Search Results
Found 512 results
510(k) Data Aggregation
(261 days)
Elecsys sFlt-1 will be marketed with the following devices:
-
CalSet sFlt-1; secondary, calibrator; JIT, 862.1150
Elecsys PlGF will be marketed with the following devices: -
CalSet PlGF; secondary, calibrator; JIT, 862.1150
Immunoassays for the in vitro quantitative determination of the soluble fms like tyrosine kinase-1/placental growth factor (sFlt-1/PlGF) ratio in human serum.
The sFlt-1/PlGF ratio is indicated as an aid in the risk assessment of pregnant women, with a singleton pregnancy (23+0 to 34+6/7 weeks' gestation) hospitalized for hypertensive disorders of pregnancy (preeclampsia, chronic hypertension with or without superimposed preeclampsia, or gestational hypertension), to develop preeclampsia with severe features within two weeks from testing. The sFit-1/PlGF ratio should be used in conjunction with clinical assessment and routine laboratory testing.
The electrochemiluminescence immunoassay "ECLIA" is intended for use on cobas e immunoassay analyzers.
The Elecsys sFlt-1 and Elecsys PlGF assays employ a sandwich principle using electrochemiluminescence immunoassay "ECLIA" technology. The total duration of each assay is 18 minutes. Samples are incubated with biotinylated and ruthenium-labeled monoclonal antibodies specific to sFlt-1 or PlGF, forming a sandwich complex. Streptavidin-coated microparticles are added, binding the complex to the solid phase. The microparticles are magnetically captured, unbound substances are removed, and a voltage is applied to induce chemiluminescent emission, which is measured by a photomultiplier. Results are determined via a calibration curve generated by 2-point calibration and a master curve provided via the reagent barcode. The reagents for each assay are combined in a "rackpack".
Here's a breakdown of the acceptance criteria and study information for the Elecsys sFlt-1 and Elecsys PlGF assays, based on the provided document.
Acceptance Criteria and Device Performance
Acceptance Criteria Category | Specific Criteria | Reported Device Performance |
---|---|---|
Clinical Performance | Risk Assessment for Preeclampsia with Severe Features within two weeks from testing (Cutoff: >38): high Negative Predictive Value (NPV) and acceptable Positive Predictive Value (PPV) for pregnant women with singleton pregnancy (23+0 to 34+6/7 weeks' gestation) hospitalized for hypertensive disorders of pregnancy. | Overall Intended Use Population (N=556): |
- Sensitivity: 91.40% (95% CI: 86.41, 95.00)
- Specificity: 77.30% (95% CI: 72.68, 81.47)
- NPV (ratio ≤ 38): 94.70% (95% CI: 91.54, 96.94)
- PPV (ratio > 38): 66.93% (95% CI: 60.77, 72.68) |
| Non-Clinical Performance | Precision: Low coefficients of variation (CV) for repeatability (within-run) and intermediate precision (within-laboratory). | Elecsys PlGF (N=84 per sample type): - Repeatability CV: 1.0% - 5.7%
- Intermediate precision CV: 1.4% - 9.9%
Elecsys sFlt-1 (N=84 per sample type): - Repeatability CV: 0.9% - 2.4%
- Intermediate precision CV: 1.7% - 3.7%
Ratio (N=84 per sample type): - Repeatability CV: 1.1% - 4.9%
- Intermediate precision CV: 1.4% - 7.0% |
| | Linearity/Assay Reportable Range: Measurements are linear across the claimed measuring range. | - Elecsys sFlt-1: 80-85000 pg/mL (claimed range) - Elecsys PlGF: 10-5400 pg/mL (claimed range)
(Study concludes measurements are linear across these ranges) |
| | Limit of Blank (LoB): ≤ 2 pg/mL for PlGF and 26.4 mg/dL can cause up to 10% decrease in ratio. - Hemoglobin, Intralipid, Rheumatoid Factors, Biotin (up to 1200 ng/mL): No significant interference reported (implies within acceptable limits though quantitative data not listed).
- Common Drugs (15 tested): No interference.
- Additional Substances (13 tested): No interference.
- Heparin: Interference with Elecsys PlGF for concentrations > 500 U/L. |
| | Analytical Specificity/Cross-Reactivity: Highly specific for sFlt-1 and PlGF, with minimal cross-reactivity with related substances. | - sFlt-1 cross-reactivity:
Ask a specific question about this device
(418 days)
immunological test system
- 21 CFR 862.1660 Quality Control material (assayed and unassayed)
- 21 CFR 862.1150
B·R·A·H·M·S™ CgA II KRYPTOR™ is an automated immunofluorescent assay using Time-Resolved Amplified Cryptate Emission (TRACE™) technology for quantitative determination of Chromogranin A concentration in human serum.
B·R·A·H·M·S™ CgA II KRYPTOR™ is to be used in conjunction with other clinical methods as an aid in monitoring of disease progression during the course of disease and treatment in patients with gastroentero-pancreatic neuroendocrine tumors (GEP-NETs, grade 1 and grade 2).
The B-R-A-H-M-S CgA II KRYPTOR assay is based on the formation of a complex comprised of a Chromogranin A (CgA) analyte "sandwiched" between two monoclonal mouse anti-CgA antibodies. One of the antibodies (537/H2) is directed at the epitope AA124–144 and labelled with DiSMP cryptate, the other antibody (541/E2) binds to AA280-301 and is labelled with Alexa Fluor®647.
The measurement principle is based on a non-radiative energy transfer from a donor (cryptate) to an acceptor (Alexa Fluor™647) when they are part of an immunocomplex (TRACE technology (Time-Resolved Amplified Cryptate Emission)).
The fluorescent signal is proportional to the concentration of the analyte to be measured.
With this principle B-R-A-H-M-S CgA II KRYPTOR is a homogenous one-step immunoassay for the quantification of CgA II in human serum. The linear direct measuring range of the assay is from 20-3,000 ng/mL, going up to 1,000,000 ng/mL with automated dilution. Results can be retrieved after a 29 min incubation time.
Here's an analysis of the acceptance criteria and study findings for the B.R.A.H.M.S CgA II KRYPTOR device, based on the provided FDA 510(k) summary:
Acceptance Criteria and Reported Device Performance
Note: The provided document primarily describes analytical performance criteria and clinical performance measures (sensitivity, specificity) rather than explicit "acceptance criteria" in a pass/fail format for clinical decision-making. However, the sensitivity and specificity values obtained from the clinical study serve as the reported device performance against which implicit clinical acceptance would be judged. The analytical performance metrics are generally presented as numerical results meeting industry standards (CLSI guidelines).
Acceptance Criteria Category | Specific Metric | Acceptance Threshold (Implicit/Standard) | Reported Device Performance |
---|---|---|---|
Analytical Performance | Precision (Repeatability CV) | Generally low CVs for quantitative assays (e.g., 3,000 ng/mL, extending range up to 1,000,000 ng/mL. | |
Interference | Bias ≤ 10% for common endogenous and exogenous interfering substances. | Substances evaluated were found not to affect test performance (bias ≤ 10%) at clinically relevant concentrations. | |
Cross-Reactivity | Low cross-reactivity with structurally similar substances. | Between -21.6% - 0.03% (for various CgA fragments and related proteins). | |
Clinical Performance | Clinical Sensitivity (for tumor progression based on ΔCgA > 50% & >100 ng/mL cutoff) | Sufficient to aid monitoring, balancing with specificity given the intended use (aid, not standalone diagnosis). | 34.4% (95% CI: 23.2% - 45.5%) |
Clinical Specificity (for tumor progression based on ΔCgA > 50% & >100 ng/mL cutoff) | Sufficient to aid monitoring, balancing with sensitivity given the intended use (aid, not standalone diagnosis). | 93.4% (95% CI: 90.2% - 96.0%) | |
Positive Predictive Value (PPV) | Relevant for clinical utility given prevalence. | 57.9% (95% CI: 40.5% - 73.6%) | |
Negative Predictive Value (NPV) | Relevant for clinical utility given prevalence. | 84.3% (95% CI: 79.3% - 89.1%) |
Study Details:
-
Sample size used for the test set and the data provenance:
- Clinical Study (for Sensitivity and Specificity): 153 adult GEP-NET patients (grade 1 and 2), with 459 total observations (likely reflecting multiple monitoring visits per patient). The study was described as a prospective study.
- Clinical Cut-off Derivation: 102 patients with diagnosed well-differentiated G1 and G2 GEP-NETs. This was a retrospective, bicentric observational pilot study.
- Reference Range Determination: 206 samples from self-declared healthy individuals. Data provenance is USA.
- Analytical studies: Various sample sizes were used, often involving replicates of pooled or individual human serum samples. For example, LoQ used 420 total replicates from 7 different pools of human serum samples.
- Provenance for analytical samples: Not explicitly stated but generally implied to be from human subjects, for instance, "human serum samples".
-
Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience):
- For the clinical study, tumor progression was classified by RECIST 1.1 criteria. This implies that experts (typically radiologists or oncologists) were involved in interpreting imaging (CT/MRI) according to these established criteria to determine the ground truth for tumor progression.
- The document does not specify the direct number of experts or their specific qualifications (e.g., "radiologist with 10 years of experience"). However, RECIST 1.1 is an internationally recognized standard for evaluating cancer treatment response based on imaging, implying adjudication by qualified personnel.
-
Adjudication method (e.g. 2+1, 3+1, none) for the test set:
- The ground truth for tumor progression in the clinical studies was established using RECIST 1.1 criteria based on standard imaging (CT/MRI).
- The document does not explicitly state an adjudication method like "2+1" or "3+1" for discordant interpretations if multiple readers were involved in RECIST assessment. However, RECIST guidelines themselves are designed to standardize interpretation, and clinical trials often employ independent central review or consensus panels for definitive RECIST ratings, though this specific detail is not provided here.
-
If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
- No, an MRMC comparative effectiveness study was not done. This device is an in vitro diagnostic (IVD) for quantitative determination of Chromogranin A concentration in human serum, intended to be used in conjunction with other clinical methods as an aid in monitoring. It is not an AI-assisted imaging device or a device that directly aids human readers in interpreting images.
-
If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:
- This is an IVD assay, which functions as a "standalone" measurement of a biomarker in serum. The results are generated by the automated instrument (B.R.A.H.M.S KRYPTOR compact PLUS analyzer) without direct human interpretation of the measurement itself. However, the device's output (CgA concentration) is explicitly stated to not be used for standalone diagnosis or monitoring but "in conjunction with other clinical methods." So while the analytical measurement is standalone, the clinical interpretation for decision-making is not.
-
The type of ground truth used (expert consensus, pathology, outcomes data, etc.):
- For the clinical performance evaluation (sensitivity and specificity for tumor progression), the ground truth was imaging-based tumor assessment using RECIST 1.1 criteria. This is a form of expert assessment based on a standardized methodology, often relying on radiologists and oncologists to interpret imaging studies.
-
The sample size for the training set:
- This document describes an IVD device submission, not a machine learning/AI device. Therefore, the concept of a "training set" for an algorithm in the typical AI sense does not directly apply. The development and validation of the assay itself would have involved numerous samples for optimization and establishment of analytical performance characteristics, but these are not referred to as a "training set" here.
-
How the ground truth for the training set was established:
- As addressed above, the concept of a "training set" in the context of machine learning/AI is largely inapplicable here. The development of the assay's analytical characteristics (e.g., linearity, precision, detection limits) would be established through standard laboratory practices and reference materials, for which "ground truth" is defined by known concentrations or established analytical methods.
Ask a specific question about this device
(382 days)
The Anti-Tg CalSet is regulated under product code JIS and 21 CFR 862.1150 and is exempt from Premarket
Immunoassay for the in vitro quantitative determination of antibodies to thyroglobulin in human serum and plasma. The anti-Tg autoantibodies determination is used as an aid in the detection of autoimmune thyroid diseases in conjunction with other laboratory and clinical findings.
The electrochemiluminescence immunoassay "ECLIA" is intended for use on the cobas e 411 immunoassay analyzer.
The Elecsys Anti-Tg immunoassay makes use of a competitive test principle using biotinylated human antigen and monoclonal human anti-Tg antibodies labeled with a ruthenium complex. The Elecsys Anti-Tg immunoassay is intended for the quantitative determination of antibodies to thyroglobulin in human serum and plasma. It is intended for use on the cobas e immunoassay analyzers.
Results are determined via a calibration curve which is instrument-specifically generated by 2 point calibration and a master curve provided via the reagent barcode or e barcode.
The provided text describes the performance evaluation of the Elecsys Anti-Tg immunoassay, a diagnostic device, and its acceptance criteria. Here's a breakdown of the requested information based on the text:
1. A table of acceptance criteria and the reported device performance
The document does not explicitly present a single table labeled "Acceptance Criteria" with corresponding "Reported Device Performance" in a direct side-by-side format. Instead, it describes various performance evaluations and states whether "All predefined acceptance criteria was met" for each. However, we can reconstruct a table based on the provided details for the non-clinical performance evaluation.
Performance Characteristic | Acceptance Criteria (Implied) | Reported Device Performance |
---|---|---|
Precision | All predefined acceptance criteria met (specific numerical criteria not detailed in this section for repeatability/intermediate precision and lot-to-lot reproducibility) | Repeatability (within-run precision) and Intermediate Precision (within-laboratory precision) measured according to CLSI guideline EP05-A3. Accepted for all samples. |
Lot-to-lot Reproducibility measured according to CLSI guideline EP05-A3 using three reagent lots. Accepted. | ||
Analytical Sensitivity | ||
Limit of Blank (LoB) | LoB claim in labeling set to 9 IU/mL (implied acceptance) | Determined according to CLSI EP17-A2. LoB claim in labeling will be set to 9 IU/mL. |
Limit of Detection (LoD) | LoD claim in labeling set to 10 IU/mL (implied acceptance) | Determined according to CLSI EP17-A2. LoD claim in labeling will be set to 10 IU/mL. |
Limit of Quantitation (LoQ) | LoQ claim in labeling set to 15 IU/mL (implied acceptance) | Determined according to CLSI EP17-A2. LoQ claim in labeling will be set to 15 IU/mL. |
Linearity | Confirmed to support the measuring range of 15 - 4000 IU/mL (implied acceptance) | Assessed according to CLSI EP06-Ed2, study design B, using weighted linear regression analysis. Linearity confirmed to support measuring range of 15 - 4000 IU/mL. |
Endogenous Interferences | All predefined acceptance criteria met (specific numerical criteria not detailed) | Evaluated for Biotin, Lipemia, Hemoglobin, Bilirubin, Rheumatoid Factor, Tg. All predefined acceptance criteria met. Proposed labeling claims: Biotin ≤ 1200 ng/mL, Lipemia ≤ 2000 mg/dL, Hemoglobin ≤ 600 mg/dL (for conc. ≤ 115 IU/mL), Bilirubin ≤ 66 mg/dL, Rheumatoid Factor ≤ 300 IU/mL, Tg ≤ 100 ng/mL. |
Analytical Specificity/Cross-Reactivity | No cross-reaction with Anti-TPO detected (implied acceptance) | Cross-reactivity study conducted with anti-TPO. No cross-reaction with Anti-TPO detected. |
Exogenous Interferences | All predefined acceptance criteria met for all drugs tested (implied acceptance) | Evaluated 17 commonly and 14 specially used pharmaceutical compounds. All predefined acceptance criteria met, and no interference observed. |
Sample Matrix Comparison | Results within specification and support use of specified matrices (implied acceptance) | Compared values from serum, K2-EDTA, and K3-EDTA plasma. Results were within specification and support the use of Serum (standard or separating gel), K2-EDTA, and K3-EDTA plasma. |
Method Comparison to Predicate | Not explicitly stated as pass/fail; presented as regression results | Compared to predicate device (K053426) using 129 human serum samples. |
Linear Regression: y = 0.905x + 48.0, r = 0.990 | ||
Passing Bablok: y = 0.974x + 1.72, τ = 0.930 | ||
Reagent Stability | ||
Reagent On-board Stability | 42 days (6 weeks) on-board (implied acceptance) | Tested on one cobas e 411 analyzer. Stable for up to 42 days (6 weeks). |
Reagent Stability After First Opening | 42 days (6 weeks) after first opening (implied acceptance) | Tested on one cobas e 411 analyzer. Stable for up to 42 days (6 weeks) when stored at 2-8°C. |
Calibration Stability | ||
Lot Calibration Stability | Calibration every 28 days (1 month) (implied acceptance) | Tested on one cobas e 411 analyzer. Recommended calibration frequency of every 28 days (1 month). |
On-board Calibration Stability | 7 days without new calibration (implied acceptance) | Tested on one cobas e 411 analyzer. Stable for up to 7 days without a new calibration. |
2. Sample sizes used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)
- Test Set Sample Sizes:
- Precision: Not explicitly stated, but includes "Human serum 1-5" and "PC THYRO1-2" (presumably replicates for each, as per CLSI EP05-A3 which requires sufficient replicates).
- Lot-to-lot Reproducibility: "three reagent lots" (number of samples per lot not specified).
- Analytical Sensitivity (LoB, LoD, LoQ): Not explicitly stated, determined according to CLSI EP17-A2 which has sample size recommendations.
- Linearity: "Six dilution series" using "native human serum samples and sample pools" (number of samples/pools not specified).
- Endogenous Interferences: Not explicitly stated per substance, but mentions "Six endogenous substances."
- Analytical Specificity/Cross-Reactivity: Not explicitly stated (for anti-TPO).
- Exogenous Interferences: "17 commonly and 14 specially used pharmaceutical compounds" (number of samples not stated).
- Sample Matrix Comparison: "blood from 13 donors" (tested across serum, K2-EDTA, K3-EDTA plasma, and serum separation tubes from 3 manufacturers).
- Method Comparison to Predicate: "total of 129 human serum samples."
- Data Provenance: The document does not specify the country of origin for the data or whether the studies were retrospective or prospective. It is a "510(k) Summary" for an FDA submission, reporting on laboratory performance studies. Given they are "non-clinical performance evaluation," these are typically controlled laboratory studies.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)
This device is an in-vitro diagnostic (IVD) immunoassay. The ground truth for such devices is established through analytical testing against reference materials, established methods, and clinical samples with known characteristics, not typically by expert consensus in the same way as an imaging AI. The "ground truth" here is the precise concentration or presence/absence of the analyte (thyroglobulin antibodies) as determined by the study's reference method or expected values for standards/controls.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
Not applicable for this type of IVD analytical performance study. Adjudication methods like 2+1 or 3+1 are common in clinical trials or imaging studies where expert readers interpret results, but not for direct quantitative measurements from an immunoassay. The acceptance criteria are based on statistical analysis of quantitative results (e.g., precision, linearity, recovery, regression).
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
Not applicable. This is an immunoassay for determining antibody levels, not an imaging device or an AI intended to assist human readers. Hence, no MRMC study was performed, and human reader improvement with AI assistance is not relevant to this device's evaluation.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
This entire non-clinical performance evaluation section (4. NON-CLINICAL PERFORMANCE EVALUATION
) describes the standalone performance of the Elecsys Anti-Tg immunoassay (a device, not an AI algorithm). The measurements are performed by the "cobas e 411 immunoassay analyzer," which acts as the "algorithm" or automated system. There's no human "in the loop" for the direct measurement results themselves, though human operators are involved in running the tests and interpreting the results in a clinical setting.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)
The ground truth for the analytical performance studies (precision, linearity, sensitivity, interferences, stability) is based on:
- CLSI guidelines: The studies adhere to specific Clinical and Laboratory Standards Institute (CLSI) guidelines (e.g., EP05-A3, EP17-A2, EP06-Ed2), which define how such analytical characteristics are determined using reference materials, spiked samples, and statistical methods.
- Reference Standards/Materials: Implied in sections like "Traceability/Standardization" against the NIBSC 65/93 Standard, and the use of calibrators (Anti-Tg CalSet) and controls (PreciControl ThyroAB).
- Known Sample Characteristics: For linearity, samples with varying known concentrations are typically used. For interference studies, samples spiked with known interferents are used.
- Predicate Device Comparison: For method comparison, the predicate device's results serve as a comparative reference.
8. The sample size for the training set
Not applicable in the context of an immunoassay. This device is an in-vitro diagnostic test kit (reagents) used on an existing analyzer. It does not involve a "training set" in the machine learning sense. The "development" or "training" of such a diagnostic involves optimizing the chemical and biological components of the assay (reagents, antibodies, detection method) and calibrating the system across a range of known concentrations.
9. How the ground truth for the training set was established
As there is no "training set" in the AI/machine learning sense, this question is not applicable. The development process for an immunoassay involves extensive research and development to create reagents that accurately quantify the target analyte. Calibration is done using reference materials with assigned values, and the assay's performance characteristics (as detailed in section 4) are then rigorously validated.
Ask a specific question about this device
(323 days)
II Classification Name: Clinical Chemistry Test Systems: Calibrator Governing Regulation: 21 CFR § 862.1150
The HAVAb IgG II assay is a chemiluminescent microparticle immunoassay (CMA) used for the qualitative detection of IgG antibody to hepatitis A virus (IgG anti-HAV) in human adult and pediatric (4 through 21 years) serum (collected in serum and serum separator tubes) and plasma (collected in sodium heparin, lithium heparin separator, dipotassium EDTA, and tripotassium EDTA tubes) from patients with signs and symptoms or at risk for hepatitis A on the Alinity i system.
The HAVAb IgG II assay is used to determine the immune status of individuals to hepatitis A virus (HAV) infection. Warning: This assay has not been cleared for use in screening blood, plasma, or tissue donors. This assay camot be used for the diagnosis of acute HAV infection.
Assay performance characteristics have not been established when the HAVAb IgG II assay is used in conjunction with other hepatitis assays.
The HAVAb IgG II assay is a chemiluminescent microparticle immunoassay (CMIA) used for the qualitative detection of IgG antibody to hepatitis A virus (IgG anti-HAV) in human adult and pediatric (4 through 21 years) serum (collected in serum and serum separator tubes) and plasma (collected in sodium heparin, lithium heparin separator, dipotassium EDTA, and tripotassium EDTA tubes) from patients with signs and symptoms or at risk for hepatitis A on the Alinity i system. The kit includes reagents (Microparticles, Conjugate, Assay Diluent), Calibrator, and Controls. The assay is an automated, two-step immunoassay.
Considering the provided document, the device described is an in vitro diagnostic (IVD) assay (HAVAb IgG II) for the qualitative detection of IgG antibody to hepatitis A virus (IgG anti-HAV). The FDA 510(k) summary focuses on establishing substantial equivalence to a predicate device, not on meeting specific acceptance criteria in the context of an AI/ML medical device's performance evaluation against clinical ground truth.
Therefore, many of the requested criteria (like ground truth establishment by experts, adjudication methods, multi-reader multi-case studies, and separate training/test sets with their ground truth establishment) are generally not applicable to the performance claims made for this in vitro diagnostic device in this FDA submission. The document describes analytical and clinical performance studies typical for an IVD, focusing on agreement with a predicate device and reproducibility/precision, rather than predictive performance of an AI model against a complex clinical endpoint established by human experts.
Based on the provided text, here's a breakdown of the information that is applicable and a clear indication where the requested information is not present or relevant to this type of device submission:
1. A table of acceptance criteria and the reported device performance
The document doesn't explicitly state "acceptance criteria" in a table format that would typically be found for an AI/ML model for diagnostic imaging (e.g., target sensitivity/specificity values). Instead, it presents performance data for comparison to a predicate device and for reproducibility. The implicit "acceptance criterion" for a 510(k) is demonstrating "substantial equivalence" to a legally marketed predicate device.
However, we can infer some performance metrics presented as evidence of equivalence:
Performance Metric | Reported Device Performance (HAVAb IgG II) | Predicate Device Performance (ARCHITECT HAVAB-G) (for comparison, not acceptance criteria) |
---|---|---|
PPA (Positive Percent Agreement) with Predicate: | ||
- Increased Risk of HAV Infection Population (n=250) | 96.75% (95% CI: 91.94%, 98.73%) | N/A (this is agreement with the predicate) |
- Signs and Symptoms of Hepatitis Infection (n=499) | 95.39% (95% CI: 92.42%, 97.24%) | N/A |
- Pediatric Population (n=105) | 100.00% (95% CI: 95.91%, 100.00%) | N/A |
NPA (Negative Percent Agreement) with Predicate: | ||
- Increased Risk of HAV Infection Population (n=250) | 98.43% (95% CI: 94.44%, 99.57%) | N/A |
- Signs and Symptoms of Hepatitis Infection (n=499) | 98.97% (95% CI: 96.34%, 99.72%) | N/A |
- Pediatric Population (n=105) | 93.33% (95% CI: 70.18%, 98.81%) | N/A |
Within-Laboratory Precision (20-Day) | ||
- High Negative Panel 1 (0.71 S/CO) | SD: 0.028 (Range 0.026-0.045) | N/A (Predicate's Within-Lab Precision: 0.029-0.050 SD for = 1.00 S/CO) |
- Negative Control (0.09 S/CO) | SD: 0.015 (Range 0.011-0.035) | N/A |
- Positive Control (2.19 S/CO) | %CV: 2.9 (Range 2.5-4.0) | N/A |
System Reproducibility (Multi-site) | ||
- High Negative Panel A (0.66 S/CO) | SD: 0.053 | N/A (Predicate's Reproducibility: 0.023-0.116 SD) |
- Low Positive Panel B (1.32 S/CO) | %CV: 5.2 | N/A (Predicate's Reproducibility: 4.6-10.8 %CV) |
- Negative Control (0.11 S/CO) | SD: 0.046 | N/A |
- Positive Control (2.26 S/CO) | %CV: 4.7 | N/A |
2. Sample sizes used for the test set and the data provenance
-
Clinical Performance Test Set (Agreement with Predicate):
- Individuals at Increased Risk of HAV Infection: n=250
- Individuals with Signs and Symptoms of Hepatitis Infection: n=499
- Pediatric Population: n=105
- Total for Agreement Study: 250 + 499 + 105 = 854 specimens.
- Data Provenance: Prospective multi-center study.
- Increased Risk of HAV: collected in California, Colorado, Florida, Illinois, Massachusetts, North Carolina, and Texas.
- Signs and Symptoms of Hepatitis: collected in California, Colorado, Florida, Illinois, Massachusetts, and Texas.
- Pediatric Population: collected in the US (California and Massachusetts) and Belgium.
-
Precision/Reproducibility Test Sets:
- Within-Laboratory Precision: n=80 per sample/control for a representative combination (tested over 20 days, 2 replicates/day). Total tested for this study was 4 reagent/calibrator/instrument combinations.
- System Reproducibility: n=360 per sample/control (tested at 3 sites, with 4 replicates/run, 2 runs/day, 5 days).
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
- Not applicable / Not stated. For this IVD device, the "ground truth" for the clinical performance study was the result produced by the legally marketed predicate device (ARCHITECT HAVAB-G). This is a common practice for 510(k) submissions for IVDs. There were no human experts adjudicating results for the purpose of establishing a "ground truth" beyond what the predicate device determined.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
- Not applicable. As the ground truth was the predicate device's result, no human adjudication method (like 2+1, 3+1) was used or described.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
- Not applicable. This is an in vitro diagnostic assay, not an imaging AI device intended to assist human readers. Therefore, an MRMC study and evaluation of human reader improvement with AI assistance were not performed.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
- Yes, in spirit. The device (HAVAb IgG II assay) functions as a standalone test; it's an automated immunoassay that generates a qualitative result (Reactive or Nonreactive) without human interpretation in the loop influencing the output beyond sample collection and instrument operation. Its performance was assessed directly against the predicate device's results.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)
- The "ground truth" for the clinical performance comparison was the results from a legally marketed predicate device (ARCHITECT HAVAB-G assay). In essence, the performance of the new device was compared to the established performance of the predicate. This is a common form of "truth" in demonstrating substantial equivalence for IVDs.
8. The sample size for the training set
- Not explicitly stated in terms of a "training set" for model development. This is an immunoassay, not an AI/ML model where a distinct training dataset size is typically reported. The document describes analytical verification and clinical performance studies, not model training.
9. How the ground truth for the training set was established
- Not applicable. As this is not an AI/ML device in the sense of a machine learning model requiring a training set with a ground truth established by experts for supervised learning, this information is not provided. The development process for an immunoassay does not involve "training data" in the same way an AI/ML model does.
Ask a specific question about this device
(209 days)
Classification: Class II, 510(k) Exempt Classification Name: Calibrator Governing Regulation: 21 CFR § 862.1150
The ARCHITECT CMV IgG assay is a chemiluminescent microparticle immunoassay (CMIA) used for the qualitative detection of IgG antibodies to cytomegalovirus in human serum, serum separator, and plasma tubes (lithium heparin, lithium heparin separator, and tripotassium EDTA) on the ARCHITECT i System.
The ARCHITECT CMV IgG assay is to be used as an aid in the diagnosis of infection with cytomegalovirus and as an aid in the determination of serological status to cytomegalovirus in individuals including women of child-bearing age.
The ARCHITECT CMV IgG assay has not been cleared for use in screening blood, plasma, or tissue donors.
The ARCHITECT CMV IgG assay is a chemiluminescent microparticle immunoassay (CMIA) for the qualitative detection of IgG antibodies to cytomegalovirus. The reagent kit contains microparticles coated with CMV virus lysate, murine anti-human IgG acridinium-labeled conjugate, and assay diluent. The assay is performed on the ARCHITECT i System. The presence or absence of anti-CMV IgG is determined by comparing the chemiluminescent relative light unit (RLU) in the reaction to a cutoff RLU from an active calibration.
Here's an analysis of the acceptance criteria and study information for the ARCHITECT CMV IgG device, based on the provided text:
1. Table of Acceptance Criteria and Reported Device Performance
The acceptance criteria are not explicitly stated as numerical targets in the document. However, based on the non-clinical and clinical performance summaries, we can infer performance goals for the device's accuracy and precision. The "Predicate Device" (ADVIA Centaur CMV IgG Assay, K181213) serves as a benchmark for substantial equivalence.
Performance Metric | Acceptance Criteria (Inferred from Study Design & Predicate Equivalence) | Reported Device Performance (ARCHITECT CMV IgG) |
---|---|---|
Within-Laboratory Precision (Repeatability) | Low variability; %CV values typically +0.6 AU/mL) noted for Triglycerides at 2475 mg/dL on 4.0 AU/mL sample. | |
Drugs/Other Substances: No significant interference observed for Acetaminophen, Ascorbic Acid, Beta Carotene, Biotin, Cidofovir, Diphenhydramine, Folic Acid, Foscarnet, Gangciclovir, Ibuprofen, Valganciclovir. | ||
Analytical Specificity (Other Conditions) | Low rate of false positives/reactives with specimens from individuals with unrelated medical conditions. | 1 reactive result out of 187 specimens tested across various medical conditions (Toxoplasmosis (IgG) specimen was reactive). 1 equivocal parvovirus B19 (IgG) specimen. |
CDC Panel Agreement (Positive) | High positive percent agreement, ideally >90% | 100% (91.59%, 100%) |
CDC Panel Agreement (Negative) | High negative percent agreement, ideally >90% | 92.11% (78.62%, 98.34%) |
CDC Panel Agreement (Overall) | High overall percent agreement, ideally >90% | 96.25% (89.43%, 99.22%) |
Clinical Percent Agreement (Positive) - Routine Order | High positive percent agreement with comparator, ideally >90% | 97.7% (96.1%, 98.7%) |
Clinical Percent Agreement (Negative) - Routine Order | High negative percent agreement with comparator, ideally >90% | 99.2% (97.3%, 99.8%) |
Clinical Percent Agreement (Positive) - Pregnant Females | High positive percent agreement with comparator, ideally >90% | 99.0% (94.5%, 99.8%) |
Clinical Percent Agreement (Negative) - Pregnant Females | High negative percent agreement with comparator, ideally >90% | 100.0% (96.4%, 100.0%) |
2. Sample Size Used for the Test Set and Data Provenance
- Within-Laboratory Precision (20-Day): 117-120 replicates per panel/control across 3 reagent lots/calibrator lot combinations.
- Analytical Specificity (Interference): Not explicitly stated, but each substance was tested at 3 analyte levels.
- Analytical Specificity (Other Conditions): 187 specimens.
- CDC Panel Agreement: 80 samples (either CMV IgG negative or CMV IgG positive). Data provenance is the Centers for Disease Control and Prevention (CDC). The document doesn't specify if this was retrospective or prospective, but the description "masked, characterized serum panel" suggests it was a curated, retrospective panel.
- Clinical Percent Agreement:
- Routine Order: 591 specimens collected in the US and 198 specimens collected outside of the US.
- Pregnant Females: 200 specimens collected in the US.
- Further Evaluation: 4 specimens (3 routine order, 1 pregnant female) with equivocal/grayzone results by the comparator assay.
- Data provenance for the clinical study is multi-site (Indianapolis Indiana, Lewisville Texas, and Palo Alto California) for the US samples, and unspecified international locations for the "outside of the US" routine order specimens. This was a clinical study (method comparison), suggesting prospective collection relative to the comparison against the investigational assay.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts
There is no information provided about experts being used to establish a ground truth for the test set in the conventional sense (e.g., for image interpretation).
- For the CDC Panel Agreement, the CDC provided "their result interpretation for each sample," implying the CDC's own characterization (likely by reference methods/consensus) served as the ground truth. No specific number or qualifications of experts are given.
- For the Clinical Percent Agreement, the ground truth was established by a "current, FDA-cleared, commercially available anti-CMV IgG assay" (the comparator assay), essentially establishing a reference standard from an already approved diagnostic device. For 4 equivocal/grayzone samples, "consensus testing" with "2 additional current, FDA-cleared, commercially available anti-CMV IgG assays" was used. The document does not mention human experts establishing ground truth for these studies.
4. Adjudication Method for the Test Set
- CDC Panel Agreement: The results were compared directly to CDC's result interpretation. No explicit adjudication method is described beyond this direct comparison.
- Clinical Percent Agreement: The primary comparison was against a single comparator assay. For specific equivocal/grayzone cases (4 specimens), a form of adjudication was used where the result was "negative by consensus testing" or "negative based on the consensus result from the comparator assay and 2 additional current, FDA-cleared, commercially available anti-CMV IgG assays." This implies a form of multi-comparator consensus for ambiguous cases, but not necessarily human expert adjudication in the traditional sense.
5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
No MRMC comparative effectiveness study was done. This device is an automated in vitro diagnostic (IVD) assay (chemiluminescent microparticle immunoassay), not an AI-powered diagnostic imaging device or tool that assists human readers. Therefore, the concepts of human readers, AI assistance, and effect size in improving human reader performance are not applicable.
6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done
The device itself, the ARCHITECT CMV IgG assay on the ARCHITECT i System, is a standalone algorithm/device for qualitative detection. Its performance is evaluated entirely as an automated system without human interpretation in the loop influencing the direct output of IgG antibody detection. The "human-in-the-loop" would be the clinician interpreting the result provided by the device (e.g., Reactive, Nonreactive, Grayzone/Equivocal) in the context of a patient's overall clinical picture, but not in generating the result itself.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)
- CDC Panel Agreement: Ground truth was based on CDC's own "result interpretation" of their characterized serum panel. This likely represents a highly robust reference method or internal consensus.
- Clinical Percent Agreement: Ground truth was primarily established by a "current, FDA-cleared, commercially available anti-CMV IgG assay" (comparator assay). For a few ambiguous cases, consensus results from multiple FDA-cleared comparator assays were used. This is a form of reference method comparison.
8. The Sample Size for the Training Set
The document does not explicitly mention a "training set" in the context of machine learning or AI. As this is a traditional immunoassay, it undergoes development and validation using various biological samples. However, no specific number is provided for samples used during the development or optimization phases that could be analogized to a training set. The clinical study samples described are specifically for performance evaluation (test set).
9. How the Ground Truth for the Training Set Was Established
As no explicit training set is defined (because it's not an AI/ML device), this question is not applicable. The assay's parameters (e.g., cutoff values) would have been established during product development using a large set of characterized samples and optimized for performance, but this is a different process than establishing ground truth for an AI training set.
Ask a specific question about this device
(486 days)
Controls Levels 1 and 2 ●
- 21 CFR 862.1660, Product Code JJX o
- Hemoglobin A1c Calibrator Set
- 21 CFR 862.1150
|
| Regulation # | 21 CFR 862.1373
21 CFR 864.7470
21 CFR 862.1150
- 21 CFR 862.1150
The Tosoh Automated Glycohemoglobin Analyzer HLC-723G8 is intended for in vitro diagnostic use for the measurement of % hemoglobin A1c (HbA1c) (DCCT/NGSP) and mmol/mol hemoglobin A1c (IFCC) in venous whole blood specimens using ion-exchange high-performance liquid chromatography (HPLC). This test is an aid in diagnosis of diabetes and identifying patients who may be at risk for developing diabetes, and for monitoring of long-term blood glucose control in individuals with diabetes mellitus.
The Tosoh Automated Glycohemoglobin Analyzer HLC-723G8 is an automated High-Performance Liguid Chromatography (HPLC) system that separates and reports stable hemoglobin A1c (sA1c) percentage in venous whole blood. The operational portion of the G8 is composed of a sampling unit, liquid pump, degasser, column, detector, microprocessors, sample loader, smart media card, operation panel, and a printer. The Tosoh Automated Glycohemoglobin Analyzer HLC-723G8 uses ion-exchange HPLC for rapid, accurate, and precise separation of the stable form of HbA1c (sA1c) from other hemoglobin fractions. The G8 uses a non-porous cation exchange column and separates the hemoglobin components in the blood. Separation is achieved by utilizing differences in ionic interactions between the cation and exchange group on the column resin surface and the hemoglobin components in a step gradient elution. The hemoglobin fractions (designated as A1a. A1b. F. LA1c+, SA1c, A0, and, if present, H-V0, H-V2, H-V2 and H-V3) are subsequently removed from the column by performing a step-wise elution gradient using the varied salt concentrations in the Variant Elution Buffers HSi 1, 2 and 3. The peaks, H-V0, H-V1, H-V2 and H-V3 are typically presumptive HbAD, HbAS, HbAC and HbAE respectively. The software compares the retention times of hemoglobin fractions in a sample to the expected "windows of retention" and labels each fraction that correctly elutes within a defined expected window of retention. The software designates a hemoglobin fraction as POX (where X is the order of the peak as it elutes from the column) if it does not match a defined window of retention. All automated processes in the G8 are controlled by internal microprocessors, using software downloaded via a smart media card. The result report is printed and can be stored on the instrument. The data can be transmitted to a host computer through a bi-directional interface. The result report includes the sample ID, date, percentage and retention time of each fraction of hemoglobin, sA1c percentage and total A1 percentage, along with a chromatogram of the elution pattern of the hemoglobin fractions. If a sample contains a hemoglobin variant, the column elutes the fraction depending upon its charge.
The provided text describes the non-clinical performance testing of the Tosoh Automated Glycohemoglobin Analyzer HLC-723G8 (subject device) to support its substantial equivalence to a predicate device. This document focuses on the analytical performance of a diagnostic device rather than an AI/ML powered device, so some of the specific questions regarding AI/ML study design (e.g., number of experts, adjudication methods, MRMC studies) are not applicable.
Here's the information extracted from the document:
1. Acceptance Criteria and Reported Device Performance
The acceptance criteria are generally implied by the statement "All performance testing results met their pre-determined acceptance criteria." While explicit numerical acceptance criteria for each test are not listed in a consolidated table, the discussion throughout the "Summary of Non-Clinical Performance Testing" implicitly defines them through the methodology and results. For example, for precision/repeatability, the claim of "imprecision at ≤ 2%" was a pre-established criterion. Similarly, for hemoglobin variant interference, "Non-clinically significant interference was defined as
Ask a specific question about this device
(91 days)
JIT |
| Regulation Number | 21 CFR 862.1150
|
| Regulation Number | 21 CFR 862.1150
The Atellica® IM BRAHMS Procalcitonin (PCT) assay is for in vitro diagnostic use in the quantitative determination of procalcitonin in human serum and plasma (EDTA, lithium heparin, and sodium heparin) using the Atellica® IM Analyzer.
The Atellica IM BRAHMS PCT assay is intended for use, in conjunction with other laboratory findings and clinical assessments, as an aid in:
· The risk assessment of critically ill patients on their first day of ICU admission for progression to severe sepsis and septic shock.
• Assessing the cumulative 28-day risk of all-cause mortality for patients diagnosed with severe sepsis or septic shock in the ICU or when obtained in the emergency department or other medical wards prior to ICU admission, using percent change in PCT level over time.
· Decision-making on antibiotic therapy for patients with suspected or confirmed lower respiratory tract infections (LRTI) – defined as community-acquired pneumonia (CAP), acute bronchitis, and acute exacerbation of chronic obstructive pulmonary disease (AECOPD) - in an inpatient setting or an emergency department.
- · Decision-making on antibiotic discontinuation for patients with suspected or confirmed sepsis.
The Atellica IM BRAHMS PCT assay is a 2-site sandwich immunoassay using direct chemiluminescent technology that uses 3 mouse monoclonal antibodies specific for PCT. The first antibody, in the Lite Reagent, is a mouse monoclonal anti-PCT antibody labeled with acridinium ester. The second and third antibodies, in the ancillary reagent, are mouse monoclonal anti-PCT antibodies labeled with fluorescein. The immunocomplex formed with PCT is captured with mouse monoclonal anti-fluorescein antibody coupled to paramagnetic particles in the Solid Phase.
A direct relationship exists between the amount of PCT present in the patient sample and the amount of relative light units (RLUs) detected by the system.
The provided text describes the Siemens Healthcare Diagnostics Atellica IM B.R.A.H.M.S Procalcitonin (PCT) assay, an in vitro diagnostic device. The study presented focuses on the analytical and clinical performance of this assay, comparing it to a predicate device and evaluating its utility in specific clinical scenarios related to sepsis and respiratory tract infections.
Here's an analysis of the acceptance criteria and the study proving the device meets them:
Acceptance Criteria and Reported Device Performance
The acceptance criteria for this device are implicitly derived from the performance goals demonstrated in the analytical and clinical studies. Since this is a 510(k) submission, the primary goal is to show substantial equivalence to a legally marketed predicate device (B.R.A.H.M.S PCT sensitive KRYPTOR). The performance characteristics are presented as meeting industry standards (CLSI guidelines) and showing comparable results to the predicate.
Here's a table summarizing the analytical performance with implicit acceptance criteria (typically, these would be defined before the study in a protocol, often as a deviation tolerance from the predicate or a specific measure):
Performance Characteristic | Acceptance Criteria (Implicit/Standard Expectations) | Reported Device Performance (Atellica IM BRAHMS PCT) |
---|---|---|
Precision | Low variability (%CV) across different PCT concentrations, meeting CLSI EP05-A3 guidelines. | Repeatability (Within-Run): %CVs ranging from 1.2% to 9.7% for serum samples (0.05 - 19.14 ng/mL). Within-Lab (Total Precision): %CVs ranging from 2.1% to 12.1%. Modeling analysis showed Total Error of 4%-32% at various PCT levels. |
Linearity/Assay Measuring Range | Deviation from linear fit ≤ 10%; demonstrated linearity across the claimed measuring range. | Linearity demonstrated in the range of 0.03 – 63.24 ng/mL. Claimed measuring range is 0.04 – 50.00 ng/mL. |
Dilution Recovery | Recoveries close to 100% when samples are diluted manually and automatically. | Manual dilution recoveries: 96% to 102% (mean 99%). Auto-dilution recoveries: 92% to 107%. |
Hook Effect | No significant hook effect within the expected upper range of analyte concentrations. | No hook effect observed up to 2000 ng/mL; samples as high as 2000 ng/mL reported > 50.00 ng/mL. |
Detection Limits (LoB, LoD, LoQ) | LoB 0.95). | N=522 samples (range 0.06-49.20 ng/mL): Weighted Deming Regression: Slope = 1.02 (95% CI: 0.99-1.05), Intercept = -0.02 (95% CI: -0.03 to -0.01), r = 0.98. Passing & Bablok Regression: Slope = 1.06 (95% CI: 1.04-1.09), Intercept = -0.04 (95% CI: -0.06 to -0.03), r = 0.98. |
Method Comparison with Predicate Device (Concordance) | High positive and negative percent agreement (PPA/NPA) at clinical cutoffs (e.g., typically >90-95%). | At 0.1 ng/mL: PPA = 99.3%, NPA = 95.0%, Overall = 98.7%. |
At 0.25 ng/mL: PPA = 99.0%, NPA = 94.6%, Overall = 98.1%. | ||
At 0.50 ng/mL: PPA = 96.7%, NPA = 97.4%, Overall = 97.0%. | ||
At 2.0 ng/mL: PPA = 97.2%, NPA = 97.6%, Overall = 97.4%. | ||
Matrix Comparison | No significant matrix effect between different sample types (serum, plasma). | High correlation (r=1.00) and minor differences in regression equations between Serum vs. EDTA, Li Heparin, and Na Heparin plasma samples across the range of 0.05-44.72 ng/mL. |
Expected Values/Reference Interval | Establishment of normal reference interval. | 99th percentile for PCT value in normal healthy subjects (N=144) was 80%) significantly associated with 28-day mortality (Fisher's Exact Test p=0.009). Cox proportional hazards regression showed HR of 1.82 (95% CI: 1.14-2.89; p=0.012) for ΔPCT positive vs. negative. Consistent utility for risk stratification based on initial PCT levels and ICU disposition. |
Clinical Performance (Antibiotic Guidance) | Support for use in decision-making on antibiotic therapy and discontinuation. | Supported by systematic literature reviews and meta-analyses demonstrating reduction in antibiotic initiation and exposure without negative effects on mortality, complications, or length of stay. |
Detailed Study Information:
-
Sample sizes used for the test set and the data provenance:
- Analytical Performance:
- Precision: 5 contrived human serum samples, 80 replicates per sample.
- Linearity: 9 samples (prepared from high and low human serum pools).
- Dilution Recovery: 5 human serum samples.
- Detection Limits (LoD): 360 determinations (160 blank, 200 low-level replicates).
- Endogenous/Therapeutic Drug/Cross-Reactivity Interference: Not specified beyond "serum pools" and "serum and plasma."
- Heterophile Interference: HAMA and RF positive patient samples + control samples.
- Method Comparison: 623 native human serum samples with assigned values from the predicate device.
- Matrix Comparison: 51 matched specimen sets in 4 tube types (Serum, EDTA plasma, Lithium Heparin Plasma, Sodium Heparin Plasma).
- Expected Values/Reference Interval: 144 serum samples from normal subjects.
- Clinical Studies:
- 28-day Mortality: 858 adult patients (>18 years) diagnosed with severe sepsis or septic shock admitted to ICU. The analysis population comprised 598 subjects.
- Antibiotic Therapy/Discontinuation Support: Systematic literature reviews and meta-analyses:
- Study-level meta-analysis (Initiation/Discontinuation): 11 RCTs (4090 patients) published 2004-2016.
- Patient-level meta-analysis (Initiation): 13 RCTs (3142 patients) published 2004-2011.
- Patient-level meta-analysis (Discontinuation): 5 RCTs (598 patients) published 2008-2010.
- Data Provenance:
- Analytical Data: In-house or contracted lab studies (implied to be prospective given the detailed methodology descriptions).
- Clinical Data: "Banked specimens that were collected from subjects... and included as part of the intention-to-diagnose population from the BRAHMS MOSES study (DEN150009)." This indicates retrospective use of previously collected clinical trial data. The original MOSES study subjects were recruited across 13 investigational sites in the US. The meta-analyses for antibiotic guidance consolidated data from previously published RCTs.
- Analytical Performance:
-
Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
- For the analytical test sets, ground truth is established by standard laboratory methods and reference materials (e.g., recombinant PCT, serum pools), not human experts.
- For the clinical study, the ground truth for patient outcomes (e.g., severe sepsis/septic shock diagnosis, 28-day all-cause mortality, need for ICU care) was presumably established by the clinical assessments and medical records from the original BRAHMS MOSES study that provided the banked specimens. The document does not specify the number or qualifications of clinicians/experts who established these initial diagnoses and outcomes for the MOSES study, but it's implied to be standard clinical practice within a multi-site clinical trial.
- For the meta-analyses, the "ground truth" for antibiotic management decisions and patient outcomes (mortality, LOS, etc.) derived from the various RCTs where those decisions and outcomes were prospectively recorded by the clinical teams involved in those trials.
-
Adjudication method (e.g., 2+1, 3+1, none) for the test set:
- For the analytical studies, the "ground truth" values are quantitative measurements or characteristics of samples, typically not subject to human adjudication in the same way as, for example, image interpretation. Internal laboratory quality control and statistical methods (e.g., regression analysis, CV calculations) serve as "adjudication."
- For the clinical study, the data were obtained from previously collected "banked specimens" and "clinical assessments" from the BRAHMS MOSES study. The document does not describe a new adjudication process for this particular 510(k) submission's analysis of this retrospective data. The rigor of the original MOSES study's data collection and endpoint ascertainment would be relevant but is not detailed here.
- The meta-analyses relied on data from published RCTs; any adjudication would have occurred within the original trials.
-
If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
- No. This device is an in vitro diagnostic (IVD) assay that measures a biomarker (Procalcitonin) in human specimens. It is not an AI-assisted diagnostic imaging or pathology device that requires human "readers." Therefore, an MRMC study is not applicable and was not performed.
-
If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:
- Yes, implicitly. The performance characteristics (precision, linearity, detection limits, interference, method comparison) are "standalone" results of the assay itself, demonstrating its analytical performance when run on the Atellica IM Analyzer. The clinical studies demonstrate the performance of the assay's results when used in conjunction with "other laboratory findings and clinical assessments," but the assay itself generates the PCT value independently of human interpretation of that specific run.
-
The type of ground truth used (expert consensus, pathology, outcomes data, etc.):
- Analytical Studies: "Ground truth" for analytical studies is typically established by meticulously prepared reference materials, spike-in methods, and comparison to a well-validated predicate device using a large sample size of native patient samples.
- Clinical Studies:
- For the 28-day mortality prediction, the ground truth was outcomes data (survival status at 28 days) derived from a pre-existing clinical study (BRAHMS MOSES study).
- For antibiotic guidance, the insights were derived from meta-analyses of published Randomized Controlled Trials (RCTs). The "ground truth" here is the aggregated, prospectively collected patient outcomes (antibiotic use, duration, mortality, complications, LOS) from these trials where patients were managed either by standard care or PCT-guided protocols.
-
The sample size for the training set:
- This document describes a 510(k) premarket notification for an IVD assay, not an AI/ML device that typically requires a distinct "training set." The performance studies for an IVD device like this are primarily for validation and demonstrate clinical and analytical performance. There is no explicit "training set" in the context of machine learning.
- However, if one were to consider the development of the assay's methodologies, the calibration and control materials (e.g., PCT MCM) would be used for standardization and quality control, which could be seen as analogous to internal "training" or optimization data during product development, but no specific sample size for this is detailed as "training set."
-
How the ground truth for the training set was established:
- As noted above, the concept of a "training set" with ground truth establishment in the ML sense is not directly applicable to this type of IVD device submission. The assay is a chemical measurement system. Its "truth" is established via analytical validation showing its accuracy, precision, and comparability to reference methods/predicate devices.
Ask a specific question about this device
(263 days)
| |
| Regulation Number | 862.1150
QUANTA Flash Calprotectin is a chemiluminescent immunoassay for the quantitative determination of fecal calprotectin in extracted human stool samples. Elevated levels of fecal calprotectin, in conjunction with clinical findings and other laboratory tests, can aid in the diagnosis of inflammatory bowel disease (IBD) (ulcerative colitis and Crohn's disease), and in the differentiation of IBD from irritable bowel syndrome (IBS).
QUANTA Flash Calprotectin Calibrators are intended for use with the QUANTA Flash Calprotectin Reagents for the determination of fecal calprotectin levels in extracted stool samples. Each calibrator establishes a point of reference for the working curve that is used to calculate unit values.
QUANTA Flash Calprotectin Controls are intended for use with the QUANTA Flash Calprotectin Reagents for quality control in the determination of fecal calprotectin levels in extracted stool samples.
QUANTA Flash Calprotectin Extraction Buffer is intended for use with the QUANTA Flash Calprotectin Reagents as sample extraction solution.
The principle of the assay is chemiluminescent microparticle immunoassay, a variation of solid phase immunoassay. The QUANTA Flash® Calprotectin assay is designed to run on the BIO-FLASH® instrument. This platform is a fully automated closed system with continuous load and random access capabilities that automatically processes the samples, runs the assay and reports the results. It includes liquid handling hardware, luminometer and computer with software-user interface. The QUANTA Flash® Calprotectin assay utilizes a reagent cartridge format, which is compatible with the BIO-FLASH® instrument.
Calprotectin-specific capture antibodies are coated on to paramagnetic beads, which are stored in the reagent cartridge under conditions that preserve the antibody in its reactive state. Prior to use in the BIO-FLASH® system, the reagent pack containing all the necessary assay reagents is mixed thoroughly by being inverted several times. The sealed reagent tubes are pierced with the reagent cartridge lid, and the reagent cartridge is loaded onto the instrument. Reagents are calibrated when the lot is first used. A patient extracted stool sample is prediluted by the BIO-FLASH® with sample buffer in a disposable plastic cuvette. Small amounts of the diluted patient extracted stool, the beads, and the assay buffer are all combined into a second cuvette, and mixed. This cuvette is then incubated at 37°C. The beads are magnetized and washed several times. Isoluminol conjugated monoclonal antibodies are then added to the cuvette, and again incubated at 37°C. The beads are magnetized and washed repeatedly. The isoluminol conjugate is oxidized when Trigger 1 (Fe(II)coproporphyrin in sodium hydroxide solution) and Trigger 2 (urea-hydrogen peroxide in sodium chloride solution) are added to the cuvette, and the flash of light produced from this reaction is measured as Relative Light Units (RLU) by the BIO-FLASH® optical system. The measured RLU is proportional to the amount of bound isoluminol conjugate, which is in turn proportional to the amount of calprotectin antigen captured by the antibodies (anti-calprotectin polyclonal antibodies in this case) on the beads. For quantitation, the QUANTA Flash® Calprotectin will utilize a predefined lot specific Master Curve that is uploaded onto the instrument through the reagent cartridge barcode. The Master Curve is generated by Inova Diagnostics for each reagent pack lot with in-house Standards with assigned unit values (ng/mL). The RLU and assigned ng/mL values of the Standards are used to create a 4 parameter logistic curve. These four parameters are embedded in the reagent pack barcode. When the lot is used the first time, the Calibrators are run, and based on the results obtained on the Calibrators, an instrument specific Working Curve is created; The Working Curve is used to calculate units (ng/mL) based on RLU values obtained on each sample. The obtained ng/mL values will be converted to mg/kg by a calculation that takes into account the dilution of the samples. This unit conversion is calculated automatically by the software.
Here's an analysis of the provided text, extracting the requested information about acceptance criteria and the study proving device performance:
1. Table of Acceptance Criteria and Reported Device Performance
The device is an in vitro diagnostic (IVD) test system, so performance metrics like sensitivity and specificity are evaluated, alongside analytical performance criteria common for laboratory assays.
Performance Characteristic | Acceptance Criteria | Reported Device Performance |
---|---|---|
Precision (Within-Laboratory) | Total %CV: |
Ask a specific question about this device
(158 days)
|
| Common or Usual Name: | Calibrator, Multi-Analyte Mixture
per 21 CFR 862.1150
N Latex FLC kappa and lambda assays: In-vitro diagnostic reagents for the quantitative determination of free light chains (FLC), type kappa or type lambda, in human serum and EDTA plasma by means of particle-enhanced immunonephelometry using the BN Systems. FLC measurements are used as an aid in the diagnosis of multiple myeloma (MM) and amyloidosis (AL).
N FLC Supplementary Reagent: Supplementary reagent for the immunonephelometric determination of free light chains (FLC), type kappa and type lambda on BN Systems. A mixture of both supplementary reagents is used to suppress interference by rheumatoid factors and human anti-mouse antibodies (HAMA).
N FLC Standard SL: Establishment of reference curves for the determination of free light chains (FLC), type kappa and type lambda on the BN Systems.
N FLC Controls SL1 and SL2: The N FLC Controls SL1 and SL2 are for use as assayed accuracy controls in the determination of free light chains (FLC), type kappa and type lambda by immunonephelometry with the BN Systems.
The N Latex FLC (free light chain) assays are in vitro diagnostic reagents for the quantitative determination of free light chains, type kappa or type lambda, in human serum and EDTA plasma by means of particle-enhanced immunonephelometry using the BN™ II and BN ProSpec® Systems. Used in conjunction with other clinical and laboratory findings, FLC measurements are used as an aid in the diagnosis of multiple myeloma (MM) and amyloidosis (AL). Used in conjunction with the assay reagents, N FLC Standard SL is for use in the establishment of reference curves for the determination of free light chains, type kappa and type lambda on the BN™ II and BN ProSpec® Systems. The N FLC Control SL 1 and 2 products are for use as assayed accuracy controls and precision controls in the determination of free light chains, type kappa and type lambda by immunonephelometry with the BN™ II and BN ProSpec® Systems. The FLC test systems are based upon the principles of particle-enhanced immunonephelometry. Polystyrene particles coated with monoclonal antibodies to human free light chains, type kappa or lambda, respectively, are agglutinated when mixed with samples containing FLC. These aggregates scatter a beam of light passed through the sample. The intensity of the scattered light is proportional to the concentration of the respective protein in the sample. The result is evaluated by comparison with a standard of known concentration.
The provided text describes the Siemens N Latex FLC kappa and N Latex FLC lambda assays, along with their associated calibrators and controls. These devices are intended for the quantitative determination of free light chains (FLC) in human serum and EDTA plasma, used as an aid in diagnosing multiple myeloma (MM) and amyloidosis (AL).
Here's an analysis of the acceptance criteria and the study that proves the device meets them:
1. Table of Acceptance Criteria and Reported Device Performance
The acceptance criteria are generally implied by the performance characteristics presented in the study. For analytical performance, typical acceptance limits for precision (CV%), linearity, and interference are industry standards for IVD devices. For clinical performance, the reported sensitivity and specificity values against clinical diagnosis are the acceptance metrics.
Acceptance Criteria Category | Specific Metric | Acceptance Criteria (Implied/Standard) | Reported Device Performance and Remarks |
---|---|---|---|
Analytical Performance | Precision (Total CV%) | Typically |
Ask a specific question about this device
(264 days)
|
| Regulation Number | 862.1150
QUANTA Flash LKM-1 is a chemiluminescent immunoassay for the semi-quantitative determination of IgG anti-liver/ kidney microsome type 1 antibodies in human serum. The presence of anti-liver/kidney microsome type 1 antibodies, in conjunction with clinical findings and other laboratory tests, is an aid in the diagnosis of autoimmune hepatitis type 2.
QUANTA Flash LKM-1 Calibrators are intended for use with the QUANTA Flash LKM-1 Reagents for the determination of IgG anti-LKM-1 autoantibodies in human serum. Each calibrator establishes a point of reference for the working curve that is used to calculate unit values.
QUANTA Flash LKM-1 Controls are intended for use with the QUANTA Flash LKM-1 Reagents for quality control in the determination of IgG anti-LKM-1 autoantibodies in human serum.
The QUANTA Flash LKM-1 assay is designed to run on the BIO-FLASH® instrument. This platform is a fully automated closed system with continuous load and random access capabilities that automatically processes the samples, runs the assay and reports the results. It includes liquid handling hardware, luminometer and computer with software-user interface. The QUANTA Flash LKM-1 assay utilizes a reagent cartridge format, which is compatible with the BIO-FLASH instrument.
Recombinant cytochrome P450 2D6 (LKM-1) antigen is coated onto paramagnetic beads. The bead suspension is lyophilized and stored in the bead tube. Prior to use in the BIO-FLASH system, the sealed reagent tubes are pierced with the reagent cartridge lid and the beads are rehydrated and resuspended using resuspension buffer by pipetting up and down with a transfer pipette. The reagent cartridge is then loaded onto the BIO-FLASH instrument. Samples are also loaded onto the instrument in sample racks. Serum samples are diluted by the BIO-FLASH with system rinse in a small disposable plastic cuvette. Small amounts of the diluted patient serum, the beads, and assay buffer are combined into a second cuvette, and mixed. This cuvette is then incubated at 37°C. The beads are magnetized and washed several times. Isoluminol conjugated anti-human IgG antibodies are then added to the cuvette, and again incubated at 37°C. The beads are magnetized and washed repeatedly. The isoluminol conjugate is oxidized when Trigger 1 (Fe(II)coproporphyrin in sodium hydroxide solution) and Trigger 2 (urea-hydrogen peroxide in sodium chloride solution) are added to the cuvette, and the flash of light produced from this reaction is measured as Relative Light Units (RLU) by the BIO-FLASH optical system. The RLU are proportional to the amount of isoluminol conjugate that is bound to the human IgG, which is in turn proportional to the amount of anti-LKM-1 antibodies bound to the corresponding beads.
For quantitation, the QUANTA Flash LKM-1 assay utilizes a predefined lot specific Master Curve that is uploaded onto the instrument through the reagent cartridge barcode. Every new lot number of reagent cartridge must be calibrated before first use, with the QUANTA Flash LKM-1 Calibrators. Based on the results obtained with the three Calibrators included in the Calibrator Set (sold separately), an instrument specific Working Curve is created, which is used to calculate chemiluminescent units (CU) from the instrument signal (RLU) obtained for each sample.
The QUANTA Flash LKM-1 Reagents kit contains the following materials:
One (1) QUANTA Flash LKM-1 Reagent Cartridge One (1) vial of Resuspension buffer One (1) Transfer pipette
The QUANTA Flash LKM-1 reagent cartridge contains the following reagents for 50 determinations:
- a. LKM-1 coated paramagnetic beads, lyophilized.
- b. Assay buffer - colored pink, containing Tris-buffered saline, Tween 20, protein stabilizers and preservatives.
- C. Tracer IgG - Isoluminol labeled anti-human IgG antibodies in buffer, containing protein stabilizers and preservative.
The QUANTA Flash LKM-1 Calibrators kit contains two vials each of Calibrator 2, and Calibrator 3:
QUANTA Flash LKM-1 Calibrators:
- । QUANTA Flash LKM-1 Calibrator 1: Two (2) barcode labeled tubes containing 0.3 mL prediluted, ready to use reagent. Calibrators contain human antibodies to LKM-1 in stabilizers and preservatives.
- -QUANTA Flash LKM-1 Calibrator 2: Two (2) barcode labeled tubes containing 0.3 mL prediluted, ready to use reagent. Calibrators contain human antibodies to LKM-1 in stabilizers and preservatives.
- -QUANTA Flash LKM-1 Calibrator 3: Two (2) barcode labeled tubes containing 0.3 mL prediluted, ready to use reagent. Calibrators contain human antibodies to LKM-1 in stabilizers and preservatives.
The QUANTA Flash LKM-1 Controls kit contains two vials of Negative Control and two vials of Positive Control:
QUANTA Flash LKM-1 Controls:
- । QUANTA Flash LKM-1 Negative Control: Two (2) barcode labeled tubes containing 0.5 mL, ready to use reagent. Controls contain human antibodies to LKM-1 in stabilizers and preservatives.
- । QUANTA Flash LKM-1 Positive Control: Two (2) barcode labeled tubes containing 0.5 mL, ready to use reagent. Controls contain human antibodies to LKM-1 in stabilizers, and preservatives.
This document describes the regulatory submission for the QUANTA Flash® LKM-1 Reagents, Calibrators, and Controls, a chemiluminescent immunoassay used for the semi-quantitative determination of IgG anti-liver/kidney microsome type 1 antibodies in human serum. This product aids in the diagnosis of autoimmune hepatitis type 2 (AIH-2) when used with clinical findings and other laboratory tests.
1. Table of Acceptance Criteria and the Reported Device Performance
The acceptance criteria are generally implied within the performance studies described in the document, rather than explicitly stated in a single summarized table. The document provides detailed performance data for analytical and clinical characteristics. Here's a partial summary based on the provided information, focusing on quantifiable metrics:
Performance Characteristic | Acceptance Criteria (Implied) | Reported Device Performance |
---|---|---|
Precision | Total %CV: |
Ask a specific question about this device
Page 1 of 52