K Number
K181381
Date Cleared
2018-11-23

(183 days)

Product Code
Regulation Number
872.3630
Panel
Dental
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

A.B. DENTAL DEVICES® Dental Implants System is indicated for use in surgical and restorative applications for placement in the bone of the upper or lower jaw to provide support for prosthetic devices, such as artificial teeth, in order to restore the patient's chewing function. A.B. DENTAL DEVICES® Dental Implants System is indicated also for immediate loading when good primary stability is achieved and with appropriate occlusal loading.

Device Description

A.B. DENTAL DEVICES® Dental Implants System consists of one and two stage endosseous form dental implants, internal hexagonal and one piece implants systems. Narrow platform implants are to be used with straight abutments only. The following components are being added:

  • . P0-3 narrow platform healing caps, 4 and 7 mm length
  • PON-3.75 standard platform, narrow healing caps, 4 and 7 mm length .
  • P3-5 standard platform, straight abutment, 5 mm diameter, 5 mm length ●
  • P3S-3 narrow platform, anatomic, straight abutment, 1, 2, 3 mm shoulder lengths
  • . P3SW-3.75 standard platform, anatomic, straight abutment, wide diameter, 1, 2, 3 mm shoulder lengths
  • P3W-3 narrow platform, straight abutment, wide diameter, 9 mm length
  • P5-3 narrow platform, ball attachment, 4, 5, 6 mm height
  • P5-3.75-20 standard platform, 20º ball attachment, 4 and 6 mm height ●
  • P9HG-3.75 standard platform, composed abutment, gold alloy
  • P9R and P9HR -3.75 standard platform, composed abutment, CoCr alloy
  • . P25-3 narrow platform, A.B. LOC abutments, 0, 1, 2, 3, 4 5 mm height
  • . P25-3.75 standard platform. A.B. LOC abutments. 0. 1. 2 mm height
  • . P64-3.75 standard platform, straight adaptor abutment, 1, 2, 3, 4, 5 mm heights
  • P64 plastic and Ti alloy sleeve
  • P64 healing cap ●
  • PK standard platform, anatomic anti-rotation abutment, 1, 2, 3, 4 mm height
  • PK plastic sleeve
  • . PK healing cap
AI/ML Overview

This document is a 510(k) premarket notification for the A.B. Dental Devices® Dental Implants System. It describes the device, its indications for use, and a comparison to predicate devices, but it does not contain information about acceptance criteria or a study proving the device meets those criteria in a typical medical device AI context.

The "Performance Data" section lists several types of engineering analysis and biocompatibility/sterilization/fatigue testing, but these are for the physical dental implant system itself, not for an AI/algorithm-based diagnostic or assistive device. There is no mention of a "device performance" in terms of clinical metrics like sensitivity, specificity, accuracy, or an effect size for human readers.

Therefore, for the specific questions asked about acceptance criteria and a study proving device performance (especially questions related to AI/algorithm performance), the provided text does not contain the requested information.

Here's why and what's missing:

  1. A table of acceptance criteria and the reported device performance: Not present. The document focuses on comparing the new components (healing caps, abutments) to previously cleared predicate devices based on material, diameter, platform, length, design, and sterility. There are no performance metrics or acceptance criteria for a diagnostic/assistive function.

  2. Sample size used for the test set and the data provenance: Not applicable as there is no "test set" for an AI algorithm. The performance data mentioned relates to physical testing (e.g., fatigue testing), but not clinical trials or data for an algorithm.

  3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable. There is no ground truth established for an AI algorithm.

  4. Adjudication method for the test set: Not applicable.

  5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable. This document is not for an AI-assisted device.

  6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done: Not applicable. This is not an AI algorithm.

  7. The type of ground truth used: Not applicable.

  8. The sample size for the training set: Not applicable.

  9. How the ground truth for the training set was established: Not applicable.

In summary, this document is a 510(k) submission for conventional dental implant components, not an AI-powered medical device. As such, it does not contain the information requested about acceptance criteria and studies proving the performance of an AI/algorithm-based device.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logos of the U.S. Department of Health & Human Services and the U.S. Food & Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the FDA acronym in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.

November 23, 2018

A.B Dental Device Ltd. % John J. Smith, M.D., J.D. Partner Hogan Lovells US LLP 555 13th St. NW Washington, District of Columbia 20004

Re: K181381

Trade/Device Name: A.B. Dental Devices® Dental Implants System Regulation Number: 21 CFR 872.3630 Regulation Name: Endosseous Dental Implant Abutment Regulatory Class: Class II Product Code: NHA Dated: October 25, 2018 Received: October 25, 2018

Dear John J. Smith, M.D., J.D.:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Andrew I. Steen -S

for Tina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use

Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement on last page

510(k) Number (if known)

K181381

Device Name

A.B. Dental Devices® Dental Implants System

Indications for Use (Describe)

A.B. DENTAL DEVICES® Dental Implants System is indicated for use in surgical and restorative applications for placement in the bone of the upper or lower jaw to provide support for prosthetic devices, such as artificial teeth, in order to restore the patient's chewing function. A.B. DENTAL DEVICES® Dental Implants System is indicated also for immediate loading when good primary stability is achieved and with appropriate occlusal loading.

Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D)□ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW *

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

FORM FDA 3881 (7/17)

Page 1 of 1 FDA

PSC Publishing Services (301) 443-6740 EF

{3}------------------------------------------------

K181381 510(k) SUMMARY A.B. DENTAL DEVICES® Dental Implants System

Sponsor:

A.B. Dental Devices Ltd. 19 Hayalomim St., Ashdod 7761117 Israel

Contact Person:

Gabi Krauss RA Manager Phone: 972-8-8531388 Facsimile: 972-8-8522562 Email: gabi@ab-dent.com

Date Prepared: November 23, 2018

Name of Device:A.B. Dental Devices® Dental Implants System
Common or Usual Name:Abutment, implant, dental, endosseous
Classification Name:Endosseous dental implant abutment; 21 CFR §872.3630.
Product Code:NHA

Predicate Devices

K162482 A.B. DENTAL DEVICES® Dental Implants System (primary) K132125 A.B. DENTAL DEVICES® Dental Implants System (reference) K112440, DENTAL DEVICES® Dental Implants System (reference)

Indications for Use

A.B. DENTAL DEVICES® Dental Implants System is indicated for use in surgical and restorative applications for placement in the bone of the upper or lower jaw to provide support for prosthetic devices, such as artificial teeth, in order to restore the patient's chewing function. A.B. DENTAL DEVICES® Dental Implants System is indicated also for immediate loading when good primary stability is achieved and with appropriate occlusal loading.

Device Description and Technological Characteristics

A.B. DENTAL DEVICES® Dental Implants System consists of one and two stage endosseous form dental implants, internal hexagonal and one piece implants systems. Narrow platform implants are to be used with straight abutments only. The following components are being added:

  • . P0-3 narrow platform healing caps, 4 and 7 mm length
  • PON-3.75 standard platform, narrow healing caps, 4 and 7 mm length .

{4}------------------------------------------------

  • P3-5 standard platform, straight abutment, 5 mm diameter, 5 mm length ●
  • P3S-3 narrow platform, anatomic, straight abutment, 1, 2, 3 mm shoulder lengths
  • . P3SW-3.75 standard platform, anatomic, straight abutment, wide diameter, 1, 2, 3 mm shoulder lengths
  • P3W-3 narrow platform, straight abutment, wide diameter, 9 mm length
  • P5-3 narrow platform, ball attachment, 4, 5, 6 mm height
  • P5-3.75-20 standard platform, 20º ball attachment, 4 and 6 mm height ●
  • P9HG-3.75 standard platform, composed abutment, gold alloy
  • P9R and P9HR -3.75 standard platform, composed abutment, CoCr alloy
  • . P25-3 narrow platform, A.B. LOC abutments, 0, 1, 2, 3, 4 5 mm height
  • . P25-3.75 standard platform. A.B. LOC abutments. 0. 1. 2 mm height
  • . P64-3.75 standard platform, straight adaptor abutment, 1, 2, 3, 4, 5 mm heights
  • P64 plastic and Ti alloy sleeve
  • P64 healing cap ●
  • PK standard platform, anatomic anti-rotation abutment, 1, 2, 3, 4 mm height
  • PK plastic sleeve
  • . PK healing cap

Technological Characteristics Comparison

The purpose of this 510(k) is to expand the current product line to include two new abutments listed above. No substantive changes are being made to the indications as compared to the most recent clearance, and the indications are provided below. Aside for K112440, the indications differ only in respect to indications related to the specific components contained within each submission. K112440 has the same intended use as it allows for implantation in the same anatomic location and, when appropriate, for immediate loading. Comparison tables for each modification are provided below.

K181381K162482K132125K112440
A.B. DENTAL DEVICES®A.B. DENTAL DEVICES®A.B. DENTAL DEVICES®The AB Dental Devices
Dental Implants System isDental Implants System isDental Implants System isimplants are intended for
indicated for use inindicated for use inindicated for use insurgical' placement in the
surgical and restorativesurgical and restorativesurgical and restorativemaxillary mandibular arch
applications for placementapplications for placementapplications for placementto support crowns, bridges,
in the bone of the upper orin the bone of the upper orin the bone of the upper oror overdentures in
lower jaw to providelower jaw to providelower jaw to provideedentulous or partially
support for prostheticsupport for prostheticsupport for prostheticand/or edentulous patients.
devices, such as artificialdevices, such as artificialdevices, such as artificialI7 Integral implant, I5
teeth, in order to restoreteeth, in order to restoreteeth, in order to restoreConical implant, P12-T,L
the patient's chewingthe patient's chewingthe patient's chewingTemporary flat connection
function. A.B. DENTALfunction. A.B. DENTALfunction. A.B. DENTALabutment, connection
DEVICES® DentalDEVICES® DentalDEVICES® Dentalabutment, and P16
Implants System isImplants System isImplants System isStraight adaptor are
indicated also forindicated also forindicated also forappropriate for immediate
immediate loading whenimmediate loading whenimmediate loading whenloading when good primary
good primary stability isgood primary stability isgood primary stability isstability is achieved and
achieved and withachieved and withachieved and withwith appropriate occlusal
appropriate occlusalappropriate occlusalappropriate occlusalloading.
loading.loading.loading.
Two Stage Implants: I22,Two Stage Implants:
I5, I55, I10.I2,I5,I6BI.
One Stage: I6, I6b, I6B.
P4 and P14 angledOne Stage & One-Piece
abutments are to be used3.0 mm diameter implants:
only with standard platformI6, I6B, I6BI, are intended
implants 3.5 mm infor placement at the
diameter or larger.mandibular central and

{5}------------------------------------------------

lateral incisors andmaxillary and lateralincisors. Indicated also fordenture stabilization usingmultiple implants.One stage & One-Piece2.4 mm diameter implantsfor temporary use or longterm use: I6, I6b, permitimmediate splint stabilityand long term fixation ofnew or existing crown,bridge and prosthesis.P14 Angulated AbutmentAdapter is to be used withimplant diameter 4.2mmand higher.
---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------

1. Healing Caps (P0)

Subject P0-3,(4,7)Predicate P0-3Predicate P0-3.75
K-numberK181381K132125K132125
MaterialTi-6Al-4V ELI with goldcolored anodizationTi-6Al-4V ELI with goldcolored anodizationTi-6Al-4V ELI
Diameter, mm4.04.04.5
PlatformNarrowNarrowStandard
Length, mm4, 72, 3, 5, 60.5 - 7
DesignStandardStandardStandard
AllowedangulationNo correction allowedNo correction allowedNo correction allowed
SterilityNon-SterileNon-SterileNon-Sterile
Subject P0N-3.75,(4,7)Predicate P0N-3.75Predicate P0-3.75
K-numberK181381K132125K132125
MaterialTi-6Al-4V ELITi-6Al-4V ELITi-6Al-4V ELI
Diameter, mm3.753.754.5
PlatformStandardStandardStandard
Length, mm4, 73, 5, 60.5 - 7
DesignStandardStandardStandard
Allowed angulationNo correction allowedNo correction allowedNo correction allowed
SterilityNon-SterileNon-SterileNon-Sterile

2. Cement Retained, Straight Abutments (P3)

Subject P3-5,5Predicate P3-5,(7,9)
K-numberK181381K162482
MaterialTi-6Al-4V ELITi-6Al-4V ELI
Diameter, mm5.05.0
PlatformStandardStandard
Length, mm57, 9
DesignAnti -rotationAnti-rotation
Allowed angulationNo correction allowedNo correction allowed

{6}------------------------------------------------

SterilityNon-SterileNon-Sterile
-------------------------------------
Subject P3S-3Predicate P3S-3.75Predicate P3-3
K-numberK181381K112440K112440
MaterialTi-6Al-4V ELI with goldcolored anodizationTi-6Al-4V ELITi-6Al-4V ELI with goldcolored anodization
Diameter4.54.54.5
PlatformNarrowStandardNarrow
Shoulder Height1, 2, 31, 2, 3
DesignAnatomic Anti-rotationAnatomic Anti-rotationAnti-rotation
AllowedangulationNo correction allowedNo correction allowedNo correction allowed
SterilityNon-SterileNon-SterileNon-Sterile
Subject P3S-Peek-3Predicate P3S-Peek-3.75
K-numberK181381K132125
MaterialPEEKPEEK
Diameter4.54.5
PlatformNarrowStandard
Shoulder Height1, 2, 31, 2, 3
DesignAnatomic Anti -rotationAnatomic Anti -rotation
Allowed angulationNo correction allowedNo correction allowed
SterilityNon-SterileNon-Sterile
Subject P3SW-3.75Predicate P3S-3.75Predicate P3W-3.75
K-numberK181381K112440K112440
MaterialTi-6Al-4V ELITi-6Al-4V ELITi-6Al-4V ELI
Diameter5.53.755.5
PlatformStandardStandardStandard
Shoulder Height1, 2, 31, 2, 31, 2, 3
DesignAnatomic Anti-rotationWideAnatomic, Anti-rotationAnti-rotation
Allowed angulationNo correction allowedNo correction allowedNo correction allowed
SterilityNon-SterileNon-SterileNon-Sterile
Subject P3W-3Predicate P3-3Predicate P3W-3.75
K-numberK181381K112440K112440
MaterialTi-6Al-4V ELI with gold coloranodizationTi-6Al-4V ELI with gold coloranodizationTi-6Al-4V ELI
Diameter5.53.755.5
PlatformNarrowNarrowStandard
Length99, 129, 12
DesignAnti-rotationAnti-rotationAnti-rotation

{7}------------------------------------------------

AllowedangulationNo correction allowedNo correction allowedNo correctionallowed
SterilityNon-SterileNon-SterileNon-Sterile

3. Overdenture Ball Attachment Abutments (P5)

Subject P5-3,(4,5,6)Predicate P5-3,(1,2,3)Predicate P5-3.75
K-numberK181381K132125K132125
MaterialTi-6Al-4V ELI with gold color anodizationTi-6Al-4V ELI with gold color anodizationTi-6Al-4V ELI
PlatformNarrowNarrowStandard
Length, mm4, 5, 61, 2, 31, 2, 3, 4, 5, 6
DesignBall AttachmentBall AttachmentBall Attachment
Allowed angulationNo correction allowedNo correction allowedNo correction allowed
SterilityNon-SterileNon-SterileNon-Sterile
Subject P5-3.75,20-(4,6)Predicate P5-3.75,20-(1,2,3,5)
K-numberK181381K132125
MaterialTi-6Al-4V ELITi-6Al-4V ELI
PlatformStandardStandard
Length, mm4, 61, 2, 3, 5
Angle, °2020
DesignBall AttachmentBall Attachment
SterilityNon-SterileNon-Sterile

4. Composed (UCLA) Abutments

Subject P9R-3.75,11, P9HR-3.75,11Predicate P9-3.75,11, P9H-3.75,11
K-numberK181381K132125
MaterialFWM 1058 Cobalt Chrome Alloy per ASTM F1058Ti-6Al-4V ELI
Connection Method/PlatformStandardStandard
Diameter, mm3.753.75
Allowed angulationNo correction allowedNo correction allowed
Length, mm1111
DesignComposed AbutmentComposed Abutment
SterilityNon-SterileNon-Sterile
Subject P9HG-3.75,11Predicate P9G-3.75,11Predicate P9H-3.75,11
K-numberK181381K132125K132125
MaterialGold Alloy 6019Gold Alloy 6019Ti-6Al-4V
\PlatformStandardStandardStandard
Diameter, mm3.753.753.75
AllowedNo correction allowedNo correction allowedNo correction allowed

{8}------------------------------------------------

angulation
Length, mm111111
DesignComposed HexAbutmentComposed Free RotatingAbutmentComposed Hex Abutment
SterilityNon-SterileNon-SterileNon-Sterile

5. A.B. LOC Abutments (P25)

Subject P25-3.75,(0,1,2)Predicate P25-3.75,(3,4,5,6,7)
K-numberK181381K132125
MaterialTi-6Al-4V ELITi-6Al-4V ELI
PlatformStandardStandard
Diameter, mm3.753.75
Allowed angulationNo correction allowedNo correction allowed
Length, mm0, 1, 23, 4, 5, 6, 7
DesignIntegrated screw, attachmentIntegrated screw, attachment
SterilityNon-SterileNon-Sterile
Subject P25-3Predicate P25-3.75
K-numberK181381K132125
MaterialTi-6Al-4V ELI with gold coloranodizationTi-6Al-4V ELI
PlatformNarrowStandard
Diameter, mm33.75
Allowed angulationNo correction allowedNo correction allowed
Length, mm0, 1, 2, 3, 4, 53,4,5,6,7
DesignIntegrated screw, attachmentIntegrated screw, attachment
SterilityNon-SterileNon-Sterile

6. P64 Straight Adaptor

Subject P64-3.75Predicate P16-3.75
K-numberK181381K112440
MaterialTi-6Al-4V ELITi-6Al-4V ELI
PlatformStandardStandard
AllowedangulationNo correction allowedNo correction allowed
Height1, 2, 3, 4, 51, 2, 3, 4, 5
DesignIntegrated screw, screw retainedIntegrated screw, screw retained
SterilityNon-SterileNon-Sterile
Subject P64bPredicate P14b
K-numberK181381K112440
MaterialDelrinDelrin
ConnectionMethod/ PlatformTaper fit+ screw /P64 platformTaper fit+ screw /P16 platform
Allowed angulationNo correction allowedNo correction allowed
Length, mm1010

{9}------------------------------------------------

DesignPlastic Sleeve for AdaptorPlastic Sleeve for Adaptor
SterilityNon-SterileNon-Sterile
Subject P64-bTPredicate P14-bT
K-numberK181381K132125
MaterialTi-6Al-4V ELITi-6Al-4V ELI
ConnectionTaper fit+ screw /Taper fit+ screw /
Method/ PlatformP64 platformP16 platform
Allowed anqulationNo correction allowedNo correction allowed
Length, mm1212
DesignTitanium Sleeve for AdaptorTitanium Sleeve for Adaptor
SterilityNon-SterileNon-Sterile
Subject P0-P64,5Predicate P0-P14
K-numberK181381K132125
MaterialTi-6Al-4V ELITi-6Al-4V ELI
ConnectionP64P14/P16
Method/Platform
Allowed angulationNo correction allowedNo correction allowed
Height52.5, 4, 5, 7
DesignTemporary healing capTemporary healing cap
SterilityNon-SterileNon-Sterile

7. PK Abutment

Subject PK-P3-3.75Predicate P3-3.75Predicate P3S-3.75,(1,2,3)
K-numberK181381K112440K112440
MaterialTi-6Al-4V ELITi-6Al-4V ELITi-6Al-4V ELI
ConnectionMethod/PlatformStandardStandardStandard
AllowedangulationNo correction allowedNo correction allowedNo correction allowed
Height/Length,mm1, 2, 3, 45, 7, 91, 2, 3
DesignAnti-Rotation AbutmentAnti Rotation AbutmentAnatomic Anti-Rotation Abutment
SterilityNon-SterileNon-SterileNon-Sterile
Subject PK-P0-3.75Predicate P0-P14
K-numberK181381K132125
MaterialTi-6Al-4V ELITi-6Al-4V ELI
ConnectionPKP14/P16
Method/Platform
Allowed angulationNo correction allowedNo correction allowed
HeightN/A2.5, 4, 5, 7
DesignTemporary healing capTemporary healing cap
SterilityNon-SterileNon-Sterile

{10}------------------------------------------------

Subject PK-P2-3.75Predicate P14b
K-numberK181381K112440
MaterialDelrinDelrin
ConnectionTaper fit/Taper fit+ screw /
Method/ PlatformPK platformP16 platform
Allowed angulationNo correction allowedNo correction allowed
Length, mm1010
DesignPlastic Sleeve for StraightAdaptorPlastic Sleeve for Adaptor
SterilityNon-SterileNon-Sterile
Subject PK-P2H-3.75Predicate P14b
K-numberK181381K112440
MaterialDelrinDelrin
Connection Method/PlatformTaper fit/PK platformTaper fit+ screw /P16 platform
Allowed angulationNo correction allowedNo correction allowed
Length, mm1010
DesignPlastic Sleeve with Hexfor Straight AdaptorPlastic Sleeve forAdaptor
SterilityNon-SterileNon-Sterile

Performance Data

  • Engineering analysis to determine if additional components constitute a new worst-case. ●
  • . Biocompatibility testing per ISO 10993-1 per ISO 10993 as originally presented in K162482
  • Cytotoxicity testing per ISO 10993-5 .
  • Sterilization validation as per ANSI/AAMI/ISO 17665-1: 2006 as originally presented in . K162482
  • Fatigue testing as per ISO 14801 as originally presented in K162482

Substantial Equivalence

The A.B. Dental Devices® Dental Implants System has the same indications and similar technological characteristics and principles of operation as its predicate device. Performance data demonstrate that the A.B. Dental Devices® Dental Implants System is substantially equivalent.

Conclusions

The A.B. Dental Devices® Dental Implants System is substantially equivalent to the predicate device.

§ 872.3630 Endosseous dental implant abutment.

(a)
Identification. An endosseous dental implant abutment is a premanufactured prosthetic component directly connected to the endosseous dental implant and is intended for use as an aid in prosthetic rehabilitation.(b)
Classification. Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)