    Intended Use

    Software: The Precision AI Planning Software is intended to be used as a pre-surgical planner for simulation of surgical interventions for shoulder joint arthroplasty. The software is used to assist in the positioning of shoulder components by creating a 3D bone construct of the joint and allows the surgeon to visualize, measure, reconstruct, annotate and edit presurgical plan data. The software leads to the generation of a surgery report along with a pre-surgical plan data file which can be used as input data to design the Precision AI Shoulder Guide and Biomodels.

    Hardware: The Precision AI Planning System Guides and Biomodels are intended to be used as patient-specific surgical instruments to assist in the intraoperative positioning of shoulder implant components used with total and reverse shoulder arthroplasty by referencing anatomic landmarks of the shoulder that are identifiable on preoperative CT-imaging scans. The Glenoid Guide is used to place the k-wire and the Humeral Guide is used to place humeral pins for humeral head resection. The Precision AI Guides and Biomodels are indicated for single use only. The Precision AI Surgical Planning System is indicated for use on adult patients that have been consented for shoulder joint arthroplasty. Both humeral and glenoid guides are suitable for a delto-pectoral approach only. The Precision AI Surgical Planning System is indicated for total and reverse shoulder arthroplasty using the following implant systems and their compatible components: Enovis and Lima.

    Device Description

    The Precision AI Surgical Planning System is a patient-specific medical device that is designed to be used to assist the surgeon in the placement of shoulder components during total anatomic and reverse shoulder replacement surgery. This can be done by generating a pre-surgical shoulder plan and, if requested by the surgeon, by manufacturing a patient-specific guides and models to transfer the plan to surgery. The subject device is a system composed of the following: The Precision AI Surgical Planning System Software will create a 3D construct/render of the patient's shoulder joint for the surgeon to plan the operatively then create a physical Patient Specific Instrument (or Guide), using 3D printing by selective laser sintering. The patient's CT scan images are the design input for this to be created and are auto segmented via a locked, or static, artificial intelligence algorithm. The surgeon can visualise the deformity of the diseased joint, on this 3D render and CT scan images, and determine the inherent deformity of the joint. They are then able to virtually place the artificial implants in an optimal position to correct the measured deformity for that specific patient. The Precision AI Guides, which are a patient-specific guide and models that are based on a pre-surgical plan. This pre-surgical plan is generated using the software component. Patient-specific guide and models will be manufactured if the surgeon requests patient-specific guides to transfer the plan to surgery.

    AI/ML Overview

    The provided text is an FDA 510(k) summary for the Precision AI Surgical Planning System (PAI-SPS). It describes the device and its intended use, but it does NOT contain specific acceptance criteria, detailed study designs, or performance results in terms of metrics like sensitivity, specificity, accuracy, or effect sizes for human reader improvement.

    The document states that "Software verification and validation were performed, and documentation was provided following the 'Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices.' This includes verification against defined requirements and validation against user needs." It also mentions that "Design verification and validation testing demonstrated that the PAI-SPS meets all design requirements and is as safe and effective as its predicate device (K233992)."

    However, it does not explicitly lay out the acceptance criteria (e.g., "model must achieve 90% accuracy") or the results against those criteria. It focuses more on demonstrating substantial equivalence to a predicate device (PAI-SPS K233992) by showing that the technology and intended use are similar, with the main differences being the addition of compatibility with more implant systems.

    Therefore, I cannot fully complete the requested table and answer all questions based solely on the provided text. I will provide information based on what is available and indicate where information is missing.


    Description of Acceptance Criteria and Study to Prove Device Meets Criteria

    Based on the provided FDA 510(k) summary, the PAI-SPS device is being cleared primarily through demonstrating substantial equivalence to an already cleared predicate device (PAI-SPS, K233992). The key claim for equivalence rests on similar intended use, fundamental scientific technology, design, functionality, operating principles, and materials, with the primary difference being expanded compatibility with additional implant systems.

    The document implicitly suggests that the "acceptance criteria" are tied to demonstrating that these technological differences "do not raise any different questions of safety and effectiveness." The studies cited are primarily focused on software verification and validation, ensuring the new compatible implant systems do not negatively impact the established safety and effectiveness.

    Here is a breakdown of the available information:

    1. Table of Acceptance Criteria and Reported Device Performance

    Note: The provided document does not explicitly state quantitative acceptance criteria (e.g., minimum accuracy percentages, specific error bounds) or detailed performance metrics. The performance is largely framed as demonstrating that the device "meets all design requirements and is as safe and effective as its predicate device."

    Acceptance Criteria (Implied)Reported Device Performance
    Software:
    Functions as a pre-surgical planner for shoulder joint arthroplasty (visualization, measurement, reconstruction, annotation, editing of plan data)."The planning functionality, including visualization, measurement, reconstruction, annotation, and editing of pre-surgical plan data, is the same in the subject and predicate device."
    "Software verification and validation were performed, and documentation was provided following the 'Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices.'"
    "Design verification and validation testing demonstrated that the PAI-SPS meets all design requirements and is as safe and effective as its predicate device (K233992)."
    Automated segmentation via artificial intelligence algorithm is locked/static and accurate for 3D bone construct creation."The patient's CT scan images are the design input for this to be created and are auto segmented via a locked, or static, artificial intelligence algorithm."
    (No specific numerical accuracy or precision metrics are reported for segmentation).
    Expanded compatibility with new Enovis and Lima implant systems does not introduce new safety/effectiveness concerns."The non-clinical performance data has demonstrated that the subject software technological differences between the subject and predicate device do not raise any different questions of safety and effectiveness."
    Hardware (Guides & Biomodels):
    Assists in intraoperative positioning of shoulder implant components by referencing anatomic landmarks."Testing verified that the accuracy and performance of the system is adequate to perform as intended."
    "The stability of the device placement, surgical technique, intended use and functional elements of the subject device are the same as that of the predicate device of Precision AI Surgical Planning System (K233992) and therefore previous cadaver testing and composite bone model testing on the previously cleared device are considered applicable to the subject device."
    Expanded compatibility with new Enovis and Lima implant systems does not introduce new safety/effectiveness concerns."The main difference between the subject device hardware and the predicate device is the extension of compatibility of the Precision AI Guides and Models with additional Enovis' and Lima's implant systems and their compatible components... [demonstrated not to raise new safety/effectiveness questions based on previous testing for predicate]."
    Biocompatibility, sterility, cleaning, debris, dimensional stability, and packaging are adequate."Previous testing for biocompatibility, sterility, cleaning, debris, dimensional stability and packaging are applicable to the subject device." (Implies these aspects were re-verified or deemed unchanged/covered by predicate testing).

    2. Sample Size Used for the Test Set and Data Provenance

    • The document does not specify the sample size for any test set (e.g., for software validation or hardware accuracy).
    • Data Provenance: Not explicitly stated for specific test sets. Given the company is "Precision AI Pty Ltd" in Australia, and the document discusses "previous cadaver testing and composite bone model testing," it's likely a mix of lab-based/simulated data and potentially some retrospective clinical imaging data for initial AI development/testing, but this is not detailed. The document implies that new testing was not extensively conducted for this submission, relying heavily on the predicate device's prior validation and the minor changes to compatibility.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Their Qualifications

    • The document does not specify the number of experts or their qualifications used to establish ground truth for any test set.
    • It mentions that the software allows a "qualified surgeon" to approve pre-surgical plan data, implying that expert surgical review is part of the workflow.

    4. Adjudication Method for the Test Set

    • The document does not specify any adjudication method (e.g., 2+1, 3+1) for establishing ground truth or evaluating test results.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done

    • No, an MRMC comparative effectiveness study was not explicitly mentioned or described. The focus of this 510(k) is substantial equivalence based on technological similarity and expanded compatibility, not a comparative study against human readers or performance improvement with AI assistance.

    6. If a Standalone (i.e. algorithm only without human-in-the loop performance) was done

    • The document states that the AI algorithm performs "auto segmentation via a locked, or static, artificial intelligence algorithm." While this indicates a standalone AI component, the document does not provide standalone performance metrics for this AI segmentation. The overall system is described as a "pre-surgical planner" where the surgeon can "visualize, measure, reconstruct, annotate and edit pre-surgical plan data," suggesting a human-in-the-loop workflow.

    7. The Type of Ground Truth Used

    • For software, the implicit ground truth appears to be expert consensus or approved surgical plans for judging the accuracy of the software's representations and planning capabilities. The document states "The software allows a qualified surgeon to visualize, measure, reconstruct, annotate, edit and approve pre-surgical plan data."
    • For hardware, "previous cadaver testing and composite bone model testing" were used, implying physical measurements against a known standard or "true" position established in these models.

    8. The Sample Size for the Training Set

    • The document does not specify the sample size used for the training set for the AI segmentation algorithm.

    9. How the Ground Truth for the Training Set was Established

    • The document does not specify how the ground truth for the AI training set was established. It only mentions that the AI algorithm for auto-segmentation is "locked, or static," implying it was trained previously.
    Ask a Question

    Ask a specific question about this device

    Why did this record match?
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Hardware:
    The Materialise Shoulder Guide and Models are intended to be used as a surgical instrument to assist in the intraoperative positioning of glenoid components used with total and reverse shoulder arthroplasty by referencing anatomic landmarks of the shoulder that are identifiable on preoperative CT-imaging scans.

    The Materialise Shoulder Guide and Models are single use only.

    The Materialise Shoulder Guide and Models can be used in conjunction with the following total and reverse shoulder implants systems and their respective compatible components:

    Software:
    SurgiCase Shoulder Planner is intended to be used as a pre-surgical planner for simulation of surgical interventions for shoulder orthopedic surgery. The software is used to assist in the positioning of shoulder components. SurgiCase Shoulder Planner allows the surgeon to visualize, measure, reconstruct, annotate and edit pre-surgical plan data. The software leads to the generation of a surgery report along with a pre-surgical plan data file which can be used as input data to design the Materialise Shoulder Guide and Models.

    Device Description

    Materialise Shoulder System™ is a patient-specific medical device that is designed to be used to assist the surgeon in the placement of shoulder components during total anatomic and reverse shoulder replacement surgery. This can be done by generating a pre-surgical shoulder plan and, if requested by the surgeon, by manufacturing a patient-specific glenoid guide and models to transfer the glenoid plan to surgery. The device is a system composed of the following:

    • a software component, branded as SurgiCase Shoulder Planner. This software is a planning tool used . to generate a pre-surgical plan for a specific patient.
    • Materialise Shoulder Guide and Models, which are a patient-specific quide and models that are based ● on a pre-surgical plan. This pre-surgical plan is generated using the software component. Patientspecific glenoid guide and models will be manufactured if the surgeon requests patient-specific guides to transfer the glenoid plan to surgery. The Materialise Shoulder Guide is designed and manufactured to fit the anatomy of a specific patient. A bone model of the scapula is delivered with the Materialise Shoulder Guide. A graft model can be delivered with the Materialise Shoulder Guide. The graft model visualizes the graft-space between implant and bone, based on the pre-operative planning of the surgeon. The graft model serves as a visual reference for the surgeon in the OR.
    AI/ML Overview

    The provided FDA 510(k) summary (K242813) for the Materialise Shoulder System™ describes a submission seeking substantial equivalence to a previously cleared device (K241143). This submission is primarily for adding compatibility with new implant components rather than introducing a completely new AI capability or significant software change that would necessitate extensive new performance data. Therefore, the document does not contain the detailed information typically found in a study proving a device meets acceptance criteria for an AI/ML product.

    Specifically, the document states:

    • "The non-clinical performance data has demonstrated that the subject software technological differences between the subject and predicate devices do not raise any different questions of safety and effectiveness." (Page 9)
    • "Software verification and validation were performed, and documentation was provided following the 'Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices.' This includes verification against defined requirements, and validation against user needs." (Page 10)
    • "Previous testing for biocompatibility, sterility, cleaning, debris, dimensional stability and packaging are applicable to the subject device. Testing verified that the accuracy and performance of the system is adequate to perform as intended. The stability of the device placement, surgical technique, intended use and functional elements of the subject device are the same as that of the predicate device of Materialise Shoulder System™ K241143 and previously cleared devices... therefore previous simulated surgeries using rapid prototyped bone models and previous cadaver testing on previously cleared devices K153602 and K131559 are considered applicable to the subject device." (Page 10)

    Given this, I cannot provide detailed answers to many of your questions as the submission relies on the substantial equivalence principle and prior testing rather than new, extensive performance studies for AI/ML.

    However, I can extract what is available:

    1. A table of acceptance criteria and the reported device performance

    The document does not provide a specific table of quantitative acceptance criteria and reported device performance for the current submission (K242813), as it relies on the previous clearance and the assessment that the changes (adding implant compatibility) do not raise new safety or effectiveness concerns.

    The general acceptance criterion mentioned is that the "accuracy and performance of the system is adequate to perform as intended." This was verified through previous testing, including "simulated surgeries using rapid prototyped bone models and previous cadaver testing."

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    The document mentions "previous simulated surgeries using rapid prototyped bone models and previous cadaver testing on previously cleared devices K153602 and K131559." It does not specify the sample size for these tests, nor the country of origin of the data or whether it was retrospective or prospective.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    This information is not provided in the document.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    This information is not provided in the document.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    No such MRMC study is mentioned. The device is a "pre-surgical planner" and "surgical instrument" designed to assist the surgeon, but the provided text does not detail comparative effectiveness studies of human readers (surgeons) with and without the AI (planning software) assistance.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    The software (SurgiCase Shoulder Planner) generates a pre-surgical plan which the "qualified surgeon" can "visualize, measure, reconstruct, annotate, edit and approve" (Page 9). This indicates a human-in-the-loop process. Standalone performance of the algorithm without human interaction is not discussed as it's not the intended use.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    The document generally refers to "verification against defined requirements, and validation against user needs" and "accuracy and performance of the system is adequate to perform as intended" based on "simulated surgeries using rapid prototyped bone models and previous cadaver testing." This suggests a ground truth established through expert-defined surgical planning parameters and comparison to physical outcomes in the simulated/cadaveric environment, but specifics on how this ground truth was formalized (e.g., expert consensus on optimal planning, precise measurement validation) are not detailed.

    8. The sample size for the training set

    This device is a surgical planning tool and guides, not a deep learning AI model that requires a "training set" in the conventional sense for image classification or similar tasks. It is based on algorithms that process CT-imaging scans and anatomical landmarks to generate personalized plans and guides. Therefore, the concept of a "training set" for AI/ML is not applicable here in the way it would be for a pattern recognition AI. The software's robustness and accuracy are likely validated through extensive testing against various patient anatomies and surgical scenarios.

    9. How the ground truth for the training set was established

    As explained above, the concept of a training set as typically understood for AI/ML models is not directly applicable to this device based on the provided information.

    Ask a Question

    Ask a specific question about this device

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Hardware:
    The Materialise Shoulder Guide and Models are intended to be used as a surgical instrument to assist in the intraoperative positioning of glenoid components used with total and reverse shoulder arthroplasty by referencing anatomic landmarks of the shoulder that are identifiable on preoperative CT-imaging scans.
    The Materialise Shoulder Guide and Models are single use only.
    The Materialise Shoulder Guide and Models can be used in conjunction with the following total and reverse shoulder implants systems and their respective compatible components:

    Software:
    SurgiCase Shoulder Planner is intended to be used as a pre-surgical planner for simulation of surgical interventions for shoulder orthopedic surgery. The software is used to assist in the positioning of shoulder components. SurgiCase Shoulder Planner allows the surgeon to visualize, measure, reconstruct, annotate and edit pre-surgical plan data. The software leads to the generation of a surgery report along with a pre-surgical plan data file which can be used as input data to design the Materialise Shoulder Guide and Models.

    Device Description

    Materialise Shoulder System™ is a patient-specific medical device that is designed to be used to assist the surgeon in the placement of shoulder components during total anatomic and reverse shoulder replacement surgery. This can be done by generating a pre-surgical shoulder plan and, if requested by the surgeon, by manufacturing a patient-specific glenoid guide and models to transfer the glenoid plan to surgery. The device is a system composed of the following:

    • a software component, branded as SurgiCase Shoulder Planner. This software is a planning tool used to generate a pre-surgical plan for a specific patient.
    • Materialise Shoulder Guide and Models, which are a patient-specific guide and models that are based on a pre-surgical plan. This pre-surgical plan is generated using the software component. Patient-specific glenoid guide and models will be manufactured if the surgeon requests patient-specific guides to transfer the glenoid plan to surgery. The Materialise Shoulder Guide is designed and manufactured to fit the anatomy of a specific patient. A bone model of the scapula is delivered with the Materialise Shoulder Guide. A graft model can be delivered with the Materialise Shoulder Guide. The graft model visualizes the graft-space between implant and bone, based on the pre-operative planning of the surgeon. The graft model serves as a visual reference for the surgeon in the OR.
    AI/ML Overview

    The provided text describes a 510(k) submission for the Materialise Shoulder System™, Materialise Shoulder Guide and Models, and SurgiCase Shoulder Planner. It indicates that this is a special 510(k) submission, meaning it's for a modification to a previously cleared device. Therefore, much of the performance data refers back to the predicate device and prior clearances.

    Here's an analysis of the acceptance criteria and supporting study information based on the provided text:

    1. A table of acceptance criteria and the reported device performance

    The document does not explicitly state quantitative acceptance criteria or a direct table showing "acceptance criteria vs. reported device performance" for this specific 510(k) submission. Instead, for this special 510(k), the performance data mainly focuses on demonstrating that the changes (addition of new implant components to the software and hardware compatibility) do not raise new questions of safety and effectiveness compared to the predicate device.

    The "performance data (non-clinical)" section highlights that:

    • Hardware: Previous testing for biocompatibility, cleaning, debris, dimensional stability, and packaging is applicable. Accuracy and performance of the system were "adequate to perform as intended." Previous simulated surgeries and cadaver testing on earlier cleared devices are considered applicable.
    • Software: Software verification and validation were performed "against defined requirements" and "against user needs," following FDA guidance.

    Since this is a special 510(k) for an incremental change (adding compatibility with specific new implants), it's implied that the acceptance criteria are met if these additions do not negatively impact the established safety and effectiveness of the existing device, and the software development process meets regulatory standards.

    2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    • Hardware (previous testing cited): The document mentions "previous simulated surgeries using rapid prototyped bone models" and "previous cadaver testing on previously cleared devices K153602 and K131559." It does not specify the sample size for these previous studies (e.g., number of bone models or cadavers) or their provenance (country of origin), nor does it state if they were retrospective or prospective.
    • Software (verification and validation): The document states "Software verification and validation were performed," but does not specify a separate "test set" in the context of clinical data or specific performance metrics with sample sizes for this particular submission. The V&V activities would involve testing against requirements and user needs, which could include various test cases and scenarios, but these are not quantified here as a "test set" size.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    This information is not provided in the document. The studies cited for hardware ("simulated surgeries" and "cadaver testing") and software ("verification and validation") do not detail the involvement of experts in establishing ground truth, their number, or specific qualifications. The software's function is to assist surgeons in planning, implying surgeon input in its use, but not explicitly in establishing a ground truth for a test set described in this submission.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    This information is not provided in the document.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    The document does not mention any MRMC comparative effectiveness studies. The device (SurgiCase Shoulder Planner software component and Materialise Shoulder Guide and Models hardware component) is designed to assist surgeons in planning and component positioning, but the provided text does not contain data on whether human readers/surgeons improve with or without this specific AI assistance or effect sizes.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    The document states that the SurgiCase Shoulder Planner is "intended to be used as a pre-surgical planner for simulation of surgical interventions for shoulder orthopedic surgery. The software is used to assist in the positioning of shoulder components." It also mentions, "SurgiCase Shoulder Planner allows the surgeon to visualize, measure, reconstruct, annotate and edit pre-surgical plan data." This indicates that the software is a human-in-the-loop device, where the surgeon is actively involved in the planning process and responsible for approving the plan. Therefore, a standalone (algorithm only) performance assessment, without human input, is unlikely to be the primary method of evaluation described or required for this type of device. The document does not provide such standalone performance data.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    The document does not explicitly state the type of ground truth used for any specific test set related to this submission. For the hardware (guides and models), the "accuracy and performance" implies a comparison to a known standard or ideal, perhaps derived from anatomical models or surgical goals. For the software, "verification against defined requirements, and validation against user needs" suggests that the ground truth for V&V would be the successful adherence to these requirements and user expectations, which could involve internal expert review or adherence to pre-defined medical/engineering specifications. However, specific types of ground truth like pathology or long-term outcomes data are not mentioned.

    8. The sample size for the training set

    This information is not provided in the document. As the submission is for a special 510(k) updating compatibility, it's possible that the core algorithms were developed and trained previously, and details of their original training are not part of this specific submission. The focus here is on the impact of the changes to the device.

    9. How the ground truth for the training set was established

    This information is not provided in the document. Similar to the training set size, the specifics of how the ground truth was established for the original training of any underlying algorithms are not included in this special 510(k).

    Ask a Question

    Ask a specific question about this device

    K Number
    K230831
    Manufacturer
    Date Cleared
    2023-11-13

    (231 days)

    Product Code
    Regulation Number
    888.3670
    Why did this record match?
    Reference Devices :

    K212933, K182039, K092122, K113254

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Anatomic Total Shoulder or Hemi-Shoulder
    The INHANCE SHOULDER SYSTEM with the humeral stemless anchor is intended for use in anatomic total shoulder replacement procedures to address the following:

    • Osteoarthritis
    • Post-traumatic arthrosis
    • Focal avascular necrosis of the humeral head
    • Previous surgeries of the shoulder that do not compromise the fixation

    The INHANCE SHOULDER SYSTEM with a humeral stem is intended for use in anatomic total or hemi-shoulder replacement procedures to address the following:

    • Non-inflammatory degenerative joint disease including osteoarthritis and avascular necrosis.
    • Rheumatoid arthritis.
    • Revision where other devices or treatments have failed.
    • Correction of functional deformity.
    • Fractures of the humeral head (with Short Humeral Stems).
    • Fractures of the humeral head and proximal humerus, where other methods of treatments are deemed inadequate (with Standard or Long Stems).
    • Difficult clinical management problems where other methods of treatment may not be inadequate.

    Reverse Total Shoulder
    The INHANCE SHOULDER SYSTEM Reverse Total Shoulder with a humeral stem is indicated for primary, fracture or revision total reverse shoulder replacement procedures the following. The system is indicated for use in patients whose shoulder joint has a gross rotator cuff deficiency. The patient must be anatomically and structurally suited to receive the implants and a functional deltoid muscle is necessary. The system is also indicated from an anatomic to reverse shoulder prosthesis without the removal of a well-fixed INHANCE humeral stem.

    • A severely painful, disabling, arthritic joint
    • Fractures of the humeral head (with Short Humeral Stems)
    • Fractures of the humeral head and proximal humerus (with Standard or Long Stems)
    • Revisions of previously failed shoulder joint replacements
    Device Description

    The INHANCE™ Convertible Glenoid system consists of a Baseplate with minor modifications compared to the previously cleared Baseplate in K212737 and a poly insert that locks atop the Baseplate to allow for an anatomic procedure. The Convertible Glenoid Insert implants are offered in four sizes: Small (24.0mm), Medium (26.5mm), Large (29.0mm), and X-Large (31.5mm). The Convertible Glenoid Insert implants consist of a Cross-linked, Vitamin E Ultra High Molecular Weight Polyethylene (Cross-linked, VE UHMWPE) articulation surface and an interrupted fixation ring along with a finned central fixation post to facilitate poly locking to the Convertible Glenoid Baseplate.

    The INHANCETM Convertible Glenoid Insert Implants have a lateral surface that is concave and designed to articulate with the Humeral Heads from the INHANCE Anatomic Stemmed and Stemless Shoulder Systems that are indicated for use in total shoulder arthroplasty.

    The INHANCE™ Convertible Glenoid Baseplates, previously cleared in K212737, were modified to provide for a means of mechanical fixation between the Convertible Glenoid Inserts and Baseplates.

    The INHANCE Convertible Glenoid implants are compatible with the implants and instruments previously cleared for use in the INHANCE Anatomic Shoulder System (K202716), the INHANCE Stemless Anatomic Shoulder System (K203108), and the INHANCE Reverse Shoulder System (K212737).

    AI/ML Overview

    This document describes the regulatory approval for the INHANCE™ Convertible Glenoid device. It is a medical device, specifically a component of a shoulder arthroplasty system. Therefore, the "acceptance criteria" and "device performance" refer to the successful completion of specific non-clinical (mechanical, material, and biocompatibility) tests according to established standards, rather than the performance of an AI algorithm based on clinical data.

    Here's a breakdown of the requested information based on the provided text:


    Acceptance Criteria and Device Performance

    1. Table of Acceptance Criteria and Reported Device Performance

    Test/Evaluation CategoryAcceptance Criteria (Implied by Standard)Reported Device Performance
    Range of Motion (RoM)Meet established specifications per ASTM F1378."The RoM targets were met."
    Biocompatibility AssessmentsFound to be biocompatible per ISO 10993-1 and FDA Guidance Document Use of International Standard ISO 10993-1."The devices were found to be biocompatible."
    Porous Structure CharacterizationIdentical to previously cleared porous structures (K202716 and K203108)."The porous structure used for the subject device is identical to the porous structure that was applied to the implants cleared under K202716 and K203108."
    Characterization of AO-HXLPEConforms to ASTM F2695, and identical in base resin, blending concentration of antioxidant, and crosslinking irradiation dose to VE UHMWPE material in K202716."The antioxidant highly crosslinked UHMWPE (AO-HXLPE) was fully characterized and conforms to ASTM F2695. The Vitamin E Ultra High Molecular Weight Polyethylene (Crosslinked, VE UHMWPE) material used for the INHANCE™ Convertible Glenoids is identical in base resin, blending concentration of antioxidant, and crosslinking irradiation dose to the VE UHMWPE material that was used on the devices cleared under K202716."
    Evaluation of Glenoid Wear RateDoes not represent a new worst-case for wear compared to predicate devices."An engineering justification demonstrated that the subject device does not represent a new worst-case for wear of the articulating surfaces when compared to the predicate devices."
    Dynamic Evaluation of Convertible Glenoid Loosening/DisassociationMeet acceptance criteria per ASTM F2028."The acceptance criteria were met."
    Static Evaluation of Anatomic Glenoid Locking Mechanism in ShearMeet acceptance criteria per ASTM F1829."The acceptance criteria were met."
    Glenoid Fatigue Resistance EvaluationMeet acceptance criteria for dissociation of VE XLPE Glenoid Inserts and Ti6Al4V Baseplate."The acceptance criteria were met."
    MRI CompatibilityQuantitative data obtained per ASTM F2052-15 (Force), ASTM F2213-17 (Torque), ASTM F2182-19e2 (Heating), ASTM F2119-07 (Image Quality) to inform labeling."Quantitative data was obtained to inform Magnetic Resonance Imaging (MRI) Conditional Labeling through the following evaluations: Force: Static Magnetic Field Induced Displacement Force per ASTM F2052-15; Torque: Static Magnetic Field Induced Torque per ASTM F2213-17; Heating: Radiofrequency field (RF) induced heating per ASTM F2182-19e2; Image Quality: Susceptibility induced image artifacts per ASTM F2119-07" (Implied acceptance by informing conditional labeling).
    Shelf-Life EvaluationEstablish a specified shelf life per ISO 11607-1 and ISO 11607-2."A five-year shelf life was established based on the resultant data."
    Sterilization ValidationAchieve a Sterility Assurance Level (SAL) of 10-6 using the VDmax method described in ISO 11137-1 and ISO 11137-2."Sterilization validation was completed using the VDmax method specified in ISO 11137-1 and ISO 11137-2. The Sterility Assurance Level (SAL) was found to be 10-9." (This exceeds the typical 10-6 requirement, indicating successful validation.)

    2. Sample Size Used for the Test Set and the Data Provenance

    This document describes the regulatory approval for a medical device (orthopedic implant), not an AI algorithm. Therefore, the concepts of "test set" and "data provenance" specifically for AI evaluation are not applicable in this context. The "test sets" here refer to the samples of the physical device or its materials subjected to various non-clinical (mechanical, material, biological) tests as outlined above. The provenance of these test articles would be the manufacturing process of the INHANCE™ Convertible Glenoid components.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts

    This information is not applicable. For physical medical devices, "ground truth" is established by adherence to recognized engineering standards (e.g., ASTM, ISO), which dictate testing methodologies and performance thresholds. These standards are developed and maintained by expert committees in relevant engineering and medical fields, but there isn't a "number of experts" establishing ground truth for a specific test set in the same way clinical data is evaluated for AI.

    4. Adjudication Method for the Test Set

    This information is not applicable. "Adjudication method" usually refers to resolving disagreements among human reviewers (e.g., radiologists) in AI or clinical studies. In the context of device testing, results are typically objective measurements against a standard, which are then analyzed and interpreted by qualified engineers/scientists.

    5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    This information is not applicable. An MRMC study is relevant to AI/diagnostic imaging devices. This document concerns the mechanical and material performance of an orthopedic implant.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

    This information is not applicable. This is not an AI algorithm.

    7. The Type of Ground Truth Used (expert consensus, pathology, outcomes data, etc.)

    For this medical device, the "ground truth" is based on:

    • Adherence to established international and national standards (e.g., ASTM F1378, ISO 10993-1, ASTM F2028, ASTM F1829, ASTM F2695, ISO 11607-1, ISO 11607-2, ISO 11137-1, ISO 11137-2).
    • Engineering justifications and characterizations demonstrating equivalence to predicate devices and acceptable performance (e.g., non-worst-case for wear, identical material properties).
    • Quantitative data derived from physical testing.

    8. The Sample Size for the Training Set

    This information is not applicable. There is no AI algorithm and thus no training set.

    9. How the Ground Truth for the Training Set Was Established

    This information is not applicable. There is no AI algorithm and thus no training set or ground truth for it.


    Summary of Device-Specific Information:

    The INHANCE™ Convertible Glenoid is a shoulder arthroplasty component. Its acceptance for market (substantial equivalence to predicate devices) was based on a comprehensive series of non-clinical tests and evaluations covering:

    • Range of Motion
    • Biocompatibility
    • Porous Structure Characterization
    • AO-HXLPE Material Characterization
    • Glenoid Wear Rate Evaluation
    • Dynamic Evaluation of Loosening/Disassociation
    • Static Evaluation of Locking Mechanism
    • Glenoid Fatigue Resistance
    • MRI Compatibility
    • Shelf-Life
    • Sterilization Validation

    All these tests met their respective acceptance criteria as defined by national and international standards (ASTM, ISO) and FDA guidance, ensuring the device's safety and effectiveness. Clinical testing was explicitly stated as "not necessary to demonstrate substantial equivalence."

    Ask a Question

    Ask a specific question about this device

    Why did this record match?
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Hardware: The Materialise Shoulder Guide and Models are intended to be used as a surgical instrument to assist in the intraoperative positioning of glenoid components used with total and reverse shoulder arthroplasty by referencing anatomic landmarks of the shoulder that are identifiable on preoperative CT-imaging scans. The Materialise Shoulder Guide and Models are single use only. The Materialise Shoulder Guide and Models can be used in conjunction with the following total and reverse shoulder implants systems and their respective compatible components: DePuy Synthes', Enovis' (DJO), Smith+Nephew's, Lima's, Stryker's.

    Software: SurgiCase Shoulder Planner is intended to be used as a pre-surgical planner for simulation of surgical interventions for shoulder orthopedic surgery. The software is used to assist in the positioning of shoulder components. SurgiCase Shoulder Planner allows the surgeon to visualize, measure, reconstruct, annotate and edit pre-surgical plan data. The software leads to the generation of a surgery report along with a pre-surgical plan data file which can be used as input data to design the Materialise Shoulder Guide and Models.

    Device Description

    Materialise Shoulder System™ is a patient-specific medical device that is designed to be used to assist the surgeon in the placement of shoulder components during total anatomic and reverse shoulder replacement surgery. This can be done by generating a pre-surgical shoulder plan and, if requested by the surgeon, by manufacturing a patient-specific glenoid guide and models to transfer the glenoid plan to surgery. The device is a system composed of the following: a software component, branded as SurgiCase Shoulder Planner. This software is a planning tool used to generate a pre-surgical plan for a specific patient. Materialise Shoulder Guide and Models, which are a patient-specific guide and models that are based on a pre-surgical plan. This pre-surgical plan is generated using the software component. Patient-specific glenoid guide and models will be manufactured if the surgeon requests patient-specific guides to transfer the glenoid plan to surgery. The Materialise Shoulder Guide is designed and manufactured to fit the anatomy of a specific patient. A bone model of the scapula is delivered with the Materialise Shoulder Guide. A graft model can be delivered with the Materialise Shoulder Guide. The graft model visualizes the graft-space between implant and bone, based on the pre-operative planning of the surgeon. The graft model serves as a visual reference for the surgeon in the OR.

    AI/ML Overview

    The provided text describes the regulatory clearance for the Materialise Shoulder System™ and mentions performance data, but it does not contain a detailed study proving the device meets specific acceptance criteria in the format requested.

    The document is a 510(k) summary, which focuses on demonstrating substantial equivalence to a legally marketed predicate device rather than presenting a comprehensive standalone study with detailed effectiveness metrics.

    Here's an analysis of the information that can be extracted, and what is missing based on your request:

    1. Table of Acceptance Criteria and Reported Device Performance:

    • Acceptance Criteria: Not explicitly stated in a quantitative manner for the performance of the AI component (SurgiCase Shoulder Planner) or the hardware (Materialise Shoulder Guide and Models) beyond ensuring it performs "as intended" and maintains accuracy.
    • Reported Device Performance: The document states that "Testing verified that the accuracy and performance of the system is adequate to perform as intended." However, no specific performance metrics (e.g., accuracy, precision, sensitivity, specificity, or error margins) are provided for either the software for planning or the hardware for guiding.

    2. Sample size used for the test set and the data provenance:

    • Test Set Sample Size: Not specified for the software component (SurgiCase Shoulder Planner).
    • Data Provenance: Not specified for any software testing.
    • For Hardware: It refers to "simulated surgeries using rapid prototyped bone models and previous cadaver testing." No specific number of models or cadavers is provided, nor is the country of origin or whether it was retrospective or prospective.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • Number of Experts: Not specified.
    • Qualifications of Experts: Not specified.

    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

    • Not specified.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • No MRMC study is mentioned. The document focuses on the planning and guidance aspect for surgeons, but not on a comparative study of human readers (surgeons) with and without AI assistance for diagnosis or planning accuracy.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    • A standalone performance evaluation of the software's planning functionality is implied through the statement "Software verification and validation were performed, and documentation was provided following the 'Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices.'" However, no specific metrics or study details are provided. The software is described as a "planning tool used to generate a pre-surgical plan" and for "assisting the surgeon in positioning shoulder components," suggesting a human-in-the-loop workflow.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    • Ground Truth Type: Not explicitly stated for the software. For the hardware, the use of "rapid prototyped bone models and previous cadaver testing" implies a physical ground truth for accuracy validation.

    8. The sample size for the training set:

    • The document does not mention a training set sample size, which is typical for AI/ML models. This suggests the software functionality described (planning, visualization, measurement) might not be based on deep learning/machine learning that requires explicit training data in the same way as, for example, an image classification algorithm. It seems to be a rules-based or physics-based planning software.

    9. How the ground truth for the training set was established:

    • As no training set is mentioned, the method for establishing its ground truth is also not provided.

    In summary:

    The provided text from the FDA 510(k) summary states that non-clinical performance testing indicates the device is as safe and effective as its predicate. It mentions software verification and validation and previous hardware testing (biocompatibility, sterility, cleaning, debris, dimensional stability, packaging, simulated surgeries, cadaver testing). However, it lacks the specific quantitative acceptance criteria and detailed study results (such as sample sizes, expert qualifications, clear performance metrics, ground truth establishment for software, and formal comparative effectiveness study results) that are typically expected when describing a study proving specific acceptance criteria in detail. This information is usually found in separate, more detailed technical documentation submitted to the FDA, not in the public 510(k) summary.

    Ask a Question

    Ask a specific question about this device

    K Number
    K221758
    Date Cleared
    2023-03-17

    (274 days)

    Product Code
    Regulation Number
    888.3660
    Why did this record match?
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The SMR Stemless Anatomic is indicated for total primary or revision shoulder joint replacement in patients suffering from disability due to:

    · non-inflammatory degenerative joint disease including osteoarthritis;

    • revision of previous surgeries of the shoulder that do not compromise the fixation (such as a failed SMR resurfacing implant);

    · glenoid arthrosis without excessive glenoid bone loss: A1, A2 and B1 according to Walch classification (SMR TT Hybrid Glenoid only).

    The SMR Stemless Anatomic is intended for uncemented use.

    Device Description

    The SMR Stemless Anatomic is a modular system comprised of a stemless core and humeral head adaptor taper. The modular components are available in various sizes and are interchangeable allowing for independent sizing and positioning. The SMR humeral heads were previously cleared (K161476, K100858), and the SMR Stemless Anatomic is compatible with the previously cleared Cemented SMR metal back Glenoid Components (K113254, K133349, K143256), Cemented SMR all polyethylene glenoid components (K100858, K130642, K153722), and SMR TT Hybrid Glenoid System (K163397).

    AI/ML Overview

    This document is a 510(k) clearance letter from the FDA for a medical device called the "SMR Stemless Anatomic." It's not a study report of an AI/ML powered medical device, and therefore does not contain the information requested in the prompt regarding acceptance criteria, study methodologies for AI performance, sample sizes, expert qualifications, or ground truth establishment relevant to an AI/ML product.

    The document discusses performance testing for a mechanical orthopedic implant, specifically a shoulder joint prosthesis, covering aspects like fatigue, micromotion, and pull-out strength. It also mentions "Clinical Data" related to the device's success in patients, comparing it to a predicate device. This is a traditional medical device clearance, not an AI/ML software as a medical device (SaMD) or AI-powered medical device.

    Therefore, I cannot extract the requested information (table of acceptance criteria with AI performance, sample sizes for AI test sets, expert details for AI ground truth, MRMC studies, etc.) from this document. The concepts and methodologies described in the prompt are specific to the evaluation and clearance of AI/ML-driven medical devices, which this document does not concern.

    Ask a Question

    Ask a specific question about this device

    K Number
    K223876
    Date Cleared
    2023-02-03

    (42 days)

    Product Code
    Regulation Number
    888.3670
    Reference & Predicate Devices
    Why did this record match?
    Reference Devices :

    K212800, K210717, K113254, K133349, K163397

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The SMR Shoulder System is intended for partial or total, primary or revision shoulder joint replacement.

    The SMR Anatomic Shoulder System is indicated for partial or total, primary or revision shoulder joint replacement in patients suffering from disability due to:

    • non-inflammatory degenerative joint disease including osteoarthritis and avascular necrosis;
    • inflammatory degenerative joint disease such as rheumatoid arthritis; ●
    • treatment of acute fractures of the humeral head that cannot be treated with other fracture fixation methods;
    • revision of a failed primary implant; in case of SMR Short Stems only if sufficient bone stock remains);
    • . cuff tear arthropathy (CTA Heads only);
    • glenoid arthrosis without excessive glenoid bone loss: A1, A2 and B1 according to Walch classification (SMR TT Hybrid ● Glenoid only).

    The SMR Reverse Shoulder System is indicated for primary, fracture or revision total shoulder replacement in a grossly rotator cuff deficient ioint with severe arthropathy (disabled shoulder). The patient's ion t must be anatomically suited to receive the selected implants and a functional deltoid muscle is neces sary to use the device.

    The SMR TT Hybrid Glenoid Reverse Baseplate must not be used in cases of excessive glenoid bone loss and/or when bone graft is needed.

    Device Description

    The SMR Shoulder System is a complete system intended to be used in primary or revision total shoulder joint replacement in either anatomic or reverse configurations. The SMR Shoulder System was cleared via several 510(k) submission, up to the latest approval under K220792.

    The new compatibilities introduced with this 510(k) are related to the system when used in reverse shoulder configuration only.

    The SMR Reverse Shoulder System is indicated for primary, fracture or revision total shoulder replacement in a grossly rotator cuff deficient joint with severe arthropathy (disabled shoulder). The patient's joint must be anatomically and structurally suited to receive the selected implants and a functional deltoid muscle is necessary to use the device. The SMR Shoulder System in reverse configuration consists of humeral stems, reverse humeral bodies, reverse liners, glenospheres, metal back glenoid components and pegs. The metal back glenoid component, when used as part of a reverse shoulder replacement, is intended for cementless fixation with bone screws.

    Glenoid components of the SMR Shoulder System are cleared for use also with the humeral components of the PRIMA Humeral System (K212800). When used in combination with the PRIMA Humeral System, the device consists of a humeral stem, tray, reverse insert, glenospheres, metal back glenoid components and pegs.

    With this Traditional 510(k) submission, new compatibilities between already cleared devices of the SMR Shoulder System are introduced:

    • Bone Screws dia. 5.0 mm (K210717) compatible with SMR Metal Back Glenoid -(K113254) and SMR TT Baseplate (K133349);
    • -SMR TT Hybrid Glenoid Reverse Baseplate (K163397) compatible with SMR Glenosphere dia. 42 mm (K212800).

    No changes in indications for use, materials, manufacturing processes, packaging and sterilization are introduced with this 510(k) on already cleared devices.

    AI/ML Overview

    The provided document is a 510(k) summary for the SMR Shoulder System, a medical device. It focuses on demonstrating substantial equivalence to a predicate device based on non-clinical testing and comparison of technological characteristics. It does not contain information about acceptance criteria or a study that proves the device meets those criteria in the context of an AI/ML device.

    This document specifically states: "Clinical testing was not necessary to demonstrate substantial equivalence of the SMR Shoulder System to the predicate device." This indicates that the device's performance was evaluated through non-clinical means (mechanical tests), not through studies involving clinical data or AI performance metrics.

    Therefore, I cannot provide the requested information for acceptance criteria and a study proving those criteria, as the document does not describe such a study for an AI/ML device.

    However, I can extract the information related to the device and its testing as presented:

    Device: SMR Shoulder System

    1. A table of acceptance criteria and the reported device performance:

    Acceptance CriteriaReported Device Performance
    Fulfilled intended useMechanical tests demonstrated that device performance fulfilled the intended use.
    Substantially equivalent to predicate devicesMechanical tests demonstrated that the device is substantially equivalent to the predicate devices.
    Dynamic Evaluation of Glenoid Loosening or Disassociation (ASTM F2028)Performed on worst-case components. (Specific results or acceptance thresholds not provided in this summary.)

    2. Sample size used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective):

    • Sample Size: Not specified for mechanical testing. The phrase "worst case components" suggests a targeted selection rather than a large statistical sample.
    • Data Provenance: Not applicable. The testing was non-clinical (mechanical).

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • Not applicable. This was a non-clinical mechanical test, not an evaluation requiring expert-established ground truth for an AI/ML device.

    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

    • Not applicable. This was a non-clinical mechanical test, not an evaluation requiring adjudication for an AI/ML device.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • No, a multi-reader multi-case (MRMC) comparative effectiveness study was not done. The document explicitly states: "Clinical testing was not necessary to demonstrate substantial equivalence." This type of study is relevant for AI/ML devices involving human readers interpreting clinical data, which is not the case here.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    • Not applicable. This device is a shoulder implant system, not a software algorithm or AI. The testing was on the physical components.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    • Not applicable in the context of clinical or AI/ML ground truth. The "ground truth" for non-clinical mechanical testing would be the engineering specifications and performance standards (e.g., ASTM F2028).

    8. The sample size for the training set:

    • Not applicable. This is not an AI/ML device; therefore, there is no training set.

    9. How the ground truth for the training set was established:

    • Not applicable. This is not an AI/ML device; therefore, there is no training set or associated ground truth establishment.
    Ask a Question

    Ask a specific question about this device

    Why did this record match?
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Materialise Shoulder Guide and Models are intended to be used as a surgical instrument to assist in the intraoperative positioning of glenoid components used with total and reverse shoulder arthroplasty by referencing anatomic landmarks of the shoulder that are identifiable on preoperative CT-imaging scans.

    The Materialise Shoulder Guide and Models are single use only.

    The Materialise Shoulder Guide and Models can be used in conjunction with the following total and reverse shoulder implants systems and their respective compatible components:

    SurgiCase Shoulder Planner is intended to be used as a pre-surgical planner for simulation of surgical interventions for shoulder orthopedic surgery. The software is used to assist in the positioning of shoulder components. SurgiCase Shoulder Planner allows the surgeon to visualize, measure, reconstruct, annotate and edit pre-surgical plan data. The software leads to the generation of a surgery report along with a pre-surgical plan data file which can be used as input data to design the Materialise Shoulder Guide and Models.

    Device Description

    Materialise Shoulder System™ is a patient-specific medical device that is designed to be used to assist the surgeon in the placement of shoulder components during total anatomic and reverse shoulder replacement surgery. This can be done by generating a pre-surgical shoulder plan and, if requested by the surgeon, by manufacturing a patient-specific glenoid guide and models to transfer the glenoid plan to surgery. The device is a system composed of the following:

    a software component, branded as SurgiCase Shoulder Planner. This software is . a planning tool used to generate a pre-surgical plan for a specific patient.

    Materialise Shoulder Guide and Models, which are a patient-specific quide and . models that are based on a pre-surgical plan. This pre-surgical plan is generated using the software component. Patient-specific glenoid guide and models will be manufactured if the surgeon requests patient-specific quides to transfer the glenoid plan to surgery. The Materialise Shoulder Guide is designed and manufactured to fit the anatomy of a specific patient. A bone model of the scapula is delivered with the Materialise Shoulder Guide. A graft model can be delivered with the Materialise Shoulder Guide. The graft model visualizes the graft-space between implant and bone, based on the pre-operative planning of the surgeon. The graft model serves as a visual reference for the surgeon in the OR.

    AI/ML Overview

    I am sorry, but the provided text primarily consists of an FDA 510(k) clearance letter and summary for the Materialise Shoulder System. While it describes the device, its indications for use, and technological characteristics compared to a predicate device, crucial details regarding the acceptance criteria and the study proving the device meets those criteria (such as specific performance metrics, sample sizes for test sets, ground truth establishment, or clinical study designs like MRMC studies) are not present in the provided document.

    The "Performance Data (non-clinical)" section briefly mentions software verification and validation, but it does not provide the specific "acceptance criteria" for performance or the details of the "study that proves the device meets the acceptance criteria" in the format requested. It mainly relies on demonstrating substantial equivalence to a predicate device based on similar technology and previous testing.

    Therefore, I cannot fulfill your request to describe the acceptance criteria and the study proving the device meets them using only the information given in the input text. The information required for the table and the detailed study description is simply not provided in this regulatory document.

    Ask a Question

    Ask a specific question about this device

    K Number
    K201905
    Date Cleared
    2020-12-17

    (161 days)

    Product Code
    Regulation Number
    888.3670
    Why did this record match?
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The SMR Shoulder System is intended for partial or total, primary or revision shoulder joint replacement.

    The SMR Anatomic Shoulder System is indicated for partial or total, primary or revision shoulder joint replacement in patients suffering from disability due to:

    • non-inflammatory degenerative joint disease including osteoarthritis and avascular necrosis;
    • . inflammatory degenerative joint disease such as rheumatoid arthritis;
    • treatment of acute fractures of the humeral head that cannot be treated with other fracture fixation methods;
    • revision of a failed primary implant; in case of SMR Short Stems only if sufficient bone stock remains); .
    • cuff tear arthropathy (CTA Heads only); .
    • glenoid arthrosis without excessive glenoid bone loss: A1, A2 and B1 according to Walch classification (SMR TT Hybrid Glenoid only). .

    The SMR Reverse Shoulder System is indicated for primary, fracture or revision total shoulder replacement in a grossly rotator coff deficient joint with severe arthropathy (disabled shoulder). The patients joint must be anatomically suited to receive the selected implants and a functional deltoid muscle is necessary to use the device.

    The SMR TT Hybrid Glenoid Reverse Baseplate must not be used in cases of excessive glenoid bone loss and/or when bone graft is needed.

    The Modular SMR Shoulder System allows the assembly of components in various humeral and glenoid constructs are intended for cemented and uncemented use as specified in the following table.

    In the Anatomic shoulder the humeral consists of the humeral stem, the humeral body, the adaptor taper and the humeral head. In the Reverse shoulder the humeral consists of the humeral stem, the reverse humeral body and the reverse liner. On the humeral side the fixation of the humeral stem determines if the construct is cemented or uncemented.

    The Anatomic glenoid construct consists of an all polyethylene glenoid with metal peg or a metal back assembled with a liner; the Reverse glenoid consists of a metal back/connector/glenosphere construct or of a peg/baseplate/glenosphere construct. On the glenoid side, the fixation of the all polyethylene glenoid with metal peg or the metal back determines if the construct is cemented or uncemented.

    Device Description

    The SMR 140° Reverse Humeral Bodies (SMR 140° Reverse Humeral Body, SMR 140° Short Reverse Humeral Body, SMR 140° Finned Reverse Humeral Body) are manufactured from Ti6Al4V alloy (ASTM F1472 - ISO 5832-3); they are for tapercoupling with previously cleared Humeral extension (K113523), SMR CTA Head Adaptor for Reverse Humeral Body (K131112), Reverse Liners (K110598, K142139) and Humeral stems (K100858, K101263, K111212, K191963).

    AI/ML Overview

    This document is a 510(k) premarket notification for the SMR 140° Reverse Humeral Body, a component of a shoulder replacement system. As such, it does not contain information about acceptance criteria or a study proving that an AI device meets acceptance criteria.

    The document describes a medical device (SMR 140° Reverse Humeral Body for shoulder replacement) and its substantial equivalence to predicate devices, supported by non-clinical mechanical testing. It explicitly states that clinical testing was not necessary to demonstrate substantial equivalence.

    Therefore, the requested information points regarding AI acceptance criteria and studies (sample size, data provenance, expert ground truth, adjudication, MRMC studies, standalone performance, ground truth type for training/test sets, and training sample size) are not applicable to this document.

    Ask a Question

    Ask a specific question about this device

    Why did this record match?
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Hardware: The Materialise Glenoid Positioning System is intended to be used as a surgical instrument to assist in the intraoperative positioning of glenoid components used with total and reverse shoulder arthroplasty by referencing anatomic landmarks of the shoulder that are identifiable on preoperative CT-imaging scans.

    Software: SurgiCase Shoulder Planner is intended to be used as a pre-surgical planner for simulation of surgical interventions for shoulder orthopedic surgery. The software is used to assist in the positioning of glenoid components. SurgiCase Shoulder Planner allows the surgeon to visualize, measure, reconstruct, annotate and edit pre-surgical plan data. The software leads to the generation of a surgery report along with a pre-surgical plan data file which can be used as input data to design the Materialise Glenoid Positioning Guide and Models.

    Device Description

    Materialise Glenoid Positioning System is a patient-specific medical device that is designed to assist the surgeon in the placement of glenoid components.

    This can be done by generating a pre-surgical plan or by generating a pre-surgical plan and manufacturing a patientspecific guide and models to transfer the plan to surgery. The device is a system composed of the following:

    • a software component, branded as SurgiCase Shoulder Planner. This software is a planning tool used to ● generate a pre-surgical plan for a specific patient.
    • a hardware component, branded as the Materialise Glenoid Positioning System™ Guide and Models, which is a patient specific guide and models that are based on a pre-surgical plan. This pre-surgical plan is generated using the software component. Patient-specific guide and models will be manufactured if the surgeon requests patient-specific guides to transfer the plan to surgery. The Materialise Glenoid Positioning System Guide is designed and manufactured to fit the anatomy of a specific patient. A bone model of the scapula is delivered with the Materialise Glenoid Positioning System Guide. A graft model can be delivered with the Materialise Glenoid Positioning System Guide.
    AI/ML Overview

    The provided document (K190286) is a 510(k) Premarket Notification for the Materialise Glenoid Positioning System. This document describes the device, its intended use, and claims substantial equivalence to a predicate device (K172054). However, it does not contain the detailed acceptance criteria and the study results proving the device meets these criteria in the format requested.

    The "Performance Data" section (page 5-6) explicitly states:
    "Previous testing for biocompatibility, cleaning, debris, dimensional stability and packaging are applicable to the subject device and demonstrate substantial equivalence with the predicate device. Testing verified that the accuracy and performance of the system is adequate to perform as intended. The stability of the device placement, surgical technique, intended use and functional elements of the same as that of the predicate Materialise Glenoid Positioning System K172054 and previously cleared devices K170893, K1536559, and therefore previous cadaver testing on previously cleared devices K153602 and K131559 is considered applicable to the subject device."

    This statement indicates that performance testing was conducted, and its results were deemed adequate. However, the details of those tests, including specific acceptance criteria, reported performance values, sample sizes, ground truth establishment, expert qualifications, and adjudication methods, are not present in this document. The document primarily relies on demonstrating substantial equivalence to previously cleared devices and states that "previous cadaver testing on previously cleared devices K153602 and K131559 is considered applicable to the subject device." This suggests that the detailed study results might be found in the 510(k) submissions for those predicate devices, rather than being explicitly laid out in this current submission.

    Therefore, I cannot populate the table or provide specific details for most of your questions based solely on the text provided in K190286. The document focuses on regulatory justification for substantial equivalence, not a detailed scientific report of device performance trials.