Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K073193
    Date Cleared
    2007-12-20

    (37 days)

    Product Code
    Regulation Number
    880.6250
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Powder free vinyl patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.

    Device Description

    powder free vinyl patient examination gloves, Clear(non-colored) that meets all of the requirements of ASTM standard D 5250-0008.

    AI/ML Overview

    The provided text describes the acceptance criteria and the study for a medical device, specifically "Powder Free Vinyl Patient Examination Gloves, Clear (Non-Colored)". The study's purpose is to demonstrate substantial equivalence to a predicate device, which is a common pathway for Class I medical devices in the US.

    Here's the breakdown of the requested information:

    1. Table of acceptance criteria and the reported device performance

    CharacteristicsAcceptance Criteria (Standard)Reported Device Performance
    DimensionASTM standard D 5250-00e4Meets
    Physical PropertiesASTM standard D 5250-00e4Meets
    Freedom from pinholes21 CFR 800.20Meets
    Powder ResidualASTM standard D 5250-00e4 and D6124-01<2mg/glove
    Biocompatibility (Primary Skin Irritation)Primary Skin Irritation in rabbitsPasses (Not a Primary Skin Irritation)
    Biocompatibility (Dermal Sensitization)Dermal sensitization in the guinea pigPasses (Not a Dermal Sensitization)

    2. Sample size used for the test set and the data provenance

    The document does not explicitly state the sample sizes used for each specific test (e.g., dimension, physical properties, pinholes). It refers to the standards (ASTM D 5250-00e4, ASTM D6124-01, 21 CFR 800.20, ISO10993-10) which would contain the prescribed sampling plans. The data provenance is not specified beyond being "submitted, reference, or relied on". Given the context of a 510(k) submission from a Chinese manufacturer, the testing likely occurred at facilities approved to conduct tests according to these international standards, but specific country of origin for the data is not mentioned. The study is a non-clinical evaluation, not a retrospective or prospective clinical study.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    Not applicable. This is a non-clinical device performance and biocompatibility study, not a study requiring expert clinical judgment to establish ground truth. The "ground truth" is established by adherence to the specified technical standards and regulations.

    4. Adjudication method for the test set

    Not applicable. As this is a non-clinical performance test, there isn't an "adjudication method" in the sense of reconciling human expert opinions. The results are based on objective measurements and observations according to standardized test protocols.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This device is a patient examination glove, a Class I medical device. It is not an AI-powered diagnostic or assistive technology, and therefore, an MRMC comparative effectiveness study involving human readers and AI is irrelevant to its evaluation.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. As stated above, this is not an AI algorithm.

    7. The type of ground truth used

    The "ground truth" in this context is defined by the technical specifications and requirements outlined in recognized standards and regulations:

    • ASTM standard D 5250-00e4 (for Dimension, Physical Properties, and Powder Residual)
    • 21 CFR 800.20 (for Freedom from pinholes - Waterleak Test)
    • ASTM D6124-01 (for Powder Residual)
    • ISO10993-10 (for Biocompatibility, specifically Primary Skin Irritation and Dermal Sensitization)

    These standards specify the acceptable limits and test methods, which serve as the objective ground truth against which the device's performance is measured.

    8. The sample size for the training set

    Not applicable. This is not a machine learning or AI-based device, so there is no training set. The evaluation focuses on the physical and chemical properties of the manufactured gloves.

    9. How the ground truth for the training set was established

    Not applicable, as there is no training set for this device.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1