K Number
K130642

Validate with FDA (Live)

Date Cleared
2013-06-12

(93 days)

Product Code
Regulation Number
888.3660
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The SMR Anatomic Shoulder System is indicated for partial or total, primary or revision shoulder joint replacement in patients suffering from disability due to:

  • Non-inflammatory degenerative joint disease including osteoarthritis and avascular necrosis: .
  • Inflammatory degenerative joint disease such as rheumatoid arthritis: .
  • Treatment of acure fractures of the humeral head that cannot be treated with other fracture fixation . methods:
  • Revision of a failed primary implant; .
  • . " Cuff tear arthropathy (CTA Heads only).

The Large Resection Stems (not available in the US) are indicated for oncology applications.

The SMR Reverse Shoulder System is indicated for primary, fracture or revision total shoulder replacement in a grossly rotator cuff deficient joint with severe arthropathy (disabled shoulder). The patient's joint must be anatomically and structurally suited to receive the selected implants and a functional deltoid muscle is necessary to use the device.

The Modular SMR Shoulder System allows the assembly of components in various humeral and glenoid constructs. The constructs are intended for cemented use as specified in the following table.

In the Anatomic shoulder the humeral consists of the humeral stem. the humeral body, the adaptor taper and the humeral head. In the Reverse shoulder the humeral consists of the humeral stem, the reverse humeral body and the reverse liner. On the humeral side the fixation of the humeral stem determines if the construct is cemented or uncemented.

The Anatomic glenoid construct consists of an all polyethylene glenoid or a metal back assembled with a liner while the Reverse glenoid consists of the metal back, the connector and the glenosphere. On the glenoid side, the fixation of the all polyethylene glenoid or the metal-back-determines if the construct is cemented or uncemented.

Device Description

The SMR 3-Pegs Glenoids are made from standard UHMWPE ). They are intended for cemented fixation only.

The SMR 3-Pegs Glenoids are available in two sizes (Small and Standard). They are characterized by an articulating surface with a radius of curvature greater than the corresponding humeral head. This mismatch allows for translation of the head in the superior/inferior and anterior/posterior directions. The back surface of the component is spherical in geometry and has three pegs for fixation in the glenoid.

The SMR 3-Pegs Glenoids are designed to articulate with the Limacorporate SMR humeral heads indicated for use in total shoulder replacement.

AI/ML Overview

This document describes the premarket notification (510(k)) for the Limacorporate SMR 3-Pegs Glenoids, a component for shoulder joint replacement. The submission aims to demonstrate substantial equivalence to previously cleared predicate devices.

1. Table of Acceptance Criteria and Reported Device Performance

The device is a medical implant, and its performance is evaluated based on mechanical and material properties, rather than diagnostic accuracy as would be common for AI devices. Therefore, the "acceptance criteria" here refer to meeting established mechanical testing standards and material specifications, and "reported device performance" refers to the results of those tests confirming compliance.

Acceptance Criteria (Mechanical/Material)DescriptionReported Device Performance
ASTM F1829 ComplianceStandard Test Method for Static Evaluation of Glenoid Locking Mechanism in Shear. This standard assesses the static mechanical shear strength of the glenoid component and its ability to resist translation relative to the bone or fixation.Mechanical testing performed. Results demonstrated the device's ability to perform under expected clinical conditions. (Exact numerical results not provided in summary)
ASTM F2028 ComplianceStandard Test Methods for Dynamic Evaluation of Glenoid Loosening or Disassociation. This standard assesses the dynamic mechanical fatigue strength of the glenoid component and its resistance to loosening or disassociation under physiological loading.Mechanical testing performed. Results demonstrated the device's ability to perform under expected clinical conditions. (Exact numerical results not provided in summary)
Range of Motion SimulationEnsure the device design does not overly limit the range of motion of the shoulder joint when implanted.Range of Motion simulation performed. Results ensured that the device design does not overly limit range of motion.
Material SpecificationsMade from standard Ultra-High Molecular Weight Polyethylene (UHMWPE) or UHMWPE X-Lima.The SMR 3-Pegs Glenoids are made from UHMWPE, or UHMWPE X-Lima, the same material as other cleared SMR System cemented glenoids.
SterilizationSterilized by ethylene oxide (EtO) with a Sterility Assurance Level (SAL) of 10-6 and a shelf life of 5 years.Sterilization method (EtO), SAL, and shelf life are identical to cleared cemented glenoids of the SMR Systems.

2. Sample Sizes Used for the Test Set and Data Provenance

This is a medical device approval based on non-clinical testing (mechanical and materials), not an AI/software device. Therefore, the concepts of "test set sample size" and "data provenance" (country of origin, retrospective/prospective) are not directly applicable in the same way they would be for an AI study.

  • Test Set Sample Size: "All mechanical testing was done on worst case components or constructs." The specific number of physical units tested is not provided in this summary.
  • Data Provenance: Not applicable as it's mechanical testing of physical components, not patient data.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

Not applicable. Ground truth for mechanical testing is established by conforming to predefined engineering standards (ASTM standards) and manufacturing specifications, not by expert consensus on clinical cases.

4. Adjudication Method for the Test Set

Not applicable, as this refers to expert review of clinical data, which was not performed.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

Not applicable. This is not an AI device or an imaging device requiring human reader interpretation. No MRMC study was performed.

6. Standalone (Algorithm Only) Performance Study

Not applicable. This is a physical medical device. There is no algorithm, and therefore no standalone algorithm performance to evaluate.

7. Type of Ground Truth Used

The "ground truth" for this medical device submission is established through:

  • Engineering Standards: Compliance with established ASTM (American Society for Testing and Materials) standards (F1829 and F2028) for mechanical properties relevant to glenoid components.
  • Material Specifications: Adherence to specifications for UHMWPE and UHMWPE X-Lima.
  • Design Specifications: Ensuring the dimensions and design features (e.g., radius of curvature, peg design) match those of legally marketed predicate devices and achieve their intended purpose (e.g., range of motion).

8. Sample Size for the Training Set

Not applicable. There is no "training set" in the context of an AI/machine learning model for this physical medical device approval. Performance is demonstrated through direct mechanical testing and comparison to predicate device characteristics.

9. How the Ground Truth for the Training Set Was Established

Not applicable, as there is no training set.

{0}------------------------------------------------

K130642 (1/4)

. JUN 1 2 2013

Summary of Safety and Effectiveness

Date: February 28, 2013

Manufacturer: Limacorporate S.p.A. Via Nazionale, 52 33038 - Villanova di San Daniele Udine - Italy

U.S. Contact Person: Stefano Adami Regulatory Affairs Manager Phone: +390432945511

ProductProduct CodeRegulation and Classification Name
SMR 3-Pegs GlenoidsKWSShoulder joint metal/polymer semi-constrainedcemented prosthesis per 21 CFR 888.3660

Description:

The SMR 3-Pegs Glenoids are made from standard UHMWPE ). They are intended for cemented fixation only.

The SMR 3-Pegs Glenoids are available in two sizes (Small and Standard). They are characterized by an articulating surface with a radius of curvature greater than the corresponding humeral head. This mismatch allows for translation of the head in the superior/inferior and anterior/posterior directions. The back surface of the component is spherical in geometry and has three pegs for fixation in the glenoid.

The SMR 3-Pegs Glenoids are designed to articulate with the Limacorporate SMR humeral heads indicated for use in total shoulder replacement.

Intended Use:

The SMR Anatomic Shoulder System is indicated for partial or total, primary or revision shoulder joint replacement in patients suffering from disability due to:

  • Non-inflammatory degenerative joint disease including osteoarthritis and avascular . necrosis:
  • Inflammatory degenerative joint disease such as rheumatoid arthritis; .
  • Treatment of acute fractures of the humeral head that cannot be treated with other . fracture fixation methods:
  • Revision of a failed primary implant; .
  • Cuff tear arthropathy (CTA Heads only).

The Large Resection Stems (not available in the US) are indicated for oncology applications.

The SMR Reverse Shoulder System is indicated for primary, fracture or revision total shoulder replacement in a grossly rotator cuff deficient joint with severe arthropathy

Traditional 510(k) - SMR 3-Pegs Glenoids February 28, 2013

Section 5 - Page 1 of 4

{1}------------------------------------------------

(disabled shoulder). The patient's joint must be anatomically and structurally suited to receive the selected implants and a functional deltoid muscle is necessary to use the device.

The Modular SMR Shoulder System allows the assembly of components in various humeral and glenoid constructs. The constructs are intended for cemented and uncemented use as specified in the following table.

In the Anatomic shoulder the humeral construct consists of the humeral stem, the humeral body, the adaptor taper and the humeral head. In the Reverse shoulder the humeral construct consists of the humeral stem, the reverse humeral body and the reverse liner. On the humeral side the fixation of the humeral stem determines if the construct is cemented or uncemented.

The Anatomic glenoid construct consists of an all polyethylene glenoid or a metal back assembled with a liner while the Reverse glenoid construct consists of the metal back, the connector and the glenosphere. On the glenoid side, the fixation of the all polyethylene glenoid or the metal back determines if the construct is cemented or uncemented.

SystemComponentsMaterialUseAvailablein US
AnatomicReverseCemNotCem
SMR Stems (Cemented, Cemented Revision)Ti6Al4VX
SMR Stems (Cementless Finned, CementlessRevision)Ti6Al4VX
SMR Large Resection stemsTi6Al4VX
SMR Modular AugmentsTi6Al4VX
SMR Humeral Bodies (Trauma, Finned)Ti6Al4VXX
SMR Reverse Humeral BodyTi6Al4VXX
SMR Reverse HA Coated Humeral BodyTi6Al4V+HAX
SMR Humeral Heads (Standard, CTA)CoCrMoXX
SMR Adaptor Tapers (Neutral, Eccentric)Ti6Al4VXX
SMR CTA Head Adaptor for Reverse Humeral BodyTi6Al4VXX
SMR GlenospheresCoCrMoX
Ti6Al4VX
UHMWPE X-Lima+Ti6Al4VX
Ti6Al4V
SMR ConnectorsTi6Al4VX
Reverse LinersUHMWPEXX
UHMWPE X-LimaXX
CoCrMoXX
AluminaXX
SMR Cemented GlenoidsUHMWPEX
SMR 3 Pegs Cemented GlenoidsUHMWPE X-LimaUHMWPEX
SMR Metal Back GlenoidsTi6Al4V+PoroTiX*X*
Ti6Al4V+PoroTi+HAX
SMR Metal Back LinerUHMWPEUHMWPE X-LimaX*X*
•*SMR Bone screwsTi6Al4VX
SMR Glenoid PlatesTiX

Traditional 510(k) - SMR 3-Pegs Glenoids February 28, 2013

Section 5 - Page 2 of 4

{2}------------------------------------------------

*NOTE :

  • . In the US, the SMR Metal Glenoid/Liner construct, used as part of the SMR Anatomic Shoulder Replacement, is intended for use with bone cement and should be used without bone screws.
  • . Outside the US, the SMR Metal Backed Glenoid/Liner construct, used as part of the SMR Anatomic Shoulder Replacement, is intended for uncemented use with the addition of screws for fixation.
  • The SMR Metal Backed Glenoid/Connector/Glenosphere construct, used as part of the SMR . Reverse Shoulder replacement, is intended for uncemented use with the addition of screws for fixation.

Predicate Devices:

  • Limacorporate SMR Cemented Glenoids included in the following cleared 510(k): .
    • SMR Cemented Shoulder System (K100858); ।
    • . SMR Uncemented Shoulder System (K101263);
    • SMR CTA Humeral Heads (K110847); -
    • SMR Revision Stems (K111212). -
  • . Global Shoulder Crosslink Glenoid (DePuy, K052472);
  • Glenoid Component for the Foundation Total Shoulder System (Encore DJO, . K960906).

Comparable Features to Predicate Device(s):

The intended use and indications for the Limacorporate SMR 3-Pegs Glenoids are similar to those of the referenced predicate devices and identical to those of the SMR cemented glenoids (K100858, K101263, K110847 - "rotator cuff tear arthropathy" and K I 1212 - "revision of a failed primary component"). As for all the referenced predicate devices where a glenoid component is used, the SMR 3-Pegs glenoids are used in total shoulder replacements and are intended for use with bone cement only.

The surface of the SMR 3-Pegs Glenoids articulating with the humeral heads and the spherical backside surface are identical to those of the already cleared SMR glenoids. As for all the referenced predicate devices, the SMR 3-Pegs glenoids have a radius of curvature greater than the radius of the articulating humeral heads. The three pegs used for the fixation by means of bone cement are very similar to those of Foundation Total Shoulder glenoid (Encore - DJO) and to the peripheral pegs of the Global Shoulder Crosslink glenoid (DePuv). The presence of grooves on the fixation peg has the same functionality (enhance cement adhesion) for SMR 3-Pegs Glenoids as for the other cemented glenoids of Limacorporate and for the Foundation Total Shoulder glenoid (Encore - DJO).

The UHMWPE used for the SMR 3-Pegs Glenoids is the same material used for the SMR System cemented glenoids (Limacorporate). The two sizes of glenoids available for the SMR 3-Pegs Glenoids have the same identical dimensions of the sizes Small and Standard of the cleared SMR cemented glenoids.

Traditional 510(k) - SMR 3-Pegs Glenoids February 28, 2013

Section 5 - Page 3 of 4

{3}------------------------------------------------

All SMR 3-Pegs Glenoids and referenced Predicate Devices are sterile, single use devices. Sterilization of the SMR 3-Pegs Glenoids performed by means of ethylene oxide (EtO), the Sterility Assurance Level (SAL of 100), the shelf life of 5 years are identical to that of the cleared cemented glenoids of the SMR Systems.

Non-Clinical Testing:

The SMR 3-Pegs Glenoid has undergone mechanical tests according to ASTM F1829 (Standard Test Method for Static Evaluation of Glenoid Locking Mechanism in Shear) and ASTM F2028 (Standard Test Methods for Dynamic Evaluation of Glenoid Loosening or Disassociation). A Range of Motion simulation has been performed to ensure the device design does not overly limit range of motion. All mechanical testing was done on worst case components or constructs. Where possible, standard test methods were used to allow comparison to testing of the predicate devices. The testing results demonstrated the device's ability to perform under expected clinical conditions.

Clinical Testing:

Clinical testing was not necessary to demonstrate substantial equivalence of the SMR 3-Pegs Glenoids to the predicate device(s).

Traditional 510(k) - SMR 3-Pegs Glenoids February 28, 2013

Section 5 - Page 4 of 4

{4}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized symbol resembling a bird or abstract human figure, with three curved lines forming its body. The symbol is positioned to the right of a circular arrangement of text that reads "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA".

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

Letter Dated: June 12, 2013

Limacorporate S.p.A. % Hastings Regulatory, LLC Ms. Cheryl Hastings Principal Consultant P.O. Box 696 Winona Lake, Indiana 46590

Re: K130642

Trade/Device Name: SMR 3-Pegs Glenoids Regulation Number: 21 CFR 888.3660 Regulation Name: Shoulder joint metal/polymer semi-constrained cemented prosthesis Regulatory Class: Class II Product Code: KWS Dated: May 4, 2013 Received: May 6, 2013

Dear Ms. Hastings:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA mav publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set

{5}------------------------------------------------

Page 2 - Ms. Cheryl Hastings

forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Laurence D. Coyne -A

Mark N. Melkerson Director . Division of Orthopedic Devices Office_of Device_Evaluation_ Center for Devices and Radiological Health

Enclosure

{6}------------------------------------------------

510(k) Number (if known): K130642 Device Name: SMR 3-Pegs Glenoids Indications for Use:

SMR 3-Pegs Glenoids Indications for Use

The SMR Anatomic Shoulder System is indicated for partial or total, primary or revision shoulder joint replacement in patients suffering from disability due to:

  • Non-inflammatory degenerative joint disease including osteoarthritis and avascular necrosis: .
  • Inflammatory degenerative joint disease such as rheumatoid arthritis: .
  • Treatment of acure fractures of the humeral head that cannot be treated with other fracture fixation . methods:
  • Revision of a failed primary implant; .
  • . " Cuff tear arthropathy (CTA Heads only).

The Large Resection Stems (not available in the US) are indicated for oncology applications.

The SMR Reverse Shoulder System is indicated for primary, fracture or revision total shoulder replacement in a grossly rotator cuff deficient joint with severe arthropathy (disabled shoulder). The patient's joint must be anatomically and structurally suited to receive the selected implants and a functional deltoid muscle is necessary to use the device.

The Modular SMR Shoulder System allows the assembly of components in various humeral and glenoid constructs. The constructs are intended for cemented use as specified in the following table.

In the Anatomic shoulder the humeral consists of the humeral stem. the humeral body, the adaptor taper and the humeral head. In the Reverse shoulder the humeral consists of the humeral stem, the reverse humeral body and the reverse liner. On the humeral side the fixation of the humeral stem determines if the construct is cemented or uncemented.

The Anatomic glenoid construct consists of an all polyethylene glenoid or a metal back assembled with a liner while the Reverse glenoid consists of the metal back, the connector and the glenosphere. On the glenoid side, the fixation of the all polyethylene glenoid or the metal-back-determines if the construct is cemented or uncemented.

SystemComponentsMaterialUseAvailablein US
AnatomicReverseCemNotCem
SMR Stems (Cemented, Cemented Revision)Ti6Al4VX
SMR Stems (Cementless Finned, Cementless Revision)Ti6Al4VX
SMR Large Resection stemsTi6Al4VX
SMR Modular AugmentsTi6Al4VX
SMR Humeral Bodies (Trauma, Finned)Ti6Al4VXX
SMR Reverse Humeral BodyTi6Al4VXX
SMR Reverse HA Coated Humeral BodyTi6Al4V+HAXX
SMR Humeral Heads (Standard, CTA)CoCrMoXX
SMR Adaptor Tapers (Neutral, Eccentric)Ti6Al4VXX
SMR CTA Head Adaptor for Reverse Humeral BodyTi6Al4VXX

Page 1 of 2

Traditional 510(k) - SMR 3-Pegs Glenoids February 28, 2013

Section 4 - Page 15 of 436

{7}------------------------------------------------

AnatomicReverseComponentsMaterialCemNot CemAvailable in US
SMR GlenospheresCoCrMoX
SMR GlenospheresTi6Al4VX
SMR GlenospheresUHMWPE X-Lima+Ti6Al4VX
SMR ConnectorsTi6Al4VX
Reverse LinersUHMWPEXX
Reverse LinersUHMWPE X-LimaXX
Reverse LinersCoCrMoXX
Reverse LinersAluminaXX
SMR Cemented GlenoidsUHMWPEX
SMR 3 Pegs Cemented GlenoidsUHMWPE X-LimaX
SMR 3 Pegs Cemented GlenoidsUHMWPEX
SMR Metal Back GlenoidsTi6Al4V+PoroTiX*X*
SMR Metal Back GlenoidsTi6Al4V+PoroTi+HAX
SMR Metal Back LinerUHMWPEX*X*
SMR Metal Back LinerUHMWPE X-LimaX
•*SMR Bone screwsTi6Al4VX
SMR Glenoid PlatesTiX

Alumina (ISO 6474) - PoroTi Titanium Coating (ASTM F1580) - HA Hyroxyapatite Coating (ISO 13779)

*NOTE :

  • In the US, the SMR Metal Backed Glenoid/Liner construct, used as part of the SMR � Anatomic Shoulder Replacement, is intended for use with bone cement and should be used without bone screws.
  • . Outside the US, the SMR Metal Backed Glenoid/Liner construct, used as part of the SMR Anatomic Shoulder Replacement, is intended for uncemented use with the addition of screws for fixation.
  • The SMR Metal Backed Glenoid/Connector/Glenosphere construct, used as part of the SMR . Reverse Shoulder replacement, is intended for uncemented use with the addition of screws for fixation.

Prescription Use X (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Image /page/7/Picture/11 description: The image shows the text "Casey E. Hanley, Ph.D." above a horizontal line. Below the line, the text "Division of Orthopedic Devices" is displayed. There are some graphical elements above and below the text, but they are not clearly identifiable.

Page 2 of 2

Traditional 510(k) - SMR 3-Pegs Glenoids February 28, 2013

Section 4 - Page 15 of 436

§ 888.3660 Shoulder joint metal/polymer semi-constrained cemented prosthesis.

(a)
Identification. A shoulder joint metal/polymer semi-constrained cemented prosthesis is a device intended to be implanted to replace a shoulder joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. This generic type of device includes prostheses that have a humeral resurfacing component made of alloys, such as cobalt-chromium-molybdenum, and a glenoid resurfacing component made of ultra-high molecular weight polyethylene, and is limited to those prostheses intended for use with bone cement (§ 888.3027).(b)
Classification. Class II. The special controls for this device are:(1) FDA's:
(i) “Use of International Standard ISO 10993 ‘Biological Evaluation of Medical Devices—Part I: Evaluation and Testing,’ ”
(ii) “510(k) Sterility Review Guidance of 2/12/90 (K90-1),”
(iii) “Guidance Document for Testing Orthopedic Implants with Modified Metallic Surfaces Apposing Bone or Bone Cement,”
(iv) “Guidance Document for the Preparation of Premarket Notification (510(k)) Application for Orthopedic Devices,” and
(v) “Guidance Document for Testing Non-articulating, ‘Mechanically Locked’ Modular Implant Components,”
(2) International Organization for Standardization's (ISO):
(i) ISO 5832-3:1996 “Implants for Surgery—Metallic Materials—Part 3: Wrought Titanium 6-aluminum 4-vandium Alloy,”
(ii) ISO 5832-4:1996 “Implants for Surgery—Metallic Materials—Part 4: Cobalt-chromium-molybdenum casting alloy,”
(iii) ISO 5832-12:1996 “Implants for Surgery—Metallic Materials—Part 12: Wrought Cobalt-chromium-molybdenum alloy,”
(iv) ISO 5833:1992 “Implants for Surgery—Acrylic Resin Cements,”
(v) ISO 5834-2:1998 “Implants for Surgery—Ultra-high Molecular Weight Polyethylene—Part 2: Moulded Forms,”
(vi) ISO 6018:1987 “Orthopaedic Implants—General Requirements for Marking, Packaging, and Labeling,” and
(vii) ISO 9001:1994 “Quality Systems—Model for Quality Assurance in Design/Development, Production, Installation, and Servicing,” and
(3) American Society for Testing and Materials':
(i) F 75-92 “Specification for Cast Cobalt-28 Chromium-6 Molybdenum Alloy for Surgical Implant Material,”
(ii) F 648-98 “Specification for Ultra-High-Molecular-Weight Polyethylene Powder and Fabricated Form for Surgical Implants,”
(iii) F 799-96 “Specification for Cobalt-28 Chromium-6 Molybdenum Alloy Forgings for Surgical Implants,”
(iv) F 1044-95 “Test Method for Shear Testing of Porous Metal Coatings,”
(v) F 1108-97 “Specification for Titanium-6 Aluminum-4 Vanadium Alloy Castings for Surgical Implants,”
(vi) F 1147-95 “Test Method for Tension Testing of Porous Metal,”
(vii) F 1378-97 “Standard Specification for Shoulder Prosthesis,” and
(viii) F 1537-94 “Specification for Wrought Cobalt-28 Chromium-6 Molybdenum Alloy for Surgical Implants.”