(231 days)
The Excelsior XT-27 Microcatheter and Excelsior XT-27 Flex Microcatheter are intended to assist in the delivery of embolization particles, diagnostic agents (such as contrast media), and interventional devices (such as stents) that are indicated for use in the neurovasculature and with a catheter of 0.027 inches in inner diameter.
The Excelsior XT-27 Pre-Shaped Microcatheter and Excelsior XT-27 Flex Pre-Shaped Microcatheter are intended to assist in the delivery of embolization particles, diagnostic agents (such as contrast media), and interventional devices (such as stents) that are indicated for use in the neurovasculature and with a catheter of 0.027 inches in inner diameter.
The subject devices hereafter referred to as Excelsior® XT-27® Microcatheters are sterile, single lumen, 0.027 inch inner diameter (ID) microcatheters with one tip marker designed to aid the physician in accessing distal neurovasculature when used with a guide catheter and steerable guidewire. Graded shaft stiffness ranging from a highly flexible tip to a semi-rigid proximal section aids the physician in tracking over selectively placed guidewires. A luer fitting located on the microcatheter hub is used for the attachment of accessories. One radiopaque tip marker is positioned at the distal tip of the device to facilitate fluoroscopic visualization® XT-27® Microcatheters are coated on the outer surface with Hydrolene™ coating which reduces friction during manipulation in the vessel.
The Excelsior® XT-27® Microcatheters are available in effective lengths of both 135 cm (53.1 inch) and 150 cm (59.1 inch), with two distal shaft configurations achieved through distal shaft lengths of 6 cm (XT-27 model) and 18 cm (XT-27 Flex model). Both straight tip and pre-shaped versions are available.
The provided text is a 510(k) summary for a medical device (Excelsior XT-27 Microcatheters) and details its acceptance criteria and the study performed to demonstrate substantial equivalence, but it does not describe an AI/ML device. Therefore, the questions related to AI/ML specific criteria (such as sample sizes for test and training sets, number of experts for ground truth, adjudication methods, multi-reader multi-case studies, standalone performance, and data provenance) are not applicable to this document.
However, I can provide the acceptance criteria and study information that is available for this mechanical microcatheter from the document.
1. Table of acceptance criteria and the reported device performance:
Test | Acceptance Criteria (Inferred from "Pass") | Reported Device Performance |
---|---|---|
Guide Catheter Compatibility with 1.17 mm (0.046") ID | Maximum forces required to completely deliver and retrieve the microcatheter inside a 0.046" ID guide catheter with worst-case sized ancillary devices must be acceptable/pass. | Pass |
Tensile Strength for Joints and Marker Band | Tensile strength after preconditioning by simulated use with worst-case sized ancillary and interventional devices must be acceptable/pass. | Pass |
Particulate and Coating Integrity | Hydrophilic coating integrity and particulate generation under simulated use conditions with a 0.046" ID guide catheter, and comparison to cleared comparator devices, must be acceptable/pass. | Pass |
2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective):
- Sample Size for Test Set: The document does not specify the exact number of samples (e.g., number of microcatheters tested) used for each bench test. It mentions testing with "ancillary devices that represent worst-case sizes" for the Guide Catheter Compatibility test and "worst-case sized ancillary devices and interventional devices" for the Tensile Strength test.
- Data Provenance: The study is described as "Bench testing," indicating it was conducted in a laboratory setting. There is no information regarding country of origin or whether it was retrospective or prospective, as these terms are typically applied to clinical studies involving human or animal subjects.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g., radiologist with 10 years of experience):
- Not applicable as this is a mechanical device performance study, not an AI/ML diagnostic or prognostic study requiring expert opinion for ground truth. The "ground truth" here is the physical performance of the device against engineering specifications.
4. Adjudication method (e.g., 2+1, 3+1, none) for the test set:
- Not applicable for the same reason as point 3. Bench testing results are typically adjudicated by meeting predefined engineering specifications and criteria, not by expert consensus in the same way as diagnostic reads.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
- No, a multi-reader multi-case (MRMC) comparative effectiveness study was not performed. This type of study is relevant for AI-assisted diagnostic devices, which is not what this device is.
6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done:
- Not applicable. This is a physical medical device (microcatheter), not an algorithm.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):
- The "ground truth" in this context is based on engineering specifications and performance criteria for the physical device. For example, for tensile strength, the ground truth would be a defined minimum force the joints and marker band must withstand without failure. For compatibility, it would be the ability to successfully deliver and retrieve the microcatheter within the specified guide catheter with acceptable force.
8. The sample size for the training set:
- Not applicable. This is not an AI/ML device, so there is no training set.
9. How the ground truth for the training set was established:
- Not applicable for the same reason as point 8.
§ 870.1250 Percutaneous catheter.
(a)
Identification. A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.(b)
Classification. Class II (performance standards).