(259 days)
The SMR Shoulder system is intended for partial or total primary shoulder joint replacement. The components are intended for use in cemented applications.
Total or hemi-shoulder replacement is indicated for patients suffering from disability due to:
- Non-inflammatory degenerative joint disease including osteoarthritis and avascular necrosis:
- Inflammatory degenerative joint disease such as rheumatoid arthritis;
- Treatment of acute fractures of the humeral head that cannot be treated with other fracture fixation methods
The SMR Shoulder System consists of a humeral stem, a humeral body, an adaptor taper, a humeral head and a glenoid component. Components are offered for cemented use for hemi or total shoulder joint arthroplasty, in primary trauma surgery.
Humeral stems are designed for cemented use and are characterized by an outline with a triple conicity to provide optimal fixation and fit independent of canal morphology. The stems are made from Ti6Al4V (ISO 5832-3, ASTM F1472). The surface in contact with cement is polished to reduce shear stresses on the cement mantle and avoid abrasion resulting from micromovements at the stem-cement interface. The distal part of the stems is characterized by a cylindrical cross-section while the proximal part is fluted to prevent rotation of the component relative to bone. The stems are provided with a male Morse taper in the proximal part for coupling with the humeral body.
Humeral bodies are made from Ti6Al4V and are coupled with the humeral stem via a female Morse-taper connection; a locking screw is provided to aid in initial mating of the stem / body assembly. Cylindrical marks are designed at the base of the Morse-taper to provide correct alignment of the eccentricity of the humeral head during surgery. A male Morsetaper connection is designed for the coupling between the humeral body and the humeral head by means of specific adaptor tapers: an angle of 45° between the axis of this Morsetaper and the axis of the stem gives the correct varus-valgus alignment to the joint. A system of holes on the external surface allows anatomical attachment of tuberosities using sutures or wires in trauma applications.
Adaptor tapers (neutral and eccentrical with different heights), are made from Ti6Al4V. They allow coupling between the humeral body and the humeral head. These devices are designed to adjust the centre of rotation of the joint and provide the required offset to the humeral head to achieve the correct tensioning to the soft tissues, optimizing joint stability.
The humeral heads are made from CoCrMo (ISO 5832-12, ASTM F1537). They are intended to articulate with the glenoid bone in hemi-arthroplasty or with the glenoid component in total shoulder joint arthroplasty. The surface is polished to aid in the reduction of wear.
The glenoid components are manufactured from Ultra-High Molecular Weight Polyethylene (UHMWPE ISO 5834-2, ASTM F648). The articulating surface has a radius of curvature greater than the corresponding humeral head, which allows translation in the superior/inferior and anterior/posterior directions. The back surface of the component is spherical in geometry and has a single central peg which is inserted in the hole drilled in the glenoid cavity during surgery. The peg surface has three grooves to provide enhanced cement fixation.
Here's an analysis of the provided text regarding the acceptance criteria and study for the SMR Shoulder System:
The provided document, K100858 for the SMR Shoulder System, is a Traditional 510(k) premarket notification. Traditional 510(k)s typically demonstrate substantial equivalence to a predicate device through non-clinical performance data. They generally do not require extensive clinical studies with specified acceptance criteria in the same way a PMA (Pre-Market Approval) or certain De Novo applications might.
Based on the provided text, the "acceptance criteria" are implicitly met by demonstrating that the device performs comparably to the predicate devices and meets established performance standards for orthopedic implants.
Here's a breakdown of the information requested, based solely on the provided document:
1. A table of acceptance criteria and the reported device performance
| Acceptance Criteria (Implied from testing) | Reported Device Performance |
|---|---|
| Static pull-out strength of modular connections | "All mechanical testing was done on worst case components or constructs... The testing results demonstrated the device's ability to perform under expected clinical conditions." |
| Torsional testing of modular connections | "All mechanical testing was done on worst case components or constructs... The testing results demonstrated the device's ability to perform under expected clinical conditions." |
| Fatigue strength of humeral stem | "All mechanical testing was done on worst case components or constructs... The testing results demonstrated the device's ability to perform under expected clinical conditions." |
| Post-fatigue strength of modular connections | "All mechanical testing was done on worst case components or constructs... The testing results demonstrated the device's ability to perform under expected clinical conditions." |
| Static shear strength of glenoid component | "All mechanical testing was done on worst case components or constructs... The testing results demonstrated the device's ability to perform under expected clinical conditions." |
| Torsional strength of glenoid component | "All mechanical testing was done on worst case components or constructs... The testing results demonstrated the device's ability to perform under expected clinical conditions." |
| Fatigue strength of glenoid component | "All mechanical testing was done on worst case components or constructs... The testing results demonstrated the device's ability to perform under expected clinical conditions." |
| Material Properties (e.g., Ti6Al4V, CoCrMo, UHMWPE conforming to ISO/ASTM standards) | Components are manufactured from specified materials (Ti6Al4V, CoCrMo, UHMWPE) conforming to ISO and ASTM standards (e.g., ISO 5832-3, ASTM F1472 for stems; ISO 5832-12, ASTM F1537 for heads; ISO 5834-2, ASTM F648 for glenoids). This implicitly means they meet the properties defined by those standards. |
| Clinical performance (implied by substantial equivalence to predicates) | "Clinical testing was not necessary to demonstrate substantial equivalence of the SMR Shoulder System to the predicate device(s)." The device is deemed substantially equivalent based on intended use, indications, design, and materials. |
2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)
- Test Set Sample Size: The document only mentions "worst case components or constructs" were used for non-clinical (mechanical) testing. It does not specify the exact number of samples for each test.
- Data Provenance: The document does not explicitly state the country of origin for the non-clinical test data. The manufacturer is Lima-Lto S.p.A. in Italy, suggesting the testing may have been conducted in Italy or by a contracted lab. The testing is prospective in nature, as it involves physical testing of newly manufactured components.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)
- Not Applicable. This information is relevant for studies involving human interpretation (e.g., image analysis, diagnostic algorithms). The provided document describes non-clinical mechanical testing of an orthopedic implant. Ground truth for mechanical tests is established by physical measurement against engineering specifications and industry standards, not by expert consensus in the diagnostic sense.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
- Not Applicable. As explained above, this document describes non-clinical mechanical testing, not a study requiring human adjudication of results.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
- Not Applicable. This document describes a medical device (shoulder implant), not an AI-assisted diagnostic tool. No MRMC study was performed, and thus no effect size related to AI assistance is mentioned.
6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done
- Not Applicable. This document describes a medical device (shoulder implant), not an algorithm or AI system. Therefore, no standalone algorithm performance study was relevant or conducted.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)
- The "ground truth" for the non-clinical testing appears to be engineering specifications, established industry standards (ISO, ASTM), and historical performance data of predicate devices. The tests (static pull-out, torsional, fatigue, shear) are designed to assess physical properties and durability against these benchmarks.
8. The sample size for the training set
- Not Applicable. This document describes a device, not an AI or machine learning model that would require a training set.
9. How the ground truth for the training set was established
- Not Applicable. As per point 8, there is no training set for an AI/ML model described in this document.
{0}------------------------------------------------
p 1/3
K100858
DEC 1 0 2010
Summary of Safety and Effectiveness
Date: October 12, 2010
Manufacturer: Lima-Lto S.p.A. Via Nazionale, 52 33038 - Villanova di San Daniele Udine - Italy
U.S. Contact Person: Cheryl Hastings Principal Consultant Phone: 574-527-4220
| Product | Product Code | Regulation and Classification Name |
|---|---|---|
| SMR Shoulder System | KWT | Shoulder joint metal/polymer non-constrainedcemented prosthesis per 21 CFR 888.3650 |
| HSD | Shoulder joint humeral (hemi-shoulder) metallicuncemented prosthesis per 21 CFR 888.3690 |
Device Description:
The SMR Shoulder System consists of a humeral stem, a humeral body, an adaptor taper, a humeral head and a glenoid component. Components are offered for cemented use for hemi or total shoulder joint arthroplasty, in primary trauma surgery.
Humeral stems are designed for cemented use and are characterized by an outline with a triple conicity to provide optimal fixation and fit independent of canal morphology. The stems are made from Ti6Al4V (ISO 5832-3, ASTM F1472). The surface in contact with cement is polished to reduce shear stresses on the cement mantle and avoid abrasion resulting from micromovements at the stem-cement interface. The distal part of the stems is characterized by a cylindrical cross-section while the proximal part is fluted to prevent rotation of the component relative to bone. The stems are provided with a male Morse taper in the proximal part for coupling with the humeral body.
Humeral bodies are made from Ti6Al4V and are coupled with the humeral stem via a female Morse-taper connection; a locking screw is provided to aid in initial mating of the stem / body assembly. Cylindrical marks are designed at the base of the Morse-taper to provide correct alignment of the eccentricity of the humeral head during surgery. A male Morsetaper connection is designed for the coupling between the humeral body and the humeral head by means of specific adaptor tapers: an angle of 45° between the axis of this Morsetaper and the axis of the stem gives the correct varus-valgus alignment to the joint. A system of holes on the external surface allows anatomical attachment of tuberosities using sutures or wires in trauma applications.
Traditional 510(k) - SMR Shoulder System October 12, 2010
Company Confidential Page 2 - Section 5
{1}------------------------------------------------
p. 2/3
K100858
Adaptor tapers (neutral and eccentrical with different heights), are made from Ti6Al4V. They allow coupling between the humeral body and the humeral head. These devices are designed to adjust the centre of rotation of the joint and provide the required offset to the humeral head to achieve the correct tensioning to the soft tissues, optimizing joint stability.
The humeral heads are made from CoCrMo (ISO 5832-12, ASTM F1537). They are intended to articulate with the glenoid bone in hemi-arthroplasty or with the glenoid component in total shoulder arthroplasty. The surface is polished to aid in the reduction of wear.
The glenoid components are manufactured from Ultra-High Molecular Weight Polyethylene (UHMWPE ISO 5834-2, ASTM F648). The articulating surface has a radius of curvature greater than the corresponding humeral head, which allows translation in the superior/inferior and anterior/posterior directions. The back surface of the component is spherical in geometry and has a single central peg which is inserted in the hole drilled in the glenoid cavity during surgery. The peg surface has three grooves to provide enhanced cement fixation.
Intended Use / Indications:
The SMR Shoulder system is intended for partial or total primary shoulder joint replacement. The components are intended for use in cemented applications.
Total or hemi-shoulder replacement is indicated for patients suffering from disability due to:
- . Non-inflammatory degenerative joint disease including osteoarthritis and avascular necrosis:
- Inflammatory degenerative joint disease such as rheumatoid arthritis;
- Treatment of acute fractures of the humeral head that cannot be treated with other � fracture fixation methods
Predicate Devices:
- Promos modular shoulder system (PlusOrthopedics, K063578) .
- . Anatomical shoulder system (Centerpulse, K030259)
- Modular shoulder system (Acumed, K992525) ●
Comparable Features to Predicate Device(s):
The SMR Shoulder System is similar to the predicate devices in terms of intended use, indications, design and materials. The SMR Shoulder System and the predicates are all intended for partial or total primary shoulder joint replacement. The SMR humeral stems are intended for cemented use only while some of the predicate humeral stems are intended for cemented or cementless use. The SMR and predicate glenoids are all intended for cemented
Traditional 510(k) - SMR Shoulder System October 12, 2010
Company Confidential Page 3 - Section 5
{2}------------------------------------------------
use only. The indications of all of the systems are similar with the exception that the SMR Shoulder System is not indicated for revision procedures.
Like the Promos Modular and Acumed Modular Shoulder Systems, the SMR Shoulder System provides modular humeral stem and humeral body components. The Anatomical Shoulder System is designed as a one-piece humeral stem / body component. The SMR Shoulder System and the Promos Modular Shoulder System include modular adaptor tapers for adjustment of the humeral head. The Anatomical Shoulder System and the Acumed Modular Shoulder System do not include adaptor tapers.
The components of the SMR Shoulder System are manufactured from the same or similar materials as the predicate devices. The SMR, Promos and Acumed humeral stems are manufactured from Ti6A14V while the Anatomical cemented humeral stems are manufactured from cast CoCrMo.
Non-Clinical Testing: The SMR Shoulder System has undergone static pull-out and/or torsional testing of all modular connections; fatigue testing to demonstrate both the strength of the humeral stem and the post-fatigue strength of the modular connections; static shear and torsional testing of the glenoid component; and fatigue testing of the glenoid component. All mechanical testing was done on worst case components or constructs. Where possible, standard test methods were used to allow comparison to testing of the predicate devices. The testing results demonstrated the device's ability to perform under expected clinical conditions.
Clinical Testing: Clinical testing was not necessary to demonstrate substantial equivalence of the SMR Shoulder System to the predicate device(s).
Traditional 510(k) - SMR Shoulder System October 12, 2010
Company Confidential Page 4 - Section 5
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the Department of Health & Human Services USA. The logo features a stylized eagle with three swooping lines representing its wings. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" is arranged in a circular fashion around the eagle.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Lima-Lto- S.P.A % Hastings Regulatory Consulting, LLC. Ms. Cheryl Hastings Principal Consultant P.O. Box 696 Winona Lake, Indiana 46590
DEC 1 0 2010
Re: K100858
Trade/Device Name: SMR Shoulder System Regulation Number: 21 CFR 888.3650 Regulation Name: 21 CF R 80015860
Regulation Name: Shoulder joint metal/polymer non constrained cemented prothesis Regulatory Class: Class II Product Code: KWT, HSD Dated: November 9, 2010 Received: November 12, 2010
Dear Ms. Hastings:
We have reviewed your Section 510(k) premarket notification of intent to market the indication We have reviewed your Section 510(x) premaint is substantially equivalent (for the indications
referenced above and have determined the device is subsess marketed in interst for use stated in the enclosure) the arrested predicate devices marketed in interstate for use stated in the encrosuly manoned people to the Medical Device Amendments, or to commerce prior in May 28, 1970, the chance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA).
and Cosmetic Act (Act) that do not require approval contraks approval provision and Cosment Act (Act) that to hot require approval controls provisions of the Act. The Act. The You may, therefore, market the device, belyev to the governments for among mishranding and general controls provisions of the frectined of the adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it
and he If your device is classified (Sec aros). Existing major regulations affecting your device can be
may be subject to additional controls. may be subject to additional controls: Existing Mayor of 800 to 898. In addition, FDA may
found in the Code of Federal Regulations, Title 21, Forts 800 to 898. Register found in the Code of Federal Regarations, Four device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean Please be advised that FDA s issualite of a substitution with other requirements of the Act
that FDA has made a determination that your device and agencies. You must that FDA has made a delemination inal your are rederal agencies. You must or any Federal statutes and regulations and limited to: registration and listing (21
comply with all the Act's requirements, including, but not reporting of medical comply with all the Act Stequirements, and more reporting (teporting of medical
CFR Part 807); labeling (21 CFR Part 800); medical device requirements as CFR Part 807); labeling (21 CFR Part 807); incultur are repeative requirements as set
device-related adverse events) (21 CFR 803); good manufacturing practice requirements as
{4}------------------------------------------------
Page 2 – Ms. Cheryl Hastings
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic forth in the quality systems (QS) regulation (21-542 of the Act); 21 CFR 1000-1050.
product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please
ly of the specific advice of any and and and CDDLUGDRUOfficae(vem 115800 h go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/Offices/ucm115809.html for and go to http://www.fda.gov/AboulrDA/Cchicr.sOffices Office of Compliance. Also, please
the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. (21CFP P the Center for Devices and Radionogical realing by reference to premarket notification" (2) (CFR Patt note the regulation entitled, "Misoranding of reference to premaint of the MDR regulation (21 CFR Part 803), please go to
CFR Part 803), please go to http://www.ida.gov/wiodiatics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the You may obtain other general information on your responsibility of the manager Division of Sman Manazones - 100 or at its Internet address (800) 038-2041 01 (301) 770-1100 et an an an an an an an an an an an an an an an an an an an an an an an an an an an an an an an an an an an an an an an an an an an an an an
Sincerely yours,
Sincerely yours,
Mark N. Melkerson Director Division of Surgical, Orthopedic, and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
510(k) Number (if known): Unknown
Device Name: SMR Shoulder System
DEC 1 0 2010
Indications for Use:
SMR Shoulder System Indications for Use
The SMR Shoulder system is intended for partial or total primary shoulder joint replacement. The components are intended for use in cemented applications.
Total or hemi-shoulder replacement is indicated for patients suffering from disability due to:
- Non-inflammatory degenerative joint disease including osteoarthritis and avascular . necrosis;
- . Inflammatory degenerative joint disease such as rheumatoid arthritis;
- . Treatment of acute fractures of the humeral head that cannot be treated with other fracture fixation methods
Prescription Use X (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
for Mr. Melkerson
(Division Sign-Off) (Division of Surgical, Orthopedic, and Restorative Devices
Page 1 of 1
K100858 510(k) Number
Traditional 510(k) - SMR Shoulder System March 15, 2010
Company Confidential Section 4 - Page 15 of 279
N/A