K Number
K123131
Date Cleared
2013-02-08

(127 days)

Product Code
Regulation Number
862.1825
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

ST AIA-PACK 25-OH Vitamin D is designed for in vitro diagnostic use only for the quantitative measurement of total 25-hydroxyvitamin D (25-OH vitamin D) in human serum, Na heparinized or EDTA plasma on TOSOH AIA System Analyzers. The Tosoh ST AIA-PACK 25-OH Vitamin D is intended as an aid in the determination of Vitamin D sufficiency.

The ST AIA-PACK 25-OH Vitamin D Calibrator Set is intended for In Vitro Diagnostic Use Only for the calibration of the ST AIA-PACK 25-OH Vitamin D assay.

The AIA-PACK 25-OH Vitamin D Control Set is.intended for IN VITRO DIAGNOSTIC USE ONLY for performing quality control procedures with the ST AIA-PACK 25-OH Vitamin D assay.

Device Description

The ST AIA-PACK 25-OH Vitamin D is a one-step delayed competitive enzyme immunoassay which, after sample pretreatment, is performed entirely in the ST AIA-PACK 25-OH Vitamin D test cup. Sample pretreatment reagents (containing sodium hydroxide) disassociate 25-OH vitamin D from its binding proteins in the test sample. 25-0H vitamin D present in the pretreated sample is bound to 25-OH vitamin D-specific monoclonal antibody immobilized on magnetic beads. After that, the enzyme-labeled 25-OH vitamin D is added to the reaction mixture. The enzyme-labeled 25-0H vitamin D competes with 25-0H vitamin D for binding to the antibody on magnetic beads in the reaction mixture.

After the second incubation, the magnetic beads are washed to remove the unbound enzymelabeled 25-OH vitamin D and are then incubated with a fluorogenic substrate, 4methylumbelliferyl phosphate (4MUP). The amount of enzyme-labeled 25-OH vitamin D that binds to the beads is inversely proportional to the 25-OH vitamin D concentration in the test sample. A standard curve is constructed, and unknown 25-OH vitamin D concentrations are calculated using this curve.

AI/ML Overview

Here's a breakdown of the acceptance criteria and the study details for the ST AIA-PACK 25-OH Vitamin D device, based on the provided 510(k) summary:

1. Table of Acceptance Criteria and Reported Device Performance

The document doesn't explicitly state "acceptance criteria" for precision, linearity, or correlation. Instead, it presents the results of these performance characteristics studies. The implied acceptance is that the reported performance is acceptable for the intended use and demonstrates substantial equivalence to the predicate device. For the purpose of this table, I will use the "Differences" table from the substantial equivalence section as a proxy for some acceptance criteria where direct comparisons were made to the predicate. For other performance characteristics, the reported results are presented.

Performance CharacteristicAcceptance Criteria (Implied/Predicate Comparison)Reported Device Performance
Assay TechnologyFluorescence Immunoassay (or comparable)Fluorescence Immunoassay
Incubation TimeShorter than predicate (110 min)10 minute cycle
Reference RangeComparable to predicate (9.0 - 37.6 ng/mL)10.8 to 54.75 ng/mL
Limit of DetectionComparable to predicate (1.5 ng/mL)2.6 ng/mL
Detection RangeComparable to predicate (4.8 - 100.0 ng/mL)4.0 ng/mL - 120.0 ng/mL
Calibrator FormatLyophilized (or comparable to predicate)Lyophilized
Calibrator Range0-100 ng/mL (or comparable to predicate)0-165 ng/mL
Controls FormatLyophilized (or comparable to predicate)Lyophilized
Controls LevelsSimilar to predicate (e.g., 2 levels)Two levels at approximately 20 and 80 ng/mL
Intra-assay Precision (CV%)Not explicitly stated, typically <10-15%Ranged from 1.2% to 7.3%
Total Precision (CV%)Not explicitly stated, typically <15-20%Ranged from 2.0% to 7.4%
Linearity RangeN/A (demonstrated)4 ng/mL to 150 ng/mL
Method Comparison (CorrelationSlope ~1, Intercept ~0, R > 0.95 (typical)Slope: 0.934, Intercept: 2.53, R: 0.944 (Deming)
Matrix Comparison (Serum vs Na Heparin)Slope ~1, Intercept ~0, R > 0.95 (typical)Slope: 0.993, Intercept: -0.390, R: 0.995 (Deming)
Matrix Comparison (Serum vs EDTA Plasma)Slope ~1, Intercept ~0, R > 0.95 (typical)Slope: 1.041, Intercept: 0.091, R: 0.994 (Deming)
Cross-reactivityMinimal for non-target analytes (<10-20%)25-OH D2: 101.1%, 25-OH D3: 99.2%, 3-epi 25-OH D2: 131.3%, 3-epi 25-OH D3: 107.7%, 24,25-(OH)2 D2: 5.2%, others <0.1-18.1%
InterferenceRecovery within 100 +/- 10%Hemoglobin, bilirubin, lipemia, protein, ascorbic acid, EDTA.2K, Sodium Heparin, Rheumatoid factor did not interfere.
Limit of Blank (LoB)N/A (determined)1.6 ng/mL
Limit of Detection (LoD)N/A (determined)2.6 ng/mL
Limit of Quantification (LoQ)N/A (determined, 20% CV)2.9 ng/mL
Reportable RangeN/A (defined)4.0 - 120 ng/mL

2. Sample Size Used for the Test Set and Data Provenance

  • Precision Studies (Intra-assay and Total):
    • Three levels of pooled and spiked Na heparinized plasma, EDTA plasma, and serum specimens.
    • 2 replicates per run, 2 times a day for 20 non-consecutive days for each specimen/reagent/instrument combination.
    • Total of 40 runs and 80 determinants for each data set.
    • Three data sets at each level for each specimen type.
    • Provenance: Not explicitly stated, but typically these samples are prepared in-house or sourced from commercial vendors for method validation. It is not indicated if they are from a specific country or if they are retrospective/prospective clinical samples.
  • Linearity:
    • Not specified, but linearity studies typically involve diluting samples to create a range of concentrations.
    • Provenance: Not specified.
  • Correlation (Method Comparison):
    • 156 unaltered serum specimens.
    • Provenance: Not explicitly stated, but implied to be clinical human serum samples. No country of origin is mentioned, and it's not specified if they are retrospective or prospective.
  • Correlation (Matrix Comparison - Serum vs Na Heparinized Plasma):
    • 115 unaltered specimens.
    • Provenance: Not explicitly stated, but implied to be clinical human samples.
  • Correlation (Matrix Comparison - Serum vs EDTA Plasma):
    • 115 unaltered specimens.
    • Provenance: Not explicitly stated, but implied to be clinical human samples.
  • Cross Reactivity:
    • Known quantities of cross-reactants spiked into serum specimens.
    • Provenance: Not explicitly stated, but typically involves commercially available purified substances.
  • Reference Ranges:
    • 233 samples.
    • Provenance: Not explicitly stated, but implied to be human samples from a normal population.
  • Interference:
    • Human specimens with added interfering substances (hemoglobin, bilirubin, triglycerides, human albumin, ascorbic acid, EDTA.2K, Sodium Heparin, Rheumatoid factor).
    • Provenance: Not specified, but implied to be human biological samples.
  • Limit of Detection and Limit of Quantification:
    • LoB: 60 measurements of 3 different blank specimens.
    • LoD: 10 measurements of 6 low-level samples.
    • LoQ: 10 measurements of 18 low-level samples.
    • Provenance: Not specified, but likely prepared in-house or commercially sourced.

3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

This device is an in vitro diagnostic (IVD) assay for quantitative measurement of a biomarker. For such devices, "ground truth" is typically established by:

  • Measurement by a validated reference method (e.g., the predicate device in the method comparison study).
  • Gravimetric preparation for linearity and analytical sensitivity.
  • Spiking studies for cross-reactivity and interference.

There is no mention of experts (like radiologists) being used to establish a subjective "ground truth" for the test set, as would be common for diagnostic imaging or AI-assisted diagnosis devices. The "ground truth" is the actual concentration of 25-OH Vitamin D as measured by the comparative method or by known analytical preparation.

4. Adjudication Method for the Test Set

Not applicable. As this is a quantitative chemical assay, not a subjective diagnostic interpretation, there is no need for expert adjudication methods like 2+1 or 3+1. The results are numerical values obtained from the instrument.

5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study Was Done

No, an MRMC comparative effectiveness study was not done. This type of study is relevant for diagnostic devices where human readers interpret results (e.g., medical images), and an AI might assist them. The ST AIA-PACK 25-OH Vitamin D is an automated in vitro diagnostic assay, so human interpretation is not involved in the primary measurement.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

Yes, the performance characteristics studies (precision, linearity, correlation, cross-reactivity, interference, LoD/LoQ) represent the standalone algorithm only performance of the ST AIA-PACK 25-OH Vitamin D assay. The device itself performs the quantitative measurement without human intervention beyond sample loading and initiation.

7. The Type of Ground Truth Used

The ground truth for the various studies was established through:

  • Comparative Method: For method comparison, the DiaSorin 25-Hydroxyvitamin D 125I RIA Kit served as the comparative method.
  • Known Concentrations/Preparations:
    • For linearity, samples were prepared with a known concentration range.
    • For LoD/LoQ, samples were prepared at low known concentrations.
    • For cross-reactivity, specific concentrations of potentially interfering substances were spiked into serum.
    • For interference, known quantities of interfering substances were added to human specimens.
  • Statistical Analysis: Reference intervals were determined using statistical analysis of a population of samples.

8. The Sample Size for the Training Set

The document does not explicitly mention a "training set" in the context of machine learning or AI algorithm development. This device is a traditional immunoassay, not an AI/ML-based device. Therefore, the concept of a training set as understood in AI is not directly applicable. The device's performance is based on its chemical reactions and detection system, which are designed and validated through analytical studies.

9. How the Ground Truth for the Training Set Was Established

Since there is no "training set" in the AI/ML sense for this traditional immunoassay, this question is not applicable. The assay's "ground truth" is inherent to its analytical design and validation against reference methods and known concentrations.

{0}------------------------------------------------

r

. <.

!や

·

.

K123131

FEB 0 8 2013

:

510(k) Summary

.

and the comments of the country

ST AIA-PACK 25-OH Vitamin D

1 of 13

{1}------------------------------------------------

=123131

510(k) Summary

ST AIA-PACK 25-OH Vitamin D

FEB 0 8 2013

Date:

.

ייא נ

です

Submitter:

Contact Person:

Device Name: Classification:

Device Name: Classification

Device Name: Classification

Predicate Device:

October 3, 2012 Amended: January 7, 2013

Tosoh Bioscience, Inc 3600 Gantz Road Grove City, OH 43123

Robert L. Wick Regulatory Specialist 6000 Shoreline Ct., Ste. 101 South San Francisco, CA 94080 Phone: 650-636-8117 650-636-8121 Fax: Email: Robert.Wick@Tosoh.com

ST AIA-PACK 25-OH Vitamin D Class II MRG Clinical Chemistry 21 CFR 862.1825

ST AIA-PACK 25-OH Vitamin D Calibrator Set Class II JIT Clinical Chemistry 21 CFR 862.1150

AIA-PACK 25-OH Vitamin D Control Set Class II JJX Clinical Chemistry 21 CFR 862.1660

K983617 DiaSorin Inc. 25-Hydroxyvitamin D 1261 RIA Kit

{2}------------------------------------------------

..

. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

510(k) Summary

ST AIA-PACK 25-OH Vitamin D

According to the requirements of 21 CFR 807.92, the following information provides sufficient detail to understand the basis for a determination of substantial equivalence.

Device Description:

The ST AIA-PACK 25-OH Vitamin D is a one-step delayed competitive enzyme immunoassay which, after sample pretreatment, is performed entirely in the ST AIA-PACK 25-OH Vitamin D test cup. Sample pretreatment reagents (containing sodium hydroxide) disassociate 25-OH vitamin D from its binding proteins in the test sample. 25-0H vitamin D present in the pretreated sample is bound to 25-OH vitamin D-specific monoclonal antibody immobilized on magnetic beads. After that, the enzyme-labeled 25-OH vitamin D is added to the reaction mixture. The enzyme-labeled 25-0H vitamin D competes with 25-0H vitamin D for binding to the antibody on magnetic beads in the reaction mixture.

After the second incubation, the magnetic beads are washed to remove the unbound enzymelabeled 25-OH vitamin D and are then incubated with a fluorogenic substrate, 4methylumbelliferyl phosphate (4MUP). The amount of enzyme-labeled 25-OH vitamin D that binds to the beads is inversely proportional to the 25-OH vitamin D concentration in the test sample. A standard curve is constructed, and unknown 25-OH vitamin D concentrations are calculated using this curve.

The following products are required to use the ST AIA-PACK 25-OH Vitamin D P/N 025234 :

ST AIA-PACK 25-OH Vitamin D Calibrator Set025334
ST AIA-PACK 25-OH Vitamin D Pretreatment Set025734
AIA-PACK 25-OH Vitamin D Control Set025434

Device Intended Use:

ST AIA-PACK 25-OH Vitamin D is designed for in vitro diagnostic use only for the quantitative measurement of total 25-hydroxyvitamin D (25-OH vitamin D) in human serum, Na heparinized or EDTA plasma on TOSOH AIA System Analyzers. The Tosoh ST AIA-PACK 25-OH Vitamin D is intended as an aid in the determination of Vitamin D sufficiency.

Calibrators:

The ST AIA-PACK 25-OH Vitamin D Calibrator Set is intended for In Vitro Diagnostic Use Only for the calibration of the ST AIA-PACK 25-OH Vitamin D assay.

Controls:

The AIA-PACK 25-OH Vitamin D Control Set is.intended for IN VITRO DIAGNOSTIC USE ONLY for performing quality control procedures with the ST AIA-PACK 25-OH Vitamin D assay. ·

{3}------------------------------------------------

Substantial Equivalence:

.

Comparison between the Tosoh ST AIA-PACK 25-OH Vitamin D and DiaSorin 25-Hydroxyvitamin D 1251 RIA Kit

Similarities

:

・・

ParameterST AIA-PACK 25-OH VitaminDDiaSorin 25-Hydroxyvitamin D125I RIA Kit
Intended useST AIA-PACK 25-OH Vitamin Dis designed for in vitro diagnosticuse only for the quantitativemeasurement of total 25-hydroxyvitamin D (25-OHvitamin D) in human serum, Naheparinized or EDTA plasma onTOSOH AIA System Analyzers.The Tosoh ST AIA-PACK 25-OH Vitamin D is intended as anaid in the determination ofVitamin D sufficiency.This kit is intended for thequantitative determination of the25-hydroxy vitamin D (25-OH-D)and other hydroxylated vitamin Dmetabolites in human serum orplasma to be used in theassessment of vitamin Dsufficiency. Assay results shouldbe used in conjunction with otherclinical and laboratory data toassist the clinician in makingindividual patient managementdecisions in an adult population.
Specimen typeEDTA plasma, Na heparinizedplasma or serumSerum or plasma

Calibrator Set

ParameterST AIA-PACK 25-OH VitaminD Calibrator Set25-OH-D3 Calibrators forDiasorin 125I RIA Kit
Intended useThe ST AIA-PACK 25-OHVitamin D Calibrator Set isintended for IN VITRODIAGNOSTIC USE ONLY forthe calibration of the ST AIA-PACK 25-OH Vitamin D assay.The 25-OH-D3 Calibrators isused to calibrate the Diasorin 125IRIA Kit.
MatrixHuman serumHuman serum
Number of Calibrators66

Control Set

ParameterAIA-PACK 25-OH Vitamin DControl Set25-OH-D Controls for Diasorin125I RIA Kit
Intended useThe AIA-PACK 25-OH VitaminD Control Set is intended for INVITRO DIAGNOSTIC USEONLY for performing qualitycontrol procedures with the STAIA-PACK 25-OH Vitamin Dassay.The 25-OH-D Controls is used forperforming quality control withthe Diasorin 125I RIA Kit.
MatrixHuman serumHuman serum
Number of levels22

{4}------------------------------------------------

Differences

ﺮ ﺍﻟﻤﺮﺍﺟﻊ

ParameterST AIA-PACK 25-OH VitaminDDiaSorin 25-Hydroxyvitamin D125I RIA Kit
Assay TechnologyFluorescence ImmunosassayRadioimmunoassay
Incubation Time10 minute cycle110 Minute Cycle
Reference range10.8 to 54.75 ng/mL9.0 - 37.6 ng/mL
Limit of detection2.6 ng/mL1.5 ng/mL
Assay Range4.0 ng/mL - 120.0 ng/mL4.8 ng/mL - 100.0 ng/mL

Calibrator Set

ParameterST AIA-PACK 25-OH VitaminD Calibrator Set25-OH-D3 Calibrators forDiasorin 125I RIA Kit
FormatLyophilizedReady to use liquid
Calibrator Range0-165 ng/mL0-100 ng/mL
ParameterAIA-PACK 25-OH Vitamin DControl Set25-OH-D Controls for Diasorin125I RIA Kit
FormatLyophilizedReady to use liquid
Controls LevelsTwo levels at approximately 20and 80 ng/mL of 25-OH vitaminDTwo levels at the low normal (9.0ng/mL) and high normal (37.6ng/mL)

PERFORMANCE CHARACTERISTICS

Precision

The precision for ST AIA-PACK 25-OH Vitamin D was determined based on guidance from CLSI Protocol EP5-A2.

Within- run precision was assessed by assaying three levels of pooled and spiked Na heparinized plasma, EDTA plasma and serum specimens with each lot number and instrument. Estimates of total and within-run precision were obtained from measurements of 2 replicates in a single run, 2 times a day for 20 non-consecutive days for each specimen/reagent/ instrument combination. This equaled a total of 40 runs and 80 determinants for each data set. There were three (3) data sets at each level for each specimen type.

{5}------------------------------------------------

:

イ..

ﻗﺮﺭ ﺍﻟﻤﺴﺎﺑﻘﺎﺕ ﺍﻟﻤﺴﺘﻘﻠﺔ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪ

·

Intra-assay (within run) Precision

SampleMean(ng/mL)StandardDeviation(ng/mL)Coefficient ofVariation(%)
Serum A116.01.27.3
Serum B136.01.02.7
Serum C194.11.51.6
Serum A221.80.73.3
Serum B244.41.63.6
Serum C298.01.92.0
Serum A321.80.94.3
Serum B346.51.83.9
Serum C3103.72.12.0
Na Hep Plasma A115.70.84.8
Na Hep Plasma B132.90.82.4
Na Hep Plasma C175.41.21.6
Na Hep Plasma A221.51.04.5
Na Hep Plasma B247.01.22.6
Na Hep Plasma C2102.81.91.9
Na Hep Plasma A321.81.15.2
Na Hep Plasma B350.01.63.1
Na Hep Plasma C3108.91.81.7
EDTA Plasma A113.70.74.9
EDTA Plasma B134.70.92.7
EDTA Plasma C171.11.21.7
EDTA Plasma A224.40.83.3
EDTA Plasma B247.40.92.0
EDTA Plasma C2105.21.51.4
EDTA Plasma A325.61.14.2
EDTA Plasma B350.31.01.9
EDTA Plasma C3110.91.31.2

ST AIA-PACK 25-OH Vitamin D

6 of 13

. :

{6}------------------------------------------------

Total Precision

:

ﻨﺔ .

11

SampleMean(ng/mL)StandardDeviation(ng/mL)Coefficient ofVariation(%)
Serum A116.01.27.4
Serum B136.01.33.7
Serum C194.12.22.3
Serum A221.80.83.8
Serum B244.41.53.3
Serum C298.02.02.0
Serum A321.81.35.8
Serum B346.51.83.9
Serum C3103.72.42.3
Na Hep Plasma A115.71.06.4
Na Hep Plasma B132.91.44.1
Na Hep Plasma C175.42.22.9
Na Hep Plasma A221.51.04.4
Na Hep Plasma B247.01.53.3
Na Hep Plasma C2102.82.72.7
Na Hep Plasma A321.81.15.2
Na Hep Plasma B350.02.04.0
Na Hep Plasma C3108.93.33.0
EDTA Plasma A113.70.96.6
EDTA Plasma B134.71.33.9
EDTA Plasma C171.11.92.7
EDTA Plasma A224.41.04.1
EDTA Plasma B247.41.42.9
EDTA Plasma C2105.22.32.2
EDTA Plasma A325.61.24.7
EDTA Plasma B350.31.63.2
EDTA Plasma C3110.92.82.5

Controls Precision

:

Intra-assay (within run) Precision

Mean(ng/mL)Pooled SD(ng/mL)CV(%)
Level 120.00.793.9
Level 279.11.632.1

·

:

{7}------------------------------------------------

. .

  • প্
Mean(ng/mL)Pooled SD(ng/mL)CV(%)
Level 120.01.165.8
Level 279.12.222.8

Linearity:

The linearity for ST AIA-PACK 25-OH Vitamin D was determined, based on guidance from CLSI protocol EP6-A. The linearity was measured on the AIA-2000 instrument and has been demonstrated to be linear from 4 ng/mL to 150 ng/mL.

Correlation

The methods comparison study was determined based on guidance from (EP9-A2).

  • a. Method Comparison
    A total of 156 unaltered serum specimens were assayed in singleton utilizing the ST AIA-PACK 25-OH Vitamin D assay on the AIA-2000 analyzer and the 25-Hydroxyvitamin D 1251 RIA Kit (Diasorin). The regression analysis for the correlation between the alternate method (x) and the ST AIA-PACK 25-OH Vitamin D (y) is as follows:
DemingRegular
Slope:0.934 (0.884 to 0.984)0.885 (0.836 to 0.935)
Intercept:2.53 (0.75 to 4.31)3.97 (2.21 to 5.73)
Corr Coef (R):0.944
Result Ranges:Diasorin 4.4 to 103.0 ng/mLTosoh 6.2 to 103.2 ng/mL

*95% Confidence Intervals are shown in parentheses

b. Matrix Comparison

The correlation between serum (x) and Na heparinized plasma (y) on the ST AIA-PACK 25-OH Vitamin D was carried out using 115 unaltered specimens.

DemingRegular
Slope:0.993 (0.974 to 1.012)0.988 (0.978 to 1.007)
Intercept:-0.390 (-1.13 to 0.36)-0.20 (-0.94 to 0.55)
Corr Coef (R):0.995
Result Ranges:Na Heparin 11.3 to 112.8 ng/mLSerum 11.0 - 111.1 ng/mL

*95% Confidence Intervals are shown in parentheses

{8}------------------------------------------------

.

. →

The correlation between serum (x) and EDTA Plasma (y) on the ST AIA-PACK 25-OH Vitamin D was carried out using 115 unaltered specimens.

DemingRegular
Slope:1.041 (1.020 to 1.062)1.034 (1.013 to 1.056)
Intercept:0.091 (0.09 to 1.73)1.14 (0.032 to 1.96)
Corr Coef (R):0.994
Result Ranges:EDTA 11.6 to 111.9 ng/mLSerum 11.0 - 111.1 ng/mL

*95% Confidence Intervals are shown in parentheses

Cross Reactivity

.

The following cross reactants were tested on the ST AIA-PACK 25-OH Vitamin D assay by spiking known quantities of the cross reactant in serum specimen.

Cross-reactantsCross-reactivity*[mol%]
25-OH Vitamin D2 (30 ng/mL)101.1
25-OH Vitamin D3 (158.2 ng/mL)99.2
3-epi 25-OH Vitamin D2 (30 ng/mL)131.3
3-epi 25-OH Vitamin D3 (30 ng/mL)107.7
24,25-(OH)2 Vitamin D2 (100 ng/mL)5.2
24,25-(OH)2 Vitamin D3 (100 ng/mL)18.1
Vitamin D2 (1000 ng/mL)0.5
Vitamin D3 (1000 ng/mL)<0.1
1, 25-(OH)2 Vitamin D2 (1.0 ng/mL)<0.1
1, 25-(OH)2 Vitamin D3 (1.0 ng/mL)<0.1
Paricalcitol (2.0 ng/mL)<0.1

{9}------------------------------------------------

Conclusion:

The ST AIA-PACK 25-OH Vitamin D recovers >/= 100 % for both 25-OH Vitamin D2 and 3epi- 25-OH Vitamin D3 at 30 ng/mL.

The ST AIA-PACK 25-OH Vitamin D recovers approximately 100 % for 25-OH Vitamin D3 at - 158.2 ng/mL.

The ST AIA-PACK 25-OH Vitamin D had a cross reactivity of 5.2% at 100 ng/mL of 24,25-Dihydroxyvitamin-D2.

The ST AIA-PACK 25-OH Vitamin D did not significantly cross react with 24, 25-(OH)2 Vitamin D3 at 100 ng/mL.

The ST AIA-PACK 25-OH Vitamin D did not significantly cross react with Vitamin D3 and Vitamin D2 at 1000 ng/mL.

The ST AIA-PACK 25-OH Vitamin D did not significantly cross react with 1, 25-(OH)2 Vitamin D3, 1, 25-(OH)2 Vitamin D2 and Paricalcitol at 2.0 ng/mL.

Reference Ranges

Number of Samples (n) 233 Reference Interval 10.8 - 54.75 ng/mL

The central 95% of the reference range was used to determine the reference interval.

Interference

Interference is defined, for the purposes of this study, with recovery outside of 100 +/- 10% of the known concentration of the specimen after the following substances are added to human specimens. Three studies were conducted using EDTA plasma, Na heparinized plasma and serum.´

  • Added hemoglobin (up to 11.4 mg/dL) does not interfere with the assay. .
  • Conjugated bilirubin (up to 18.4 mg/dL) and free bilirubin (up to 16.8 mg/dL) does not . interfere with the assay.
  • Lipemia, as indicated by triglyceride concentration (up to 357 mg/dL), does not interfere . with the assay.

{10}------------------------------------------------

  • Protein, as indicated by human albumin concentration (up to 8.5 g/dL), does not interfere . with the assay.
  • Ascorbic acid (up to 20 mg/dL) does not interfere with the assay. .
  • EDTA.2K (up to 10 mg/mL) does not interfere with the assay. .
  • Sodium Heparin (up to 100 U/mL) does not interfere with the assay. .
  • Rheumatoid factor (up to 500 IU/mL) does not interfere with the assay. .

Limit of Detection and Limit of Quantification

The LoB and LoQ of the ST AIA-PACK 25-OH Vitamin D was determined based on CLSI guideline EP17-A.

The LoB was determined by 60 measurements of 3 different blank specimens. The LoB was the value at the 95th percentile. In this case the LoB was determined to be 1.6 ng/mL.

The LoD was determined by 10 measurements of 6 low level samples. The sample range was chosen to be between LoB and 4xLoB and 3 lots of reagents were utilized. The LoD was determined to be 2.6 ng/mL.

The LoQ was determined by 10 measurements of 18 low level samples and the CV% for each sample was calculated. The samples were selected to be in the range of LoB to 4xLoB. A precision profile was plotted for concentration vs. the CV%. The functional sensitivity at 20% CV was calculated and determined as the LoQ. The LoQ was determined to be 2.9 ng/mL.

The reportable range for the assay is 4.0 - 120 ng/mL.

Standards:

NumberFDARecognitionNumberRevisionDateTitle
EP5-A27-11010/31/2005Evaluation of Precision Performance ofQuantitative Measurement Methods;Approved Guideline-Second Edition
EP6-A7-19303/18/2009Evaluation of the Linearity of QuantitativeMeasurement Procedures: a StatisticalApproach: Approved Guideline
C28-A37-20209/08/2009Defining, Establishing, and verifyingReference Intervals in the Clinical Laboratory;Approved Guideline - Third Edition
EP17-A7-19403/18/2009Protocols for Determination of Limits ofDetection and Limits of Quantitation;Approved Guideline
EP9-A27-9203/08/2004Method Comparison and Bias EstimationUsing Patient Samples; Approved Guideline-Second Edition
EP21-A7-17403/18/2009Estimation of Total Analytical Error for

ST AIA-PACK 25-OH Vitamin D

{11}------------------------------------------------

.

------------------------------------------------------------------------------------------------------------------------------------------------------------------------------A & R. HE MENSE & BE- Banks A - B - B - 0 - 0 - 0 - 0 - 0 - 0 - 0 - 0 - 0 - 0 - 0 - 0 - 0 - 0 - 0 - 0 - 0 - 0 - 0 - 0 - 0 - 0 - 0 - 0 - 0 - 0 - 0 - 0 - 0 - 0 - 0 - 0 - 0 - 0 - 0 - 0 - 0 - 0 - 0Canadian.marrier-----------------------------------------------------------------------------------------------------Company Child Child Child A. Market A.
.
------------------------------------------------------------------------------------------------------------------------------------------------------------------------------CONCOLOGICAL

ST AIA-PACK 25-OH Vitamin D

:

:

100 million in the state

.

:

.

.

{12}------------------------------------------------

Conclusion:

t.

ﺎﺭ ﺍﻟﻤﺮﺍﺟﻊ ﻓﻲ

The Tosoh Bioscience, Inc. ST AIA-PACK 25-OH Vitamin D is substantially equivalent to the DiaSorin Inc. (k)983617-Hydroxyvitamin D 1251 RIA Kit for in vitro diagnostic use only for the quantitative measurement of total 25-hydroxyvitamin D (25-OH vitamin D) in human serum, Na heparinized or EDTA plasma on TOSOH AIA System Analyzers. The Tosoh ST AIA-PACK 25-OH Vitamin D is intended as an aid in the determination of Vitamin D sufficiency

{13}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/13/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the seal is a stylized symbol resembling an abstract human figure or a caduceus, rendered in a simple, flowing line.

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

February 8, 2013

Tosoh Bioscience, Inc c/o Robert L. Wick 6000 Shoreline Court, Suite 101 South San Francisco, CA 94080

Re: K123131

Trade/Device Name: ST AIA-PACK 25-OH Vitamin D, ST AIA-PACK 25-OH Vitamin D Calibrator Set, AIA-PACK 25-OH Vitamin D Control Set Regulation Number: 21 CFR 862.1825 Regulation Name: Vitamin D test system

Regulatory Class: II

Product Code: MRG, JIT, JJX

Dated: January 08, 2013

Received: January 09, 2013

Dear Mr. Wick:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device; subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{14}------------------------------------------------

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Carol C. Benson for

Courtney H. Lias, Ph.D. Director Division of Chemistry and Toxicology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{15}------------------------------------------------

Indications for Use

510(k) Number (if known): K123131

Device Name:

ST AIA-PACK 25-OH Vitamin D

ST AIA-PACK 25-OH Vitamin D Calibrator Set

AIA-PACK 25-OH Vitamin D Control Set

Indications For Use:

ST AIA-PACK 25-OH Vitamin D is designed for in vitro diagnostic use only for the quantitative measurement of total 25-hydroxyvitamin D (25-OH vitamin D) in human serum, Na-heparinized plasma or EDTA plasma on TOSOH AIA System Analyzers. The Tosoh ST AIA-PACK 25-OH Vitamin D is intended as an aid in the determination of Vitamin D sufficiency.

The ST AIA-PACK 25-OH Vitamin D Calibrator Set is intended for in vitro diagnostic use only for the calibration of the ST AIA-PACK 25-OH Vitamin D assay.

The AIA-PACK 25-OH Vitamin D Control Set is intended for in vitro diagnostic use only for performing quality control procedures with the ST AIA-PACK 25-OH Vitamin D assay.

Prescription Use X (21 CFR Part 801 Subpart D)

And/Or

Over the Counter Use (21 CFR Part 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE; CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of In Vitro Diagnostic Device Evaluation and Safety (OIVD)

Yung W.Chan -S

Division Sign-Off Office of In Vitro Diagnostic Device Evaluation and Safety

K123131 510(k)

Page __ 1_ of __ 1_

§ 862.1825 Vitamin D test system.

(a)
Identification. A vitamin D test system is a device intended for use in clinical laboratories for the quantitative determination of 25-hydroxyvitamin D (25-OH-D) and other hydroxylated metabolites of vitamin D in serum or plasma to be used in the assessment of vitamin D sufficiency.(b)
Classification. Class II (special controls). Vitamin D test systems must comply with the following special controls:(1) Labeling in conformance with 21 CFR 809.10 and
(2) Compliance with existing standards of the National Committee on Clinical Laboratory Standards.