Search Results
Found 1 results
510(k) Data Aggregation
(48 days)
Powder free vinyl patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.
powder free vinyl patient examination gloves, Clear(non-colored) that meets all of the requirements of ASTM standard D 5250-06.
The provided text describes the acceptance criteria and performance of "Powder Free Vinyl Patient Examination Gloves, Clear(non-colored)" manufactured by Zhang Jia Gang Fengyuan Plastic Products Co.,Ltd. This is a medical device submission, and the "study" described is the testing conducted to demonstrate substantial equivalence to a predicate device and adherence to relevant standards.
Here's a breakdown of the requested information:
1. Table of Acceptance Criteria and Reported Device Performance
| Characteristics | Acceptance Criteria | Device Performance |
|---|---|---|
| Dimension | ASTM standard D 5250-06 | Meets |
| Physical Properties | ASTM standard D 5250-06 | Meets |
| Freedom from pinholes | 21 CFR 800.20 (Water leak test on pinhole AQL) | Meets |
| Powder Residual | ASTM standard D 5250-06 and D6124-06 | <2mg/glove |
| Biocompatibility (Primary Skin Irritation) | FDA recognition number 2-87: AAMI / ANSI / ISO 10993-10:2002 (Primary Skin Irritation in rabbits) | Pass (Not a Primary Skin Irritation) |
| Biocompatibility (Dermal Sensitization) | FDA recognition number 2-87: AAMI / ANSI / ISO 10993-10:2002 (Dermal sensitization in the guinea pig) | Pass (Not a Dermal sensitization) |
2. Sample Size Used for the Test Set and Data Provenance
The document does not explicitly state the specific sample sizes used for each test (e.g., how many gloves were tested for pinholes, or how many rabbits/guinea pigs were used for biocompatibility). The testing is described as meeting ASTM standards and CFR regulations, which inherently include statistically significant sample size requirements for their respective tests.
- Provenance: This is a submission from China (Zhang Jia Gang Fengyuan Plastic Products Co.,Ltd, Jiangsu, China). The data appears to be prospective in the sense that the manufacturer conducted these tests specifically to demonstrate compliance for their device before submitting for FDA clearance.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications
The document does not mention the use of experts to establish a "ground truth" for the test set in the context of human annotation or interpretation. The tests are for physical and biological properties of the gloves, where the ground truth is established by objective measurements against predefined standards (ASTM, CFR, ISO). These tests are typically performed by laboratory technicians or qualified personnel following standard operating procedures, rather than expert clinicians establishing ground truth through consensus.
4. Adjudication Method for the Test Set
Not applicable. The tests are based on objective measurements and compliance with specified standards, not on human interpretation that would require adjudication.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study
No. This type of study (MRMC) is generally relevant for imaging or diagnostic devices where human readers interpret results, and the AI's impact on their performance is evaluated. This submission is for patient examination gloves, which do not involve such interpretation.
6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study
Not applicable. This is not an algorithmic or AI-based device, so the concept of standalone performance doesn't apply. The device itself (the glove) is being tested for its physical and biological properties.
7. Type of Ground Truth Used
The ground truth used is based on:
- Predefined Standards: ASTM standard D 5250-06, ASTM D6124-06.
- Regulatory Requirements: 21 CFR 800.20, 21 CFR 801, 21 CFR 803, 21 CFR 807, 21 CFR 820.
- International Standards: FDA recognition number 2-87: AAMI / ANSI / ISO 10993-10:2002.
These standards and regulations define the objective criteria and methodologies for determining dimensions, physical properties, freedom from pinholes, powder residue, and biocompatibility.
8. Sample Size for the Training Set
Not applicable. This is not an AI/machine learning device, so there is no concept of a "training set."
9. How the Ground Truth for the Training Set Was Established
Not applicable, as there is no training set for this device.
Ask a specific question about this device
Page 1 of 1