Search Results
Found 2 results
510(k) Data Aggregation
(64 days)
Powder free vinyl patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.
powder free vinyl patient examination gloves, white(non-colored) that meets all of the requirements of ASTM standard D 5250-00e4
The provided document refers to the 510(k) summary for "FUGUAN (Brand) Powder Free Vinyl Patient Examination Gloves". This is a medical device application for gloves, not a complex AI-powered diagnostic or therapeutic device. As such, the study design and acceptance criteria are focused on physical and biological properties of the gloves, rather than diagnostic performance metrics (e.g., sensitivity, specificity, accuracy, F1 score) or human-AI interaction in a clinical setting.
Therefore, the requested information categories typically applicable to AI/ML device studies (e.g., sample size for test/training set, number of experts, adjudication methods, MRMC studies, standalone performance, ground truth types) are not relevant to this specific product.
Here's a breakdown of the acceptance criteria and performance as described in the document, tailored to the nature of this device:
Acceptance Criteria and Device Performance for FUGUAN Powder Free Vinyl Patient Examination Gloves
| Characteristics | Acceptance Criteria (Standard) | Reported Device Performance |
|---|---|---|
| Physical Properties | ||
| Dimension | ASTM standard D 5250-00e4. | Meets |
| Physical Properties | ASTM standard D 5250-00e4. | Meets |
| Barrier Integrity | ||
| Freedom from pinholes | 21 CFR 800.20 | Meets |
| Chemical Properties | ||
| Powder Residual | ASTM standard D 5250-00e4 | <2mg/glove |
| Biocompatibility | (Not explicitly stated as a numerical standard in table, but referenced in narrative) | |
| Primary Skin Irritation | (Implied by standard testing, ISO10993-10 mentioned) | Passes (Not a Primary Skin Irritation) |
| Dermal Sensitization | (Implied by standard testing, ISO10993-10 mentioned) | Passes (Not a Dermal Sensitization) |
Study Details:
- Sample size used for the test set and the data provenance: The document does not specify a distinct "test set" in the context of an AI/ML study. The evaluation appears to be based on quality control testing of representative samples of the manufactured gloves for each characteristic. No specific sample sizes for these tests are provided in this summary. Data provenance is implied by the manufacturer's location (Shijiazhuang Fuguan Plastic Products Co., Ltd., China) and adherence to international standards (e.g., ASTM, ISO) and US regulations (e.g., 21 CFR). The testing is generally prospective, as it involves evaluating manufactured products against predetermined standards.
- Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable. Ground truth for these types of devices is established by objective measurements against recognized industry standards (e.g., ASTM for physical properties, 21 CFR for pinholes, ISO for biocompatibility). Human expert consensus is not a factor for these objective measures.
- Adjudication method for the test set: Not applicable. The "ground truth" is determined by direct measurement and adherence to established pass/fail criteria from the standards.
- If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable. This device is a patient examination glove, not an AI-powered diagnostic tool requiring human reader studies.
- If a standalone (i.e. algorithm only without human-in-the loop performance) was done: Not applicable. This device is a physical product, not an algorithm.
- The type of ground truth used:
- Objective Standards: The "ground truth" is established by publicly available and internationally recognized technical standards and regulations:
- ASTM standard D 5250-00e4 (for dimensions, physical properties, powder residual)
- 21 CFR 800.20 (for freedom from pinholes/waterleak test)
- ASTM D6124-01 (implied for powder residual measurement)
- ISO10993-10 (for biocompatibility, specifically irritation and sensitization)
- Objective Standards: The "ground truth" is established by publicly available and internationally recognized technical standards and regulations:
- The sample size for the training set: Not applicable. There is no "training set" in the context of an AI/ML device. Manufacturing processes are optimized through engineering and quality control, not machine learning training.
- How the ground truth for the training set was established: Not applicable.
Summary of the Study:
The study demonstrating the device meets the acceptance criteria is a series of non-clinical bench tests and biocompatibility assessments. These tests are performed on samples of the manufactured gloves to ensure they conform to established safety and performance standards. The conclusions are drawn from these non-clinical tests, affirming that the FUGUAN (Brand) Powder Free Vinyl Patient Examination Gloves meet the requirements of ASTM standard D 5250-00e4, ASTM D6124-01, 21 CFR 800.20, and ISO10993-10. This substantiates that the device is as safe and effective as the predicate device (Shijiazhuang Great Eagle Plastics Products Co., Ltd. K000071). Clinical data was deemed not necessary for this type of device.
Ask a specific question about this device
(64 days)
Powdered vinyl patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.
powdered vinyl patient examination gloves, white(non-colored) that meets all of the requirements of ASTM standard D 5250-0064
This document describes the acceptance criteria and the study conducted to demonstrate that the medical device in question, "FUGUAN (Brand) Powdered Vinyl Patient Examination Gloves, White (Non-colored)," meets these criteria.
1. Table of Acceptance Criteria and Reported Device Performance
| Characteristics | Acceptance Criteria (Standard) | Reported Device Performance |
|---|---|---|
| Dimension | ASTM standard D 5250-00e4. | Meets |
| Physical Properties | ASTM standard D 5250-00e4. | Meets |
| Freedom from pinholes | 21 CFR 800.20 | Meets |
| Powder Amount | ASTM standard D 5250-00e4 | <10mg/dm² |
| Biocompatibility | Primary Skin Irritation in rabbits | Passes (Not a Primary Skin Irritation) |
| Dermal sensitization in the guinea pig | Passes (Not a Dermal sensitization) |
2. Sample Size Used for the Test Set and Data Provenance
The document does not explicitly state the specific sample sizes used for each test. However, it references established industry standards from ASTM and FDA regulations, which typically prescribe statistically significant sample sizes for their respective testing methodologies.
- Data Provenance: The document does not specify the country of origin of the data or whether it was retrospective or prospective. The testing was conducted by the manufacturer, Shijiazhuang Fuguan Plastic Products Co., Ltd., which is based in Hebei Province, P.R. China.
3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications
The document does not mention the use of experts to establish ground truth for the test set. The acceptance criteria are based on established regulatory standards (21 CFR 800.20) and industry standards (ASTM D 5250-00e4, ASTM D6124-01, ISO10993-10). The device's performance was evaluated against these predefined objective criteria rather than expert consensus on individual cases.
4. Adjudication Method for the Test Set
Not applicable. The tests are based on objective measurements against established standards, not on subjective assessments requiring adjudication.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study
No. This type of study is not applicable to the assessment of patient examination gloves. MRMC studies are typically used for diagnostic or screening devices where human interpretation is involved.
6. Standalone (Algorithm Only) Performance Study
No. This is a physical medical device (gloves), not an algorithm or software. Therefore, a standalone algorithm performance study is not relevant.
7. Type of Ground Truth Used
The ground truth for evaluating the device's performance is based on established regulatory and industry standards. These standards define the acceptable physical, mechanical, and biocompatibility properties for patient examination gloves.
- ASTM D 5250-00e4: Standard Specification for Poly(Vinyl Chloride) Patient Examination Gloves. This standard dictates requirements for dimensions, physical properties, and powder amount.
- 21 CFR 800.20: This FDA regulation covers "Indirectly Heated Autoclaves," but in this context, it is referenced for "Freedom from pinholes," which is a specific test for barrier integrity.
- ASTM D 6124-01: Standard Test Method for Residual Powder on Medical Gloves.
- ISO 10993-10: Biological evaluation of medical devices – Part 10: Tests for irritation and skin sensitization.
8. Sample Size for the Training Set
Not applicable. This device is a physical product and does not involve AI algorithms that require a "training set." The testing performed is for product quality and compliance with standards.
9. How the Ground Truth for the Training Set was Established
Not applicable, as there is no training set for this type of device.
Ask a specific question about this device
Page 1 of 1