(264 days)
ST AIA-PACK 25-OH Vitamin D is designed for in vitro diagnostic use only for the quantitative measurement of total 25-hydroxyvitamin D (25-OH vitamin D) in human serum, Na-heparinized plasma or EDTA plasma on TOSOH AIA System Analyzers. The Tosoh ST AIA-PACK 25-OH Vitamin D is intended as an aid in the determination of Vitamin D sufficiency.
The ST AIA-PACK 25-OH Vitamin D Calibrator Set is intended for in vitro diagnostic use only for the calibration of the ST AIA-PACK 25-OH Vitamin D assay.
The ST AIA-PACK 25-OH Vitamin D is a one-step delayed competitive enzyme immunoassay which, after sample pretreatment, is performed entirely in the ST AIA-PACK 25-OH Vitamin D test cup. Sample pretreatment reagents (containing sodium hydroxide) disassociate 25-OH vitamin D from its binding proteins in the test sample. 25-OH vitamin D present in the pretreated sample is bound to 25-OH vitamin D-specific monoclonal antibody immobilized on magnetic beads. After that, the enzymelabeled 25-OH vitamin D is added to the reaction mixture. The enzyme-labeled 25-OH vitamin D competes with 25-OH vitamin D for binding to the antibody on magnetic beads in the reaction mixture.
After the second incubation, the magnetic beads are washed to remove the unbound enzymelabeled 25-OH vitamin D and are then incubated with a fluorogenic substrate, 4- methylumbellifery phosphate (4MUP). The amount of enzyme-labeled 25-OH vitamin D that binds to the beads is inversely proportional to the 25-0H vitamin D concentration in the test sample. A standard curve is constructed, and unknown 25-OH vitamin D concentrations are calculated using this curve.
The ST AIA-PACK 25-OH Vitamin D Calibrator Set contains human sera with assigned levels of 25-OH Vitamin D. The calibrator set consists of six calibrators with assigned concentrations of approximately 0, 8, 17, 33, 66 and 135 ng/mL. Each level contains the assigned concentration of the 25-OH vitamin D (described on each vial) with sodium azide as a preservative.
This looks like a 510(k) Summary for a medical device called "ST AIA-PACK 25-OH Vitamin D" and its associated calibrator set. The document describes the device, its intended use, and provides evidence of its performance and substantial equivalence to a predicate device.
Let's break down the requested information based on the provided text.
1. A table of acceptance criteria and the reported device performance
The document doesn't explicitly state "acceptance criteria" in a separate section with pass/fail thresholds against which the performance is measured. Instead, it presents performance characteristics and largely uses the performance of the predicate device or established clinical standards (like CLSI guidelines) as a benchmark for comparison or to demonstrate expected performance.
However, we can infer some criteria from the presented performance characteristics.
| Performance Characteristic | Acceptance Criteria (inferred/implied) | Reported Device Performance |
|---|---|---|
| Precision | Equal to or exceed predicate calibrator traceability (K123131). | Within-run precision: 1.3 - 3.9% Between run precision: 1.5 - 3.1% Between day precision: 2.2 - 4.1% Total precision: 2.6 - 4.9% Between lot precision: 1.1 - 3.1% Statement: "The precision equals or exceeds the precision obtained with the predicate calibrator traceability for (k)123131." |
| Linearity | Demonstrated to be linear within the assay range. | Linear from 4.0 ng/mL to 120 ng/mL. Statement: "There is no change in linearity from K123131." |
| Standardization & Traceability | Traceable to ID-LC/MS/MS 25-OH Vitamin D RMP (University of Ghent). Bias should be acceptable. | Weighted Deming regression analysis (vs. RMP): Slope: 0.98 (0.92 - 1.05) Intercept: -0.48 (-1.89 - 0.92) Corr Coef (R): 0.965 Bias: -0.23 (-0.71%) The study concludes that the assay is traceable to the RMP. |
| Method Comparison (vs. Predicate) | Demonstrate acceptable correlation and bias against predicate. | Deming regression analysis (vs. Predicate): Slope: 0.808 (0.801 to 0.816) Intercept: 0.26 (-0.15 to 0.67) Corr Coef (R): 0.9979 Bias: -8.90 (-20.25%) This comparison demonstrates the change in bias introduced by the standardization process. |
| Reference Ranges | Establish a representative reference interval. | Reference Interval: 12.3 – 60.0 ng/mL (based on 252 healthy individuals). |
| Limit of Detection (LoD) | Determined according to CLSI guideline EP17-A. | LoD: 3.2 ng/mL (determined by 12 measurements of 5 low-level samples across 3 lots) |
| Limit of Quantitation (LoQ) | Determined according to CLSI guideline EP17-A. | Functional sensitivity (LoQ): 3.3 ng/mL at 20% CV. |
| Limit of Blank (LoB) | Determined according to CLSI guideline EP17-A. | LoB: 1.6 ng/mL (determined by 60 measurements of 3 different blank specimens) |
2. Sample sizes used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)
- Precision:
- Sample Size: Three levels of unaltered serum specimens (Serum-B and Serum-C) and three spiked serum specimens (Serum-E and Serum-F) were assayed. Each level involved 2 replicates in a single run, 2 times a day for 20 non-consecutive days, across 2 analyzers and 2 lots of reagents. This totaled 40 runs and 80 determinants per lot.
- Data Provenance: Not explicitly stated, but typically clinical laboratory studies are prospective and involve samples collected for internal validation or purchased from biorepositories. No country of origin is mentioned.
- Linearity:
- Sample Size: Not explicitly stated, but typically involves a series of diluted samples to cover the assay range.
- Data Provenance: Not explicitly stated.
- Standardization and Traceability (Method Comparison vs. RMP):
- Sample Size: 111 serum specimens.
- Data Provenance: The RMP (Reference Measurement Procedure) was performed at the University of Ghent, in Ghent, Belgium. The samples were value-assigned by the Ghent University ID-LC-MS/MS RMP. This suggests these were likely retrospective samples used for standardization purposes in a prospective study design to validate the device's traceability.
- Method Comparison (vs. Predicate):
- Sample Size: 181 unaltered serum specimens.
- Data Provenance: Not explicitly stated, but likely from various sources for routine clinical testing. Implied to be a prospective comparison study.
- Reference Ranges:
- Sample Size: 252 serum samples (111 females, 141 males).
- Data Provenance: Specimens were obtained from "apparently healthy individuals" in Maryland, Pennsylvania, Wisconsin, and Southern California (USA). Samples were collected in March, May, June, and July, indicating a prospective collection over several months.
- Limit of Detection and Limit of Quantitation:
- LoB: 60 measurements of 3 different blank specimens.
- LoD: 12 measurements of 5 low-level sample ranges, using 3 lots of reagents.
- LoQ: Series of low Vitamin D samples assayed in replicates of 8 for 5 days (total 40 replicates per sample).
- Data Provenance: Not explicitly stated.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)
- For this in vitro diagnostic device, "ground truth" is primarily established by highly accurate reference methods or clinical standards, rather than expert consensus on interpretation.
- Standardization and Traceability: The ground truth for this segment was the ID-LC/MS/MS 25-OH Vitamin D Reference Measurement Procedure (RMP) at the University of Ghent. This is a highly specialized analytical chemistry method considered the "gold standard" for measuring 25-OH Vitamin D, not typically performed by individual "experts" in the clinical sense, but rather by trained laboratory personnel following precise protocols. The qualifications of the individuals performing the RMP are implied to be high-level analytical chemists/technicians experienced in ID-LC/MS/MS.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
- Adjudication methods like 2+1 or 3+1 are typically used in clinical imaging studies where subjective interpretation is involved.
- For an in vitro diagnostic device measuring a quantitative analyte like 25-OH Vitamin D, the "adjudication" is inherent in the analytical method's precision, accuracy, and traceability to a recognized reference method (like the ID-LC/MS/MS RMP). Discordant results would be investigated through re-testing or troubleshooting, not through a consensus of human reviewers. Therefore, the concept of a multi-observer adjudication method is not applicable here.
5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
- This is an in vitro diagnostic device for measuring a biochemical marker (Vitamin D). The concept of "multi-reader multi-case (MRMC) comparative effectiveness study" for human readers improving with or without AI assistance is not applicable to this type of device. There is no "reading" of images or complex data by human experts being assisted by AI in the context of this submission. The device performs an automated quantitative measurement.
6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done
- This device is an automated, standalone assay system. Its performance characteristics (precision, linearity, LoD, LoQ, method comparison) are its standalone performance without continuous human intervention in the interpretive phase. While humans operate the instrument and perform quality control, the measurement itself is algorithmic. Therefore, the presented studies (precision, linearity, method comparison, etc.) represent the standalone performance of the device algorithm.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)
- The primary ground truth used for validating the accuracy and standardization of the device is the Isotope Dilution-Liquid Chromatography/Tandem Mass Spectrometry (ID-LC/MS/MS) 25-OH Vitamin D Reference Measurement Procedure (RMP). This is a highly accurate and precise analytical method recognized as the gold standard for 25-OH Vitamin D measurement, traceable to NIST Standard Reference Material (SRM) 2972.
8. The sample size for the training set
- This document describes a 510(k) submission for a diagnostic assay, not a machine learning or AI algorithm in the context of typical training/test sets. The "training" for such an assay involves the development and optimization of the reagents, assay protocol, and calibration procedures using various samples. There isn't a single, defined "training set" sample size in the sense of AI model development.
- The "training" data, in an analogous sense, would be all the samples and experiments conducted during the development phase to establish the assay's performance characteristics, optimize reagent concentrations, and set initial calibrator values. This information is typically not detailed in a 510(k) summary as a single "training set."
9. How the ground truth for the training set was established
- As mentioned above, the concept of a distinct "training set" with ground truth in the AI sense is not directly applicable here. However, the development and initial calibration of the assay would have relied on highly characterized samples, likely correlated with established reference methods or primary standards (such as those traceable to NIST SRMs) to ensure accuracy and consistency from the earliest stages of development. The subsequent "standardization" step (as described in the document, traceable to the University of Ghent's ID-LC/MS/MS RMP) represents a formal validation and adjustment process to ensure consistency with recognized reference methods.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of a caduceus, which is a symbol of medicine. The caduceus is made up of three intertwined snakes and a pair of wings.
October 26,2015
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
TOSOH BIOSCIENCE, INC. ROBERT WICK REGULATORY SPECIALIST 6000 SHORELINE COURT, SUITE 101 SOUTH SAN FRANCISCO CA 94080
Re: K150270
Trade/Device Name: ST AIA-PACK 25-OH Vitamin D, ST AIA-PACK 25-OH Vitamin D Calibrator Set Regulation Number: 21 CFR 862.1825 Regulation Name: Vitamin D test system Regulatory Class: II Product Code: MRG, JIT Dated: September 10, 2015 Received: September 14, 2015
Dear Robert Wick:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
If you desire specific advice for your device on our labeling regulations (21 CFR Parts 801 and 809), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Katherine Serrano -S
For : Courtney H. Lias, Ph.D. Director Division of Chemistry and Toxicology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Indications for Use
510(k) Number (if known) K150270
Device Name
ST AIA-PACK 25-OH Vitamin D, ST AIA-PACK 25-OH Vitamin D Calibrator Set
Indications for Use (Describe)
ST AIA-PACK 25-OH Vitamin D is designed for in vitro diagnostic use only for the quantitative measurement of total 25-hydroxyvitamin D (25-OH vitamin D) in human serum, Na-heparinized plasma or EDTA plasma on TOSOH AIA System Analyzers. The Tosoh ST AIA-PACK 25-OH Vitamin D is intended as an aid in the determination of Vitamin D sufficiency.
The ST AIA-PACK 25-OH Vitamin D Calibrator Set is intended for in vitro diagnostic use only for the calibration of the ST AIA-PACK 25-OH Vitamin D assay.
Type of Use (Select one or both, as applicable)
| Prescription Use (Part 21 CFR 801 Subpart D) | ☑ |
|---|---|
| Over-The-Counter Use (21 CFR 801 Subpart C) | ☐ |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
510(k) Summary
ST AIA-PACK 25-OH Vitamin D K150270
| Date: | October 22, 2015 |
|---|---|
| Submitter: | Tosoh Bioscience, Inc3600 Gantz RoadGrove City, OH 43123 |
| Contact Person: | Robert L. WickRegulatory Specialist6000 Shoreline Ct., Ste. 101South San Francisco, CA 94080Phone: 650-636-8117Fax:650-636-8121Email: Robert.Wick@Tosoh.com |
| Device Name:Classification: | ST AIA-PACK 25-OH Vitamin DClass IIMRGClinical Chemistry21 CFR 862.1825 |
| Device Name:Classification | ST AIA-PACK 25-OH Vitamin D Calibrator SetClass IIJITClinicalChemistry21 CFR 862.1150 |
| Predicate Device: | K123131Tosoh Bioscience, Inc.ST AIA-PACK 25-OH Vitamin D AssayST AIA-PACK 25-OH Vitamin D Calibrator Set |
According to the requirements of 21 CFR 807.92, the following information provides sufficient detail to understand the basis for a determination of substantial equivalence.
Device Description:
The ST AIA-PACK 25-OH Vitamin D is a one-step delayed competitive enzyme immunoassay which, after sample pretreatment, is performed entirely in the ST AIA-PACK 25-OH Vitamin D test cup. Sample pretreatment reagents (containing sodium hydroxide) disassociate 25-OH vitamin D from its binding proteins in the test sample. 25-OH vitamin D present in the pretreated sample is bound to 25-OH vitamin D-specific monoclonal antibody immobilized on magnetic beads. After that, the enzymelabeled 25-OH vitamin D is added to the reaction mixture. The enzyme-labeled 25-OH vitamin D competes with 25-OH vitamin D for binding to the antibody on magnetic beads in the reaction mixture.
After the second incubation, the magnetic beads are washed to remove the unbound enzymelabeled 25-OH vitamin D and are then incubated with a fluorogenic substrate, 4- methylumbellifery phosphate (4MUP). The amount of enzyme-labeled 25-OH vitamin D that binds to the beads is inversely proportional to the 25-0H vitamin D concentration in the test sample. A standard curve is
ST AIA-PACK 25-OH Vitamin D
{4}------------------------------------------------
Tosoh Bioscience, Inc. constructed, and unknown 25-OH vitamin D concentrations are calculated using this curve.
The ST AIA-PACK 25-OH Vitamin D Calibrator Set contains human sera with assigned levels of 25-OH Vitamin D. The calibrator set consists of six calibrators with assigned concentrations of approximately 0, 8, 17, 33, 66 and 135 ng/mL. Each level contains the assigned concentration of the 25-OH vitamin D (described on each vial) with sodium azide as a preservative.
The following products are required to use the ST AIA-PACK 25-OH Vitamin D P/N 025234: ST
| AIA-PACK 25-OH Vitamin D Calibrator Set | 025334 |
|---|---|
| ST AIA-PACK 25-OH Vitamin D Pretreatment Set | 025734 |
Device Intended Use:
ST AIA-PACK 25-OH Vitamin D is designed for in vitro diagnostic use only for the quantitative measurement of total 25-hydroxyvitamin D (25-OH vitamin D) in human serum, Na heparinized or EDTA plasma on TOSOH AIA System Analyzers. The Tosoh ST AIA-PACK 25-OH Vitamin D is intended as an aid in the determination of Vitamin D sufficiency.
Calibrators:
The ST AIA-PACK 25-OH Vitamin D Calibrator Set is intended for In Vitro Diagnostic Use Only for the calibration of the ST AIA-PACK 25-OH Vitamin D assay.
Device Indications for Use:
Same as Intended Use
Substantial Equivalence:
Comparison between the Tosoh ST AIA-PACK 25-OH Vitamin D (standardized) and Tosoh AIA-PACK 25- OH Vitamin D Assay (unstandardized)
Similarities and Differences
| Parameter | ST AIA-PACK 25-OH VitaminD Test Cups (modified calibrator) | ST AIA-PACK 25-OH VitaminD Test Cups K123131 |
|---|---|---|
| Intended use | ST AIA-PACK 25-OH Vitamin D is designed for in vitrodiagnostic use only for thequantitative measurement oftotal 25-hydroxyvitamin D (25-OH vitamin D) in humanserum, Na heparinized orEDTA plasma on TOSOH AIASystem Analyzers. The TosohST AIA-PACK 25-OH VitaminD is intended as an aid in thedetermination of Vitamin Dsufficiency. | Same |
| Indications for Use | Same as Intended Use. | Same |
| Sample type | EDTA plasma, Na heparinizedplasma or serum | Same |
| Assay Range | 4.0 ng/mL - 120.0 ng/mL | Same |
| Assay Technology | Fluorescent Immunoassay | Same |
| Incubation Time | 10 minute cycle | Same |
| Reference range | 12.3 ng/mL to 60.0 ng/mL | 10.8 to 54.75 ng/mL |
{5}------------------------------------------------
| Parameter | ST AIA-PACK 25-OH VitaminD Calibrator Set | ST AIA-PACK 25-OH VitaminD Calibrator Set K123131 |
|---|---|---|
| Intended use | The ST AIA-PACK 25-OHVitamin D Calibrator Set isintended for IN VITRODIAGNOSTIC USE ONLY forthe calibration of the ST AIA-PACK 25-OH Vitamin D assay. | Same |
| Matrix | Human serum | Same |
| Number of Calibrators | 6 | Same |
| Format | Lyophilized | Same |
| Standardization/Traceability | Traceable to the ID-LC/MS/MS25 (OH) vitamin D referencemeasurement procedure(University of Ghent) via patientsample correlation. | Traceable to an Internalreference standard |
| Concentration 25-OH vit D | Approximately 0-135 ng/mL (0,8, 17, 33, 66, 135 ng/mL) | Approximately 0-165 ng/mL (0,10, 20, 40, 80, 165 ng/mL) |
Calibrator Set പ വിഷ Similarities
PERFORMANCE CHARACTERISTICS
Precision
Precision was assessed by assaying three levels of unaltered serum specimens (Serum-B and Serum-C) and three spiked serum specimens (Serum-E and Serum-F) on two analyzers and two lots of reagents. Estimates of total and within-run precision were obtained from measurements of 2 replicates in a single run, 2 times a day for 20 non-consecutive days. This equaled a total of 40 runs and 80 determinants per lot. In addition, all of the data were combined to assess the within run, between run, between lot and total precision. The following results were obtained when testing for precision:
| Within-run precision | 1.3 - 3.9% |
|---|---|
| Between run precision | 1.5 - 3.1% |
| Between day precision | 2.2 - 4.1% |
| Total precision | 2.6 - 4.9% |
| Between lot precision | 1.1 - 3.1% |
Combined Lot Summary Table
| Sample | Serum A | Serum B | Serum C | Serum D | Serum E | Serum F | |
|---|---|---|---|---|---|---|---|
| Mean Conc.ng/mL | 16.9 | 20.9 | 24.7 | 55.2 | 74.6 | 95.3 | |
| WithinRun | SD | 0.62 | 0.82 | 0.77 | 1.28 | 1.55 | 1.24 |
| %CV | 3.7 | 3.9 | 3.1 | 2.3 | 2.1 | 1.3 |
{6}------------------------------------------------
Tosoh Bioscience. Inc
| BetweenRun | SD | 0.53 | 0.65 | 0.52 | 1.07 | 1.26 | 1.40 |
|---|---|---|---|---|---|---|---|
| %CV | 3.1 | 3.1 | 2.1 | 1.9 | 1.7 | 1.5 | |
| BetweenDay | SD | 0.64 | 0.68 | 1.02 | 1.48 | 1.88 | 2.07 |
| %CV | 3.8 | 3.3 | 4.1 | 2.7 | 2.5 | 2.2 | |
| BetweenLot | SD | 0.53 | 0.36 | 0.76 | 1.16 | 1.57 | 1.03 |
| %CV | 3.1 | 1.7 | 3.1 | 2.1 | 2.1 | 1.1 | |
| Total | SD | 0.82 | 0.96 | 1.19 | 1.81 | 2.28 | 2.43 |
| %CV | 4.9 | 4.6 | 4.8 | 3.3 | 3.1 | 2.6 |
The precision equals or exceeds the precision obtained with the predicate calibrator traceability for (k)123131.
Linearity:
The linearity for ST AIA-PACK 25-OH Vitamin D Calibrator was determined, based on the CLSI protocol entitled: Evaluation of the Linearity of Qualitative Measurement Methods; Approved Guidelines (EP6-A). The linearity was measured on the AIA-2000 instrument and has been demonstrated to be linear from 4.0 ng/mL to 120 ng/mL.
There is no change in linearity from K123131.
Standardization and Traceability
This assay is traceable to the isotope dilution-liquid chromatography/tandem mass spectrometry (ID-LC-/MS/MS) 25-OH Vitamin D Reference Method Procedure (RMP) that is traceable to the National Institute of Standards and Technology (NIST) Standard Reference Material (SRM) 2972.
The Tosoh ST-AIA PACK 25 OH Vitamin D assay was originally cleared under 510(k)123131. Tosoh modified the assay by standardizing the cleared Vitamin D assay in accordance with the Vitamin D Standardization and Certification Program (VDSCP).
Please refer to http://ods.od.nih.gov/Research/Vitamin D.aspx for more information on this program.
In accordance with the recommendations of the program, the traceability of the Tosoh ST-AIA PACK 25-OH Vitamin D Calibrator Set was verified by completing a method comparison and bias estimation using the University of Ghent's ID-LC/MS/MS 25-OH vitamin D Reference Measurement Procedure (RMP).
The relationship between the Tosoh ST-AIA PACK 25-OH Vitamin D assay and the ID-LC/MS/MS 25-OH Vitamin D RMP is described below in the method comparison section.
Value Assignment
The values of the primary reference materials were adjusted with correlation study results with the ID-LC-MS/MS method. The values were verified by comparing measured results with those obtained with ID-LC-MS/MS for patient samples at the University of Ghent, in Ghent, Belgium.
The values of the secondary reference material were assigned using the primary reference
ST AIA-PACK 25-OH Vitamin D
{7}------------------------------------------------
materials as calibrators. The values were verified by comparing measured results with those obtained with the previous lot for patient samples.
The commercial calibrator values were assigned with the secondary reference materials as calibrators. The values were verified by comparing measured results with those obtained with the previous lot for control materials.
Correlation
Method Comparison
The methods comparison study was conducted with reference to the CLSI protocol entitled: Method Comparison and Bias Estimation Using Patient Samples: Approved Guideline (EP9-A2)
A total of 111 serum specimens in the range of 5.0 -114 ng/mL, as value assigned by the Ghent University ID-LC-MS/MS reference measurement procedure (RMP), were assayed in singleton utilizing the ST AIA-PACK 25-OH Vitamin D assay to assess the ST-AIA-PACK 25-OH Vitamin D Assay traceability to the RMP.
The Weighted Deming regression analysis between the RMP method and the ST AIA-PACK 25-OH Vitamin D Assay is listed below:
| Weighted Deming * | |
|---|---|
| Slope: | 0.98 (0.92 - 1.05) |
| Intercept: | -0.48 (-1.89 - 0.92) |
| Standard Error Estimate | 0.708 |
| Corr Coef (R): | 0.965 |
| Bias: | -0.23 (-0.71%) |
| Points (Plotted/Total): | 111/111 |
| Result Ranges: | ST AIA-PACK 25-OH Vitamin D 4.7 to 120 ng/mLID-LC-MS/MS (Ghent University) 5.0 to 114 ng/mL |
*95% Confidence Intervals are shown in parentheses
To demonstrate the bias between the standardized assay (candidate) as compared to the original assay (predicate), one hundred eighty-one (181) unaltered serum specimens were assayed in singleton utilizing the ST AIA-PACK 25-OH Vitamin D assay on the AIA-2000 analyzer using calibrations from ST AIA-PACK 25-OH Vitamin D Calibrator (predicate and candidate) and comparing the results. The results are listed below:
| Deming * | Regular * | |
|---|---|---|
| Slope | 0.808 (0.801 to 0.816) | 0.807 (0.7999 to 0.815) |
| Intercept | 0.26 (-0.15 to 0.67) | 0.33 (-0.08 to 0.73) |
| Standard Error Estimate | 1.17 | 1.17 |
| Corr Coef ® | 0.9979 | |
| Bias | -8.90 (-20.25%) | |
| Points (Plotted/Total) | 181/181 | |
| Result Ranges | Predicate calibrator 11.4 to111.8 ng/mLCandidate calibrator 9.4 to 92.3 ng/mL |
*95% Confidence Intervals are shown in parentheses
Matrix Comparison
See K123131
{8}------------------------------------------------
Cross Reactivity
See K123131
Reference Ranges
A reference range study was conducted based on guidance from Clinical and Laboratory Standards Institute (CLSI) Protocol C28-A3.
To determine the reference range for the ST-AIA-PACK 25-OH Vitamin D Assay, a total of 252 serum samples, from 111 females and 141 males, ages 21 to 86 years, were assayed in singleton utilizing the ST AIA-PACK 25-OH Vitamin D assay on the AIA-2000 analyzer. The specimens were obtained from apparently healthy individuals with normal PTH, TSH, calcium, magnesium and phosphorus levels. The samples were collected in Maryland, Pennsylvania, Wisconsin and Southern California in March, May, June, and July respectively.
The reference range specimens had a normal distribution and the 95% confidence interval was defined as the ref. range interval.
Demographics
| Caucasian | 48% | non-Caucasian | 52% | |
|---|---|---|---|---|
| Female | 44% | Male | 56% | |
| Supplements | 26% | No Supplements | 56% | Not specified 18% |
There was no significant difference between values obtained for male and females. The following reference range was obtained for the adult population:
| Number of Samples (n) | 252 |
|---|---|
| Reference Interval | 12.3 – 60.0 ng/mL |
Interference
See K123131
Limit of Detection and Limit of Quantification
The LoB and LoD of the ST AIA-PACK 25-OH Vitamin D was determined based on the CLSI guideline entitled: Protocols for Determination of Limits of Detection and Limits of Quantitation; Approved Guideline EP17-A.
The LoB was determined by 60 measurements of 3 different blank specimens. The LoB was the value at the 95th percentile. In this case the LoB was determined to be 1.6 ng/mL.
The LoD was determined by 12 measurements of 5 low level sample range was chosen to be between LoB and 4xLoB and 3 lots of reagents were utilized. The LoD was determined to be 3.2 ng/mL.
The LoQ was determined by measuring a series of low Vitamin D samples that were prepared by diluting a specimen with known Vitamin C concentration at different dilution ratios. The prepared samples were assayed in replicates of eight 98) every day for five (5) days on one instrument for a total of 40 replicates per sample.
{9}------------------------------------------------
The mean, SD and %CV were established and plotted to calculate the functional sensitivity. The functional sensitivity is 3.3 ng/mL at 20% CV.
Standards:
| Number | FDARecognitionNumber | RevisionDate | Title |
|---|---|---|---|
| EP5-A2 | 7-110 | 10/31/2005 | Evaluation of Precision Performance ofQuantitative Measurement Methods;Approved Guideline-Second Edition |
| EP6-A | 7-193 | 03/18/2009 | Evaluation of the Linearity of QuantitativeMeasurement Procedures: a StatisticalApproach: Approved Guideline |
| EP28-A3 | 7-224 | 09/08/2009 | Defining, Establishing, and verifyingReference Intervals in the Clinical Laboratory;Approved Guideline - Third Edition |
| EP17-A | 7-194 | 03/18/2009 | Protocols for Determination of Limits ofDetection and Limits of Quantitation;Approved Guideline |
| EP9-A3 | 7-92 | 03/08/2004 | Method Comparison and Bias EstimationUsing Patient Samples; Approved Guideline-Second Edition |
Conclusion:
The Tosoh Bioscience, Inc. ST AIA-PACK 25-OH Vitamin D Assay and Calibrator Set is substantially equivalent to the Tosoh K123131- ST AIA-PACK 25-OH Vitamin D Assay and Calibrator Set for in vitro diagnostic use only for the quantitative measurement of total 25hydroxyvitamin D (25-OH Vitamin D) in human serum, Na heparinized or EDTA plasma on Tosoh AIA System Analyzers.
§ 862.1825 Vitamin D test system.
(a)
Identification. A vitamin D test system is a device intended for use in clinical laboratories for the quantitative determination of 25-hydroxyvitamin D (25-OH-D) and other hydroxylated metabolites of vitamin D in serum or plasma to be used in the assessment of vitamin D sufficiency.(b)
Classification. Class II (special controls). Vitamin D test systems must comply with the following special controls:(1) Labeling in conformance with 21 CFR 809.10 and
(2) Compliance with existing standards of the National Committee on Clinical Laboratory Standards.