K Number
K141071
Date Cleared
2014-07-28

(94 days)

Product Code
Regulation Number
880.6850
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

KIMGUARD* ONE-STEP* Sterilization Wrap (KC100, KC200, KC300, KC400, KC500 and KC600) are intended to enclose another medical device that is to be sterilized by a healthcare provider using:

  • Gravity Steam at 250°F/121°C for 30 minutes. .
    KIMGUARD* ONE-STEP* Sterilization Wrap (KC100, KC200, KC300, KC400, KC500 and KC600) are intended to allow sterilization of the enclosed medical device(s) and also maintain sterility of the enclosed device(s) until used.
Device Description

KIMGUARD* ONE-STEP* Sterilization Wrap is comprised of two sheets of KIMGUARD* Sterilization Wrap that is ultrasonically seamed on two edges. This seamed configuration allows for convenient wrapping of an article using two sheets simultaneously.

The sheets of sterilization wrap are square or rectangular fabric produced using a threelayer SMS (spunbound-meltblown-spunbound) process. The wrap fabric is composed of polypropylene with the addition of less than 2% by weight of phthalocyanine blue pigment, less than 1% by weight titanium dioxide pigment, and less than 0.008% by weight of antistatic treatment. The wrap allows a sterilized package to be opened aseptically.

AI/ML Overview

Here's an analysis of the provided text regarding the acceptance criteria and study for the KIMGUARD ONE-STEP Sterilization Wrap:

Acceptance Criteria and Device Performance for KIMGUARD® ONE-STEP® Sterilization Wrap

1. Table of Acceptance Criteria and Reported Device Performance

Acceptance Criteria CategoryReported Device Performance
Sterilization Efficacy (Gravity Steam)Passed
Sterilant PenetrationPassed
Package Integrity/Physical PropertiesPassed
Drying and AerationPassed
Maintenance of Package IntegrityPassed
Biocompatibility (pre- and post-sterilization)Passed
Maintenance of Package Sterility (Duration)30 days

Note: The document states that "All results of testing met the same acceptance criteria as the predicate devices demonstrating substantial equivalence to the predicate devices." While specific, numerical acceptance criteria are not explicitly detailed (e.g., minimum tensile strength), the "Passed" designation for each category indicates successful fulfillment of the established criteria, which are identical to those of the predicate devices (K082177 and K091685). The specific loads and weights for each wrap model also represent acceptance criteria for their respective uses.


2. Sample Size Used for the Test Set and Data Provenance

The document does not explicitly state the specific numerical sample size used for the test sets (e.g., how many wraps were tested for sterilant penetration). However, it lists the types of loads used in the "Sterility Maintenance Validation Study" for each wrap model:

  • KC100: 1 tray liner 20" x 25", 12.5" x 9" x 1" Tray, 1 lb of metal mass
  • KC200: 1 tray liner 20" x 25", 10" x 20" x 3 ½ Tray, 3 lbs of metal mass
  • KC300: 1 tray liner 20" x 25", 10" x 20" x 3 ½Tray, 6 lbs of metal mass
  • KC400: 1 tray liner 20" x 25", 10" x 20" x 3 ½ Tray, 10 lbs of metal mass
  • KC500: 1 tray liner 20" x 25", 11" x 22" x 3 ½Tray, 12 lbs of metal mass
  • KC600: 1 tray liner 20" x 25", 11" x 22" x 3 ½Tray, 20 lbs of metal mass

The document does not provide information on the country of origin of the data or whether the study was retrospective or prospective. It is implied to be prospective testing conducted for regulatory submission.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

This information is not provided in the document. The tests described are nonclinical, laboratory-based performance tests rather than studies requiring human expert interpretation for ground truth establishment.

4. Adjudication Method (e.g., 2+1, 3+1, none) for the Test Set

This information is not applicable as the tests are nonclinical performance evaluations rather than interpretive studies requiring adjudication.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and Effect Size

No, an MRMC comparative effectiveness study was not conducted. This type of study is typically for evaluating diagnostic or interpretive AI systems, not packaging materials.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

This is not applicable. The device is a physical sterilization wrap, not an algorithm or AI system. The performance evaluation is inherently "standalone" in the sense that it assesses the physical product's characteristics and functionality.

7. The Type of Ground Truth Used

The ground truth for the nonclinical tests appears to be established scientific and regulatory standards/methods for sterilization packaging performance. This would include methods for assessing sterilant penetration, package integrity (e.g., burst strength, seal integrity), drying, aeration, and maintenance of sterility (e.g., microbial barrier testing). The "results of testing met the same acceptance criteria as the predicate devices," implying a comparison against established performance benchmarks for similar products.

8. The Sample Size for the Training Set

This is not applicable. The KIMGUARD ONE-STEP Sterilization Wrap is a physical medical device, not an AI or Machine Learning algorithm that requires a "training set" in the computational sense.

9. How the Ground Truth for the Training Set Was Established

This is not applicable for the reasons stated above.

{0}------------------------------------------------

510(k) Summary

K141071 11111 28 2014

Applicant's Name, Address, Telephone, FAX, Contact Person

Applicant Name: Kimberly-Clark Health Care 1400 Holcomb Bridge Road Roswell, GA 30076-2190, USA Address: Establishment Registration #: . 1033422

Contact Name:

Thomas Kozma, Director of Regulatory Affairs e-mail: thomas.kozma@kcc.com phone: (770) 587-8393 (920) 225-3408 fax:

Date: July 11, 2014

1. CLASSIFICATION, COMMON OR USUAL NAME, DEVICE

NAME

.

Classification:Class II per 21 CFR 880.6850
Classification Name:Sterilization Wrap
Common/Usual Name:Sterilization Wrap
Product Code:FRG
Proprietary NameKIMGUARD® ONE-STEP® Sterilization Wrap(Models: KC100, KC200, KC300, KC400,KC500, and KC600)

2. PREDICATE DEVICES

KIMGUARD* ONE-STEP* Sterilization Wrap, which is currently manufactured and distributed by Kimberly-Clark Corporation, ( K082177 and K091685).

3. INDICATIONS FOR USE

KIMGUARD* ONE-STEP* Sterilization Wrap (KC100, KC200, KC300, KC400, KC500 and KC600) are intended to enclose another medical device that is to be sterilized by a healthcare provider using:

  • Gravity Steam at 250°F/121°C for 30 minutes. .
    KIMGUARD* ONE-STEP* Sterilization Wrap (KC100, KC200, KC300, KC400, KC500 and KC600) are intended to allow sterilization of the enclosed medical device(s) and also maintain sterility of the enclosed device(s) until used.

Test results validated that KIMGUARD* ONE-STEP* Sterilization Wrap (KC100. KC200. KC300, KC400, KC500 and KC600) allowed sterilization of the enclosed devices by the Gravity Steam Sterilization Systems (250°F/121°C for 30 minutes).

KIMGUARD* ONE-STEP* Sterilization Wrap Recommendations for Use within Gravity Steam Systems are provided in Table 1.

510(k) Summary - K141071

Page 1 of 4

{1}------------------------------------------------

TABLE 1: Recommended Loads for KIMGUARD* ONE-STEP* Sterilization Wrap for use with GravitySteam Sterilization Systems (250°F/121°C for 30 minutes)
KIMGUARD*ONE-STEP*SterilizationWrapModelsIntended Loads¹Maximum WrappedPackage ContentWeights Used in SterilityMaintenance ValidationStudyDescription of LoadsUsed in SterilityMaintenance ValidationStudy²
KC100Very Light Weight Package(for example batteries)3 lbs.1 tray liner 20" x 25"12.5" x 9" x 1" Tray1 lb of metal mass
KC200Light Weight Package (forexample telescope with lightcord)6 lbs.1 tray liner 20" x 25"10" x 20" x 3 ½ Tray3 lbs of metal mass
KC300Light to Moderate WeightPackage (for example: generaluse medical instruments)9 lbs.1 tray liner 20" x 25"10" x 20" x 3 ½Tray6 lbs of metal mass
KC400Moderate to Heavy WeightPackage (for example: generaluse medical instruments)13 lbs.1 tray liner 20" x 25"10" x 20" x 3 ½ Tray10 lbs of metal mass
KC500Heavyweight Package (forexample: general use medicalinstruments)17 lbs.1 tray liner 20" x 25"11" x 22" x 3 ½Tray12 lbs of metal mass
KC600Very Heavy Weight Package(for example: general usemedical instruments)25 lbs.1 tray liner 20" x 25"11" x 22" x 3 ½Tray20 lbs of metal mass

'individual results may differ due to factors such as variations in handling practices, wrapping techniques, and folding methods. Results may also differ due to the use of irregularly shaped contents, which may put added stress on the wrap. Each healthcare facility should determine for itself which wrap model is the most appropriate for each intended use. 11 is recommended to not exceed the maximum wrapped package content weights indicated for each wrap model. Furthermore, it is recommended to not exceed the number, weight, and size of individual content types that were validated.

4. Device Description

KIMGUARD* ONE-STEP* Sterilization Wrap is comprised of two sheets of KIMGUARD* Sterilization Wrap that is ultrasonically seamed on two edges. This seamed configuration allows for convenient wrapping of an article using two sheets simultaneously.

The sheets of sterilization wrap are square or rectangular fabric produced using a threelayer SMS (spunbound-meltblown-spunbound) process. The wrap fabric is composed of polypropylene with the addition of less than 2% by weight of phthalocyanine blue pigment, less than 1% by weight titanium dioxide pigment, and less than 0.008% by weight of antistatic treatment. The wrap allows a sterilized package to be opened aseptically.

5. Substantial Equivalence to Predicate Device

KIMGUARD ONE-STEP* Sterilization Wrap (i.e., subject of this Premarket Notification) is substantially equivalent to the predicate Kimberly-Clark KIMGUARD* ONE-STEP* Sterilization Wraps (K082177 and K091685) in technology, design, and materials.

{2}------------------------------------------------

CharacteristicsPredicate Devices:KIMGUARD* ONE-STEP* SterilizationWrap (K082177 and K091685)Proposed Device:KIMGUARD* ONE-STEP* Sterilization Wrap
ManufacturerKimberly-Clark CorporationKimberly Clark Corporation
Regulation/ProductCodeSterilization Wrap: 880.6850 / FRGSterilization Wrap: 880.6850 / FRG[SAME]
Indications for UseThe device is intended to be used to encloseanother medical device that is to besterilized by a healthcare provider. It isintended to allow sterilization of the enclosedmedical device(s) and also maintain sterilityof the enclosed device(s) until opened. Notindicated for use for gravity steamsterilization.KIMGUARD* ONE-STEP* Sterilization Wrap(KC100, KC200, KC300, KC400, KC500 andKC600) are intended to enclose anothermedical device that is to be sterilized by ahealthcare provider using:• Gravity Steam at 250°F/121°C for 30minutes.KIMGUARD* ONE-STEP* Sterilization Wrap(KC100, KC200, KC300, KC400, KC500 andKC600) are intended to allow sterilization ofthe enclosed medical device(s) and alsomaintain sterility of the enclosed device(s)until used.
SterilizationParametersPre-Vacuum Steam at 270°F/132°C for 4minutes, EO with a concentration of 725-735mg/L at 131°F/55°C and 40%-80% relativehumidity for 60 minutes, and Amsco V-ProTM 1 and V-Pro 1 Plus Low TemperatureSterilization SystemsGravity Steam at 250°F/121°C for 30 minutes
Maintenance ofPackage SterilityFor models KC100, KC200, KC300, K400,KC500, and KC600 for up to 30 days.Real-time testing following sterilization usingGravity Steam supports maintenance ofpackage sterility for 30 days for all models ofKIMGUARD* ONE-STEP* Sterilization Wrap.
TechnologyTortuous sheet material used to enclosemedical devices that are to be sterilized by ahealthcare provider to allow sterilization ofthe enclosed medical device(s) and maintainsterility of the enclosed device(s) until used.Tortuous sheet material used to enclosemedical devices that are to be sterilized by ahealthcare provider to allow sterilization ofthe enclosed medical device(s) and maintainsterility of the enclosed device(s) until used.[SAME]
Device DesignTwo sheets of nonwoven polypropylenefabric. Each sheet is composed of threethermally- bonded layers consisting of aMeltblown polypropylene layer surrounded bySpunbound polypropylene layers (SMS)Two sheets of nonwoven polypropylenefabric. Each sheet is composed of threethermally- bonded layers consisting of aMeltblown polypropylene layer surroundedby Spunbound polypropylene layers (SMS)[SAME]
Method for bondingSMS layersThermal bonding with round pin, hexagonal,triangle bond pattern ("daisy" pattern)Thermal bonding with round pin, hexagonal,triangle bond pattern ("daisy" pattern)[SAME]
MaterialsPolypropylene with blue and whitepigmentsPolypropylene with blue and whitepigments[SAME]
DistributionNon-Sterile andOver-the-CounterNon-Sterile andOver-the-Counter[SAME]
Single Use DeviceYesYes[SAME]

Device Comparison Table (Technological, Design, & Materials)

:

510(k) Summary – K141071

.

·

Page 3 of 4

.

{3}------------------------------------------------

6. SUMMARY OF NONCLINICAL TESTS:

Performance of KIMGUARD* ONE-STEP* Sterilization Wrap (KC100, KC200, KC300, KC400, KC500, KC600) has been tested as summarized below. All results of testing met the same acceptance criteria as the predicate devices demonstrating substantial equivalence to the predicate devices.

Summary of Testing Performed(Gravity Steam Sterilization)Results
Sterilant PenetrationPassed
Package Integrity/Physical PropertiesPassed
Drying and AerationPassed
Maintenance of Package IntegrityPassed
Biocompatibility (pre- and post-sterilization)Passed

7. OVERALL PERFORMANCE CONCLUSIONS

The nonclinical studies demonstrate that the KIMGUARD* ONE-STEP* Sterilization Wrap performs as intended as a sterilization packaging system of medical devices when terminally sterilized in Gravity Steam Sterilization Systems. These studies demonstrate that the KIMGUARD* ONE-STEP* Sterilization Wrap met the same criteria as the predicate devices and are substantially equivalent.

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the circle is an abstract symbol that resembles a stylized human figure.

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

July 28, 2014

Kimberly-Clark Health Care Dr. Thomas Kozma Director of Regulatory Affairs 1400 Holcomb Bridge Road Roswell, GA 30076-2190

Re: K141071

Trade/Device Name: KIMGUARD® ONE-STEP® Sterilization Wrap (Models: KC100, KC200, KC300, KC400, KC500, and KC600) Regulation Number: 21 CFR 880.6850 Regulation Name: Sterilization Wrap Regulatory Class: II Product Code: FRG Dated: June 17, 2014 Received: June 18, 2014

Dear Dr. Kozma:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{5}------------------------------------------------

Page 2 – Dr. Kozma

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours,

Mary S. Runner -S

Erin I. Keith, M.S. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{6}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

Indications for Use

Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below

510(k) Number (if known)

K141071

Device Name

KIMGUARD* ONE-STEP* Sterlization Wrap (Models KC100, KC300, KC300, KC500, and KC600)

Indications for Use (Describe)

KIMGUARD* ONE-STEP* Sterilization Wrap (KC100, KC300, KC500 and KC600) are intended to enclose another medical device that is to be sterilized by a healthcare provider using:

  • Gravity Steam at 250°F/121°C for 30 minutes. ◆
    KIMGUARD* ONE-STEP* Sterilization Wrap (KC100, KC200, KC500 and KC600) are intended to allow sterilization of the enclosed medical device(s) and also maintain sterlity of the enclosed device(s) until used.

Test results validated that KIMGUARD* ONE-STEP* Sterliization Wrap (KC100, KC300, KC300 and KC600) allowed sterlization of the enclosed devices by the Gravity Steam Sterlization Systems (250°F/121°C for 30 minutes).

KIMGUARD* ONE-STEP* Sterilization Wrap Recommendations for Use within Gravity Steam Systems are provided in Table 1.

{7}------------------------------------------------

TABLE 1: Recommended Loads for KIMGUARD* ONE-STEP* Sterilization Wrap for use with Gravity Steam SterilizationSystems (250°F/121°C for 30 minutes)
KIMGUARD* ONE-STEP* SterilizationWrap ModelsIntended Loads1Maximum Wrapped PackageContent Weights Used in SterilityMaintenance Validation StudyDescription of Loads Used inSterility MaintenanceValidation Study2
KC100Very Light Weight Package (forexample batteries)3 lbs.1 tray liner 20" x 25"12.5" x 9" x 1" Tray1 lb of metal mass
KC200Light Weight Package (forexample telescope with light cord)6 lbs.1 tray liner 20" x 25"10" x 20" x 3 ½ Tray3 lbs of metal mass
KC300Light to Moderate WeightPackage (for example: generaluse medical instruments)9 lbs.1 tray liner 20" x 25"10" x 20" x 3 ½ Tray6 lbs of metal mass
KC400Moderate to Heavy WeightPackage (for example: generaluse medical instruments)13 lbs.1 tray liner 20" x 25"10" x 20" x 3 ½ Tray10 lbs of metal mass
KC500Heavyweight Package (forexample: general use medicalinstruments)17 lbs.1 tray liner 20" x 25"11" x 22" x 3 ½ Tray12 lbs of metal mass
KC600Very Heavy Weight Package (forexample: general use medicalinstruments)25 lbs.1 tray liner 20" x 25"11" x 22" x 3 ½ Tray20 lbs of metal mass

incividual results may differ due to factors such as variations in handing pracices, wraping techniques, and folding methods. Results may also differ due to the use of impact contents, which may put added stress on the wap. Each healthcare facilly should delemine for itself which wrap model is the most appropriate for each intended use.

It is recommended to not exceed the maximum weights indicated for each wrap model. Furthermore, it is recommended to not exceed the number, weight, and size of Individual content types that were validated.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

K | Over-The-Counter Use (21 CFR 801 Subpart C)

PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.

FOR FDA USE ONLY

Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)

Elizabeth F. Claverie -S

DN: c=US, o=U.S. Government, ou=HHS, ou=FDA, ou=People, 0.9.2342.19200300.100.1.1=1300055864, cn=Ellzabeth F. Claverie -S Date: 2014.07.28 16:06:17 -04'00'

Digitally signed by Elizabeth F. Claverie -S

This section applies only to requirements of the Paperwork Reduction Act of 1995.

*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW."

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

§ 880.6850 Sterilization wrap.

(a)
Identification. A sterilization wrap (pack, sterilization wrapper, bag, or accessories, is a device intended to be used to enclose another medical device that is to be sterilized by a health care provider. It is intended to allow sterilization of the enclosed medical device and also to maintain sterility of the enclosed device until used.(b)
Classification. Class II (performance standards).