Search Results
Found 80 results
510(k) Data Aggregation
(85 days)
Versana Premier; Versana Premier Lotus; LOGIQ F
The Versana Premier/Versana Premier Lotus/LOGIQ F is a general-purpose diagnostic ultrasound system intended for use by qualified and trained healthcare professionals for ultrasound imaging, measurement, display and analysis of the human body and fluid.
Versana Premier/Versana Premier Lotus/LOGIQ F clinical applications include: Fetal/Obstetrics, Abdominal, Gynecology, Urology, Pediatric, Small Parts (includes breast, testes, thyroid), Cardiac Adult, Cardiac Pediatric, Vascular/Peripheral Vascular, Musculoskeletal Conventional, Musculoskeletal Superficial, Thoracic/Pleural, Transcranial, Transrectal, Transvaginal, Transesophageal, Interventional guidance (includes tissue biopsy, fluid drainage, vascular and nonvascular access).
Modes of operation include: B, M, PW Doppler, CW Doppler, Color Doppler, Color M Doppler, Power Doppler, Harmonic Imaging, Coded Pulse, 3D/4D Imaging mode and Combined modes: B/M, B/Color, B/PWD, B/Color/PWD, B/Power/PWD.
Versana Premier/Versana Premier Lotus/LOGIQ F is intended to be used in a hospital or medical clinic
The Versana Premier, Versana Premier Lotus and LOGIQ F is a general purpose, Track 3, diagnostic ultrasound system for use by qualified and trained healthcare professionals. The system is a mobile console that includes an operator control panel, display monitor and transducers. The console provides digital acquisition, processing, and display capability. The system has an internal battery to allow for acquisition while the system is not plugged into a power source. Acquisition can also be done while the system is connected to an AC power source. The operator control panel includes function keys, trackball, and a touch panel with a digital keyboard (physical keyboard as an option) as input sources of the device. The variety of transducers include convex, linear, sector, Bi-plane probe and mechanical 4D transducers. The access types include trans- body surface, transrectal, transvaginal, transcranial and transesophageal. The Versana Premier, Versana Premier Lotus and LOGIQ F share a common software and hardware platform. There may be different configurations commercially offered, however they are all within the overall design of the product.
The provided text is a 510(k) Summary for a new medical device submission (K242005) for the GE Versana Premier, Versana Premier Lotus, and LOGIQ F ultrasound systems. It focuses on demonstrating substantial equivalence to predicate devices rather than presenting a study proving that the device meets specific acceptance criteria for performance.
Therefore, the document does not contain the information requested regarding acceptance criteria and a study proving the device meets those criteria, specifically:
- A table of acceptance criteria and the reported device performance: Not provided. The document outlines a comparison to predicate devices, but no specific performance metrics with acceptance criteria are listed.
- Sample size used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective): Not provided, as no clinical study demonstrating performance against acceptance criteria was conducted.
- Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not provided.
- Adjudication method (e.g., 2+1, 3+1, none) for the test set: Not provided.
- If a multi-reader multi-case (MRMC) comparative effectiveness study was done, if so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not provided. The device is a diagnostic ultrasound system and not an AI-assisted diagnostic tool as described.
- If a standalone (i.e., algorithm only without human-in-the-loop performance) was done: Not provided.
- The type of ground truth used (expert consensus, pathology, outcomes data, etc.): Not provided.
- The sample size for the training set: Not provided.
- How the ground truth for the training set was established: Not provided.
Summary of what the document does state regarding meeting requirements:
The document states:
- "The subject of this premarket submission, Versana Premier/Versana Premier Lotus/LOGIO F did not require clinical studies to support substantial equivalence." (Page 8)
- It attests to the device's conformance with recognized performance standards and safety standards through non-clinical tests (acoustic output, biocompatibility, cleaning and disinfection effectiveness, thermal, electrical, electromagnetic, and mechanical safety).
- It lists several voluntary standards the device complies with (e.g., AAMI/ANSI ES60601-1, IEC 60601-1-2, ISO 10993-1, ISO 14971, NEMA PS 3.1 - 3.20, IEC 62359).
- It outlines quality assurance measures applied during development, including Risk Analysis, Requirements Reviews, Design Reviews, and various levels of testing (unit, integration, system, performance, safety).
In essence, the submission relies on demonstrating substantial equivalence to previously cleared predicate devices and compliance with recognized safety and performance standards for ultrasound systems, rather than on a new clinical study with specific acceptance criteria met by the device's performance.
Ask a specific question about this device
(251 days)
GEM Premier 7000 with IQM3
The GEM Premier 7000 with iQM3 is a portable critical care system for use by health care professionals to rapidly analyze lithium heparinized whole blood samples at the point of health care delivery in a clinical setting and in a central laboratory. The instrument provides quantitative measurements of pH, pCO2, sodium, potassium, chloride, ionized calcium, glucose, lactate, hematocrit, total bilirubin, and CO-Oximetry (tHb, O2Hb, MetHb, HHb, sO2*) parameters from arterial, venous, or capillary lithium heparinized whole blood. These parameters, along with derived parameters, aid in the diagnosis of a patient's acid/base status, electrolyte and metabolite balance and oxygen delivery capacity.
*s02 = ratio between the concentration of oxyhemoglobin and oxyhemoglobin plus deoxyhemoglobin.
- · pH, pCO2, and pO2 measurements in whole blood are used in the diagnosis and treatment of life-threatening acid- base disturbances.
- · Electrolytes in the human body have multiple roles. Nearly all metabolic processes depend on or vary with electrolytes:
- Sodium (Na+) measurements are used in the diagnosis and treatment of aldosteronism, diabetes insividus, adrenal hypertension, Addison's disease, dehydration, inappropriate antidiuretic secretion, or other diseases involving electrolyte imbalance.
- Potassium (K+) measurements are used to monitor electrolyte balance in the diagnosis and treatment
- of disease conditions characterized by low or high blood potassium levels.
- Ionized calcium (Ca++) measurements are used in the diagnosis and treatment of parathyroid disease, a variety of bone diseases, chronic renal disease, and tetany.
- Chloride (Cl-) measurements are used in the diagnosis and treatment of electrolyte and metabolic disorders, such as cystic fibrosis and diabetic acidosis.
- · Hematocrit (Hct) measurements in whole blood of the packed red cell volume of a blood sample are used to distinguish normal from abnormal states, such as anemia and erythrocytosis (an increase in the number of red cells).
- · Glucose (Glu) measurement is used in the diagnosis, monitoring and treatment of carbohydrate metabolism
- disturbances including diabetes mellitus, neonatal hypoglycemia, idiopathic hypoglycemia, and pancreatic islet cell carcinoma.
- · Lactate (Lac) measurement is used:
- to evaluate the acid-base status of patients suspected of having lactic acidosis;
- to monitor tissue hypoxia and strenuous physical exertion;
- in the diagnosis of hyperlactatemia.
- · Total Bilirubin (tBili) measurement is used to aid in assessing the risk of kernicterus and hyperbilirubinemia in neonates.
• CO-Oximetry (tHb, COHb, MetHb, O2Hb, HHb, and sO2) evaluates the ability of the blood to carry oxygen by measuring total hemoglobin and determining the percentage of functional and dysfunctional hemoglobin species.
– Total Hemoglobin (tHb): Total hemoglobin measurements are used to measure the hemoglobin content of whole blood for the detection of anemia.
-
COHo: Carboxyhemoglobin measurements are used to determine the carboxyhemoglobin content of human blood as an aid in the diagnosis of carbon monoxide poisoning.
-
MetHb: Methemoglobin measurements are used to determine different conditions of methemoglobinemia.
-
HHb: Deoxyhemoglobin, as a fraction of total hemoglobin, is used in combination with oxyhemoglobin to measure oxygen status.
-
O2Hb: Oxyhemoglobin, as a fraction of total hemoglobin, is used in combination with deoxyhemoglobin to measure oxygen status.
-
sO2: Oxygen saturation, more specifically the ratio between the concentration of oxyhemoglobin and oxyhemoglobin plus deoxyhemoglobin, is used to measure oxygen status.
The GEM Premier 7000 with iQMs system provides health care professionals with quantitative measurements of lithium heparinized whole blood pH, pCO2, pO2, Na*, K*, Ch, Ca**, glucose, lactate, Hct, total bilirubin and CO-Oximetry (tHb, O2Hb, COHb, MetHb, HHb, sO₂*) from arterial, venous or capillary samples at the point of health care delivery in a clinical setting and in a central laboratory.
*sO₂ = Ratio between the concentration of oxyhemoglobin plus deoxyhemoglobin plus deoxyhemoglobin.
Key Components:
Instrument: It employs a unique touch-sensitive color screen and a simple set of menus and buttons for user interaction. The analyzer guides operators through the sampling process with simple, clear messages and prompts.
PAK (Cartridge): All required components for sample analysis are contained in the GEM PAK, including sensors, optical cell for CO-Oximetry and total bilirubin, sampler, pump tubing, distribution valve, waste container and Process Control Solutions. The GEM PAK is an entirely closed analytical system. The operator cannot introduce changes to the analytical process before or during the GEM PAK's use-life on board the instrument. The GEM PAK has flexible menus and test volume options to assist facilities in maximizing efficiency. The EEPROM on the GEM PAK includes all solution values and controls the analyte menu and number of tests. The setup of the instrument consists of inserting the GEM PAK into the instrument. The instrument will perform an automated GEM PAK start-up during which the following is performed: warm-up (15 minutes), sensor conditioning (10 minutes), Process Control Solution (PCS) performance (15 minutes), all of which take about 40 minutes. After GEM PAK start-up, Auto PAK Validation (APV) process is automatically completed: two completely independent solutions traceable to NIST standards, CLSI procedures or internal standards, containing two levels of concentration for each analyte (PC Solution D and E), are run by the analyzer to validate the integrity of the PC Solutions and the overall performance of the analytical system. Note: GEM PAKs that include tBili analyte will require the successful performance of CVP 5 tBili. Includes all necessary components for hemolysis detection, such as an acoustofluidic flow cell, an LED light source and an optical detector, for appropriate flagging of potassium measurements in whole blood samples without additional sample volume or sample processing steps.
Intelligent Quality Management (iQM3): iQM3 is used as the quality control and assessment system for the GEM Premier 7000 system. iQM3 is an active quality process control program designed to provide continuous monitoring of the analytical process before, during and after sample measurement with real-time, automatic error detection, automatic correction of the system and automatic documentation of all corrective actions, replacing the use of traditional external QC. iQM3 introduces hemolysis detection in whole blood samples, enhancing quality assessment in the pre-analytical phase of testing.
Based on the provided text, the device in question is the GEM Premier 7000 with iQM3, which is a portable critical care system for analyzing blood samples. The document describes its comparison to a predicate device, the GEM Premier 5000, and discusses its performance studies.
Here's an analysis of the acceptance criteria and the study proving the device meets them:
1. A table of acceptance criteria and the reported device performance
The document does not provide a direct table of specific numerical acceptance criteria for each analyte's performance (e.g., pH, pCO2, Na+, etc.) nor does it list the reported device performance in those exact terms. Instead, it states that "All verification activities were performed in accordance to established plans and protocols and design control procedures. Testing verified that all acceptance criteria were met."
The "Performance Summary" section lists the types of studies conducted to demonstrate that the modifications (specifically the new iQM quality check/Hemolysis detection module) do not impact the performance data represented in the Operators Manual, aligning with recognized guidelines. This implies the acceptance criteria are tied to maintaining performance comparable to the predicate device and being within acceptable ranges as defined by the mentioned CLSI guidelines.
Therefore, a table of explicit numerical acceptance criteria and reported performance values for each analyte is NOT AVAILABLE in the provided text. The document broadly states that the device met its acceptance criteria.
2. Sample size used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective)
The document mentions several types of performance studies:
- Verification (Internal Method Comparison, Internal Whole Blood Precision, Hemolysis Interference on Potassium, Hemolysis Verification)
- Shelf-life and Use-life studies
However, the specific sample sizes used for these test sets are NOT provided in the text. There is also no information about the data provenance (e.g., country of origin of the data, retrospective or prospective).
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
This information is NOT available in the provided text. The device is an in-vitro diagnostic (IVD) instrument that provides quantitative measurements of various blood parameters. The "ground truth" for such devices typically comes from reference methods, calibrated standards, or comparative analyses with established, highly accurate laboratory instruments, rather than human expert consensus on interpretations like with imaging.
4. Adjudication method (e.g., 2+1, 3+1, none) for the test set
Given that this is an IVD device for quantitative measurements of blood parameters, the concept of "adjudication" by multiple human readers (like in imaging studies) does not directly apply. Performance is assessed through analytical accuracy, precision, and interference studies against known standards or reference methods. Therefore, no adjudication method in the sense of expert consensus on interpretations is described or implied.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
There is no indication that a multi-reader multi-case (MRMC) comparative effectiveness study was performed. This type of study is relevant for AI-assisted diagnostic tools where human interpretation is part of the workflow. The GEM Premier 7000 with iQM3 is described as an analytical instrument providing direct quantitative measurements, not an AI system assisting human readers with interpretation. The "iQM3" refers to Intelligent Quality Management, which is an automated quality control system for the instrument itself, not an AI for human diagnostic assistance.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
The device itself is a standalone analytical instrument. The performance studies described (Internal Method Comparison, Internal Whole Blood Precision, Hemolysis Verification, etc.) essentially represent "standalone" performance, as they evaluate the accuracy and precision of the instrument's measurements directly. The iQM3 system is an internal quality control mechanism for the device's measurements. Therefore, yes, a standalone performance evaluation of the device's analytical capabilities was implicitly done.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)
For a device that provides quantitative measurements of blood parameters, the "ground truth" for the test set would typically be established using:
- Reference methods: Highly accurate and precise laboratory methods for measuring each analyte.
- Calibrated standards: Solutions with precisely known concentrations of the target analytes.
- Comparison to predicate device: As this is a 510(k) submission, a primary method of establishing "ground truth" performance for the new device is by comparing its measurements against those of a legally marketed predicate device (GEM Premier 5000), which itself would have been validated against reference methods and standards.
The text mentions "two completely independent solutions traceable to NIST standards, CLSI procedures or internal standards" for "Auto PAK Validation (APV)". This strongly suggests that traceable standards and potentially CLSI-defined reference methods were used to establish the ground truth for performance evaluation.
8. The sample size for the training set
The document describes the GEM Premier 7000 with iQM3 as a medical device for quantitative measurements, not explicitly as a machine learning/AI model that requires a "training set" in the conventional sense (i.e., for supervised learning). The iQM3 is an "active quality process control program" with "Pattern Recognition (PR) software." While pattern recognition might involve some form of "training" or calibration, the document does not specify a separate "training set" in terms of data volume for such a process. It focuses on the validation of the device's analytical performance. Therefore, the concept of a "training set" sample size as applicable to AI/ML devices is not explicitly discussed or provided.
9. How the ground truth for the training set was established
As noted above, the primary function of GEM Premier 7000 with iQM3 is quantitative measurement. If the "iQM3" component involved training for its "Pattern Recognition (PR) software," the document does not detail how a specific ground truth for such training was established. It primarily discusses the use of "Process Control Solutions (PCS)" and "Calibration Valuation Product (CVP 5)" for system checks and validation ("Auto PAK Validation (APV) process"). These solutions, traceable to NIST or CLSI standards, function as internal reference points for the device's operational checks and quality control, which could be considered an ongoing form of "ground truth" to maintain analytical performance, rather than a one-time "training set" for model development.
Ask a specific question about this device
(157 days)
ReliOn Premier BLU Blood Glucose Monitoring System
The ReliOn Premier BLU Blood Glucose Monitoring System is comprised of the ReliOn Premier BLU Blood Glucose Meter and the ReliOn Premier Blood Glucose Test Strips.
The ReliOn Premier BLU Blood Glucose Monitoring System is intended for the quantitative measurement of glucose in fresh capillary whole blood samples drawn from the fingertips.
The ReliOn Premier BLU Blood Glucose Monitoring System is intended for self-testing outside the body (in vitro diagnostic use) by people with diabetes at home as an aid to monitor the effectiveness of diabetes control.
The system is intended to be used by a single person and should not be shared.
The system is not intended for use on neonates and is not for the diagnosis or screening of diabetes.
The ReliOn Premier BLU Blood Glucose Monitoring System is a blood glucose meter with Bluetooth Low Energy technology that the meter communicates to smartphones wirelessly. The ReliOn Premier BLU BGMS has a target range indicator that upon consulting with a healthcare professional, users have the option to activate the function that displays blood glucose measurement results on a colored backlight display. By default, this indicator function is set to OFF. However, it is recommended that users consult with a healthcare professional to determine the target range for their individual needs.
This document describes the FDA's 510(k) clearance for the ReliOn Premier BLU Blood Glucose Monitoring System, comparing it to a predicate device. The information required for a comprehensive study description, such as that for an AI-based medical device, is not provided in this document. This submission focuses on engineering modifications of an existing device and its equivalence to a predicate, not on a new algorithm's performance or a comparative effectiveness study involving human readers and AI.
Therefore, many of the requested details, particularly regarding AI-specific studies (e.g., MRMC studies, standalone algorithm performance, training data details, expert consensus for ground truth on AI models), are not applicable to this submission.
However, I can extract the information relevant to the device's technical specifications and the non-clinical testing performed.
1. A table of acceptance criteria and the reported device performance
The document does not provide a table of precise quantitative acceptance criteria (e.g., specific thresholds for accuracy, precision) or detailed reported device performance data as would be seen for a new algorithm study. Instead, it describes equivalence to a predicate device and mentions that the non-clinical tests indicated the device "performed as intended and fulfilled the specifications."
The key performance characteristics are compared to the predicate device:
Feature | Predicate device CareSens N Premier BT (K170614) | Candidate device ReliOn Premier BLU Blood Glucose Monitoring System | Performance (as stated) |
---|---|---|---|
Measurement range | 20-600 mg/dL | Same | Same |
Sample size | Minimum 0.5 µL | Same | Same |
Test time | 5 seconds | Same | Same |
Sample type | Fresh capillary whole blood | Same | Same |
Calibration | Plasma-equivalent | Same | Same |
Battery life | 1,000 tests | Same | Same |
Power | Two 3.0V lithium batteries (disposable, type CR2032) | Same | Same |
Memory | 1,000 test results | Same | Same |
Target range indicator | N/A | Red, Green, Blue colored screens for below, within, and above target range. | New feature added; functionality verified. |
Acceptance Criteria (Implied from the document):
The acceptance criteria are implicitly that the ReliOn Premier BLU Blood Glucose Monitoring System demonstrates "substantial equivalence" to the predicate device, CareSens N Premier BT Blood Glucose Monitoring System (K170614), and that the modifications (target range function, LCD, PCB, labeling) perform as intended and fulfill specifications. Specific quantitative criteria for accuracy, precision, or other performance metrics (e.g., ISO standards for blood glucose meters) are not detailed in this summary.
Reported Device Performance:
"The results of these tests indicated that the ReliOn Premier BLU Blood Glucose Monitoring System, with the proposed changes outlined in this document, performed as intended and fulfilled the specifications."
2. Sample size used for the test set and the data provenance
The document does not specify the sample size for any test set (e.g., number of subjects or samples used in the non-clinical tests like software function, power consumption, etc.).
The data provenance (e.g., country of origin, retrospective or prospective) is also not specified for the non-clinical tests mentioned. Given the manufacturer's location, testing likely occurred in South Korea, but this is not confirmed.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
This is not applicable. The ground truth for a blood glucose monitoring system is typically established against a laboratory reference method (e.g., YSI analyzer) for glucose concentration, not by human experts interpreting results. The document does not mention any expert involvement in establishing ground truth for the non-clinical tests.
4. Adjudication method for the test set
This is not applicable. Adjudication methods are typically used in studies involving human interpretation or subjective assessments, often in imaging or clinical trial contexts. This document describes non-clinical engineering and performance tests for a blood glucose meter.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
This is not applicable. This document describes a continuous glucose monitoring system, not an AI imaging or diagnostic tool that assists human readers. No MRMC study was conducted or mentioned.
6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done
This is not applicable in the context of an AI algorithm. This device is a measurement system; its "standalone performance" refers to its accuracy in measuring blood glucose, which is evaluated through specific performance studies (e.g., accuracy against a reference method, precision studies). While these types of studies were undoubtedly performed as part of the overall testing (likely consistent with ISO 15197 for blood glucose monitoring systems, though not explicitly stated in this summary), the details of such quantitative performance studies are not provided in this 510(k) summary. The summary only broadly states that the device "performed as intended and fulfilled the specifications."
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)
For a blood glucose monitoring system, the ground truth for glucose measurements is typically established using a laboratory reference method (e.g., YSI glucose analyzer) on venous plasma or serum samples. This is a highly accurate chemical measurement. The document does not explicitly state the ground truth method but this is standard practice for blood glucose meter validation.
8. The sample size for the training set
This is not applicable. This document describes a medical device (blood glucose meter), not an AI model that requires a training set.
9. How the ground truth for the training set was established
This is not applicable. As above, this is an engineering device, not an AI model with a training set.
Ask a specific question about this device
(338 days)
Premier Resolution System
The Premier Resolution System is an automated High Performance Liquid Chromatography (HPLC) system which performs the separation of hemoglobin species in venous whole blood samples for the quantitative analysis of normal hemoglobin (A, A2, and F), and the qualitative detection of major variant hemoglobin S, C, D-Los Angeles, and E in adult, adolescent, children and infant populations. The assays are performed on venous whole blood samples collected in tubes containing K2EDTA as anticoagulant.
The Premier Resolution System is intended for Professional Laboratory Use only.
The Premier Resolution System is intended for use with analytical components and reagents provided by Trinity Biotech.
The Premier Resolution System is intended to be used in conjunction with other laboratory and clinical findings.
For In Vitro Diagnostic Use.
The Premier Resolution System consists of a high performance liquid chromatographic analyzer, reagents, analytical column and software which allows for the fractionation and quantitation of fetal hemoglobin (Hb F), and hemoglobin A2 (Hb A2), and with fractionation and presumptive identification of abnormal hemoglobin variants. This is accomplished using the principles of ion-exchange (IEX) high performance liquid chromatography (HPLC).
This document describes the performance data for the "Premier Resolution System," an automated High Performance Liquid Chromatography (HPLC) system for hemoglobin analysis. The study aims to demonstrate that the device is substantially equivalent to a predicate device, the Bio-Rad Variant II ß-thalassemia (K991127).
1. Acceptance Criteria and Reported Device Performance:
The document outlines comparative performance against a predicate device (Bio-Rad Variant II ß-thalassemia) and precision studies. The acceptance criteria are implicitly defined by demonstrating comparability and acceptable precision, rather than explicit thresholds for each metric. The reported device performance includes:
Acceptance Criteria (Implicit) | Reported Premier Resolution System Performance (Quick Scan Assay) | Reported Premier Resolution System Performance (High Resolution Assay) |
---|---|---|
Correlation (Method Comparison) - Mean Bias vs. Predicate (Bio-Rad Variant II) | ||
HbF comparability | -0.3 bias (1.1 to 48.9% interval) with 160 patient results | -0.4 bias (1.1 to 46.6% interval) with 158 patient results |
HbA comparability | 0.7 bias (2.5 to 89.7% interval) with 682 patient results | 2.4 bias (3.5 to 90.5% interval) with 586 patient results |
HbA2 comparability | 0.1 bias (1.6 to 6.1% interval) with 602 patient results | 0.1 bias (1.6 to 6.0% interval) with 598 patient results |
HbS comparability | 0.3 bias (6.8 to 67.1% interval) with 106 patient samples | 1.3 bias (1.9 to 67.9% interval) with 110 patient samples |
HbC comparability | -1.1 bias (9.5 to 82.8% interval) with 49 patient results | -1.0 bias (10.2 to 82.5% interval) with 49 patient results |
HbD-LA comparability | 1.6 bias (11.6 to 82.7% interval) with 17 patient results | 2.7 bias (11.7 to 84.1% interval) with 17 patient results |
HbE comparability | -3.0 bias (5.5 to 70.4% interval) with 25 patient results | -4.9 bias (5.3 to 66.7% interval) with 25 patient results |
Precision (Single Site) - Within-Laboratory %CV | ||
HbA (High) | 3.61% | 3.63% |
HbA (Mid) | 0.89% | 1.37% |
HbA2 (Mid) | 2.23% | 7.22% |
HbA2 (Low) | 5.99% | 7.15% |
HbF (High) | 1.05% | 3.49% |
HbF (Mid) | 3.26% | 9.10% |
HbS (High) | 0.89% | 1.33% |
HbS (Mid) | 0.98% | 1.69% |
HbC (High) | 0.78% | 1.15% |
HbC (Mid) | 1.75% | 2.03% |
HbD (High) | 1.88% | 2.20% |
HbD (Mid) | 2.04% | 1.32% |
HbE (High) | 2.84% | 4.63% |
HbE (Mid) | 3.06% | 3.14% |
LoD/LoQ (Quick Scan) | ||
HbF LoD | 0.2% | 0.1% |
HbF LoQ | 1.1% | 1.1% |
HbA LoD | 0.1% | 0.7% |
HbA LoQ | 2.3% | 2.2% |
HbA2 LoD | 0.1% | 0.2% |
HbA2 LoQ | 1.5% | 1.5% |
HbS LoD | 0.1% | 0.3% |
HbS LoQ | 1.0% | 0.9% |
HbC LoD | 0.1% | 0.3% |
HbC LoQ | 1.0% | 1.7% |
HbD-LA LoD | 0.1% | 0.1% |
HbD-LA LoQ | 1.5% | 1.4% |
HbE LoD | 0.1% | 0.6% |
HbE LoQ | 1.5% | 2.7% |
2. Sample Size and Data Provenance (Test Set):
- Correlation (Method Comparison): A total of 780 unique patient samples were collected and analyzed. The data provenance is described as being collected and analyzed at three (3) professional external laboratory sites. It is implicit that these were retrospective real-world samples, as they are referred to as "patient samples" used for method comparison against an existing device. The country of origin is not explicitly stated, but given the FDA submission, it is likely the US.
- Precision (Single Site): The sample size for each analyte was 80 data points, generated by a 20x2x2 study design over 20 days, with two runs per day and two replicates per run. These were "samples of varying concentrations" and not explicitly patient samples.
- Precision (Multisite): The sample size for each analyte was 75 data points (3x5x5 study design across three external sites, over five days with five replicates per day). These were "precision samples."
- Limits of Detection: 60 determinations of low-level samples for each hemoglobin type. These were "human whole blood samples" with varying levels of hemoglobins.
3. Number of Experts and Qualifications for Ground Truth (Test Set):
This device is an in-vitro diagnostic (IVD) instrument for quantitative and qualitative analysis of hemoglobin species. The ground truth for such devices is typically established by well-characterized reference methods or by comparison to a legally marketed predicate device. In this submission, the primary method for establishing the device's performance is:
- Correlation (Method Comparison): Comparison against the Bio-Rad Variant II ß-thalassemia, which serves as the "ground truth" or reference for establishing substantial equivalence. No human experts are described as establishing ground truth for the test set, as the ground truth is the measurement from the predicate device.
4. Adjudication Method for the Test Set:
Not applicable. The study design involves direct comparison of quantitative measurements from two analytical instruments (device under review vs. predicate device), and precision of the device itself. There is no subjective interpretation requiring adjudication by experts.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:
Not applicable. This is an in-vitro diagnostic device that provides quantitative and qualitative measurements, not an image-based diagnostic system requiring human interpretation with or without AI assistance. Therefore, an MRMC study is not relevant.
6. Standalone Performance (Algorithm Only without Human-in-the-Loop):
Yes, the entire study focuses on the standalone performance of the "Premier Resolution System" as an automated HPLC system. There is no human-in-the-loop component described for its routine operation or for the performance studies presented. The device performs the analysis and provides results independently.
7. Type of Ground Truth Used:
The ground truth for the device's performance is established mainly through:
- Comparison to a legally marketed predicate device: The Bio-Rad Variant II ß-thalassemia, for method comparison (correlation).
- Internal consistency and statistical measures: For precision (repeatability, within-laboratory, reproducibility) and limits of detection/quantitation. This relies on the inherent analytical capabilities and controls of the new device itself.
8. Sample Size for the Training Set:
The document does not explicitly describe a "training set" in the context of machine learning or AI models. This device is an automated HPLC system, where performance is based on chemical separation principles, not a learning algorithm that requires a separate training phase with labeled data in the AI sense. Its "training" or development would involve chemical and engineering optimization.
9. How the Ground Truth for the Training Set Was Established:
Not applicable, as there is no mention of a "training set" in the context of an AI/ML model for this device. The development of an HPLC system involves instrument design, reagent formulation, and software development, which are validated through analytical performance studies like those presented (precision, linearity, method comparison, etc.).
Ask a specific question about this device
(94 days)
Premier HpSA Flex (619096)
The Premier HpSA Flex enzyme immunoasay (EIA) is an in vitro qualitative procedure for the detection of Helicobacter pylori antigens in human stool. The test is intended for use with unpreserved stool specimens or preserved stool specimens in transport media. Test results are intended to aid in the diagnosis of H. pylori infection and to monitor response during and post- therapy in patients. Accepted medical practice recommends that testing by any current method, to confirm eradication, be done at least four weeks following completion of therapy.
Meridian Bioscience has modified its FDA-cleared PREMIER Platinum HpSA® PLUS assay (K182559), a qualitative, in vitro diagnostic test for the detection of Helicobacter pylori antigens present in unpreserved human stool specimens. This modification, to be marketed under new device trade name Premier HpSA® Flex upon FDA clearance, is the addition of a new specimen type claim to the intended use of the previously cleared device (K182559) whereby specimens may be preserved in Cary-Blair or Culture and Sensitivity (C&S) transport media.
The Premier HpSA Flex test is a microwell-based enzyme immunoassay that detects H. pylori antigens present in human stool specimens, either unpreserved in transport media. The test uilizes a plurality (mixture) of monoclonal anti-H. pylori capture antibodies adsorbed to microwells. Diluted patient samples and an enzyme conjugate reagent are added to the microwells and incubated for one hour at room temperature. A wash is performed to remove unbound material. Substrate is added and incubated for 10 minutes at room temperature. Color develops in the presence of bound enzyme. Stop solution is added and the results are interpreted visually or spectrophotometrically.
Here's a breakdown of the acceptance criteria and study that proves the device meets them, based on the provided text.
Acceptance Criteria and Device Performance
The core of this submission is about adding a new specimen type claim (preserved stool in Cary-Blair or C&S transport media) to an already FDA-cleared device. Therefore, the "acceptance criteria" revolve around demonstrating that the device performs equivalently with these new specimen types as it did with the original unpreserved stool and that the performance remains robust.
Here's a summary of the performance characteristics presented as implicit acceptance criteria and the reported device performance:
Acceptance Criteria (Implicit by Study Design) | Reported Device Performance (Premier HpSA Flex with Preserved Stool) |
---|---|
Analytical Sensitivity (Limit of Detection - LoD): Demonstrate a specific LoD for H. pylori antigen in preserved stool. | LoD = 12 ng/ml in Cary-Blair or C&S transport media. (Previously established LoD for unpreserved stool was 4.66 ng/mL). Equivalence between Cary-Blair and C&S media at LoD and below LoD antigen concentrations was determined. |
Precision/Reproducibility: Demonstrate consistent results across different laboratories, operators, and kit lots with preserved stool samples. | Overall agreement between assay result and expected result was 100.0% (95% CI: 98.9-100.0%). (Reproducibility with unpreserved stool was previously evaluated under K182559). |
Specimen Storage Stability: Demonstrate stability of preserved stool specimens under various temperature and duration conditions. | Specimens stable up to 120 hours at 2-8°C or 19-27°C, or up to 14 days frozen (-20°C and/or -80°C). |
Freeze/Thaw Stability: Demonstrate robustness of preserved stool specimens to multiple freeze/thaw cycles. | Stable for up to two (2) freeze/thaw cycles when stored frozen (≤ -20°C). |
Analytical Specificity/Interference: Show no interference from common chemical and biological substances found in stool. | No interference observed for any of the evaluated substances (TUMS, Mylanta, Pepto-Bismol, Tagamet, Prilosec OTC, Barium Sulfate, Whole Blood, Leukocytes, Mucin, Hemoglobin, Stearic Acid, Palmitic Acid, NSAID, Ibuprofen) at their respective test concentrations. (Same substances previously evaluated for predicate). |
Analytical Specificity/Cross-Reactivity: Show no cross-reactivity with common microorganisms or interference with H. pylori detection. | No cross-reactivity or microbial interference observed with any of the tested bacteria, fungi, and viral strains. (Same organisms previously evaluated for predicate). |
Method Comparison (Clinical Performance with a Comparator Device): Achieve acceptable positive and negative percent agreement with an FDA-cleared comparator device using preserved stool. | Positive Agreement: 100.0% (49/49) [95% CI: 92.7% - 100.0%] |
Negative Agreement: 98.5% (131/133) [95% CI: 94.7% - 99.6%] |
Study Details
-
Sample sizes used for the test set and the data provenance:
- Analytical Sensitivity (LoD): Not explicitly stated how many samples per lot were used in the LoD study, but it mentions "Three lots" and "positive results >= 95% of the time."
- Precision/Reproducibility: 360 samples (10 panels x 12 blinded samples x 3 laboratories). The samples were "contrived stool samples" with H. pylori antigen spiked in. Data provenance is implied to be domestic (USA) due to the submission context, and it's a prospective study looking at controlled, contrived samples.
- Preserved Specimen Storage Stability: Not explicitly stated how many samples were used, but the study was designed to validate stability claims.
- Freeze/Thaw Stability: Not explicitly stated how many samples were used.
- Analytical Specificity/Interference: Not explicitly stated how many samples were used, but testing was performed in the presence of various substances.
- Analytical Specificity/Cross-reactivity: Not explicitly stated how many samples were used, but each organism was tested with a true negative and a contrived low positive sample at specified concentrations.
- Method Comparison (Clinical Performance): 200 archived stool specimens were enrolled, of which 182 were evaluable and used for the comparison. Data provenance is of archived specimens from patients, suggesting retrospective data. The specific country of origin is not mentioned.
-
Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
- This device is an in-vitro diagnostic (IVD) for detecting antigens, not an imaging device requiring expert interpretation for ground truth.
- For the Precision/Reproducibility study, "expected assay result" was used as ground truth for contrived samples. This implies the ground truth was based on the known concentration of spiked antigen.
- For the Method Comparison study, the comparison was against an "FDA-cleared comparator device" and "Standard of Care (SoC) testing using an FDA-cleared commercial assay." The "ground truth" for clinical performance appears to be established by the results of these existing FDA-cleared methods. No human expert consensus was used to define the ground truth for individual samples.
-
Adjudication method (e.g., 2+1, 3+1, none) for the test set:
- No human adjudication method was mentioned or implied, as the device is an IVD detecting antigens, and ground truth was established either by known concentrations in contrived samples or by results from existing FDA-cleared assays.
-
If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
- No MRMC comparative effectiveness study was conducted. This type of study is typically performed for AI-assisted diagnostic imaging devices where human interpretation directly impacts results. This device is an immunoassay (IVD).
-
If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:
- Yes, the performance data provided (LoD, Reproducibility, Interference, Cross-reactivity, Method Comparison) represent the standalone performance of the Premier HpSA Flex assay. It's an automated or semi-automated EIA, not an algorithm requiring human interaction for its direct output.
-
The type of ground truth used (expert consensus, pathology, outcomes data, etc.):
- For analytical performance studies (LoD, Precision, Interference, Cross-Reactivity), the ground truth was known concentrations of H. pylori antigen in contrived samples or known presence/absence of interfering/cross-reacting substances/organisms.
- For the method comparison (clinical performance), the ground truth was established by results from an FDA-cleared comparator device and Standard of Care (SoC) testing using an FDA-cleared commercial assay.
-
The sample size for the training set:
- This is a traditional in-vitro diagnostic device (immunoassay), not a machine learning/AI algorithm that requires a "training set" in the conventional sense. The device's components and parameters are developed through standard chemistry and assay development processes, not through iterative training on a dataset.
-
How the ground truth for the training set was established:
- As stated above, there is no "training set" for an immunoassay in the context of AI/ML. The "ground truth" for the development of the assay itself would align with standard analytical validation methods, ensuring the assay accurately detects the target analyte (H. pylori antigens) at various concentrations and in the presence of relevant interferents, against known standards.
Ask a specific question about this device
(119 days)
GEM Premier ChemSTAT
The GEM Premier ChemSTAT is a portable critical care system for use by health care professionals to rapidly analyze lithium heparinized whole blood samples at the point of health care delivery in a clinical setting and in a central laboratory. The instrument provides quantitative measurements of sodium (Na+), Potassium (K+), Ionized Calcium (Ca++), Chloride (Cl-), Glucose (Glu), Lactate (Lac), Hematocrit (Hct), Creatinine (Crea), Blood Urea Nitrogen (BUN), Total Carbon Dioxide (tCO2), pH, and partial pressure of carbon dioxide (pCO2) from arterial and venous heparinized whole blood. These parameters, along with derived parameters, aid in the diagnosis of a patient's acid/base status, electrolyte and metabolite balance.
Electrolytes in the human body have multiple roles. Nearly all metabolic processes depend on or vary with electrolytes:
· Sodium (Na+) measurements are used in the diagnosis and treatment of aldosteronism, diabetes insipidus, adrenal hypertension, Addison's disease, dehydration, inappropriate antidiuretic secretion, or other diseases involving electrolyte imbalance.
· Potassium (K+) measurements are used to monitor electrolyte balance in the diagnosis and treatment of disease conditions characterized by low or high blood potassium levels.
· Ionized calcium (Ca++) measurements are used in the diagnosis and treatment of parathyroid disease, a variety of bone diseases, chronic renal disease and tetany. · Chloride (Cl-) measurements are used in the diagnosis and treatment of electrolyte and metabolic disorders, such as cystic fibrosis and diabetic acidosis.
· Glucose (Glu) measurement is used in the diagnosis, monitoring and treatment of carbohydrate metabolism disturbances including diabetes mellitus, neonatal hypoglycemia, idiopathic hypoglycemia, and of pancreatic islet cell carcinoma.
· Lactate (Lac) measurement is used to evaluate the acid-base status of patients suspected of having lactic acidosis, to monitor tissue hypoxia and strenuous physical exertion, and in the diagnosis of hyperlactatemia.
· Hematocrit (Hct) measurements in whole blood of the packed red cell volume of a blood sample are used to distinguish normal from abnormal states, such as anemia and erythrocytosis (an increase in the number of red cells).
· Creatinine (Crea) measurements are used in the diagnosis and treatment of renal diseases and in monitoring renal dialysis.
· Blood Urea Nitrogen (BUN) or urea measurements are used for the diagnosis, monitoring, and treatment of certain renal and metabolic diseases.
· Total carbon dioxide/tCO2 (also referred to as bicarbonate/HCO3-) is used in the diagnosis, monitoring, and treatment of numerous potentially serious disorders associated with changes in body acid-base balance.
· pH and pCO2 measurements in whole blood are used in the diagnosis and treatment of life-threatening acid-base disturbances.
The GEM Premier ChemSTAT system provides fast, accurate, quantitative measurements of Sodium (Na"), Potassium (K*), Ionized Calcium (Ca*), Chloride (Cl·), Glucose (Glu), Lactate (Lac), Hematocrit (Hct), Creatinine (Crea), Blood Urea Nitrogen (BUN), Total Carbon Dioxide (tCO2), pH, and partial pressure of carbon dioxide (pCO2) from arterial and venous lithium heparinized whole blood.
The provided text describes a Special 510(k) submission for an upgrade to the operating system of the GEM Premier ChemSTAT device. The device itself is an in vitro diagnostic (IVD) system for quantitative measurements of various blood parameters. The submission focuses on the software upgrade rather than a change in the device's fundamental function or performance.
Therefore, the "acceptance criteria" and "reported device performance" in this context refer to the successful verification and validation of the software upgrade and the continued adherence to the established performance of the unmodified device, as the indications for use and performance claims remain unchanged. The study proving this essentially consists of the software verification and validation activities.
Here's the information extracted from the document, tailored to the context of a software upgrade:
1. Table of Acceptance Criteria and Reported Device Performance
Since this is a software upgrade with no changes to the performance claims of the device, the general acceptance criteria are that the upgraded software performs as intended without adversely affecting the device's established performance specifications. The reported device performance is that these criteria were met.
Acceptance Criteria (Software Upgrade) | Reported Device Performance (Software Upgrade) |
---|---|
All identified risks associated with the design changes for the modified device are mitigated. | Risk assessments were performed in compliance with ISO 14971:2019, and identified risks were mitigated. |
All software verification and validation activities are completed according to established plans and protocols. | All verification and validation activities were performed in accordance with established plans and protocols and Design Control procedures. |
All acceptance criteria for software verification and validation are met. | Testing verified all acceptance criteria were met. |
Cybersecurity vulnerabilities are identified, assessed, and compensating controls are implemented. | Cybersecurity assessments were performed, vulnerabilities identified and assessed, and compensating controls implemented to mitigate threats and safeguard data. |
No changes to indications for use or intended use. | No changes to indications for use or intended use. |
No changes to the fundamental scientific technology. | No changes to the fundamental scientific technology. |
No changes to operating principle. | No changes to operating principle. |
No changes to labeled performance claims. | No changes to labeled performance claims. |
2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)
The document does not specify a "test set sample size" or "data provenance" in the traditional sense for evaluating diagnostic performance. The focus is on software verification and validation. Therefore, the "sample" for testing the software functionality would be the various test cases and scenarios designed to validate the operating system upgrade and its interaction with the GEM Premier ChemSTAT application software.
The document states: "Performance data is limited to Software Verification and Validation as the scope of this Special 510(k) is specific to an operating system upgrade from Fedora 17 Linux to WindRiver LTS 18 Linux."
Further details on the specific number of test cases, the nature of the data (e.g., simulated, actual runs on the device), or its origin are not provided in this summary.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)
This information is not applicable to a software operating system upgrade as described. "Ground truth" in the context of expert consensus is typically relevant for diagnostic performance studies where human interpretation or a gold standard reference is needed (e.g., pathology for an imaging device). Here, the "ground truth" is the proper functioning of the software and its integration with the hardware, which is evaluated through engineering and software testing.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
This information is not applicable for a software operating system upgrade. Adjudication methods like 2+1 or 3+1 are used in clinical studies to resolve discrepancies in expert interpretation of diagnostic results. Software verification and validation typically rely on predefined test outcomes and engineering assessments.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
This information is not applicable. An MRMC comparative effectiveness study is used to evaluate the impact of an AI algorithm on human reader performance, usually for diagnostic tasks. This submission is for a software operating system upgrade for an existing IVD device, not for a new AI algorithm.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
The concept of "standalone performance" in the context of an algorithm's diagnostic capability (like an AI algorithm) is not directly applicable here. The device itself (GEM Premier ChemSTAT) operates to provide quantitative measurements. The software upgrade ensures the continued, correct operation of the device. The verification and validation activities demonstrate that the upgraded software performs its functions correctly as part of the overall device system.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)
For this software upgrade, the "ground truth" is the expected behavior and functionality of the software and the device. This is established through:
- Functional specifications: The software is expected to perform according to its design specifications.
- Risk analysis: The software should not introduce new risks or fail to mitigate existing ones.
- Cybersecurity standards: The software should meet cybersecurity requirements.
- Established device performance: The software upgrade should not negatively impact the established analytical and clinical performance of the GEM Premier ChemSTAT device (which relies on the physical and chemical principles of its measurements).
The document explicitly states that the changes "do not introduce...changes to labeled performance claims." This implies that the performance of the device (e.g., accuracy, precision of Na+, K+, Glu measurements) remains the same as previously cleared, and the software upgrade was validated not to alter these.
8. The sample size for the training set
This information is not applicable. Training sets are used for machine learning models. This submission describes a conventional software operating system upgrade (Fedora 17 Linux to WindRiver LTS 18 Linux) for an existing IVD device, not the development or retraining of a machine learning algorithm.
9. How the ground truth for the training set was established
This information is not applicable, as there is no training set for a machine learning model; it is a software operating system upgrade.
Ask a specific question about this device
(1 days)
Premier's MultiMatch Universal Chameleon Restorative Composite
Premier's MultiMatch Universal Chameleon Restorative Composite is a nano-hybrid dental restorative intended for direct placement in all cavity classes in anterior teeth. Additional indications include repair of enamel defects, repair of provisionals, and repair of porcelain restorations, minor occlusal build-ups, and incisal abrasions.
Premier's MultiMatch Universal Chameleon Restorative Composite is a light-cured, radiopaque, resin-based nanohybrid restorative composite. It is formulated for direct placement in any type of anterior and posterior cavity form. It is intended to be used with a dental bonding agent and can be sculpted prior to curing. Once light cured, the composite can be easily finished and polished to achieve a highly esthetic and durable restoration. Available in three chameleon shades, MultiMatch can be dispensed via a unit-dose capsule or syringe to provide a restoration with an excellent visual color match to the surrounding dentition.
This document is a 510(k) Summary for Premier's MultiMatch Universal Chameleon Restorative Composite. It seeks to demonstrate substantial equivalence to a predicate device, SimpliShade (K162257).
The document does not describe a study involving an AI/Machine Learning algorithm for medical image analysis or similar diagnostic purposes where an "acceptance criteria" for algorithm performance would be relevant in the way outlined in the prompt. Instead, it describes a dental restorative material and its performance testing as compared to a predicate device.
Therefore, the prompt asks for information that is not present or applicable to the provided document. The document focuses on:
- Device Description: A light-cured, radiopaque, resin-based nanohybrid restorative composite.
- Intended Use/Indications for Use: Direct placement in all cavity classes in anterior and posterior teeth, repair of enamel defects, provisionals, porcelain restorations, minor occlusal build-ups, core build-ups, and incisal abrasions.
- Technological Characteristics: Comparison of composition, curing mechanism, storage, etc., with a predicate device.
- Non-Clinical Performance Data: Mechanical strength (flexural strength, compressive strength), water sorption and solubility, depth of cure, light sensitivity, radiopacity, polymerization shrinkage, volumetric wear, and color stability, as well as biocompatibility testing.
Since this 510(k) is for a dental restorative material and not an AI-powered diagnostic device, there are no "acceptance criteria" related to AI performance metrics (like sensitivity, specificity, AUC), no "test set" of images, no "experts" establishing ground truth for AI, no "adjudication methods" for AI interpretations, no "MRMC studies," no "standalone algorithm performance," and no "training set" in the context of machine learning.
The "acceptance criteria" mentioned in the document relate to the physical and chemical properties of the composite material, tested against established ISO standards for dental materials, and comparison to the predicate device's performance.
To answer the prompt directly based on the provided text, it's necessary to state that the requested information (related to AI model acceptance criteria and study design) is not relevant to this specific premarket notification.
Ask a specific question about this device
(175 days)
Multi-Lancet Device 2, ReliOn Premier Lancing Device
The Multi-Lancet Device 2 and ReliOn Premier Lancing Device are reusable lancing devices intended to be used with sterile, single-use compatible lancet blades to obtain a capillary blood sample from the fingertip or alternate sites for blood glucose testing or other tests that require small amounts of blood. The Multi-Lancet Device 2 and ReliOn Premier Lancing Device are intended for single person use only.
The Multi-Lancet Device 2 and ReliOn Premier Lancing Device are used with a compatible single-use lancet to puncture the skin and release a small amount of blood. The products are distributed individually or packaged together with blood glucose monitoring systems to obtain capillary whole blood samples required for testing blood glucose levels.
The provided text is an FDA 510(k) premarket notification approval letter for a Multi-Lancet Device 2 and ReliOn Premier Lancing Device. It focuses on demonstrating substantial equivalence to a predicate device (Accu-Chek Softclix Blood Lancing System) through comparisons of intended use, technological characteristics, and non-clinical performance testing.
Crucially, this document focuses on a medical device (lancing device) and not an AI/ML-driven medical device. Therefore, a significant portion of the requested information, such as "number of experts used to establish ground truth," "adjudication method," "MRMC comparative effectiveness study," "standalone algorithm performance," "training set size," and "how ground truth for training set was established," are not applicable to the information provided.
The document primarily describes the physical device and its mechanical and safety performance, not the performance of an AI algorithm.
Here's a breakdown of the available information based on your request, noting where information is not applicable (N/A) due to the nature of the device:
1. A table of acceptance criteria and the reported device performance
The document mentions "Non-Clinical bench testing was performed per the special controls (878.4850). Cleaning/Disinfecting and Mechanical Robustness verification testing was completed to ensure the safety and usability for the duration of the claimed service life." It also states, "Performance testing on the proposed Multi-Lancet Device 2 and ReliOn Premier Lancing Device demonstrate that the devices meet the performance requirements for their intended use."
However, specific numerical acceptance criteria (e.g., maximum allowed force for puncture, minimum number of uses before failure) and the reported device performance against these criteria are not detailed in this public FDA letter. The letter only confirms that such testing was done and met the requirements.
Table of Acceptance Criteria and Reported Device Performance (Based on provided text)
Category | Acceptance Criteria (Stated or Implied) | Reported Device Performance |
---|---|---|
Safety & Usability | Ensure safety and usability for the duration of the claimed service life, addressing all identified risks. | "Non-Clinical bench testing was performed per the special controls (878.4850). Cleaning/Disinfecting and Mechanical Robustness verification testing was completed to ensure the safety and usability for the duration of the claimed service life. The risk analysis contained in the Risk Management Report confirms that all identified risks were addressed and mitigated appropriately. All residual risks after mitigation were acceptable and communicated in the instructions for use." |
Performance | Meet performance requirements for their intended use. | "Performance testing on the proposed Multi-Lancet Device 2 and ReliOn Premier Lancing Device demonstrate that the devices meet the performance requirements for their intended use." |
Labeling | Adequately communicate intended use, safety precautions, and directions for use, meeting 21 CFR § 801 and 21 CFR 878.4850 (c)(vi). | "Labeling adequately communicates to the user the device intended use, safety precautions and directions for use. The labeling meets the requirements documented in the: 1. Regulation for Medical Device Labeling (21 CFR § 801) 2. Regulation for multiple use blood lancet for single patient use only (21 CFR 878.4850 (c)(vi))" |
Substantial Equivalence | Demonstrate substantial equivalence to predicate device based on classifications, intended use, and technological characteristics. | "The Multi-Lancet Device 2 and ReliOn Premier Lancing Device are substantially equivalent to the predicate Accu-Chek Softclix Lancing Device based on comparisons of the device classifications, intended use, and technological characteristics. Non-clinical performance testing successfully confirmed the suitability of the Multi-Lancet Device 2 and ReliOn Premier Lancing Device for the intended uses and demonstrated the devices are as safe, as effective, and perform as well as the predicate device as required per 21 CFR § 807.92(b)(3)." |
2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)
The document mentions "Non-Clinical bench testing." This typically involves laboratory testing of the physical device rather than a "test set" of clinical data (like medical images). The sample size for these bench tests is not specified, nor is the "data provenance" as it's not a clinical data study. The applicant's address is ARKRAY Factory, Inc., 1480 Koji, Konan-cho, Koka-shi, Shiga 520-3306 Japan, suggesting the company is based in Japan, but this doesn't explicitly state the location of the bench testing.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)
Not applicable. This device is a mechanical lancing device, not an AI/ML-driven diagnostic tool that requires ground truth established by experts interpreting data (e.g., radiologists interpreting images). The "ground truth" for this device relates to its mechanical performance and safety, verified through specified testing protocols.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
Not applicable. There is no "test set" of patient data that would require an adjudication method.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
Not applicable. This is not an AI/ML device, and no MRMC study would be conducted. The "assistance" provided by this device is mechanical (blood sampling), not interpretative or cognitive.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
Not applicable. This is a mechanical device, not an algorithm.
7. The type of ground truth used (expert concensus, pathology, outcomes data, etc)
The "ground truth" for this device's performance, as implied, is established by bench test standards and regulatory requirements (e.g., 21 CFR 878.4850 special controls) regarding mechanical robustness, cleaning/disinfection effectiveness, and general safety. It's not a "truth" derived from clinical data like pathology or expert consensus.
8. The sample size for the training set
Not applicable. As this is not an AI/ML device, there is no "training set."
9. How the ground truth for the training set was established
Not applicable. As this is not an AI/ML device, there is no "training set" or corresponding ground truth establishment method.
Ask a specific question about this device
(210 days)
Versana Premier
The Versana Premier is a general purpose diagnostic ultrasound system intended for use by qualified and trained healthcare professionals for ultrasound imaging, measurement, display and analysis of the human body and fluid. Versana Premier clinical applications include : Fetal/Obstetrics, Abdominal, Gynecology, Urology, Pediatric, Small Parts (includes breast, testes, thyroid), Cardiac Pediatric, VascularPeripheral Vascular, Musculoskeletal Conventional, Musculosketal Superficial, Thoracic/Pleural, Transrectal, Transvaginal, Interventional guidance (includes tissue biopsy, fluid drainage, vascular access). Modes of operation include : B. M. PW Doppler, CW Doppler, Color M Doppler, Power Doppler, Harmonic Imaging, Coded Pulse, 3D/4D Imaging mode and Combined modes: B/M, B/Color, B/PWD, B/Color/PWD, B/Power/PWD. Versana Premier is intended to be used in a hospital or medical clinic.
The Versana Premier is a general purpose, Track 3, diagnostic ultrasound system for use by qualified and trained healthcare professionals. The system is a mobile console that includes an operator control panel, display monitor and transducers. The console provides digital acquisition, processing and display capability. The system has an internal battery to allow for acquisition while the system is not plugged into a power source. The operator control panel includes function keys, trackball, an alfa-numeric keyboard and a touch panel as input sources of the device. The variety of transducers include convex, linear, sector, dual and mechanical 4D transducers. The access types include trans- body surface, transrectal, transvaginal and transcranial. Data can be imported or exported by DVD, USB, LAN or WiFi if the USB wireless adapter is connected to the system. An external ECG module has been verified to use as input for gating during scanning. The system has a HDMI port, VGA connection port, S-Video port, and a Composite Out port connection. The system has an external AC outlet to allow connection of a printer box of the console and an option for external Printer USB Isolator for other printers to connect. The system supports one way, Bluetooth communication capability from the system to a personal device to allow for sharing of the patient's data/images when the external Bluetooth USB adapter is connected to the system.
The provided text is a 510(k) Premarket Notification Submission for the GE Versana Premier ultrasound system. This document focuses on demonstrating substantial equivalence to existing predicate devices, rather than proving that a new AI-powered diagnostic device meets specific performance acceptance criteria through a dedicated clinical study.
Therefore, the information required to populate the fields related to acceptance criteria, ground truth, sample size, expert adjudication, and comparative effectiveness studies is not present in this document. The document explicitly states: "The subject of this premarket submission, Versana Premier, did not require clinical studies to support substantial equivalence."
The only mention of an "Artificial Intelligence (AI) feature" is "Whizz Label," and no performance data or study details are provided for this feature. The document indicates that the Versana Premier employs "the same fundamental scientific technology as its predicate device and reference devices," and that it is "substantially equivalent ... with regard to intended use, imaging capabilities, technological characteristics, imaging modes, hardware, and safety effectiveness."
Here's what can be extracted from the document regarding the AI feature, acknowledging the absence of the requested study details:
- AI Feature Mentioned: Whizz Label is an Artificial Intelligence (AI) feature that is being added.
All other requested information cannot be found in the provided document:
- A table of acceptance criteria and the reported device performance: Not provided.
- Sample sized used for the test set and the data provenance: Not provided.
- Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not provided.
- Adjudication method (e.g. 2+1, 3+1, none) for the test set: Not provided.
- If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Explicitly stated that clinical studies were not required.
- If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not provided.
- The type of ground truth used (expert concensus, pathology, outcomes data, etc): Not provided.
- The sample size for the training set: Not provided.
- How the ground truth for the training set was established: Not provided.
Ask a specific question about this device
(235 days)
GEM Premier 5000
The GEM Premier 5000 is a portable critical care system for use by health care professionals to rapidly analyze heparinized whole blood samples at the point of health care delivery in a clinical setting and in a central laboratory. The instrument provides quantitative measurements of pH, pCO2, pO2, sodium, chloride, ionized calcium, glucose, lactate, hematocrit, total bilirubin and CO-Oximetry (tHb, O2Hb, COHb, MHb, sO2*) parameters from arterial, venous or capillary heparinized whole blood. These parameters, along with derived parameters, aid in the diagnosis of a patient's acid/base status, electrolyte and metabolite balance and oxygen delivery capacity.
*sO2 = ratio between the concentration of oxyhemoglobin plus deoxyhemoglobin plus deoxyhemoglobin.
· pH, pCO2, and pO2 measurements in whole blood are used in the diagnosis and treatment of life-threatening acid-base disturbances.
· Electrolytes in the human body have multiple roles. Nearly all metabolic processes depend on or vary with electrolytes:
· Sodium (Na+) measurements are used in the diagnosis and treatment of aldosteronism, diabetes insipidus, adrenal hypertension, Addison's disease, dehydration, inappropriate antidiuretic secretion, or other diseases involving electrolyte imbalance.
· Potassium (K+) measurements are used to monitor electrolyte balance in the diagnosis and treatment of disease conditions characterized by low or high blood potassium levels.
· Ionized calcium (Ca++) measurements are used in the diagnosis and treatment of parathyroid disease, a variety of bone diseases, chronic renal disease and tetany.
· Chloride (Cl-) measurements are used in the diagnosis and treatment of electrolyte and metabolic disorders, such as cystic fibrosis and diabetic acidosis.
· Hematocrit (Hct) measurements in whole blood of the packed red cell volume of a blood sample are used to distinguish normal from abnormal states, such as anemia and erythrocytosis (an increase in the number of red cells).
· Glucose (Glu) measurement is used in the diagnosis, monitoring and treatment of carbohydrate metabolism disturbances including diabetes mellitus, neonatal hypoglycemia, idiopathic hypoglycemia, and pancreatic islet cell carcinoma.
- · Lactate (Lac) measurement is used:
- · to evaluate the acid-base status of patients suspected of having lactic acidosis;
- · to monitor tissue hypoxia and strenuous physical exertion;
- in the diagnosis of hyperlactatemia.
· Total Bilirubin (tBili) measurement is used to aid in assessing the risk of kernicterus and hyperbilirubinemia in neonates.
· CO-Oximetry (tHb, COHb, MetHb, O2Hb. HHb, and sO2) evaluates the ability of the blood to carry oxygen by measuring total hemoglobin and determining the percentage of functional hemoglobin species.
• Total Hemoglobin (tHb): Total hemoglobin measurements are used to measure the hemoglobin content of whole blood for the detection of anemia.
· COHb: Carboxyhemoglobin measurements are used to determine the carboxyhemoglobin content of human blood as an aid in the diagnosis of carbon monoxide poisoning.
3
· MetHb: Methemoglobin measurements are used to determine different conditions of methemoglobinemia.
· HHb: Deoxyhemoglobin, as a fraction of total hemoglobin, is used in combination with oxyhemoglobin to measure oxygen status.
· O2Hb: Oxyhemoglobin, as a fraction of total hemoglobin, is used in combination with deoxyhemoglobin to measure oxygen status.
• sO2: Oxygen saturation, more specifically the ratio between the concentration of oxyhemoglobin and oxyhemoglobin plus deoxyhemoglobin, is used to measure oxygen status.
The GEM Premier 5000 system provides fast, accurate, quantitative measurements of heparinized whole blood pH, pCO2, pO2, Na+, K+, Cl-, Ca++, glucose, lactate, Hct, total bilirubin and CO-Oximetry (tHb, O2Hb, COHb, MetHb, HHb, sO2) from arterial, venous or capillary samples.
The provided text is a 510(k) summary for the GEM Premier 5000 device, detailing an operating system upgrade. This document is a regulatory submission for a device change and does not contain the information requested regarding acceptance criteria, device performance tables, study specifics (sample size, data provenance, expert qualifications, adjudication methods, MRMC studies, standalone performance), or ground truth establishment.
The submission is a Special 510(k), which indicates a modification to an already cleared device, not a de novo clearance requiring extensive clinical performance studies. The core of this submission is a software update (operating system change from Fedora 17 Linux to WindRiver LTS 18 Linux) with the stated reason to "accommodate long-term support of resolutions for common vulnerability exposures."
The document explicitly states:
- "Performance data is limited to Software Verification as the scope of this Special 510(k) is specific to an operating system upgrade..."
- "The changes in this submission do not introduce: Changes to indications for use or intended use, Changes to the fundamental scientific technology, Changes to operating principle, Changes to labeled performance claims."
Therefore, the requested information, which typically pertains to the establishment of initial clinical performance and effectiveness, is not present in this regulatory document for this specific submission. The focus here is on ensuring the device continues to meet its previously established performance claims after a technical software upgrade, rather than demonstrating new performance capabilities.
Ask a specific question about this device
Page 1 of 8