K Number
K250995
Manufacturer
Date Cleared
2025-05-01

(30 days)

Product Code
Regulation Number
866.3981
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Xpert Xpress CoV-2/Flu/RSV plus test, performed on the GeneXpert Xpress System, is an automated multiplexed real-time reverse transcriptase polymerase chain reaction (RT-PCR) test intended for use in the simultaneous in vitro qualitative detection and differentiation of severe acute respiratory syndrome coronavirus (SARS-CoV-2), influenza A, influenza B, and/or respiratory syncytial virus (RSV) viral RNA in nasopharyngeal swab and anterior nasal swab specimens collected from individuals with signs and symptoms of respiratory tract infection. Clinical signs and symptoms of respiratory tract infection due to SARS-CoV-2, influenza A, influenza B, and RSV can be similar.

The Xpert Xpress CoV-2/Flu/RSV plus test is intended for use in the differential detection of SARS-CoV-2, influenza A, influenza B and/or RSV RNA and aids in the diagnosis of COVID-19, influenza and/or RSV infections if used in conjunction with other clinical and epidemiological information, and laboratory findings. SARS-CoV-2, influenza A, influenza B, and RSV viral RNA are generally detectable in nasopharyngeal swab and anterior nasal swab specimens during the acute phase of infection.

Positive results are indicative of the presence of the identified virus, but do not rule out bacterial infection or co-infection with other pathogens not detected by the test. The agent (s) detected by the Xpert Xpress CoV-2/Flu/RSV plus test may not be the definite cause of the disease.

Negative results do not preclude SARS-CoV-2, influenza A, influenza B and/or RSV infection. The results of this test should not be used as the sole basis for diagnosis, treatment or other patient management decisions.

Device Description

The Xpert Xpress CoV-2/Flu/RSV plus test is an automated in vitro diagnostic test for the simultaneous qualitative detection and differentiation of SARS-CoV-2, Flu A, Flu B, and RSV viral RNA in nasopharyngeal swab (NPS) and anterior nasal swab (NS) specimens collected from individuals showing signs and symptoms of respiratory viral infection.

The Xpert Xpress CoV-2/Flu/RSV plus test is performed on GeneXpert Xpress System, which consist of a GeneXpert IV instrument that executes sample preparation, nucleic acid amplification and real-time fluorescent signal detection for the tests, and a GeneXpert Hub with preloaded GeneXpert Xpress software for running the tests and viewing the test results. The GeneXpert Hub accessory integrates the computer, touchscreen monitor and barcode scanner. Each of the GeneXpert modules in the GeneXpert IV instrument can perform independent sample preparation and testing. The GeneXpert Xpress System requires the use of single-use disposable cartridges that hold the RT-PCR reagents and host sample purification, nucleic acid amplification, and detection of the target sequences. Because the cartridges are self-contained, cross-contamination between samples is minimized.

The Xpert Xpress CoV-2/Flu/RSV plus test includes reagents for the detection of SARS-CoV-2, Flu A, Flu B and RSV viral RNA from NPS and NS specimens. The primers and probes in the Xpert Xpress CoV-2/Flu/RSV plus test are designed to amplify and detect unique sequences in the genes that encode the following proteins: SARS-CoV-2 nucleocapsid (N), SARS-CoV-2 envelope (E), SARS-CoV-2 RNA-dependent RNA polymerase (RdRP), influenza A matrix (M), influenza A basic polymerase (PB2), influenza A acidic protein (PA), influenza B matrix (M), influenza B non-structural protein (NS), and the RSV A and RSV B nucleocapsid.

A Sample Processing Control (SPC) and a Probe Check Control (PCC) are also included in the cartridge utilized by the GeneXpert instrument. The SPC is present to control for adequate processing of the sample and to monitor for the presence of potential inhibitor(s) in the RT-PCR reaction. The SPC also ensures that the RT-PCR reaction conditions (temperature and time) are appropriate for the amplification reaction and that the RT-PCR reagents are functional. The PCC verifies reagent rehydration, PCR tube filling, and confirms that all reaction components are present in the cartridge including monitoring for probe integrity and dye stability.

The Xpert Xpress CoV-2/Flu/RSV plus test is designed for use with NPS or NS specimens collected with nylon flocked swabs and placed into viral transport medium (VTM), Universal Transport Medium (UTM), or eNAT®. The ancillary specimen collection kits, swabs and transport media validated for use with the Xpert Xpress CoV-2/Flu/RSV plus test included:

Nasopharyngeal Sample Collection Kit for Viruses

  • Copan UTM® 3C057N (Flexible Minitip Flocked Swab with UTM® Medium without Beads)
  • Copan eNAT® Molecular Collection and Preservation Medium P/N 6U074S01 (Flexible Minitap Flocked Swab with eNAT® Medium)

Nasal Sample Collection Kit for Viruses

  • Copan UTM® 3C064N (Regular Flocked Swab with UTM® Medium without Beads)
  • Copan eNAT® Molecular Collection and Preservation Medium P/N 6U073S01 (Regular Flocked Swab with eNAT® Medium)

Alternatively, swabs and transport media can be obtained separately:

  • Nylon flocked swab (Copan P/N 502CS01, 503CS01)
  • Viral Transport Medium, 3 mL (Copan P/N 330C, 3C047N, BD Universal Transport Medium, Remel M4RT, or Remel M5)

The ancillary reagents allow NPS and NS specimens from patients to be collected, preserved and transported to laboratory prior to analysis with the Xpert Xpress CoV 2/Flu/RSV plus test.

AI/ML Overview

Based on the provided FDA 510(k) clearance letter and summary, here's an analysis of the acceptance criteria and study that proves the device meets them:

Important Note: This document describes a "Special 510(k) submission." This type of submission is used when changes are made to a previously cleared device that do not affect its fundamental technology, intended use, or safety/effectiveness. In this specific case, the changes were to the Assay Definition File (ADF) – essentially software parameter settings. Therefore, the "study that proves the device meets the acceptance criteria" largely relies on re-analysis of existing studies from the original device clearance (K242071) rather than entirely new, large-scale clinical trials.

1. Table of Acceptance Criteria and Reported Device Performance

The provided document doesn't explicitly state quantitative acceptance criteria in a dedicated table. However, since this is a Special 510(k) for software parameter changes, the primary "acceptance criteria" appear to be demonstrating non-inferiority or equivalency to the previously cleared predicate device in terms of:

  • Valid Test Runs: The number of tests that yield a valid result.
  • Non-Determinate (ND) Test Results: The number of tests that do not yield a definitive positive or negative result (e.g., "NO RESULT").
  • Performance Claims: That the modifications did not negatively impact the overall analytical and clinical performance claims established for the predicate device (sensitivity, specificity for detecting SARS-CoV-2, Flu A, Flu B, and RSV).

Inferred Acceptance Criteria and Reported Performance:

Acceptance Criterion (Inferred)Reported Device Performance (with Updated ADF)
Maintain number of valid test runs (vs. predicate)"The comparison of 16264 test results demonstrated that the numbers of valid test runs... were the same between the original ADF and the updated ADF with Flu B SLD Off."
Maintain number of non-determinate (ND) test results (vs. predicate)"The comparison of 16264 test results demonstrated that... non-determinate (ND) test results were the same between the original ADF and the updated ADF with Flu B SLD Off." "All non-determinate GeneXpert ERROR test results with error code 5011 (SLD) were changed to NO RESULT." (This indicates a change in reporting for specific errors to a more appropriate "NO RESULT" category, but the number of non-determinate results remained the same.)
Correctly implement revised algorithm for SARS-CoV-2 only mode"The comparison of 25 test results demonstrated that the number of valid test runs and ND test results were the same between the original and the updated ADF." (for the SARS-CoV-2 only test mode) "The revised algorithm produced the expected NO RESULT – REPEAT TEST GeneXpert test results for the Xpress SARS-CoV-2 plus test mode." (for simulated conditions)
No negative impact on established performance claims"The assessment of the re-analysis results determined that the performance claims of the Xpert Xpress CoV-2/Flu/RSV plus test were not impacted by the modifications made to the predicate device."

2. Sample Size Used for the Test Set and Data Provenance

  • Sample Size:
    • Main analysis: 16,264 test results. This number represents the re-analysis of "verification, validation and flex studies data from the original studies."
    • SARS-CoV-2 only test mode analysis: 25 test results.
    • Simulated conditions for revised algorithm: The document mentions "Test cartridges were simulated to generate SARS-CoV-2 INVALID and SPC FAIL conditions," but does not explicitly state the number of simulated tests.
  • Data Provenance: The data used for re-analysis originated from the "original studies" (K242071 submission). The document does not specify the country of origin or whether these original studies were retrospective or prospective, though typical clinical validation studies for IVDs are often prospective. Given it references "flex study," those can sometimes include both retrospective and prospective elements.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

The document does not provide information on the number or qualifications of experts used to establish the ground truth. This is likely because the ground truth was established during the original predicate device clearance (K242071) and this Special 510(k) focused on demonstrating equivalence in specific software-related metrics through re-analysis. For IVD devices like this, ground truth is typically established by comparing against a highly sensitive and specific reference method (e.g., an FDA-cleared laboratory developed test or a combination of clinical diagnosis and other accepted testing methods).

4. Adjudication Method for the Test Set

The document does not specify any adjudication method (e.g., 2+1, 3+1, none) for the test set. Again, this specific submission involved re-analysis of existing data rather than new clinical trials where such adjudication might be more explicitly detailed if human reads were involved. For an automated RT-PCR test, adjudication as typically understood for image-based AI would not be directly applicable.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not conducted or is not mentioned in this document. This is an in vitro diagnostic device (RT-PCR test), not an AI-assisted diagnostic imaging system that would typically involve human readers interpreting images with or without AI assistance. The device is fully automated.

6. Standalone Performance (Algorithm Only)

Yes, the performance evaluated here is inherently standalone (algorithm only). The Xpert Xpress CoV-2/Flu/RSV plus is an automated RT-PCR test system. Its output is directly generated by the instrument and its embedded software (ADF). The "performance data" detailed in section 18.5 describes how the device's algorithmic and software changes affected its ability to produce valid results and interpret them correctly, without human intervention in the result generation.

7. Type of Ground Truth Used

The document does not explicitly state the type of ground truth used for the original studies, beyond stating "analytical, clinical and flex studies data." For RT-PCR assays, the ground truth for clinical performance is typically established by:

  • Clinical Reference Method: Comparison against a highly sensitive and specific laboratory reference method (e.g., another FDA-cleared or EUA-authorized RT-PCR test, or a composite reference standard using multiple methods).
  • Clinical Diagnosis/Outcomes Data (less common for purely diagnostic tests): While the device aids in diagnosis ("aids in the diagnosis of COVID-19, influenza and/or RSV infections if used in conjunction with other clinical and epidemiological information, and laboratory findings"), the performance claims themselves are based on detection of viral RNA, rather than patient outcomes as primary ground truth.

Given the nature of a molecular diagnostic test for viral RNA, the ground truth for the performance claims (sensitivity, specificity) of the original device would have been established by comparing the device's results against a highly reliable reference molecular test, often considered the "gold standard" for pathogen detection.

8. Sample Size for the Training Set

The document does not provide information on the sample size for the training set. This is because:

  • This is an RT-PCR test, not a deep learning AI model that undergoes "training" in the typical machine learning sense. The "algorithm" here refers to the pre-defined logical rules and parameters within the Assay Definition File (ADF) for interpreting RT-PCR signals.
  • The re-analysis was performed on verification, validation, and flex studies data, which are typically considered test or validation sets in the context of device development, not training sets.

9. How the Ground Truth for the Training Set Was Established

As noted above, a "training set" and associated ground truth establishment in the context of deep learning AI are not applicable to this RT-PCR device. The "ground truth" for the device's performance relies on the known characteristics of the viral targets and the performance of the RT-PCR chemistry against reference methods.

FDA 510(k) Clearance Letter - Xpert Xpress CoV-2/Flu/RSV plus

Page 1

U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov

Doc ID # 04017.07.05

May 1, 2025

Cepheid
Bansari Doshi
Regulatory Affairs Specialist - New Product Development
904 Caribbean Dive
Sunnyvale, California 94089

Re: K250995
Trade/Device Name: Xpert Xpress CoV-2/Flu/RSV plus
Regulation Number: 21 CFR 866.3981
Regulation Name: Device To Detect And Identify Nucleic Acid Targets In Respiratory Specimens From Microbial Agents That Cause The SARS-CoV-2 Respiratory Infection And Other Microbial Agents When In A Multi-Target Test
Regulatory Class: Class II
Product Code: QOF
Dated: March 31, 2025
Received: April 1, 2025

Dear Bansari Doshi:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Page 2

K250995 - Bansari Doshi Page 2

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-

Page 3

K250995 - Bansari Doshi Page 3

assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Joseph Briggs -S

Joseph Briggs, Ph.D.
Deputy Division Director
Division of Microbiology Devices
OHT7: Office of In Vitro Diagnostics
Office of Product Evaluation and Quality
Center for Devices and Radiological Health

Enclosure

Page 4

FORM FDA 3881 (8/23) Page 1 of 1 PSC Publishing Services (301) 443-6740 EF

DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration

Indications for Use

Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.

510(k) Number (if known): K250995
Device Name: Xpert Xpress CoV-2/Flu/RSV plus

Indications for Use (Describe)

The Xpert Xpress CoV-2/Flu/RSV plus test, performed on the GeneXpert Xpress System, is an automated multiplexed real-time reverse transcriptase polymerase chain reaction (RT-PCR) test intended for use in the simultaneous in vitro qualitative detection and differentiation of severe acute respiratory syndrome coronavirus (SARS-CoV-2), influenza A, influenza B, and/or respiratory syncytial virus (RSV) viral RNA in nasopharyngeal swab and anterior nasal swab specimens collected from individuals with signs and symptoms of respiratory tract infection. Clinical signs and symptoms of respiratory tract infection due to SARS-CoV-2, influenza A, influenza B, and RSV can be similar.

The Xpert Xpress CoV-2/Flu/RSV plus test is intended for use in the differential detection of SARS-CoV-2, influenza A, influenza B and/or RSV RNA and aids in the diagnosis of COVID-19, influenza and/or RSV infections if used in conjunction with other clinical and epidemiological information, and laboratory findings. SARS-CoV-2, influenza A, influenza B, and RSV viral RNA are generally detectable in nasopharyngeal swab and anterior nasal swab specimens during the acute phase of infection.

Positive results are indicative of the presence of the identified virus, but do not rule out bacterial infection or co-infection with other pathogens not detected by the test. The agent (s) detected by the Xpert Xpress CoV-2/Flu/RSV plus test may not be the definite cause of the disease.

Negative results do not preclude SARS-CoV-2, influenza A, influenza B and/or RSV infection. The results of this test should not be used as the sole basis for diagnosis, treatment or other patient management decisions.

Type of Use (Select one or both, as applicable)

☒ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

Page 5

Xpert® Xpress CoV-2/Flu/RSV plus Special 510(k) Submission

510(k) Summary for Xpert Xpress CoV-2/Flu/RSV plus

Page 6

Xpert® Xpress CoV-2/Flu/RSV plus Special 510(k) Submission

TABLE OF CONTENTS

18 510(K) SUMMARY ............................................................................................................... 3

18.1 DEVICE DESCRIPTION................................................................................................... 4
18.2 DEVICE INTENDED USE................................................................................................ 5
18.3 TECHNICAL CHARACTERISTICS ................................................................................ 6
18.4 SUBSTANTIAL EQUIVALENCE.................................................................................... 6
18.5 SUMMARY OF PERFORMANCE DATA....................................................................... 8
18.6 CONCLUSION................................................................................................................... 9

Page 7

Xpert® Xpress CoV-2/Flu/RSV plus Special 510(k) Submission

18 510(k) Summary

As required by 21 CFR Section 807.92(c).

Submitted by: Cepheid
904 Caribbean Drive
Sunnyvale, CA 90489
Phone number: (408) 541-4191

Contact: Bansari Doshi, M.S.

Date of Preparation: April 28, 2025

Device:

Trade nameXpert® Xpress CoV-2/Flu/RSV plus
Common nameXpert Xpress CoV-2/Flu/RSV plus
Type of TestQualitative real-time reverse transcription polymerase chain reaction (RT-PCR) and detection test
Regulation number Classification name Product code21 CFR 866.3981, Multi-Target Respiratory Specimen Nucleic Acid Test Including SARS-CoV-2 and Other Microbial Agents. QOF
Secondary Product CodeOOI
Classification Advisory PanelMicrobiology (83)
Prescription UseYes
Predicate Device Assay:Xpert Xpress CoV-2/Flu/RSV plus (K242071)

Page 8

Xpert® Xpress CoV-2/Flu/RSV plus Special 510(k) Submission

18.1 Device Description

The Xpert Xpress CoV-2/Flu/RSV plus test is an automated in vitro diagnostic test for the simultaneous qualitative detection and differentiation of SARS-CoV-2, Flu A, Flu B, and RSV viral RNA in nasopharyngeal swab (NPS) and anterior nasal swab (NS) specimens collected from individuals showing signs and symptoms of respiratory viral infection.

The Xpert Xpress CoV-2/Flu/RSV plus test is performed on GeneXpert Xpress System, which consist of a GeneXpert IV instrument that executes sample preparation, nucleic acid amplification and real-time fluorescent signal detection for the tests, and a GeneXpert Hub with preloaded GeneXpert Xpress software for running the tests and viewing the test results. The GeneXpert Hub accessory integrates the computer, touchscreen monitor and barcode scanner. Each of the GeneXpert modules in the GeneXpert IV instrument can perform independent sample preparation and testing. The GeneXpert Xpress System requires the use of single-use disposable cartridges that hold the RT-PCR reagents and host sample purification, nucleic acid amplification, and detection of the target sequences. Because the cartridges are self-contained, cross-contamination between samples is minimized.

The Xpert Xpress CoV-2/Flu/RSV plus test includes reagents for the detection of SARS-CoV-2, Flu A, Flu B and RSV viral RNA from NPS and NS specimens. The primers and probes in the Xpert Xpress CoV-2/Flu/RSV plus test are designed to amplify and detect unique sequences in the genes that encode the following proteins: SARS-CoV-2 nucleocapsid (N), SARS-CoV-2 envelope (E), SARS-CoV-2 RNA-dependent RNA polymerase (RdRP), influenza A matrix (M), influenza A basic polymerase (PB2), influenza A acidic protein (PA), influenza B matrix (M), influenza B non-structural protein (NS), and the RSV A and RSV B nucleocapsid.

A Sample Processing Control (SPC) and a Probe Check Control (PCC) are also included in the cartridge utilized by the GeneXpert instrument. The SPC is present to control for adequate processing of the sample and to monitor for the presence of potential inhibitor(s) in the RT-PCR reaction. The SPC also ensures that the RT-PCR reaction conditions (temperature and time) are appropriate for the amplification reaction and that the RT-PCR reagents are functional. The PCC verifies reagent rehydration, PCR tube filling, and confirms that all reaction components are present in the cartridge including monitoring for probe integrity and dye stability.

The Xpert Xpress CoV-2/Flu/RSV plus test is designed for use with NPS or NS specimens collected with nylon flocked swabs and placed into viral transport medium (VTM), Universal Transport Medium (UTM), or eNAT®. The ancillary specimen collection kits, swabs and transport media validated for use with the Xpert Xpress CoV-2/Flu/RSV plus test included:

Nasopharyngeal Sample Collection Kit for Viruses

  • Copan UTM® 3C057N (Flexible Minitip Flocked Swab with UTM® Medium without Beads)
  • Copan eNAT® Molecular Collection and Preservation Medium P/N 6U074S01 (Flexible Minitap Flocked Swab with eNAT® Medium)

Page 9

Xpert® Xpress CoV-2/Flu/RSV plus Special 510(k) Submission

  • Becton Dickinson Universal Viral Transport Kit P/N 220531 (Flexible Minitip Flocked Swab with UVT Medium)

Nasal Sample Collection Kit for Viruses

  • Copan UTM® 3C064N (Regular Flocked Swab with UTM® Medium without Beads)
  • Copan eNAT® Molecular Collection and Preservation Medium P/N 6U073S01 (Regular Flocked Swab with eNAT® Medium)

Alternatively, swabs and transport media can be obtained separately:

  • Nylon flocked swab (Copan P/N 502CS01, 503CS01)
  • Viral Transport Medium, 3 mL (Copan P/N 330C, 3C047N, BD Universal Transport Medium, Remel M4RT, or Remel M5)

The ancillary reagents allow NPS and NS specimens from patients to be collected, preserved and transported to laboratory prior to analysis with the Xpert Xpress CoV 2/Flu/RSV plus test.

18.2 Device Intended Use

The Xpert® Xpress CoV-2/Flu/RSV plus test, performed on the GeneXpert® Xpress System, is an automated multiplexed real-time reverse transcriptase polymerase chain reaction (RT-PCR) test intended for use in the simultaneous in vitro qualitative detection and differentiation of severe acute respiratory syndrome coronavirus (SARS-CoV-2), influenza A, influenza B, and/or respiratory syncytial virus (RSV) viral RNA in nasopharyngeal swab and anterior nasal swab specimens collected from individuals with signs and symptoms of respiratory tract infection. Clinical signs and symptoms of respiratory tract infection due to SARS-CoV-2, influenza A, influenza B, and RSV can be similar.

The Xpert Xpress CoV-2/Flu/RSV plus test is intended for use in the differential detection of SARS-CoV-2, influenza A, influenza B and/or RSV RNA and aids in the diagnosis of COVID-19, influenza and/or RSV infections if used in conjunction with other clinical and epidemiological information, and laboratory findings. SARS-CoV-2, influenza A, influenza B, and RSV viral RNA are generally detectable in nasopharyngeal swab and anterior nasal swab specimens during the acute phase of infection.

Positive results are indicative of the presence of the identified virus, but do not rule out bacterial infection or co-infection with other pathogens not detected by the test. The agent (s) detected by the Xpert Xpress CoV-2/Flu/RSV plus test may not be the definite cause of the disease.

Negative results do not preclude SARS-CoV-2, influenza A, influenza B and/or RSV infection. The results of this test should not be used as the sole basis for diagnosis, treatment or other patient management decisions.

Page 10

Xpert® Xpress CoV-2/Flu/RSV plus Special 510(k) Submission

18.3 Technical Characteristics

The Xpert Xpress CoV-2/Flu/RSV plus test has the same technological characteristics as the predicate device.

18.4 Substantial Equivalence

The purpose of this Special 510(k) submission is to support design changes to the Xpert Xpress CoV-2/Flu/RSV plus test which require modifications of the parameter settings in the Assay Definition File (ADF). These modifications include turning off the Signal Loss Detection (SLD) setting for the Flu B channel and revising the result reporting algorithm for the SARS-CoV-2 only test mode.

Table 18-1 shows the similarities and Table 18-2 shows the differences between the Xpert Xpress CoV-2/Flu/RSV plus test and predicate device. The differences between Xpert Xpress CoV-2/Flu/RSV plus test and predicate device do not raise different questions of safety and effectiveness.

Table 18-1. Similarities Between Xpert Xpress CoV-2/Flu/RSV plus and the Predicate Device

AttributeSubject DevicePredicate Device
Xpert® Xpress CoV-2/Flu/RSV plus (Modified Design)Xpert® Xpress CoV-2/Flu/RSV plus (K242071)
RegulationSame21 CFR 866.3981 Devices to detect and identify nucleic acid targets in respiratory samples from microbial agents that cause the SARS-CoV-2 respiratory infection and other microbial agents when in a multi-analyte test
Product CodeSameQOF Multi-target respiratory specimen nucleic acid test including SARS-CoV-2 and other microbial agents
Device ClassSameII (Special Controls)
Technology/ DetectionSameReal-time reverse transcription polymerase chain reaction (RT-qPCR)
Intended UseXpert® Xpress CoV-2/Flu/RSV plus test, performed on the GeneXpert® Xpress Systems, is an automated multiplexed real-time reverse transcriptase polymerase chain reaction (RT-PCR) test intended for use in the simultaneous in vitro qualitative detection and differentiation of severe acute respiratory syndrome coronavirus (SARS-CoV-2), influenza A, influenza B, and/or respiratory syncytial virus (RSV)Xpert® Xpress CoV-2/Flu/RSV plus test, performed on the GeneXpert® Xpress Systems, is an automated multiplexed real-time reverse transcriptase polymerase chain reaction (RT-PCR) test intended for use in the simultaneous in vitro qualitative detection and differentiation of severe acute respiratory syndrome coronavirus (SARS-CoV-2), influenza A, influenza B, and/or respiratory syncytial virus (RSV)

Page 11

Xpert® Xpress CoV-2/Flu/RSV plus Special 510(k) Submission

AttributeSubject DevicePredicate Device
Xpert® Xpress CoV-2/Flu/RSV plus (Modified Design)Xpert® Xpress CoV-2/Flu/RSV plus (K242071)
viral RNA in nasopharyngeal swab and anterior nasal swab specimens collected from individuals with signs and symptoms of respiratory tract infection. Clinical signs and symptoms of respiratory tract infection due to SARS-CoV-2, influenza A, influenza B, and RSV can be similar. The Xpert Xpress CoV-2/Flu/RSV plus test is intended for use in the differential detection of SARS-CoV-2, influenza A, influenza B and/or RSV RNA and aids in the diagnosis of COVID-19, influenza and/or RSV infections if used in conjunction with other clinical and epidemiological information, and laboratory findings. SARS-CoV-2, influenza A, influenza B, and RSV viral RNA are generally detectable in nasopharyngeal swab and anterior nasal swab specimens during the acute phase of infection. Positive results are indicative of the presence of the identified virus, but do not rule out bacterial infection or co-infection with other pathogens not detected by the test. The agent(s) detected by the Xpert Xpress CoV-2/Flu/RSV plus test may not be the definite cause of disease. Negative results do not preclude SARS-CoV-2, influenza A, influenza B and/or RSV infection. The results of this test should not be used as the sole basis for diagnosis, treatment, or other patient management decisions.viral RNA in nasopharyngeal swab and anterior nasal swab specimens collected from individuals with signs and symptoms of respiratory tract infection. Clinical signs and symptoms of respiratory tract infection due to SARS-CoV-2, influenza A, influenza B, and RSV can be similar. The Xpert Xpress CoV-2/Flu/RSV plus is intended for use in the differential detection of SARS-CoV-2, influenza A, influenza B and/or RSV RNA and aids in the diagnosis of COVID-19, influenza and/or RSV infections if used in conjunction with other clinical and epidemiological information, and laboratory findings. SARS-CoV-2, influenza A, influenza B, and RSV viral RNA are generally detectable in nasopharyngeal swab and anterior nasal swab specimens during the acute phase of infection. Positive results are indicative of the presence of the identified virus, but do not rule out bacterial infection or co-infection with other pathogens not detected by the test. The agent(s) detected by the Xpert Xpress CoV-2/Flu/RSV plus test may not be the definite cause of disease. Negative results do not preclude SARS-CoV-2, influenza A, influenza B and/or RSV infection. The results of this test should not be used as the sole basis for treatment, or other patient management decisions.
Assay TargetsSameSARS-CoV-2, Influenza A, Influenza B, RSV viral RNA
Specimen TypeSame• Nasopharyngeal swab (NPS) • Anterior nasal swab (NS)
Transport MediaSame• Universal Transport Medium (UTM) / Viral Transport Medium (VTM) • eNAT
Test FormatSameSingle Use
AutomationSameAutomated Nucleic Acid Extraction, Detection and Results Interpretation

Page 12

Xpert® Xpress CoV-2/Flu/RSV plus Special 510(k) Submission

AttributeSubject DevicePredicate Device
Xpert® Xpress CoV-2/Flu/RSV plus (Modified Design)Xpert® Xpress CoV-2/Flu/RSV plus (K242071)
Combinatorial Assay Definition Files (ADFs)Same• Xpress SARS-CoV-2 Flu RSV plus • Xpress SARS-CoV-2 Flu plus • Xpress SARS-CoV-2 plus • Xpress Flu plus • Xpress Flu RSV plus Additionally, a video to guide user through ordering and running the test
Assay ResultsSameQualitative
Non-Determinate Results TextSameNO RESULTS – REPEAT TEST INSTRUMENT ERROR
Internal ControlSameSample Processing Control (SPC) Probe Check Control (PCC)
Instrument SystemsSameCepheid GeneXpert® Xpress System
GeneXpert SoftwareSameGeneXpert Xpress software
Time to ResultSame< 36 min for sample preparation and RT-PCR

Table 18-2. Differences Between Xpert Xpress CoV-2/Flu/RSV plus and the Predicate Device

AttributeSubject DevicePredicate Device
Xpert® Xpress CoV-2/Flu/RSV plus (Modified Design)Xpert® Xpress CoV-2/Flu/RSV plus (K242071)
Assay Definition File• Signal Loss Detection (SLD) Turned off in the Flu B channel• Signal Loss Detection (SLD) Turned on in the Flu B channel
• The rules table for Xpress SARS-CoV-2 only test mode reports a NO RESULT – REPEAT TEST GeneXpert test result if the analyte result for SARS-CoV-2 is INVALID and for SPC is FAIL (i.e., the SPC does not have a valid Ct)• The rules table for Xpress SARS-CoV-2 only test mode reports a SARS-CoV-2 NEGATIVE GeneXpert test result if the analyte result for SARS-CoV-2 is INVALID and for SPC is FAIL (i.e., the SPC does not have a valid Ct)

18.5 Summary of Performance Data

The performance of the Xpert Xpress CoV-2/Flu/RSV plus with the updated ADF was assessed by re-analyzing verification, validation and flex studies data from the original studies. The analytical, clinical and flex studies data generated with the original design (K242071), were reanalyzed and compared to data generated using the updated ADF with the SLD setting for the Flu B channel turned off. The comparison of 16264 test results demonstrated that the numbers of valid test runs and non-determinate (ND) test results were the same between the original ADF

Page 13

Xpert® Xpress CoV-2/Flu/RSV plus Special 510(k) Submission

and the updated ADF with Flu B SLD Off. All non-determinate GeneXpert ERROR test results with error code 5011 (SLD) were changed to NO RESULT. For the specific tests in the flex study where the Xpress SARS-CoV-2 plus only test mode of the ADF was used, the original data were reanalyzed and compared to the data generated using the updated Xpress SARS-CoV-2 plus only test mode. The comparison of 25 test results demonstrated that the number of valid test runs and ND test results were the same between the original and the updated ADF.

Additionally, a study was conducted to verify the revised algorithm for the Xpress SARS-CoV-2 plus only test mode. Test cartridges were simulated to generate SARS-CoV-2 INVALID and SPC FAIL conditions. The revised algorithm produced the expected NO RESULT – REPEAT TEST GeneXpert test results for the Xpress SARS-CoV-2 plus test mode.

The assessment of the re-analysis results determined that the performance claims of the Xpert Xpress CoV-2/Flu/RSV plus test were not impacted by the modifications made to the predicate device.

18.6 Conclusion

The results of the re-analysis of verification and validation studies demonstrated that the Xpert Xpress CoV-2/Flu/RSV plus test when performed with an updated ADF is substantially equivalent to the original design of the Xpert Xpress CoV-2/Flu/RSV plus (K242071, the predicate device).

§ 866.3981 Device to detect and identify nucleic acid targets in respiratory specimens from microbial agents that cause the SARS-CoV-2 respiratory infection and other microbial agents when in a multi-target test.

(a)
Identification. A device to detect and identify nucleic acid targets in respiratory specimens from microbial agents that cause the SARS-CoV-2 respiratory infection and other microbial agents when in a multi-target test is an in vitro diagnostic device intended for the detection and identification of SARS-CoV-2 and other microbial agents when in a multi-target test in human clinical respiratory specimens from patients suspected of respiratory infection who are at risk for exposure or who may have been exposed to these agents. The device is intended to aid in the diagnosis of respiratory infection in conjunction with other clinical, epidemiologic, and laboratory data or other risk factors.(b)
Classification. Class II (special controls). The special controls for this device are:(1) The intended use in the labeling required under § 809.10 of this chapter must include a description of the following: Analytes and targets the device detects and identifies, the specimen types tested, the results provided to the user, the clinical indications for which the test is to be used, the specific intended population(s), the intended use locations including testing location(s) where the device is to be used (if applicable), and other conditions of use as appropriate.
(2) Any sample collection device used must be FDA-cleared, -approved, or -classified as 510(k) exempt (standalone or as part of a test system) for the collection of specimen types claimed by this device; alternatively, the sample collection device must be cleared in a premarket submission as a part of this device.
(3) The labeling required under § 809.10(b) of this chapter must include:
(i) A detailed device description, including reagents, instruments, ancillary materials, all control elements, and a detailed explanation of the methodology, including all pre-analytical methods for processing of specimens;
(ii) Detailed descriptions of the performance characteristics of the device for each specimen type claimed in the intended use based on analytical studies including the following, as applicable: Limit of Detection, inclusivity, cross-reactivity, interfering substances, competitive inhibition, carryover/cross contamination, specimen stability, precision, reproducibility, and clinical studies;
(iii) Detailed descriptions of the test procedure(s), the interpretation of test results for clinical specimens, and acceptance criteria for any quality control testing;
(iv) A warning statement that viral culture should not be attempted in cases of positive results for SARS-CoV-2 and/or any similar microbial agents unless a facility with an appropriate level of laboratory biosafety (
e.g., BSL 3 and BSL 3+, etc.) is available to receive and culture specimens; and(v) A prominent statement that device performance has not been established for specimens collected from individuals not identified in the intended use population (
e.g., when applicable, that device performance has not been established in individuals without signs or symptoms of respiratory infection).(vi) Limiting statements that indicate that:
(A) A negative test result does not preclude the possibility of infection;
(B) The test results should be interpreted in conjunction with other clinical and laboratory data available to the clinician;
(C) There is a risk of incorrect results due to the presence of nucleic acid sequence variants in the targeted pathogens;
(D) That positive and negative predictive values are highly dependent on prevalence;
(E) Accurate results are dependent on adequate specimen collection, transport, storage, and processing. Failure to observe proper procedures in any one of these steps can lead to incorrect results; and
(F) When applicable (
e.g., recommended by the Centers for Disease Control and Prevention, by current well-accepted clinical guidelines, or by published peer-reviewed literature), that the clinical performance may be affected by testing a specific clinical subpopulation or for a specific claimed specimen type.(4) Design verification and validation must include:
(i) Detailed documentation, including performance results, from a clinical study that includes prospective (sequential) samples for each claimed specimen type and, as appropriate, additional characterized clinical samples. The clinical study must be performed on a study population consistent with the intended use population and compare the device performance to results obtained using a comparator that FDA has determined is appropriate. Detailed documentation must include the clinical study protocol (including a predefined statistical analysis plan), study report, testing results, and results of all statistical analyses.
(ii) Risk analysis and documentation demonstrating how risk control measures are implemented to address device system hazards, such as Failure Modes Effects Analysis and/or Hazard Analysis. This documentation must include a detailed description of a protocol (including all procedures and methods) for the continuous monitoring, identification, and handling of genetic mutations and/or novel respiratory pathogen isolates or strains (
e.g., regular review of published literature and periodic in silico analysis of target sequences to detect possible mismatches). All results of this protocol, including any findings, must be documented and must include any additional data analysis that is requested by FDA in response to any performance concerns identified under this section or identified by FDA during routine evaluation. Additionally, if requested by FDA, these evaluations must be submitted to FDA for FDA review within 48 hours of the request. Results that are reasonably interpreted to support the conclusion that novel respiratory pathogen strains or isolates impact the stated expected performance of the device must be sent to FDA immediately.(iii) A detailed description of the identity, phylogenetic relationship, and other recognized characterization of the respiratory pathogen(s) that the device is designed to detect. In addition, detailed documentation describing how to interpret the device results and other measures that might be needed for a laboratory diagnosis of respiratory infection.
(iv) A detailed device description, including device components, ancillary reagents required but not provided, and a detailed explanation of the methodology, including molecular target(s) for each analyte, design of target detection reagents, rationale for target selection, limiting factors of the device (
e.g., saturation level of hybridization and maximum amplification and detection cycle number, etc.), internal and external controls, and computational path from collected raw data to reported result (e.g., how collected raw signals are converted into a reported signal and result), as applicable.(v) A detailed description of device software, including software applications and hardware-based devices that incorporate software. The detailed description must include documentation of verification, validation, and hazard analysis and risk assessment activities, including an assessment of the impact of threats and vulnerabilities on device functionality and end users/patients as part of cybersecurity review.
(vi) For devices intended for the detection and identification of microbial agents for which an FDA recommended reference panel is available, design verification and validation must include the performance results of an analytical study testing the FDA recommended reference panel of characterized samples. Detailed documentation must be kept of that study and its results, including the study protocol, study report for the proposed intended use, testing results, and results of all statistical analyses.
(vii) For devices with an intended use that includes detection of Influenza A and Influenza B viruses and/or detection and differentiation between the Influenza A virus subtypes in human clinical specimens, the design verification and validation must include a detailed description of the identity, phylogenetic relationship, or other recognized characterization of the Influenza A and B viruses that the device is designed to detect, a description of how the device results might be used in a diagnostic algorithm and other measures that might be needed for a laboratory identification of Influenza A or B virus and of specific Influenza A virus subtypes, and a description of the clinical and epidemiological parameters that are relevant to a patient case diagnosis of Influenza A or B and of specific Influenza A virus subtypes. An evaluation of the device compared to a currently appropriate and FDA accepted comparator method. Detailed documentation must be kept of that study and its results, including the study protocol, study report for the proposed intended use, testing results, and results of all statistical analyses.
(5) When applicable, performance results of the analytical study testing the FDA recommended reference panel described in paragraph (b)(4)(vi) of this section must be included in the device's labeling under § 809.10(b) of this chapter.
(6) For devices with an intended use that includes detection of Influenza A and Influenza B viruses and/or detection and differentiation between the Influenza A virus subtypes in human clinical specimens in addition to detection of SARS-CoV-2 and similar microbial agents, the required labeling under § 809.10(b) of this chapter must include the following:
(i) Where applicable, a limiting statement that performance characteristics for Influenza A were established when Influenza A/H3 and A/H1-2009 (or other pertinent Influenza A subtypes) were the predominant Influenza A viruses in circulation.
(ii) Where applicable, a warning statement that reads if infection with a novel Influenza A virus is suspected based on current clinical and epidemiological screening criteria recommended by public health authorities, specimens should be collected with appropriate infection control precautions for novel virulent influenza viruses and sent to State or local health departments for testing. Viral culture should not be attempted in these cases unless a BSL 3+ facility is available to receive and culture specimens.
(iii) Where the device results interpretation involves combining the outputs of several targets to get the final results, such as a device that both detects Influenza A and differentiates all known Influenza A subtypes that are currently circulating, the device's labeling must include a clear interpretation instruction for all valid and invalid output combinations, and recommendations for any required followup actions or retesting in the case of an unusual or unexpected device result.
(iv) A limiting statement that if a specimen yields a positive result for Influenza A, but produces negative test results for all specific influenza A subtypes intended to be differentiated (
i.e., H1-2009 and H3), this result requires notification of appropriate local, State, or Federal public health authorities to determine necessary measures for verification and to further determine whether the specimen represents a novel strain of Influenza A.(7) If one of the actions listed at section 564(b)(1)(A) through (D) of the Federal Food, Drug, and Cosmetic Act occurs with respect to an influenza viral strain, or if the Secretary of Health and Human Services determines, under section 319(a) of the Public Health Service Act, that a disease or disorder presents a public health emergency, or that a public health emergency otherwise exists, with respect to an influenza viral strain:
(i) Within 30 days from the date that FDA notifies manufacturers that characterized viral samples are available for test evaluation, the manufacturer must have testing performed on the device with those influenza viral samples in accordance with a standardized protocol considered and determined by FDA to be acceptable and appropriate.
(ii) Within 60 days from the date that FDA notifies manufacturers that characterized influenza viral samples are available for test evaluation and continuing until 3 years from that date, the results of the influenza emergency analytical reactivity testing, including the detailed information for the virus tested as described in the certificate of authentication, must be included as part of the device's labeling in a tabular format, either by:
(A) Placing the results directly in the device's labeling required under § 809.10(b) of this chapter that accompanies the device in a separate section of the labeling where analytical reactivity testing data can be found, but separate from the annual analytical reactivity testing results; or
(B) In a section of the device's label or in other labeling that accompanies the device, prominently providing a hyperlink to the manufacturer's public website where the analytical reactivity testing data can be found. The manufacturer's website, as well as the primary part of the manufacturer's website that discusses the device, must provide a prominently placed hyperlink to the website containing this information and must allow unrestricted viewing access.