is intended for use in interbody fusions in the lumbar spine from L2 to S1, following discectomy in the treatment of symptomatic degenerative disc disease (DDD) or degenerative spondylolisthesis at one or two adjacent levels. DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. The 3DP Interfixated ALIF System implants can also be used as an adjunct to fusion in patients diagnosed with multilevel degenerative scoliosis and sagittal deformity; however, when used in these patients at multiple levels and for patients with degenerative spondylolisthesis, the 3DP Interfixated ALIF System must be used with a supplemental internal spinal fixation system (e.g., pedicle screw system) cleared by the FDA for use in the lumbar spine in addition to the integrated screws.

F. Technological Characteristics

As was established in this submission, the subject NuVasive Thoracolumbar Interbody Systems are substantially equivalent to other predicate devices cleared by the FDA for commercial distribution in the United States. The subject device was shown to be substantially equivalent and have the same technological characteristics to its predicate

{21}------------------------------------------------

devices through comparison in areas including design, intended use, material composition, function, and range of sizes. This device does not contain software or electrical equipment.

G. Performance Data

The purpose of the submission is to expand indications to include the treatment of multilevel sagittal deformity in the thoracolumbar spine. A clinical literature analysis of thoracolumbar sagittal deformities treated with the subject device was performed. Based on the clinical data, it was determined that the NuVasive Thoracolumbar Interbody Systems used in the treatment of thoracolumbar sagittal deformities has a safety and effectiveness profile similar to the predicate device.

No new NuVasive Thoracolumbar Interbody Systems implant designs are being introduced to the previously cleared CoRoent Thoracolumbar System (K153419, K161230, K170962), Brigade System (K161230, K181386), BASE Interfixated Titanium System (K170592), Vertera Spine Coalesce Thoracolumbar Interbody Fusion System (K173153), Cohere Thoracolumbar Interbody System (K181860, K193541), Modulus Interbody System (K172341, K192760) and 3DP Interfixated ALIF System (K193593) implants. Nevertheless, minor design modifications were made to certain devices within the CoRoent Thoracolumbar System (K161230), Coalesce Thoracolumbar Interbody System (K163506), Cohere Thoracolumbar Interbody System (K181860) and Modulus TLIF Interbody System (K172341) systems via add-to-file. We include the confirmatory mechanical testing that shows that the minor design modifications do not create a new worst case that would require new or additional testing. Since multilevel sagittal deformity use does not change the biomechanical stresses placed upon the individual implants, additional non-clinical testing is unwarranted. Therefore, no performance testing was performed for this 510(k) submission and the worst case devices included with the subject system were tested and cleared in predicate 510(k) submissions. Only previously cleared devices are the subject of this submission, i.e., there is no new worst case device.

H. Conclusions

Based on the indications for use, technological characteristics, performance testing, and comparison to predicate devices, the subject NuVasive Thoracolumbar Interbody Systems have been shown to be substantially equivalent to legally marketed predicate devices.

§ 888.3080 Intervertebral body fusion device.

(a)
Identification. An intervertebral body fusion device is an implanted single or multiple component spinal device made from a variety of materials, including titanium and polymers. The device is inserted into the intervertebral body space of the cervical or lumbosacral spine, and is intended for intervertebral body fusion.(b)
Classification. (1) Class II (special controls) for intervertebral body fusion devices that contain bone grafting material. The special control is the FDA guidance document entitled “Class II Special Controls Guidance Document: Intervertebral Body Fusion Device.” See § 888.1(e) for the availability of this guidance document.(2) Class III (premarket approval) for intervertebral body fusion devices that include any therapeutic biologic (e.g., bone morphogenic protein). Intervertebral body fusion devices that contain any therapeutic biologic require premarket approval.
(c)
Date premarket approval application (PMA) or notice of product development protocol (PDP) is required. Devices described in paragraph (b)(2) of this section shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.