    Intended Use

    HALYARD® SMART-FOLD Sterilization Wraps are intended to be used to enclose another medical device that is to be sterilized by a healthcare provider using:

    • Pre-vacuum steam at 270°F/132°C for 4 minutes. The wrap was validated for dry times of 30 minutes.
    • 100% ethylene oxide (EO) with a concentration of 725-735 mg/L at 131°F/55°C and 40%-80% relative humidity for 60 minutes. The wrap was validated for aeration times for EO sterilization of 8 hours at 110° F/43.3ºC.

    STERIS V-PRO® Low Temperature Sterilization Systems.

    • · STERIS® V-PRO® 60 (Lumen, Non-Lumen and Flexible Cycles)

    • · STERIS® V-PRO® s2 (Lumen, Non-Lumen and Flexible Cycles)

    • · STERIS® V-PRO® 1 (Lumen Cycle)

    • · STERIS® V-PRO® 1 Plus (Lumen and Non-Lumen Cycle)

    • · STERIS® V-PRO® maX (Lumen, Non-Lumen and Flexible Cycle)

    • · STERIS® V-PRO® maX 2 (Lumen, Non-Lumen and Flexible Cycle)

    • Advanced Sterilization Products STERRAD® Sterilization System

    • STERRAD® 100S

    • · STERRAD® NX®, (Standard Cycle, Advanced Cycle)

    • · STERRAD® NX® with ALLClear® Technology, (Standard Cycle, Advanced Cycle)

    • · STERRAD® 100NX® (Standard Cycle, Flex Cycle, EXPRESS Cycle, DUO Cycle, ULTRA GL Cycle)

    • STERRAD® 100NX® with ALLClear® Technology, (Standard Cycle, EXPRESS Cycle, DUO Cycle, ULTRA GI Cycle)

    HALYARD* ONE-STEP, QUICK CHECK and SEQUENTIAL Sterilization Wraps are intended to be used to enclose another medical device that is to be sterilized by a healthcare provider using:

    · Pre-vacuum steam at 270°F/132°C for 4 minutes. The wrap was validated for dry times of 20 minutes for Models 100 and 200, and for 30 minutes for Models 300, 400, 500, and 600.

    • · 100% ethylene oxide (EO) with a concentration of 725-735 mg/L at 131°F/55°C and 40%-80% relative humidity for 60 minutes. The wrap was validated for aeration times for EO sterilization of 8 hours at 43.3℃.
      · STERIS V-PRO® Low Temperature Sterilization Systems. The wrap was validated to be effectively aerated during the pre-programmed cycles.

    • · STERIS V-PRO® 60 (Lumen, Non-Lumen and Flexible Cycles)

    • · STERIS V-PRO® s2 (Lumen, Non-Lumen and Flexible Cycles)

    • · STERIS V-PRO® 1 (Lumen Cycle)

    • · STERIS V-PRO® 1 Plus (Lumen and Non-Lumen Cycle)

    • · STERIS V-PRO® maX (Lumen, Non-Lumen and Flexible Cycle)

    • · STERIS V-PRO® maX 2 (Lumen, Non-Lumen and Flexible Cycle)

    · Gravity steam at 250°F/121°C for 30 minutes (25 minute dry time for Models 100, 200 and 30 minute dry time for Models 400, 500 and 600)

    • · Advanced Sterilization Products STERRAD® Sterilization System
    • · STERRAD® 50, 100S, and 200
    • · STERRAD® NX®, (Standard Cycle, Advanced Cycle)
    • · STERRAD® NX® with ALLClear® Technology, (Standard Cycle, Advanced Cycle)
    • · STERRAD® 100NX® (Standard Cycle, Flex Cycle, EXPRESS Cycle, DUO Cycle)
    • · STERRAD® 100NX® with ALLClear® Technology, (Standard Cycle, EXPRESS Cycle, DUO Cycle)
    • · STERILUCENT® HC80TT Sterilization Cycles (Lumen and Flexible Cycles)
    • · Stryker Sterizone® VP4 Sterilizer Cycle 1
    Device Description

    HALYARD SEQUENTIAL Sterilization Wrap is supplied to the customer as bulk packages of single sheets, where in accordance with standard hospital practices, two sheets are then used to wrap a medical device or a collection of medical devices for sterilization. HALYARD ONE-STEP, HALYARD QUICK CHECK and HALYARD SMART-FOLD Sterilization Wraps are comprised of two sheets of HALYARD* SEQUENTIAL Sterilization Wrap. This allows for convenient wrapping with two sheets simultaneously.

    The HALYARD ONE-STEP, QUICK CHECK and HALYARD SMART-FOLD Sterilization Wraps are intended to enclose another medical device that is to be sterilized by a healthcare provider.

    AI/ML Overview

    The document describes the regulatory submission for HALYARD sterilization wraps. It focuses on demonstrating substantial equivalence to a predicate device through performance testing. No AI/ML algorithm is involved in the device. Therefore, a complete response is not possible for all aspects of your request.

    Here's a breakdown of the available information:

    1. A table of acceptance criteria and the reported device performance

    PurposeTestAcceptance CriteriaResults
    Sterilant Penetration/EfficacyANSI/AAMI ST79, ANSI/AAMI/ISO 11138-7Achieving a 10^-6 sterility assurance level following processing in a worst-case half-cyclePassed
    Performance Testing (Non-sterile/Sterile)ANSI/AAMI/ISO 11607-1 Annex B, ISO 13938-2, ASTM D4966-12, CPSC 1610Complies with the selected physical propertiesPassed
    Maintenance of Package IntegrityANSI/AAMI/ISO 11607-1, ANSI/AAMI ST79Maintain sterility for 1 year, except for gravity steam (30 days), STERRAD ULTRA GI (3 months) and Sterilucent (6 months).Passed
    BiocompatibilityISO 10993-1, ISO 10993-5, ISO 10993-10, ISO 10993-7Non-cytotoxic, Non-irritating, Non-sensitizingPassed
    ResidualsSteris V-PRO Low Temperature Sterilization System H2O2 residuals, Sterizone VP4 Cycle 1 residual H2O2, ISO 14937 (ULTRA GI Cycle only), ISO 10993-7 EtO and ECHH2O2 ≤ 0.000218 mg/cm2, H2O2 ≤ 0.56 µg/cm2, H2O2 ≤ 9100 µg/cm2, EO