(270 days)
HALYARD® SMART-FOLD Sterilization Wraps are intended to be used to enclose another medical device that is to be sterilized by a healthcare provider using:
- Pre-vacuum steam at 270°F/132°C for 4 minutes. The wrap was validated for dry times of 30 minutes.
- 100% ethylene oxide (EO) with a concentration of 725-735 mg/L at 131°F/55°C and 40%-80% relative humidity for 60 minutes. The wrap was validated for aeration times for EO sterilization of 8 hours at 110° F/43.3ºC.
STERIS V-PRO® Low Temperature Sterilization Systems.
-
· STERIS® V-PRO® 60 (Lumen, Non-Lumen and Flexible Cycles)
-
· STERIS® V-PRO® s2 (Lumen, Non-Lumen and Flexible Cycles)
-
· STERIS® V-PRO® 1 (Lumen Cycle)
-
· STERIS® V-PRO® 1 Plus (Lumen and Non-Lumen Cycle)
-
· STERIS® V-PRO® maX (Lumen, Non-Lumen and Flexible Cycle)
-
· STERIS® V-PRO® maX 2 (Lumen, Non-Lumen and Flexible Cycle)
-
Advanced Sterilization Products STERRAD® Sterilization System
-
STERRAD® 100S
-
· STERRAD® NX®, (Standard Cycle, Advanced Cycle)
-
· STERRAD® NX® with ALLClear® Technology, (Standard Cycle, Advanced Cycle)
-
· STERRAD® 100NX® (Standard Cycle, Flex Cycle, EXPRESS Cycle, DUO Cycle, ULTRA GL Cycle)
-
STERRAD® 100NX® with ALLClear® Technology, (Standard Cycle, EXPRESS Cycle, DUO Cycle, ULTRA GI Cycle)
HALYARD* ONE-STEP, QUICK CHECK and SEQUENTIAL Sterilization Wraps are intended to be used to enclose another medical device that is to be sterilized by a healthcare provider using:
· Pre-vacuum steam at 270°F/132°C for 4 minutes. The wrap was validated for dry times of 20 minutes for Models 100 and 200, and for 30 minutes for Models 300, 400, 500, and 600.
-
· 100% ethylene oxide (EO) with a concentration of 725-735 mg/L at 131°F/55°C and 40%-80% relative humidity for 60 minutes. The wrap was validated for aeration times for EO sterilization of 8 hours at 43.3℃.
· STERIS V-PRO® Low Temperature Sterilization Systems. The wrap was validated to be effectively aerated during the pre-programmed cycles. -
· STERIS V-PRO® 60 (Lumen, Non-Lumen and Flexible Cycles)
-
· STERIS V-PRO® s2 (Lumen, Non-Lumen and Flexible Cycles)
-
· STERIS V-PRO® 1 (Lumen Cycle)
-
· STERIS V-PRO® 1 Plus (Lumen and Non-Lumen Cycle)
-
· STERIS V-PRO® maX (Lumen, Non-Lumen and Flexible Cycle)
-
· STERIS V-PRO® maX 2 (Lumen, Non-Lumen and Flexible Cycle)
· Gravity steam at 250°F/121°C for 30 minutes (25 minute dry time for Models 100, 200 and 30 minute dry time for Models 400, 500 and 600)
- · Advanced Sterilization Products STERRAD® Sterilization System
- · STERRAD® 50, 100S, and 200
- · STERRAD® NX®, (Standard Cycle, Advanced Cycle)
- · STERRAD® NX® with ALLClear® Technology, (Standard Cycle, Advanced Cycle)
- · STERRAD® 100NX® (Standard Cycle, Flex Cycle, EXPRESS Cycle, DUO Cycle)
- · STERRAD® 100NX® with ALLClear® Technology, (Standard Cycle, EXPRESS Cycle, DUO Cycle)
- · STERILUCENT® HC80TT Sterilization Cycles (Lumen and Flexible Cycles)
- · Stryker Sterizone® VP4 Sterilizer Cycle 1
HALYARD SEQUENTIAL Sterilization Wrap is supplied to the customer as bulk packages of single sheets, where in accordance with standard hospital practices, two sheets are then used to wrap a medical device or a collection of medical devices for sterilization. HALYARD ONE-STEP, HALYARD QUICK CHECK and HALYARD SMART-FOLD Sterilization Wraps are comprised of two sheets of HALYARD* SEQUENTIAL Sterilization Wrap. This allows for convenient wrapping with two sheets simultaneously.
The HALYARD ONE-STEP, QUICK CHECK and HALYARD SMART-FOLD Sterilization Wraps are intended to enclose another medical device that is to be sterilized by a healthcare provider.
The document describes the regulatory submission for HALYARD sterilization wraps. It focuses on demonstrating substantial equivalence to a predicate device through performance testing. No AI/ML algorithm is involved in the device. Therefore, a complete response is not possible for all aspects of your request.
Here's a breakdown of the available information:
1. A table of acceptance criteria and the reported device performance
Purpose | Test | Acceptance Criteria | Results |
---|---|---|---|
Sterilant Penetration/Efficacy | ANSI/AAMI ST79, ANSI/AAMI/ISO 11138-7 | Achieving a 10^-6 sterility assurance level following processing in a worst-case half-cycle | Passed |
Performance Testing (Non-sterile/Sterile) | ANSI/AAMI/ISO 11607-1 Annex B, ISO 13938-2, ASTM D4966-12, CPSC 1610 | Complies with the selected physical properties | Passed |
Maintenance of Package Integrity | ANSI/AAMI/ISO 11607-1, ANSI/AAMI ST79 | Maintain sterility for 1 year, except for gravity steam (30 days), STERRAD ULTRA GI (3 months) and Sterilucent (6 months). | Passed |
Biocompatibility | ISO 10993-1, ISO 10993-5, ISO 10993-10, ISO 10993-7 | Non-cytotoxic, Non-irritating, Non-sensitizing | Passed |
Residuals | Steris V-PRO Low Temperature Sterilization System H2O2 residuals, Sterizone VP4 Cycle 1 residual H2O2, ISO 14937 (ULTRA GI Cycle only), ISO 10993-7 EtO and ECH | H2O2 ≤ 0.000218 mg/cm2, H2O2 ≤ 0.56 µg/cm2, H2O2 ≤ 9100 µg/cm2, EO |