Search Filters

Search Results

Found 25 results

510(k) Data Aggregation

    K Number
    K250995
    Device Name
    Xpert Xpress CoV-2/Flu/RSV plus
    Manufacturer
    Cepheid®
    Date Cleared
    2025-05-01

    (30 days)

    Product Code
    QOF
    Regulation Number
    866.3981
    Why did this record match?
    Product Code :

    QOF

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP Authorized
    Intended Use
    The Xpert Xpress CoV-2/Flu/RSV plus test, performed on the GeneXpert Xpress System, is an automated multiplexed real-time reverse transcriptase polymerase chain reaction (RT-PCR) test intended for use in the simultaneous in vitro qualitative detection and differentiation of severe acute respiratory syndrome coronavirus (SARS-CoV-2), influenza A, influenza B, and/or respiratory syncytial virus (RSV) viral RNA in nasopharyngeal swab and anterior nasal swab specimens collected from individuals with signs and symptoms of respiratory tract infection. Clinical signs and symptoms of respiratory tract infection due to SARS-CoV-2, influenza A, influenza B, and RSV can be similar. The Xpert Xpress CoV-2/Flu/RSV plus test is intended for use in the differential detection of SARS-CoV-2, influenza A, influenza B and/or RSV RNA and aids in the diagnosis of COVID-19, influenza and/or RSV infections if used in conjunction with other clinical and epidemiological information, and laboratory findings. SARS-CoV-2, influenza A, influenza B, and RSV viral RNA are generally detectable in nasopharyngeal swab and anterior nasal swab specimens during the acute phase of infection. Positive results are indicative of the presence of the identified virus, but do not rule out bacterial infection or co-infection with other pathogens not detected by the test. The agent (s) detected by the Xpert Xpress CoV-2/Flu/RSV plus test may not be the definite cause of the disease. Negative results do not preclude SARS-CoV-2, influenza A, influenza B and/or RSV infection. The results of this test should not be used as the sole basis for diagnosis, treatment or other patient management decisions.
    Device Description
    The Xpert Xpress CoV-2/Flu/RSV plus test is an automated in vitro diagnostic test for the simultaneous qualitative detection and differentiation of SARS-CoV-2, Flu A, Flu B, and RSV viral RNA in nasopharyngeal swab (NPS) and anterior nasal swab (NS) specimens collected from individuals showing signs and symptoms of respiratory viral infection. The Xpert Xpress CoV-2/Flu/RSV plus test is performed on GeneXpert Xpress System, which consist of a GeneXpert IV instrument that executes sample preparation, nucleic acid amplification and real-time fluorescent signal detection for the tests, and a GeneXpert Hub with preloaded GeneXpert Xpress software for running the tests and viewing the test results. The GeneXpert Hub accessory integrates the computer, touchscreen monitor and barcode scanner. Each of the GeneXpert modules in the GeneXpert IV instrument can perform independent sample preparation and testing. The GeneXpert Xpress System requires the use of single-use disposable cartridges that hold the RT-PCR reagents and host sample purification, nucleic acid amplification, and detection of the target sequences. Because the cartridges are self-contained, cross-contamination between samples is minimized. The Xpert Xpress CoV-2/Flu/RSV plus test includes reagents for the detection of SARS-CoV-2, Flu A, Flu B and RSV viral RNA from NPS and NS specimens. The primers and probes in the Xpert Xpress CoV-2/Flu/RSV plus test are designed to amplify and detect unique sequences in the genes that encode the following proteins: SARS-CoV-2 nucleocapsid (N), SARS-CoV-2 envelope (E), SARS-CoV-2 RNA-dependent RNA polymerase (RdRP), influenza A matrix (M), influenza A basic polymerase (PB2), influenza A acidic protein (PA), influenza B matrix (M), influenza B non-structural protein (NS), and the RSV A and RSV B nucleocapsid. A Sample Processing Control (SPC) and a Probe Check Control (PCC) are also included in the cartridge utilized by the GeneXpert instrument. The SPC is present to control for adequate processing of the sample and to monitor for the presence of potential inhibitor(s) in the RT-PCR reaction. The SPC also ensures that the RT-PCR reaction conditions (temperature and time) are appropriate for the amplification reaction and that the RT-PCR reagents are functional. The PCC verifies reagent rehydration, PCR tube filling, and confirms that all reaction components are present in the cartridge including monitoring for probe integrity and dye stability. The Xpert Xpress CoV-2/Flu/RSV plus test is designed for use with NPS or NS specimens collected with nylon flocked swabs and placed into viral transport medium (VTM), Universal Transport Medium (UTM), or eNAT®. The ancillary specimen collection kits, swabs and transport media validated for use with the Xpert Xpress CoV-2/Flu/RSV plus test included: • **Nasopharyngeal Sample Collection Kit for Viruses** - Copan UTM® 3C057N (Flexible Minitip Flocked Swab with UTM® Medium without Beads) - Copan eNAT® Molecular Collection and Preservation Medium P/N 6U074S01 (Flexible Minitap Flocked Swab with eNAT® Medium) • **Nasal Sample Collection Kit for Viruses** - Copan UTM® 3C064N (Regular Flocked Swab with UTM® Medium without Beads) - Copan eNAT® Molecular Collection and Preservation Medium P/N 6U073S01 (Regular Flocked Swab with eNAT® Medium) • **Alternatively, swabs and transport media can be obtained separately:** - Nylon flocked swab (Copan P/N 502CS01, 503CS01) - Viral Transport Medium, 3 mL (Copan P/N 330C, 3C047N, BD Universal Transport Medium, Remel M4RT, or Remel M5) The ancillary reagents allow NPS and NS specimens from patients to be collected, preserved and transported to laboratory prior to analysis with the Xpert Xpress CoV 2/Flu/RSV plus test.
    Ask a Question
    K Number
    K250996
    Device Name
    Xpert Xpress CoV-2/Flu/RSV plus
    Manufacturer
    Cepheid®
    Date Cleared
    2025-05-01

    (30 days)

    Product Code
    QOF
    Regulation Number
    866.3981
    Why did this record match?
    Product Code :

    QOF

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP Authorized
    Intended Use
    The Xpert Xpress CoV-2/Flu/RSV plus test, performed on the GeneXpert Dx and GeneXpert Infinity Systems, is an automated multiplexed real-time reverse transcriptase polymerase chain reaction (RT-PCR) test intended for use in the simultaneous in vitro qualitative detection and differentiation of severe acute respiratory syndrome coronavirus (SARS-CoV-2), influenza A, influenza B, and/or respiratory syncytial virus (RSV) viral RNA in nasopharyngeal swab and anterior nasal swab specimens collected from individuals with signs and symptoms of respiratory tract infection. Clinical signs and symptoms of respiratory tract infection due to SARS-CoV-2, influenza A, influenza B, and RSV can be similar. The Xpert Xpress CoV-2/Flu/RSV plus test is intended for use in the differential detection of SARS-CoV-2, influenza A, influenza B and/or RSV RNA and aids in the diagnosis of COVID-19, influenza and/or RSV infections if used in conjunction with other clinical and epidemiological information, and laboratory findings. SARS-CoV-2, influenza A, influenza B, and RSV viral RNA are generally detectable in nasopharyngeal swab and anterior nasal swab specimens during the acute phase of infection. Positive results are indicative of the presence of the identified virus, but do not rule out bacterial infection or co-infection with other pathogens not detected by the test. The agent (s) detected by the Xpert Xpress CoV-2/Flu/RSV plus test may not be the definite cause of the disease. Negative results do not preclude SARS-CoV-2, influenza A, influenza B and/or RSV infection. The results of this test should not be used as the sole basis for diagnosis, treatment or other patient management decisions.
    Device Description
    The Xpert Xpress CoV-2/Flu/RSV plus test is an automated in vitro diagnostic test for the simultaneous qualitative detection and differentiation of SARS-CoV-2, Flu A, Flu B, and RSV viral RNA in nasopharyngeal swab (NPS) and anterior nasal swab (NS) specimens collected from individuals showing signs and symptoms of respiratory viral infection. The Xpert Xpress CoV-2/Flu/RSV plus test is performed on GeneXpert Instrument Systems (GeneXpert Dx, GeneXpert Infinity-48s and GeneXpert Infinity-80 systems, GeneXpert System with Touchscreen), which consist of an instrument, computer or touchscreen, and preloaded software for running tests and viewing the results. The GeneXpert Instrument Systems automate and integrate sample preparation, nucleic acid extraction and amplification, and detection of the target sequences in simple or complex samples using real-time reverse transcription (RT)-polymerase chain reaction (PCR) and PCR technology. Depending on the instrument, the GeneXpert Instrument Systems can have from 1 and up to 80 randomly accessible modules, each capable of performing separate sample preparation and real-time RT-PCR and PCR tests. Each module contains a syringe drive for dispensing fluids (i.e., the syringe drive activates the plunger that works in concert with the rotary valve in the cartridge to move fluids between chambers), an ultrasonic horn for lysing cells or spores, and a proprietary I-CORE® thermocycler for performing real-time RT-PCR and PCR as well as detection. The systems require the use of single-use disposable cartridges that hold the RT-PCR reagents and host sample purification, nucleic acid amplification, and detection of the target sequences. Because the cartridges are self-contained, cross-contamination between samples is minimized. The Xpert Xpress CoV-2/Flu/RSV plus test includes reagents for the detection of SARS-CoV-2, Flu A, Flu B and RSV viral RNA from NPS and NS specimens. The primers and probes in the Xpert Xpress CoV-2/Flu/RSV plus test are designed to amplify and detect unique sequences in the genes that encode the following proteins: SARS-CoV-2 nucleocapsid (N), SARS-CoV-2 envelope (E), SARS-CoV-2 RNA-dependent RNA polymerase (RdRP), influenza A matrix (M), influenza A basic polymerase (PB2), influenza A acidic protein (PA), influenza B matrix (M), influenza B non-structural protein (NS), and the RSV A and RSV B nucleocapsid. A Sample Processing Control (SPC) and a Probe Check Control (PCC) are also included in the cartridge utilized by the GeneXpert instrument. The SPC is present to control for adequate processing of the sample and to monitor for the presence of potential inhibitor(s) in the RT-PCR reaction. The SPC also ensures that the RT-PCR reaction conditions (temperature and time) are appropriate for the amplification reaction and that the RT-PCR reagents are functional. The PCC verifies reagent rehydration, PCR tube filling, and confirms that all reaction components are present in the cartridge including monitoring for probe integrity and dye stability. The Xpert Xpress CoV-2/Flu/RSV plus test is designed for use with NPS or NS specimens collected with nylon flocked swabs and placed into viral transport medium (VTM), Universal Transport Medium (UTM), or eNAT®.
    Ask a Question
    K Number
    K243400
    Device Name
    cobas liat SARS-CoV-2 & Influenza A/B v2 nucleic acid test
    Manufacturer
    Roche Molecular Systems, Inc.
    Date Cleared
    2025-04-25

    (176 days)

    Product Code
    QOF
    Regulation Number
    866.3981
    Why did this record match?
    Product Code :

    QOF

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP Authorized
    Intended Use
    The cobas liat SARS-CoV-2 & Influenza A/B v2 nucleic acid test is an automated rapid multiplex real-time reverse transcription polymerase chain reaction (RT-PCR) test intended for the simultaneous qualitative detection and differentiation of severe acute respiratory syndrome coronavirus 2 (SARS-CoV-2), influenza A virus and influenza B virus nucleic acids in anterior nasal (nasal) and nasopharyngeal swab specimens from individuals exhibiting signs and symptoms of respiratory tract infection. Clinical signs and symptoms of respiratory tract infection due to SARS-CoV-2 and influenza can be similar. This test is intended to aid in the differential diagnosis of SARS-CoV-2, influenza A and influenza B infections in humans and is not intended to detect influenza C virus infections. Nucleic acids from the viral organisms identified by this test are generally detectable in nasopharyngeal and nasal swab specimens during the acute phase of infection. The detection and identification of specific viral nucleic acids from individuals exhibiting signs and symptoms of respiratory tract infection are indicative of the presence of the identified virus, and aid in diagnosis if used in conjunction with other clinical and epidemiological information and laboratory findings. The results of this test should not be used as the sole basis for diagnosis, treatment, or other patient management decisions. Positive results do not rule out coinfection with other organisms. The organism(s) detected by the cobas liat SARS-CoV-2 & Influenza A/B v2 nucleic acid test may not be the definite cause of disease. Negative results do not preclude SARS-CoV-2, influenza A virus or influenza B virus infections.
    Device Description
    The cobas liat SARS-CoV-2 & Influenza A/B v2 nucleic acid test is performed on the cobas liat analyzer which automates and integrates sample purification, nucleic acid amplification, and detection of the target sequence in biological samples using real-time PCR assays. The assay targets both the ORF1 a/b non-structural region and membrane protein gene that are unique to SARS-CoV-2, a well-conserved region of the matrix gene of influenza A (Flu A target), and the nonstructural protein 1 (NS1) gene of influenza B (Flu B target). An Internal Control (IC) is included to control for adequate processing of the target virus through all steps of the assay process and to monitor the presence of inhibitors in the RT-PCR processes.
    Ask a Question
    K Number
    K243406
    Device Name
    cobas liat SARS-CoV-2, Influenza A/B & RSV nucleic acid test
    Manufacturer
    Roche Molecular Systems, Inc.
    Date Cleared
    2025-04-25

    (175 days)

    Product Code
    QOF
    Regulation Number
    866.3981
    Why did this record match?
    Product Code :

    QOF

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP Authorized
    Intended Use
    The cobas liat SARS-CoV-2, Influenza A/B & RSV nucleic acid test is an automated rapid multiplex real-time reverse transcription polymerase chain reaction (RT-PCR) test intended for the simultaneous qualitative detection and differentiation of severe acute respiratory syndrome coronavirus 2 (SARS-CoV-2), influenza A virus, influenza B virus and respiratory syncytial virus (RSV) nucleic acids in anterior nasal (nasal) and nasopharyngeal swab specimens from individuals exhibiting signs and symptoms of respiratory tract infection. Clinical signs and symptoms of respiratory viral infection due to SARS-CoV-2, influenza and RSV can be similar. This test is intended to aid in the differential diagnosis of SARS-CoV-2, influenza A, influenza B, and RSV infections in humans and is not intended to detect influenza C virus infections. Nucleic acids from the viral organisms identified by this test are generally detectable in nasopharyngeal and nasal swab specimens during the acute phase of infection. The detection and identification of specific viral nucleic acids from individuals exhibiting signs and symptoms of respiratory tract infection are indicative of the presence of the identified virus, and aid in diagnosis if used in conjunction with other clinical and epidemiological information, and laboratory findings. The results of this test should not be used as the sole basis for diagnosis, treatment, or other patient management decisions. Positive results do not rule out coinfection with other organisms. The organism(s) detected by the cobas liat SARS-CoV-2, Influenza A/B & RSV nucleic acid test may not be the definite cause of disease. Negative results do not preclude SARS-CoV-2, influenza A virus, influenza B virus, or RSV infections.
    Device Description
    The cobas liat SARS-CoV-2, Influenza A/B & RSV nucleic acid test is performed on the cobas liat analyzer which automates and integrates sample purification, nucleic acid amplification, and detection of the target sequence in biological samples using real-time PCR assays. The assay targets both the ORF1 a/b non-structural region and membrane protein gene that are unique to SARS-CoV-2, a well-conserved region of the matrix gene of influenza A (Flu A target), the nonstructural protein 1 (NS1) gene of influenza B (Flu B target) and the matrix gene of RSV (RSV target). An Internal Control (IC) is included to control for adequate processing of the target virus through all steps of the assay process and to monitor the presence of inhibitors in the RT-PCR processes.
    Ask a Question
    K Number
    K242526
    Device Name
    Visby Medical Respiratory Health Test
    Manufacturer
    Visby Medical, Inc.
    Date Cleared
    2025-02-19

    (177 days)

    Product Code
    QOF, OOF
    Regulation Number
    866.3981
    Why did this record match?
    Product Code :

    QOF

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP Authorized
    Intended Use
    The Visby Medical Respiratory Health Test is a single-use (disposable), fully integrated, automated Reverse Transcription Polymerase Chain Reaction (RT-PCR) in vitro diagnostic test intended for the simultaneous qualitative detection and differentiation of severe acute respiratory syndrome coronavirus 2 (SARS-CoV-2), influenza A, and influenza B RNA in nasopharyngeal swab and anterior nasal swab specimens from individuals with signs and symptoms of respiratory tract infection. Clinical signs and symptoms of respiratory tract infection due to SARS-CoV-2, influenza A, and influenza B can be similar. The Visby Medical Respiratory Health Test is intended for use as an aid in the differential diagnosis of SARS-CoV-2, influenza A, and influenza B infection if used in conjunction with other clinical and epidemiological information, and laboratory findings. SARS-CoV-2, influenza A, and influenza B viral RNA are generally detectable in nasopharyngeal swab and anterior nasal swab specimens during the acute phase of infection. This test is not intended to detect influenza C virus infections. Positive results are indicative of the identified virus, but do not rule out bacterial infection or co-infection with other organisms not detected by the test. The agent(s) detected by the Visby Medical Respiratory Health Test may not be the definitive cause of disease. Negative results do not preclude SARS-CoV-2, influenza A, or influenza B infection. The results of this test should not be used as the sole basis for diagnosis, treatment, or other patient management decisions.
    Device Description
    The Visby Medical Respiratory Health Test is a single-use (disposable), fully integrated, compact device containing a reverse transcription polymerase chain reaction (RT-PCR) based assay for qualitative detection of influenza B, and/or SARS-CoV-2 viral RNA in upper respiratory tract specimens. The device automatically performs all steps required to complete lysis, reverse transcription (RT), PCR amplification, and detection. Specimen collected using nasopharyngeal (NP) or anterior nasal (AN) swabs (without transport media) are placed in the Visby Medical Respiratory Health Buffer and then transferred into the sample port of the device using the provided fixed volume pipette. The sample enters a lysis module and rehydrates the RT enzyme and RT primers. The mixture then moves through a sample preparation module where viruses and human cells are simultaneously lysed, and RNA is reverse transcribed. The resulting fluid (containing cDNA) is then mixed with lyophilized PCR reagents containing the DNA polymerase enzyme and PCR primers. The PCR mixture (containing cDNA template and reagents) is then thermal cycled to amplify the targets, including human beta-2 microglobulin (B2M) RNA, which serves as a process control. After PCR, the biotinylated product is hybridized to covalently bound capture probes at specific locations along a flow channel. The flow channel is configured to facilitate an enzymatic reaction that uses streptavidin bound horseradish peroxidase (HRP) and a colorimetric substrate that forms a purple precipitate. The operator observes a color change at the specific locations indicating the presence of an amplified target. Test results can be expected in approximately 30 minutes: illumination of a "DONE" status light on the front of the device and a purple color in the "RESULTS VALID" spot, indicate a successful test. A purple spot adjacent to "Flu A", "Flu B", and/or "COVID-19" signifies the presence of, influenza A, influenza B, and/or SARS-CoV-2 viral RNA.
    Ask a Question
    K Number
    K241573
    Device Name
    Alinity m Resp-4-Plex
    Manufacturer
    Abbott Molecular Inc
    Date Cleared
    2025-02-14

    (256 days)

    Product Code
    QOF
    Regulation Number
    866.3981
    Why did this record match?
    Product Code :

    QOF

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP Authorized
    Intended Use
    Alinity m Resp-4-Plex is a multiplexed real-time in vitro reverse transcription polymerase chain reaction (RT-PCR) assay for use with the automated Alinity m System for the qualitative detection and differentiation of Severe Acute Respiratory Syndrome Coronavirus 2 (SARS-CoV-2), influenza A virus, influenza B virus and Respiratory Syncytial Virus (RSV) in nasopharyngeal swab specimens collected from patients with signs and symptoms of respiratory tract infection. Clinical signs and symptoms of respiratory tract infection due to SARS-CoV-2, influenza B, and RSV can be similar. The Alinity m Resp-4-Plex assay is intended for use in the differential detection of SARS-CoV-2, influenza B and/or RSV RNA and aids in the diagnosis of COVID-19, influenza and/or RSV infections if used in conjunction with other clinical and epidemiological information, and laboratory findings. SARS-CoV-2, influenza B and RSV viral RNA are generally detectable in nasopharyngeal swab specimens during the acute phase of infection. This test is not intended to detect influenza C virus infections. Positive results are indication of the identified virus, but do not rule out bacterial infection or co-infection with other pathogens not detected by the test. The agent(s) detected by the Alinity m Resp-4-Plex assay may not be the definite cause of disease. Negative results do not preclude SARS-CoV-2, influenza B and/or RSV infections and should not be used as the sole basis for diagnosis, treatment or other patient management decisions.
    Device Description
    The Alinity m Resp-4-Plex assay requires two separate assay-specific kits: Alinity m Resp-4-Plex AMP Kit and Alinity m Resp-4-Plex CTRL Kit. The assay utilizes real-time PCR to amplify and detect genomic RNA sequences of influenza A (flu A), influenza B (flu B), RSV, and/or SARS-CoV-2 from nasopharyngeal (NP) swab specimens. The assay targets 2 different genes within the SARS-CoV-2 genome. Fluorescently labeled probes allow for simultaneous detection and differentiation of amplified products of all 4 viruses and Internal Control (IC) in a single reaction vessel. All steps of the assay procedure are executed automatically by the Alinity m System, which is a continuous random-access analyzer. The system performs automated sample preparation using magnetic microparticle technology. The IC is introduced into each specimen at the beginning of sample preparation. Purified RNA is combined with activation and amplification/detection reagents and transferred to a reaction vessel for reverse transcription, PCR amplification, and real-time fluorescence detection. A positive and negative control are tested to ensure performance. Patient results are automatically reported. The assay also utilizes the Alinity m Resp-4-Plex Assay Application Specification File, Alinity m System and System Software, Alinity m Sample Prep Kit 2, Alinity m Tubes and Caps, and Alinity m System Solutions.
    Ask a Question
    K Number
    K242071
    Device Name
    Xpert Xpress CoV-2/Flu/RSV plus
    Manufacturer
    Cepheid
    Date Cleared
    2025-01-10

    (178 days)

    Product Code
    QOF
    Regulation Number
    866.3981
    Why did this record match?
    Product Code :

    QOF

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP Authorized
    Intended Use
    The Xpert Xpress CoV-2/Flu/RSV plus test, performed on the GeneXpert Xpress System, is an automated multiplexed real-time reverse transcriptase polymerase chain reaction (RT-PCR) test intended for use in the simultaneous in vitro qualitative detection and differentiation of severe acute respiratory syndrome coronavirus (SARS-CoV-2), influenza A, influenza B, and/or respiratory syncytial virus (RSV) viral RNA in nasopharyngeal swab anterior nasal swab specimens collected from individuals with signs and symptoms of respiratory tract infection. Clinical signs and symptoms of respiratory tract infection due to SARS-CoV-2, influenza B, and RSV can be similar. The Xpert Xpress CoV-2/Flu/RSV plus is intended for use in the differential detection of SARS-CoV-2, influera A, influenza B, and/or RSV RNA and aids in the diagnosis of COVID-19, influenza, and/or RSV infections if used in conjunction with other clinical and epidemiological information, and laboratory findings. SARS-CoV-2, influenza A. influenza B, and RSV viral RNA are generally detectable in nasopharyngeal swab and anterior nasal swab specimens during the acute phase of infection. Positive results are indicative of the identified virus, but do not rule out bacterial infection or co-infection with other pathogens not detected by the test. The agent (s) detected by the Xpert Xpress CoV-2/Flu/RSV plus test may not be the definite cause of the disease. Negative results do not preclude SARS-CoV-2, influenza A virus, and/or RSV infection. The results of this test should not be used as the sole basis for treatment or other patient management decisions.
    Device Description
    The Xpert Xpress CoV-2/Flu/RSV plus test is an automated in vitro diagnostic test for the simultaneous qualitative detection and differentiation of SARS-CoV-2. Flu A. Flu B. and RSV viral RNA in nasopharyngeal swab (NPS) and anterior nasal swab (NS) specimens collected from individuals showing signs and symptoms of respiratory viral infection. The Xpert Xpress CoV-2/Flu/RSV plus test is performed on GeneXpert Xpress System, which consist of a GeneXpert IV instrument that executes sample preparation, nucleic acid amplification and real-time fluorescent signal detection for the tests, and a GeneXpert Hub with preloaded GeneXpert Xpress software for running the tests and viewing the test results. The GeneXpert Hub accessory integrates the computer, touchscreen monitor and barcode scanner. Each of the GeneXpert modules in the GeneXpert IV instrument can perform independent sample preparation and testing. The GeneXpert Xpress System requires the use of single-use disposable cartridges that hold the RT-PCR reagents and host sample purification, nucleic acid amplification, and detection of the target sequences. Because the cartridges are self-contained, cross-contamination between samples is minimized. The Xpert Xpress CoV-2/Flu/RSV plus test cartridge includes reagents for the detection of SARS-CoV-2, Flu A, Flu B and RSV viral RNA from NPS and NS specimens. The primers and probes in the Xpert Xpress CoV-2/Flu/RSV plus test are designed to amplify and detect unique sequences in the genes that encode the following proteins: SARS-CoV-2 nucleocapsid (N), SARS-CoV-2 envelope (E), SARS-CoV-2 RNA-dependent RNA polymerase (RdRP), influenza A matrix (M), influenza A basic polymerase (PB2), influenza A acidic protein (PA), influenza B matrix (M), influenza B non-structural protein (NS), and the RSV A and RSV B nucleocapsid. A Sample Processing Control (SPC) and a Probe Check Control (PCC) are also included in the cartridge utilized by the GeneXpert Xpress System. The SPC is present to control for adequate processing of the sample and to monitor for the presence of potential inhibitor(s) in the RT-PCR reaction. The SPC also ensures that the RT-PCR reaction conditions (temperature and time) are appropriate for the amplification reaction and that the RT-PCR reagents are functional. The PCC verifies reagent rehydration, PCR tube filling, and confirms that all reaction components are present in the cartridge including monitoring for probe integrity and dye stability. The Xpert Xpress CoV-2/Flu/RSV plus test is designed for use with NPS or NS specimens collected with nylon flocked swabs and placed into viral transport medium (VTM), Universal Transport Medium (UTM), or eNAT®. The ancillary specimen collection kits, swabs and transport media validated for use with the Xpert Xpress CoV-2/Flu/RSV plus test include: - Nasopharyngeal Sample Collection Kit for Viruses - Copan UTM® 3C057N (Flexible Minitip Flocked Swab with UTM® Medium O without Beads) - Copan eNAT® Molecular Collection and Preservation Medium P/N 6U074S01 о (Flexible Minitip Flocked Swab with eNAT® Medium) - Becton Dickinson Universal Viral Transport Kit P/N 220531 (Flexible Minitip o Flocked Swab with UVT Medium) - Nasal Sample Collection Kit for Viruses - Copan UTM® 3C064N (Regular Flocked Swab with UTM® Medium without O Beads) - Copan eNAT® Molecular Collection and Preservation Medium P/N 6U073S01 O (Regular Flocked Swab with eNAT® Medium) - Alternatively, swabs and transport media can be obtained separately: ● - Nylon flocked swab (Copan P/N 502CS01, 503CS01) o - Viral transport medium, 3 mL (Copan P/N 330C, 3C047N, BD Universal O Transport Medium, Remel M4RT or Remel M5) These ancillary reagents allow NPS and NS specimens from patients to be collected, preserved and transported to laboratory prior to analysis with the Xpert Xpress CoV-2/Flu/RSV plus test.
    Ask a Question
    K Number
    K241806
    Device Name
    Applied Biosystems™ TaqPath™ COVID-19, Flu A, Flu B, RSV Select Panel
    Manufacturer
    Life Technologies Corporation
    Date Cleared
    2025-01-08

    (201 days)

    Product Code
    QOF
    Regulation Number
    866.3981
    Why did this record match?
    Product Code :

    QOF

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP Authorized
    Intended Use
    The Applied Biosystems TaqPath COVID-19, Flu A, Flu B, RSV Select Panel is a multiplex, real-time reverse transcription polymerase chain reaction (RT-PCR) in vitro diagnostic test for the qualitative detection and differentiation of severe acute respiratory syndrome coronavirus 2 (SARS-CoV-2), influenza B virus, and respiratory syncytial virus (RSV). Nucleic acids are isolated and purified from nasopharyngeal (NP) swab and anterior nasal (AN) swab specimens obtained from individuals exhibiting signs and symptoms of a respiratory tract infection. Clinical signs and symptoms of respiratory viral infection due to SARS-CoV-2, influenza, and RSV can be similar. This test is intended to aid in the differential diagnosis of SARS-CoV-2, influenza B, and RSV A/B (undifferentiated) infections in humans and is not intended to detect influenza C virus infections. Nucleic acids from the viral organisms identified by this test are generally detectable in NP and AN swab specimens during the acute phase of infection and identification of specific viral nucleic acids from individuals exhibiting signs and symptoms of respiratory tract infective of the presence of the identified virus and aids in diagnosis if used in conjunction with other clinical and epidemiological information, and laboratory findings. The results of this test should not be used as the sole basis for diagnosis, treatment, or other patient management decisions. Positive results do not rule out coinfection with other organism(s) detected by the Applied Biosystems TaqPath COVID-19, Flu A, Flu B, RSV Select Panel may not be the definite cause of disease. Negative results do not preclude SARS-CoV-2, influenza A virus, influenza B virus, or RSV infections. The Applied Biosystems TaqPath COVID-19, Flu A, Flu B, RSV Select Panel is intended for use by qualified and trained clinical laboratory personnel specifically instructed and trained in the techniques of real-time PCR and in vitro diagnostic procedures.
    Device Description
    The Applied Biosystems™ TaqPath™ COVID-19, Flu A, Flu B, RSV Select Panel is a multiplex, real-time reverse transcription polymerase chain reaction (RT-PCR) test. The severe acute respiratory syndrome coronavirus 2 (SARS-CoV-2), influenza A, influenza B, respiratory syncytial virus (RSV) A/B and RNase P primer and probe sets are designed to detect viral RNA in nasopharyngeal (NP) and anterior nasal (AN) swab specimens from individuals exhibiting signs and symptoms of a respiratory tract infection. Each TaqPath™ COVID-19, Flu A, Flu B, RSV Select Panel includes the following components: - TaqPath™ COVID-19, Flu A, Flu B, RSV Select Assay-Multiplex assays that contain . primer and probe sets specific to the following targets: - Three SARS-CoV-2 targets (Orfla, Orf1b, and N genes) . - . Two influenza A virus targets (PB1 & M genes) - I Two influenza B virus targets (M & NS genes) - . Three RSV targets (NP, M, and L protein genes) - l RNase P (internal human sample collection control) - TaqPath™ COVID-19, Flu A, Flu B, RSV Select Positive Control—Inactivated viral . control that contains SARS-CoV-2, influenza A, influenza B, and RSV. - TaqPath™ COVID-19, Flu A, Flu B, RSV Select Negative Control—MS2 packaged RNA . control that contains targets specific to RNase P genomic regions targeted by the assay. - TaqPath™ 1-Step Select Master Mix (No ROX)—Ready-to-use PCR mix, including . reverse transcriptase, polymerase, dNTPs, salts, and buffer. - . Package Insert -- Provides the instructions and the link to download the instructions for use and other assets (including the ADF) - An Assay Definition File (ADF) applicable to the instrument used in the workflow . (available via download). In addition to the SARS-CoV-2, influenza A, influenza B and RSV viral assay targets, the assay portion of the panel includes RNase P, which serves as an endogenous internal process control to monitor extraction and amplification of each clinical sample. The TaqPath™ COVID-19, Flu A, Flu B, RSV Select Panel also contains external process positive and negative controls. The positive control (PC) component included is an inactivated viral control that contains SARS-CoV-2, influenza A, influenza B, and RSV viruses. The PC monitors extraction and real-time RT-PCR by demonstrating that each of the four viruses can be detected when present and that RNase P is not detected when absent. The negative control (NC) component included is an MS2 packaged RNA control that contains targets specific to RNase P genomic regions targeted by the assay. The NC also monitors extraction and real-time RT-PCR by demonstrating RNase P can be detected when present and that the four viruses are not detected when absent. The TaqPath™ 1-Step Select Master Mix (No ROX) included as a component of the kit is a ready-to-use PCR mix which contains a deoxyribonucleotide triphosphate mix (dNTPs), enzymes, and other components to permit reverse transcription and amplification of the assay targets. The TagPath™ 1-Step Select Master Mix (No ROX) also contains ribonuclease (RNase) inhibitors as well as deoxyuridine triphosphate (dUTP) and uracil N-glycosylase (known as UNG or UDG). The TaqPath™ COVID-19, Flu A, Flu B, RSV Select Panel is provided in two overall kit configurations: either 200 reactions or 1.000 reactions.
    Ask a Question
    K Number
    K241652
    Device Name
    DASH® SARS-CoV-2 & Flu A/B Test
    Manufacturer
    Nuclein, LLC
    Date Cleared
    2024-12-20

    (196 days)

    Product Code
    QOF, OOF
    Regulation Number
    866.3981
    Why did this record match?
    Product Code :

    QOF

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP Authorized
    Intended Use
    The DASH® SARS-CoV-2 & Flu A/B Test is a rapid reverse transcription polymerase chain reaction (RT-PCR) assay performed on the DASH Rapid PCR Instrument and is intended for the simultaneous in vitro qualitative detection and differentiation of SARS-CoV-2, influenza A and influenza B virus ribonucleic acid (RNA) in anterior nasal swab specimens from patients with signs and symptoms of respiratory tract infection. The test is intended to aid in the differential diagnosis of SARS-CoV-2, influenza B in humans in conjunction with other clinical, epidemiologic and laboratory findings. Positive results of a specific target are indicative of that viral RNA and may not be the definite cause of disease. Positive results do not rule out co-infection with other pathogens. Negative results do not preclude SARS-COV-2, influenza A or influenza B infection and should not be used as the sole basis for patient management decisions.
    Device Description
    The DASH® SARS-CoV-2 & Flu A/B Test is a rapid, polymerase chain reaction (PCR) assay performed on the DASH Rapid PCR Instrument (DASH Instrument) with the DASH External Controls. The external control materials and DASH Instrument are sold and distributed separately from the DASH SARS-CoV-2 & Flu A/B Test. The DASH SARS-CoV-2 & Flu A/B Test (for use with the DASH Rapid PCR System components) uses reverse transcription polymerase chain reaction (RT-PCR) for rapid qualitative detection and differentiation of SARS-CoV-2, Flu A and Flu B from nasal swabs. The test combines the technologies of sequence specific capture sample preparation and RT-PCR amplification. The DASH SARS-CoV-2 & Flu A/B Test cartridge contains all reagents necessary to perform the test. An anterior nares nasal swab with a 30-mm breakpoint is used to collect a specimen. The nasal swab specimen is added directly to the DASH SARS-CoV-2 & Fly A/B Test cartridge sample chamber. The cartridge is capped and inserted into the DASH Rapid PCR Instrument to initiate the test, and all subsequent test steps are performed automatically by the DASH Instrument.
    Ask a Question
    K Number
    K242465
    Device Name
    Panther Fusion SARS-CoV-2/Flu A/B/RSV assay
    Manufacturer
    Hologic, Inc.
    Date Cleared
    2024-11-15

    (87 days)

    Product Code
    QOF, OOI
    Regulation Number
    866.3981
    Why did this record match?
    Product Code :

    QOF

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP Authorized
    Intended Use
    The Panther Fusion® SARS-CoV-2/Flu A/B/RSV assay is a fully automated multiplexed real-time polymerase chain reaction (RT-PCR) in vitro diagnostic test intended for the qualitative detection and differentiation of severe acute respiratory syndrome coronavirus 2 (SARS-CoV-2), influenza A virus (Flu A), influenza B virus (Flu B), and respiratory syncytial virus (RSV). Nucleic acids are isolated and purified from nasopharyngeal (NP) swab specimens and anterior nasal (AN) swab specimens obtained from individuals exhibiting signs and symptoms of a respiratory tract infection. Clinical signs and symptoms of respiratory viral infection due to SARS-CoV-2, influenza, and RSV can be similar. This assay is intended to aid in the differential diagnosis of SARS-CoV-2, Flu A, Flu B, and RSV infections in humans and is not intended to detect influenza C virus infections. Nucleic acids from the viral organisms identified by this test are generally detectable in NP and AN swab specimens during the acute phase of infection and identification of specific viral nucleic acids from individuals exhibiting signs and symptoms of respiratory tract infective of the presence of the identified virus and aids in diagnosis if used in conjunction with other clinical and epidemiological information, and laboratory findings. The results of this test should not be used as the sole basis for diagnosis, treatment, or other patient management decisions. Positive results do not rule out coinfection with other organism(s) detected by the Panther Fusion SARS-CoV-2/Flu A/B/RSV assay may not be the definite cause of disease. Negative results do not preclude SARS-CoV-2, influenza A virus, influenza B virus, or RSV infections. This assay is designed for use on the Panther Fusion system.
    Device Description
    The Panther Fusion SARS-CoV-2/Flu A/B/RSV assay is classified as a Class II in vitro diagnostic device per 21 CFR 866.3981 and has product code QOF. The Panther Fusion SARS-CoV-2/Flu A/B/RSV assay is designed for use on the fully automated Panther Fusion System. The Panther Fusion system is a class II exempt device under 21CFR 862.2570 that has product code OOI. The Panther Fusion SARS-CoV-2/Flu A/B/RSV assay is a multiplex real-time reverse transcriptase PCR (RT-PCR) in vitro diagnostic test developed for use on the fully automated Panther Fusion system to detect and differentiate SARS-CoV-2, influenza A, influenza B, and respiratory syncytial virus (RSV) directly from nasopharyngeal and anterior nasal swab specimens, from individuals exhibiting signs and symptoms of a respiratory tract infection. The Panther Fusion SARS-CoV-2/Flu A/B/RSV assay involves the following steps: sample lysis, nucleic acid capture and elution transfer, and multiplex RT-PCR where analytes (when present) are simultaneously amplified, detected and differentiated. Nucleic acid capture and elution takes place in a single tube on the Panther Fusion system. The eluate is transferred to the Panther Fusion system reaction tube containing the assay reagents. Multiplex RT-PCR is then performed for the eluted nucleic acid on the Panther Fusion system. Sample lysis, nucleic acid capture, and elution: Prior to processing and testing on the Panther Fusion system, specimens are transferred to a Specimen Lysis Tube containing specimen transport media (STM). Alternatively, samples can be collected with the RespDirect Collection kit which contains enhanced specimen transport media (eSTM). STM and eSTM lyse the cells, release target nucleic acid and protect them from degradation during storage. The Internal Control-S (IC-S) is added to each test specimen and controls via the working Panther Fusion Capture Reagent-S (wFCR-S). The IC-S is the reagent used to monitor specimen processing, amplification and detection. Magnetic particles with covalently bound oligonucleotides mediate the nucleic acid capture. Capture oligonucleotides hybridize to nucleic acid in the test specimen. Hybridized nucleic acid is then separated from the lysed specimen in a magnetic field. Wash steps remove extraneous components from the reaction tube. The elution step elutes purified nucleic acid. During the nucleic acid capture and elution step, total nucleic acid is isolated from specimens. Elution transfer and RT-PCR: During the elution transfer step, eluted nucleic acid is transferred to a Panther Fusion tube already containing oil and reconstituted mastermix. Target amplification occurs via RT-PCR. A reverse transcriptase generates a DNA copy of the target sequence. Target specific forward and reverse primers and probes then amplify targets while simultaneously detecting and discriminating multiple target types via multiplex RT-PCR. The Panther Fusion system compares the fluorescence signal to a predetermined cut-off to produce a qualitative result for the presence or absence of the analyte. The positive result for each analyte will be accompanied by the cycle threshold (Ct value).
    Ask a Question

    Page 1 of 3