(231 days)
The Lactate test, as part of the epoc Blood Analysis System, is intended for use by trained medical professionals as an in vitro diagnostic device for the quantitative testing of samples of heparinized or un-anticoagulated arterial, venous or capillary whole blood in the laboratory or at the point of care in hospitals, nursing homes or other clinical care institutions.
Lactate measurements from the epoc Blood Analysis System are used to evaluate the acid-base status and are used in the diagnosis and treatment of lactic acidosis (abnormally high acidity of the blood).
The epoc Lactate Test is being added as an additional sensor to the existing single use test card that is used with the epoc Blood Analysis System. This test card is inserted into the epoc Reader and all analytical steps are performed automatically. Patient and user information may be entered into the mobile computing device (epoc Host) during the automated analysis cycle.
The epoc Blood Analysis System is an in vitro analytical system comprising a network of one or more epoc Readers designed to be used at the point of care (POC). The readers accept an epoc single use test card containing a group of sensors that perform diagnostic testing on whole blood. The blood test results are transmitted wirelessly to an epoc Host, which displays and stores the test results.
Here's a summary of the acceptance criteria and study information for the epoc Lactate Test, based on the provided document:
1. Table of Acceptance Criteria and Reported Device Performance (Lactate only)
| Acceptance Criteria Category | Specific Metric (Lactate) | Acceptance Criteria (Implicit) | Reported Device Performance (Lactate) |
|---|---|---|---|
| Precision (Aqueous Controls) | Within-Device CV% | Not explicitly stated but expected to be low for clinical accuracy. | L1: 4.9% (WD), 6.3% (Total); L3: 3.1% (WD), 4.7% (Total) |
| Precision (Blood Samples - Site 1) | %CV | Not explicitly stated. | WB L1: 6.0% (Phlebotomist 1), 3.3% (Phlebotomist 2) |
| Precision (Aqueous Controls - Site 1) | %CV | Not explicitly stated. | L3: 3.3% (RN 1), 2.9% (Anesthesia Tech); L2: 1.8% (RN 2), 2.8% (Resp Therapist) |
| Linearity/Reportable Range | Test Range (mM) | 0.3 - 20 mmol/L (as specified in technology comparison) | 0.001 - 20.1 mM (appears to be "test range" based on the table, indicating it covers the specified range) |
| Traceability | To NIST standards | Device calibrated and QC materials traceable to NIST. | Explicitly states: "epoc System is calibrated is against methods traceable to NIST standards." and "Calibration verification uses commercially available calibration verification fluids whose concentration values are traceable to NIST standards." |
| Detection Limit | Statistically discernable from Limit of Blank | Test's low end of reportable range >= Limit of Detection | Low end of reportable range (0.30 mM) is >= Limit of detection and statistically discernable from Limit of Blank (0.21 mM). |
| Analytical Specificity (Interference Bias) | Unacceptable bias defined as >5% significant error. | Max 5% significant error. | Several interfering substances listed with specific bias values; most found to be "insignificant" at tested levels. |
| Method Comparison (vs. Predicate) | R² | Generally >0.95 for good correlation. | 0.9711 (overall), 0.9769 (venous), 0.9829 (arterial), 0.9653 (capillary) |
| Method Comparison (vs. Predicate) | Slope | Ideally close to 1. | 0.967 (overall) |
| Method Comparison (vs. Predicate) | Intercept | Ideally close to 0. | 0.132 (overall) |
| Method Comparison (vs. Predicate) - Bias | Average Bias for decision levels | Expected to be small, with narrow 95% CI. | At 2.2mM: 0.061 (all) with 95% CI ± 0.119; At 5.0mM: -0.031 (all) with 95% CI ± 0.084 |
| Effect of Anticoagulant (vs. Predicate) | R² | Generally >0.95 for good correlation. | 0.9916 |
| Effect of Anticoagulant (vs. Predicate) | Slope | Ideally close to 1. | 1.036 |
| Effect of Anticoagulant (vs. Predicate) | Intercept | Ideally close to 0. | -0.045 |
2. Sample Size and Data Provenance
- Test Set for Method Comparison:
- Sample Size: 373 patient samples for overall method comparison; broken down into 126 venous, 73 arterial, 174 capillary samples for matrix effects.
- Data Provenance: Field trials at several hospitals on "patient samples of whole blood at various locations." This indicates prospective, real-world data from multiple sites (likely within North America given the FDA submission).
- Test Set for Blood Precision:
- Sample Size: 15 samples per user at Site 1, and unspecified number at Site 2 (likely similar).
- Data Provenance: Field trials at two (2) hospitals on "volunteer samples of whole blood by potential end users." This indicates prospective, real-world data from multiple sites.
- Test Set for Aqueous Precision:
- Sample Size: 15 samples per user per QC level at Site 1, and unspecified number at Site 2 (likely similar).
- Data Provenance: Field trials at two (2) hospitals on commercially available control fluids by potential end users.
- Test Set for Anticoagulant Effect:
- Sample Size: 60 samples (43 from hospital POC sites, 17 from in-house studies).
- Data Provenance: Patient samples from hospital POC sites and in-house studies.
3. Number of Experts and their Qualifications (for Test Set Ground Truth)
- The document describes the predicate device as the "gold standard" for comparison. The ground truth for the method comparison studies was established by the predicate device (i-Stat™ Lactate Test using i-Stat™ Model 300 Portable Clinical Analyzer).
- No specific number of human experts or their qualifications for establishing ground truth are mentioned, as the comparison is against another established device. However, the predicate device itself would have undergone its own validation with expert input.
4. Adjudication Method
- Given that the ground truth is established by a predicate analytical device, there is no human adjudication method (like 2+1 or 3+1) described or applicable in this context. The comparison is quantitative against readings from the predicate device.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study
- A formal MRMC comparative effectiveness study, as typically performed for diagnostic imaging devices involving human readers, was not conducted or described.
- The studies involved different "users" (phlebotomists, RNs, Anesthesia Techs, Resp Therapists) performing tests, which is a form of multi-reader study, but it's focused on device precision/reproducibility across different operators rather than assessing AI assistance for human diagnostic performance. Therefore, there is no effect size reported for human readers improving with/without AI assistance.
6. Standalone Performance Study
- Yes, a standalone performance study was done. The entire document describes the standalone performance of the epoc Lactate Test (algorithm and device combined) against a predicate device and established analytical standards. The reported device performance metrics in the tables (precision, linearity, method comparison slope, intercept, R², bias) are all measures of the device's standalone performance. There is no human-in-the-loop component for result interpretation.
7. Type of Ground Truth Used
- The primary ground truth for the clinical and non-clinical studies is:
- Readings from a legally marketed predicate device (i-Stat™ Lactate Test using i-Stat™ Model 300 Portable Clinical Analyzer) for method comparison studies.
- NIST traceable standards for calibration, quality control, and linearity studies.
- Pooled human serum and blood samples (spiked with known interferents or aged to increase lactate) for analytical specificity and blood precision studies.
8. Sample Size for the Training Set
- The document does not explicitly state the sample size for the training set for the epoc Lactate Test development. This type of submission (510(k) for an IVD) typically focuses on validation data rather than internal development/training data for the algorithm.
9. How the Ground Truth for the Training Set was Established
- As the training set size is not disclosed, the method for establishing its ground truth is also not explicitly described in this document. However, given the nature of the device (a quantitative sensor measurement system), the training/development likely involved:
- Controlled reference materials with known lactate concentrations.
- Comparison to reference laboratory methods known to be accurate and traceable to NIST standards.
{0}------------------------------------------------
JUN - 9-2010
ероса
2060 Walkley Road Ottawa Ontario, Canada K1G 3P5
510(k) SUMMARY
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.
The assigned 510(k) number is: k-093297.
Summary Prepared: June 07, 2010
| Submitted by: | Epocal Inc.2060 Walkley Road, Ottawa, Ontario, Canada K1G 3P5Telephone: (613) 738-6192Fax: (613) 738-6195 |
|---|---|
| Contact: | Roy Layer |
Director of Quality Assurance and Regulatory Affairs.
5.1 Identification of the Device
| Device Name: | Acid, Lactic, Enzymatic Method |
|---|---|
| Proprietary / Trade Name: | epoc Lactate Test |
| Common Name: | Lactate acid test system |
| Classification Name: | Acid, Lactic, Enzymatic Method |
| Device Classification: | I (Class II with limitation of exemption) |
| Regulation Number: | 862.1450 |
| Panel: | Clinical Chemistry |
| Product Code: | KHP |
Identification of the Predicate Device 5.2
i-Stat™ Lactate Test using i-Stat™ Model 300 Portable Clinical Analyzer
Description of the New Device 5.3
The epoc Lactate Test is being added as an additional sensor to the existing single use test card that is used with the epoc Blood Analysis System. This test card is inserted into the epoc Reader and all analytical steps are performed automatically. Patient and user information may be entered into the mobile computing device (epoc Host) during the automated analysis cycle.
The epoc Blood Analysis System is an in vitro analytical system comprising a network of one or more epoc Readers designed to be used at the point of care (POC). The readers accept an epoc single use test card containing a group of sensors that perform diagnostic testing on whole blood. The blood test results are transmitted wirelessly to an epoc Host, which displays and stores the test results.
{1}------------------------------------------------
The epoc System is intended for use by trained medical professionals as an in vitro diagnostic device for the quantitative testing of samples of whole blood.
The test card panel configuration currently includes sensors for Sodium Na, Potassium K, Ionized Calcium iCa, pH, pCO2, pO2, Glucose and Hematocrit Hct. This submission adds Lactate (Lact) to this list of approved tests.
To perform a blood test, a new test card is inserted into a card reader's card slot with white label face down. When fully inserted, the test card is automatically engaged in the reader.
The card insertion process:
- Brings the cards sensor module into contact with the reader's electrical contact . array;
- Brings the card's measurement region, which is the fluidic channel above the . sensor array, into thermal contact with the reader's heater assembly for heating the measurement region to 37℃;
- Actuates the opening of the fluidic valve in the card and causes delivery of . calibrator fluid from the reservoir to the measurement region.
After calibration, and upon a prompt by the reader (LED visual and audio beep), the user introduces a blood sample for measurement through the blood sample port to the card's measurement region. When sensors are contacted by the blood sample they generate electrical signals proportional to analyte concentrations in the blood sample, which are transmitted wirelessly by the Reader to the epoc Host displays and stores the blood test results.
Changes to the epoc Blood Analysis System required to introduce the Lactate test include:
- Developing a new Lactate sensor and adding it to the existing epoc test card, . which was already designed to accommodate additional sensors;
- Modifications to the existing EpocHost software application to accommodate the . new test:
- . Labeling changes including indications for use for the Lactate test.
Comparison of Technological Characteristics To Predicate 5.4 Device
| 510(k) # | epoc Blood Analysis System | i-STAT Model 300 | Same / Different |
|---|---|---|---|
| Item | Device | Predicate | |
| Intended use | The Lactate test as part of epoc BloodAnalysis System is intended for use bytrained medical professionals as an invitro diagnostic device for thequantitative testing of samples ofheparinized or un-anticoagulatedarterial, venous or capillary wholeblood using the BGEM (Blood GasElectrolyte and Metabolytes) test cardpanels. | The i-STAT Model 300 Portable ClinicalAnalyzer is intended to be used bytrained medical professionals for usewith i-STAT test cartridges. i-STATcartridges comprise a variety ofclinical chemistry tests and testpanels. | same |
| Where used | hospital | hospital | same |
{2}------------------------------------------------
| Measuredparameters | pH, pCO2, pO2, Na, K, iCa, Hct, Gluc, Lact | pH, pCO2, pO2, Na, K, iCa, Hct, Gluc, Lact | same | |||
|---|---|---|---|---|---|---|
| Calculatedparameters | TCO2, HCO3, BE, SO2, Hgb | TCO2, HCO3, BE, SO2, Hgb | same | |||
| Sample type | Venous, arterial and capillary whole blood | Venous, arterial and capillary whole blood | same | |||
| Reportableranges | pH | 6.5 - 8.0pH units | pH | 6.5 - 8.2pH units | different | |
| pCO2 | 5 - 250mm Hg | pCO2 | 5 - 130mm Hg | different | ||
| pO2 | 5 - 750mm Hg | pO2 | 5 - 800mm Hg | same | ||
| Na | 85 - 180mmol/L | Na | 100 - 180mmol/L | different | ||
| K | 1.5 - 12mmol/L | K | 2.0 - 9.0mmol/L | different | ||
| iCa | 0.25 - 4mmol/L | iCa | 0.25 - 2.5mmol/L | different | ||
| Hct | 10 - 75%PCV | Hct | 10 - 75%PCV | same | ||
| Gluc | 20 - 700mg/dL | Gluc | 20 - 700mg/dL | same | ||
| Lact | 0.3 - 20mmol/L | Lact | 0.3 - 20mmol/L | same | ||
| TCO2 | 1 - 85mmol/L | TCO2 | 5 - 50mmol/L | different | ||
| HCO3 | 1 - 85mmol/L | HCO3 | 1 - 85mmol/L | same | ||
| BEecf | -30 - +30mmol/L | BEecf | -30 - +30mmol/L | same | ||
| BEB | -30 - +30mmol/L | BEB | -30 - +30mmol/L | same | ||
| SO2 | 0 - 100% | SO2 | 0 - 100% | same | ||
| Hb | 3.3 - 25g/dL | Hb | 3 - 26g/dL | same | ||
| Samplevolume | Non-volumetric over 95 µL | 100µL | same | |||
| Test card | Unit-use card withon-board calibrator in sealedreservoiran electrochemical multi-sensor arrayport for sample introductionfluid waste chamber | Unit-use cartridge withon-board calibrator in sealedreservoiran electrochemical multi-sensor arrayport for sample introductionfluid waste chamber | same | |||
| Test cardstorage | Room temperature until expiry date | Fridge storage until expiry dateincluding max 2 weeks at room temperature | different | |||
| Sensor array | A laminated foil sensor module | A micro-fabricated chip-set | different | |||
| Tests/sensorcomponents | pH - PVC ion selective electrodepCO2 - QH modified Severinghaus typepO2 - membrane coated gold cathodeNa - PVC ion selective electrodeK - PVC ion selective electrodeiCa - PVC ion selective electrodeGlu - glucose oxidase basedamperometric peroxide detectionLact - lactate oxidase basedamperometric peroxide detectionHct - conductivity, gold electrodes | pH - PVC ion selective electrodepCO2 - QH modified Severinghaus typepO2 - membrane coated gold cathodeNa - PVC ion selective electrodeK - PVC ion selective electrodeiCa - PVC ion selective electrodeGlu - glucose oxidase basedamperometric peroxide detectionLact - lactate oxidase basedamperometric peroxide detectionHct - conductivity, gold electrodes | same | |||
| Analyzercomponents | Two housings;1 - The reader comprisingOrifice for test card introductionelectrical connector to cardheater for 37°C operationmechanical card engagement device for- making electrical contact to card's sensors- for rupture of calibrator reservoir- moving calibrator to sensors- engaging heaters with cardop-amp sensor signal detectorsiQC monitoring devices | A single housing comprisingOrifice for test card introductionelectrical connector to cardheater for 37°C operationmechanical card engagement device for- making electrical contact to card's sensors- for rupture of calibrator reservoir- moving calibrator to sensors- engaging heaters with cardop-amp sensor signal detectorsiQC monitoring devices | different |
{3}------------------------------------------------
| MUXA/DBluetooth stack for wirelesstransmission of digitized rawsensor signals to computingdevicebar code scanner for acquiringcard infointernal electronic reader self-testcircuit2 - The computing device comprisinga PDAmicroprocessor | MUXA/Dwire transmission of digitized rawsensor signals to computingsubsystem in same housingn/ainternal and external electronicreader self-test circuitmicroprocessormemory | samesamedifferentdifferentdifferentsamesame | |
|---|---|---|---|
| memorycolor LCD display | monochrome LCD display | different | |
| keyboard | keyboard | same | |
| i/o for communicating test results | i/o for communicating test results | same | |
| to other devices | to other devices | ||
| software to control the test andcalculate analytical values fromraw sensor signals | software to control the test andcalculate analytical values fromraw sensor signals | same | |
| battery operated withrechargeable batteries via plug inplug-in power supply | battery operated withrechargeable batteries viaexternal power supply indownloader cradle | same | |
| Measurementtemperature | 37°C | 37°C | same |
| Measurement Calibrate test card-introduce sample- | Introduce sample-calibrate test | different | |
| sequence | measure | cartridge-measure | |
| Measurementtime | 35sec from sample introduction | 200 sec from sample introduction | different |
| Error | iQC system to detect user errors | iQC system to detect user errors | same |
| detection | iQC system for reader self-check | iQC system for reader self-check | same |
| iQC system to detect card non- | iQC system to detect card non- | same | |
| conformance | conformance |
Figure 5.1 - Table - Comparing epoc Device Performance Characteristics With Predicate Device
The epoc System has the same intended use and utilizes the same test methodologies as the predicate device. Most of the system components are very similar to the predicate device. Differences between the epoc device and the predicate device have no significant effect on the safety or effectiveness of the system.
{4}------------------------------------------------
Summary of Non-Clinical Test Performance in Support of 5.5 Substantial Equivalence
5.5.1 Aqueous precision
.
Experiments were performed in-house to demonstrate the precision of the epoc test methods. The table below shows the results of a twenty day precision study using performed on 4 lots using aqueous controls at two levels L1 and L3 for the blood gases, electrolytes and metabolytes.
| Lactate | All | |
|---|---|---|
| mM | L1 | L3 |
| N | 320 | 320 |
| Mean | 7.99 | 0.94 |
| SWD | 0.39 | 0.03 |
| SDD | 0.32 | 0.03 |
| ST | 0.51 | 0.04 |
| WD CV% | 4.9% | 3.1% |
| Total CV% | 6.3% | 4.7% |
Figure 5.2 - Table - 20 Day Precision Study Data
Linearity/Reportable Range 5.5.2
This study was performed in-house using blood samples as per CLSI EP6-A recommendations for evaluation of linearity. A total of nine blood samples were prepared starting with two pools of blood, which were evaluated versus an in-house standard method with traceability to NIST standards. Regression analysis was performed as per CLSI EP6-A. The summary is given in the table in Figure 5.3.
| " est Range "Slope | C | |||
|---|---|---|---|---|
| CmMl20.1ﺗﺄﺳﻴﺴﺎﺕ ﺳﻨﺔ 1999 ﻓﻲ ﺇﺳﺒﺎﻧﻴﺎ ﻓﻲ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤ | .001 | 0.77 | n qaqaﻟﻠﺴﻠﺔ ﺍﻟﻤﺴﺎﺣﺔ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ |
Figure 5.3 – Table - In House Whole Blood Linearity
5.5.3 Traceability
The epoc System is calibrated is against methods traceable to NIST standards.
The epoc System's test card comprises an on-board calibration material, prepared gravimetrically and assayed on reference systems calibrated with traceability to NIST standards.
Calibration verification uses commercially available calibration verification fluids whose concentration values are traceable to NIST standards.
Quality control materials are commercially available fluids with concentrations traceable to NIST standards.
5.5.4 Detection Limit
This study was performed in-house as per CLSI EP6-A recommendations for evaluation limits of detection and quantification. The low end of the reportable range for the EPOC lactate test (0.30 mM) is greater than or equal to the limit of detection and is statistically discernable from the limit of blank ( 0.21 mM).
{5}------------------------------------------------
Analytical Specificity 5.5.5
Interference testing4 was performed in-house on the epoc lactate sensor. In each of these tests a pooled human serum was aliguoted into two samples. The test sample was spiked by addition of interferent, while the control sample was spiked by the addition of the solvent of the interferent. The lactate bias between the mean of six replicates on both the control sample and the test sample with added interferent was calculated.
Unacceptable interference bias was defined as producing a significant error more than 5% of the time.
Significant interfering substances are itemized below:
- · Acetaminophen will have no significant effect up to 0.81 mM after which it will increase the lactate reading up to 306 µM/mM Tylenol. Because the therapeutic upper limit for acetaminophen is 0.20 mM, interfering levels of acetaminophen should only be encountered in overdose situations
- · Iodide will decrease the lactate reading up to -1.2mM/mM of Iodide up to an Iodide concentration of 0.67 mM. Above 0.67 mM Iodide the decrease will be -1.2mM.
- Bromide will have no significant effect up to 25.4 mM after which it will decrease the lactate reading up to 14.6 µM/mM Bromide.
- · Thiocyanate will have no significant effect up to 2.7 mM after which it will decrease the lactate reading by up to 96.6 µM/mM thiocyanate.
- N-Acetylcysteine will have no significant effect up to 3.7 mM after which it will decrease the lactate reading by up to 96.3 uM/mM N-Acetylcysteine.
Ethylene glycol ingestion and metabolism has been shown to produce falsely elevated lactate measurements*. Ethylene glycol plus three metabolism products -Glycolic Acid, Glyoxylic Acid and Oxalic Acid - were tested for interference. Ethylene Glycol and Oxalic Acid do not interfere significantly.
- · Glycolic Acid will have no significant effect up to 0.87 mM after which it will increase the lactate reading up to 142 µM/mM glycolic acid.
- Glyoxylic Acid will have no significant effect up to 0.85 mM after which it will increase the lactate reading up to 373 uM/mM qlyoxylic acid.
- CMAJ, April 10, 2007, 176(8), p.1097 "Falsely elevated point-of-care lactate measurement after ingestion of ethylene glycol"
The following levels of exogenous interferences were tested and found to be insignificant: 1.66mM (25mg/dL) acetaminophen, 630µmol/L (12.5mg/dL) Na ascorbate. 20mmol/L (588 mg/dL) citrate. 100 umol/L (~2mg/dL) L-dopa, 9mmol/L (263mg/dL) EDTA, 4.84mmol/L (30mg/dL) ethylene glycol, 105 µmmol/L (0.441mg/dL) Na fluoride, 71 µmol/L Methyldopa, 2.55mmol/L oxidized glutathione, 2.55mmol/L reduced glutathione, 132 umol/L (1.0mg/dL) hydroxyurea, 292µmol/L (4mg/dL) isoniazide (nydrazid), 81 µmol/L (1.5 mg/dL) K Oxalate, 0.037 mmol/L (1.2 mg/dL) Quinidine.
The following levels of endogenous interferences were tested and found to be insignificant: +342µmol/L (+29.0mg/dL) bilirubin conjugated, +342 (+20.1mg/dL) bilirubin unconjuqated, +13mmol/L (+503.1mg/dL) cholesterol, +1500umol/L (+18mg/dL) L-cysteine, +0.8% lipids, pH (+0.4, -0.4), 3% to 10% total protein, 1.4 mM (+ 23.5 mg/dL) Uric Acid.
{6}------------------------------------------------
Low hematocrit did not interfere down to a level of 21 % hematocrit and high hematocrit did not interfere up to a level of 61 % hematocrit. Triglycerides did not show significant interference up to a level of 37 mM (1430 ma/dL).
Summary of Clinical Tests Submitted in Support of 5.6 Substantial Equivalence
5.6.1 Method comparison with Predicate Device
The method comparison studies were performed in field trials at several hospitals on patient samples of whole blood at various locations. Patient specimens were venous, arterial and capillary. The method comparison was against the predicate device.
| epoc Lactate vs. i-STAT | |
|---|---|
| N | 373 |
| Sxx | 0.215 |
| Syy | 0.530 |
| intercept | 0.132 |
| slope | 0.967 |
| Syx | 0.948 |
| X min | 0.48 |
| X max | 19.95 |
| R² | 0.9711 |
Figure 5.6 - Table of Method Comparison Summary against Predicate Device
5.6.2 Blood Precision
Blood precision studies were performed in field trials at two (2) hospitals on volunteer samples of whole blood by potential end users. One (1) sample was obtained and tested fresh (WB L2). Another sample was obtained and held for several hours to increase lactate concentration (WB L1). This sample was introduced via epoc Care-Fill Capillary Tubes.
Site 1
| User | QC Level | N | Avg | SD | %CV | lot |
|---|---|---|---|---|---|---|
| Phlebotomist 1 | WB L1 | 15 | 10.24 | 0.62 | 6.0% | 09231/09230 |
| Phlebotomist 2 | WB L1 | 15 | 10.27 | 0.34 | 3.3% | 09231/09230 |
Figure 5.7 - Table - Blood Precision Study Summary (Site 1)
Site 2
| ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------ER ESSEE INVESTIENT FORMERCar | 4-4-4-4-4-4-4-4-4-4-4-4-4-4-4-4-4-4-4-4-4-4-4-4-4-4-4-4-4-4-4-4-4-4-4-4-4-4-4-4-4-4-4-4-4-4-4-4-4-4-4-4-4-4-4-4-4-4-4-4-4-4-4-4-4-4-4-4-4-4-4-4-4-4-4-4-4-4-4-4-4-4-4-4-4-4-4-A MINILES EN LANDINGER MIRANIL SEMENTE MILLING MALE(A MIN R LEN & COMMENT IN OR LE | ----------- | |||
|---|---|---|---|---|---|
| ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------ | ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------ | 1 | 11. And A CHINESS - A Charles An An- | 1 | പ്രാമം. |
| Station of Arms Andrew Associated on Area aAnd All Andיחמם/S | A-4-60-100 | ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------ | 10727CAR AN ANNUAL CONNECT CHARACT CHANAL |
Figure 5.8 - Table - Blood Precision Study Summary (Site 2) - Sample Introduced with Capillary Tubes
{7}------------------------------------------------
5.6.3 Aqueous precision
Aqueous precision studies were performed in field trials by potential end users at two (2) hospitals on commercially available blood gas, electrolytes and metabolites control fluids, L1, L2 and L3 (Eurotrol, The Netherlands).
Site 1
| User | QC Level | N | Avg | SD | %CV | lot |
|---|---|---|---|---|---|---|
| RN 1 | L3 | 15 | 0.95 | 0.031 | 3.3% | 09229 |
| Anesthesia Tech | L3 | 15 | 0.94 | 0.027 | 2.9% | 09229 |
| RN 2 | L2 | 14 | 2.88 | 0.05 | 1.8% | 09229 |
| Resp Therapist | L2 | 15 | 2.91 | 0.08 | 2.8% | 09229 |
Figure 5.9 - Table - Aqueous Precision Study Summary (Site 1)
Site 2
| ------------------------------------------------------------------------------------------------------------iser | A STATE CONSULTION CONSULTION OF CONSULTION OF CONSULTION OF CONSULTION OF CONSULTION OF CONSULTION OF CONSULTION OF CONSULTION OF CONSULTION OF CONSULTION OF CONSULTION OF C | - Partificant Mitter ------ | -------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------- | 01 PM610 %ANNUAL AND AN AND CONSULER OF CONSULER OF CONSULERS OF CONSULERS OF CONSULERS OF CONSULERS OF CONSULERS OF CONSULERS OF CONSULERS OF CONSULERS OF CONSULERS OF CONSULERS OF CO | -------------------------------------- | |
|---|---|---|---|---|---|---|
| AND A BOOM A CALL ANDC | ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------ | 1 | Property Autor | C1------------------------------------------------------------------------------------------------------------------------------------------------------------------------------ | വാറ്റംat and the lastA | |
| A L & MINNE A----------------------1ﺮ | . | FRANCHILLAND AND AND AND1. | 500I1 1 100 | uanesسال الساب |
Figure 5.10 - Table - Aqueous Precision Study Summary (Site 2)
Matrix Effects 5.6.4
The method comparison studies were performed in field trials at several hospitals on patient samples of whole blood at various locations. Patient specimens were venous, arterial and capillary. The method comparison was against the predicate device.
| epoc Lactate vs. i-STAT | ||||
|---|---|---|---|---|
| venous | arterial | capillary | all | |
| N | 126 | 73 | 174 | 373 |
| Sxx | 0.113 | 0.116 | 0.290 | 0.215 |
| Syy | 0.586 | 0.455 | 0.517 | 0.530 |
| intercept | 0.211 | -0.165 | 0.257 | 0.132 |
| slope | 0.937 | 1.032 | 0.955 | 0.967 |
| Syx | 0.750 | 0.831 | 1.062 | 0.948 |
| X min | 0.66 | 0.57 | 0.48 | 0.48 |
| X max | 19.88 | 19.95 | 19.57 | 19.95 |
| R2 | 0.9769 | 0.9829 | 0.9653 | 0.9711 |
Figure 5.11 - Table of Method Comparison Summary Against Predicate Device By Sample Matrix Type
{8}------------------------------------------------
| Lactate, mM | |||
|---|---|---|---|
| matrix | Decision level | 2.2 | 5.0 |
| venous | Average Bias | 0.073 | -0.103 |
| 95% Confidence Interval ± | 0.165 | 0.113 | |
| arterial | Average Bias | -0.094 | -0.004 |
| 95% Confidence Interval ± | 0.223 | 0.162 | |
| capillary | Average Bias | 0.158 | 0.031 |
| 95% Confidence Interval ± | 0.198 | 0.142 | |
| all | Average Bias | 0.061 | -0.031 |
| 95% Confidence Interval ± | 0.119 | 0.084 |
Figure 5.12 - Table of Method Comparison Summary Against Predicate Device -Consolidated Bias by Sample Matrix Type
5.6.4.1 Effect of Anticoagulant
The effect of anticoagulant was evaluated on patient samples that were collected using heparinized and non-heparinized collection devices. This study was performed at various POC sites of a hospital (43 samples) and supplemented with in-house studies (17 samples). The data was analyzed using EP9-2A methodology.
| epoc Lactate | |
|---|---|
| No heparin vs. Heparinized | |
| N | 60 |
| Sxx | 0.091 |
| Syy | 0.160 |
| intercept | -0.045 |
| slope | 1.036 |
| Syx | 0.232 |
| X min | 0.52 |
| X max | 11.21 |
| R2 | 0.9916 |
Figure 5.13 - Table of Heparinized Versus Non-Heparinized Samples
Summary of Conclusions Drawn from Non Clinical and 5.7 Clinical Tests
We conclude from the data presented in section 5.5 that the device performs effectively. We conclude from the data section 5.6 that the clinical performance of the device is equivalent to the predicate device: i-Stat Model 300 Portable Clinical Analyzer.
{9}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Image /page/9/Picture/2 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized eagle with three curved lines representing its wings and tail. The eagle is positioned to the right of a circular seal that contains the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter.
Epocal, Inc. c/o Mr. Roy Layer Director of Quality Assurance and Regulatory Affairs 2060 Walkley Road Ottawa, Ontario Canada K1G-3P5
Food & Drug Administration 10903 New Hampshire Avenue Building 66 Silver Spring, MD 20993
JUN 0 9 2000
Re: K093297
Trade Name: epoc Lactate test Regulation Number: 21 CFR §862.1450 Regulation Name: Lactic acid test system. Regulatory Class: Class I, meets limitations of exemptions, 21 CFR §862.9 (c)(9) Product Codes: KHP Dated: May 13, 2010 Received: May 17, 2010
Dear Mr. Layer:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); and good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820).
{10}------------------------------------------------
Page 2
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at (301) 796-5760. For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or ( 301 ) 796-5680 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
CA
Courtney C. Harper, Ph.D. Director Division of Chemistry and Toxicology Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health
Enclosure
{11}------------------------------------------------
Indication for Use
k = 93297 510(k) Number (if known):
Device Name: epoc Lactate test
Indication For Use:
The Lactate test, as part of the epoc Blood Analysis System, is intended for use by trained medical professionals as an in vitro diagnostic device for the quantitative testing of samples of heparinized or un-anticoagulated arterial, venous or capillary whole blood in the laboratory or at the point of care in hospitals, nursing homes or other clinical care institutions.
Lactate measurements from the epoc Blood Analysis System are used to evaluate the acid-base status and are used in the diagnosis and treatment of lactic acidosis (abnormally high acidity of the blood).
Prescription Use __ X (21 CFR Part 801 Subpart D)
And/Or
Over the Counter Use _ (21 CFR Part 801 Subpart C) .
(PLEASE DO NOT WRITE BELOW THIS LINE; CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostic Device Evaluation and Safety (OIVD)
Carol C. Benson
Division Sign-Off Office of In Vitro Diagnostic Device Evaluation and Safety
510(k) k 0 93217
Page 1 of 1
§ 862.1450 Lactic acid test system.
(a)
Identification. A lactic acid test system is a device intended to measure lactic acid in whole blood and plasma. Lactic acid measurements that evaluate the acid-base status are used in the diagnosis and treatment of lactic acidosis (abnormally high acidity of the blood).(b)
Classification. Class I (general controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 862.9.