K Number
K162257
Manufacturer
Date Cleared
2016-11-07

(88 days)

Product Code
Regulation Number
872.3690
Panel
Dental
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

P1145, a nanohybrid dental restorative, is indicated for direct placement in all cavity classes in anterior and posterior teeth. Additional indications include: repair of enamel defects, repair of provisionals, repair of porcelain restorations, minor occlusal build-ups, core build-ups and incisal abrasions.

Device Description

P1145 is a light-cured, esthetic, resin-based nanohybrid dental restorative designed for direct placement in both anterior and posterior restorations. It is designed to be used after placement of a dental adhesive. Once placed, the composite can be easily sculpted to the desired shape, and after curing, can easily be finished and polished to achieve a highly esthetic restoration.

AI/ML Overview

The provided text is a 510(k) premarket notification for a dental restorative device (P1145 Dental Restorative). It describes the device, its intended use, and argues for its substantial equivalence to a predicate device (Herculite Ultra).

However, the document does not describe acceptance criteria for an AI/algorithm-based device, nor does it detail a study proving such a device meets acceptance criteria.

The "Non-Clinical Performance Data" section refers to tests for a conventional dental material (mechanical strength, polishability, water sorption, etc.) against established standards like ISO 4049:2009. These are physical and chemical property tests for the restorative material itself, not performance metrics for an AI or algorithm.

Therefore, I cannot extract the information required for the prompt as it pertains to an AI/algorithm-based device and its performance study. The document is about a physical dental material, not a digital diagnostic or prognostic tool.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized symbol of three human profiles facing right, stacked on top of each other, with flowing lines beneath them.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

November 7, 2016

Kerr Corporation % Mohammad Ansari Regulatory Affairs Specialist Sybron Dental Specialties 1717 W. Collins Ave Orange, California 92867

Re: K162257

Trade/Device Name: P1145 Dental Restorative Regulation Number: 21 CFR 872.3690 Regulation Name: Tooth Shade Resin Material Regulatory Class: Class II Product Code: EBF Dated: August 5, 2016 Received: August 11, 2016

Dear Mohammad Ansari:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in

{1}------------------------------------------------

the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely,

Susan Russo, DDS, MA

Tina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) (Not yet assigned)

Device Name

P1145 Dental Restorative

Indications for Use (Describe)

P 1145, a nanohybrid dental restorative, is indicated for direct placement in all cavity classes in anterior and posterior teeth. Additional indications include: repair of enamel defects, repair of provisionals, repair of porcelain restorations, minor occlusal build-ups, core build-ups and incisal abrasions.

Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for Sybron Dental Specialties. The logo consists of the letters "sds" in a stylized, bold, blue font. Above the letters is a curved blue line. Below the letters is the text "Sybron Dental Specialties" in a smaller, less bold, blue font.

510(k) SUMMARY for P1145 Dental Restorative

In accordance with Title 21 of the Code of Federal Regulations, Part 807, and in particular 21 CFR §807.92, the following summary of information is provided:

    1. Submitter Information:
      Sybron Dental Specialties 1717 W. Collins Ave. Orange CA, 92867
Contact Person:Wendy Garman
Telephone Number:909-962-5666
Fax Number:909-962-5694

Date Prepared: November 7, 2016

    1. Device Name:
  • Proprietary Name: P1145 Dental Restorative ●

    • Classification Name: Tooth Shade Resin Material
  • CFR Number: .

    • 872.3690 2
  • Device Class: . Product Code: EBF .

    1. Predicate Device:

The P1145 product line is substantially equivalent to the legally marketed device Herculite Ultra (K082671) cleared on December 11, 2008, product code EBF.

4. Description of Device:

P1145 is a light-cured, esthetic, resin-based nanohybrid dental restorative designed for direct placement in both anterior and posterior restorations. It is designed to be used after placement of a dental adhesive. Once placed, the composite can be easily sculpted to the desired shape, and after curing, can easily be finished and polished to achieve a highly esthetic restoration.

  • న. Indications for Use:
    P1145, a nanohybrid dental restorative, is indicated for direct placement in all cavity classes in anterior and posterior teeth. Additional indications include: repair of enamel defects, repair of provisionals, repair of porcelain restorations, minor occlusal build-ups, core build-ups and incisal abrasions.

{4}------------------------------------------------

Description of Safety and Substantial Equivalence: 6.

Technological Characteristics

The technological characteristics of P1145 are virtually identical to those of the predicate device, Herculite Ultra (K082671). Both formulas mostly contain solid filler-particles, mixed with a resin system, in order to form a high viscosity paste which can be directly applied into a restoration preparation.

The filler systems in predicate Herculite Ultra (K082671) and P1145 are formulated in order to provide the composites its mechanical strength and durability, while maintaining excellent handling and esthetic characteristics. Additionally, the composites are formulated with methacrylate-based resin systems in combination with a photo-initiation system, which results in a polymerization of the composite when irradiated using a dental curing light.

Non-Clinical Performance Data

Non-clinical performance data included testing for mechanical strength (flexural strength, compressive strength, diametral tensile strength, Rockwell hardness), polishability, water sorption and solubility, depth of cure, light sensitivity, radiopacity, volumetric shrinkage, and color stability. Working time and gloss were also performed. The data analyzed from the various tests substantiate that P1145 is safe and effective as the predicate Herculite Ultra (K082671).

The following standards were utilized for non-clinical performance testing of P1145:

  • . Guidance for Industry and FDA Staff: Dental Composite Resin Devices - Premarket Notification [510(k)] Submissions, November 27, 1998
  • ISO 10993-1: 2009 Biological evaluation of medical devices
  • . ISO 10993-3:2003 Biological Evaluation of Medical Devices- Part 3: Tests for Genotoxicity, Carcinogenicity, and Reproductive Toxicity
  • . ISO 10993-5:2009 Biological Evaluation of Medical Devices- Part 5: Tests for in Vitro Cytotoxicity
  • . ISO 10993-6:2007 Biological Evaluation of Medical Devices- Part 6: Tests for Local Effects after Implantation
  • . ISO 10993-10:2010 Biological Evaluation of Medical Devices- Part 10: Tests for Irritation and Skin Sensitization
  • . ISO 10993-11:2011 Biological Evaluation of Medical Devices- Part 11: Tests for Systemic Toxicity
  • . ISO 14971:2007 Risk Management
  • . ISO 4049:2009 Dentistry- Polymer based Filling, Restorative, and Luting Materials.

{5}------------------------------------------------

ElementPredicateProposed
Herculite UltraP1145
510(k)K121824To be assigned
Trade NameHerculite UltraP1145
Target UsersLicensed dental professionalsLicensed dental professionals
Device DescriptionHerculite Ultra is a light-cured, esthetic, resin-baseddental restorative.P1145 is a light-cured, esthetic, resin-based nanohybrid dental restorativedesigned for direct placement in bothanterior and posterior restorations. It isdesigned to be used after placement ofa dental adhesive. Once placed, thecomposite can be easily sculpted to thedesired shape, and after curing, caneasily be finished and polished toachieve a highly esthetic restoration.
Intended UseDental RestorativeDental Restorative
Common NameDental Composite RestorativeMaterialDental Composite Restorative Material
Classification NameTooth Shade Resin MaterialTooth Shade Resin Material
Class22
Product CodeEBFEBF
StorageStore at room temperature.Store at room temperature.
Indications for UseHerculite Ultra is a dentalcomposite restorative materialintended to be used in allclasses of cavities.P1145, a nanohybrid dental restorative,is indicated for direct placement in allcavity classes in anterior and posteriorteeth. Additional indications include:repair of enamel defects, repair ofprovisionals, repair of porcelainrestorations, minor occlusal build-ups,core build-ups and incisal abrasions.
Curing MechanismPhoto-initiationPhoto-initiation
MaterialCompatibilityMeets BiocompatibilityRequirementsMeets Biocompatibility Requirements
Shelf LifeMinimum 36 months based onreal time dataMinimum 24 months based onaccelerated data
Particle SizeDistributionNanohybrid:Proprietary formulation ofnano and non-nano particlesizes of fillerNanohybrid:Same

Table 5.1: Predicate and Proposed Device Comparison Table

{6}------------------------------------------------

Clinical Performance Data

Clinical performance testing has not been performed for P1145.

Conclusion as to Substantial Equivalence

The technological characteristics of P1145 are virtually identical to the predicate Herculite Ultra (K082671). These proposed P1145 has similarities in select performance characteristics as compared to the predicate. The proposed P1145 is substantially equivalent to the predicate Herculite Ultra (K082671) based on the similarities in the two products in design, device performance, biocompatibility testing, and the same intended use. The proposed product P1145 and Herculite Ultra (K082671) have the same intended use and are to be used in dental restorative procedures. The proposed device P1145 has more specific indications, as compared to Herculite Ultra (K082671), to clearly define and describe the device applications and types of restorative procedures as well as describe the nanohybrid filler composition. Any noted differences in technological characteristics between the proposed and predicate products do not raise new questions of safety and effectiveness and demonstrate the proposed product is at least as safe and effective as the legally marketed predicate device. Both Herculite Ultra (K082671) and P1145 utilize a nanohybrid filler system and a similar methacrylate-based resin system to achieve the same function. Herculite Ultra (K082671) and P1145. Based on this reasoning and the results of ISO 4049:2009 performance testing, including mechanical strength, depth of cure, shrinkage, polishability, color stability, and biocompatibility testing, P1145 is substantially equivalent to the predicate Herculite Ultra (K082671).

§ 872.3690 Tooth shade resin material.

(a)
Identification. Tooth shade resin material is a device composed of materials such as bisphenol-A glycidyl methacrylate (Bis-GMA) intended to restore carious lesions or structural defects in teeth.(b)
Classification. Class II.