Search Results
Found 64 results
510(k) Data Aggregation
(297 days)
EBC
The Parkell Pit and Fissure Sealant is indicated for prophylactic sealing of pits and fissures as well as for micro-restorative procedures for composite restorations.
The Device is a light-cured, flowable, resin-based sealant for prophylactic sealing of pits and fissures as well as for micro-restorative procedures. The Device is light-curable under standard power (600 mW/cm²) and contains filling agents which cause the Device to release and recharge beneficial ions (fluoride and calcium).
The provided text describes the 510(k) clearance for the "Parkell Pit and Fissure Sealant." However, the clearance letter and accompanying 510(k) summary pertain to a physical dental device (a sealant), not an AI/software medical device.
Therefore, many of the requested criteria regarding AI/software performance studies (such as sample size for test/training sets, data provenance, expert ground truth, adjudication methods, MRMC studies, standalone performance, etc.) are not applicable and are not mentioned in this document.
The document focuses on the physical and chemical properties of the sealant and compares them to a predicate device. The performance data is based on non-clinical, in-vitro testing of these physical properties.
Below is the information that can be extracted from the provided text, adapted as best as possible to the requested format, with clear indications where information is not applicable (N/A) due to the nature of the device.
Acceptance Criteria and Device Performance for Parkell Pit and Fissure Sealant
This device is a physical dental sealant, not an AI/software medical device. Therefore, the performance evaluation focuses on physical and chemical properties and biocompatibility, not algorithmic performance on image data.
1. Table of Acceptance Criteria and Reported Device Performance
Property | Acceptance Criteria (Pass/Fail) | Reported Device Performance (Parkell Pit and Fissure Sealant) |
---|---|---|
Bond Strength to Dry Enamel | ≥10 MPa | ≥10 MPa |
Compressive Strength | ≥200 MPa | ≥200 MPa |
Flexural Strength | ≥90 MPa | ≥90 MPa |
Depth of Cure | 1.5-2.5 mm | 1.5-2.5 mm |
Shrinkage | ≤8% | ≤8% |
Viscosity (initial) | 1-5 Pa.s | 1-5 Pa.s |
Shelf Life | Not explicitly stated | 2 years (established by testing) |
Material Biocompatibility | Substantially equivalent to predicate device | Substantially equivalent to predicate device |
Beneficial Ion Release | Fluoride & Calcium | Fluoride & Calcium (vs. Predicate's Fluoride, Calcium, Phosphate) |
2. Sample size used for the test set and the data provenance
- Sample Size: Not specified in the document. The testing was non-clinical (in-vitro).
- Data Provenance: Non-clinical physical property testing. Not applicable for geographical or retrospective/prospective distinctions.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
- N/A. This is a physical dental material. "Ground truth" in this context typically refers to objective measurements from laboratory testing equipment (e.g., universal testing machines for strength, viscometers for viscosity, etc.), not expert interpretation of data like in AI/imaging studies.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
- N/A. As above, this is laboratory-based physical property testing, not subjective assessment requiring adjudication.
5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
- N/A. This is not an AI-assisted device. No MRMC study was performed.
6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done
- N/A. This is not an algorithm. Standalone performance is not applicable.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)
- Laboratory Standard Measurements / In-vitro physical property testing. The performance criteria (e.g., strength in MPa, shrinkage in %, viscosity in Pa.s) are based on standardized tests for dental materials.
8. The sample size for the training set
- N/A. This is not an AI/machine learning device. No training set was used.
9. How the ground truth for the training set was established
- N/A. As above, no training set was used.
Ask a specific question about this device
(1 days)
EBC
UltraSeal XT plus bioprotection by Nobio sealant - UltraSeal XT plus sealant for prophylactic sealing of pits and fissures and can be used where the use of a flowable composite is indicated. The bioprotection by Nobio reduces demineralization, which is part of the caries formation process.
UltraSeal XT hydro bioprotection by Nobio sealant - Use UltraSeal XT hydro sealant for prophylactic sealing of pits and fissures. The addition of the bioprotection by Nobio reduces demineralization, which is part of the caries formation process.
UltraSeal XT plus bioprotection by Nobio sealant and flowable composite – UltraSeal XT plus sealant is a 58% filled, light-cured, radiopaque, methacrylate-based, thixotropic resin sealant. It is used with PrimaDry drying and priming agent to aid in avoiding moisture contamination which can cause microleakage and poor retention. It is a Type 1 Class 2 Group 1 polymer-based dental restorative flowable resin which is suitable for occlusal surfaces. It can be used where the use of a flowable is indicated. It is compatible with adhesives and bonding agents. UltraSeal XT plus sealant has a radiopacity value of 1.44 mm of aluminum. Aluminum has a radiopacity equivalent to that of dentin. Thus, 1 mm of material (having a radiopacity equivalent to 1 mm of aluminum) has a radiopacity equivalent to that of dentin-and 2 mm of aluminum is equivalent to enamel. The addition of the bioprotection by Nobio reduces demineralization, which is part of the caries formation process.
UltraSeal XT hydro bioprotection by Nobio sealant – UltraSeal XT hydro sealant is a Class 2, 53% filled, light-cured, radiopaque, methacrylate-based, thixotropic resin sealant. It is partly hydrophobic and hydrophilic before it is cured, hydrophobic once cured, and has a selfadhesive quality. UltraSeal XT hydro sealant chemistry provides the option of a visual verification for marginal retention-with the use of a black light upon placement and at recall visits. The UV light is not included in the UltraSeal XT hydro sealant kit. The addition of the bioprotection by Nobio reduces demineralization, which is part of the caries formation process.
Some of the silica-based filler particles (max 1.2% wt/wt of the total formulation) contained in all the UltraSeal XT plus Bioprotection by Nobio and UltraSeal XT hydro Bioprotection by Nobio are decorated with quaternary ammonium functional groups that are covalently bound to the filler particles' silica core via a silane linker (herein referred to as "QASi"). Following light induced polymerization, QASi particles (produced by Nobio) remain permanently entrapped within the sealant and destroy microorganisms by contact. As the cell membrane of microorganisms is negatively charged and the QASi particles have strong positive charges, microorganisms are electrostatically attracted to the positively charged cured material's surface resulting in disruption of their cell membrane and immediate microbial lysis.
Both subject devices have a depth of cure ≥ 1.5mm, are curable at wavelengths of 385-515nm and will remain uncured after light exposure while in their primary container. UltraSeal XT plus Bioprotection by Nobio is also indicated for use as a flowable composite. It has a flexural strength of ≥ 80 MPa, Water Sortpion ≤ 40 µ g/mm3, Water Solubility ≤ 7.5 µ g/mm3, Radio-Opacity, ≥ 1mm of Al, Surface Hardness ≥ 20 HK and Ambient light sensistivity resulting in the product being physically homogeneous after 60 seconds.
The provided text focuses on the substantial equivalence determination for dental sealants and does not contain information about a medical device involving AI or machine learning. Therefore, I cannot extract acceptance criteria and study details related to AI/ML device performance from this document.
The document discusses physical and chemical properties of dental sealants and their biocompatibility, which are typically evaluated through bench testing and ISO standards, not through clinical studies with human readers or AI algorithms. As such, information regarding sample sizes for test sets, data provenance, expert adjudication, MRMC studies, standalone performance, training sets, or ground truth establishment in the context of AI/ML is absent.
Ask a specific question about this device
(154 days)
EBC
Proseal Enamel Coating Resin is a light curing material for sealing the enamel pits and fissures of teeth.
Proseal Enamel Coating Resin is a light-curing fissure sealant and is mainly composed of methacrylate resin, photo initiator system and inorganic fillers. It is packaged in syringes or bottles.
This document is a 510(k) Summary for the Proseal Enamel Coating Resin. It focuses on demonstrating substantial equivalence to a predicate device rather than presenting a study proving a device meets specific acceptance criteria for performance in a diagnostic or clinical setting.
Therefore, many of the requested fields cannot be filled from the provided text as they relate to the performance of an AI/diagnostic device, which is not what this document describes.
Here's an analysis based on the available information:
1. A table of acceptance criteria and the reported device performance
The document states that in-vitro bench tests were performed and that the "test results indicated that the Proseal Enamel Coating Resin meets the pass/fail criteria and supports substantial equivalence when compared to the predicate device on physical properties." It also notes that the device is classified as Class 2 sealant in accordance with ISO 6874:2015.
The specific acceptance criteria (pass/fail criteria) are implied to be derived from ISO 6874:2015 and FDA guidance "Dental Composite Resin Devices - Premarket Notification [510(k)] Submissions Document issued on: October 26, 2005". However, the document does not list the specific numerical acceptance criteria.
The reported device performance is only described qualitatively as "meets the pass/fail criteria." Specific numeric performance values for each test (depth of cure, flexural strength, shear bonding strength, compressive strength, water sorption, solubility, film thickness, and elastic modulus) are not provided in this summary.
Acceptance Criteria (Derived from ISO 6874:2015 & FDA Guidance) | Reported Device Performance |
---|---|
(Specific numerical criteria not detailed in the document) | Meets criteria |
Depth of Cure | Meets criteria |
Flexural Strength | Meets criteria |
Shear Bonding Strength | Meets criteria |
Compressive Strength | Meets criteria |
Water Sorption | Meets criteria |
Solubility | Meets criteria |
Film Thickness | Meets criteria |
Elastic Modulus | Meets criteria |
2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)
The document mentions "in-vitro bench tests" for physical properties and "Biocompatibility tests" but does not specify the sample size for these tests. Data provenance (country of origin, retrospective/prospective) is not mentioned for the testing.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)
Not applicable. This is a material science study, not a study evaluating human expert performance or diagnostic accuracy.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
Not applicable. This is a material science study.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
Not applicable. This is a material science product, not an AI or diagnostic device that would involve human readers.
6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done
Not applicable. This is a material science product.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)
For the physical property tests, the "ground truth" would be the scientifically established and standardized measurement protocols and evaluation criteria defined by ISO 6874:2015 and FDA guidance for dental composite resin devices. For biocompatibility tests, the ground truth would be established by ISO 10993 standards.
8. The sample size for the training set
Not applicable. There is no AI algorithm being trained.
9. How the ground truth for the training set was established
Not applicable. There is no AI algorithm being trained.
Ask a specific question about this device
(206 days)
EBC
BA Pit & Fissure Sealant is intended to seal pit and fissure depressions/faults in the biting surfaces of teeth. For dental professional use only.
Pit and Fissure sealants are used as a preventive measure to fill defects and prevent decay within crevices, voids or fissures on the surface of teeth. These voids are more challenging to keep free of substances that can lead to caries and eventual breakdown of tooth structure. BA Pit & Fissure Sealant is a material designed to fill these voids with a wear resistant material, ultimately preserving natural tooth structure.
BA Pit & Fissure Sealant is a light activated / photopolymerizable, glass filled, urethane dimethacrylate material that is placed onto an etched enamel surface and light activated for 20 seconds. The result after application is a tooth that has been sealed with a tough, wear resistant material.
Here's a breakdown of the acceptance criteria and the study that proves the device meets them, based on the provided text:
1. A table of acceptance criteria and the reported device performance
The document does not explicitly state "acceptance criteria" with numerical targets for each test. Instead, it presents the results of various non-clinical performance tests for the subject device (BA Pit & Fissure Sealant) and then compares them to a predicate device (Pulpdent (Activa) Pit & Fissure Sealant with MCP) and sometimes a reference device (Prevent Seal) to establish substantial equivalence. The "acceptance criterion" in this context is generally that the subject device's performance is "similar" or "equivalent" to the predicate device, or that any differences do not raise additional concerns regarding safety or efficacy.
Test | Acceptance Criteria (implied - "similar/equivalent to predicate/reference" or "no additional concerns") | Reported Device Performance (BA Pit & Fissure Sealant) |
---|---|---|
Compressive Strength | Similar to predicate and reference devices. | Exhibited very similar compressive strengths to the predicate device and reference device. |
Diametral Tensile Strength | Similar to predicate device. | Exhibited similar diametral tensile strength compared to the predicate device. |
Elastic Modulus | Level of flexibility that allows absorption of occlusal forces without damage, supporting efficacious use. | Material is flexible and will absorb significant forces without reaching permanent deformation; deformation limited by tooth structure allows absorption of occlusal forces without damage. |
Working Time | Equivalent to predicate device. | Unlimited working time, equivalent to the predicate device. |
Depth of Cure | Very similar to predicate device; minor differences acceptable for occlusal surface/thin layers and do not impact safety/efficacy. | Exhibited a very similar depth of cure to the predicate device. Minor difference not a concern for pit and fissure sealants. |
Film Thickness | Nearly equivalent to reference device. | Nearly equivalent to the reference device. |
Cytotoxicity | Same result as predicate device. | Provided the same result as the predicate device. |
Accelerated Shelf-Life (interim) | "Passing" results at 18 months, supporting an 18-month shelf-life. | All test points to-date have yielded "passing" results, supporting an 18-month shelf-life (interim analysis). |
Ion Release (Calcium, Phosphate, Fluoride) | Measurable ions released; minor differences to predicate acceptable as oral environment naturally fluctuates, and concentrations are very low. | Released measurable calcium, phosphate, and fluoride. Released more than predicate, but differences are low (ppm/ppb) and oral environment fluctuates, so no concerns. Substantial equivalence supported. |
2. Sample sizes used for the test set and the data provenance
The document describes non-clinical benchtop testing. For these types of tests, sample sizes are typically specified within the referenced ISO or ADA standards. However, the exact sample sizes for each specific test conducted for K190998 are not explicitly mentioned in the provided text.
The data provenance is retrospective, as the tests were conducted prior to the 510(k) submission to demonstrate substantial equivalence. The country of origin of the data is not specified, but given the applicant is Apex Dental Materials, Inc. with an address in Lake Zurich, Illinois, it is likely the studies were conducted in the USA or by contract labs.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
This section is not applicable as the document describes non-clinical benchtop performance testing, not studies involving human interpretation or expert-established ground truth. The "ground truth" for these tests would be the measurement obtained from the laboratory instruments according to the specified standards.
4. Adjudication method (e.g., 2+1, 3+1, none) for the test set
This section is not applicable as the document describes non-clinical benchtop performance testing, not studies involving human reading or adjudication.
5. If a multi-reader, multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
This section is not applicable. The device is "BA Pit & Fissure Sealant," which is a material used for dental sealing, not an AI-powered diagnostic or assistive tool for human readers. No MRMC study or AI component is mentioned.
6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done
This section is not applicable. The device is a physical material (dental sealant), not an algorithm or software device.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)
The "ground truth" for the non-clinical tests is based on measurements obtained following standardized test methods (e.g., ISO, ADA standards). For example, for compressive strength, the ground truth is the force at which the material fractures, as measured by a testing machine according to the specified standard.
8. The sample size for the training set
This section is not applicable. The device is a physical material, not an AI model, and therefore does not have a "training set."
9. How the ground truth for the training set was established
This section is not applicable. The device is a physical material, not an AI model, and therefore does not have a "training set" or ground truth relevant to AI model training.
Ask a specific question about this device
(165 days)
EBC
Helioseal F Plus is used to seal fissures, pits and foramina caeca.
Helioseal® F Plus is a light-curing, white shaded fissure sealant featuring fluoride release. Helioseal F Plus is supplied in either a syringe or cavifil delivery form. The fissure sealant is composed of dimethacrylates. In the handling technique, the dental professional will clean the enamel surface to be sealed, isolate the working field, conditioning is performed, a thorough rinse and dry to remove the conditioner, before the application of Helioseal F Plus to the tooth surface.
This document is a 510(k) premarket notification for a dental device, specifically a pit and fissure sealant. It demonstrates substantial equivalence to a predicate device, rather than providing a performance study proving a device meets specific acceptance criteria as would be typical for an AI/ML medical device.
Therefore, the requested information regarding acceptance criteria, study design for AI/ML performance, sample sizes, expert involvement, ground truth establishment, and MRMC studies is not applicable to this document.
The document primarily focuses on demonstrating equivalence through:
- Comparison to Predicate Device: Detailed comparison of indications for use, contraindications, technology, principles of operation, and physical properties between Helioseal F Plus (subject device) and Helioseal F (predicate device).
- Bench Testing: Non-clinical performance testing for physical properties (flexural strength, curing depth, light intensity, wavelength for curing, water sorption, water solubility, radio-opacity) in accordance with FDA Guidance and EN ISO 6874:2015.
- Biocompatibility Testing: Evaluation for cytotoxicity and genotoxicity according to ISO 10993-1, ISO 7405, and ISO 14971.
The conclusion states that the devices are substantially equivalent, implying that the new device performs similarly to the already legally marketed predicate device. There are no acceptance criteria, in the sense of accuracy/recall/precision metrics for an AI/ML algorithm, or a human reader study, included in this type of submission.
Ask a specific question about this device
(171 days)
EBC
Pulpdent's Activa Pit and Fissure Sealant with MCP is intended to seal pit and fissure depressions/faults in the enamel, in the biting surfaces of teeth. For dental professional use only.
Activa Pit and Fissure Sealant is a light cured, flowable dental sealant material, available in a single barrel syringe. This sealant is designed with MCP (Modified Calcium Phosphate); it releases calcium, phosphate, and fluoride ions into the oral environment. It contains no Bisphenol A, no BisGMA, and no BPA derivatives.
This document is a 510(k) premarket notification for a dental pit and fissure sealant. It does not describe an AI/ML-driven medical device, and therefore the requested information regarding acceptance criteria and study data for an AI/ML device is not present in the provided text.
The document discusses substantial equivalence to a predicate device based on material composition, intended use, and physical properties, not on performance metrics of an AI algorithm.
Therefore, I cannot extract the information required by your prompt from the provided text.
Ask a specific question about this device
(126 days)
EBC
Nanova Sealant is indicated for use as a pit and fissure sealant
Not Found
This document is a 510(k) clearance letter for the Nanova™ Sealant, a pit and fissure sealant. It confirms that the device has been found substantially equivalent to a predicate device.
However, the provided text does NOT contain any information about acceptance criteria, device performance studies, sample sizes, ground truth establishment, expert qualifications, adjudication methods, or MRMC studies.
The document is a regulatory communication from the FDA, and while it confirms the device can be legally marketed, it does not detail the technical performance studies that would typically be included in a 510(k) submission to demonstrate substantial equivalence. Such performance data would be in the original 510(k) submission (K171330), not in the clearance letter itself.
Therefore, I cannot provide the requested information from the given text.
Ask a specific question about this device
(246 days)
EBC
Pit and fissure sealant
HAPI Seal is a light-cured, fluoride containing, resin-based pit & fissure sealant designed for sealing the enamel pits and fissures of teeth.
The provided text is a 510(k) summary for a dental device, HAPI Seal, and does not describe a study involving an AI algorithm or human readers. Therefore, several of the requested categories (2, 3, 4, 5, 6, 7, 8, 9) are not applicable.
Here's the information that can be extracted from the document:
1. Table of acceptance criteria and the reported device performance:
Acceptance Criteria / Requirement | Reported Device Performance |
---|---|
Depth of Cure | HAPI Seal meets the requirements of ISO 6874:2015. |
Flexural Strength | HAPI Seal meets the requirements of ISO 4049:2009 for Flexural Strength. |
Shear Bond Strength (Modified ISO 29022) | HAPI Seal is equivalent to the predicate. |
Light Curability (Cure time) | HAPI Seal is equivalent to the predicate. |
Ambient Light Sensitivity | HAPI Seal is equivalent to the predicate. |
Flexural Modulus (ANSI/ADA Spec No. 27) | HAPI Seal is equivalent to the predicate. |
Cumulative Fluoride Release | HAPI Seal is equivalent to the predicate. |
Compressive Strength (ANSI/ADA Spec No. 27) | HAPI Seal is equivalent to the predicate. |
Surface Hardness (Barcol-Colman 934-1) | HAPI Seal is equivalent to the predicate. |
Shrinkage (Volumetric) | HAPI Seal is equivalent to the predicate. |
Biocompatibility (Cytotoxicity, Oral Toxicity) | HAPI Seal was not toxic in the oral toxicity test (10 mice, 14 days). An evaluation using ISO 7405:2008 and ISO 10993-1 was conducted. |
2. Sample size used for the test set and the data provenance:
- Oral Toxicity Study: 10 mice were used for a 14-day study.
- Other performance tests: The document mentions that HAPI Seal "meets" or is "equivalent" to the predicate for various physical/mechanical properties based on specific ISO or ANSI/ADA standards. However, it does not explicitly state the sample sizes used for these tests. The data provenance is not specified (e.g., country of origin, retrospective or prospective), but these are laboratory tests done to regulatory standards.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
This information is not applicable as the studies described are laboratory performance and biocompatibility tests, not studies requiring expert interpretation of data.
4. Adjudication method (e.g., 2+1, 3+1, none) for the test set:
This information is not applicable.
5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
This information is not applicable.
6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:
This information is not applicable.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):
The ground truth for the physical/mechanical properties and biocompatibility was established through adherence to recognized international and national standards (e.g., ISO 6874:2015, ISO 4049:2009, ISO 29022:2013, ANSI/ADA Spec No. 27, ISO 7405:2008, ISO 10993-1).
8. The sample size for the training set:
This information is not applicable as this is not an AI algorithm study.
9. How the ground truth for the training set was established:
This information is not applicable.
Ask a specific question about this device
(132 days)
EBC
Premier Sealant, a prescription only medical device used for prophylactic sealing of pits and fissures. It may also be used for micro-restorative or "initial layer" of composite restorations.
Not Found
I am sorry, but the provided text from the FDA 510(k) premarket notification for the "Premier Sealant" does not contain any information about a medical device that uses artificial intelligence (AI) or machine learning.
The document is a standard FDA clearance letter for a dental pit and fissure sealant, which is a physical product, not a software-based diagnostic or assistive device that would involve acceptance criteria for AI performance.
Therefore, I cannot provide the requested information regarding:
- A table of acceptance criteria and reported device performance.
- Sample sizes for test set or data provenance.
- Number of experts or their qualifications for ground truth.
- Adjudication methods.
- MRMC comparative effectiveness study or human reader improvement with AI.
- Standalone algorithm performance.
- Type of ground truth used.
- Training set sample size.
- How ground truth for the training set was established.
This information is simply not present in the provided text as the device in question is a traditional dental material, not an AI-powered diagnostic tool.
Ask a specific question about this device
(257 days)
EBC
PacSeal™ Pit & Fissure Sealant is intended for use as:
- Seal the pits and fissures in teeth
PacSeal™ Pit & Fissure Sealant is a fluoride releasing, light-cured acrylate resin designed to fill and seal the pits and fissures of teeth.
Here's a breakdown of the acceptance criteria and study information for the PacSeal™ Pit & Fissure Sealant as derived from the provided document:
This document is a 510(k) summary, which typically focuses on demonstrating substantial equivalence to a predicate device rather than presenting a full clinical study with specific acceptance criteria in the manner of a novel device. Therefore, the "acceptance criteria" are implied by the comparison to the predicate device's performance characteristics.
1. Table of Acceptance Criteria and Reported Device Performance
The document doesn't explicitly state "acceptance criteria" as pass/fail thresholds. Instead, it compares the subject device's performance to that of predicate devices, implicitly suggesting that performance comparable to or better than the predicate is acceptable.
Performance Characteristic | Acceptance Criteria (Implied by Predicate) | Reported Device Performance (PacSeal™ Pit & Fissure Sealant) |
---|---|---|
Depth of Cure (ISO 6874-2005) | Comparable to or better than 1.86 mm (3M™ Clinpro™ Sealant) | Complies (with ISO 6874-2005) |
Compressive Strength | Comparable to or better than 210 MPa (3M™ Clinpro™ Sealant) | 181 MPa |
Diametral Tensile Strength | Comparable to or better than 37 MPa (3M™ Clinpro™ Sealant) | Subject device is equivalent to the predicate device |
Film Thickness | Comparable to or better than 10 µm (3M™ Clinpro™ Sealant) | Subject device is equivalent to the predicate device |
Flexural Strength (ISO 4049-2009) | Comparable to or better than 110 MPa (3M™ Clinpro™ Sealant) | Complies (with ISO 4049-2009) |
Light Curing Time (ISO 6874-2005) | 30 sec (matches 3M™ Clinpro™ Sealant) | 30 sec |
Shear Bond Strength to Etched Enamel | Comparable to or better than 8.12 MPa (3M™ Clinpro™ Sealant) | Subject device is equivalent to the predicate device |
Note: For "Compressive Strength," the subject device (181 MPa) is lower than the predicate (210 MPa). The document still concludes substantial equivalence, suggesting that this difference was deemed acceptable in the context of overall performance and "equivalent to the predicate" claim for other properties.
2. Sample Size Used for the Test Set and Data Provenance
The document does not specify the sample sizes used for the performance tests (e.g., number of specimens for depth of cure, strength tests).
The data provenance is not explicitly stated beyond being presented in a 510(k) summary by Pac-Dent International, Inc. These types of tests are typically in-vitro laboratory tests, not clinical studies with patient data.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts
This type of information is not applicable to the data provided. The "ground truth" for these physical property tests is based on standardized laboratory measurements (e.g., ISO standards) and not on expert consensus from human readers or clinicians.
4. Adjudication Method for the Test Set
This is not applicable as the tests are for physical properties measured in a lab, not for diagnostic or interpretive tasks requiring adjudication.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done
No, an MRMC comparative effectiveness study was not done. This document describes the physical and chemical characteristics of a dental sealant, not an AI or imaging device that would typically undergo MRMC studies.
6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done
No, this is not relevant to the device described. The device is a physical dental sealant, not an algorithm.
7. The Type of Ground Truth Used
The "ground truth" for the performance characteristics (e.g., depth of cure, strength, film thickness) is based on standardized laboratory test methods (e.g., ISO 6874-2005, ISO 4049-2009).
8. The Sample Size for the Training Set
This is not applicable. The device is a physical product, not an algorithm that requires a training set.
9. How the Ground Truth for the Training Set Was Established
This is not applicable for the same reason as above.
Ask a specific question about this device
Page 1 of 7