K Number
K200386
Device Name
Z5-BL
Manufacturer
Date Cleared
2020-10-14

(239 days)

Product Code
Regulation Number
872.3640
Panel
Dental
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Z5-BL implants are designed for surgical implantation into the upper and lower jaw for the attachment of prosthodontic appliances to replace missing teeth. Z5-BL implants are suitable for patients with metal allergies and the chronic diseases resulting from them. Z5-BL implants are intended for delayed loading.

Device Description

Z5-BL is a dental implant system that includes root-form, endosseous threaded dental implants indicated for bone level placement. The subject device includes dental implants provided in one body diameter of 3.6 mm in lengths of 8 mm, 10 mm. The subject device also includes a gingiva former, straight abutments for single-unit cemented restorations, and a 15° angled abutment also for single-unit cemented restorations.

The subject device is compatible with components cleared previously in K190243, including a healing cap, straight abutments for multi-unit cemented restorations, Locator-type abutments for retention of overdentures, and occlusal (abutment) screws.

The subject device dental implants are manufactured from Y-TZP zirconia conforming to ISO 13356. The subject device abutments are manufactured from Y-TZP zirconia conforming to ISO 13356, or from polyetheretherketone (PEEK).

The subject device implants are provided sterilized by plasma gas. All other components are to be sterilized by the end-user by moist heat (steam).

AI/ML Overview

The provided text describes a 510(k) premarket notification for a dental implant system (Z5-BL). The submission aims to demonstrate substantial equivalence to previously cleared predicate devices. The information primarily focuses on non-clinical performance data and a comparison of technological characteristics, rather than reporting on a study demonstrating the device meets specific acceptance criteria in terms of clinical performance or human-AI effectiveness.

Therefore, many of the requested details, such as acceptance criteria for clinical performance, sample sizes for test and training sets, expert qualifications for ground truth, adjudication methods, multi-reader multi-case studies, and standalone algorithm performance, are not applicable or not present in the provided document.

Here's a breakdown of the available information based on your request:

1. A table of acceptance criteria and the reported device performance:

The document does not specify quantitative clinical acceptance criteria for the Z5-BL device. Instead, it relies on demonstrating substantial equivalence to predicate devices through non-clinical performance data and technological characteristics.

The "performance data" section lists the types of non-clinical tests performed and their respective standards. The implied acceptance criterion for these tests is conformance to the referenced international or national standards, which is then used to support the claim of substantial equivalence.

Acceptance Criteria (Implied Non-Clinical)Reported Device Performance (Summary)
Conformance to ISO 14937, ISO 17665-1Sterilization validated
Conformance to ISO 11607-1, ISO 11607-2, ASTM F1886/F1886M, ASTM D3078Sterile product shelf life tested
Conformance to USP 40-NF 35 <85> (LAL test)Bacterial endotoxin testing performed
Conformance to ISO 13356Biocompatibility of Y-TZP zirconia confirmed (referenced)
Conformance to ISO 10993-5, ISO 10993-12, ISO 10993-10, ISO 10993-11, ISO 10993-18, USP Class VIBiocompatibility of PEEK material confirmed (referenced)
Conformance to ISO 14801Static compression and compression fatigue testing performed; fatigue limit data demonstrated sufficient strength
Conformance to ISO 18130Assessment of abutment screw loosening and removal torque testing performed
Examination by light microscopy and SEM/EDXAssessment of wear particles from Ti-6A1-4V alloy abutment screw performed

2. Sample size used for the test set and the data provenance:

  • Not applicable. This submission did not include clinical data or a "test set" in the context of clinical performance evaluation or AI model validation. The performance data listed are non-clinical, mechanical, and material tests.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

  • Not applicable. No clinical test set or ground truth established by experts is mentioned.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

  • Not applicable. No clinical test set.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

  • Not applicable. This document describes a dental implant and does not involve AI or human reader improvement studies.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

  • Not applicable. This document describes a dental implant and does not involve an algorithm.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

  • For the non-clinical performance data, the "ground truth" is established by adherence to the respective international and national standards (e.g., ISO, ASTM, USP) for material properties, sterility, biocompatibility, and mechanical performance. These standards define the accepted methods and criteria for evaluating device properties.

8. The sample size for the training set:

  • Not applicable. No training set for an AI model is mentioned.

9. How the ground truth for the training set was established:

  • Not applicable. No training set for an AI model is mentioned.

In summary, this 510(k) submission for the Z5-BL dental implant relies on demonstrating substantial equivalence through comprehensive non-clinical testing according to established standards. It explicitly states, "No clinical data were included in this submission" (page 5). Therefore, the requested information pertaining to clinical studies, AI performance, ground truth, and expert evaluation for a test or training set is not present.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA acronym along with the full name of the agency on the right. The FDA part of the logo is in blue, with the acronym in a bold, sans-serif font and the full name in a smaller, sans-serif font below it. The overall design is clean and professional, reflecting the agency's role in regulating food and drugs in the United States.

October 14, 2020

Z-Systems AG % Kevin Thomas Vice President and Director of Regulatory Affairs PaxMed International, LLC 12264 El Camino Real, Suite 400 San Diego, California 92130

Re: K200386

Trade/Device Name: Z5-BL Regulation Number: 21 CFR 872.3640 Regulation Name: Endosseous dental implant Regulatory Class: Class II Product Code: DZE, NHA Dated: September 11, 2020 Received: September 14, 2020

Dear Kevin Thomas:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

for

Srinivas Nandkumar, Ph.D. Director DHT1B: Division of Dental Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known)

K200386

Device Name

Z5-BL

Indications for Use (Describe)

Z5-BL implants are designed for surgical implantation into the upper and lower jaw for the attachment of prosthodontic appliances to replace missing teeth. Z5-BL implants are suitable for patients with metal allergies and the chronic diseases resulting from them. Z5-BL implants are intended for delayed loading.

Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C) Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C)
Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services

Food and Drug Administration

Office of Chief Information Officer

Paperwork Reduction Act (PRA) Staff

PRAStaff @fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

FORM FDA 3881 (7/17)

PSC Publishing Services (301) 443-6740 EF

{3}------------------------------------------------

510(k) Summary

K200386

Z5-BL

Z-Systems AG

September 11, 2020

kthomas@paxmed.com flarson@paxmed.com

ADMINISTRATIVE INFORMATION

Manufacturer NameZ-Systems AGWerkhofstrasse 5CH-4702 OensingenSwitzerlandTelephone: +41 62 388 69 69
Official ContactRubino DiGirolamo, CEO

Representative/Consultant Kevin A. Thomas, PhD Floyd G. Larson, MS, MBA PaxMed International, LLC 12264 El Camino Real, Suite 400 San Diego, CA 92130 Telephone: +1 858-792-1235 Fax: +1 858-792-1236

Email:

DEVICE NAME AND CLASSIFICATION

Trade/Proprietary NameZ5-BL
Common NameDental implant
Regulation Number21 CFR 872.3640
Regulation NameEndosseous dental implant
Regulatory ClassClass II
Product CodeDZE
Secondary Product CodeNHA
Classification PanelDental
Reviewing OfficeOffice of Health Technology 1 (Ophthalmic, Anesthesia, Respiratory,ENT and Dental Devices)
Reviewing DivisionDivision of Health Technology 1 B (Dental Devices)

PREDICATE DEVICE INFORMATION

Primary Predicate Device: K190243, Z5-BL, Z-Systems AG Reference Devices: K132881, Z5c, Z-Systems AG K131701, Z5mlb and Z5mlc, Z-Systems AG K120793, Z-Look3 Evo SLM, Z-Systems AG

{4}------------------------------------------------

INDICATIONS FOR USE STATEMENT

Z5-BL implants are designed for surgical implantation into the upper and lower jaw for the attachment of prosthodontic appliances to replace missing teeth. Z5-BL implants are suitable for patients with metal allergies and the chronic diseases resulting from them. Z5-BL implants are intended for delayed loading.

SUBJECT DEVICE DESCRIPTION

Z5-BL is a dental implant system that includes root-form, endosseous threaded dental implants indicated for bone level placement. The subject device includes dental implants provided in one body diameter of 3.6 mm in lengths of 8 mm, 10 mm. The subject device also includes a gingiva former, straight abutments for single-unit cemented restorations, and a 15° angled abutment also for single-unit cemented restorations.

The subject device is compatible with components cleared previously in K190243, including a healing cap, straight abutments for multi-unit cemented restorations, Locator-type abutments for retention of overdentures, and occlusal (abutment) screws.

The subject device dental implants are manufactured from Y-TZP zirconia conforming to ISO 13356. The subject device abutments are manufactured from Y-TZP zirconia conforming to ISO 13356, or from polyetheretherketone (PEEK).

The subject device implants are provided sterilized by plasma gas. All other components are to be sterilized by the end-user by moist heat (steam).

PERFORMANCE DATA

Non-clinical data submitted to demonstrate substantial equivalence included:

sterilization validation according to ISO 14937 (referenced from K132881) and ISO 17665-1 (referenced from K120793);

sterile product shelf life testing according to ISO 11607-1, ISO 11607-2, ASTM F1886/F1886M, and ASTM D3078 (referenced from K132881);

bacterial endotoxin testing for subject devices provided sterile (Limulus Amebocyte Lysate (LAL) test according to USP 40-NF 35 <85>, referenced from K190243);

biocompatibility of the yttria-stabilized zirconia (Y-TZP) conforming to ISO 13356 used to manufacture the subject device dental implants and abutments (referenced from identical material used for dental implants and abutments cleared in K190243);

biocompatibility endosseous threaded surface of the subject device implants (referenced from identical Al2O3 grit blasted and laser modified surface applied to the dental implants cleared in K190243);

biocompatibility of the PEEK material according to ISO 10993-5, ISO 10993-12, ISO 10993-10, ISO 19003-11. ISO 10993-18, and USP Class VI testing according to USP 37-NF32 <88> (referenced from K190243);

static compression and compression fatigue testing according to ISO 14801 of the subject device zirconia implants and zirconia abutments, with both zirconia and titanium alloy abutment screws;

assessment of abutment screw loosening and abutment screw removal torque testing (according to ISO 18130) after insertion to the torque recommended in the subject device labeling, and after dynamic testing

{5}------------------------------------------------

according to ISO 14801, including microscopic examination of the abutment screws and internal threads of the implant bodies (referenced from K190243);

assessment of wear particles associated with the Ti-6A1-4V alloy abutment screw used with zirconia implant bodies and zirconia abutments after dynamic testing according to ISO 14801, including examination by light microscopy and scanning electron microscopy with energy dispersive X-ray analysis (SEM / EDX) (referenced from K190243).

No clinical data were included in this submission.

EQUIVALENCE TO MARKETED DEVICES

The subject device is substantially equivalent in indications and design principles to the primary predicate device and the reference devices listed above. Provided at the end of this summary is a table comparing the Indications for Use Statements and the technological characteristics of the subject device, the primary predicate device, and the reference devices.

The subject device is substantially equivalent in intended use to the primary predicate K190243, and to the reference devices K132881 and K131701. All are intended for use in the maxilla and mandible to provide functional and esthetic rehabilitation of the edentulous maxilla and mandible. The subject device and the primary predicate device K190243 are made of identical materials (implants and abutments). The subject device dental implants and the primary predicate device K190243 dental implants have an identical endosseous threaded surface (grit-blasted and laser modified), have the same internal implant connection features, and the same implant lengths. The subject device and the primary predicate device K190243 also include straight and angled abutments for cement-retained, single-unit restorations. Differences between subject device and the primary predicate device K190243 include the implant diameter (subject device 3.6 mm) and implant platform diameter (subject device also 3.6 mm). None of these minor design differences impact safety or effectiveness or change the intended use of the device.

The reference device K132881 is for support of substantial equivalence of the subject device in terms of sterilization and shelf life (also referenced in K190243). The subject device dental implants are provided to the end user sterilized by plasma gas. The subject device components that are provided sterile use the same packaging and have the same sterile barrier shelf life as the sterile components cleared in K132881 (referenced in K190243). The processes to monitor, control, and test for bacterial endotoxins for the subject device are the same as K190243.

The reference device K131701 is for support of substantial equivalence of the subject device in terms of the smaller dental implant body diameter and corresponding implant platform diameter (both 3.6 mm). The subject device dental implants and the reference device K131701 dental implants also are made of identical materials, and have the identical endosseous threaded surface (grit-blaster and laser modified).

Mechanical performance testing of the subject device was performed according to ISO 14801 Dentistry -Implants – Dynamic fatigue test for endosseous dental implants. The fatigue limit data demonstrated that constructs of the device have sufficient strength for their intended use.

Minor differences in the designs, dimensions, sizes, or materials among the subject device, the primary predicate device, and the reference devices do not affect substantial equivalence. These minor differences do not impact safety or effectiveness because these differences are related to the specific designs features and system components, and are mitigated by the mechanical performance testing.

{6}------------------------------------------------

CONCLUSION

The subject device, the primary predicate device, and the reference devices have the same intended use, have similar technological characteristics, and are made of identical materials. The subject device, the primary predicate, and reference devices encompass the same range of physical dimensions, are packaged in similar materials, and are sterilized using similar methods. The data included in this submission demonstrate substantial equivalence to the predicate devices listed above.

{7}------------------------------------------------

Table of Substantial Equivalence

ComparisonSubject DevicePrimary PredicateReference DeviceReference Device
Indications for Use StatementK200386Z5-BLZ-Systems AGZ5-BL implants are designed for surgical implantationinto the upper and lower jaw for the attachment ofprosthodontic appliances to replace missing teeth.Z5-BL implants are suitable for patients with metalallergies and the chronic diseases resulting from them.Z5-BL implants are intended for delayed loading.K190243Z5-BLZ-Systems AGZ5-BL implants are designed for surgical implantationinto the upper and lower jaw for the attachment ofprosthodontic appliances to replace missing teeth.Z5-BL implants are suitable for patients with metalallergies and the chronic diseases resulting from them.Z5-BL implants are intended for delayed loading.K132881Z5cZ-Systems AGZ5c implants are designed for surgical implantationinto the upper and lower jaw for the attachment ofprosthodontic appliances to replace missing teeth.Z5c implant system is also suitable for patients withmetal allergies and the chronic diseases resulting fromthem. Z5c implants are intended for delayed loading.K131701Z5mlb and Z5mlcZ-Systems AGZ5mlb:Z5mlb implants arc designed for surgical implantationinto the edentulous upper and lower jaw for theattachment of dentures to replace missing teeth. Z5mlbimplant system is also suitable for patients with metalallergies and the chronic diseases resulting from them.Z5mlc:Z5mlc implants are designed for surgical implantationinto the edentulous upper and lower jaw for theattachment of dentures to replace missing teeth. Z5mlcimplant system is also suitable for patients with metalallergies and the chronic diseases resulting from them.
Intended UseFunctional and esthetic rehabilitation of the edentulousmaxilla and mandibleFunctional and esthetic rehabilitation of the edentulousmaxilla and mandibleFunctional and esthetic rehabilitation of the edentulousmaxilla and mandibleFunctional and esthetic rehabilitation of the edentulousmaxilla and mandible
Reason for Predicate/Reference DeviceNot applicableIFU Statement; implant design, material, surface;abutment designReference for sterilization and shelf lifeReference for implant diameter (3.6 mm)
Product CodesDZE, NHADZE, NHADZE, NHADZE
ComponentsImplants, abutmentsImplants, abutments,abutment screwsImplants, abutments(abutment cemented to implant)One-piece implants
Implant Designs
Implant Diameter, mm3.64.0, 5.04.0, 5.03.6, 4.0
Implant Endosseous Length, mm8, 10, 128, 10, 128, 10, 128, 10
Platform diameter, mm3.64.0, 5.04.0, 5.0Not applicable
Implant MaterialY-TZPY-TZPY-TZPY-TZP
Implant SurfaceGrit-blasted and laser modifiedGrit-blasted and laser modifiedGrit-blasted and laser modifiedGrit-blasted and laser modified
Abutment Designs
FeaturesSubject deviceIndexed, straight and angledPreviously cleared compatibleNon-indexed straight;Indexed and Non-indexed Locator-typeIndexed, straight and angled;Non-indexed straight;Indexed and Non-indexed Locator-typeNon-indexed straightNon-indexed angled 15°Ball-type attachment;Locator-type attachment
Prosthesis AttachmentSubject deviceCement-retainedPreviously cleared compatibleCement-retained;Locator-type attachmentCement-retained;Locator-type attachmentCement-retainedBall-type attachment;Locator-type attachment
RestorationSingle-unit; Multi-unitSingle-unit; Multi-unitSingle-unit; Multi-unitMulti-unit
ComparisonSubject DevicePrimary PredicateReference DeviceReference Device
K200386Z5-BLZ-Systems AGK190243Z5-BLZ-Systems AGK132881Z5cZ-Systems AGK131701Z5mlb and Z5mlcZ-Systems AG
Prosthetic platform diameter, mmSubject device3.8 mm – 4.3 mmPreviously cleared compatible4.5 mm4.5 mm – 6.5 mm3.6 mmNot applicable
Abutment angleStraight (0°), 15°Straight (0°), 15°Straight (0°), 15°Straight (0°)
Abutment materialY-TZPY-TZPY-TZPY-TZP
Abutment Screw MaterialPreviously cleared compatibleY-TZPTi-6Al-4V alloyY-TZPTi-6Al-4V alloyNot applicableNot applicable
Sterilization Status/Method
ImplantsSterile / plasma gasSterile / plasma gasSterile / plasma gasSterile / plasma gas
AbutmentsNon-sterile / moist heat (steam)Healing Cap: Sterile / plasma gasAll other:Non-sterile / moist heat (steam)Non-sterile / Dry heatNot applicable
Other ComponentsNot applicableNon-sterile / moist heat (steam)Non-sterile / moist heat (steam)Not applicable

{8}------------------------------------------------

510(k) Summary
Page 6 of 6

§ 872.3640 Endosseous dental implant.

(a)
Identification. An endosseous dental implant is a prescription device made of a material such as titanium or titanium alloy that is intended to be surgically placed in the bone of the upper or lower jaw arches to provide support for prosthetic devices, such as artificial teeth, in order to restore a patient's chewing function.(b)
Classification. (1) Class II (special controls). The device is classified as class II if it is a root-form endosseous dental implant. The root-form endosseous dental implant is characterized by four geometrically distinct types: Basket, screw, solid cylinder, and hollow cylinder. The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)(2)
Classification. Class II (special controls). The device is classified as class II if it is a blade-form endosseous dental implant. The special controls for this device are:(i) The design characteristics of the device must ensure that the geometry and material composition are consistent with the intended use;
(ii) Mechanical performance (fatigue) testing under simulated physiological conditions to demonstrate maximum load (endurance limit) when the device is subjected to compressive and shear loads;
(iii) Corrosion testing under simulated physiological conditions to demonstrate corrosion potential of each metal or alloy, couple potential for an assembled dissimilar metal implant system, and corrosion rate for an assembled dissimilar metal implant system;
(iv) The device must be demonstrated to be biocompatible;
(v) Sterility testing must demonstrate the sterility of the device;
(vi) Performance testing to evaluate the compatibility of the device in a magnetic resonance (MR) environment;
(vii) Labeling must include a clear description of the technological features, how the device should be used in patients, detailed surgical protocol and restoration procedures, relevant precautions and warnings based on the clinical use of the device, and qualifications and training requirements for device users including technicians and clinicians;
(viii) Patient labeling must contain a description of how the device works, how the device is placed, how the patient needs to care for the implant, possible adverse events and how to report any complications; and
(ix) Documented clinical experience must demonstrate safe and effective use and capture any adverse events observed during clinical use.