Search Filters

Search Results

Found 19 results

510(k) Data Aggregation

    K Number
    K211199
    Date Cleared
    2021-11-08

    (200 days)

    Product Code
    Regulation Number
    862.1117
    Reference & Predicate Devices
    Why did this record match?
    Applicant Name (Manufacturer) :

    Tosoh Bioscience, Inc.

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Tosoh ST AIA-PACK BNP assay is designed for IN VITRO DIAGNOSTIC USE ONLY for the quantitative measurement of BNP in human (K2EDTA) plasma on Tosoh AIA System Analyzers. BNP is used as an aid in the diagnosis of heart failure (HF) in patients presenting to the emergency department (ED) with clinical suspicion of new onset HF, acutely decompensated or exacerbated HF.

    Device Description

    The ST AIA-PACK BNP is a two-site immunoenzymometric assay which is performed entirely in the ST AIA-PACK BNP test cups. BNP present in the test sample is bound with monoclonal antibody immobilized on magnetic beads and enzyme-labeled monoclonal antibody. The magnetic beads are washed to remove unbound enzyme-labeled monoclonal antibody and are then incubated with a fluorogenic substrate, 4-methylumbelliferyl phosphate (4MUP). The amount of enzyme-labeled monoclonal antibody that binds to the beads is directly proportional to the BNP concentration in the test sample. A standard curve is constructed, and unknown sample concentrations are calculated using the curve.

    AI/ML Overview

    The Tosoh ST AIA-PACK BNP Assay aims to extend its measuring interval beyond 2,000 pg/mL through manual and automated 1:5 and 1:10 dilutions. This is a special 510(k) submission, meaning it refers to a device that has already been cleared (K192380) and modifications were made to it that do not significantly alter its performance or safety (e.g., modified firmware, revised labeling, minor material changes, etc.). Since this is a Special 510(k) and not a de novo submission, this document is a justification for substantial equivalence to its predicate device (K192380), and does not contain detailed information for how the predicate device was evaluated.

    Therefore, the study summary below only refers to the performance of the modified device compared to its predicate device, and not a full evaluation of the predicate device's performance.


    1. Acceptance Criteria and Reported Device Performance

    Acceptance CriteriaReported Device Performance
    Manual Dilution
    Recovery value102% for 1:5 dilution
    Recovery value100% for 1:10 dilution
    On-board (automated) Dilution
    Recovery value95% for 1:5 dilution
    Recovery value95% for 1:10 dilution

    2. Sample size used for the test set and data provenance

    The document does not specify the sample sizes used for the manual and automated dilution studies, nor does it specify the country of origin or whether the data was retrospective or prospective.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    Not applicable. The ground truth method does not involve human experts; it relies on direct measurement.

    4. Adjudication method for the test set

    Not applicable. The ground truth method does not involve human experts.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, if so, what was the effect size of how much human readers improve with AI versus without AI assistance

    Not applicable. This is not an AI-assisted diagnostic device, but an in vitro diagnostic assay.

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done

    Not applicable. This device is an in vitro diagnostic assay, not an algorithm.

    7. The type of ground truth used

    The ground truth used for both manual and automated dilution studies is the recovery value of the BNP concentration after dilution, indicating the accuracy of the dilution process. This is a direct measurement based on the expected concentration after dilution.

    8. The sample size for the training set

    Not applicable. This device is an in vitro diagnostic assay and does not involve machine learning or a training set.

    9. How the ground truth for the training set was established

    Not applicable. This device is an in vitro diagnostic assay and does not involve machine learning or a training set.

    Ask a Question

    Ask a specific question about this device

    K Number
    K200904
    Date Cleared
    2021-08-05

    (486 days)

    Product Code
    Regulation Number
    862.1373
    Reference & Predicate Devices
    Why did this record match?
    Applicant Name (Manufacturer) :

    Tosoh Bioscience, Inc.

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Tosoh Automated Glycohemoglobin Analyzer HLC-723G8 is intended for in vitro diagnostic use for the measurement of % hemoglobin A1c (HbA1c) (DCCT/NGSP) and mmol/mol hemoglobin A1c (IFCC) in venous whole blood specimens using ion-exchange high-performance liquid chromatography (HPLC). This test is an aid in diagnosis of diabetes and identifying patients who may be at risk for developing diabetes, and for monitoring of long-term blood glucose control in individuals with diabetes mellitus.

    Device Description

    The Tosoh Automated Glycohemoglobin Analyzer HLC-723G8 is an automated High-Performance Liguid Chromatography (HPLC) system that separates and reports stable hemoglobin A1c (sA1c) percentage in venous whole blood. The operational portion of the G8 is composed of a sampling unit, liquid pump, degasser, column, detector, microprocessors, sample loader, smart media card, operation panel, and a printer. The Tosoh Automated Glycohemoglobin Analyzer HLC-723G8 uses ion-exchange HPLC for rapid, accurate, and precise separation of the stable form of HbA1c (sA1c) from other hemoglobin fractions. The G8 uses a non-porous cation exchange column and separates the hemoglobin components in the blood. Separation is achieved by utilizing differences in ionic interactions between the cation and exchange group on the column resin surface and the hemoglobin components in a step gradient elution. The hemoglobin fractions (designated as A1a. A1b. F. LA1c+, SA1c, A0, and, if present, H-V0, H-V2, H-V2 and H-V3) are subsequently removed from the column by performing a step-wise elution gradient using the varied salt concentrations in the Variant Elution Buffers HSi 1, 2 and 3. The peaks, H-V0, H-V1, H-V2 and H-V3 are typically presumptive HbAD, HbAS, HbAC and HbAE respectively. The software compares the retention times of hemoglobin fractions in a sample to the expected "windows of retention" and labels each fraction that correctly elutes within a defined expected window of retention. The software designates a hemoglobin fraction as POX (where X is the order of the peak as it elutes from the column) if it does not match a defined window of retention. All automated processes in the G8 are controlled by internal microprocessors, using software downloaded via a smart media card. The result report is printed and can be stored on the instrument. The data can be transmitted to a host computer through a bi-directional interface. The result report includes the sample ID, date, percentage and retention time of each fraction of hemoglobin, sA1c percentage and total A1 percentage, along with a chromatogram of the elution pattern of the hemoglobin fractions. If a sample contains a hemoglobin variant, the column elutes the fraction depending upon its charge.

    AI/ML Overview

    The provided text describes the non-clinical performance testing of the Tosoh Automated Glycohemoglobin Analyzer HLC-723G8 (subject device) to support its substantial equivalence to a predicate device. This document focuses on the analytical performance of a diagnostic device rather than an AI/ML powered device, so some of the specific questions regarding AI/ML study design (e.g., number of experts, adjudication methods, MRMC studies) are not applicable.

    Here's the information extracted from the document:

    1. Acceptance Criteria and Reported Device Performance

    The acceptance criteria are generally implied by the statement "All performance testing results met their pre-determined acceptance criteria." While explicit numerical acceptance criteria for each test are not listed in a consolidated table, the discussion throughout the "Summary of Non-Clinical Performance Testing" implicitly defines them through the methodology and results. For example, for precision/repeatability, the claim of "imprecision at ≤ 2%" was a pre-established criterion. Similarly, for hemoglobin variant interference, "Non-clinically significant interference was defined as

    Ask a Question

    Ask a specific question about this device

    K Number
    K160113
    Date Cleared
    2016-02-17

    (29 days)

    Product Code
    Regulation Number
    862.1150
    Reference & Predicate Devices
    Why did this record match?
    Applicant Name (Manufacturer) :

    TOSOH BIOSCIENCE, INC.

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The ST AIA-PACK hsE2 Calibrator Set is intended for In Vitro Diagnostic Use Only for the calibration of the ST AIA-PACK hsE2 assay.

    Device Description
    • 2 x 1 mL ST AIA-PACK hsE2 Calibrator (1) 0 pg/mL Human serum containing no detectable concentration of estradiol with sodium azide as a preservative.
    • 2 x 1 mL ST AIA-PACK hsE2 Calibrator (2) 25 pg/mL (approx.)
    • ST AIA-PACK hsE2 Calibrator (3) 50 pg/mL (approx.)
    • ST AIA-PACK hsE2 Calibrator (4) 100 pg/mL (approx.)
    • ST AIA-PACK hsE2 Calibrator (5) 500 pg/mL (approx.)
    • ST AIA-PACK hsE2 Calibrator (6) 1,100 pg/mL (approx.)
      Human serum containing the assigned concentration of estradiol (described on each vial) with sodium azide as a preservative.
      ST AIA-PACK hsE2 Calibrator Set P/N # 025325
      The ST AIA-PACK hsE2 Calibrator Set is designed specifically for use on the Tosoh AIA System Analyzers which have been previously cleared as a family of instruments under K971103. Only materials obtained from Tosoh should be used. Materials obtained elsewhere should not be substituted since assay performance is characterized based strictly on Tosoh materials.
      The ST AIA-PACK hsE2 Calibrator Set is designed for use with ST AIA-PACK hsE2 and ST AIA-PACK hsE2 Sample Diluting Solution.
    AI/ML Overview

    The provided document describes the ST AIA-PACK hsE2 Calibrator Set and its performance.

    Here's an analysis of the acceptance criteria and the study proving the device meets them:

    1. Table of Acceptance Criteria and Reported Device Performance:

    The document details two types of stability studies, each with its own acceptance criteria and results for recovery and reproducibility (CV%).

    Test TypeAcceptance Criteria (Recovery)Reported Device Performance (Recovery)Acceptance Criteria (Reproducibility - CV%)Reported Device Performance (Reproducibility - CV%)
    Real Time Testing100% +/- 10%Not explicitly stated; "Current Real Time Studies support a 12 month Shelf life at 2-8°C" implies criteria were met
    Ask a Question

    Ask a specific question about this device

    K Number
    K153417
    Date Cleared
    2015-12-18

    (23 days)

    Product Code
    Regulation Number
    862.1150
    Reference & Predicate Devices
    Why did this record match?
    Applicant Name (Manufacturer) :

    TOSOH BIOSCIENCE, INC.

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The ST AIA-PACK PROG III Calibrator Set is intended for In Vitro Diagnostic Use Only for the calibration of the ST AIA-PACK PROG III assay.

    Device Description
    • 2 x 1 mL ST AIA-PACK PROG III Calibrator (1) 0 nq/mL Bovine protein matrix containing no detectable concentration of progesterone with sodium azide as a preservative (Liquid).
    • 2 x 1 mL ST AIA-PACK PROG III Calibrator (2) 0.5 ng/mL (approx.)
    • ST AIA-PACK PROG III Calibrator (3) 1.5 ng/mL (approx.)
    • ST AIA-PACK PROG III Calibrator (4) 5.0 ng/mL (approx.)
    • ST AIA-PACK PROG III Calibrator (5) 15 ng/mL (approx.)
    • ST AIA-PACK PROG III Calibrator (6) 45 ng/mL (approx.)
      Bovine protein matrix containing the assigned concentration of progesterone (described on each vial) with sodium azide as a preservative.
      The ST AIA-PACK PROG III Calibrator Set is designed specifically for use on the Tosoh AIA System Analyzers which have been previously cleared as a family of instruments under K971103. Only materials obtained from Tosoh should be used. Materials obtained elsewhere should not be substituted since assay performance is characterized based strictly on Tosoh materials.
      The ST AIA-PACK PROG III Calibrator Set is designed for use with ST AIA-PACK PROG III and ST AIA-PACK PROG III Sample Diluting Solution.
    AI/ML Overview

    Here's an analysis of the provided text regarding the acceptance criteria and the study that proves the device meets them, formatted as requested:

    Acceptance Criteria and Device Performance Study for ST AIA-PACK PROG III Calibrator Set

    This document describes the ST AIA-PACK PROG III Calibrator Set, a device intended for in vitro diagnostic use for the calibration of the ST AIA-PACK PROG III assay. The information provided heavily focuses on stability and value assignment studies to demonstrate substantial equivalence to a predicate device.

    1. Table of Acceptance Criteria and Reported Device Performance

    Study TypeAcceptance CriteriaReported Device Performance
    Real Time Stability (Shelf Life)Recovery: 100 +/- 10%Supports an 8-month shelf life at 2-8°C.
    Reproducibility (CV%): ≤ 10%(Specific CV% values not provided for shelf life, but implied to meet criteria)
    Open Vial StabilityRecovery: 100 +/- 10%Supports a 1-day reconstituted claim when stored at 2-8°C.
    Reproducibility (CV%): ≤ 10%(Specific CV% values not provided for open vial, but implied to meet criteria)
    Value Assignment PrecisionPrecision (CV%): Within 10% (for each calibrator level)Cal (2): 4.8% CV
    Cal (3): 2.2% CV
    Cal (4): 2.2% CV
    Cal (5): 2.0% CV
    Cal (6): 2.9% CV
    Value Assignment Recovery(Not explicitly stated as an acceptance criterion for individual calibrator levels but implied by "assigned value" and "Reference Value")Cal (2) Mean: 0.504 ng/mL (Reference: 0.5 ng/mL)
    Cal (3) Mean: 1.53 ng/mL (Reference: 1.5 ng/mL)
    Cal (4) Mean: 5.08 ng/mL (Reference: 5.0 ng/mL)
    Cal (5) Mean: 15.0 ng/mL (Reference: 15 ng/mL)
    Cal (6) Mean: 45.5 ng/mL (Reference: 45 ng/mL)

    2. Sample Size Used for the Test Set and Data Provenance

    • Test Set Sample Size:
      • Real Time Stability: Not explicitly stated, but calibrator sets were assayed at 3, 6, and 9 months (implied multiple samples/replicates at each time point).
      • Open Vial Stability: Not explicitly stated, but samples were "reconstituted and stored... for 2 days and tested."
      • Value Assignment: 5 replicates of the test calibrator were analyzed for each calibrator level (Cal 2-6).
    • Data Provenance: Not specified, but generally, such studies for regulatory submission are prospective and conducted in-house or by contract research organizations under the manufacturer's control. No country of origin is explicitly mentioned for the data itself, beyond the US location of the submitter.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

    • This device is a calibrator set for an in vitro diagnostic assay, not an AI or imaging device requiring human expert ground truth. The "ground truth" for calibrators is related to the reference materials and their assigned concentrations.
    • For Value Assignment: The primary reference material was USP (United States Pharmacopeia) Standard (Lot #I1J239). The progesterone value of this reference material was assigned gravimetrically, which is a chemical/analytical method, not typically involving human expert consensus in the diagnostic sense.

    4. Adjudication Method for the Test Set

    • Not applicable. This is a calibrator set for an in vitro diagnostic device, and adjudication methods (like 2+1, 3+1 for medical image interpretation) are not relevant here. The evaluation relies on analytical performance criteria (recovery, CV%) against established reference values.

    5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was Done

    • No. This is not an AI or imaging device; therefore, an MRMC study is not relevant.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was Done

    • The device itself is a calibrator, which is a reagent used by an instrument. Its performance is evaluated analytically, which is inherently "standalone" in the sense that its function (calibration) is determined by its chemical and physical properties and how it interacts with the instrument. There isn't an "algorithm only" performance metric in the typical sense of AI. The "algorithm" here is the assay procedure itself on the Tosoh AIA System Analyzers, and the calibrator's performance helps ensure the accuracy of that system.

    7. The Type of Ground Truth Used

    • Analytical/Reference Standard: The ground truth for the calibrator concentrations is established by referring to the USP (United States Pharmacopeia) Standard (Lot #I1J239). The values were assigned gravimetrically (for the primary reference material). This is a highly controlled chemical/analytical method, not expert consensus, pathology, or outcomes data.

    8. The Sample Size for the Training Set

    • Not applicable. This is an in vitro diagnostic calibrator, not an AI model that requires a training set of data. The "training" in this context refers to the development and manufacturing of the calibrator itself, where the concentrations are established and verified against reference standards.

    9. How the Ground Truth for the Training Set was Established

    • Not applicable, as there is no "training set" in the AI sense for this type of device. The "ground truth" for the calibrator's assigned values is based on traceability to the USP (United States Pharmacopeia) Standard (Lot #I1J239), with primary reference material values assigned gravimetrically. These are highly standardized analytical procedures to ensure accuracy and consistency. The product calibrator values were then assigned using AIA-2000 instruments with the secondary reference material as calibrator.
    Ask a Question

    Ask a specific question about this device

    Why did this record match?
    Applicant Name (Manufacturer) :

    TOSOH BIOSCIENCE, INC.

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    ST AIA-PACK 25-OH Vitamin D is designed for in vitro diagnostic use only for the quantitative measurement of total 25-hydroxyvitamin D (25-OH vitamin D) in human serum, Na-heparinized plasma or EDTA plasma on TOSOH AIA System Analyzers. The Tosoh ST AIA-PACK 25-OH Vitamin D is intended as an aid in the determination of Vitamin D sufficiency.

    The ST AIA-PACK 25-OH Vitamin D Calibrator Set is intended for in vitro diagnostic use only for the calibration of the ST AIA-PACK 25-OH Vitamin D assay.

    Device Description

    The ST AIA-PACK 25-OH Vitamin D is a one-step delayed competitive enzyme immunoassay which, after sample pretreatment, is performed entirely in the ST AIA-PACK 25-OH Vitamin D test cup. Sample pretreatment reagents (containing sodium hydroxide) disassociate 25-OH vitamin D from its binding proteins in the test sample. 25-OH vitamin D present in the pretreated sample is bound to 25-OH vitamin D-specific monoclonal antibody immobilized on magnetic beads. After that, the enzymelabeled 25-OH vitamin D is added to the reaction mixture. The enzyme-labeled 25-OH vitamin D competes with 25-OH vitamin D for binding to the antibody on magnetic beads in the reaction mixture.

    After the second incubation, the magnetic beads are washed to remove the unbound enzymelabeled 25-OH vitamin D and are then incubated with a fluorogenic substrate, 4- methylumbellifery phosphate (4MUP). The amount of enzyme-labeled 25-OH vitamin D that binds to the beads is inversely proportional to the 25-0H vitamin D concentration in the test sample. A standard curve is constructed, and unknown 25-OH vitamin D concentrations are calculated using this curve.

    The ST AIA-PACK 25-OH Vitamin D Calibrator Set contains human sera with assigned levels of 25-OH Vitamin D. The calibrator set consists of six calibrators with assigned concentrations of approximately 0, 8, 17, 33, 66 and 135 ng/mL. Each level contains the assigned concentration of the 25-OH vitamin D (described on each vial) with sodium azide as a preservative.

    AI/ML Overview

    This looks like a 510(k) Summary for a medical device called "ST AIA-PACK 25-OH Vitamin D" and its associated calibrator set. The document describes the device, its intended use, and provides evidence of its performance and substantial equivalence to a predicate device.

    Let's break down the requested information based on the provided text.

    1. A table of acceptance criteria and the reported device performance

    The document doesn't explicitly state "acceptance criteria" in a separate section with pass/fail thresholds against which the performance is measured. Instead, it presents performance characteristics and largely uses the performance of the predicate device or established clinical standards (like CLSI guidelines) as a benchmark for comparison or to demonstrate expected performance.

    However, we can infer some criteria from the presented performance characteristics.

    Performance CharacteristicAcceptance Criteria (inferred/implied)Reported Device Performance
    PrecisionEqual to or exceed predicate calibrator traceability (K123131).Within-run precision: 1.3 - 3.9%
    Between run precision: 1.5 - 3.1%
    Between day precision: 2.2 - 4.1%
    Total precision: 2.6 - 4.9%
    Between lot precision: 1.1 - 3.1%
    Statement: "The precision equals or exceeds the precision obtained with the predicate calibrator traceability for (k)123131."
    LinearityDemonstrated to be linear within the assay range.Linear from 4.0 ng/mL to 120 ng/mL.
    Statement: "There is no change in linearity from K123131."
    Standardization & TraceabilityTraceable to ID-LC/MS/MS 25-OH Vitamin D RMP (University of Ghent). Bias should be acceptable.Weighted Deming regression analysis (vs. RMP):
    Slope: 0.98 (0.92 - 1.05)
    Intercept: -0.48 (-1.89 - 0.92)
    Corr Coef (R): 0.965
    Bias: -0.23 (-0.71%)
    The study concludes that the assay is traceable to the RMP.
    Method Comparison (vs. Predicate)Demonstrate acceptable correlation and bias against predicate.Deming regression analysis (vs. Predicate):
    Slope: 0.808 (0.801 to 0.816)
    Intercept: 0.26 (-0.15 to 0.67)
    Corr Coef (R): 0.9979
    Bias: -8.90 (-20.25%)
    This comparison demonstrates the change in bias introduced by the standardization process.
    Reference RangesEstablish a representative reference interval.Reference Interval: 12.3 – 60.0 ng/mL (based on 252 healthy individuals).
    Limit of Detection (LoD)Determined according to CLSI guideline EP17-A.LoD: 3.2 ng/mL (determined by 12 measurements of 5 low-level samples across 3 lots)
    Limit of Quantitation (LoQ)Determined according to CLSI guideline EP17-A.Functional sensitivity (LoQ): 3.3 ng/mL at 20% CV.
    Limit of Blank (LoB)Determined according to CLSI guideline EP17-A.LoB: 1.6 ng/mL (determined by 60 measurements of 3 different blank specimens)

    2. Sample sizes used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    • Precision:
      • Sample Size: Three levels of unaltered serum specimens (Serum-B and Serum-C) and three spiked serum specimens (Serum-E and Serum-F) were assayed. Each level involved 2 replicates in a single run, 2 times a day for 20 non-consecutive days, across 2 analyzers and 2 lots of reagents. This totaled 40 runs and 80 determinants per lot.
      • Data Provenance: Not explicitly stated, but typically clinical laboratory studies are prospective and involve samples collected for internal validation or purchased from biorepositories. No country of origin is mentioned.
    • Linearity:
      • Sample Size: Not explicitly stated, but typically involves a series of diluted samples to cover the assay range.
      • Data Provenance: Not explicitly stated.
    • Standardization and Traceability (Method Comparison vs. RMP):
      • Sample Size: 111 serum specimens.
      • Data Provenance: The RMP (Reference Measurement Procedure) was performed at the University of Ghent, in Ghent, Belgium. The samples were value-assigned by the Ghent University ID-LC-MS/MS RMP. This suggests these were likely retrospective samples used for standardization purposes in a prospective study design to validate the device's traceability.
    • Method Comparison (vs. Predicate):
      • Sample Size: 181 unaltered serum specimens.
      • Data Provenance: Not explicitly stated, but likely from various sources for routine clinical testing. Implied to be a prospective comparison study.
    • Reference Ranges:
      • Sample Size: 252 serum samples (111 females, 141 males).
      • Data Provenance: Specimens were obtained from "apparently healthy individuals" in Maryland, Pennsylvania, Wisconsin, and Southern California (USA). Samples were collected in March, May, June, and July, indicating a prospective collection over several months.
    • Limit of Detection and Limit of Quantitation:
      • LoB: 60 measurements of 3 different blank specimens.
      • LoD: 12 measurements of 5 low-level sample ranges, using 3 lots of reagents.
      • LoQ: Series of low Vitamin D samples assayed in replicates of 8 for 5 days (total 40 replicates per sample).
      • Data Provenance: Not explicitly stated.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    • For this in vitro diagnostic device, "ground truth" is primarily established by highly accurate reference methods or clinical standards, rather than expert consensus on interpretation.
    • Standardization and Traceability: The ground truth for this segment was the ID-LC/MS/MS 25-OH Vitamin D Reference Measurement Procedure (RMP) at the University of Ghent. This is a highly specialized analytical chemistry method considered the "gold standard" for measuring 25-OH Vitamin D, not typically performed by individual "experts" in the clinical sense, but rather by trained laboratory personnel following precise protocols. The qualifications of the individuals performing the RMP are implied to be high-level analytical chemists/technicians experienced in ID-LC/MS/MS.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    • Adjudication methods like 2+1 or 3+1 are typically used in clinical imaging studies where subjective interpretation is involved.
    • For an in vitro diagnostic device measuring a quantitative analyte like 25-OH Vitamin D, the "adjudication" is inherent in the analytical method's precision, accuracy, and traceability to a recognized reference method (like the ID-LC/MS/MS RMP). Discordant results would be investigated through re-testing or troubleshooting, not through a consensus of human reviewers. Therefore, the concept of a multi-observer adjudication method is not applicable here.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • This is an in vitro diagnostic device for measuring a biochemical marker (Vitamin D). The concept of "multi-reader multi-case (MRMC) comparative effectiveness study" for human readers improving with or without AI assistance is not applicable to this type of device. There is no "reading" of images or complex data by human experts being assisted by AI in the context of this submission. The device performs an automated quantitative measurement.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

    • This device is an automated, standalone assay system. Its performance characteristics (precision, linearity, LoD, LoQ, method comparison) are its standalone performance without continuous human intervention in the interpretive phase. While humans operate the instrument and perform quality control, the measurement itself is algorithmic. Therefore, the presented studies (precision, linearity, method comparison, etc.) represent the standalone performance of the device algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    • The primary ground truth used for validating the accuracy and standardization of the device is the Isotope Dilution-Liquid Chromatography/Tandem Mass Spectrometry (ID-LC/MS/MS) 25-OH Vitamin D Reference Measurement Procedure (RMP). This is a highly accurate and precise analytical method recognized as the gold standard for 25-OH Vitamin D measurement, traceable to NIST Standard Reference Material (SRM) 2972.

    8. The sample size for the training set

    • This document describes a 510(k) submission for a diagnostic assay, not a machine learning or AI algorithm in the context of typical training/test sets. The "training" for such an assay involves the development and optimization of the reagents, assay protocol, and calibration procedures using various samples. There isn't a single, defined "training set" sample size in the sense of AI model development.
    • The "training" data, in an analogous sense, would be all the samples and experiments conducted during the development phase to establish the assay's performance characteristics, optimize reagent concentrations, and set initial calibrator values. This information is typically not detailed in a 510(k) summary as a single "training set."

    9. How the ground truth for the training set was established

    • As mentioned above, the concept of a distinct "training set" with ground truth in the AI sense is not directly applicable here. However, the development and initial calibration of the assay would have relied on highly characterized samples, likely correlated with established reference methods or primary standards (such as those traceable to NIST SRMs) to ensure accuracy and consistency from the earliest stages of development. The subsequent "standardization" step (as described in the document, traceable to the University of Ghent's ID-LC/MS/MS RMP) represents a formal validation and adjustment process to ensure consistency with recognized reference methods.
    Ask a Question

    Ask a specific question about this device

    K Number
    K143075
    Date Cleared
    2015-07-02

    (248 days)

    Product Code
    Regulation Number
    862.1680
    Reference & Predicate Devices
    Why did this record match?
    Applicant Name (Manufacturer) :

    TOSOH BIOSCIENCE, INC.

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    ST AIA-PACK SHBG is designed for In Vitro Diagnostic Use Only for the quantitative measurement of sex hormone binding globulin (SHBG) in human serum or Na heparinized plasma on Tosoh AIA System Analyzers. The ST AIA-PACK SHBG assay is intended for use as an aid in the diagnosis of androgen disorders.

    The ST AIA-PACK SHBG Calibrator Set is intended for In Vitro Diagnostic Use Only for the calibration of the ST AIA-PACK SHBG assay.

    Device Description

    The ST AIA-PACK SHBG is a two-site immunoenzymometric assay which is performed entirely in the ST AIA-PACK SHBG test cups. SHBG present in the test sample is bound with monoclonal antibody immobilized on a magnetic solid phase and enzyme-labeled monoclonal antibody in the test cups. The magnetic beads are washed to remove unbound enzyme-labeled antibody and are then incubated with a fluorogenic substrate, 4-methylumbelliferyl phosphate (4MUP). The amount of enzyme-labeled monoclonal antibody that binds to the beads is directly proportional to the SHBG concentration in the test sample. A standard curve is constructed, and unknown sample concentrations are calculated using this curve.

    AI/ML Overview

    The provided text describes the performance characteristics of the ST AIA-PACK SHBG device, which is an in vitro diagnostic assay for measuring Sex Hormone Binding Globulin (SHBG). The studies conducted for this device do not involve AI or human readers for image interpretation, but rather focus on analytical performance criteria typical for a laboratory diagnostic assay. Therefore, questions related to AI assistance, human reader improvement, and adjudication methods are not applicable.

    Here's an analysis of the acceptance criteria and study information provided:

    1. Table of Acceptance Criteria and Reported Device Performance

    For analytical devices like the ST AIA-PACK SHBG, acceptance criteria are typically specified in terms of performance metrics like precision (CV%), linearity (measuring range), limit of detection (LoD), and correlation with a predicate device. The document generally presents the performance metrics achieved by the device without explicitly stating pre-defined "acceptance criteria" numerical targets. However, comparisons to the predicate device and CLSI guidelines serve as implicit benchmarks for acceptability.

    Performance CharacteristicImplicit Acceptance Criteria (based on CLSI guidelines/predicate comparison/industry standards)Reported Device Performance (ST AIA-PACK SHBG)
    Precision (Total Precision CV%)Typically 0.95, slope near 1, intercept near 0).Deming Slope: 0.949 (0.926 to 0.972), Intercept: -0.64 (-2.61 to 1.34), Corr Coef (R): 0.991. These values demonstrate excellent correlation with the predicate.
    Matrix Comparison (Serum vs. Plasma)High correlation between different sample matrices.Deming Slope: 0.977 (0.964 to 0.991), Intercept: 0.269 (-0.629 to 1.168), Corr Coef (R): 0.997. These values demonstrate excellent correlation between serum and Na heparinized plasma.
    InterferenceNo significant interference from common interfering substances (e.g., hemoglobin, bilirubin, lipids, HAMA). Recovery typically within +/-10%.No interference observed from: Hemoglobin (up to 446 mg/dL), free and conjugated bilirubin (up to 17.6 mg/dL and 18.5 mg/dL respectively), lipemia (triglycerides up to 1,667 mg/dL), ascorbic acid (up to 20 mg/dL), protein (albumin up to 5.00 g/dL), Na Heparin (up to 100.0 U/mL), Rheumatoid factor (up to 550 IU/mL), HAMA (up to 24,269 ng/mL). This meets the non-interference expectation.
    Cross-ReactivityMinimal to no cross-reactivity with related or common substances.Generally N.D. (not detectable) or very low percentage (e.g., 0.003% for Cortisol, 0.019% for Testosterone). Thyroglobulin had a 2.544% cross-reactivity, which may be noted but specific acceptance for such levels is not provided.
    Calibrator Stability (Real Time)Recovery within 100 +/- 10%; Reproducibility (CV%)
    Ask a Question

    Ask a specific question about this device

    K Number
    K143296
    Date Cleared
    2014-12-16

    (29 days)

    Product Code
    Regulation Number
    862.1660
    Reference & Predicate Devices
    Why did this record match?
    Applicant Name (Manufacturer) :

    Tosoh BioScience, Inc.

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The AIA-PACK C-Peptide Control Set is intended for In Vitro Diagnostic Use Only for performing quality control procedures with the ST AIA-PACK C-Peptide II assay.

    Device Description

    2 x 2 mL AIA-PACK C-Peptide Control Level 1 Buffered bovine serum albumin containing approximately 2 nq/mL of C-peptide (Lyophilized). See the vial label for the assigned concentration range. AIA-PACK C-Peptide Control Level 2 2 x 2 mL Buffered bovine serum albumin containing approximately 20 ng/mL of C-peptide (Lyophilized). See the vial label for the assigned concentration range. The AIA-PACK C-Peptide Control Set contains buffered bovine serum albumin with assigned levels of C-peptide. The C-peptide controls shall be run at the beginning of each day on which C-peptide assays are scheduled.

    AI/ML Overview

    This document describes the TOSOH BIOSCIENCE, INC. AIA-PACK C-Peptide Control Set, a quality control material. The provided text primarily focuses on demonstrating substantial equivalence to a predicate device and summarizing stability and value assignment studies. It does not contain information about a study proving the device meets clinical acceptance criteria for diagnostic performance, as it is a control set, not a diagnostic device itself.

    Therefore, many of the requested items (e.g., sample size for test set, data provenance, number of experts for ground truth, MRMC study, standalone performance) are not applicable or discussed for this type of device (a quality control material).

    Here's a breakdown of the available information:

    1. Table of Acceptance Criteria and Reported Device Performance

    CriterionAcceptance CriteriaReported Performance
    Stability (Shelf Life)Recovery within 100 ± 10%. Reproducibility (CV%) ≤ 10%.6, 12, and 13 months after manufacture: Recovery was within 100 ± 10% and reproducibility (CV%) was ≤ 10%.
    Conclusion: Shelf life set at 12 months at 2-8°C.
    Stability (In-Use, Reconstituted)Recovery within 100 ± 10%. Reproducibility (CV%) ≤ 10%.15 days at refrigerator temperature (2-8°C) after reconstitution: Recovery and reproducibility met acceptance criteria.
    Conclusion: In-use stability set at 14 days after reconstitution, provided vials are kept tightly sealed and refrigerated (2-8°C).
    Value Assignment (Control Set)Mean values fell within the value range. All CV% fell within pre-determined acceptance criteria. Assigned control value range is ± 20% of the target value. Target levels: approximately 2 ng/mL and 20 ng/mL of C-peptide.For 2 lots of control levels 1 and 2, using two AIA 2000 analyzers: Measurements in 5 replicates showed mean values fell within the value range and all CV% fell within pre-determined acceptance criteria. The assigned concentration range is determined lot-by-lot to provide target control levels of approximately 2 and 20 ng/mL of C-peptide, with a range of +/- 20% of the target value.

    2. Sample size used for the test set and the data provenance

    • Test Set Sample Size:
      • Stability Studies: Three different lots of the AIA-PACK C-Peptide Control Set were used as samples for the study.
      • In-Use Stability: A single lot of control set was used. Each specimen was assayed in 5 replicates.
      • Value Assignment: Two lots for control levels 1 and 2 were used. Measurements were made in 5 replicates.
    • Data Provenance: Not explicitly stated, but clinical laboratory studies involving control materials are typically conducted internally by the manufacturer or by contract labs in the country where the manufacturer is based (USA in this case, based on the address provided). The studies described appear to be prospective studies specifically designed to assess the performance of the control set under various conditions.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    • Not applicable. This device is a quality control material. Its performance is assessed against defined analytical criteria (recovery, reproducibility) rather than by expert clinical interpretation. The "ground truth" for the assigned values is established through a traceable process using reference materials, not expert consensus on diagnostic images or patient outcomes.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    • Not applicable. This is a quality control material. Its performance is based on analytical measurements against pre-defined quantitative criteria.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • Not applicable. This device is a quality control material and not an AI-powered diagnostic device or a system intended for human reader assistance.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • Not applicable. This device is a quality control material; it does not involve an algorithm for diagnostic performance.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    • For the stability and in-use studies, the "ground truth" is the expected original concentration of C-peptide in the control material, against which recovery and reproducibility are measured.
    • For value assignment, the ground truth for the primary reference material was assigned based on C-Peptide of Human Insulin, International Reference Reagent (WHO International Reference Reagent). The secondary reference material and the control set values were then assigned using this primary reference material as a calibrator, following a hierarchical traceability scheme.

    8. The sample size for the training set

    • Not applicable. This is a quality control material; it does not involve a "training set" in the context of machine learning.

    9. How the ground truth for the training set was established

    • Not applicable. There is no training set for this type of device.
    Ask a Question

    Ask a specific question about this device

    K Number
    K143144
    Date Cleared
    2014-12-03

    (30 days)

    Product Code
    Regulation Number
    862.1150
    Reference & Predicate Devices
    Why did this record match?
    Applicant Name (Manufacturer) :

    Tosoh BioScience, Inc.

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The ST AIA-PACK PROG II Calibrator Set is intended for In Vitro Diagnostic Use Only for the calibration of the ST AIA-PACK PROG II assay.

    Device Description

    2 x 1 mL ST AIA-PACK PROG II Calibrator (1) 0 ng/mL Protein matrix containing no detectable concentration of PROG with sodium azide as a preservative (Liquid).
    2 x 1 mL ST AIA-PACK PROG II Calibrator (2) 0.5 ng/mL (approx.)
    ST AIA-PACK PROG II Calibrator (3) 1.5 ng/mL (approx.)
    ST AIA-PACK PROG II Calibrator (4) 5.0 ng/mL (approx.)
    ST AIA-PACK PROG II Calibrator (5) 15 ng/mL (approx.)
    ST AIA-PACK PROG II Calibrator (6) 45 ng/mL (approx.)
    Human serum containing the assigned concentration of progesterone (described on each vial) with sodium azide as a preservative.

    The ST AIA-PACK PROG II Calibrator Set is designed specifically for use on the Tosoh AIA System Analyzers which have been previously cleared as a family of instruments under K971103. Only materials obtained from Tosoh should be used. Materials obtained elsewhere should not be substituted since assay performance is characterized based strictly on Tosoh materials.

    The ST AIA-PACK PROG II Calibrator Set is designed for use with ST AIA-PACK PROG II and ST AIA-PACK PROG II Sample Diluting Solution.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and study information for the ST AIA-PACK PROG II Calibrator Set, based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criterion (Stability)RequirementReported Performance (ST AIA-PACK PROG II Calibrator Set)
    Real-Time StabilityRecovery within 100 ± 10%Met (supports 6-month shelf life)
    Real-Time StabilityReproducibility (CV%) ≤ 10%Met (supports 6-month shelf life)
    Open Vial StabilityRecovery within 100 ± 10%Met (supports 1-day reconstituted claim)
    Open Vial StabilityReproducibility (CV%) ≤ 10%Met (supports 1-day reconstituted claim)
    Value Assignment PrecisionPrecision (CV%) ≤ 10%Met (all calibrator levels were ≤ 5.8%)

    Acceptance Criterion (Value Assignment - Precision)Test for Calibrator levels (Mean CV%)
    CV% ≤ 10% (for 5 replicates)Cal (2): 5.8%
    Cal (3): 3.5%
    Cal (4): 1.9%
    Cal (5): 2.2%
    Cal (6): 1.8%

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size:
      • For stability studies (Real-Time and Open Vial): The exact number of individual samples tested isn't explicitly stated, but it mentions "AIA-PACK Progesterone Calibrator Set were stored..." and "Samples were reconstituted..." implying multiple sets or vials.
      • For value assignment: 5 replicates for each of the 5 calibrator levels (Cal 2-6) were analyzed.
    • Data Provenance: The document does not specify the country of origin for the data. The data appears to be prospective as it involves the testing of the newly formulated ST AIA-PACK PROG II Calibrator Set under defined conditions (stability studies, value assignment protocol).

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    • The document does not mention the use of experts to establish the ground truth for the test set.
    • The ground truth for the calibrators is established through a chain of traceability:
      • Primary reference material uses USP (United States Pharmacopeia) Standard as calibrator.
      • Secondary reference material is assigned against the primary reference material.
      • Product calibrators are assigned against the secondary reference material.
    • The "experts" are the "Tosoh AIA-2000" instruments and the established calibration process rooted in the USP Standard.

    4. Adjudication Method for the Test Set

    • No human adjudication method (e.g., 2+1, 3+1) is mentioned or implied for the test set.
    • The determination of acceptance is based on quantitative measurements against predefined criteria (recovery % and CV%).

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve with AI vs Without AI Assistance

    • No, an MRMC comparative effectiveness study was not done.
    • This device is a calibrator set for an in vitro diagnostic assay, not an AI-powered diagnostic tool that assists human readers. Therefore, this type of study is not applicable.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    • This device is a calibrator set, not an algorithm. Therefore, "standalone algorithm performance" is not applicable in the context of AI.
    • However, the performance of the calibrator itself is tested in a "standalone" fashion in that its stability and assigned values are determined within the analytical system (Tosoh AIA-2000 with the specified assay) without human interpretation being part of the performance measurement itself. The results are quantitative and objective based on instrument readings.

    7. The Type of Ground Truth Used

    • The ground truth for the calibrator values is established based on a reference standard and hierarchical assignment process:
      • The highest level of truth is the USP (United States Pharmacopeia) Standard for progesterone.
      • Subsequent calibrator lots and levels are then assigned values through a traceable process against this primary reference and internal secondary standards.

    8. The Sample Size for the Training Set

    • This document describes a calibrator set and its performance validation, not a machine learning or AI model. Therefore, there is no training set in the context of AI. The "training" for this device would be analogous to the manufacturing and value assignment process itself, ensuring each lot meets specifications.

    9. How the Ground Truth for the Training Set Was Established

    • N/A, as there is no training set in the AI context. The ground truth for the calibrator's assigned values is established through traceability to the USP Standard as described in point 7.
    Ask a Question

    Ask a specific question about this device

    K Number
    K140648
    Date Cleared
    2014-04-10

    (28 days)

    Product Code
    Regulation Number
    862.1150
    Reference & Predicate Devices
    Why did this record match?
    Applicant Name (Manufacturer) :

    Tosoh BioScience, Inc.

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The ST AIA-PACK C-Peptide II Calibrator Set is intended for In Vitro Diagnostic Use Only for the calibration of the ST AIA-PACK C-Peptide II assay.

    Device Description

    2 x 1 mL ST AIA-PACK C-Peptide II Calibrator (1) 0 ng/mL Protein matrix containing no detectable concentration of C-peptide with sodium azide as a preservative (Liquid).
    2 x 1 mL ST AIA-PACK C-Peptide II Calibrator (2) 0.5 ng/mL (approx.)
    ST AIA-PACK C-Peptide II Calibrator (3) 2 ng/mL (approx.)
    ST AIA-PACK C-Peptide II Calibrator (4) 6 ng/mL (approx.)
    ST AIA-PACK C-Peptide II Calibrator (5) 15 ng/mL (approx.)
    ST AIA-PACK C-Peptide II Calibrator (6) 33 ng/mL (approx.)
    Protein matrix containing the assigned concentration of C-peptide (described on each vial) (Lyophilized).
    The ST AIA-PACK C-Peptide II Calibrator Set is designed specifically for use on the Tosoh AIA Svstem Analyzers which have been previously cleared as a family of instruments under K971103. Only materials obtained from Tosoh should be used. Materials obtained elsewhere should not be substituted since assay performance is characterized based strictly on Tosoh materials.
    The ST AIA-PACK C-Peptide II Calibrator Set is designed for use with ST AIA-PACK C-Peptide II, ST AIA-PACK C-Peptide Sample Diluting Solution, AIA-PACK C-Peptide Control Set.

    AI/ML Overview

    Here's an analysis of the provided text regarding the ST AIA-PACK C-Peptide II Calibrator Set, focusing on acceptance criteria and supporting studies.

    Based on the provided 510(k) Summary, this document primarily discusses the substantial equivalence of a calibrator set, not a diagnostic device with performance metrics like accuracy, sensitivity, or specificity in relation to a disease state. Therefore, many standard questions about AI device performance (like MRMC studies, standalone performance, training set details, expert qualifications for ground truth, etc.) are not applicable to this type of submission.

    The "acceptance criteria" and "device performance" in this context refer to the stability and reproducibility of the calibrator set itself, not its diagnostic accuracy in patients.


    1. Table of Acceptance Criteria and Reported Device Performance

    This calibrator set's "performance" is assessed based on its stability and reproducibility rather than diagnostic accuracy.

    Performance MetricAcceptance CriteriaReported Device Performance
    Shelf-Life StabilityRecovery: 100 ± 10%Recovery: Within 100 ± 10%
    (Unopened vial, 2-8°C)Reproducibility (CV%): ≤ 10%Reproducibility (CV%): ≤ 10% at 13 months
    In-Use StabilityRecovery: 100 ± 10%Recovery: Met criteria for 2 days at refrigerator temperature
    (Opened/Reconstituted vial,Reproducibility (CV%): ≤ 10%Reproducibility (CV%): Met criteria for 2 days at refrigerator temperature
    2-8°C)

    The conclusions from these studies allowed the manufacturer to set:

    • Shelf-life: 12 months at 2-8°C (based on 13-month data meeting criteria)
    • In-use stability: 1 day (24 hours) after reconstitution at 2-8°C (based on 2-day data meeting criteria)

    2. Sample Size Used for the Test Set and Data Provenance

    • Test Set (Stability Studies) Sample Size:
      • Shelf-Life Stability: Three different lots of calibrator set. Each specimen (presumably from each lot at each time point) was assayed in 5 replicates. This refers to the number of lots tested, not patient samples.
      • In-Use Stability: One lot of calibrator material. Three sets of calibrator material were opened and reconstituted. Urine, serum, and EDTA plasma specimens were chosen for this study. Each specimen was assayed in 5 replicates. The number of patient specimens (urine, serum, EDTA plasma) is not explicitly stated, only that such specimens were chosen.
    • Data Provenance: Not specified, but generally for such in-vitro diagnostic (IVD) calibrators, the studies are conducted internally by the manufacturer. The text does not mention country of origin or whether the studies were external. The nature of the study (stability of a reagent) suggests it's prospective testing in a lab setting.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    • Not applicable. This device is a calibrator set, not a diagnostic device requiring expert interpretation of results to establish a "ground truth" for clinical performance. Its "ground truth" (assigned C-peptide concentrations) is established through an analytical traceability chain to international standards (WHO 1ª IRP 84/510).

    4. Adjudication Method for the Test Set

    • Not applicable. There is no "adjudication" in the context of stability and reproducibility testing for a calibrator. The performance is assessed against predefined recovery and CV% criteria.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

    • No. This is not an AI-powered diagnostic device, and thus MRMC studies are not relevant.

    6. If a Standalone (i.e. algorithm only without human-in-the loop performance) Was Done

    • No. This is a calibrator product for an in-vitro diagnostic assay, not an algorithm.

    7. The Type of Ground Truth Used

    • Analytical Traceability to International Standard complemented by Internal Referencing: The assigned value of C-peptide in the primary reference material (which is then used to assign values to the calibrators) was based on WHO 1ª IRP 84/510. This is an internationally recognized reference preparation for C-peptide. Subsequent secondary reference materials and product calibrators are assigned values using Tosoh AIA instruments, comparing measured results with those obtained with previous lots or a primary reference.

    8. The Sample Size for the Training Set

    • Not applicable. This is a calibrator set, not an AI or machine learning model that requires a training set. The values are assigned based on analytical methods and traceability, not learned from data.

    9. How the Ground Truth for the Training Set Was Established

    • Not applicable, as there is no "training set." The "ground truth" for the calibrator values is established through the analytical traceability described in point 7.
    Ask a Question

    Ask a specific question about this device

    K Number
    K131580
    Date Cleared
    2014-01-23

    (237 days)

    Product Code
    Regulation Number
    862.1373
    Reference & Predicate Devices
    Why did this record match?
    Applicant Name (Manufacturer) :

    Tosoh BioScience, Inc.

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Tosoh Automated Glycohemoglobin Analyzer HLC-723G8 is intended for in vitro diagnostic use for the quantitative measurement of % hemoglobin A Ic (HbA 1c) (DCCT/NGSP) and mmol/mo hemoglobin A1c (IFCC) in whole blood specimens. This test is to be used as an aid in diagnosis of diabetes and as an aid in identifying patients who may be at risk for developing diabetes.

    Device Description

    The Tosoh Automated Glycohemoglobin Analyzer HLC-723G8 is an automated High Performance Liquid Chromatography (HPLC) system that separates and reports stable A 1c (sA1c) percentage in whole blood. The operational portion of the G8 is composed of a sampling unit, liquid pump, degasser, column, detector, microprocessors, sample loader, floppy disk drive unit, operation panel and a printer. The Tosoh Automated Glycohemoglobin Analyzer HLC-723G8 uses non-porous ion- exchange high performance liquid chromatography (HPLC) for rapid, accurate and precise separation of the stable form of HbA Ic from other hemoglobin fractions. The G8 uses a cation exchange column and separates the usual hemoglobin components in the blood into six fractions, A 1a, A 1b, F, LA 1c, sA Ic, and A0. The separation is done by eluting the hemoglobins from the column with a stepwise elution of three elution buffers containing different salt concentrations. The result report includes a sample ID, date, percentage and retention time of each fraction, sA1c percentage and total A1 percentage (A1a + A1b + sA1c), along with a chromatogram of the elution pattern of the hemoglobin fractions. If a sample contains a hemoglobin variant, the column elutes the material depending upon its charge.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and the study details for the Tosoh Automated Glycohemoglobin Analyzer HLC-723G8, based on the provided 510(k) summary:

    Acceptance Criteria and Device Performance

    CriterionAcceptance Criteria (Stated or Implied)Reported Device Performance
    Linearity / Reportable RangePreviously established for 4.0% - 16.9% HbA1c (from predicate device comparison table)4.0% - 16.9% HbA1c
    InterferenceVariance greater than assigned value x 1.00 ± 5% for potential interferents. No interference from HbA2, HbS, HbC, HbD. Flag for HbE.No Significant Interference Observed at Specified Concentrations:
    Acetylated Hb: 50 mg/dL
    Albumin: 5000 mg/dL
    Aldehyde Hb: 25 mg/dL
    Ascorbic Acid: 25 mg/dL
    Carbamylated Hb: 25 mg/dL
    Bilirubin C: 21 mg/dL
    Labile Hb: 1000 mg/dL
    Lipemia: 1000 mg/dL
    Rheumatoid Factor: 550 IU/mL
    Bilirubin F: 18 mg/dL
    Hemoglobin Variant Interference:
    HbS: No interference (at ~6% and ≥8% HbA1c levels)
    HbC: No interference (at ~6% and ≥8% HbA1c levels)
    HbE: Yes, interference. A flag is displayed, and %HbA1c result is not reported.
    HbD: No interference (at ~6% and ≥8% HbA1c levels)
    HbF: No interference (at ~6% and ≥8% HbA1c levels)
    HbA2: No interference (at ~6% and ≥8% HbA1c levels)
    PrecisionNot explicitly stated as a numerical criterion for this device, but predicate device states "equal to or less than 1.5% for concentrations in the range of 5.3% to 12.1% HbA1c" for between-analyzer and between-lot. For this device, "0.7 - 1.8% (total precision) and 0.3 - 0.9% (within run, precision)" is mentioned in the comparison table as the performance of the new device.Within-run Precision (CV%): Ranges from 0.18% (12% D1) to 1.00% (5% A2).
    Total Precision (CV%): Ranges from 0.68% (6.5% B1) to 1.83% (5% A1).
    Correlation (Method Comparison)r ≥ 0.991; Y ≈ 1.0x + 0.0y (implied from predicate, which showed Y = 1.015x + 0.015)Compared to Primus ultra2:
    Slope (Deming): 0.996 (0.971 to 1.021)
    Intercept (Deming): 0.073 (-0.110 to 0.255)
    Correlation Coefficient (R): 0.991
    Bias: 0.047 (0.657%)
    Total Error (TE)%TE of ≤6% is acceptable. Goal at 6% for (TE)95 (e.g., 0.30 at 5.0% Decision Level, 0.72 at 12.0% Decision Level)%TE (at 5%, 6.5%, 8%, 12%): 5.8%, 2.8%, 3.0%, 3.1% respectively (for the first method of calculation)
    (TE)95 (at 5%, 6.5%, 8%, 12%): 0.23, 0.13, 0.20, 0.34 respectively (for the second method of calculation)

    Study Details for the HLC-723G8

    The provided document describes the performance characteristics and studies conducted for the Tosoh Automated Glycohemoglobin Analyzer HLC-723G8.

    1. Sample size used for the test set and the data provenance:

      • Interference Studies:
        • Various numbers of samples for each potential interferent, tested at two HbA1c levels (e.g., Acetylated Hb, Albumin, etc., were tested at certain concentrations, some up to 1000 mg/dL to 5000 mg/dL).
        • Hemoglobin Variant Interference: 10 samples for each variant (HbS, HbC, HbE, HbD, HbF, HbA2), tested at two HbA1c levels (approximately 6% and ≥8%).
        • Data Provenance: Not explicitly stated, but clinical samples are implied as "known concentrations of HbA1c" were spiked. The samples are likely clinical specimens from an unspecified region, retrospectively used for these studies.
      • Precision Study: Four concentrations of unaltered EDTA whole blood specimens. Tested with 3 lots and 3 instruments.
        • Each specimen/reagent combination had 2 replicates in a single run, 2 times a day for 20 nonconsecutive days, totaling 40 runs and 80 determinants per data set. This implies a large number of measurements, but the exact number of unique patient blood specimens for the "four concentrations" is not specified.
        • Data Provenance: Unaltered EDTA whole blood specimens. Likely clinical samples, but country of origin/retrospective/prospective is not specified.
      • Correlation (Method Comparison) Study:
        • Sample Size: 120 unaltered EDTA whole blood specimens.
        • Data Provenance: "Frozen specimens were utilized for this study." Implies retrospective samples, but country of origin is not specified.
        • Sample Distribution: The 120 samples covered a range of HbA1c levels:
          • ≤ 5%: 5 samples (4.2%)
          • 5 – 6%: 15 samples (12.5%)
          • 6 – 6.5%: 30 samples (25.0%)
          • 6.5 – 7%: 30 samples (25.0%)
          • 7 – 8%: 20 samples (16.7%)
          • 8 – 9%: 10 samples (8.3%)
          • 9%: 10 samples (8.3%)

    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

      • For this in vitro diagnostic device, "ground truth" is established by a reference method/device, not human experts.
      • Interference Study: Ground truth was based on "known concentrations of HbA1c" (likely measured by a standardized reference method) and evaluated against a "variance greater than the assigned value x 1.00 ± 5%".
      • Hemoglobin Variant Interference Study: Ground truth was established by "reference values" for %HbA1c and IFCC mmol/mol, likely from a reference method.
      • Correlation Study: The comparative method was the Primus ultra2, analyzed by a "secondary NGSP reference laboratory." The Primus ultra2 is an established HPLC system for HbA1c. The qualification of individuals operating this reference laboratory is not specified, but its designation as an "NGSP reference laboratory" implies adherence to high standards for HbA1c measurement.
    3. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

      • Not applicable in this context. Adjudication by multiple human experts is typically used for subjective assessments (e.g., image interpretation). For quantitative diagnostic devices, performance is typically assessed against a 'gold standard' reference method.
    4. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

      • Not applicable. This is an automated in vitro diagnostic device, not an AI-assisted human reader system.
    5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

      • Yes, this is a standalone device. The entire submission describes the algorithm-only performance of the Automated Glycohemoglobin Analyzer HLC-723G8. It is designed to quantitatively measure HbA1c without human intervention in the measurement process itself, beyond sample loading and general operation.
    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc):

      • The ground truth for performance evaluation was based on reference methods and certified standards.
        • For correlation/comparison, the Primus ultra2 (analyzed by an NGSP reference laboratory) served as the comparative method.
        • The device's assigned HbA1c values are certified via the National Glycohemoglobin Standardization Program (NGSP) and are traceable to both the International Federation of Clinical Chemistry (IFCC) and the Diabetes Control and Complications Trial (DCCT). This indicates that the ground truth is based on established and internationally recognized standardization programs and reference measurement procedures for HbA1c.
    7. The sample size for the training set:

      • The document does not explicitly mention a "training set" in the context of machine learning or AI. This is a traditional HPLC system. Its development likely involved internal optimization and validation studies, but these are not described as distinct "training" and "test" sets in the way an AI/ML algorithm would. The linearity of the assay was "previously established" under a prior 510(k) (K071132).
    8. How the ground truth for the training set was established:

      • As a traditional HPLC system, there isn't a "training set" in the common understanding of AI/ML. The device's calibration and method development would rely on known standards and reference materials, traceable to NGSP/IFCC, to ensure accurate and reproducible measurements. The "ground truth" for developing such a system would involve calibrated reference materials with known HbA1c concentrations.
    Ask a Question

    Ask a specific question about this device

    Page 1 of 2