K Number
K251968
Date Cleared
2025-07-24

(28 days)

Product Code
Regulation Number
864.7340
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

For the quantitative determination of fibrinogen, based on the Clauss method, in human citrated plasma on IL Coagulation Systems.

Device Description

Several congenital abnormalities of fibrinogen result in impaired conversion of fibrinogen to fibrin during blood coagulation. Fibrinogen is also a useful marker in the evaluation of several disease states including disseminated intravascular coagulation, liver disease, inflammatory diseases and malignancy. High levels of fibrinogen are associated with an increased risk for cardiovascular disease. Increased levels are also found during pregnancy and oral contraceptive use, while reduced levels are found during thrombolytic therapy.

The HemosIL Fibrinogen-C kit and HemosIL Fibrinogen-C XL kit use an excess of thrombin to convert fibrinogen to fibrin in diluted plasma. At high thrombin and low fibrinogen concentration, the rate of reaction is a function of fibrinogen concentration.

AI/ML Overview

This document, an FDA 510(k) clearance letter for HemosIL Fibrinogen-C and HemosIL Fibrinogen-C XL, does not contain the information requested in your prompt regarding acceptance criteria and a study proving the device meets those criteria.

The 510(k) summary clearly states:

  • Reason for Submission: "This Special 510(k) is being submitted to remove the reconstituted reagent frozen stability claim of 1 month at -20°C in the original vial for HemosIL Fibrinogen-C and HemosIL Fibrinogen-C XL."
  • Data Requirement: "No new performance data are needed to remove the reconstituted reagent frozen stability claim."
  • No new performance claims: "Changes to labeled performance claims, except to remove the reconstituted reagent frozen stability claim from the Instructions for Use and add 'Do not freeze reconstituted reagent.'"

Therefore, the document explicitly states that no new performance data or studies were required or submitted for this particular 510(k) clearance. The clearance is based on a minor labeling change related to reagent stability, not on new performance validation.

To provide the information you requested (acceptance criteria, study details, sample sizes, expert involvement, ground truth, etc.), you would need a different type of submission document, such as an original 510(k) application or a PMA, where initial and comprehensive performance validation studies are typically included.

FDA 510(k) Clearance Letter - HemosIL Fibrinogen-C; HemosIL Fibrinogen-C XL

Page 1

U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov

Doc ID # 04017.08.00

July 24, 2025

Instrumentation Laboratory (IL) Co.
Kirivann Chhoeun
Regulatory Affairs Specialist III
180 Hartwell Rd
Bedford, Massachusetts 01730

Re: K251968
Trade/Device Name: HemosIL Fibrinogen-C; HemosIL Fibrinogen-C XL
Regulation Number: 21 CFR 864.7340
Regulation Name: Fibrinogen Determination System
Regulatory Class: Class II
Product Code: KQJ
Dated: June 26, 2025
Received: June 26, 2025

Dear Kirivann Chhoeun:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download). Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires

Page 2

K251968 - Kirivann Chhoeun Page 2

device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Takeesha Taylor-bell -S

Takeesha Taylor-Bell
Deputy Director
Division of Immunology and Hematology Devices
OHT7: Office of In Vitro Diagnostics
Office of Product Evaluation and Quality
Center for Devices and Radiological Health

Enclosure

Page 3

FORM FDA 3881 (8/23) Page 1 of 1 PSC Publishing Services (301) 443-6740 EF

DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration

Indications for Use

Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.

510(k) Number (if known): K251968

Device Name:
HemosIL Fibrinogen-C
HemosIL Fibrinogen-C XL

Indications for Use (Describe):
For the quantitative determination of fibrinogen, based on the Clauss method, in human citrated plasma on IL Coagulation Systems.

Type of Use (Select one or both, as applicable):
☒ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

Page 4

510(k) Summary

Special 510(k): HemosIL Fibrinogen-C, HemosIL Fibrinogen-C XL Page 1 of 4

510(k) Summary

This Special 510(k) Summary of Safety and Effectiveness is being submitted in accordance with the requirements of 21 CFR 807.92.

510(k) Submitter

Instrumentation Laboratory (IL) Company
180 Hartwell Road
Bedford, MA 01730-2443 (USA)
Establishment Registration: 1217183

Primary Contact

Kirivann Chhoeun
Regulatory Affairs Specialist III
Phone: 774-254-5359
Fax: 781-861-4207
Email: kchhoeun@werfen.com

Preparation Date: June 26, 2025

Device Trade Names

AssayPart No.
HemosIL Fibrinogen-C0020301100
HemosIL Fibrinogen-C XL0020003900

Predicate Device and 510(k) Number

HemosIL Fibrinogen-C K073367

Regulatory Information

Assays
ClassificationClass II
Classification PanelHematology (81)
Regulation Section No.21 CFR 864.7340
Regulation DescriptionFibrinogen determination system
Product CodeKQJ

Page 5

510(k) Summary Special 510(k): HemosIL Fibrinogen-C, HemosIL Fibrinogen-C XL Page 2 of 4

Device Description

HemosIL Fibrinogen-C and HemosIL Fibrinogen-C XL

Several congenital abnormalities of fibrinogen result in impaired conversion of fibrinogen to fibrin during blood coagulation. Fibrinogen is also a useful marker in the evaluation of several disease states including disseminated intravascular coagulation, liver disease, inflammatory diseases and malignancy. High levels of fibrinogen are associated with an increased risk for cardiovascular disease. Increased levels are also found during pregnancy and oral contraceptive use, while reduced levels are found during thrombolytic therapy.

The HemosIL Fibrinogen-C kit and HemosIL Fibrinogen-C XL kit use an excess of thrombin to convert fibrinogen to fibrin in diluted plasma. At high thrombin and low fibrinogen concentration, the rate of reaction is a function of fibrinogen concentration.

Intended Use / Indication for Use

HemosIL Fibrinogen-C and HemosIL Fibrinogen-C XL

For the quantitative determination of fibrinogen, based on the Clauss method, in human citrated plasma on IL Coagulation Systems.

Reason Submission Qualifies as Special 510(k)

This Special 510(k) is being submitted to remove the reconstituted reagent frozen stability claim of 1 month at -20°C in the original vial for HemosIL Fibrinogen-C and HemosIL Fibrinogen-C XL.

The submission meets the criteria for a Special 510(k) based on the following:

  • The proposed change is submitted by the manufacturer legally authorized to market the existing devices.
  • No new performance data are needed to remove the reconstituted reagent frozen stability claim.

In addition, the change in this submission does not introduce:

  • Changes to indications for use or intended use
  • Changes to operating principle
  • Changes to formulation
  • Changes to labeled performance claims, except to remove the reconstituted reagent frozen stability claim from the Instructions for Use and add "Do not freeze reconstituted reagent."
  • Changes to assay algorithms or data reduction software

Page 6

510(k) Summary Special 510(k): HemosIL Fibrinogen-C, HemosIL Fibrinogen-C XL Page 3 of 4

Comparison to Predicate

ItemPredicate Devices (K073367)Subject Devices
Trade NamesHemosIL Fibrinogen-CHemosIL Fibrinogen-CHemosIL Fibrinogen-C XL
ManufacturerInstrumentation Laboratory Co.Same

Similarities

MeasurandFibrinogenSame
Assay TypeQuantitativeSame
Product CodeKQJSame
Regulation Section21 CFR 864.7340Same
ClassificationClass IISame
Intended Use/ Indications for UseFor the quantitative determination of fibrinogen, based on the Clauss method, in human citrated plasma on IL Coagulation Systems.Same
MethodologyClauss methodSame
Compositionlyophilized bovine thrombin (35 UNIH/mL) with bovine albumin, calcium chloride, buffer and stabilizers.Same
Calibrators (Sold Separately)HemosIL Calibration PlasmaSame
Controls (Sold Separately)HemosIL Normal Control AssayedHemosIL Low Abnormal Control AssayedHemosIL Low Fibrinogen ControlSame
Sample TypeHuman citrated plasmaSame
InstrumentationACL Elite/ACL Elite ProACL TOP FamilyACL TOP Family 50 SeriesACL TOP Family 70 SeriesSame

Differences

Reconstituted StabilityCurrent Insert ClaimModified Insert Claim
1 month at -20°C in the original vialDo not freeze reconstituted reagent

Page 7

510(k) Summary Special 510(k): HemosIL Fibrinogen-C, HemosIL Fibrinogen-C XL Page 4 of 4

Conclusion

HemosIL Fibrinogen-C and HemosIL Fibrinogen-C XL remain substantially equivalent to their legally marketed predicate device, HemosIL Fibrinogen-C, last FDA cleared under K073367.

This submission is limited in scope to the removal of the reconstituted reagent frozen stability claim and the addition of an instruction advising users not to freeze the reconstituted reagent.

§ 864.7340 Fibrinogen determination system.

(a)
Identification. A fibrinogen determination system is a device that consists of the instruments, reagents, standards, and controls used to determine the fibrinogen levels in disseminated intravascular coagulation (nonlocalized clotting within the blood vessels) and primary fibrinolysis (the dissolution of fibrin in a blood clot).(b)
Classification. Class II (special controls). A control or fibrinogen standard intended for use with a fibrinogen determination system is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 864.9.