K Number
K022722

Validate with FDA (Live)

Device Name
PTA AND PTV
Manufacturer
Date Cleared
2002-09-13

(28 days)

Product Code
Regulation Number
870.1250
Age Range
All
Reference & Predicate Devices
N/A
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Z-MED-X Catheter - This catheter is recommended for Percutaneous Transluminal Valvuloplasty (PTV) of the pulmonary valve. - A patient with isolated pulmonary stenosis - A patient with valvular pulmonary stenosis with other minor congenital heart disease that does not require surgical intervention. And; This catheter is recommended for Percutaneous Transluminal Angioplasty (PTA) of the femoral, illiac, and renal arteries. These catheters are not designed to be used in the coronary arteries.

Tyshak-X Catheter - This catheter is recommended for Percutaneous Transluminal Valvuloplasty (PTV) of the pulmonary valve. - A patient with isolated pulmonary stenosis - A patient with valvular pulmonary stenosis with other minor congenital heart disease that does not require surgical intervention. And; This catheter is recommended for Percutaneous Transluminal Angioplasty (PTA) of the femoral, iliac, and renal arteries. These catheters are not designed to be used in the coronary arteries.

Tyshak II-X Catheter - This catheter is recommended for Percutaneous Transluminal Valvuloplasty (PTV) of the pulmonary valve. - A patient with isolated pulmonary stenosis - A patient with valvular pulmonary stenosis with other minor congenital heart disease that does not require surgical intervention. And;

Z-MED II-X Catheter - This catheter is recommended for Percutaneous Transluminal Valvuloplasty (PTV) of the pulmonary valve. - A patient with isolated pulmonary stenosis - A patient with valvular pulmonary stenosis with other minor congenital heart disease that does not require surgical intervention. And;

Mullins-X PTA - This catheter is recommended for Percutaneous Transluminal Angioplasty (PTA) of the femoral, iliac, and renal arteries. These catheters are not designed to be used in the coronary arteries.

Device Description

The only change from the current catheter configuration to the new X line is the inner tubing configuration. The current inner tubing on all catheter models is composed of Pebax polyamide material. The tubing is sized to accept a 0.035" guidewire and provide adequate guidewire movement and catheter trackability. This tubing is also fitted with radiopaque image bands of platinum/iridium. This tubing is common to all catheters in NuMED's product lines. The proposed tubing is comprised of a multi layer extrusion of polyamide (Vestamid PA12) that surrounds a braid of 304 LV Stainless Steel. This tubing is loaded with 20% BaSO4 and is Colored blue with Pantone 295C. All materials have been certified as biocompatible. This tubing is designed to enhance guidewire movement and also to avoid any stretching of the catheter body during a tougher than normal catheter removal.

AI/ML Overview

The provided document describes bench testing performed on various balloon catheters (Tyshak-X, Tyshak II-X, Z-MED-X, Z-MED II-X, Mullins-X) to demonstrate substantial equivalence to predicate devices, primarily due to a change in the internal tubing material.

Here's the information requested, based on the provided text:

1. Table of Acceptance Criteria and Reported Device Performance

The document provides the acceptance criteria for several bench tests. It doesn't explicitly state the "reported device performance" as direct numerical results for each test (e.g., "Deflation achieved in 15 seconds"). Instead, the context of the submission (510(k) for substantial equivalence) implies that the new devices met these criteria, as the FDA granted approval. The "Comparison Information" section states, "The only change from the current catheter configuration to the new X line is the inner tubing configuration," and that "All materials have been certified as biocompatible." This suggests the tests performed on the new material confirmed its suitability and performance aligned with the predicate devices.

Bench Testing PerformedAcceptance CriteriaReported Device Performance (Implied)
2. Balloon DeflatabilityDeflation achieved in less than 20 seconds.Met acceptance criteria.
3. Inflation/ Deflation Times TestInflation and deflation achieved in less than 20 seconds.Met acceptance criteria.
5. Maximum Luminal Injection Pressure TestMust withstand 400 psi.Met acceptance criteria.
6. Bond IntegrityMinimum of 2 lbs. at all test points.Met acceptance criteria.
7. Diameter/ Profile Test and Introducer TestMust fit through rated introducer size.Met acceptance criteria.
9. Guidewire CompatibilityMinimal resistance while pushing catheter over guidewire. Force may not be great enough to bend guidewire.Met acceptance criteria.

2. Sample Size Used for the Test Set and the Data Provenance

The document does not explicitly state the sample sizes for the bench tests. It only lists the types of tests performed. The data provenance is laboratory (bench) testing, likely conducted by NuMED, Inc. in Hopkinton, NY, USA. The testing is prospective in the sense that it was performed to evaluate specific changes to the device before market entry.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts

Not applicable. This is bench testing of a medical device, not a study involving human data interpretation where experts would establish ground truth for a test set. The "ground truth" for these tests is defined by the physical properties and performance characteristics measured against the pre-defined acceptance criteria.

4. Adjudication Method for the Test Set

Not applicable. This is bench testing, not a clinical trial or study requiring expert adjudication of results. The results of the bench tests are direct measurements against specified criteria.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done

No, an MRMC comparative effectiveness study was not done. This submission pertains to bench testing of a device's physical properties.

6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done

Not applicable. This document describes the physical characteristics and performance of a medical device (balloon catheter), not an algorithm or AI system.

7. The Type of Ground Truth Used

The "ground truth" for the bench tests is defined by the pre-established acceptance criteria for each test, which are based on engineering specifications and expected device performance. For example, "Deflation achieved in less than 20 seconds" is the ground truth for the Balloon Deflatability test.

8. The Sample Size for the Training Set

Not applicable. There is no training set mentioned, as this is not an AI or machine learning study.

9. How the Ground Truth for the Training Set was Established

Not applicable. There is no training set.

{0}------------------------------------------------

SUMMARY OF SAFETY AND EFFECTIVENESS DATA

SEP 1 3 2002

August 14, 2002

Submitted By: NuMED, Inc., 2880 Main St., Hopkinton, NY 12965 (Ph) 315-328-4491 (F) 315-328-4941

Contact Person: Nichelle LaFlesh

Device Name: Tyshak-X, Tyshak II-X, Z-MED-X, Z-MED II-X, Mullins-X

Predicate Devices: Tyshak, Tyshak II, Z-MED, Z-MED II, Mullins

The only material that has changed is the material in the inner Biocompatibility Testing: tubing. It has gone from being Pebax to Vestamid with stainless steel braiding. The testing that was done on this new material in enclosed in the biocompatibility section of this submission as well as the data for the other materials that have not changed.

Bench Testing PerformedAcceptance Criteria
2. Balloon DeflatabilityDeflation achieved in less than 20 seconds.
3. Inflation/ Deflation Times TestInflation and deflation achieved in less than 20seconds.
5. Maximum Luminal Injection Pressure TestMust withstand 400 psi.
6. Bond IntegrityMinimum of 2 lbs. at all test points.
7. Diameter/ Profile Test and Introducer TestMust fit through rated introducer size.
9. Guidewire CompatibilityMinimal resistance while pushing catheter overguidewire. Force may not be great enough tobend guidewire.

Laboratory (Bench) Testing:

Comparison Information:

Change Description:

The only change from the current catheter configuration to the new X line is the inner tubing configuration. The current inner tubing on all catheter models is composed of Pebax

{1}------------------------------------------------

polyamide material. The tubing is sized to accept a 0.035" guidewire and provide adequate guidewire movement and catheter trackability. This tubing is also fitted with radiopaque image bands of platinum/iridium. This tubing is common to all catheters in NuMED's product lines.

The proposed tubing is comprised of a multi layer extrusion of polyamide (Vestamid PA12) that surrounds a braid of 304 LV Stainless Steel. This tubing is loaded with 20% BaSO4 and is Colored blue with Pantone 295C. All materials have been certified as biocompatible. This tubing is designed to enhance guidewire movement and also to avoid any stretching of the catheter body during a tougher than normal catheter removal.

Important Considerations:

The proposed inner tubing is a nylon (polyamide) tubing that is very similar to the current material. There will be no significant material changes in the tubing or the image bands. The only change will be the construction of the inner tubing in changing from a polyamide tubing to a braided polyamide tubing.

{2}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/2/Picture/1 description: The image shows the seal for the Department of Health & Human Services - USA. The seal is circular and contains the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the seal is an abstract image of an eagle.

SEP 1 3 2002

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

Numed, Inc. c/o Ms. Nichelle R. LaFlesh Regulatory Affairs Manager P.O. Box 129 Nicholville, NY 12965

Re: K022722

Trade Name: PTA and PTV Regulation Number: 21 CFR 870.1250 Regulation Name: Percutaneous catheter Regulatory Class: Class II (two) Product Code: DQY Dated: August 15, 2002 Received: August 16, 2002

Dear Ms. LaFlesh:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations. Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set

{3}------------------------------------------------

Page 2 – Ms. Nichelle R. LaFlesh

forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 21 CFR Part 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4646. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html

Sincerely yours.

Delguth

Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

Indications For Use

510(k) Number (if known):

Device Names: Z-MED-X Catheter, Tyshak-X Catheter, Tyshak II-X Catheter, Z-MED II-X Catheter, and Mullins-X PTA Catheter

Indications For Use:

Z-MED-X Catheter - This catheter is recommended for Percutaneous Transluminal Valvuloplasty (PTV) of the pulmonary valve.

  • A patient with isolated pulmonary stenosis י
  • A patient with valvular pulmonary stenosis with other minor congenital heart disease that does not l require surgical intervention. And;

This catheter is recommended for Percutaneous Transluminal Angioplasty (PTA) of the femoral, illiac, and renal arteries. These catheters are not designed to be used in the coronary arteries.

Tyshak-X Catheter - This catheter is recommended for Percutaneous Transluminal Valvuloplasty (PTV) of the pulmonary valve.

  • A patient with isolated pulmonary stenosis ■
  • 트 A patient with valvular pulmonary stenosis with other minor congenital heart disease that does not require surgical intervention. And;

This catheter is recommended for Percutaneous Transluminal Angioplasty (PTA) of the femoral, iliac, and renal arteries. These catheters are not designed to be used in the coronary arteries.

Tyshak II-X Catheter - This catheter is recommended for Percutaneous Transluminal Valvuloplasty (PTV) of the pulmonary valve.

  • A patient with isolated pulmonary stenosis B
  • A patient with valvular pulmonary stenosis with other minor congenital heart disease that does not require surgical intervention. And;

Z-MED II-X Catheter - This catheter is recommended for Percutaneous Transluminal Valvuloplasty (PTV) of the pulmonary valve.

  • A patient with isolated pulmonary stenosis 트
  • I A patient with valvular pulmonary stenosis with other minor congenital heart disease that does not require surgical intervention. And;

Mullins-X PTA - This catheter is recommended for Percutaneous Transluminal Angioplasty (PTA) of the femoral, iliac, and renal arteries. These catheters are not designed to be used in the coronary arteries.

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Prescription Use(Per 21 CFR 801.109)OROver-The-Counter-Use(Optional Format 1-2-96)
------------------------------------------------------------------------------------------------

Division of Cardiovascular & Respiratory Devices

510(k) NumberK022722
------------------------

§ 870.1250 Percutaneous catheter.

(a)
Identification. A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.(b)
Classification. Class II (performance standards).