Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K072001
    Date Cleared
    2007-08-15

    (23 days)

    Product Code
    Regulation Number
    880.6250
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Powder free vinyl patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.

    Device Description

    powder free vinyl patient examination gloves, Clear(non-colored) that meets all of the requirements of ASTM standard D 5250-00ed.

    AI/ML Overview

    Here's an analysis of the provided text regarding the acceptance criteria and study that proves the device meets those criteria:

    Device: Powder Free Vinyl Patient Examination Gloves, Clear(non-colored)


    1. Table of Acceptance Criteria and Reported Device Performance

    CharacteristicAcceptance Criteria (Standard)Reported Device Performance
    DimensionASTM standard D 5250-00e4.Meets
    Physical PropertiesASTM standard D 5250-00e4.Meets
    Freedom from pinholes21 CFR 800.20Meets
    Powder ResidualASTM standard D 5250-00e4 and D6124-01Meets, <2mg/glove
    Biocompatibility:ISO10993-10
    - Primary Skin IrritationPrimary Skin Irritation in rabbits (ISO10993-10)Passes, Not a Primary Skin Irritation
    - Dermal SensitizationDermal sensitization in the guinea pig (ISO10993-10)Passes, Not a Dermal sensitization

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not explicitly state the sample sizes used for the specific tests (e.g., freedom from pinholes, physical properties, biocompatibility). It only indicates that the device "Meets" the specified ASTM and CFR standards.

    The data provenance is not explicitly mentioned as retrospective or prospective, nor are country of origin of the data. The tests are described as being performed to meet international and US standards (ASTM, 21 CFR, ISO), which implies general laboratory testing rather than data collected from specific patient populations or geographical regions.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

    This information is not provided in the document. For device testing of physical and chemical properties and biocompatibility, the "ground truth" is typically defined by the established testing protocols and acceptance limits within the referenced standards (e.g., ASTM D 5250-00e4, 21 CFR 800.20, ISO10993-10) rather than requiring expert consensus on individual results in the same way clinical image interpretation might. The tests are designed to be objective and measurable.

    4. Adjudication Method for the Test Set

    The document does not mention any adjudication method. For the type of objective physical and chemical testing performed for gloves, adjudication methods like 2+1 or 3+1 are typically not applicable. Results are against predefined criteria from standards.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    No, a multi-reader multi-case (MRMC) comparative effectiveness study was not done. This type of study is relevant for diagnostic devices that involve human interpretation (e.g., imaging AI). This device is a medical glove, which is a physical barrier device, and its effectiveness is determined by its physical properties and biocompatibility, not by human interpretation or AI assistance.

    6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study

    This question is not applicable to the device. The device is a physical medical glove, not an algorithm or an AI-powered system. Therefore, no standalone algorithm performance study was conducted.

    7. Type of Ground Truth Used

    The "ground truth" for the device's performance is based on objective measurements and predefined criteria as established by recognized industry standards and regulations. Specifically:

    • ASTM Standard D 5250-00e4: For Dimension, Physical Properties, and Powder Residual.
    • 21 CFR 800.20: For Freedom from pinholes (which relates to the Acceptable Quality Level or AQL for water leak testing).
    • ISO 10993-10: For Biocompatibility (Primary Skin Irritation and Dermal Sensitization).

    These standards specify the methods of testing and the acceptance limits that define whether a glove "meets" the standard.

    8. Sample Size for the Training Set

    This information is not applicable and not provided. As a medical glove, the device does not employ machine learning or AI, and therefore does not have a "training set" in the context of algorithm development.

    9. How Ground Truth for the Training Set Was Established

    This information is not applicable and not provided for the same reasons mentioned under point 8.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1