K Number
K241573
Date Cleared
2025-02-14

(256 days)

Product Code
Regulation Number
866.3981
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Alinity m Resp-4-Plex is a multiplexed real-time in vitro reverse transcription polymerase chain reaction (RT-PCR) assay for use with the automated Alinity m System for the qualitative detection and differentiation of Severe Acute Respiratory Syndrome Coronavirus 2 (SARS-CoV-2), influenza A virus, influenza B virus and Respiratory Syncytial Virus (RSV) in nasopharyngeal swab specimens collected from patients with signs and symptoms of respiratory tract infection. Clinical signs and symptoms of respiratory tract infection due to SARS-CoV-2, influenza B, and RSV can be similar.

The Alinity m Resp-4-Plex assay is intended for use in the differential detection of SARS-CoV-2, influenza B and/or RSV RNA and aids in the diagnosis of COVID-19, influenza and/or RSV infections if used in conjunction with other clinical and epidemiological information, and laboratory findings. SARS-CoV-2, influenza B and RSV viral RNA are generally detectable in nasopharyngeal swab specimens during the acute phase of infection. This test is not intended to detect influenza C virus infections.

Positive results are indication of the identified virus, but do not rule out bacterial infection or co-infection with other pathogens not detected by the test. The agent(s) detected by the Alinity m Resp-4-Plex assay may not be the definite cause of disease.

Negative results do not preclude SARS-CoV-2, influenza B and/or RSV infections and should not be used as the sole basis for diagnosis, treatment or other patient management decisions.

Device Description

The Alinity m Resp-4-Plex assay requires two separate assay-specific kits: Alinity m Resp-4-Plex AMP Kit and Alinity m Resp-4-Plex CTRL Kit. The assay utilizes real-time PCR to amplify and detect genomic RNA sequences of influenza A (flu A), influenza B (flu B), RSV, and/or SARS-CoV-2 from nasopharyngeal (NP) swab specimens. The assay targets 2 different genes within the SARS-CoV-2 genome. Fluorescently labeled probes allow for simultaneous detection and differentiation of amplified products of all 4 viruses and Internal Control (IC) in a single reaction vessel. All steps of the assay procedure are executed automatically by the Alinity m System, which is a continuous random-access analyzer. The system performs automated sample preparation using magnetic microparticle technology. The IC is introduced into each specimen at the beginning of sample preparation. Purified RNA is combined with activation and amplification/detection reagents and transferred to a reaction vessel for reverse transcription, PCR amplification, and real-time fluorescence detection. A positive and negative control are tested to ensure performance. Patient results are automatically reported. The assay also utilizes the Alinity m Resp-4-Plex Assay Application Specification File, Alinity m System and System Software, Alinity m Sample Prep Kit 2, Alinity m Tubes and Caps, and Alinity m System Solutions.

AI/ML Overview

The provided text is a 510(k) Summary for the Abbott Molecular Inc. Alinity m Resp-4-Plex assay, a multiplexed real-time RT-PCR assay for the qualitative detection and differentiation of SARS-CoV-2, influenza A, influenza B, and RSV.

Here's an analysis of the acceptance criteria and the study that proves the device meets them, based on the provided document:

Acceptance Criteria and Reported Device Performance

The acceptance criteria for this device are demonstrated through various analytical and clinical studies, primarily focusing on analytical sensitivity (Limit of Detection), inclusivity, precision, reproducibility, analytical specificity (interfering substances and cross-reactants), competitive interference, carryover, and clinical performance (Positive Percent Agreement - PPA, and Negative Percent Agreement - NPA).

The document implicitly defines the acceptance criteria as the successful demonstration of performance metrics that are typically expected for such in vitro diagnostic devices, often compared to highly sensitive FDA-cleared or EUA assays. While specific numerical acceptance criteria (e.g., PPA > X%, NPA > Y%) are not explicitly stated as "acceptance criteria," they are demonstrated by the reported results. The conclusion statement (Section 5.0) explicitly states that the "analytical and clinical study results demonstrate that the Alinity m Resp-4-Plex assay... performs comparably to the predicate device... the results support a substantial equivalence decision." This implies that the demonstrated performance values met the FDA's criteria for substantial equivalence to the predicate device.

Here's a table summarizing the reported device performance, which implicitly met the acceptance criteria:

Table 1: Acceptance Criteria (as Demonstrated Performance) and Reported Device Performance

Performance CharacteristicAcceptance Criteria (Implied by Demonstrated Performance)Reported Device Performance
Analytical Sensitivity (LoD)Lowest concentration at which ≥ 95% of replicates test positive.Influenza A: 0.002 - 0.06 TCID50/mL (across 5 strains)Influenza B: 0.02 - 0.05 TCID50/mL (across 2 strains)RSV: 0.1 - 0.3 TCID50/mL (across 2 strains)SARS-CoV-2: 30 Genome Copies/mL (for 1 strain)
Inclusivity100% positive results for various strains at or below 3xLoD; >99.99% detection by in silico analysis for SARS-CoV-2.Flu A: 16 strains tested, lowest concentration yielding 100% positive results (e.g., 0.006 TCID50/mL, 3.33E+00 CEID50/mL).Flu B: 9 strains tested, lowest concentration yielding 100% positive results (e.g., 0.006 TCID50/mL, 2.78E-02 CEID50/mL).RSV: 6 strains tested, lowest concentration yielding 100% positive results (e.g., 0.03 TCID50/mL, 0.9 TCID50/mL).SARS-CoV-2: 9 strains tested, lowest concentration yielding 100% positive results (90 GC/mL or GE/mL).In silico (SARS-CoV-2): ≥ 99.99% of sequences predicted to be detected (14.8M GISAID, 7.6M NCBI).
PrecisionConsistent and reproducible results across multiple runs, days, and instruments.Flu A (Moderate/Low): Total %CV 1.8%/1.7%. 100% agreement.Flu B (Moderate/Low): Total %CV 1.1%/1.0%. 100% agreement.RSV (Moderate/Low): Total %CV 2.0%/2.2%. 100% agreement.SARS-CoV-2 (Moderate/Low): Total %CV 1.1%/1.4%. 100% agreement. (All negative samples 100% agreement).
ReproducibilityConsistent and reproducible results across external sites.Flu A (Moderate/Low): Total %CV 2.0%/2.0%. 100% agreement.Flu B (Moderate/Low): Total %CV 1.0%/2.1%. 100% agreement.RSV (Moderate/Low): Total %CV 2.8%/2.3%. 100% agreement.SARS-CoV-2 (Moderate/Low): Total %CV 1.1%/2.5%. 100% for moderate, 97.5% for low positive. Negative samples 99.7% agreement.
Analytical SpecificityNo interference from common substances; no cross-reactivity from other microorganisms.Interfering Substances: No interference observed for 34 tested substances (e.g., blood, mucin, nasal sprays, medications).Cross-Reactants: No cross-reactivity observed with 74 potential cross-reacting microorganisms (viruses, bacteria, fungi) at high concentrations. No interference on positive samples.
Competitive InterferenceAccurate detection of low concentration analytes in presence of high concentration of other analytes.All valid replicates of low concentration analytes reported positive results; high concentrations did not interfere.
CarryoverMinimal to no carryover between samples (e.g., from high positive to negative).Overall carryover rate of 0.0% (0/360) for SARS-CoV-2.
Clinical Performance (PPA/NPA) Prospective StudyHigh agreement with composite comparator.Flu A: PPA 100.0% (96.2, 100.0), NPA 99.6% (99.3, 99.8)Flu B: NPA 100.0% (99.8, 100.0) (PPA not calculated as no CC positive)RSV: PPA 98.0% (89.3, 99.6), NPA 99.7% (99.5, 99.9)SARS-CoV-2: PPA 95.3% (91.4, 97.5), NPA 96.0% (94.0, 97.4)
Clinical Performance (PPA/NPA) Retrospective StudyHigh agreement with composite comparator (especially for flu B where prospective data was limited).Flu A: NPA 99.4% (98.3, 99.8) (PPA not calculated as no CC positive)Flu B: PPA 100.0% (92.9, 100.0), NPA 98.5% (96.9, 99.3)RSV: NPA 100.0% (99.2, 100.0) (PPA not calculated as only 1 CC positive)

Study Details

Here's the breakdown of the study details as requested:

  1. A table of acceptance criteria and the reported device performance: Already provided above.

  2. Sample sizes used for the test set and the data provenance:

    • Analytical Studies (Test Set):

      • LoD: For each virus (Flu A, Flu B, RSV, SARS-CoV-2), preliminary LoD involved testing a minimum of 3 levels, each in a minimum of 3 replicates. Final LoD confirmation involved testing a minimum of 3 panel members with target concentrations bracketing the preliminary LoD, each panel member in a minimum of 20 replicates. (Total specific sample numbers not provided per virus, but this describes the method and minimums). Specimens were pooled negative NP clinical specimens.
      • Inclusivity: Each individual virus isolate or strain was tested in replicates of 5. (Total specific sample numbers not provided per virus, but 16 Flu A strains, 9 Flu B, 6 RSV, 9 SARS-CoV-2). Specimens were pooled negative clinical NP swab matrix.
      • Precision: 5 panel members (1 negative, 4 positive) tested with 4 replicates twice each day for 5 days, on 3 Alinity m Systems operated by 3 operators using 3 reagent lots. This leads to:
        • Flu A: 120 positive replicates for each level, 360 negative replicates.
        • Flu B: 120 positive replicates for each level, 357 negative replicates.
        • RSV: 120 positive replicates for moderate, 117 for low, 360 negative.
        • SARS-CoV-2: 120 positive replicates for each level, 357 negative.
      • Reproducibility: 5 panel members tested at 3 external clinical testing sites. Each site tested 2 Alinity m Resp-4-Plex AMP Kit lots, on 5 non-consecutive days for each lot. Four replicates of each panel member were tested on each of 5 days. This leads to:
        • Flu A: 120 positive replicates for each level, 360 negative replicates.
        • Flu B: 120 positive replicates for each level, 359 negative replicates.
        • RSV: 120 positive replicates for moderate, 119 for low, 360 negative.
        • SARS-CoV-2: 120 positive replicates for moderate, 117 for low, 359 negative.
      • Analytical Specificity (Interfering Substances): 34 substances evaluated in 2 different positive panel members (PM1 & PM2), each containing multiple analytes at 3xLoD. (Replicate number not specified).
      • Analytical Specificity (Cross-Reactants): 74 microorganisms added to pooled negative clinical NP swab matrix (replicate number not specified) and also to positive samples (replicate number not specified).
      • Competitive Interference: 4 panel members, each containing 3 viruses at low concentrations and one at high concentration. (Replicate number not specified).
      • Carryover: Negative and high positive samples tested in alternating positions, across 3 Alinity m Systems. 360 negative samples total.
    • Clinical Performance (Test Set):

      • Prospective Clinical Study:
        • Flu A/B, RSV: 2,753 valid results initially, 2,504 (Flu A), 2,710 (Flu B), 2,700 (RSV) used in analysis.
          • Data Provenance: Multicenter study using prospectively collected nasopharyngeal swab specimens. 4 US clinical sites for testing. Specimens collected during 2021-2022 flu season at 7 geographically distributed locations in the US and during the 2020 flu season at 1 location in the Southern Hemisphere.
        • SARS-CoV-2: 826 valid results initially, 698 used in analysis.
          • Data Provenance: Specimens collected at 10 geographically distributed locations in the US over 2 time periods (Dec 2020 - Feb 2021 and May 2023).
      • Retrospective Clinical Study:
        • Flu A/B, RSV: 515 valid results initially, 506 (Flu A), 504 (Flu B), 505 (RSV) used in analysis.
          • Data Provenance: Preselected archived flu B positive NP swab specimens in UVT or UTM collected during the 2017-2018 and 2019-2020 flu seasons. Randomly mixed with known negative specimens.
  3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • The ground truth for the clinical test sets (both prospective and retrospective) was established using a Composite Comparator (CC). This CC was based on results from "2 to 3 FDA cleared assays for flu A, flu B, and RSV" and "2 to 3 highly sensitive EUA SARS-CoV-2 molecular assays."
    • The document does not specify the number or qualifications of human experts (e.g., radiologists, pathologists) involved in establishing this ground truth. The ground truth method described is entirely based on laboratory comparator assays, not expert human interpretation of results.
  4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

    • The ground truth was established by a Composite Comparator (CC) method:
      • A specimen was categorized as CC positive if a minimum of 2 comparator positive results were reported.
      • A specimen was categorized as CC negative if a minimum of 2 comparator negative results were reported.
      • A specimen was categorized CC indeterminate if a CC could not be determined due to missing results from the comparator assays.
    • This functions as a type of "majority rule" adjudication or consensus, but strictly between other molecular assays, not human experts.
  5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • No, an MRMC comparative effectiveness study was not done.
    • This device is an in vitro diagnostic (RT-PCR assay) that provides a qualitative (positive/negative) result directly. It does not involve human "readers" interpreting images or other complex data that would typically benefit from AI assistance or an MRMC study design. Therefore, there's no data on human reader improvement with or without AI assistance.
  6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    • Yes, the primary performance evaluation is a standalone algorithm-only performance.
    • The Alinity m Resp-4-Plex assay is an automated RT-PCR system. Its performance (PPA, NPA) in both analytical and clinical studies is the performance of the "algorithm only" in generating positive/negative results from the sample. Human intervention is limited to sample collection and system operation, not interpretation of the primary diagnostic output.
  7. The type of ground truth used (expert consensus, pathology, outcomes data, etc):

    • The ground truth for the clinical studies was established using a Composite Comparator (CC) method based on the results of other FDA-cleared molecular diagnostic assays (or EUA high-sensitivity molecular assays for SARS-CoV-2).
    • It was not based on expert consensus, pathology, or outcomes data directly.
  8. The sample size for the training set:

    • The document describes a de novo device, or a device for which substantial equivalence is being sought, not an AI/ML device that requires a distinct "training set" and "test set" in the context of model development.
    • The studies presented are primarily verification and validation studies to demonstrate the device's analytical and clinical performance after its development.
    • Therefore, the concept of a separate "training set" as understood in machine learning (where data is used to train a model) is not applicable to this RT-PCR assay. The assay's "knowledge" is embedded in its reagents, primers, probes, and system parameters, which were likely optimized during development using various analytical samples, but these are not typically referred to as a "training set" in the context of a 510(k) for an RT-PCR assay.
  9. How the ground truth for the training set was established:

    • As explained in point 8, there isn't a "training set" in the AI/ML sense for this RT-PCR assay. The ground truth for any samples used during the development or optimization phases would similarly be established using well-characterized samples (e.g., cultured viruses, positive clinical samples confirmed by reference methods, synthetic nucleic acids, negative clinical samples).

{0}------------------------------------------------

February 14, 2025

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA acronym and name on the right. The Department of Health & Human Services logo is a stylized depiction of a human figure. The FDA acronym is in a blue square, and the words "U.S. FOOD & DRUG ADMINISTRATION" are in blue text to the right of the square.

Abbott Molecular Inc Stacy Ferguson Director Regulatory Affairs 1300 E Touhy Ave Des Plaines, Illinois 60018

Re: K241573

Trade/Device Name: Alinity m Resp-4-Plex Regulation Number: 21 CFR 866.3981 Regulation Name: Device To Detect And Identify Nucleic Acid Targets In Respiratory Specimens From Microbial Agents That Cause The SARS-Cov-2 Respiratory Infection And Other Microbial Agents When In A Multi-Target Test Regulatory Class: Class II Product Code: QOF Dated: May 31, 2024 Received: June 3, 2024

Dear Stacy Ferguson:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (OS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safetyreporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatory

{2}------------------------------------------------

assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Anna M. Mielech -S

Anna Mielech, PhD. Deputy Branch Chief (Acting) Viral Respiratory and HPV Branch Division of Microbiology Devices OHT7: Office of In Vitro Diagnostics Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

Indications for Use

510(k) Number (if known) K241573

Device Name

Alinity m Resp-4-Plex

Indications for Use (Describe)

Almity m Resp-4-Plex is a multiplexed real-time in vitro reverse transcription polymerase chain reaction (RT-PCR) assay for use with the automated Alinity m System for the qualitative detection and differentiation of Severe Acute Respiratory Syndrome Coronavirus 2 (SARS-CoV-2), influenza A virus, influenza B virus and Respiratory Syncytial Virus (RSV) in nasopharyngeal swab specimens collected from patients with signs and symptoms of respiratory tract infection. Clinical signs and symptoms of respiratory tract infection due to SARS-CoV-2, influenza B, and RSV can be similar.

The Alinity m Resp-4-Plex assay is intended for use in the differential detection of SARS-CoV-2, influenza B and/or RSV RNA and aids in the diagnosis of COVID-19, influenza and/or RSV infections if used in conjunction with other clinical and epidemiological information, and laboratory findings. SARS-CoV-2, influenza B and RSV viral RNA are generally detectable in nasopharyngeal swab specimens during the acute phase of infection. This test is not intended to detect influenza C virus infections.

Positive results are indication of the identified virus, but do not rule out bacterial infection or co-infection with other pathogens not detected by the test. The agent(s) detected by the Alinity m Resp-4-Plex assay may not be the definite cause of disease.

Negative results do not preclude SARS-CoV-2, influenza B and/or RSV infections and should not be used as the sole basis for diagnosis, treatment or other patient management decisions.

Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

510(k) Summary

Table of Contents
1.0Submitter2
1.1Device Information3
1.2Predicate Device3
1.3Intended Use4
2.0Device Description5
3.0Similarities and Differences to the Predicate Device8
4.0Performance Data12
4.1Specific Performance Characteristics – Analytical Studies12
4.1.1Limit of Detection (Analytical Sensitivity)
4.1.2Inclusivity
4.1.3Precision
4.1.4Reproducibility
4.1.5Analytical Specificity: Potentially Interfering Substances
4.1.6Analytical Specificity: Potential Cross-Reactants
4.1.7On-Panel Cross-Reactivity
4.1.8Competitive Interference
4.1.9Carryover
4.2Clinical Performance28
4.2.1Performance with Clinical Specimens Prospective Clinical Study
4.2.2Retrospective Clinical Study
4.2.3Expected Values
5.0Conclusion Drawn from the Studies35

{5}------------------------------------------------

510(k) Summary

This 510(k) summary of safety and effectiveness information is being submitted in accordance with the requirement of 21 CFR Section 807.92(c).

1.0 Submitter

Applicants Name and Address:Abbott Molecular Inc.1300 E. Touhy Ave Des Plaines, IL 60018
Contact Person:Stacy FergusonDirector Regulatory AffairsAbbott Molecular, Inc.1300 E. Touhy AvenueDes Plaines, IL 60018Phone: 224-361-7449
Date Prepared:February 12, 2025

{6}------------------------------------------------

1.1 Device Information

Device Trade NameRegulation NameProductCodeRegulationNumberClass
Alinity m Resp-4-PlexMulti-Target Respiratory SpecimenNucleic Acid Test IncludingSars-Cov-2 And Other MicrobialAgentsQOF21 CFR 866.3981II
1.2 Predicate Device
Predicate Device510(k)Date Cleared
Hologic™ Panther Fusion SARS-CoV-2/Flu A/B/RSV AssayK222736May 16, 2023

{7}------------------------------------------------

1.3 Intended Use

Alinity m Resp-4-Plex is a multiplexed real-time in vitro reverse transcription polymerase chain reaction (RT-PCR) assay for use with the automated Alinity m System for the qualitative detection and differentiation of Severe Acute Respiratory Syndrome Coronavirus 2 (SARS-CoV-2), influenza A virus, influenza B virus and Respiratory Syncytial Virus (RSV) in nasopharyngeal swab specimens collected from patients with signs and symptoms of respiratory tract infection. Clinical signs and symptoms of respiratory tract infection due to SARS-CoV-2, influenza A, influenza B, and RSV can be similar.

The Alinity m Resp-4-Plex assay is intended for use in the differential detection of SARS-CoV-2, influenza A, influenza B and/or RSV RNA and aids in the diagnosis of COVID-19, influenza and/or RSV infections if used in conjunction with other clinical and epidemiological information, and laboratory findings. SARS-CoV-2, influenza A, influenza B and RSV viral RNA are generally detectable in nasopharyngeal swab specimens during the acute phase of infection. This test is not intended to detect influenza C virus infections.

Positive results are indication of the presence of the identified virus, but do not rule out bacterial infection or co-infection with other pathogens not detected by the test. The agent(s) detected by the Alinity m Resp-4-Plex assay may not be the definite cause of disease.

Negative results do not preclude SARS-CoV-2, influenza B and/or RSV infections and should not be used as the sole basis for diagnosis, treatment or other patient management decisions.

{8}------------------------------------------------

2.0 Device Description

The Alinity m Resp-4-Plex assay requires two separate assay-specific kits as follows:

  • Alinity m Resp-4-Plex AMP Kit; 09N79-095, consists of 2 types of multi-well . assay trays. The amplification trays (AMP TRAY 1) contain unit-dose liquid PCR amplification/detection reagents and unit-dose liquid Internal Control (IC) in separate wells. The activation trays (ACT TRAY 2) contain liquid unit-dose activation reagent. The intended storage condition for the Alinity m Resp-4-Plex AMP Kit is -25°C to -15°C.
  • . Alinity m Resp-4-Plex CTRL Kit; 09N79-085, consists of negative controls and positive controls, each supplied as liguid in single-use tubes. The Alinity m Resp-4-Plex controls are used for validity determination of the Alinity m Resp-4-Plex on the automated Alinity m System. The intended storage condition for the Alinity m Resp-4-Plex CTRL Kit is -25°C to -15°C.

The Alinity m Resp-4-Plex assay utilizes real-time PCR to amplify and detect genomic RNA sequences of influenza A (flu A), influenza B (flu B), RSV, and/or SARS-CoV-2 from nasopharyngeal (NP) swab specimens. The Alinity m Resp-4-Plex detects assay analytes and internal control (IC) separately, using analyte specific probes. The assay targets 2 different genes within the SARS-CoV-2 genome. The fluorescently labeled probes do not generate a detectable signal unless they are specifically bound to the amplified product. All probes are labeled with analyte specific fluorophores, thus allowing for simultaneous detection and differentiation of amplified products of all 4 viruses and IC in a single reaction vessel.

The steps of the Alinity m Resp-4-Plex assay consist of sample preparation, PCR assembly, amplification/ detection, and result calculation and reporting. All steps of the Alinity m Resp-4-Plex assay procedure are executed automatically by the Alinity m System. The Alinity m System is designed to be a continuous random-access analyzer that can perform the Alinity m Resp-4-Plex assay in parallel with other Alinity m assays on the same instrument. One PCR reaction can detect 1 or more pathogens with the Alinity m Resp-4-Plex assay. Therefore, only 1 patient specimen aliquot is required for detection of the selected assay(s).

The Alinity m System performs automated sample preparation using the Alinity m Sample Prep Kit 2, Alinity m Lysis Solution, and Alinity m Diluent Solution. The Alinity m System employs magnetic microparticle technology to facilitate nucleic acid capture, wash, and

{9}------------------------------------------------

elution. The IC is introduced into each specimen at the beginning of the sample preparation process to demonstrate that the process was completed correctly for each specimen and control sample.

The resulting purified RNA is combined with liquid unit-dose Alinity m Resp-4-Plex activation reagent and liquid unit-dose Alinity m Resp-4-Plex amplification/detection reagent and transferred into a reaction vessel. Alinity m Vapor Barrier Solution is then added to the reaction vessel which is then transferred to an amplification/detection unit for reverse transcription, PCR amplification, and real-time fluorescence detection.

A positive control and a negative control are tested at or above an established minimum frequency to help ensure that instrument and reagent performance remain satisfactory. During each control event, a negative control and positive control are processed through sample preparation and PCR procedures that are identical to those used for specimens.

The target sequences for the Alinity m Resp-4-Plex assay are:

  • RdRp and N genes of the SARS-CoV-2 genome .
  • the Matrix gene of the flu A genome .
  • . the Nonstructural 1 gene of the flu B genome
  • the Matrix gene of the RSV genome

Patient results are automatically reported on the Alinity m instrument. The Alinity m Resp-4-Plex application parameters will be contained in an assay application specification file.

{10}------------------------------------------------

The Alinity m Resp-4-Plex assay also utilizes the following:

  • Alinity m Resp-4-Plex Assay Application Specification File, List No. 09N79-03A . The Alinity m Resp-4-Plex application specification file is intended for use with the Alinity m Resp-4-Plex assay on the automated Alinity m System to allow for processing of assay controls and patient samples.
  • Alinity m System and System Software, List No. 08N53-002ª
  • . Alinity m Sample Prep Kit 2, List No. 09N12-001ª
  • Alinity m Tubes and Caps, List No. 09N49ª
    • Alinity m Transport Tubes Pierceable Capped, List No. 09N49-010 ●
    • Alinity m Transport Tube, List No. 09N49-011
    • . Alinity m Aliquot Tube, List No. 09N49-013
  • . Alinity m System Solutions, List No. 09N20
    • . Alinity m Lysis Solution, List No. 09N20-001ª
    • Alinity m Diluent Solution, List No. 09N20-003ª
    • . Alinity m Vapor Barrier Solution, List No. 09N20-004a

ಡ Items previously approved by FDA under PMA P190025.

{11}------------------------------------------------

3.0 Similarities and Differences to the Predicate Device

The legally marketed predicate device chosen for the current submission is the Hologic™ Panther Fusion SARS-CoV-2/Flu A/B/RSV Assay (for use on the Panther Fusion system). The Alinity m Resp-4-Plex assay is substantially equivalent to the legally marketed predicate device, Hologic™ Panther Fusion SARS-CoV-2/Flu A/B/RSV Assay (for use on the Panther Fusion system), intended for the qualitative detection and differentiation of SARS-CoV-2, flu A virus, flu B virus, and RSV. The primary similarities between Alinity m Resp-4-Plex assay and the predicate device are presented in Table 1. The primary differences between Alinity m Resp-4-Plex and the predicate device are shown in Table 2. Both the Alinity m Resp-4-Plex assay and the predicate have the same intended use. Any technological differences that exist between Alinity m Resp-4-Plex assay and the predicate device do not raise new types of safety or effectiveness questions.

{12}------------------------------------------------

Table 1. Similarities Between Alinity m Resp-4-Plex Assay and Nucleic Acid Amplification Tests Predicate Device
FeatureCurrent SubmissionPredicate Device
Device Trade NameAlinity m Resp-4-PlexHologic™ Panther Fusion SARS-CoV-2/Flu A/B/RSV Assay(K222736)
Intended UseAlinity m Resp-4-Plex is a multiplexed real-time in vitro reversetranscription polymerase chain reaction (RT-PCR) assay for usewith the automated Alinity m System for the qualitative detectionand differentiation of Severe Acute Respiratory SyndromeCoronavirus 2 (SARS-CoV-2) influenza A virus, influenza B virusand Respiratory Syncytial Virus (RSV), in nasopharyngeal swabspecimens collected from patients with signs and symptoms ofrespiratory tract infection. Clinical signs and symptoms ofrespiratory tract infection due to SARS-CoV-2, influenza A,influenza B, and RSV can be similar.The Alinity m Resp-4-Plex assay is intended for use in thedifferential detection of SARS-CoV-2, influenza A, influenza Band/or RSV RNA and aids in the diagnosis of COVID-19,influenza and/or RSV infections if used in conjunction with otherclinical and epidemiological information, and laboratory findings.SARS-CoV-2, influenza A, influenza B and RSV viral RNA aregenerally detectable in nasopharyngeal swab specimens during theacute phase of infection. This test is not intended to detectinfluenza C virus infections.Positive results are indication of the presence of the identifiedvirus, but do not rule out bacterial infection or co-infection withother pathogens not detected by the test. The agent(s) detected bythe Alinity m Resp-4-Plex assay may not be the definite cause ofdisease.Negative results do not preclude SARS-CoV-2, influenza A,influenza B, and/or RSV infections and should not be used as thesole basis for diagnosis, treatment or other patient managementdecisions.The Panther Fusion SARS-CoV-2/Flu A/B/RSV assay is a fullyautomated multiplexed real-time polymerase chain reaction (RT-PCR) in vitro diagnostic test intended for the qualitative detection anddifferentiation of severe acute respiratory syndrome coronavirus 2(SARS-CoV-2), influenza A virus (Flu A), influenza B virus (Flu B),and respiratory syncytial virus (RSV). Nucleic acids are isolated andpurified from nasopharyngeal (NP) specimens obtained fromindividuals exhibiting signs and symptoms of a respiratory tractinfection. Clinical signs and symptoms of respiratory viral infectiondue to SARS-CoV-2, influenza, and RSV can be similar. This assay isintended to aid in the differential diagnosis of SARS-CoV-2, Flu A,Flu B, and RSV infections in humans and is not intended to detectinfluenza C virus infections.Nucleic acids from the viral organisms identified by this test aregenerally detectable in NP specimens during the acute phase ofinfection. The detection and identification of specific viral nucleicacids from individuals exhibiting signs and symptoms of respiratorytract infection are indicative of the presence of the identified virusand aids in diagnosis if used in conjunction with other clinical andepidemiological information, and laboratory findings. The results ofthis test should not be used as the sole basis for diagnosis, treatment,or other patient management decisions.Positive results do not rule out coinfection with other organisms. Theorganism(s) detected by the Panther Fusion SARS-CoV-2/ FluA/B/RSV assay may not be the definite cause of disease.Negative results do not preclude SARS-CoV-2, influenza A virus,influenza B virus, or RSV infections. This assay is designed for useon the Panther Fusion system.
Table 1. Similarities Between Alinity m Resp-4-Plex Assay and Nucleic Acid Amplification Tests Predicate Device
FeatureCurrent SubmissionPredicate Device
Device Trade NameAlinity m Resp-4-PlexHologic™ Panther Fusion SARS-CoV-2/Flu A/B/RSV Assay(K222736)
Conditions for UseFor prescription useFor in vitro diagnostic use onlySame
Organisms Detectedinfluenza A (flu A) influenza B (flu B) Respiratory Syncytial Virus (RSV) SARS-CoV-2Same
PlatformAll steps of this qualitative assay procedure areexecuted automatically by the Alinity m System.Uses Alinity m System for all steps including nucleicacid extraction, amplification, detection, and resultprocessing.No intermediate processing or transfer steps areperformed by the user.All steps of this qualitative assay procedure are executed automatically bythe Panther Fusion System. Uses Panther Fusion System for all stepsincluding nucleic acid extraction, amplification, detection, and resultprocessing.No intermediate processing or transfer steps are performed by the user.
Specimen TypesNasopharyngeal (NP) swab specimensSame
Assay ControlsNegative Control Positive Control Internal Control (IC)Same

{13}------------------------------------------------

{14}------------------------------------------------

Table 2. Differences Between Alinity m Resp-4-Plex Assay and Nucleic Acid Amplification Tests Predicate Device
FeatureCurrent SubmissionPredicate Device
Device Trade NameAlinity m Resp-4-PlexHologic™ Panther Fusion SARS-CoV-2/Flu A/B/RSV Assay (K222736)
Specimen Collectionand Transport• BDTM UVT• Copan UTM®• Same• Same• Remel MicroTest M4RT, M5, or M6 formulations• Hardy Diagnostics Viral Transport Medium• Hologic RespDirect Collection Kit

{15}------------------------------------------------

4.0 Performance Data

The following performance data were provided in support of the substantial equivalence determination.

Specific Performance Characteristics – Analytical Studies 4.1

4.1.1 Limit of Detection (Analytical Sensitivity)

Limit of Detection (LoD) studies determine the lowest concentration of Influenza A (flu A), Influenza B (flu B), RSV, and SARS-CoV-2 at which greater than or equal to 95% of replicates test positive.

The LoD was determined by testing 10 cultured viruses, including 5 flu A strains (H1N1 and H3N2 subtypes), 2 flu B strains (Victoria and Yamagata lineages), 2 RSV strains (RSV A and RSV B), and 1 SARS-CoV-2 strain diluted in pooled negative NP clinical specimens collected in viral transport media.

For each virus, the preliminary LoD was determined by testing a minimum of 3 levels, each in a minimum of 3 replicates. The final LoD was confirmed by testing a minimum of 3 panel members with target concentrations bracketing the preliminary LoD, each panel member in a minimum of 20 replicates. For each strain, LoD was defined as the lowest concentration at which greater than or equal to 95% of all replicates tested positive rate greater or equal to 95%), as summarized in Table 3.

{16}------------------------------------------------

Table 3. Limit of Detection
VirusStrain (Source)LoDMean CN at LoDa
Influenza AA/H1N1/Brisbane/59/2007(Cat No 0810244CF, Lot 323919)0.002 TCID50/mL36.29
A/H1N1/ Michigan/45/2015(Cat No 0810538CF, Lot 326537)0.004 TCID50/mL35.75
A/H3N2/Switzerland/9715293/13(Cat No 0810511CF, Lot 322440)0.015 TCID50/mL35.71
A/H3N2/Wisconsin/04/2018(Cat No FR-1692, Lot 70030518)0.06 TCID50/mL35.15
A/H3N2/ Darwin/9/21(Cat No 0810650CF, Lot 331100)0.004 TCID50/mL36.02
Influenza BB/Victoria lineage/Brisbane/33/08(Cat No 0810253CF, Lot 316752)0.02 TCID50/mL36.23
B/Yamagatalineage/Massachusetts/2/12(Cat No 0810239CF, Lot 324519)0.05 TCID50/mL36.13
RSVRSVA/Long/MD/56(Cat No VR-26PQ, Lot 70024412)0.3 TCID50/mL34.92
RSVB/West Virginia/14617/85(Cat No VR-1400, Lot 70013461)0.1 TCID50/mL35.62
SARS-CoV-2Isolate USA-WA1/2020, gammairradiated(Cat No NR-52287, Lot 70048202)30 GenomeCopies/mL35.93

a Mean cycle number (CN) was calculated for all valid replicates with a positive result in the confirmation of LoD testing. TCID50/mL = Median Tissue Culture Infectious Dose/mL.

{17}------------------------------------------------

4.1.2 Inclusivity

In addition to strains tested in LoD study, the inclusivity of Alinity m Resp-4-Plex assay for the detection of flu A, flu B, RSV, and SARS-CoV-2 was also evaluated by testing 16 strains of flu A virus (including H1N1, H3N2, and H5N15), 9 strains of flu B virus (including Victoria and Yamagata lineages), 6 strains of RSV (including RSV A and B), and 9 strains of SARS-CoV-2 (including A, B, BA.1 and B.1.1.7 lineages) in pooled negative clinical NP swab matrix. Each individual virus isolate or strain (cultured virus or viral RNA) was tested in replicates of 5 at the level equal to or lower than 3 × LoD (Table 4). For strains with reported units of measure different than units listed in LoD study, dilution series were prepared and tested. The lowest concentration tested that resulted in 100% positive results is included in Table 4.

Table 4. Inclusivity
VirusStrain/IsolateCatalogNumberaLot NumberConcentrationTestedb
Influenza AA/H1N1/Brisbane/02/20180810585CFHI3254330.006TCID50/mL
A/H1N1/Fort Monmouth/1/1947VR-1754595234911.67E+00CEID50/mL
A/H1N1/California/07/2009VR-1894700148333.33E+00CEID50/mL
A/H1N1/Puerto Rico/8/1934NR-348700131223.33E+00CEID50/mL
A/H1N1/New Jersey/8/1976VR-897588105881.00E+01CEID50/mL
A/H1N1/Wisconsin/588/2019 – Candidate VaccineVirusN/A30260543331.37E-06 HA
A/H1N1/Guangdong-Maonan/1536/2019 – CandidateVaccine VirusN/A30012939008.53E-06 HA
A/H1N1/Victoria/2570/2019 –Candidate Vaccine VirusN/A30008275195.12E-05 HA
A/H3N2/Tasmania/503/2020 –Candidate Vaccine VirusN/A30008244302.37E-07 GP
A/H3N2/Hong Kong/45/2019 –Candidate Vaccine VirusN/A30015844042.56E-06 GP
VirusStrain/IsolateCatalogNumberaLot NumberConcentrationTestedb
A/H3N2/Hong Kong/2671/2019- Candidate Vaccine VirusN/A30012924004.27E-05 GP
A/H3N2/Singapore/INFIMH-16-0019/20160810574CF3272980.045TCID50/mL
A/H3N2/Brisbane/10/20070810138CF3246940.045TCID50/mL
A/H3N2/Perth/16/20090810251CF3132190.045TCID50/mL
A/H3N2/Delaware/01/2021FR-1748700483063.64E-01FFU/mL
A/H5N1/Vietnam/1203/2004NR-12145VNV1F014A9.00E-05µg/mL
Influenza BB/Victorialineage/Brisbane/60/20080810254CF3255270.06TCID50/mL
B/Victorialineage/Colorado/6/20170810573CF3256510.02TCID50/mL
B/Victorialineage/Alabama/2/20170810572CF3255400.02TCID50/mL
B/Victorialineage/Malaysia/2506/20040810258CF3276090.006TCID50/mL
B/Victorialineage/Washington/02/2019 –Candidate Vaccine VirusN/A30260195371.42E-06 HA
B/Yamagatalineage/Phuket/3073/2013 –Candidate Vaccine VirusN/A20147686192.96E-07 HA
B/Yamagata lineage/GreatLakes/1739/1954VR-103642957161.67E+00CEID50/mL
Lee/19400810257CF3132590.06TCID50/mL
Allen/1945VR-102700212782.78E-02CEID50/mL
Table 4. Inclusivity
VirusStrain/IsolateCatalogNumberaLot NumberConcentrationTestedb
RSVRSVB/3/2015 Isolate 10810479CF3227410.1 TCID50/mL
RSVB/12/2014 Isolate 10810450CF3187980.03TCID50/mL
RSVB/11/2014 Isolate 20810451CF3187960.03TCID50/mL
RSVA/12/2014 Isolate 120810462CF3188410.9 TCID50/mL
RSVA/1/2015 Isolate 10810466CF3188430.09TCID50/mL
RSVA/2014 Isolate 3410810290CF3159110.09TCID50/mL
SARS-CoV-2USA/GA-EHC-2811C/2021(Lineage BA.1; OmicronVariant)NR-564967005069490 GC/mL
Hong Kong/VM20001061/2020(Lineage A)cNR-523887003467990 GE/mL
USA-AZ1/2020 (Lineage A)cNR-525057003525690 GE/mL
USA-IL1/2020 (Lineage A)cNR-525037003525490 GE/mL
USA-CA3/2020 (Lineage B)cNR-525077003525890 GE/mL
USA-WI1/2020 (Lineage B)cNR-525067003525790 GE/mL
Germany/BavPat1/2020(Lineage B.1)cNR-525027003618190 GE/mL
USA/CA_CDC_5574/2020(Lineage B.1.1.7; Alpha Variant)cNR-552447004256090 GE/mL
Italy-INMI1cNR-524987003526190 GE/mL

{18}------------------------------------------------

Table 4 Inclusivity

{19}------------------------------------------------

a N/A indicates Catalog Number is not applicable.

b TCIDsolmL = Median Tissue Culture Infectious Dose/mL ; CEIDso/mL = median chicken embryo infectious dose/mL; HA = Hemagglutination titer, indicates Hemagglutinin levels; FFU = focus-forming units/mL; GC/mL = Genome Copies/mL; GE/mL = Genome Equivalent/mL; GP = Glycoprotein titer, indicates Glycoprotein levels.

Genomic RNA C

In addition, in silico analyses were performed in which the sequence of each of the SARS-CoV-2 analyte primers and probes was analyzed for homology with sequences from human isolates available in GISAID and NCBI databases.

For SARS-CoV-2, in silico analysis of the primer/probe set for homology with all 14,818,776 virus sequences available in the GISAID database as of October 3, 2023 and all 7,590,332 virus Alinity m Resp-4-Plex Page 16 of 35 February 2025

{20}------------------------------------------------

sequences available in the NCBI database as of October 9, 2023 showed that greater than or equal to 99.99% of the sequences are predicted to be detected by the Alinity m Resp-4-Plex assay.

4.1.3 Precision

Alinity m Resp-4-Plex assay within-laboratory precision was evaluated by testing 5 panel members: 1 flu A/flu B/RSV/SARS-CoV-2 negative panel member, 2 positive panel members containing flu A and RSV (low positive targeted at 2 × LoD and moderate positive targeted at 5 × LoD for both viruses), and 2 positive panel members containing flu B and SARS-CoV-2 (low positive targeted at 2 × LoD and moderate positive targeted at 5 × LoD for both viruses). All panel members were prepared in simulated nasal matrix (SNM). Each panel member was tested with 4 replicates twice each day for 5 days, on 3 Alinity m Systems operated by 3 operators using 3 reagent lots (1 reagent lot and 1 operator per system). The results for each analyte at each target level are summarized in Table 5.

{21}------------------------------------------------

Table 5. Precision

NanbAgreement(n/N)MeanCNWithin-RunComponentBetween-RunComponentBetween-DayComponentWithin-LaboratorycBetween-InstrumentComponentdTotale
AnalyteTarget LevelSD% CVSD% CVSD% CVSD% CVSD% CVSD% CV
flu AModeratePositive120120100.0%31.700.5011.60.1090.30.1150.40.5261.70.2230.70.5711.8
Low Positive117117100.0%33.040.5121.50.0000.00.0000.00.5121.50.2310.70.5621.7
Negativef360360100.0%-------------
flu BModeratePositive120120100.0%28.930.1790.60.0580.20.0000.00.1880.60.2540.90.3161.1
Low Positive120120100.0%30.220.1630.50.0490.20.0580.20.1800.60.2560.80.3131.0
Negativef357357100.0%-------------
RSVModeratePositive120120100.0%29.910.5541.90.2050.70.0860.30.5972.00.1360.50.6122.0
Low Positive117117100.0%31.080.6222.00.0000.00.2530.80.6712.20.1960.60.6992.2
Negativef360360100.0%-------------
SARS-CoV-2ModeratePositive120120100.0%32.800.2910.90.0000.00.1050.30.3090.90.2160.70.3771.1
Low Positive120120100.0%34.230.3921.10.0000.00.1140.30.4091.20.2750.80.4931.4
Negativef357357100.0%-------------

a N = Total number valid replicates.

b n = Number of replicates with positive passive panel member of replicates with negative result interpretation for negative parel member; the number of replicates used in the Mean Cycle Number (CN) and SD calculations for the positive panel members.

c Within-laboratory includes Within-Run, Between-Run, and Between-Day Components.

d Almity m System, Alinity m Resp-4-Plex AMP Kit lot, and Operator are confounding effect is represented by Instrument.

e Total includes Within-Run, Between-Run, Between-Day, and Between-Instrument Components.

f The negative results included 3 panel members negative for the respective analyte.

{22}------------------------------------------------

Reproducibility 4.1.4

Alinity m Resp-4-Plex assay reproducibility was evaluated at 3 external clinical testing sites. A total of 5 panel members were tested: 1 flu A/flu B/RSV/SARS-CoV-2 negative panel member, 2 positive panel members containing flu A and RSV (low positive targeted at 2 × LoD and moderate positive targeted at 5 × LoD for both viruses), and 2 positive panel members containing flu B and SARS-CoV-2 (low positive targeted at 2 × LoD and moderate positive targeted at 5 × LoD for both viruses). All panel members were prepared in SNM. A total of 3 Alinity m Resp-4-Plex AMP Kit lots were used. Each of the 3 external sites tested 2 Alinity m Resp-4-Plex AMP Kit lots, on 5 non-consecutive days for each lot. Four replicates of each panel member were tested on each of 5 days. Each of the 3 external sites used different lots of Alinity m Resp-4-Plex CTRL Kits and Alinity m Sample Prep Kit 2. The reproducibility results are summarized in Table 6.

{23}------------------------------------------------

Table 6. Reproducibility

NanbAgreement(n/N)MeanCNWithin-Run/DayComponentBetween-Run/DayComponentWithin-LaboratorycBetween-LotComponentBetween-Site/InstrumentComponentTotald
AnalyteTarget LevelSD% CVSD% CVSD% CVSD% CVSD% CVSD% CV
flu AModeratePositive120120100.0%30.920.5681.80.1490.50.5871.90.2070.70.0000.00.6232.0
Low Positive119119100.0%32.180.5921.80.2080.60.6282.00.0000.00.0650.20.6312.0
Negativee360360100.0%-------------
flu BModeratePositive120120100.0%28.880.2200.80.1190.40.2500.90.0810.30.0810.30.2751.0
Low Positive120120100.0%30.080.5902.00.1230.40.6032.00.0000.00.1880.60.6312.1
Negativee359359100.0%-------------
RSVModeratePositive120120100.0%29.090.7362.50.3141.10.8002.80.0000.00.0000.00.8002.8
Low Positive119119100.0%30.350.6652.20.0000.00.6652.20.1200.40.1550.50.6932.3
Negativee360360100.0%-------------
SARS-CoV-2ModeratePositive120120100.0%32.560.3251.00.0910.30.3371.00.0560.20.0290.10.3431.1
Low Positive12011797.5%34.000.7762.30.2780.80.8242.40.0000.00.1750.50.8432.5
Negativee36035999.7%-------------

a N = Total number valid replicates.

b Number of replicates with positive result interpretation for plicates with negative result interpretation for regative parel member, the number of replicates used in the Mean Cycle Number (CN) and SD calculations for the positive panel members.

c Within-laboratory includes Within-Run/Day and Between-Run/Day Components.

d Total includes Within-Run/Day, Between-Run/Day, Between-Site/Instrument Variance Components.

C The negative results included 3 panel members negative for the respective analyte.

{24}------------------------------------------------

4.1.5 Analytical Specificity: Potentially Interfering Substances

A total of 34 endogenous and exogenous substances that may be encountered in respiratory specimens were assessed as potentially interfering substances for Alinity m Resp-4-Plex assay. Each of the potentially interfering substances were evaluated in 2 different positive panel members (PM) each containing multiple analytes at 3 × LoD in pooled negative clinical NP swab matrix.

  • . PM 1: influenza A (H3N2), influenza B (Victoria lineage), RSV B, and SARS-CoV-2
  • PM 2: influenza A (H31N1), influenza B (Yamagata lineage), and RSV A .

All positive samples reported positive results for flu A, flu B, RSV, and SARS-CoV-2 in the presence of the potentially interfering substances. No interference in the performance of Alinity m Resp-4-Plex assay was observed for the tested substances at the concentrations listed in Table 7.

SubstanceActive Ingredient(s)TestedConcentration
Antibacterial, SystemicTobramycin4 μg/mL
Antibiotic, Nasal Ointment - MupirocinMupirocin5 mg/mL
Anti-Viral Drug - Relenza™MZanamivir5 mg/mL
Anti-Viral Drug - RemdesivirRemdesivir27.0 µM
BloodN/A10% (v/v)
Cold Eeze® (Throat lozenges)Zincum gluconicum 2X2.5% (w/v)
Cough Syrup – Wal-Tussin®Dextromethorphan5% (v/v)
FluMist®ªLive intranasal influenza virus10% (v/v)
Human Genomic DNAN/A20 ng/uL
LeukocytesLeukocytes1.1E6 cells/mL
Liposomal NUMB520 SprayLidocaine and Phenylephrine2.68 mg/mL
Mucin-BovinePurified mucin protein5 mg/mL
Mucin-PorcinePurified mucin protein5 mg/mL
Nasal Corticosteroid - BudesonideBudesonide2% (v/v)

Table 7. Potentially Interfering Substances

{25}------------------------------------------------

SubstanceActive Ingredient(s)TestedConcentration
Nasal Corticosteroid - DexamethasoneDexamethasone0.2 mg/mL
Nasal Corticosteroid - Flonase®Sensimist™Fluticasone Furoate10% (v/v)
Nasal Corticosteroid - FlunisolideFlunisolide2% (v/v)
Nasal Corticosteroid - MometasoneMometasone2% (v/v)
Nasal Corticosteroid - QVAR®Beclomethasone2% (v/v)
Nasal Corticosteroid - TriamcinoloneTriamcinolone2% (v/v)
Nasal Decongestant - PhenylephrinePhenylephrine2% (v/v)
Nasal Gel / Homeopathic AllergyRelief Medicine - Zicam®Galphimia glauca, Histaminum,hydrochloricum, Luffa operculata, Sulfur10% (v/v)
Nasal Spray - Perrigo OxymetazolineHClOxymetazoline15% (v/v)
Nicotine PouchNicotine0.1% (w/v)
Oral rinse- Listerine® Cool Mint®Ethanol, essential oil10% (v/v)
Saline Nasal MistSodium chloride2% (v/v)
SalivaN/A10% (v/v)
Tamiflu®Oseltamivir3.3 mg/mL
Throat Lozenges, Oral Anesthetic andAnalgesic - Cepacol®Benzocaine, Menthol5 mg/mL
Throat Spray - Chloraseptic®Phenol5% (v/v)
Tobacco PouchNicotine0.1% (w/v)
ToothpasteFluoride1% (w/v)
Vaseline®Petroleum Jelly1% (w/v)
Vicks® VapoRub™Camphor-synthetic, eucalyptus oil andmenthol ointment1% (w/v)

Table 7. Potentially Interfering Substances

FluMistø was not tested with negative panel but is produce positive results for flu A and flu B due to the presence of attenuated viruses. Refer to FluMist package insert for further information.

{26}------------------------------------------------

4.1.6 Analytical Specificity: Potential Cross-Reactants

A total of 74 potential cross-reacting microorganisms (viruses, bacteria, and fungi) that are phylogenetically related to the analytes of the assay or that are commonly found in respiratory tract and pooled human nasal wash were tested with Alinity m Resp-4-Plex assay to assess analytical specificity (cross-reactivity) and microbial interference. The microorganisms were tested at 1.00E+5 units/mL for viruses and fungi and 1.00E+6 units/mL for bacteria or the highest titer available (as indicated in Table 8). The unit of measure was specific to each microorganism. The microorganisms are listed in Table 8.

Cross-reactivity was evaluated by adding each of the potential cross reactants in pooled flu A/flu B/RSV/SARS-CoV-2 negative clinical NP swab matrix.

No cross-reactivity was observed with the potential cross-reactants at the concentrations tested.

In addition, the impact of the cross-reactants listed in Table 8 on the detection of flu A, flu B, RSV, and SARS-CoV-2 was evaluated by testing them at 1.00E+5 units/mL for viruses and fungi and 1.00E+6 units/mL for bacteria or the highest titer available (as indicated in Table 8) in flu A/flu B/RSV/ SARS-CoV-2 positive samples targeted to 3 × LoD in pooled negative clinical NP swab matrix. All positive samples reported positive results for flu A, flu B, RSV, and SARS-CoV-2 in the presence of these potential cross-reactants at the concentrations tested.

{27}------------------------------------------------

Table 8. Potential Cross-Reactants

BacteriaViruses
Bordetella pertussisLegionella pneumophilaAdenovirus Type 01gHuman Metapneumovirus 9 Type 1Ag
Chlamydia pneumoniaeMoraxella catarrhalisAdenovirus Type 02gHuman Metapneumovirus 18 Type 2Ba, g
Chlamydophila psittaciMycobacterium tuberculosisAdenovirus Type 03gInfluenza A (H1N1)b, g, h
Corynebacterium diphtheriaMycoplasma pneumoniaeAdenovirus Type 04gInfluenza A (H1N1) pdm09b, g
Coxiella burnettiNeisseria elongataAdenovirus Type 05 gInfluenza A (H1N3)b, g
Cutibacterium acnesNeisseria meningitidisAdenovirus Type 07AgInfluenza A (H3N2)b, g, h
Enterococcus faecalisProteus mirabilisAdenovirus Type 11gInfluenza Bc, g
Escherichia coliPseudomonas aeruginosaAdenovirus Type 14a, gInfluenza C
Haemophilus influenzaeStaphylococcus aureusAdenovirus Type 31gMeaslesg
Klebsiella pneumoniaeStaphylococcus epidermisAdenovirus Type 34gMERS-Coronavirusg
Lactobacillus gasseriStreptococcus pneumoniaeBocavirusjMumpsg
Lactobacillus plantarumStreptococcus pyogenesCoxsackievirusgParainfluenza Virus Type 1g
Legionella longbeachaeStreptococcus salivariusCytomegalovirusgParainfluenza Virus Type 2g
FungiEchovirusgParainfluenza Virus Type 3g
Aspergillus fumigatusEnterovirus Type 68gParainfluenza Virus Type 4g
Candida albicansdEnterovirus EgParechovirus Type 3g
Pneumocystis jirovecii (PJP)eEpstein-Barr VirusgRhinovirus 1Ag, h
OtherHerpes Simplex Virus Type 1gRhinovirus B42g
Pooled Human Nasal Wash (10%)Human Coronavirus 229EgRSV Af,g
Human Coronavirus HKU1iRSV Bf, g, h
Human Coronavirus NL63gSARS-Coronavirusg
Human Coronavirus OC43gVaricella-Zoster Virusg
Human Metapneumovirus 5 Type B1g

Tested at 1.00E+04 units/mL. a

b Only the cross-reactivity of the non-flu A signals were evaluated.

c Only the cross-reactivity of the non-flu B signals were evaluated.

d Tested at 1.00E+06 units/mL

e Pneumocystis jirovecii was not measured in terms of copy number. Instead, it was measured in cycle threshold (Ct). This specimen was tested neat.

f Only the cross-reactivity of the non-RSV signals were evaluated.

as Viral lysate.

Alinity m Resp-4-Plex February 2025

Viral particles. h

I Viral RNA.

Viral DNA. j

Page 24 of 35

{28}------------------------------------------------

4.1.7 On-Panel Cross-Reactivity

The cross-reactivity of each signal channel of the Alinity m Resp-4-Plex assay for on-panel viruses (flu A, flu B, RSV, and SARS-CoV-2) was evaluated by testing 2 different flu A strains (H1N1 and H3N2), 2 different flu B strains (Yamagata and Victoria lineage), 2 different RSV strains (RSV A and RSV B), and 1 SARS-CoV-2 strain. Individual viral strains were diluted in SNM at 1.00E+05 units/mL in the absence of other assay analytes. No cross-reactivity was observed for on-panel viruses at the concentrations tested.

4.1.8 Competitive Interference

Potential competitive interference between flu A, flu B, RSV, and SARS-CoV-2 analytes was evaluated by testing the 4 panel members each containing 3 viruses at low concentrations in the presence of the fourth virus at a high concentration:

  • PM 1: flu A, flu B, and RSV at < 3 × LoD (Low) and SARS-CoV-2 at >1000 × LoD (High) .
  • PM 2: flu A, flu B, and SARS-CoV-2 at ≤ 3 × LoD (Low) and RSV at >1000 × LoD (High) .
  • PM 3: flu A, RSV, and SARS-CoV-2 at ≤ 3 × LoD (Low) and flu B at >1000 × LoD (High) .
  • PM 4: flu B, RSV, and SARS-CoV-2 at ≤ 3 × LoD (Low) and flu A at >1000 × LoD (High) .

All valid replicates of the analytes tested at the low concentrations reported positive results. None of the analytes present at the high concentrations interfered with the detection of the other analytes.

4.1.9 Carryover

The carryover rate for Alinity m Resp-4-Plex assay was evaluated by testing negative and high positive samples containing 2.00E+09 Copies/mL SARS-CoV-2 (as the representative analyte) in alternating positions, across 3 Alinity m Systems. SARS-CoV-2 RNA was not detected in any of the 360 negative samples, resulting in an overall carryover rate of 0.0% (0/360, 95% CI: 0.0%,1.1%).

{29}------------------------------------------------

Clinical Performance 4.2

4.2.1 Performance with Clinical Specimens Prospective Clinical Study

The performance of the Alinity m Resp-4-Plex assay was evaluated in a multicenter study using prospectively collected nasopharyngeal swab specimens in UVT or UTM from individuals presenting with signs and symptoms of respiratory tract infection. A total of 4 US clinical sites tested the specimens with the Alinity m Resp-4-Plex assay. All 4 sites tested specimens for flu A, flu B, and RSV performance, and 3 of the 4 sites tested specimens for SARS-CoV-2 performance.

For the evaluation of flu A, flu B, and RSV performance, the Alinity m Resp-4-Plex results were compared to a composite comparator (CC) established using 2 to 3 FDA cleared assays for flu A, flu B, and RSV. For the evaluation of SARS-CoV-2 performance, the Alinity m Resp-4-Plex results were compared to CC established using 2 to 3 highly sensitive EUA SARS-CoV-2 molecular assays. A specimen was categorized as CC positive if a minimum of 2 comparator positive results were reported. A specimen was categorized as CC negative if a minimum of 2 comparator negative results were reported. A specimen was categorized CC indeterminate if a CC could not be determined due to missing results from the comparator assays.

The specimens for flu A, flu B, and RSV performance evaluation were collected during the 2021-2022 flu season at 7 geographically distributed locations in the US and during the 2020 flu season at 1 location in the Southern Hemisphere. Valid Alinity m Resp-4- Plex results included 2,753 each for flu A and flu B analytes (from 2,453 US specimens and 300 specimens from outside the US), and 2,745 for RSV analyte (from 2,445 US specimens and 300 specimens from outside the US). Of these, 2,504 flu A analyte results, 2,710 flu B analyte results, and 2,700 RSV analyte results of the Alinity m Resp-4-Plex assay were used in the analysis. CC could not be determined for 249 specimens for flu A. 43 specimens for flu B, and 45 specimens for RSV.

{30}------------------------------------------------

The specimens for SARS-CoV-2 performance evaluation were collected at 10 geographically distributed locations in the US over 2 time periods; ie, between December 2020 and February 2021 and in May 2023. A total of 826 valid SARS-CoV-2 analyte results of the Alinity m Resp-4-Plex assay were obtained. Of these, 698 SARS-CoV-2 analyte results were used in the analysis, while 128 specimens did not have CC.

Demographic characteristics for the prospective population that were included in the performance analysis are shown in Table 9 for the evaluation of flu A, flu B, and RSV analytes and Table 10 for SARS-CoV-2 analyte.

The performance of the Alinity m Resp-4-Plex assay with prospective specimens is summarized in Table 11. Alinity m Resp-4-Plex results were compared to CC by the analysis of positive percent agreement (PPA) and negative percent agreement (NPA). Alinity m Resp-4-Plex yielded PPA of 100.0% and NPA of 99.6% for the flu A analyte, NPA of 100.0% for the flu B analyte, PPA of 98.0% and NPA of 99.7% for the RSV analyte, and PPA of 95.3% and NPA of 96.0% for the SARS-CoV-2 analyte. PPA was not calculated for the flu B analyte because there was no CC positive specimen in the prospective study.

{31}------------------------------------------------

CharacteristicStatistic
Age
n2,710
Mean37.0
Median35
SD28.02
Minimum0
Maximum103
Age Groupn (%)
Birth to 5 years600 (22.1%)
6 to 21 years359 (13.2%)
22 to 59 years1051 (38.8%)
≥ 60 years700 (25.8%)
Gendern (%)
Female1438 (53.1%)
Male1271 (46.9%)
Unknown1 (0.0%)
Ethnicityn (%)
Hispanic or Latino1104 (40.7%)
Not Hispanic or Latino1442 (53.2%)
Unknown164 (6.1%)
Racen (%)
White1070 (39.5%)
Otherᵃ672 (24.8%)
Black or African American378 (13.9%)
Asian136 (5.0%)
American Indian or Alaska Native2 (0.1%)
Native Hawaiian or Other Pacific Islander2 (0.1%)
Unknown450 (16.6%)

Table 9. Summary of Demographic Characteristics for Subjects Included in the Analysis -Prospective Population Evaluation of flu A, flu B, and RSV Analytes

a Documentation of races other than the categories listed were grouped under "Other".

{32}------------------------------------------------

CharacteristicStatistic
Age
n698
Mean41.3
Median41
SD16.88
Minimum2
Maximum87
Age Groupn (%)
Birth to 5 years5 (0.7%)
6 to 21 years80 (11.5%)
22 to 59 years497 (71.2%)
≥ 60 years116 (16.6%)
Gendern (%)
Female405 (58.0%)
Male293 (42.0%)
Ethnicityn (%)
Hispanic or Latino261 (37.4%)
Not Hispanic or Latino413 (59.2%)
Unknown24 (3.4%)
Racen (%)
White508 (72.8%)
Black or African American163 (23.4%)
Asian10 (1.4%)
Native Hawaiian or Other Pacific Islander2 (0.3%)
American Indian or Alaska Native1 (0.1%)
Unknown14 (2.0%)

Table 10. Summary of Demographic Characteristics for Subjects Included in the Analysis - Prospective Population Evaluation of SARS-CoV-2 Analyte

{33}------------------------------------------------

PPA (%)NPA (%)
AnalyteSampleTypeNCC+Alinity+CC+Alinity-CC-Alinity-CC-Alinity+Estimate(95% CI)n/NEstimate(95% CI)n/N
flu AFresh219194020889100.0 (96.1, 100.0)94/9499.6 (99.2, 99.8)2088/2097
Frozen313403090100.0 (51.0, 100.0)4/4100.0 (98.8, 100.0)309/309
Overall250498023979100.0 (96.2, 100.0)98/9899.6 (99.3, 99.8)2397/2406
flu BFresh23840023831N/AaN/A100.0 (99.8, 100.0)2383/2384
Frozen326003260N/AaN/A100.0 (98.8, 100.0)326/326
Overall27100027091N/AaN/A100.0 (99.8, 100.0)2709/2710
RSVFresh23744812318798.0 (89.3, 99.6)48/4999.7 (99.4, 99.9)2318/2325
Frozen326003260N/AaN/A100.0 (98.8, 100.0)326/326
Overall27004812644798.0 (89.3, 99.6)48/4999.7 (99.5, 99.9)2644/2651
SARS-CoV-2Fresh65118294431795.3 (91.3, 97.5)182/19196.3 (94.2, 97.7)443/460
Frozen4720423100.0 (34.2, 100.0)2/293.3 (82.1, 97.7)42/45
Overall69818494852095.3 (91.4, 97.5)184/19396.0 (94.0, 97.4)485/505

Table 11. Alinity m Resp-4-Plex Assay Performance with Prospectively Collected Specimens

PPA was not calculated because there was no CC positive specimen in this study.

a

{34}------------------------------------------------

4.2.2 Retrospective Clinical Study

Flu B prevalence was very low during the prospective clinical study; there was no comparator positive specimen to evaluate the performance of the Alinity m Resp-4-Plex assay. To supplement the results of the prospective specimen population, retrospective specimen testing was performed using preselected archived flu B positive NP swab specimens in UVT or UTM collected during the 2017-2018 and 2019-2020 flu seasons, which were randomly mixed with known negative specimens to target 10% flu B prevalence. The specimens were tested with both Alinity m Resp-4-Plex and FDA cleared comparators for flu A, flu B, and RSV. Five hundred fifteen valid Alinity m Resp-4-Plex results were obtained for each of the flu A, flu B, and RSV analytes. Of these, 506 flu A analyte results, 504 of flu B analyte results, and 505 RSV analyte results were used in the analysis. CC could not be determined for 9 specimens for flu A, 11 specimens for flu B, and 10 specimens for RSV.

Demographic characteristics for the subjects included in the analysis from the retrospective population are shown in Table 12.

The performance of the Alinity m Resp-4-Plex assay with retrospective specimens is summarized in Table 13. Alinity m Resp-4-Plex yielded NPA of 99.4% for the flu A analyte, PPA of 100.0% and NPA of 98.5% for the flu B analyte, and NPA of 100.0% for the RSV analyte. PPA was not calculated for flu A or RSV because there was no CC positive for flu A and only 1 CC positive for RSV in the retrospective study.

{35}------------------------------------------------

CharacteristicStatistic
Age
n506
Mean22.1
Median9
SD24.86
Minimum0
Maximum93
Age Groupn (%)
Birth to 5 years219 (43.3%)
6 to 21 years99 (19.6%)
22 to 59 years139 (27.5%)
≥ 60 years49 (9.7%)
Gendern (%)
Female246 (48.6%)
Male260 (51.4%)
Ethnicityn (%)
Hispanic or Latino257 (50.8%)
Not Hispanic or Latino78 (15.4%)
Unknown171 (33.8%)
Racen (%)
Othera173 (34.2%)
White109 (21.5%)
Black or African American36 (7.1%)
Asian9 (1.8%)
American Indian or Alaska Native1 (0.2%)
Unknown178 (35.2%)

Table 12. Summary of Demographic Characteristics for Subjects Included in the Final Analysis – Retrospective Population

a Documentation of races other than the categories listed were grouped under "Other".

{36}------------------------------------------------

AnalyteNCC+Alinity+CC+Alinity-CC-Alinity-CC-Alinity+Positive Percent Agreement (%)Negative Percent Agreement (%)
Estimate (95% CI)n/NEstimate (95% CI)n/N
flu A506005033N/AaN/A99.4 (98.3, 99.8)503/506
flu B5045004477100.0 (92.9, 100.0)50/5098.5 (96.9, 99.3)447/454
RSV505105040N/AaN/A100.0 (99.2, 100.0)504/504

Table 13. Alinity m Resp-4-Plex Assay Performance with Retrospective Archived Specimens

PPA was not calculated because there was no CC positive for flu A and only 1 CC positive for RSV in this study. a

{37}------------------------------------------------

4.2.3 Expected Values

The Alinity m Resp-4-Plex assay included a total of 874 prospectively collected NPS specimens that were tested for SARS-CoV-2, of which 698 results were evaluable. Influenza A, influenza B, and RSV included 2,753 each for flu A and flu B analytes and 2,745 for RSV. Of these, 2,504 flu A analyte results, 2,710 flu B analyte results, and 2,700 RSV analyte results of the Alinity m Resp-4-Plex assay were evaluable. Positivity, as determined by the Alinity m Resp-4-Plex AMP Kit, are presented in Tables 14 - 16 for SARS-CoV-2 stratified by collection site and collection period and influenza A, and RSV stratified by collection site.

Collection SiteNExpected Value
Site 1 (Florida)17138.6% (66/171)
Site 2 (Georgia)6121.3% (13/61)
Site 3 (Tennessee)10820.4% (22/108)
Site 4 (Florida)2548.0% (12/25)
Site 5 (Ohio)6437.5% (24/64)
Site 6 (California)3467.6% (23/34)
Site 7 (Ohio)5641.1% (23/56)
Site 8 (Texas)7727.3% (21/77)
Overall59634.2% (204/596)

Table 14: Alinity m Resp-4-Plex Expected Results by Specimen Collection Site for Specimens Collected from December 2020 to February 2021 for SARS-CoV-2

{38}------------------------------------------------

Collection SiteNExpected Value
Site 1 (Texas)650.0% (0/65)
Site 2 (Florida)370.0% (0/37)
Overall1020.0% (0/102)

Table 15: Alinity m Resp-4-Plex Expected Results by Specimen Collection Site for Specimens Collected in May 2023 for SARS-CoV-2

Table 16: Alinity m Resp-4-Plex Expected Results by Specimen Collection Site for Influenza A, Influenza B, and RSV
Collection SiteInfluenza AInfluenza BRSV
NExpected ValuesNExpected ValuesNExpected Values
Site 1 (Florida)3000.0% (0/300)3010.0% (0/301)3010.0% (0/301)
Site 2 (Illinois)11852.2 (26/1185)12220.1% (1/1222)12212.4% (29/1221)
Site 3 (California)2116.2% (13/211)3090.0% (0/309)3075.2% (16/307)
Site 4 (New York)34916.0% (56/349)4040.0% (0/404)4011.7% (7/401)
Site 5 (New Jersey)979.3% (9/97)970.0% (0/97)932.2% (2/93)
Site 6 (Florida)150.0% (0/15)150.0% (0/15)150.0% (0/15)
Site 7 (Georgia)565.4% (3/56)620.0% (0/62)621.6% (1/62)
Site 8a2910.0% (0/291)3000.0% (0/300)3000.0% (0/300)
Overall25044.3% (107/2504)27100.0% (1/2710)27002.0% (55/2700)

ª(OUS, Southern Hemisphere)

5.0 Conclusion Drawn from the Studies

The analytical and clinical study results demonstrate that the Alinity m Resp-4-Plex assay on the Alinity m System performs comparably to the predicate device in qualitative detection and differentiation of influenza A virus (flu A), influenza B virus (flu B), Respiratory Syncytial Virus (RSV), and Severe Acute Respiratory Syndrome Coronavirus 2 (SARS-CoV-2); the results support a substantial equivalence decision for the Alinity m Resp-4-Plex assay.

§ 866.3981 Device to detect and identify nucleic acid targets in respiratory specimens from microbial agents that cause the SARS-CoV-2 respiratory infection and other microbial agents when in a multi-target test.

(a)
Identification. A device to detect and identify nucleic acid targets in respiratory specimens from microbial agents that cause the SARS-CoV-2 respiratory infection and other microbial agents when in a multi-target test is an in vitro diagnostic device intended for the detection and identification of SARS-CoV-2 and other microbial agents when in a multi-target test in human clinical respiratory specimens from patients suspected of respiratory infection who are at risk for exposure or who may have been exposed to these agents. The device is intended to aid in the diagnosis of respiratory infection in conjunction with other clinical, epidemiologic, and laboratory data or other risk factors.(b)
Classification. Class II (special controls). The special controls for this device are:(1) The intended use in the labeling required under § 809.10 of this chapter must include a description of the following: Analytes and targets the device detects and identifies, the specimen types tested, the results provided to the user, the clinical indications for which the test is to be used, the specific intended population(s), the intended use locations including testing location(s) where the device is to be used (if applicable), and other conditions of use as appropriate.
(2) Any sample collection device used must be FDA-cleared, -approved, or -classified as 510(k) exempt (standalone or as part of a test system) for the collection of specimen types claimed by this device; alternatively, the sample collection device must be cleared in a premarket submission as a part of this device.
(3) The labeling required under § 809.10(b) of this chapter must include:
(i) A detailed device description, including reagents, instruments, ancillary materials, all control elements, and a detailed explanation of the methodology, including all pre-analytical methods for processing of specimens;
(ii) Detailed descriptions of the performance characteristics of the device for each specimen type claimed in the intended use based on analytical studies including the following, as applicable: Limit of Detection, inclusivity, cross-reactivity, interfering substances, competitive inhibition, carryover/cross contamination, specimen stability, precision, reproducibility, and clinical studies;
(iii) Detailed descriptions of the test procedure(s), the interpretation of test results for clinical specimens, and acceptance criteria for any quality control testing;
(iv) A warning statement that viral culture should not be attempted in cases of positive results for SARS-CoV-2 and/or any similar microbial agents unless a facility with an appropriate level of laboratory biosafety (
e.g., BSL 3 and BSL 3+, etc.) is available to receive and culture specimens; and(v) A prominent statement that device performance has not been established for specimens collected from individuals not identified in the intended use population (
e.g., when applicable, that device performance has not been established in individuals without signs or symptoms of respiratory infection).(vi) Limiting statements that indicate that:
(A) A negative test result does not preclude the possibility of infection;
(B) The test results should be interpreted in conjunction with other clinical and laboratory data available to the clinician;
(C) There is a risk of incorrect results due to the presence of nucleic acid sequence variants in the targeted pathogens;
(D) That positive and negative predictive values are highly dependent on prevalence;
(E) Accurate results are dependent on adequate specimen collection, transport, storage, and processing. Failure to observe proper procedures in any one of these steps can lead to incorrect results; and
(F) When applicable (
e.g., recommended by the Centers for Disease Control and Prevention, by current well-accepted clinical guidelines, or by published peer-reviewed literature), that the clinical performance may be affected by testing a specific clinical subpopulation or for a specific claimed specimen type.(4) Design verification and validation must include:
(i) Detailed documentation, including performance results, from a clinical study that includes prospective (sequential) samples for each claimed specimen type and, as appropriate, additional characterized clinical samples. The clinical study must be performed on a study population consistent with the intended use population and compare the device performance to results obtained using a comparator that FDA has determined is appropriate. Detailed documentation must include the clinical study protocol (including a predefined statistical analysis plan), study report, testing results, and results of all statistical analyses.
(ii) Risk analysis and documentation demonstrating how risk control measures are implemented to address device system hazards, such as Failure Modes Effects Analysis and/or Hazard Analysis. This documentation must include a detailed description of a protocol (including all procedures and methods) for the continuous monitoring, identification, and handling of genetic mutations and/or novel respiratory pathogen isolates or strains (
e.g., regular review of published literature and periodic in silico analysis of target sequences to detect possible mismatches). All results of this protocol, including any findings, must be documented and must include any additional data analysis that is requested by FDA in response to any performance concerns identified under this section or identified by FDA during routine evaluation. Additionally, if requested by FDA, these evaluations must be submitted to FDA for FDA review within 48 hours of the request. Results that are reasonably interpreted to support the conclusion that novel respiratory pathogen strains or isolates impact the stated expected performance of the device must be sent to FDA immediately.(iii) A detailed description of the identity, phylogenetic relationship, and other recognized characterization of the respiratory pathogen(s) that the device is designed to detect. In addition, detailed documentation describing how to interpret the device results and other measures that might be needed for a laboratory diagnosis of respiratory infection.
(iv) A detailed device description, including device components, ancillary reagents required but not provided, and a detailed explanation of the methodology, including molecular target(s) for each analyte, design of target detection reagents, rationale for target selection, limiting factors of the device (
e.g., saturation level of hybridization and maximum amplification and detection cycle number, etc.), internal and external controls, and computational path from collected raw data to reported result (e.g., how collected raw signals are converted into a reported signal and result), as applicable.(v) A detailed description of device software, including software applications and hardware-based devices that incorporate software. The detailed description must include documentation of verification, validation, and hazard analysis and risk assessment activities, including an assessment of the impact of threats and vulnerabilities on device functionality and end users/patients as part of cybersecurity review.
(vi) For devices intended for the detection and identification of microbial agents for which an FDA recommended reference panel is available, design verification and validation must include the performance results of an analytical study testing the FDA recommended reference panel of characterized samples. Detailed documentation must be kept of that study and its results, including the study protocol, study report for the proposed intended use, testing results, and results of all statistical analyses.
(vii) For devices with an intended use that includes detection of Influenza A and Influenza B viruses and/or detection and differentiation between the Influenza A virus subtypes in human clinical specimens, the design verification and validation must include a detailed description of the identity, phylogenetic relationship, or other recognized characterization of the Influenza A and B viruses that the device is designed to detect, a description of how the device results might be used in a diagnostic algorithm and other measures that might be needed for a laboratory identification of Influenza A or B virus and of specific Influenza A virus subtypes, and a description of the clinical and epidemiological parameters that are relevant to a patient case diagnosis of Influenza A or B and of specific Influenza A virus subtypes. An evaluation of the device compared to a currently appropriate and FDA accepted comparator method. Detailed documentation must be kept of that study and its results, including the study protocol, study report for the proposed intended use, testing results, and results of all statistical analyses.
(5) When applicable, performance results of the analytical study testing the FDA recommended reference panel described in paragraph (b)(4)(vi) of this section must be included in the device's labeling under § 809.10(b) of this chapter.
(6) For devices with an intended use that includes detection of Influenza A and Influenza B viruses and/or detection and differentiation between the Influenza A virus subtypes in human clinical specimens in addition to detection of SARS-CoV-2 and similar microbial agents, the required labeling under § 809.10(b) of this chapter must include the following:
(i) Where applicable, a limiting statement that performance characteristics for Influenza A were established when Influenza A/H3 and A/H1-2009 (or other pertinent Influenza A subtypes) were the predominant Influenza A viruses in circulation.
(ii) Where applicable, a warning statement that reads if infection with a novel Influenza A virus is suspected based on current clinical and epidemiological screening criteria recommended by public health authorities, specimens should be collected with appropriate infection control precautions for novel virulent influenza viruses and sent to State or local health departments for testing. Viral culture should not be attempted in these cases unless a BSL 3+ facility is available to receive and culture specimens.
(iii) Where the device results interpretation involves combining the outputs of several targets to get the final results, such as a device that both detects Influenza A and differentiates all known Influenza A subtypes that are currently circulating, the device's labeling must include a clear interpretation instruction for all valid and invalid output combinations, and recommendations for any required followup actions or retesting in the case of an unusual or unexpected device result.
(iv) A limiting statement that if a specimen yields a positive result for Influenza A, but produces negative test results for all specific influenza A subtypes intended to be differentiated (
i.e., H1-2009 and H3), this result requires notification of appropriate local, State, or Federal public health authorities to determine necessary measures for verification and to further determine whether the specimen represents a novel strain of Influenza A.(7) If one of the actions listed at section 564(b)(1)(A) through (D) of the Federal Food, Drug, and Cosmetic Act occurs with respect to an influenza viral strain, or if the Secretary of Health and Human Services determines, under section 319(a) of the Public Health Service Act, that a disease or disorder presents a public health emergency, or that a public health emergency otherwise exists, with respect to an influenza viral strain:
(i) Within 30 days from the date that FDA notifies manufacturers that characterized viral samples are available for test evaluation, the manufacturer must have testing performed on the device with those influenza viral samples in accordance with a standardized protocol considered and determined by FDA to be acceptable and appropriate.
(ii) Within 60 days from the date that FDA notifies manufacturers that characterized influenza viral samples are available for test evaluation and continuing until 3 years from that date, the results of the influenza emergency analytical reactivity testing, including the detailed information for the virus tested as described in the certificate of authentication, must be included as part of the device's labeling in a tabular format, either by:
(A) Placing the results directly in the device's labeling required under § 809.10(b) of this chapter that accompanies the device in a separate section of the labeling where analytical reactivity testing data can be found, but separate from the annual analytical reactivity testing results; or
(B) In a section of the device's label or in other labeling that accompanies the device, prominently providing a hyperlink to the manufacturer's public website where the analytical reactivity testing data can be found. The manufacturer's website, as well as the primary part of the manufacturer's website that discusses the device, must provide a prominently placed hyperlink to the website containing this information and must allow unrestricted viewing access.