(606 days)
KIMGUARD* ONE-STEP* Sterilization Wraps are intended to enclose another medical device that is to be sterilized by a healthcare provider using:
- Advanced Sterilization Products STERRAD® Sterilization Systems that include: STERRAD® 50, 100S, and 200 o
- STERRAD® NX® [Standard Cycle, Advanced Cycle] o
- STERRAD® 100NX® [Standard Cycle, Flex Cycle, EXPRESS Cycle, DUO o Cycle]
KIMGUARD* ONE-STEP* Sterilization Wraps are intended to allow sterilization of the enclosed medical device(s) and also maintain sterility of the enclosed device(s) until used.
KIMGUARD® ONE-STEP® Sterilization Wrap is comprised of two sheets of KIMGUARD® Sterilization Wrap ultrasonically seamed on two sides. This allows for convenient wrapping with two sheets simultaneously.
The sheets of sterilization wrap are square or rectangular fabric produced using a three-layer SMS (spunbound-meltblown-spunbound) process. The wrap fabric is composed of polypropylene with the addition of less than 2% by weight of phthalocyanine blue pigment, less than 1% by weight titanium dioxide pigment, and less than 0.008% by weight of a potassium phosphate anti-static treatment. The wrap allows a sterilized package to be opened aseptically.
The provided text describes the KIMGUARD® ONE-STEP® Sterilization Wrap and validation studies demonstrating its performance. However, it does not involve an AI device or a study comparing AI with human readers. Therefore, several sections of your request are not applicable.
Here's the information that can be extracted from the provided text:
Acceptance Criteria and Device Performance
| Acceptance Criteria | Reported Device Performance |
|---|---|
| Maintain sterility until used | Passed: The nonclinical studies demonstrate that the KIMGUARD® ONE-STEP® Sterilization Wrap performs as intended as a sterilization packaging system of medical devices when terminally sterilized in the STERRAD® Sterilization Systems, and maintains sterility until used. |
| Sterilant Penetration | Passed |
| Package Integrity | Passed |
| Maintenance of Package Integrity | Passed |
| Material Compatibility | Passed |
| Shelf Life | Passed |
| Biocompatibility | Passed |
| Allow sterilization of enclosed devices | Validated: Test results validated that KIMGUARD® ONE-STEP® Sterilization Wraps (KC100, KC200, KC300, KC400, KC500, and KC600) allowed sterilization of the enclosed devices by the Advanced Sterilization Products STERRAD® Sterilization Systems (STERRAD® 50, 100S, 200, NX® [Standard Cycle and Advanced Cycle] and 100NX® [Standard Cycle, Flex Cycle, EXPRESS Cycle, DUO Cycle]). All models of the KIMGUARD® ONE-STEP® Sterilization Wrap have been validated for use with the specified STERRAD® cycles for various intended loads as detailed in Table 1 and Table 2 of the Indications for Use. |
Study Details
-
Sample size used for the test set and the data provenance:
- The document does not specify the exact sample sizes (e.g., number of wraps, number of sterilization cycles, number of instruments) used for the performance tests (Sterilant Penetration, Package Integrity, Maintenance of Package Integrity, Material Compatibility, Shelf Life, Biocompatibility).
- The data provenance is not explicitly stated in terms of country of origin or whether it was retrospective or prospective. It is implied to be prospective testing conducted by the manufacturer for regulatory submission.
-
Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable. This is a performance study for a physical medical device (sterilization wrap), not an AI diagnostic device requiring expert ground truth establishment.
-
Adjudication method for the test set: Not applicable. (See point 2)
-
If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable. This is not an AI device.
-
If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable. This is not an AI device.
-
The type of ground truth used (expert consensus, pathology, outcomes data, etc.): The "ground truth" for this device's performance is established through objective engineering and microbiological testing standards for sterilization wraps. This involves:
- Verification of the ability to allow sterilant penetration.
- Demonstration of package integrity (e.g., preventing microbial ingress).
- Proof of material compatibility with sterilization methods.
- Shelf life studies to ensure long-term integrity.
- Biocompatibility testing to ensure safety of the material.
- These are based on established scientific and regulatory guidelines for sterilization products.
-
The sample size for the training set: Not applicable. This is not an AI model, so there is no training set.
-
How the ground truth for the training set was established: Not applicable. (See point 7)
{0}------------------------------------------------
K113806 510(k) Summary
Applicant's Name, Address, Telephone, FAX, Contact Person Kimberly-Clark Health Care 1400 Holcomb Bridge Road Roswell, GA 30076-2190, USA Contact Name: Thomas Kozma, Director of Regulatory Affairs E-mail: thomas.kozma@kcc.com (770) 587-8393 (Telephone) (920) 225-3408 (Fax)
Establishment Registration Number: 1033422
AUG 20 2013
Contact Person Christopher B. Swanson
Regulatory Affairs Specialist Email: cswanso@its.jnj.com (949) 789-8620 (Telephone) (949) 789-3900 (Fax)
Date: August 8, 2013
1. CLASSIFICATION, COMMON OR USUAL NAME, DEVICE NAME
Classification: Class II per 21 CFR 880.6850 Classification Name: Sterilization Wrap Common/Usual Name: Sterilization Wrap Product Code: FRG KIMGUARD® ONE-STEP® Sterilization Proprietary Name Wrap (Models: KC100, KC200, KC300, KC400, KC500, and KC600)
2. PREDICATE DEVICES
:
KIMGUARD® ONE-STEP® Sterilization Wrap, which is currently manufactured and distributed by Kimberly-Clark Corporation [510(k) Notification K092167, cleared November 2, 2009]
{1}------------------------------------------------
3. INDICATIONS FOR USE
.
KIMGUARD* ONE-STEP* Sterilization Wraps are intended to enclose another medical device that is to be sterilized by a healthcare provider using:
- Advanced Sterilization Products STERRAD® Sterilization Systems that include: STERRAD® 50, 100S, and 200 o
- STERRAD® NX® [Standard Cycle, Advanced Cycle] o
- STERRAD® 100NX® [Standard Cycle, Flex Cycle, EXPRESS Cycle, DUO o Cycle]
KIMGUARD* ONE-STEP* Sterilization Wraps are intended to allow sterilization of the enclosed medical device(s) and also maintain sterility of the enclosed device(s) until used.
Test results validated that KIMGUARD* ONE-STEP* Sterilization Wraps (KC100, KC200, KC300, KC400, KC500, and KC600) allowed sterilization of the enclosed devices by the Advanced Sterilization Products STERRAD® Sterilization Systems (STERRAD® 50, 100S, 200, NX® [Standard Cycle and Advanced Cycle] and 100NX® [Standard Cycle, Flex Cycle, EXPRESS Cycle. DUO Cycle)). (Refer to the STERRAD® Sterilizer System User's Guide for complete instructions on load(s) and cycle(s), including chamber loading instructions.)
All models of the KIMGUARD* ONE-STEP* Sterilization Wrap have been validated for use with the STERRAD® 50, STERRAD® 100S, STERRAD® 200, STERRAD® NX® and STERRAD® 100NX® cycles in Table 1.
TABLE 1: Validated Advanced Sterilization Products (ASP) STERRAD® 100S, STERRAD® 200, STERRAD® NX®, and STERRAD® 100NX® Cycles
Note: Refer to the User's Guide for complete instructions on load and cycle for each STERRAD® System. The instructions provided below are not intended to replace the detailed Instructions For Use provided with the STERRAD® Systems
| ASP STERRAD*System and Cycle | Intended Load | ||
|---|---|---|---|
| STERRAD*50 | Reusable metal and non-metal medical devices, including up to 10 lumens of the following dimensions per chamber load:An inside diameter of 1 mm or larger and a length of 125 mm or shorter of single-channel stainless steel lumens. An inside diameter of 2 mm or larger and a length of 250 mm or shorter of single-channel stainless steel lumens. An inside diameter of 3 mm or larger and a length of 400 mm or shorter of single-channel stainless steel lumens. An inside diameter of 6 mm or larger and a length of 310 mm or shorter of single-channel Teflon/Polyethylene lumens. Refer to the STERRAD® 50 Sterilizer User's Guide for complete instructions on load(s) and cycle(s), including chamber loading instructions (i.e., 10 lumens per load). | ||
| STERRAD*100S | Reusable metal and non-metal medical devices, including up to 10 lumens of the following dimensions per chamber load:An inside diameter of 1 mm or larger and a length of 125 mm or shorter of single-channel stainless steel lumens. An inside diameter of 2 mm or larger and a length of 250 mm or shorter of single-channel stainless steel lumens. An inside diameter of 3 mm or larger and a length of 400 mm or shorter of single-channel stainless steel lumens. An inside diameter of 6 mm or larger and a length of 310 mm or shorter of single-channel Teflon/Polyethylene lumens. Refer to the STERRAD® 100S Sterilizer User's Guide for complete instructions on load(s) and cycle(s), including chamber loading instructions (i.e., 10 lumens per load). | ||
| ASP STERRAD*System and Cycle | Intended Load | ||
| STERRAD*200 | Reusable metal and non-metal medical devices, including up to 12 lumens of the following dimensions perchamber load:An inside diameter of 1 mm or larger and a length of 125 mm or shorter of single-channel stainless steel lumens. An inside diameter of 2 mm or larger and a length of 250 mm or shorter of single-channel stainless steel lumens. An inside diameter of 3 mm or larger and a length of 400 mm or shorter of single-channel stainless steel lumens. An inside diameter of 6 mm or larger and a length of 310 mm or shorter of single-channel Teflon/Polyethylene lumens. Refer to the STERRAD® 200 Sterilizer User's Guide for complete instructions on load(s) and cycle(s),including chamber loading instructions (i.e., 36.48 lbs per load). | ||
| STERRAD*NX*Standard Cycle | Reusable metal and non-metal medical devices, including up to 10 lumens of the following dimensions perchamber load:An inside diameter of 1 mm or larger and a length of 150 mm or shorter of single-channel stainless steel lumens. An inside diameter of 2 mm or larger and a length of 400 mm or shorter of single-channel stainless steel lumens. Refer to the STERRAD* NX* Sterilizer User's Guide for complete instructions on load(s) and cycle(s),including chamber loading instructions (i.e., 10.7 lbs per load). | ||
| STERRAD*NX*AdvancedCycle | Reusable metal and non-metal medical devices, including up to 10 lumens of the following dimensions perchamber load:An inside diameter of 1 mm or larger and a length of 500 mm or shorter of single-channel stainless steel lumens. OROne single-channel Flexible Endoscope with or without a silicone mat and no additional load. The flexibleendoscope may contain:A single-channel Teflon*/Polyethylene lumen with an inside diameter of 1 mm or larger and alength of 850 mm or shorter. Refer to the STERRAD® NX* Sterilizer User's Guide for complete instructions on load(s) and cycle(s),including chamber loading instructions (i.e., 10.7 lbs per load). | ||
| STERRAD*100NX*Standard Cycle | Reusable metal and non-metal medical devices, including up to 10 lumens of the following dimensions perchamber load:An inside diameter of 0.7 mm or larger and a length of 500 mm or shorter of single-channel stainless steel lumens. (A maximum of two flexible endoscopes, one per tray per sterilization cycle.) Refer to the STERRAD* 100NX* Sterilizer User's Guide for complete instructions on load(s) and cycle(s),including chamber loading instructions (i.e., 21.4 lbs per load). | ||
| STERRAD*100NX*Flex Cycle | One or two single-channel Flexible Endoscope with or without a silicone mat and no additional load. Theflexible endoscope may contain:A single-channel Teflon*/Polyethylene lumen with an inside diameter of 1 mm or larger and alength of 850 mm or shorter. (A maximum of two flexible endoscopes, one per tray per sterilization cycle). Refer to the STERRAD® 100NX* Sterilizer User's Guide for complete instructions on load(s) and cycle(s),including chamber loading instructions (i.e., 12.2 lbs per load). | ||
| STERRAD*100NX*EXPRESSCycle | Non-lumened reusable metal and non-metal devices requiring surface sterilization, and sterilization ofdiffusion-restricted spaces such as the hinged portions of forceps and scissors, and rigid or semi-rigidendoscopes without lumens.Refer to the STERRAD* 100NX* User's Guide for complete instructions on load(s) and cycle(s), includingchamber loading instructions (i.e., 10.7 lbs per load). | ||
| STERRAD*100NX*DUO Cycle | One or two single-channel Flexible Endoscope with accessory devices that are normally connected to it, with orwithout a silicone mat. The flexible endoscope may contain:A single-channel Teflon*/Polyethylene lumen with an inside diameter of 1 mm or larger and alength of 875 mm or shorter. Accessory devices that are normally connected to a flexible endoscope during use. Flexible endoscopes without lumens Refer to the STERRAD* 100NX* Sterilizer User's Guide for complete instructions on load(s) and cycle(s),including chamber loading instructions (i.e., 13.2 lbs per load). |
{2}------------------------------------------------
{3}------------------------------------------------
KIMGUARD* ONE-STEP* Sterilization Wrap Recommendations for Use with the Advanced Sterilization Products STERRAD® Sterilization Systems are provided in TABLE 2.
TABLE 2: Recommended Loads for KIMGUARD* ONE-STEP* Sterilization Wrap for use with Advanced Sterilization Products (ASP) STERRAD® Sterilization Systems (STERRAD® 50, 100S, 200, NX® [Standard Cycle, Advanced Cycle], and 100NX® [Standard Cycle, Flex Cycle, EXPRESS Cycle, DUO Cycle])
| KIMGUARD*ONE-STEP*Sterilization WrapModels | Intended Load¹ | Weights of Wrapped PackageContent Used in Validation Study(Total weight including tray) | Descriptions of Loads Used in SterilityMaintenance Validation Study |
|---|---|---|---|
| KC100 | Very light weight package(e.g., batteries or towel packs) | 10.7 lbs | APTIMAX Instrument Tray(23 in. x 11 in. x 4 in.) with Tray MatMetal and non-metal instruments |
| KC200 | Light weight package(e.g., telescope with light cord) | 10.7 lbs | APTIMAX Instrument Tray(23 in. x 11 in. x 4 in.) with Tray MatMetal and non-metal instruments |
| KC300 | Light to moderate weightpackage (e.g., general usemedical instruments) | 10.7 lbs | APTIMAX Instrument Tray(23 in. x 11 in. x 4 in.) with Tray MatMetal and non-metal instruments |
| KC400 | Moderate to heavy weightpackage (e.g., general usemedical instruments) | 10.7 lbs | APTIMAX Instrument Tray(23 in. x 11 in. x 4 in.) with Tray MatMetal and non-metal instruments |
| KC500 | Heavy weight package (e.g.,general use medical instruments) | 10.7 lbs | APTIMAX Instrument Tray(23 in. x 11 in. x 4 in.) with Tray MatMetal and non-metal instruments |
| KC600 | Very heavy weight package(e.g., general use medicalinstruments) | 10.7 lbs | APTIMAX Instrument Tray(23 in. x 11 in. x 4 in.) with Tray MatMetal and non-metal instruments |
Intended loads include: Medical Instruments with and without lumens that include telescopes, cameras, light cords, and general use medical instruments
4. DESCRIPTION OF DEVICE
KIMGUARD® ONE-STEP® Sterilization Wrap is comprised of two sheets of KIMGUARD® Sterilization Wrap ultrasonically seamed on two sides. This allows for convenient wrapping with two sheets simultaneously.
The sheets of sterilization wrap are square or rectangular fabric produced using a three-layer SMS (spunbound-meltblown-spunbound) process. The wrap fabric is composed of polypropylene with the addition of less than 2% by weight of phthalocyanine blue pigment, less than 1% by weight titanium dioxide pigment, and less than 0.008% by weight of a potassium phosphate anti-static treatment. The wrap allows a sterilized package to be opened aseptically.
{4}------------------------------------------------
| TechnologicalCharacteristics | Proposed:KIMGUARD® ONE-STEP® SterilizationWrap | Predicate Device:KIMGUARD® ONE-STEP® SterilizationWrap (K092167) |
|---|---|---|
| Manufacturer | Kimberly-Clark Corporation | Kimberly Clark Corporation |
| Regulation/ProductCode | Sterilization Wrap: 880.6850 / FRG | Sterilization Wrap: 880.6850 / FRG |
| Device Design | Two sheets of medium blue nonwovenpolypropylene fabric. Each sheet offabric is composed of three thermally-bonded layers consisting of aMeltblown polypropylene layersurrounded by Spunboundpolypropylene layers (SMS) | Two sheets of medium blue nonwovenpolypropylene fabric. Each sheet offabric is composed of three thermally-bonded layers consisting of aMeltblown polypropylene layersurrounded by Spunboundpolypropylene layers (SMS) |
| Method for bondingSMS layers | Thermal bonding with round pin,hexagonal, triangle bond pattern( |
Table 3. Device Comparison Table - Technological Characteristics
5. SUMMARY OF NONCLINICAL TESTS
Performance testing was conducted to show that the KIMGUARD® ONE-STEP® Sterilization Wrap maintains sterility until used, after completion of the sterilization process in the STERRAD® Sterilization Systems.
| Table 4: Sterilization Wrap Performance Tests | |
|---|---|
| -- | ----------------------------------------------- |
| Study | Results |
|---|---|
| Sterilant Penetration | Passed |
| Package Integrity | Passed |
| Maintenance of Package Integrity | Passed |
| Material Compatibility | Passed |
| Shelf Life | Passed |
| Biocompatibility | Passed |
6. OVERALL PERFORMANCE CONCLUSIONS
The nonclinical studies demonstrate that the KIMGUARD® ONE-STEP® Sterilization Wrap performs as intended as a sterilization packaging system of medical devices when terminally sterilized in the STERRAD® Sterilization Systems. These studies show that the KIMGUARD® ONE-STEP® Sterilization Wrap met the same criteria as the predicate device and are substantially equivalent.
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES-USA" arranged around the perimeter. Inside the circle is an abstract symbol resembling an eagle or bird-like figure, composed of three curved lines.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
August 20, 2013
Kimberly-Clark Health Care Christopher B. Swanson Regulatory Affairs Specialist Email 1400 Holcomb Bridge Road ROSWELL GA 30076-2190
Re: K113806
Trade/Device Name: KIMGUARD® ONE-STEP® Sterilization Wrap (Models KC100, KC200, KC 300, KC 400, KC500, and KC600) Regulation Number: 21 CFR 880.6850 Regulation Name: Sterilization Wrap Regulatorv Class: II Product Code: FRG Dated: August 8, 2013 Received:
Dear Mr. Swanson:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{6}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Mary S. Runner -S
Kwame Ulmer M.S. Acting Division Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{7}------------------------------------------------
Indications for Use
510(k) Number (if known): K113806
KIMGUARD® ONE-STEP® Sterilization Wrap (Models_KC100, Device Name: KC200, KC300, KC500, KC500, and KC600) Indications for Use with Advanced Sterilization Products STERRAD® Sterilization Systems
Indications for Use:
KIMGUARD* ONE-STEP* Sterilization Wraps are intended to enclose another medical device that is to be sterilized by a healthcare provider using:
- · Advanced Sterilization Products STERRAD® Sterilization Systems that include:
- STERRAD® 50, 100S, and 200 o
- STERRAD® NX® [Standard Cycle, Advanced Cycle] o
- STERRAD® 100NX® [Standard Cycle, Flex Cycle, EXPRESS Cycle, DUO o Cycle]
KIMGUARD* ONE-STEP* Sterilization Wraps are intended to allow sterilization of the enclosed medical device(s) and also maintain sterility of the enclosed device(s) until used.
Test results validated that KIMGUARD* ONE-STEP* Sterilization Wraps (KC100, KC200, KC300, KC400, KC500, and KC600) allowed sterilization of the enclosed devices by the Advanced Sterilization Products STERRAD® Sterilization Systems (STERRAD® 50, 100S, 200, NX® [Standard Cycle and Advanced Cycle] and 100NX® [Standard Cycle, Flex Cycle, EXPRESS Cycle, DUO Cycle]). (Refer to the STERRAD® Sterilizer System User's Guide for complete instructions on load(s) and cycle(s), including chamber loading instructions.)
All models of the KIMGUARD* ONE-STEP* Sterilization Wrap have been validated for use with the STERRAD® 50, STERRAD® 100S, STERRAD® 200, STERRAD® NX® and STERRAD® 100NX® cycles in Table 1.
KIMGUARD* ONE-STEP* Sterilization Wrap Recommendations for Use with the Advanced Sterilization Products STERRAD® Sterilization Systems are provided in Table 2.
{8}------------------------------------------------
TABLE 1: Validated Advanced Sterilization Products (ASP) STERRAD® 50, STERRAD® 100S, STERRAD® 200, STERRAD® NX®, and STERRAD® 100NX® Cycles
| STERRAD® Systems | ||||||
|---|---|---|---|---|---|---|
| ASP STERRAD®System and Cycle | Intended Load | |||||
| STERRAD®50 | Reusable metal and non-metal medical devices, including up to 10 lumens of the following dimensions perchamber load:An inside diameter of 1 mm or larger and a length of 125 mm or shorter of single-channel stainless steel lumens. An inside diameter of 2 mm or larger and a length of 250 mm or shorter of single-channel stainless steel lumens. An inside diameter of 3 mm or larger and a length of 400 mm or shorter of single-channel stainless steel lumens. An inside diameter of 6 mm or larger and a length of 310 mm or shorter of single-channel Teflon/Polyethylene lumens. Refer to the STERRAD® 50 Sterilizer User's Guide for complete instructions on load(s) and cycle(s), including chamber loading instructions (i.e., 10 lumens per load). | |||||
| STERRAD®100S | Reusable metal and non-metal medical devices, including up to 10 lumens of the following dimensions perchamber load:An inside diameter of 1 mm or larger and a length of 125 mm or shorter of single-channel stainless steel lumens. An inside diameter of 2 mm or larger and a length of 250 mm or shorter of single-channel stainless steel lumens. An inside diameter of 3 mm or larger and a length of 400 mm or shorter of single-channel stainless steel lumens. An inside diameter of 6 mm or larger and a length of 310 mm or shorter of single-channel Teflon/Polyethylene lumens. Refer to the STERRAD® 100S Sterilizer User's Guide for complete instructions on load(s) and cycle(s), including chamber loading instructions (i.e., 10 lumens per load). | |||||
| STERRAD®200 | Reusable metal and non-metal medical devices, including up to 12 lumens of the following dimensions perchamber load:An inside diameter of 1 mm or larger and a length of 125 mm or shorter of single-channel stainless steel lumens. An inside diameter of 2 mm or larger and a length of 250 mm or shorter of single-channel stainless steel lumens. An inside diameter of 3 mm or larger and a length of 400 mm or shorter of single-channel stainless steel lumens. An inside diameter of 6 mm or larger and a length of 310 mm or shorter of single-channel Teflon/Polyethylene lumens. Refer to the STERRAD® 200 Sterilizer User's Guide for complete instructions on load(s) and cycle(s), including chamber loading instructions (i.e., 36.48 lbs per tray load). | |||||
| STERRAD®NX®Standard Cycle | Reusable metal and non-metal medical devices, including up to 10 lumens of the following dimensions perchamber load:An inside diameter of 1 mm or larger and a length of 150 mm or shorter of single-channel stainless steel lumens. An inside diameter of 2 mm or larger and a length of 400 mm or shorter of single-channel stainless steel lumens. Refer to the STERRAD® NX® Sterilizer User's Guide for complete instructions on load(s) and cycle(s), including chamber loading instructions (i.e., 10.7 lbs per load). |
Note: Refer to the User's Guide for complete instructions on load and cycle for each STERRAD® System. The instructions provided below are not intended to replace the detailed Instructions For Use provided with the
{9}------------------------------------------------
:
| TABLE 1: Validated Advanced Sterilization Products (ASP) STERRAD ® 50, STERRAD® 100S |
|---|
| STERRAD® 200, STERRAD NX®, and STERRAD® 100NX® Cycles (cont.) |
| ASP STERRAD®System and Cycle | Intended Load |
|---|---|
| STERRAD®NX®AdvancedCycle | Reusable metal and non-metal medical devices, including up to 10 lumens of the following dimensions perchamber load:• An inside diameter of 1 mm or larger and a length of 500 mm or shorter of single-channel stainlesssteel lumens.OROne single-channel Flexible Endoscope with or without a silicone mat and no additional load. The flexibleendoscope may contain:• A single-channel Teflon®/Polyethylene lumen with an inside diameter of 1 mm or larger and alength of 850 mm or shorter.Refer to the STERRAD® NX® Sterilizer User's Guide for complete instructions on load(s) and cycle(s),including chamber loading instructions (i.e., 10.7 lbs per load). |
| STERRAD®100NX®Standard Cycle | Reusable metal and non-metal medical devices, including up to 10 lumens of the following dimensions perchamber load:• An inside diameter of 0.7 mm or larger and a length of 500 mm or shorter of single-channelstainless steel lumens. (A maximum of two flexible endoscopes, one per tray per sterilization cycle.)Refer to the STERRAD® 100NX® Sterilizer User's Guide for complete instructions on load(s) and cycle(s),including chamber loading instructions (i.e., 21.4 lbs per load). |
| STERRAD®100NX®Flex Cycle | One or two single-channel Flexible Endoscope with or without a silicone mat and no additional load. Theflexible endoscope may contain:• A single-channel Teflon®/Polyethylene lumen with an inside diameter of 1 mm or larger and alength of 850 mm or shorter. (A maximum of two flexible endoscopes, one per tray per sterilizationcycle).Refer to the STERRAD® 100NX® Sterilizer User's Guide for complete instructions on load(s) and cycle(s),including chamber loading instructions (i.e., 12.2 lbs per load). |
| STERRAD®100NX®EXPRESSCycle | Non-lumened reusable metal and non-metal devices requiring surface sterilization, and sterilization ofdiffusion-restricted spaces such as the hinged portions of forceps and scissors, and rigid or semi-rigidendoscopes without lumens.Refer to the STERRAD® 100NX® User's Guide for complete instructions on load(s) and cycle(s), includingchamber loading instructions (i.e., 10.7 lbs per load). |
| STERRAD®100NX®DUO Cycle | One or two single-channel Flexible Endoscope with accessory devices that are normally connected to it, with orwithout a silicone mat. The flexible endoscope may contain:• A single-channel Teflon®/Polyethylene lumen with an inside diameter of 1 mm or larger and alength of 875 mm or shorter.• Accessory devices that are normally connected to a flexible endoscope during use.• Flexible endoscopes without lumensRefer to the STERRAD® 100NX® Sterilizer User's Guide for complete instructions on load(s) and cycle(s),including chamber loading instructions (i.e., 13.2 lbs per load). |
{10}------------------------------------------------
TABLE 2: Recommended Loads for KIMGUARD* ONE-STEP* Sterilization Wrap for use with Advanced Sterilization Products (ASP) STERRAD® Sterilization Systems (STERRAD® 50, 100S, 200, NX® |Standard Cycle, Advanced Cycle], and 100NX® [Standard Cycle, Flex Cycle, EXPRESS Cycle, DUO Cycle|)
| KIMGUARD®ONE-STEP®Sterilization WrapModels | Intended Load¹ | Weights of Wrapped PackageContent Used in Validation Study(Total weight including tray) | Descriptions of Loads Used in SterilityMaintenance Validation Study |
|---|---|---|---|
| KC100 | Very light weight package(e.g., batteries or towel packs) | 10.7 lbs | • APTIMAX® Instrument Tray(23 in. x 11 in. x 4 in.) with Tray Mat• Metal and non-metal instruments |
| KC200 | Light weight package(e.g., telescope with light cord) | 10.7 lbs | • APTIMAX® Instrument Tray(23 in. x 11 in. x 4 in.) with Tray Mat• Metal and non-metal instruments |
| KC300 | Light to moderate weightpackage (e.g., general usemedical instruments) | 10.7 lbs | • APTIMAX® Instrument Tray(23 in. x 11 in. x 4 in.) with Tray Mat• Metal and non-metal instruments |
| KC400 | Moderate to heavy weightpackage (e.g., general usemedical instruments) | 10.7 lbs | • APTIMAX® Instrument Tray(23 in. x 11 in. x 4 in.) with Tray Mat• Metal and non-metal instruments |
| KC500 | Heavy weight package (e.g.,general use medical instruments) | 10.7 lbs | • APTIMAX® Instrument Tray(23 in. x 11 in. x 4 in.) with Tray Mat• Metal and non-metal instruments |
| KC600 | Very heavy weight package(e.g., general use medicalinstruments) | 10.7 lbs | • APTIMAX® Instrument Tray(23 in. x 11 in. x 4 in.) with Tray Mat• Metal and non-metal instruments |
Intended loads include: Medical Instruments with and without lumens that include telescopes, cameras, light cords, and general use medical instruments
Prescription Use _ (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use _ V (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Sreekanth Gutala
2013.08.20 13:05:45-04'00'
Page 4 of 4
(Division Sign-Off)
Division of Anesthesiology, General Hospital
Respiratory, Infection Control and
Dental Devices
| 510(k) Number: | K113806 |
|---|---|
| ---------------- | --------- |
§ 880.6850 Sterilization wrap.
(a)
Identification. A sterilization wrap (pack, sterilization wrapper, bag, or accessories, is a device intended to be used to enclose another medical device that is to be sterilized by a health care provider. It is intended to allow sterilization of the enclosed medical device and also to maintain sterility of the enclosed device until used.(b)
Classification. Class II (performance standards).