Search Results
Found 2 results
510(k) Data Aggregation
(140 days)
The epoc® Blood Analysis System is intended for use by trained medical professionals as an in vitro diagnostic device for the quantitative testing of samples of heparinized or un-anticoagulated arterial, venous, or capillary whole blood in the laboratory or at the point of care.
The Blood Gas Electrolyte and Metabolite (BGEM) Test Card panel configuration includes sensors that quantitate pH, pCO2, oO2, Sodium, Potassium, Ionized Calcium, Chloride, Total Carbon Dioxide, Glucose, Lactate, Blood Urea Nitrogen, Creatinine, and Hematocrit.
pH, pCO2, pO2 (blood gases) measurements from the epoc Blood Analysis System are used in the diagnosis and treatment of lifethreatening acid-base disturbances.
Sodium and Potassium measurements from the epoc Blood Analysis System are treatment of diseases involving electrolyte imbalance.
lonized Calcium measurements from the epoc Blood Analysis System are treatment of parathyroid disease, a variety of bone diseases, chronic renal disease and tetany.
Chloride measurements from the epoc Blood Analysis System are used in the treatment of electrolyte and metabolic disorders.
Total Carbon Dioxide measurements from the epoc Blood Analysis System are used in the diagnosis and treatment of disorders associated with changes in body acid-base balance.
Glucose measurements from the epoc Blood Analysis System are used in the diagnosis and treatment of carbolism disorders, including diabetes mellitus and idiopathic hypodycemia, and of pancreatic islet cell tumors.
Lactate measurements from the epoc Blood Analysis System are used to evaluate the acid-base status and are used in the diagnosis and treatment of lactic acidosis (abnormally high acidity of the blood),
Blood Urea Nitrogen measurements from the epoc Blood Analysis System are used in the diagnosis and treatment of certain renal and metabolic diseases.
Creatinine measurements from the epoc Blood Analysis System are used in the diagnosis and treatment diseases and in monitoring renal dialysis.
Hematocrit measurements from the epoc Blood Analysis System are used to distinguish normal states of blood volume, such as anemia and erythrocytosis.
The epoc® Blood Analysis System is an in vitro diagnostic device system for the quantitative testing of blood gases, electrolytes, and metabolites in venous, arterial, and capillary whole blood samples. The epoc® System is comprised of three (3) major subsystems: epoc® Host, epoc® Reader and epoc® BGEM Test Card.
- epoc® Blood Gas Electrolyte Metabolite (BGEM) Test Card: single-use, device with . port for blood sample introduction which contains the sensor configurations for testing Sodium (Na+), Potassium (K+), Calcium (Ca++), Chloride (C)I-, pH, partial pressure of carbon dioxide (pCO2), partial pressure of oxygen (pO2), Glucose (Glu), Lactate (Lact), Creatinine (Crea), Hematocrit (Hct), Blood Urea Nitrogen (BUN) and Total Carbon Dioxide (TCO2).
- . epoc® Reader: portable, battery-powered device component that measures electrical signals from the test card sensors during blood testing and transmits this sensor data wirelessly via Bluetooth to the epoc Host.
- . epoc® Host: mobile computer-based device component for calculating test results from the sensor data sent by the epoc Reader and displaying these results on the graphical user interface. The epoc Host component can be physically connected to the Reader by a cradle component. The epoc Host also incorporates an internal laser barcode scanner for scanning patient and operator IDs. The epoc Host component currently runs on Microsoft® Windows Mobile 6.5 Operating System (OS).
Based on the provided text, the "epoc® Blood Analysis System with NXS Host" is being submitted as a modified device, and the submission primarily focuses on hardware and software updates to the epoc Host component. The document explicitly states that "No performance data was required to evaluate the changes introduced with the alternate epoc Host component" and that there is "no change to labeled performance claims." Therefore, there isn't a comprehensive study proving the device meets new acceptance criteria. Instead, the submission argues for substantial equivalence to a previously cleared predicate device by demonstrating that the modifications do not negatively impact safety and effectiveness.
Here's an analysis based on the provided information, addressing your points where possible:
1. Table of Acceptance Criteria and Reported Device Performance
The document does not detail specific acceptance criteria for new performance claims or provide a table of performance data because the submission states "No performance data was required to evaluate the changes introduced with the alternate epoc Host component." The device is intended to meet the same performance specifications as the predicate device. The change is in the host component (hardware and OS), not the core measurement technology or labeled performance.
The submission is essentially asserting that the "epoc® Blood Analysis System with NXS Host" (Modified Device) is substantially equivalent to the "epoc® Blood Analysis System" (Predicate Device) and thus relies on the predicate device's existing performance data and acceptance criteria.
2. Sample Size Used for the Test Set and Data Provenance
Not applicable. No new clinical performance or analytical performance study with a test set is described. The submission focuses on verification and validation activities for hardware, software, and usability to support substantial equivalence due to a change in the host component.
3. Number of Experts Used to Establish Ground Truth for the Test Set and Their Qualifications
Not applicable. As no new performance study is described, there's no mention of experts establishing ground truth for a test set.
4. Adjudication Method for the Test Set
Not applicable. No new performance study is described.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done
No. This device is an in vitro diagnostic system for quantitative testing of blood parameters, not an imaging AI device that would typically involve a multi-reader multi-case study with human readers.
6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) Was Done
The device is an analytical instrument. Its performance is inherent to the system (sensors, reader, software). While the software is a key component, the "standalone" concept as typically applied to AI in imaging doesn't directly map. However, the document states: "No performance data was required to evaluate the changes introduced with the alternate epoc Host component." This implies that the core analytical performance (algorithm) is considered unchanged from the predicate device and its previous clearances. Verification and validation activities were done on the new host component, but not necessarily a "standalone" re-evaluation of the core measurement algorithm's performance on a new dataset.
7. The Type of Ground Truth Used
Not applicable for new performance data. The device measures objective chemical and physical properties of blood samples. The ground truth for such devices is typically established through reference methods and calibrated controls, not expert consensus or pathology in the same way as an imaging AI. The submission relies on the established ground truth methodologies for the predicate device.
8. Sample Size for the Training Set
Not applicable. This document describes a modification to an existing IVD device (updating hardware and operating system for the host component). It does not describe the development of a new AI algorithm or machine learning model that would involve a "training set" in the conventional sense. The "epoc Host Application Software has been modified to support the Android-based Operating System" but this is a software porting/adaptation, not a new algorithm being trained on data.
9. How the Ground Truth for the Training Set Was Established
Not applicable, as there is no mention of a training set for a new AI algorithm.
Summary of the Study and Why it Meets (Implied) Acceptance Criteria:
The study detailed in this 510(k) submission is not a clinical performance study generating new acceptance criteria or performance data for the analytes measured. Instead, it is a Special 510(k) submission designed to demonstrate that hardware and software updates to the epoc Host component of the epoc® Blood Analysis System do not alter the safety or effectiveness of the device and thus maintain substantial equivalence to the previously cleared predicate device.
The "study" or evidence provided to meet acceptance criteria consists of:
- Verification and Validation Activities: "All software, hardware and usability verification and validation activities were performed in accordance to relevant standards, established plans and protocols and Design Control procedures."
- Meeting Acceptance Criteria: "Testing verified all acceptance criteria were met." (These are acceptance criteria related to software, hardware function, and usability for the new host component, ensuring it performs its intended role without impacting the core analytical performance).
- Risk Management: A risk management process compliant with EN ISO 14971:2012 and ISO 14971:2007 was performed, concluding that "the overall residual risk of the epoc System with the epoc NXS host is acceptable."
- Cybersecurity Information: Cybersecurity design inputs were established, risks assessed, and controls designed within the software.
The core argument for meeting acceptance criteria (by proving substantial equivalence) is that:
- There is no change to the intended use or indications for use.
- There is no change to the fundamental scientific technology (the epoc Reader and Test Card, which contain the sensors, remain unchanged).
- There is no change to labeled performance claims.
- There is no change to the principle of operation.
- There is no change to cartridge (test card) calibrator formulation and technology.
Therefore, the "acceptance criteria" here are implicitly that the new epoc NXS Host component, through its verification and validation, functions correctly, safely, and does not introduce new risks, thereby maintaining the established performance and safety profile of the predicate device.
Ask a specific question about this device
(264 days)
The Blood Urea Nitrogen and Total Carbon Dioxide tests, as part of the epoc Blood Analysis System, is intended for use by trained medical professionals as an in vitro diagnostic device for the quantitative testing of samples of heparinized or un-anticoagulated arterial, venous or capillary whole blood in the laboratory or at the point of care.
Blood Urea Nitrogen measurements from the epoc Blood Analysis System are used in the diagnosis and treatment of certain renal and metabolic diseases.
Total Carbon Dioxide measurements from the epoc Blood Analysis System are used in the diagnosis and treatment of disorders associated with changes in body acid-base balance.
The epoc Blood Analysis System is an in vitro diagnostic device system for the quantitative testing of blood gases, electrolytes, and metabolites in venous, arterial, and capillary whole blood samples. The epoc System is comprised of 3 major subsystems: epoc Host, epoc Reader and epoc BGEM Test Card. The main accessory used with the epoc System includes the epoc Care-Fill Capillary Tubes used to collect and introduce capillary blood samples into the epoc Test Card.
The epoc Blood Analysis System was previously cleared for prescription use to quantitate pH, pCO2, pO2, Na, K, iCa, Cl, Glu, Lact, Crea, and Hct in arterial, venous, and capillary blood samples per K061597, K090109, K092849, K093297, and K113726. This premarket notification submission adds blood urea nitrogen (BUN) and total carbon dioxide (TCO2) quantitation to the epoc BGEM Test Card and Blood Analysis System.
The epoc Blood Urea Nitrogen Test and epoc Total Carbon Dioxide Test, as part of the epoc Blood Analysis System, are intended for use by trained medical professionals as an in vitro diagnostic device for quantitative testing of heparinized or un-anticoagulated arterial, venous or capillary whole blood.
The acceptance criteria and device performance are described in several studies:
Acceptance Criteria and Device Performance:
| Study | Acceptance Criteria | Reported Device Performance |
|---|---|---|
| Analytical Sensitivity (LoB, LoD, LoQ per CLSI EP17-A2) | Not explicitly stated as acceptance criteria, but demonstrates detection limits. | BUN: LoB 2 mg/dL, LoD 3 mg/dL, LoQ 3 mg/dL |
| TCO2: LoB 4.0 mM, LoD 4.3 mM, LoQ 4.3 mM | ||
| Linearity (per CLSI EP06-A) | Not explicitly stated as acceptance criteria, but demonstrates linearity across reportable range. | BUN (4-119 mg/dL): Slope 1.020, Intercept 0.4, R 0.9989 |
| TCO2 (4-49 mmol/L): Slope 0.903, Intercept 3.32, R 0.9997 | ||
| Precision (Aqueous Controls) (CLSI EP05-A3) | Not explicitly stated as acceptance criteria, but demonstrates precision. | BUN High Level (51.7 mg/dL): SWR 1.01 (2.0% CV), ST 1.16 (2.3% CV) |
| BUN Low Level (7.1 mg/dL): SWR 0.30 (4.2% CV), ST 0.32 (4.5% CV) | ||
| TCO2 High Level (30.7 mmol/L): SWR 0.82 (2.7% CV), ST 0.92 (3.0% CV) | ||
| TCO2 Low Level (16.2 mmol/L): SWR 0.88 (5.4% CV), ST 1.02 (6.3% CV) | ||
| Interference (CLSI EP07-A2) | Unacceptable interference bias defined as producing a significant error more than 5% of the time. | Clinically significant interfering substances for BUN and TCO2 are itemized and reported. Various exogenous and endogenous interferences were tested and found to be clinically insignificant below certain thresholds. |
| Clinical Field Precision (Aqueous Controls) (CLSI EP05-A3) | Not explicitly stated as acceptance criteria, but demonstrates precision in a clinical setting. | BUN Level 1 (52.1 mg/dL): SWR 1.06 (2.0%), Total Reproducibility 1.54 (3.0%) |
| BUN Level 2 (17.7 mg/dL): SWR 0.45 (2.5%), Total Reproducibility 1.11 (6.3%) | ||
| BUN Level 3 (7.1 mg/dL): SWR 0.24 (3.4%), Total Reproducibility 0.26 (3.7%) | ||
| TCO2 Level 1 (15.9 mM): SWR 0.44 (2.8%), Total Reproducibility 0.50 (3.1%) | ||
| TCO2 Level 2 (19.7 mM): SWR 0.66 (3.4%), Total Reproducibility 0.78 (3.9%) | ||
| TCO2 Level 3 (30.4 mM): SWR 0.58 (1.9%), Total Reproducibility 1.05 (3.4%) | ||
| Clinical Field Precision (Whole Blood) | Not explicitly stated as acceptance criteria, but demonstrates precision in a clinical setting. | BUN Hi-Syringe (57.4 mg/dL): %CV 2.3% |
| BUN Lo-Cap Tube (7.6 mg/dL): %CV 7.0% | ||
| TCO2 Hi-Syringe (36.5 mM): %CV 1.5% | ||
| TCO2 Lo-Cap Tube (13.5 mM): %CV 3.5% | ||
| Method Comparison (BUN) (CLSI EP09-A3) | Not explicitly stated as a numerical acceptance criterion, but implies a high correlation with the reference method. | Comparing epoc BUN to Roche Cobas 8000: Slope 0.985, Intercept 0.3, R 0.998, Mean Bias at 26 mg/dL -0.1+0.2 |
| Method Comparison (TCO2) | Not explicitly stated as a numerical acceptance criterion, but implies a high correlation with the reference method. | Comparing epoc TCO2 to i-STAT-CHEM8+: Slope 1.039, Intercept -0.8, R 0.974, Mean Bias at 20 mM 0.0+0.2 |
| Matrix Comparison: Anticoagulant | No significant difference between results in Li-heparinized, Na-heparinized, and non-anticoagulated blood samples | Concluded no significant difference in BUN and TCO2 results. |
Study Information:
-
Sample sizes used for the test set and the data provenance:
- Analytical Sensitivity (LoB, LoD, LoQ): Test samples were prepared from dialyzed whole blood. The specific number of samples or runs is not explicitly stated, but the study was conducted according to CLSI EP17-A2.
- Linearity: Multiple whole blood samples were used, spanning the reportable range. Conducted per CLSI EP06-A. Specific number not provided.
- Precision (Aqueous Controls): 320 replicates for each level of both BUN and TCO2. These were in-house measurements.
- Clinical Field Precision (Aqueous Controls): N=170 for BUN Level 1, 171 for Level 2, 168 for Level 3. N=172 for TCO2 Level 1, 170 for Level 2, 169 for Level 3. Data provenance is from "three different clinical sites."
- Clinical Field Precision (Whole Blood): N=134-136 for BUN samples, N=134-139 for TCO2 samples, depending on the type (syringe/cap tube) and level (high/NB/low). Data provenance is from "three different clinical sites."
- Precision (Duplicate Epoc Test Results): Over 430 patient tests run in duplicate. "Approximately equal numbers of venous, arterial and capillary samples." Data provenance not explicitly stated (e.g., country of origin), assumed to be from clinical sites in the context of "Clinical Field Precision." This is prospective clinical data.
- Method Comparison (BUN): N=433 venous, arterial, and capillary blood samples. Performed at "three clinical sites." This is prospective clinical data.
- Method Comparison (TCO2): N=574 venous, arterial, and capillary patient samples. Performed at "three clinical sites." This is prospective clinical data.
- Matrix Comparison: Anticoagulant: Over 60 volunteer donors, with samples further aliquoted into 3 vacutainers each. Data provenance not explicitly stated.
-
Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable. This device is a quantitative diagnostic test for chemical analytes (BUN, TCO2), not an imaging or qualitative diagnostic device requiring expert interpretation for ground truth. The ground truth for analytical performance studies is typically established using reference methods (e.g., IDMS-traceable laboratory system) or prepared reference materials.
-
Adjudication method for the test set: Not applicable. The ground truth for quantitative chemical analytes is established by reference methods or gravimetric preparation, not through human adjudication.
-
If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable. This device is a diagnostic testing system for chemical analytes, not an AI-assisted diagnostic imaging or qualitative interpretation tool for human readers.
-
If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Yes, the entire performance characterization (analytical sensitivity, linearity, precision, interference, method comparison, and matrix comparison) represents standalone algorithm/device performance. The device provides quantitative results directly. Human-in-the-loop performance is about accuracy of human readers, and the clinical field precision study assesses the precision of the device in the hands of intended users, not the interpretative performance of those users.
-
The type of ground truth used (expert consensus, pathology, outcomes data, etc):
- Analytical Sensitivity, Linearity, Precision: Ground truth established via prepared reference materials (dialyzed whole blood, gravimetric mixtures of high/low samples) and aqueous controls with known concentrations.
- Method Comparison (BUN): Ground truth established by an "IDMS-traceable plasma/serum-based laboratory system" (
Roche Cobas 8000). - Method Comparison (TCO2): Ground truth established by a "whole blood point-of-care system" (
i-STAT-CHEM8+), which is also a predicate device. - Interference and Matrix Comparison: Comparisons were made against control samples (e.g., solvent added, or anticoagulant-free) to assess the impact of interfering substances or different matrices.
-
The sample size for the training set: Not applicable. This document describes the performance of a chemical analyte detection system, not a machine learning or AI model that requires a training set.
-
How the ground truth for the training set was established: Not applicable, as there is no training set for this device.
Ask a specific question about this device
Page 1 of 1