Search Filters

Search Results

Found 4 results

510(k) Data Aggregation

    K Number
    K122012
    Device Name
    Z-MED Z-MED II
    Manufacturer
    Date Cleared
    2012-10-04

    (86 days)

    Product Code
    Regulation Number
    870.1255
    Reference & Predicate Devices
    Why did this record match?
    Reference Devices :

    K022722,K081680,K014124

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Balloon Aortic Valvuloplasty

    Device Description

    The NuMED Z-MED™ catheter is a coaxial catheter recommended for Balloon Aortic Valvuloplasty. The catheter's inner and outer shafts are constructed of polyamide tubing. The catheter features a molded proximal end bifurcate with two distinct luminal passages. The inflation lumen terminates into a distally mounted balloon made of polyamide. This balloon is of the non-compliant variety. Both the shaft size and the guidewire size vary according to balloon diameter. The distal lumen terminates at the tip of the catheter and will accept the passage of the appropriate guidewire. The lumen has radiopaque platinum marker band(s), centered or under the balloon shoulders, for placement using fluoroscopy. The catheter is white in color and the balloon material is clear. The catheter balloon diameter and name is stamped onto the Y sleeve and the balloon extension is labeled with the catheter specifications and lot number. The catheter is packaged in a polyethylene sheath and is double packed in two heat sealed Tyvek pouches. The NuMED Z-MED II™ catheter is a coaxial catheter recommended for Balloon Aortic Valvuloplasty. The catheter's inner and outer shafts are constructed of polyamide tubing. The catheter features a molded proximal end bifurcate with two distinct luminal passages. The inflation lumen terminates into a distally mounted balloon made of polyamide. This balloon is of the non-compliant variety. Both the shaft size and the guidewire size vary according to balloon diameter. The distal lumen terminates at the tip of the catheter and will accept the passage of the appropriate guidewire. The lumen has radiopaque platinum marker band(s), centered or under the balloon shoulders, for placement using fluoroscopy. The catheter is white in color and the balloon material is clear. The catheter balloon diameter and name is stamped onto the Y sleeve and the balloon extension is labeled with the catheter specifications and lot number. The catheter is packaged in a polyethylene sheath and is double packed in two heat sealed Tyvek pouches.

    AI/ML Overview

    Here's a summary of the acceptance criteria and the study that proves the device meets them, based on the provided text:

    Acceptance Criteria and Device Performance Study

    The K122012 submission for the Z-MED and Z-MED II Catheters primarily relies on in-vitro (bench) testing and a review of existing clinical data for safety and effectiveness. No specific algorithms or human-in-the-loop performance studies are described as this is a medical device (catheter), not an AI/software device.

    1. Table of Acceptance Criteria and Reported Device Performance

    The provided text details in-vitro test results for the Z-MED and Z-MED II catheters, as well as two predicate devices (NuCLEUS-X and NuCLEUS). The table below focuses on the Z-MED and Z-MED II results as the subject devices.

    Test PerformedAcceptance CriteriaZ-MED ResultsZ-MED II Results
    Visual InspectionThe catheters shall be free from contamination, discoloration, and any form of damage that could impact the proper functioning of the device.All catheters were visually inspected without any anomalies.All catheters were visually inspected without any anomalies.
    Balloon Preparation TestEach catheter shall be prepped per the procedure without functional difficulties or anomalies.All catheters tested were without functional difficulties or anomalies.All catheters tested were without functional difficulties or anomalies.
    Diameter and Profile TestThe balloon diameter at rated burst pressure shall be within +/- 10% of the labeled balloon diameter and the samples should fit through the selected introducer with no problems.All catheters met the acceptance criteria.All catheters met the acceptance criteria.
    Balloon DistensibilityThe results must demonstrate that the balloon diameter is within +/- 10% of the labeled diameter at the RBP and will not be significantly increased at increasingly higher pressures.All data obtained demonstrates that the balloon diameter is within +/- 10% of the labeled diameter at the RBP. All data obtained demonstrates that the diameter of the balloons will not be significantly increased at increasingly higher pressures.All data obtained demonstrates that the balloon diameter is within +/- 10% of the labeled diameter at the RBP. All data obtained demonstrates that the diameter of the balloons will not be significantly increased at increasingly higher pressures.
    Balloon Minimum Burst StrengthThe results must show statistically that with at least 95% confidence, 99.9% of the balloons will not burst at or below the maximum recommended rated burst pressure.(Specific pressures listed for various balloon sizes, all indicating performance at or above the acceptance criteria)(Specific pressures listed for various balloon sizes, all indicating performance at or above the acceptance criteria)
    Repeated Balloon Inflation (Balloon Fatigue) TestNo breaks allowed.No Breaks.No Breaks.
    Balloon Inflation/Deflation TestInflation achieved in less than 12 seconds and deflation achieved in less than 20 seconds.All catheters met the established acceptance criteria.All catheters met the established acceptance criteria.
    Balloon Deflatability TestThere should be no interference with balloon deflation.All catheters met the established acceptance criteria.All catheters met the established acceptance criteria.
    Tip Pull and Torque TestMust withstand at least 10 turns without breaking.No breaks.No breaks.
    Bond Strength TestAll bonds must withstand at least 3 lbs. of pull strength.All bonds met the established acceptance criteria.All bonds met the established acceptance criteria.
    Catheter Body Maximum Pressure TestAll samples must withstand 30 ATM (400psi).>30 ATM>30 ATM

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size for In-Vitro Testing: The document does not explicitly state the numerical sample size for each in-vitro test (e.g., how many catheters were visually inspected or burst tested). It refers to "All catheters" or "All samples" meeting criteria, indicating that a sufficient number were tested.
    • Data Provenance (Clinical):
      • Country of Origin: The clinical evaluation report was prepared to meet the requirements of the Medical Device Directive MOD 93/42/EEC (European Union standard). The clinical use of the Z-MED and Z-MED II catheters for BAV and BMV is mentioned in "those countries where they are used," and for BPV in "the United States."
      • Retrospective/Prospective: The clinical information gathered appears to be a retrospective review of existing clinical literature, device history, previous sales, and reported complaints. It references "Numerous clinical studies reported in literature" and "200 patients" without incidents from BAV prior to TAVI.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

    This information is not applicable and not provided in the context of this submission. The tests described are in-vitro bench tests for a physical medical device (catheter), which do not typically involve experts establishing ground truth in the way it's done for diagnostic AI/software. The ground truth for in-vitro tests is based on objective physical measurements and engineering specifications.

    For the clinical context, the "ground truth" refers to observations from actual patient use, as reported in clinical literature and through device monitoring. No specific number or qualification of experts reviewing patient outcomes is provided beyond the implicit understanding that medical professionals conducted the procedures and reported outcomes.

    4. Adjudication Method for the Test Set

    This is not applicable for the in-vitro bench testing described. Adjudication methods (like 2+1, 3+1) are typically used in clinical studies or for establishing ground truth in cases where there's variability in interpretation (e.g., image analysis by multiple radiologists). The in-vitro tests are objective physical measurements.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    An MRMC study is not applicable and not mentioned. This type of study is relevant for diagnostic devices, especially those involving human interpretation of data, often comparing human performance with and without AI assistance. The Z-MED and Z-MED II are interventional catheters, not diagnostic devices or AI software. There's no AI component mentioned to compare human readers with or without.

    6. Standalone (Algorithm Only) Performance Study

    This is not applicable and not mentioned. The devices are physical catheters, not algorithms or software. Therefore, there is no standalone (algorithm only) performance to evaluate.

    7. Type of Ground Truth Used

    • In-Vitro Testing: The ground truth for the in-vitro tests is established by the predefined engineering specifications and acceptance criteria outlined in the table (e.g., balloon diameter within +/- 10% of labeled, no breaks, inflation/deflation times). These are objective, measurable physical parameters.
    • Clinical Evaluation: The "ground truth" for clinical safety and performance is based on real-world patient outcomes and reported incidents/adverse events from existing clinical use and literature reviews.

    8. Sample Size for the Training Set

    This is not applicable. The Z-MED and Z-MED II are physical medical devices undergoing traditional bench testing and clinical review, not AI models that require training sets.

    9. How the Ground Truth for the Training Set Was Established

    This is not applicable, as there is no training set for a physical medical device.

    Ask a Question

    Ask a specific question about this device

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Multi-Track Angiographic Catheter (K952984, K003902) - Recommended for use in Multi-Track Anglography of cardiovascular vessels and/or chambers. It can be used for catheterization for anglogiap. In and pressure measurement in any chamber or vessel.

    Ghost II PTA (K003972, K011557) - This catheter is recommended for Percutaneous Gransbuminal Angioplasty (PTA) of the femoral, iliac, and renal arteries. These catheters are not designed to be used in the coronary arteries.

    Z-MED Catheter (K991977, K003114, K003643, K011557) - This catheter is recommended for Percutaneous Transluminal Valvuloplasty (PTV) of the pulmonary valve.

    • A patient with isolated pulmonary stenosis .
    • A patient with valvular pulmonary stenosis with other minor congenital heart disease that 사 does not require surgical intervention. And;

    (K931009, K011557) This catheter is recommended for Percutaneous Transluminal Angioplasty (ISSIN67) Internoral, ilias, and renal arteries. These catheters are not designed to be used in the coronary arteries.

    Tyshak Mini Pediatric PTV Catheter (K003276, K011557, K032591) - This catheter is I your recuration for Percutaneous Transluminal Valvuloplasty (PTV) of the pulmonary valve in Pediatric applications.

    • A patient with isolated pulmonary stenosis. u
    • A patient with valvular pulmonary stenosis with other minor congenital heart disease that l does not require surgical intervention.

    Tyshak Catheter (K991977, K003114, K011557) - This catheter is recommended for Percutaneous Transluminal Valvuloplasty (PTV) of the pulmonary valve.

    • A patient with isolated pulmonary stenosis u
    • A patient with valvular pulmonary stenosis with other minor congenital heart disease that ■ does not require surgical intervention. And;

    (K931009, K011557) This catheter is recommended for Percutaneous Transluminal Angioplasty (RSS1007, Norths of I'ms cancenal arteries. These catheters are not designed to be used in the coronary arteries.

    Tyshak II Catheter (K003052, K011557, K030589) - This catheter is recommended for I Yshak II Catherer in Catherer (PTV) of the pulmonary valve.

    • A patient with isolated pulmonary stenosis i
    • A patient with isolated pulmonary stenosis
      A patient with valvular pulmonary stenosis with other minor congenital heart disease that I does not require surgical intervention. And;

    Z-MED II Catheter (K003052, K011557, K030589) - This catheter is recommended for Percutaneous Transluminal Valvuloplasty (PTV) of the pulmonary valve.

    • A patient with isolated pulmonary stenosis .
    • A patient with isolated pulmonary stenosis with other minor congenital heart disease that l does not require surgical intervention. And;

    High Pressure PTA (Marauder) (K010880, K011557) - This catheter is recommended for right Fressure I 17x (maraduct) (PTA) of the femoral, iliac, and renal arteries. These catheters are not designed to be used in the coronary arteries.

    Z-5 Atrioseptostomy (K960070, K011557) - Used for the palliation of several congenital heart 2-5 All loseptosition of the great arteries, total anomalous pulmonary venous drainage without delects. transposition of the Ereal arcess, mitral stenosis, mitral atresia, and pulmonary atresia with intact ventricular septum.

    Z-5 Braided Atrioseptostomy (K001804, K011557) - Used for the palliation of several 2-3 Draided Attrioseptostom. (2001 of the great arteries, total anomalous pulmonary venous congential licar delected transportion, tricuspid atresia, mitral stenosis, mitral atressa, and pulmonary atresia with intact ventricular septum.

    Tyshak X Catheter (K022722) - This catheter is recommended for Percutaneous Transluminal Valvuloplasty (PTV) of the pulmonary valve.

    • A patient with isolated pulmonary stenosis .
    • A patient with valvular pulmonary stenosis with other minor congenital heart disease that I does not require surgical intervention.

    Tyshak II X Catheter (K022722) - This catheter is recommended for Percutaneous Transluminal Valvuloplasty (PTV) of the pulmonary valve.

    • A patient with isolated pulmonary stenosis .
    • A patient with valvular pulmonary stenosis with other minor congenital heart disease that 1 does not require surgical intervention.

    Z-MED X Catheter (K022722) - This catheter is recommended for Percutaneous Transluminal Valvuloplasty (PTV) of the pulmonary valve.

    • A patient with isolated pulmonary stenosis a
    • A patient with valvular pulmonary stenosis with other minor congenital heart disease that . does not require surgical intervention.

    Z ·MED II X Catheter (K022722) - This catheter is recommended for Percutaneous Transluminal Valvuloplasty (PTV) of the pulmonary valve.

    • A patient with isolated pulmonary stenosis .
    • A patient with valvular pulmonary stenosis with other minor congenital heart disease that . does not require surgical intervention.

    COEfficient Catheter (K014124) - This catheter is recommended for Percutaneous Transluminal Valvuloplasty (PTV) of the pulmonary valve.

    • A patient with isolated pulmonary stenosis ﺎ
    • A patient with isolated pulmonary stenosis with other minor congenital heart disease that I does not require surgical intervention.

    Ghost PTA (K931009, K011557) - This catheter is recommended for Percutaneous Gross ITTA (KSJ1007) RW115872 - The femoral, iliac, and renal arteries. These catheters are not designed to be used in the coronary arteries.

    Mullins PTA (K013601) - This catheter is recommended for Percutaneous Transluminal Millins ITA (INVL001) - This Satileter and renal arteries. These catheters are not designed to be used in the coronary arteries.

    Mullins X PTA (K022722)_- This catheter is recommended for Percutaneous Transluminal Mullins A F FA (XN22122) - Frid Sains and renal arteries. These catheters are not designed to be used in the coronary arteries.

    PTS (K003320, K011557) - For use in those patients with cardiovascular defects wherein accurate measurement of the defect is important to select the appropriately sized occluder device.

    Device Description

    Not Found

    AI/ML Overview

    The provided document is a 510(k) summary for NuMED Catheters, primarily focusing on extending the shelf life of their existing devices from 3 to 5 years, and outlining their various intended uses. It does not contain information about acceptance criteria or a study proving device performance in the context of diagnostic or therapeutic efficacy.

    The document details:

    • K040830 - All NuMED Catheters: A submission for various catheters (angiographic, PTA, PTV, atrioseptostomy, etc.).
    • Purpose of Submission: To extend the shelf life validation from 3 to 5 years, indicating a focus on stability rather than primary diagnostic or therapeutic performance.
    • Predicate Devices: Numerous existing NuMED catheters.
    • Regulatory Information: Class II devices, product codes, and relevant CFR sections.
    • Indications for Use: Detailed indications for each specific catheter type (e.g., Multi-Track Angiographic Catheter for angiography, Ghost II PTA for Percutaneous Transluminal Angioplasty of femoral, iliac, and renal arteries, Z-MED Catheter for Percutaneous Transluminal Valvuloplasty of the pulmonary valve).

    Therefore, I cannot provide the requested information regarding acceptance criteria and a study proving the device meets those criteria for diagnostic or therapeutic performance because this information is not present in the provided text. The document is a regulatory submission for shelf-life extension and outlines intended uses, not a clinical performance study.

    Ask a Question

    Ask a specific question about this device

    K Number
    K041306
    Manufacturer
    Date Cleared
    2004-06-10

    (24 days)

    Product Code
    Regulation Number
    870.4450
    Reference & Predicate Devices
    Why did this record match?
    Reference Devices :

    K931009, K991977, K003320, K022722

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    For use in those patients with cardiovascular defects wherein accurate measurement . I of the defect is important to select the appropriately sized occluder device.

    Device Description

    The NuMED, Inc. PTS X™ Sizing Balloon catheter is a coaxial catheter for use in those patients with cardiovascular defects wherein accurate measurement of the defect is important to select the appropriately sized occluder device. The outer body is made of polymeric tubing, and the inner tubing is comprised of a multi layer extrusion of polyamide (Vestamid PA12) that surrounds a braid of 304 LV Stainless Steel. The catheter features a proximal end bifurcate with two distinct luminal passages. The inflation lumen terminates into a distally mounted balloon. This balloon is of the non-compliant variety and will have a typical single wall thickness of 0.0004". This balloon is designed to insert through the smallest possible introduction sleeve. The through lumen terminates at the tip of the catheter and will accept the passage of the appropriate guidewire. This lumen has radiopaque platinum marker bands under the balloon shoulders for placement using fluoroscopy. Additionally, there are two radiopaque platinum marker bands spaced at 10mm (as measured from leading edge to leading edge). These bands are located at the balloon center and are used as a distance reference. The catheter is white in color and the balloon material is clear. The catheter balloon diameter is stamped onto the Y sleeve and the balloon extension is labeled with balloon diameter x balloon length x introducer size x shaft size x usable length x guidewire size and the catheter lot number. The catheter is packaged in a polyethylene loop and is double packed in two heat sealed Tyvek pouches.

    AI/ML Overview

    The provided text is a clearance letter from the FDA for the NuMED PTS X Sizing Balloon Catheter, not a study report. Therefore, it does not contain information about acceptance criteria, device performance, sample sizes, ground truth establishment, or multi-reader multi-case studies.

    The document primarily states that the device is substantially equivalent to a legally marketed predicate device (NuMED PTS Catheter) based on its intended use and material similarity. It outlines the regulatory classification and general requirements for the device.

    Here's a breakdown of why the requested information cannot be extracted from the provided text:

    • No clinical study or performance data: The letter focuses on the regulatory clearance process and substantial equivalence, not on a detailed study proving performance against specific acceptance criteria.
    • No mention of acceptance criteria: Specific metrics for device performance (e.g., accuracy of measurement, safety rates) that would constitute "acceptance criteria" are not listed.
    • No device performance metrics: The document describes the device and its intended use but does not provide quantitative or qualitative data on how well it performs its function.
    • No sample size for test set: No clinical or laboratory test results are detailed with sample sizes.
    • No data provenance: Information on the origin of data (if any), such as geographical location or retrospective/prospective nature, is absent.
    • No expert involvement: There is no mention of experts establishing ground truth or their qualifications.
    • No adjudication method: This would be relevant to evaluating expert consensus, which isn't discussed.
    • No MRMC study: The document does not describe any study involving human readers or AI assistance.
    • No standalone algorithm performance: This device is a physical catheter, not an algorithm, so standalone algorithm performance is not applicable.
    • No type of ground truth used: Since no performance study is described, there's no mention of ground truth.
    • No sample size for training set/ground truth for training set: These concepts are typically relevant to AI/machine learning models, which this device is not. The "training" for this device would be its design and manufacturing processes, not data training.

    In summary, the provided document is a regulatory clearance letter, not a study report. It does not contain the information required to populate the table or answer the specific questions about acceptance criteria, study details, and device performance.

    Ask a Question

    Ask a specific question about this device

    K Number
    K041093
    Manufacturer
    Date Cleared
    2004-05-06

    (9 days)

    Product Code
    Regulation Number
    870.1250
    Reference & Predicate Devices
    Why did this record match?
    Reference Devices :

    K931009, K991977, K022722

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    This catheter is recommended for Percutaneous Transluminal Angioplasty (PTA) of the femoral, iliac, and renal arteries. These catheters are not designed to be used in the coronary arteries.

    Device Description

    The Mullins X™ catheter is recommended for Percutaneous Transluminal Angioplasty (PTA) of the femoral, iliac and renal arteries. The catheter is a coaxial over the wire catheter with a balloon near the distal tip. One lumen permits guidewire insertion to facilitate advancement of the catheter, while the other lumen is for balloon inflation and deflation.

    The balloons of the MULLINS X™ PTA Catheter are made of a non-compliant polymeric material. The two laminate balloon system is designed to inflate to a specific diameter at a given pressure. The change in diameter is minimal over the range of inflation pressures. The balloons are heat bonded to the shaft.

    The outer body is made of polymeric tubing, and the inner tubing is comprised of a multi layer extrusion of polyamide (Vestamid PA12) that surrounds a braid of 304 LV Stainless Steel. The area under the balloon is enhanced with five radiopaque platinum image bands. Two that are 5mm on each side of the balloon center and two more under the balloon shoulders. An additional image band is imbedded into the tip of the catheter as an additional safety measure.

    The catheter is white in color and the balloon material is clear. The catheter balloon diameter and name is stamped onto the Y sleeve and the balloon extension is labeled with balloon diameter x balloon length x introducer size x shaft size x usable length x guidewire size and the catheter lot number. The catheter is packaged in a polyethylene sheath and is double packed in two heat sealed Tyvek pouches.

    AI/ML Overview

    The provided document is a 510(k) premarket notification for a medical device (NuMED Mullins X PTA Catheter), which primarily focuses on demonstrating substantial equivalence to a predicate device based on manufacturing specifications, materials, and intended use. It does not contain information about clinical studies, device performance in terms of efficacy or diagnostic accuracy, or specific acceptance criteria and their achievement through studies in the way a diagnostic AI device submission would.

    Therefore, I cannot extract the requested information regarding acceptance criteria, study details, sample sizes, expert ground truth adjudication, MRMC studies, standalone performance, or training set details. This document describes a medical device for a physical intervention (PTA catheter), not a diagnostic algorithm or AI system.

    I reiterate that the provided text does not contain any information about acceptance criteria or studies proving device performance related to diagnostic accuracy, AI, or human reader improvement.

    The document does mention:

    • Biocompatibility Testing: Materials used are the same as previously approved devices (K931009, K991977) which were tested for biocompatibility in compliance with the Tripartite Biocompatibility Guidance for Medical Devices. Test results are on file at NuMED, Inc.
    • Laboratory (Bench) Testing: All bench testing was performed in accordance with GMP's and results are kept on file at NuMED, Inc.

    However, these are general statements about compliance and are not the type of detailed performance studies asked for in the user's prompt (which implies a diagnostic or AI-driven device context).

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1