K Number
K113726
Device Name
EPOC CHLORIDE TEST AND EPOC CREATININE TEST
Manufacturer
Date Cleared
2012-10-05

(291 days)

Product Code
Regulation Number
862.1225
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP Authorized
Intended Use
The Chloride test, as part of the epoc Blood Analysis System, is intended for use by trained medical professionals as an in vitro diagnostic device for the quantitative testing of samples of heparinized or un-anticoagulated arterial, venous or capillary whole blood in the laboratory or at the point of care. Chloride measurements from the epoc Blood Analysis System are used in the diagnosis and treatment of electrolyte and metabolic disorders. The Creatinine test, as part of the epoc Blood Analysis System, is intended for use by trained medical professionals as an in vitro diagnostic device for the quantitative testing of samples of heparinized or un-anticoagulated arterial, venous or capillary whole blood in the laboratory or at the point of care. Creatinine measurements from the epoc Blood Analysis System are used in the diagnosis and treatment of certain renal diseases and in monitoring renal dialysis.
Device Description
The epoc Blood Analysis System is an in vitro analytical system comprising a network of one or more epoc Readers designed to be used at the point of care (POC). The readers accept an epoc single use test card containing a group of sensors that perform diagnostic testing on whole blood. The blood test results are transmitted wirelessly to an epoc Host, which displays and stores the test results. The epoc System is intended for use by trained medical professionals as an in vitro diagnostic device for the quantitative testing of samples of whole blood. The test card panel configuration currently includes sensors for Sodium Na, Potassium K, Ionized Calcium iCa, pH, pCO2, pO2, Lactate, Glucose and Hematocrit Hct. This submission adds Chloride and Creatinine to this list of approved tests.
More Information

No
The document describes a standard in vitro diagnostic system for blood analysis using sensors and does not mention any AI or ML components in its description, intended use, or performance studies.

No
The device is described as an "in vitro diagnostic device" used for quantitative testing of blood samples to provide measurements that are "used in the diagnosis and treatment of electrolyte and metabolic disorders" or "certain renal diseases and in monitoring renal dialysis." It does not directly provide therapy.

Yes

Explanation: The "Intended Use / Indications for Use" section explicitly states that the epoc Blood Analysis System and its Chloride and Creatinine tests are "intended for use by trained medical professionals as an in vitro diagnostic device." The "Device Description" also reiterates that "The epoc System is intended for use by trained medical professionals as an in vitro diagnostic device".

No

The device description clearly states it is an "in vitro analytical system comprising a network of one or more epoc Readers" and "accepts an epoc single use test card containing a group of sensors". This indicates the system includes significant hardware components (readers, test cards with sensors) in addition to any software.

Yes, this device is an IVD (In Vitro Diagnostic).

Here's why:

  • Explicit Statement in Intended Use/Indications for Use: The document repeatedly states that the epoc Blood Analysis System and its tests (Chloride and Creatinine) are "intended for use by trained medical professionals as an in vitro diagnostic device".
  • Explicit Statement in Device Description: The Device Description also explicitly states that the epoc Blood Analysis System "is an in vitro analytical system" and "is intended for use by trained medical professionals as an in vitro diagnostic device".
  • Testing of Biological Samples: The system is designed to test "samples of heparinized or un-anticoagulated arterial, venous or capillary whole blood", which are biological samples.
  • Purpose of Testing: The measurements are used "in the diagnosis and treatment of electrolyte and metabolic disorders" (Chloride) and "in the diagnosis and treatment of certain renal diseases and in monitoring renal dialysis" (Creatinine). This aligns with the definition of an IVD, which is used to examine specimens derived from the human body to provide information for diagnostic, monitoring, or compatibility purposes.
  • Performance Studies: The document details various performance studies (precision, linearity, method comparison, etc.) conducted to demonstrate the analytical and clinical performance of the device, which is a requirement for IVD devices.

The presence of the phrase "in vitro diagnostic device" in multiple sections, coupled with the description of its function and intended use, clearly indicates that this device falls under the category of an IVD.

N/A

Intended Use / Indications for Use

The Chloride test, as part of the epoc Blood Analysis System, is intended for use by trained medical professionals as an in vitro diagnostic device for the quantitative testing of samples of heparinized or un-anticoagulated arterial, venous or capillary whole blood in the laboratory or at the point of care. Chloride measurements from the epoc Blood Analysis System are used in the diagnosis and treatment of electrolyte and metabolic disorders.

The Creatinine test, as part of the epoc Blood Analysis System, is intended for use by trained medical professionals as an in vitro diagnostic device for the quantitative testing of samples of heparinized or un-anticoagulated arterial, venous or capillary whole blood in the laboratory or at the point of care. Creatinine measurements from the epoc Blood Analysis System are used in the diagnosis and treatment of certain renal diseases and in monitoring renal dialysis.

Product codes (comma separated list FDA assigned to the subject device)

CGL, CGZ

Device Description

The epoc Chloride and Creatinine tests are being added as additional sensors to the existing single use test card that is used with the epoc Blood Analysis System. This test card is inserted into the epoc Reader and all analytical steps are performed automatically. Patient and user information may be entered into the mobile computing device (epoc Host) during the automated analysis cycle.

The epoc Blood Analysis System is an in vitro analytical system comprising a network of one or more epoc Readers designed to be used at the point of care (POC). The readers accept an epoc single use test card containing a group of sensors that perform diagnostic testing on whole blood. The blood test results are transmitted wirelessly to an epoc Host, which displays and stores the test results.

The epoc System is intended for use by trained medical professionals as an in vitro diagnostic device for the quantitative testing of samples of whole blood.

The test card panel configuration currently includes sensors for Sodium Na, Potassium K, Ionized Calcium iCa, pH, pCO2, pO2, Lactate, Glucose and Hematocrit Hct. This submission adds Chloride and Creatinine to this list of approved tests.

To perform a blood test, a new test card is inserted into a card reader's card slot with white label face down. When fully inserted, the test card is automatically engaged in the reader.

The card insertion process:

  • Brings the cards sensor module into contact with the reader's electrical contact array:
  • Brings the card's measurement region, which is the fluidic channel above the sensor array, into thermal contact with the reader's heater assembly for heating the measurement region to 37°C;
  • Actuates the opening of the fluidic valve in the card and causes delivery of calibrator fluid from the reservoir to the measurement region.

After calibration, and upon a prompt by the reader (LED visual and audio beep), the user introduces a blood sample for measurement through the blood sample port to the card's measurement region. When sensors are contacted by the blood sample they generate electrical signals proportional to analyte concentrations in the blood sample, which are transmitted wirelessly by the Reader to the epoc Host. The epoc Host displays and stores the blood test results.

The epoc Host will also display calculated values based on the new analytes:
Anion Gap: AGap = (Na+) - (CI- + cHCO3-)
Anion Gap, K: AGapK = (Na+ + K+) - (CI- + cHCO3-)
Estimated Glomerular Filtration Rate (IDMS-traceable MDRD type): eGFR = 175 x (Crea-1.154) x (Age-0.203) x (0.742 if female, 1 if male)
Estimated Glomerular Filtration Rate, if African American (IDMStraceable MDRD type):
eGFR-a = 175 x (Crea-1.154) x (Age-0.203) x (0.742 if female, 1 if male) x 1.212
Crea concentration is in units of mg/dL. Age and gender ("M" or "F") are user inputs. eGFR, eGFR-a are both reported on the epoc System.

Mentions image processing

Not Found

Mentions AI, DNN, or ML

Not Found

Input Imaging Modality

Not Found

Anatomical Site

Not Found

Indicated Patient Age Range

eGFR, eGFR-a values are not reported if age is less that 18 years old or greater than 70 years old.

Intended User / Care Setting

trained medical professionals / laboratory or at the point of care (POC)

Description of the training set, sample size, data source, and annotation protocol

Not Found

Description of the test set, sample size, data source, and annotation protocol

Not Found

Summary of Performance Studies (study type, sample size, AUC, MRMC, standalone performance, key results)

Summary of Non-Clinical Test Performance in Support of Substantial Equivalence:

  • Aqueous precision: A twenty-day precision study was performed in-house on 3 lots using aqueous controls at two levels (L1 and L3) for blood gases, electrolytes, and metabolites.
    • Chloride: N=240, Mean L1=76.9, Mean L3=125.0, WD %CV L1=0.5%, WD %CV L3=0.5%, Total %CV L1=0.5%, Total %CV L3=0.7%
    • Creatinine: N=239 (L1), N=241 (L3), Mean L1=0.71, Mean L3=5.50, WD %CV L1=4.2%, WD %CV L3=3.6%, Total %CV L1=4.9%, Total %CV L3=4.1%
  • Linearity/Reportable Range: Performed in-house using blood samples as per CLSI EP6-A recommendations. Nine blood samples were prepared from two pools of blood, evaluated against an in-house standard method traceable to NIST standards.
    • Creatinine (0.25 - 15.5 mg/dL): Slope=1.00, Intercept=0.07, R squared=0.99.
    • This data supports reportable ranges of 65-140 mM for Chloride and 0.3 - 15.00 mg/dL for Creatinine.
  • Traceability:
    • Chloride ion concentration values are traceable to NIST standards.
    • Creatinine concentration values are traceable to NIST standard SRM 967. The test is calibrated to an IDMS-traceable whole blood method and reports plasma equivalent concentrations.
  • Detection Limit: Performed in-house as per CLSI EP6-A. The low end of the reportable range for epoc chloride (65 mM) and creatinine (0.30 mg/dL) are greater than or equal to the limit of detection and statistically discernable from the limit of blank.
  • Analytical Specificity: Interference testing performed in-house as per CLSI document EP7-A2. Defined clinically insignificant bias for creatinine as ≤ 0.2 mg/dL for concentrations ≤2 mg/dL and ≤7.9% for concentrations >2 mg/dL; for chloride as ≤ 4.2% for concentrations ≤125 mM and ≤5.2% for concentrations >125 mM. Lists clinically significant and insignificant interfering substances.

Summary of Clinical Tests Submitted in Support of Substantial Equivalence:

  • Method comparison with Predicate and Comparative Devices: Side-by-side comparisons of the epoc System with predicate and other comparative devices in clinical field trials. Venous, arterial, capillary samples were tested by various end-users. Multiple card lots and epoc system readers/host computers were employed.
    • Chloride method comparison: Performed at two hospitals. Venous samples (N=96) compared with 2 non-point-of-care systems (Roche Cobas 6000, Siemens Advia). Venous, arterial, and capillary patient samples (N=155) compared with Abbott i-STAT 300.
      • vs. non-POC Systems: N=96, slope=0.90, intercept=9.38, R2=0.96, Mean Bias at 112 mM=-1.3.
      • vs. Abbott i-STAT: N=155, slope=0.99, intercept=0.20, R2=0.98, Mean Bias at 112 mM=-1.0.
    • Creatinine method comparison: Performed at a hospital site. Venous, arterial, and capillary patient samples (N=144) compared with a serum-based laboratory method (Roche Cobas 6000).
      • vs. Roche Cobas 6000: N=144, slope=1.03, intercept=-0.10, R2=0.99, Mean Bias at 1.25 mg/dL=-0.06.
  • Blood Precision: Whole blood precision studies performed at clinical sites to demonstrate precision when analyzed by end-users. Normal and spiked level samples were tested using both syringes and capillary tubes.
    • Chloride (All Users - All Sites):
      • Syringe (Normal): N=120, Mean=106.8, Avd SD W-R=0.63, CV% W-R=0.6%.
      • Syringe (Spiked): N=119, Mean=120.9, Avd SD W-R=0.86, CV% W-R=0.7%.
      • Capillary (Normal): N=40, Mean=106.4, Avd SD W-R=0.70, CV% W-R=0.7%.
      • Capillary (Spiked): N=40, Mean=123.4, Avd SD W-R=1.11, CV% W-R=0.9%.
    • Creatinine (All Users - All Sites):
      • Syringe (Normal): N=118, Mean=0.6, Avd SD W-R=0.05, CV% W-R=7.6%.
      • Syringe (Spiked): N=118, Mean=1.57, Avd SD W-R=0.06, CV% W-R=3.9%.
      • Capillary (Normal): N=29, Mean=0.52, Avd SD W-R=0.04, CV% W-R=6.8%.
      • Capillary (Spiked): N=30, Mean=1.45, Avd SD W-R=0.06, CV% W-R=3.9%.
  • Aqueous Precision: Aqueous precision studies performed at clinical sites using commercially available blood gas, electrolytes, and metabolites control fluids (Eurotrol, L1, L2, L3).
    • Chloride (Control Fluids):
      • QC Level 1: N=165, Mean=76.5, SD=0.50, CV% Total=0.7%, CV% R-R=0.3%, CV% W-R=0.6%.
      • QC Level 2: N=163, Mean=98.6, SD=0.56, CV% Total=0.6%, CV% R-R=0.3%, CV% W-R=0.4%.
      • QC Level 3: N=148, Mean=123.7, SD=1.06, CV% Total=0.9%, CV% R-R=0.6%, CV% W-R=0.6%.
    • Creatinine (Control Fluids):
      • QC Level 1: N=120, Mean=0.66, SD=0.05, CV% Total=6.8%, CV% R-R=3.4%, CV% W-R=6.1%.
      • QC Level 2: N=119, Mean=2.04, SD=0.13, CV% Total=6.4%, CV% R-R=5.5%, CV% W-R=3.8%.
      • QC Level 3: N=120, Mean=4.31, SD=0.27, CV% Total=6.3%, CV% R-R=4.0%, CV% W-R=4.8%.
  • Matrix Effects: Method comparison studies in field trials at two hospitals on patient samples of whole blood at various locations (venous, arterial, capillary). Comparison against the predicate device.
    • Chloride epoc vs. i-STAT by Matrix Type: Venous (N=49, R²=0.97), Arterial (N=43, R²=0.99), Capillary (N=63, R²=0.99), All (N=155, R²=0.98). Consolidated Bias by Sample Matrix Type provided at decision levels.
    • Creatinine epoc vs. Roche Cobas by Matrix Type: Venous (N=53, R²=0.99), Arterial (N=42, R²=0.99), Capillary (N=49, R²=0.99), All (N=144, R²=0.99). Consolidated Bias by Sample Matrix Type provided at decision levels.
  • Effect of Anticoagulant: Evaluated on patient samples collected using heparinized and non-heparinized devices (46 hospital samples, 29 in-house samples). Analyzed using EP9-2A methodology.
    • Chloride epoc Hep vs. No-Hep: N=76, slope=0.98, intercept=1.92, R2=0.97.
    • Creatinine epoc Hep vs. No-Hep: N=77, slope=0.99, intercept=0.02, R2=0.99.

Key Metrics (Sensitivity, Specificity, PPV, NPV, etc.)

Not Found

Predicate Device(s): If the device was cleared using the 510(k) pathway, identify the Predicate Device(s) K/DEN number used to claim substantial equivalence and list them here in a comma separated list exactly as they appear in the text. List the primary predicate first in the list.

K001387, K024098

Reference Device(s): Identify the Reference Device(s) K/DEN number and list them here in a comma separated list exactly as they appear in the text.

K061597, K090109, K093297

Predetermined Change Control Plan (PCCP) - All Relevant Information for the subject device only (e.g. presence / absence, what scope was granted / cleared under the PCCP, any restrictions, etc).

Not Found

§ 862.1225 Creatinine test system.

(a)
Identification. A creatinine test system is a device intended to measure creatinine levels in plasma and urine. Creatinine measurements are used in the diagnosis and treatment of renal diseases, in monitoring renal dialysis, and as a calculation basis for measuring other urine analytes.(b)
Classification. Class II.

0

еросаl

2060 Walkley Road Ottawa Ontario, Canada K1G 3P5

OCT 5 2012

1

510(k) SUMMARY

This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.

The assigned 510(k) number is: K113726

Summary Prepared: July 28, 2012 Submitted by: Epocal Inc. 2060 Walkley Road, Ottawa, Ontario, Canada K1G 3P5 Telephone: (613) 738-6192 Fax: (613) 738-6195 Contact: Roy Layer Vice President, Quality Assurance & Regulatory Affairs. 5.1 Identification of the Device(s) Common Name: Creatinine Test epoc™ Creatinine Test Trade Name: Classification Name: electrode, ion based, enzymatic, creatinine Device Classification: 2 Requlation Number: 862.1225 Clinical Chemistry Panel: Product Code: CGL Common Name: Chloride Test epoc™ Chloride Test Trade Name: Classification Name: electrode, ion-specific, chloride Device Classification: 2 Regulation Number: 862.1170 Panel: Clinical Chemistry Product Code: CGZ

5.2 Identification of Predicate Devices

  • i-Stat™ Chloride using i-Stat™ Model 300 Portable Clinical Analyzer .
  • . Roche Cobas c 511/512 CREP2 Creatinine Plus ver. 2 assay

5.3 Description of the New Device

The epoc Chloride and Creatinine tests are being added as additional sensors to the existing single use test card that is used with the epoc Blood Analysis System. This test card is inserted into the epoc Reader and all analytical steps are performed automatically. Patient and user information may be entered into the mobile computing device (epoc Host) during the automated analysis cycle.

epoc Creatinine & Chloride Tests - 510(k) Submission Section 05 - Page 1 of 16

1

The epoc Blood Analysis System is an in vitro analytical system comprising a network of one or more epoc Readers designed to be used at the point of care (POC). The readers accept an epoc single use test card containing a group of sensors that perform diagnostic testing on whole blood. The blood test results are transmitted wirelessly to an epoc Host, which displays and stores the test results.

The epoc System is intended for use by trained medical professionals as an in vitro diagnostic device for the quantitative testing of samples of whole blood.

The test card panel configuration currently includes sensors for Sodium Na, Potassium K, Ionized Calcium iCa, pH, pCO2, pO2, Lactate, Glucose and Hematocrit Hct. This submission adds Chloride and Creatinine to this list of approved tests.

To perform a blood test, a new test card is inserted into a card reader's card slot with white label face down. When fully inserted, the test card is automatically engaged in the reader.

The card insertion process:

  • Brings the cards sensor module into contact with the reader's electrical contact . array:
  • . Brings the card's measurement region, which is the fluidic channel above the sensor array, into thermal contact with the reader's heater assembly for heating the measurement region to 37°C;
  • . Actuates the opening of the fluidic valve in the card and causes delivery of calibrator fluid from the reservoir to the measurement region.

After calibration, and upon a prompt by the reader (LED visual and audio beep), the user introduces a blood sample for measurement through the blood sample port to the card's measurement region. When sensors are contacted by the blood sample they generate electrical signals proportional to analyte concentrations in the blood sample, which are transmitted wirelessly by the Reader to the epoc Host. The epoc Host displays and stores the blood test results.

The epoc Host will also display calculated values based on the new analytes:

Anion Gap: AGap = (Na+) - (CI- + cHCO3-)

Anion Gap, K: AGapK = (Na+ + K+) - (CI- + cHCO3-)

Estimated Glomerular Filtration Rate (IDMS-traceable MDRD type): eGFR = 175 x (Crea-1.154) x (Age-0.203) x (0.742 if female, 1 if male)

Estimated Glomerular Filtration Rate, if African American (IDMStraceable MDRD type):

eGFR-a = 175 x (Crea-1.154) x (Age-0.203) x (0.742 if female, 1 if male) x 1.212

Crea concentration is in units of mg/dL. Age and gender ("M" or "F") are user inputs. eGFR, eGFR-a are both reported on the epoc System. eGFR, eGFR-a values are not reported if age is less that 18 years old or greater than 70 years old.

Numeric values will be reported for values between 2-60 mL/min/1.73 m². Values >60 will be reported as > 60 mL/min/1.73 m². This range is based on the specific National Kidney Disease

2

Education Program (NKDEP) recommendation for reporting eGFR the following web link: values. Please refer to http://nkdep.nih.qov/lab-evaluation/gfr/reporting.shtml.

The following additional statement appears in our labeling: eGFR > 60 does not exclude the possibility of mild renal disease. Further laboratory testing may be necessary to distinguish normal renal function from mild renal disease.

Addition of the epoc Chloride and Creatinine tests comprises three (3) changes to the epoc System:

    1. Addition of the new sensors on the test card (described below);
    1. Modification of the epoc System software application to accommodate the new tests (refer to Section 16 - Software);
    1. Labeling changes including indications for use (refer to Section 04 Indications for Use Statements and Section 13 - Proposed Labeling and Packaging),

5.4 Comparison of Characteristics To Predicate Devices

Tables Figures 5.1 and 5.2 itemize key characteristics of the epoc chloride and creatinine tests as part of the epoc device and their respective predicate devices. The chloride and creatinine tests are an addition to the previously cleared system. All other system features are the same as previously submitted on K061597, K090109 and K093297.

epoc Blood Analysis Systemi-STAT Model 300Same / Different
510(k) #K113726K001387
ItemDevicePredicate
Intended useThe Chloride test, as part of the epoc
Blood Analysis System, is intended for
use by trained medical professionals as
an in vitro diagnostic device for the
quantitative testing of samples of
heparinized or un-anticoagulated
arterial, venous or capillary whole blood
in the laboratory or at the point of care.
Chloride measurements from the epoc
Blood Analysis System are used in the
diagnosis and treatment of electrolyte
and metabolic disorders.The test for chloride, as part of the i-STAT
System, is intended for use in the in vitro
quantification of chloride in arterial,
venous, or capillary whole blood.
Chloride measurements are primarily
used in the diagnosis, monitoring, and
treatment of electrolyte and metabolic
disorders including, but not limited to,
cystic fibrosis, diabetic acidosis, and
hydration disorders.similar
Where usedhospital, point of carehospital, point of caresame
Sample typeVenous, arterial and capillary whole
bloodVenous, arterial and capillary whole bloodsame
Reportable
range65 - 140 mmol/L65 - 140 mmol/Lsame
Detection
principleThe epoc Chloride test relies on ion
selective membrane potentiometry.The i-STAT Chloride test relies on ion
selective membrane potentiometry.same
Sample
volumeAt least 92 uL100μLsame

Figure 5.1 - Table Comparing epoc Chloride Test Characteristics with Predicate Device

3

epoc Blood Analysis SystemRoche Cobas
510(k) #K113726K024098
ItemDevicePredicate
Intended useThe Creatinine test, as part of the epoc
Blood Analysis System, is intended for
use by trained medical professionals as
an in vitro diagnostic device for the
quantitative testing of samples of
heparinized or un-anticoagulated
arterial, venous or capillary whole blood
in the laboratory or at the point of care .
Creatinine measurements from the epoc
Blood Analysis System are used in the
diagnosis and treatment of certain renal
diseases and in monitoring renal
dialysis.In vitro assay for the quantitative
determination of creatinine in human
serum, plasma and urine on Roche
automated clinical chemistry analyzers.
Where usedhospital, point of carehospital, laboratory
Sample typeVenous, arterial and capillary whole
bloodSerum, Plasma, Urine
Reportable
range0.3 - 15.0 mg/dL0.03 - 30 mg/dL
Detection
principleThe epoc Creatinine test relies on
enzymatic cascade reaction
(creatininase, creatinase, sarcosine
oxidase) leading to amperometric
peroxide detection.The Roche Creatinine test relies on an
enzymatic cascade reaction (creatininase,
creatinase, sarcosine oxidase) leading to
peroxidase-catalyzed chromogenic
peroxide detection.
Sample
volumeAt least 92 uL2-5 uL

Figure 5.2 - Table Comparing epoc Creatinine Test Characteristics with Predicate Device

Summary of Non-Clinical Test Performance in Support of 5.5 Substantial Equivalence

5.5.1 Aqueous precision

Experiments were performed in-house to demonstrate the precision of the epoc test methods. The table below shows the results of a twenty day precision study using performed on 3 lots using aqueous controls at two levels L1 and L3 for the blood gases, electrolytes and metabolytes.

All lotsAll lots
Chloride [mM]L1L3Creatinine [mg/dL]L1L3
N240240N239241
Mean76.9125.0Mean0.715.50
SWD0.350.61SWD0.0300.197
SDD0.180.61SDD0.0170.112
ST0.390.86ST0.0350.226
WD %CV0.5%0.5%WD %CV4.2%3.6%
Total %CV0.5%0.7%Total %CV4.9%4.1%

Figure 5.3 - Table - 20 Day Precision Study Data

4

Linearity/Reportable Range 5.5.2

This study was performed in-house using blood samples as per CLSI EP6-A recommendations for evaluation of linearity. A total of nine blood samples were prepared starting with two pools of blood, which were evaluated versus an in-house standard method with traceability to NIST standards. Regression analysis was performed as per CLSI EP6-A.

| r
AND ARRANTHISAN IN CHARACT OF CHARACTER OF CHARACT TO COLLECTION

And In a first of the county of the county
Range
  1. 24 | 1006 | merc | | |
    | | ﺎ ﺍﻟﻤﻠﻜﺔ | | 000 | |

| Creatinine

------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
Test RangeSlopeInterceptn2
0.25 - 15.5 mg/dL1.000.070.99

Figure 5.3 - Table - In House Whole Blood Linearity

This data supports the claimed reportable ranges of 65-140 mM for Chloride and 0.3 - 15.00 mg/dL for Creatinine.

5.5.3 Traceability

Chloride ion concentration values assigned to controls and calibrator fluids are traceable to NIST standards.

Creatinine concentration values assigned to controls and calibrator fluids are traceable to NIST standard SRM 967. The epoc Creatinine test is calibrated to an IDMS-traceable whole blood method and reports plasma equivalent concentrations.

5.5.4 Detection Limit

This study was performed in-house as per CLSI EP6-A recommendations for evaluation limits of detection and quantification. The low end of the reportable range for the epoc chloride test (65 mM) and the epoc creatinine test (0.30 mg/dL) are greater than or equal to the limit of detection and are statistically discernable from the limit of blank.

5.5.5 Analytical Specificity

Interference testing based on CLSI "Interference Testing in Clinical Chemistry; Approved Guideline", CLSI document EP7-A2. was performed in-house on the epoc chloride and creatinine sensors. In each of these tests a pooled human serum was aliquoted into two (2) samples. The test sample was spiked by addition of interferent, while the control sample was spiked by the addition of the solvent of the interferent. The bias between the mean of six replicates on both the control sample and the test sample with added interferent was calculated. The concentration of interfering substance considered as causing no clinically significant interference is defined as a bias (difference between the test and the control sample) of:

≤ 0.2 mg/dL for creatinine concentrations ≤2 mg/dL and ≤7.9% for creatinine concentrations >2 mg/dL :

5

≤ 4.2% for chloride concentrations ≤125 mM and ≤5.2% for chloride concentrations >125 mM.

Clinically significant interfering substances for chloride are itemized below:

  • ß-Hydroxybutyrate will have no significant effect up to 6.46 mM (67.2 . mg/dL) after which it will increase the chloride reading by up to 0.63 mM/mM ß-Hydroxybutyrate.
  • . Bromide will have no significant effect up to 3.43 mM after which it will increase the chloride reading up to 9.36 mM/mM Bromide.
  • Citrate will have no significant effect up to 2.36 (45.3 mg/dL) mM after . which it will increase the chloride reading by up to 1.37 mM /mM Citrate.
  • N-Acetylcysteine will have no significant effect up to 2.85 mM (46.4 � mg/dL) after which it will decrease the chloride reading by up to 1.34 mM/mM N-Acetvlcvsteine.
  • Salicylic acid will have no significant effect up to 2.55 mM (41 mg/dL) . after which it will increase the chloride reading up to 1.66 mM/mM Salicylic acid.
  • Thiocyanate will have no significant effect up to 2.50 (14.5 mg/dL) mM ● after which it will increase the chloride reading up to 1.66 mM/mM Thiocyanate.

The following levels of exogenous interferences were tested and found to be clinically insignificant:

1.324 mmol/L (20 mg/dL) acetaminophen, 3.62 mmol/L (65.2 mg/dL) acetylsalicy/ic acid, 342 µmol/L (6.8 mg/dL) Na ascorbate, 3.4 µmol/L (0.1 mg/dL) EDTA, 71 umol/L methyIdopa, 2.55 mmol/L (156 mg/dL) oxidized glutathione, 132 µmol/L (1.0mg/dL) hydroxyurea, 292 umol/L (4mg/dL) isoniazid (nydrazid), +0.8% intralipid, 3 µmol/L (0.1 mg/dL) dobutamine, 5.87 µmol/L (0.1 mg/dL) dopamine, 86.8 mmol/L (400 mg/dL) ethanol, 105 µmol/L (0.44 mg/dL) fluoride, 133 µmol/L (0.4 mg/dL) formaldehyde, 55 mmol/L (990 mg/dL) glucose, 0.4 mmol/L (5 mg/dL) guaiacol, 3000 U/L heparin, 2.43 mmol/L (50 mg/dL) ibuprofen, 0.1 mmol/L (2.0 mg/dL) L-Dopa, 51.2 µmol/L (1.2 mg/dL) lidocaine, 71 µmol/L (1.7 mg/dL) 2.37 mmol/L (64 mg/dL) methyldopa, 354 µmol/L (9.4 mg/dL) pentathol, tolbutamide, 2.99 mmol/L (49.6 mq/dL) Iodide.

The following levels of endogenous interferences were tested and found to be clinically insignificant:

+342 umol/L (+20.1mg/dL) bilirubin unconjugated, +342 umol/L (28.8 mg/dL) bilirubin conjugate, +382 µmol/L (5.0 mg/dL) creatine, 102 mmHg CO2, 13 mmHg CO2 + 40 mmol/L bicarbonate , pH >8.0, pH 8.0% protein, 1.4 mmol/L (23.5 mg/dL) uric acid. 6.6 mmol/L (74 mg/dL) lactate, 122 mmHg O2, 28 mmHg O2, 0.25 mmol/L (2.9 mg/dL) proline, 1 µmol/L (0.01 mg/dL) sarcosine, 42.9 mmol/L (258 mg/dL) urea.

Clinically significant interfering substances for creatinine are itemized below:

  • । Creatine will have no significant effect up to 116 µmol/L (1.52 mg/dL) after which it will increase the creatinine concentration by up to 0.0025 mg/dL creatinine per umol/L creatine. The normal range of creatine in plasma is 8 -31 umol/L (0.1- 0.4 mg/dL) in males and 15 - 53 µmol/L (0.2 - 0.7 mg/dL) in females '3.
  • Bilirubin conjugate will have no significant effect up to 104 µmol/L -(8.76 mg/dL) after which it will decrease the creatinine concentration by up to

6

0.002 mg/dL creatinine per umol/L bilirubin conjugate. The normal range of bilirubin conjugate is 0 - 3.4 umol/L (0 - 0.2 mg/dL) 2.

  • Bromide will have no significant effect up to 17.9 mmol/L after which it will decrease the creatinine concentration by up to 0.014 mg/dL creatinine per mmol/L bromide.
  • Thiocyanate will have no significant effect up to 0.93 mmol/L (5.41 mg/dL) after which it will decrease the creatinine concentration by up to 0.142 mg/dL creatinine per mmol/L thiocyanate.
  • Citrate will have no significant effect up to 19.9 mmol/L (382.1 mg/dL) after which it will decrease the creatinine concentration by up to 0.026 mg/dL creatinine per mmol/L citrate.
  • Iodide will have no significant effect up to 0.007 mmol/L iodide (0.089 mg/dL) after which it will decrease the creatinine concentration by up to 28 mq/dL creatinine per mmol/L iodide.
  • N-acetyl cysteine will have no significant effect up to 820 umol/L (13.35 mg/dL) after which it will decrease the creatinine concentration by up to 0.26 mq/dL creatinine per mmol/L N-acetyl cysteine. It has been reported that 1 mM N-acetyl cysteine is therapeutically unattainable in plasma8. The therapeutic level for N-acetyl cysteine is 0.3 mM16.

The following levels of exogenous interferences were tested and found to be clinically insignificant:

1.324 mmol/L (20 mg/dL) acetaminophen, 3.62 mmol/L (65.2 mg/dL) acetylsalicy/ic acid, 5 µmol/L (0.7 mq/dL) bacitractin, 30.2 µmol/L (1 mg/dL) ciprofloxacin, 48.6 µmol/L (1.75 mg/dL) Levofloxaxin, 342 µmol/L (6.8 mg/dL) Na ascorbate, 100 μmol/L {~2mg/dL) L-dopa, 3.4 µmol/L (0.1 mg/dL) EDTA, 105 µmol/L (0.441mg/dL) Na fluoride, 71 µmol/L (1.7 mg/dL) methyldopa, 2.55 mmol/L (156 mg/dL) oxidized glutathione, 2.55 mmol/L (78 mg/dL) reduced glutathione, 920 µmol/L (6.96 mg/dL) hydroxyurea, 292 µmol/L (4mg/dL) isoniazide (nydrazid), +0.8% (800 mg/dL) intralipid, 3 umol/L (0.1 mq/dL) dobutamine, 5.87 umol/L (0.1 mg/dL) dopamine, 86.8 mmol/L (400 mg/dL) ethanol, 133 µmol/L (0.4 mg/dL) formaldehyde, 55 mmol/L (990 mg/dL) glucose, 0.4 mmol/L (5 mg/dL) guaiacol, 3000 U/L heparin, 2.43 mmol/L (50 mg/dL) ibuprofen, 78.1 µmol/L (6.42 mg/dL) rifampicin, 51.2 umol/L (1.2 mq/dL) lidocaine, 354 µmol/L (9.4 mg/dL) pentathol, 4.34 mmol/L (70 mg/dL) salicylate, 2.37 mmol/L (64 mg/dL) tolbutamide.

The following levels of endogenous interferences were tested and found to be clinically insignificant:

+342 µmol/L (+20.1mg/dL) bilirubin unconjugated, 100.26 mmHg CO2, 15.5 mmHg C02, + 48.2 mmol/L Bicarbonate , pH >8.0, pH 8.0% Protein, 1.4 mmol/L (23.5 mg/dL) Uric Acid. 6.6 mmol/L (74 mg/dL) lactate, 132 mmHg 02, 22 mmHg 02, 0.25 mmol/L (2.9 mg/dL) proline, 1 µmol/L (0.01 mg/dL) sarcosine, 10.0 mmol/L (104 mg/dL)

ß-hydroxybutyrate, 42.9 mmol/L (258 mg/dL) urea.

7

Summary of Clinical Tests Submitted in Support of 5.6 Substantial Equivalence

5.6.1Method comparison with Predicate and Comparative Devices

We have performed side-by-side comparisons of the epoc System with the predicate devices and other comparative devices in clinical field trials. Venous, arterial, capillary samples were tested by a variety of potential end users. Multiple card lots and multiple epoc system readers and host computers were employed.

Method comparison studies for chloride were performed at two hospitals. Venous samples were compared with 2 non-point-of-care systems (2 serum methods). Venous, arterial and capillary patient samples were compared with a whole blood point-of-care system. Consolidated method comparison data versus the predicate device and comparative instruments are summarized in the tables 5.5 below.

non-POC Systems*Abbott i-STAT†
N96155
Sxx0.60.9
Syy0.60.8
slope0.900.99
intercept9.380.20
Syx2.11.9
X min7169
X max142139
R20.960.98
Mean Bias at 112 mM-1.3-1.0

Figure 5.5 – Table – Chloride Method Comparison Data

  • Pooled venous sample data. Approximate equal number vs. Roche Cobas 6000, Siemens Advia + Patient samples approximately equal numbers of venous, arterial and capillary samples versus Abbott i-STAT 300

Method comparison studies were performed at a hospital site comparing venous, arterial and capillary patient samples with a serum-based laboratory method. Consolidated method comparison data versus the predicate device are summarized in the tables 5.6 below.

Roche Cobas 6000*
N144
Sxx0.10
Syy0.30
slope1.03
intercept-0.10
Syx0.45
X min0.30
X max14.80
R20.99
Mean Bias at 1.25 mg/dL-0.06

Figure 5.6 – Table – Creatinine Method Comparison Data

  • Patient samples approximately equal numbers of venous, arterial and capillary samples

8

5.6.2 Blood Precision

Whole blood precision studies were performed at clinical sites to demonstrate precision when analyzed by end users. Normal and spiked level samples were tested injected by both syringes and capillary tubes.

UserQC LevelNMeanSD%CVLot
Special Care Nursery RN 1Normal syringe10102.111.011.0%08-11250-00
Special Care Nursery RN 2Normal syringe10102.580.530.5%08-11244-00
Labor & Delivery RN 1Normal syringe9108.801.251.2%08-11243-00
Labor & Delivery RN 2Normal syringe10109.100.990.9%08-11244-00
Special Care Nursery RN 1Spiked syringe10111.970.510.5%08-11250-00
Special Care Nursery RN 2Spiked syringe10112.290.660.6%08-11243-00
Labor & Delivery RN 1Spiked syringe10114.990.950.8%08-11243-00
Labor & Delivery RN 2Spiked syringe10114.180.490.4%08-11250-00

Figure 5.7 – Table – Chloride Blood Precision Site 1

UserQC LevelNMeanSD%CVLot
Respiratory Therapist 1Normal syringe10107.430.340.3%08-11250-00
Client Services RepNormal syringe11105.930.410.4%08-11243-00
Anesthesia Tech 1Normal syringe10105.810.390.4%08-11244-00
Phlebotomist 1Normal syringe10109.310.310.3%08-11243-00
OP Surgery Tech 1Normal syringe10106.460.400.4%08-11243-00
OP Surgery Tech 2Normal syringe10107.430.830.8%08-11243-00
Phlebotomist 2Normal syringe10104.570.330.3%08-11243-00
Phlebotomist 3Normal syringe10107.160.790.7%08-11245-00
Respiratory Therapist 1Spiked syringe10135.180.740.5%08-11243-00
Client Services RepSpiked syringe10131.070.750.6%08-11250-00
Anesthesia Tech 1Spiked syringe10131.271.791.4%08-11243-00
Phlebotomist 1Spiked syringe10136.251.551.1%08-11250-00
OP Surgery Tech 1Spiked syringe10115.860.370.3%08-11243-00
OP Surgery Tech 2Spiked syringe10115.871.191.0%08-11245-00
Phlebotomist 2Spiked syringe10115.150.550.5%08-11243-00
Phlebotomist 3Spiked syringe10117.120.810.7%08-11245-00
Client Services RepNormal capillary10106.310.560.5%08-11250-00
Anesthesia Tech 1Normal capillary10107.220.750.7%08-11244-00
OP Surgery Tech 1Normal capillary10106.740.840.8%08-11244-00
Phlebotomist 2Normal capillary10105.210.660.6%08-11243-00
Client Services RepSpiked capillary10131.322.101.6%08-11250-00
Anesthesia Tech 1Spiked capillary10132.081.110.8%08-11244-00
OP Surgery Tech 1Spiked capillary10115.190.880.8%08-11244-00
Phlebotomist 2Spiked capillary10114.820.340.3%08-11243-00

Figure 5.8 - Table - Chloride Blood Precision Site 2

9

UserSiteSampleNMeanSD%CVLot
Medical Technician 11Normal syringe100.640.035.3%07-12137-00
Medical Technician 21Normal syringe100.690.079.7%07-12137-00
Phlebotomist 12Normal syringe100.620.058.7%07-12144-00
Phlebotomist 21Normal syringe100.560.058.6%07-12136-00
Phlebotomist 32Normal syringe100.590.047.4%07-12132-00
Phlebotomist 42Normal syringe100.650.057.7%07-12137-00
Phlebotomist 52Normal syringe100.680.0710.1%07-12144-00
Respiratory Therapist 13Normal syringe90.640.034.3%07-12136-00
Respiratory Therapist 23Normal syringe100.630.058.0%07-12132-00
Respiratory Therapist 33Normal syringe100.480.048.8%07-12144-00
Respiratory Therapist 43Normal syringe100.430.023.6%07-12132-00
Specimen processor1Normal syringe90.630.058.1%07-12144-00
Medical Technician 11Spiked syringe101.700.095.2%07-12144-00
Medical Technician 21Spiked syringe101.600.063.9%07-12144-00
Phlebotomist 12Spiked syringe91.610.106.1%07-12137-00
Phlebotomist 21Spiked syringe101.500.031.7%07-12132-00
Phlebotomist 32Spiked syringe101.520.053.1%07-12136-00
Phlebotomist 42Spiked syringe101.580.053.4%07-12144-00
Phlebotomist 52Spiked syringe101.480.053.7%07-12137-00
Respiratory Therapist 13Spiked syringe101.470.074.9%07-12132-00
Respiratory Therapist 23Spiked syringe91.510.064.2%07-12136-00
Respiratory Therapist 33Spiked syringe101.690.031.6%07-12137-00
Respiratory Therapist 43Spiked syringe101.580.053.2%07-12144-00
Specimen processor1Spiked syringe101.610.105.9%07-12137-00
Phlebotomist 12Normal capillary100.500.059.5%07-12144-00
Phlebotomist 62Normal capillary90.550.046.5%07-12144-00
Specimen processor1Normal capillary100.520.024.3%07-12132-00 (n=7)
07-12136-00 (n=3)
Phlebotomist 12Spiked capillary101.420.075.1%07-12132-00
Phlebotomist 62Spiked capillary101.430.053.4%07-12136-00
Specimen processor1Spiked capillary101.520.053.4%07-12144-00

Figure 5.9 – Table – Creatinine Blood Precision

.

:

:

:

:

.......

.

:

10

Chloride
All Users - All SitesNormalSpikedNormalSpiked
Whole BloodSyringeSyringeCapillaryCapillary
N - Tests1201194040
N - Runs121244
N - Replicates10101010
Within-run
Range102.1-109.3112.0-136.3105.2-107.2114.8-132.1
Mean106.8120.9106.4123.4
Avd SD W-R0.630.860.701.11
CV% W-R0.6%0.7%0.7%0.9%
Creatinine
All Users - All SitesNormal
SyringeSpiked
SyringeNormal
CapillarySpiked
Capillary
Whole Blood
N - Tests1181182930
N - Runs121233
N - Replicates10101010
Within-run
Range0.43-0.691.47-1.700.5-0.551.42-1.52
Mean0.61.570.521.45
Avd SD W-R0.050.060.040.06
CV% W-R7.6%3.9%6.8%3.9%

. .

Figure 5.10 - Tables - Chloride and Creatinine Blood Precision Summaries

. .

:

11

5.6.3 Aqueous precision

Aqueous precision studies were performed at clinical sites to demonstrate precision when analyzed by end users. Samples tested were commercially available blood gas, electrolytes and metabolites control fluids, L1, L2 and L3 (Eurotrol, The Netherlands).

UserQC LevelNMeanSD%CVLot
Special Care Nursery RN 1L11576.650.350.5%08-11243-00
Special Care Nursery RN 2L11576.690.350.5%08-11243-00
Labor & Delivery RN 1L11576.470.851.1%08-11244-00
Labor & Delivery RN 2L11576.600.400.5%08-11243-00
Special Care Nursery RN 1L21598.200.360.4%08-11250-00
Special Care Nursery RN 2L21598.470.300.3%08-11244-00
Labor & Delivery RN 1L21598.660.430.4%08-11243-00
Labor & Delivery RN 2L21598.170.500.5%08-11250-00
Special Care Nursery RN 1L315123.300.650.5%08-11250-00
Special Care Nursery RN 2L315123.290.260.2%08-11250-00
Labor & Delivery RN 1L315123.491.861.5%08-11250-00
Labor & Delivery RN 2L315123.700.520.4%08-11243-00
UserQC LevelNMeanSD%CVLot
NICU RN 1L11576.260.710.9%08-11250-00
Respiratory Therapist 1L11576.710.300.4%08-11244-00
Client Services RepL11576.730.400.5%08-11243-00
Anesthesia Tech 1L11576.640.420.6%08-11243-00
OP Surgery Tech 1L11576.120.430.6%08-11250-00
OP Surgery Tech 2L11576.410.430.6%08-11250-00
Phlebotomist 2L11576.260.240.3%08-11243-00
NICU RN 2L21598.690.570.6%08-11250-00
Respiratory Therapist 1L21499.020.290.3%08-11244-00
Client Services RepL21598.960.650.7%08-11245-00
Anesthesia Tech 1L21499.100.180.2%08-11244-00
Phlebotomist 1L21598.180.300.3%08-11250-00
OP Surgery Tech 1L21598.380.740.8%08-11250-00
OP Surgery Tech 2L21598.450.480.5%08-11243-00
NICU RN 3L315125.390.890.7%08-11250-00
Respiratory Therapist 1L315124.520.370.3%08-11244-00
Client Services RepL314123.430.590.5%08-11250-00
Anesthesia Tech 1L315123.270.370.3%08-11243-00
OP Surgery Tech 1L315123.071.100.9%08-11250-00
OP Surgery Tech 2L314123.530.310.2%08-11243-00

Figure 5.11 - Table - Chloride Aqueous Precision Site 1

Figure 5.12 - Table - Chloride Aqueous Precision Site 2

12

UserQC LevelNMeanSD%CVLot
Obstetrics RN 1L1150.700.045.6%07-12130-00
Phlebotomist 1L1150.680.068.2%07-12129-00
Respiratory Therapist 1L1150.640.046.6%07-12136-00
POC TechnicianL1150.660.057.2%07-12129-00
Obstetrics RN 2L1140.670.023.2%07-12129-00
Phlebotomist 2L1150.640.058.3%07-12132-00
Medical TechnologistL1150.680.045.3%07-12130-00
Obstetrics RN 3L1150.710.034.9%07-12129-00
Obstetrics RN 1L2152.180.094.2%07-12132-00
Phlebotomist 1L2151.910.073.5%07-12130-00
Respiratory Therapist 1L2152.030.052.3%07-12129-00
POC TechnicianL2151.900.063.2%07-12130-00
Obstetrics RN 2L2152.120.115.0%07-12132-00
Phlebotomist 2L2152.060.094.2%07-12136-00
Medical TechnologistL2152.150.094.3%07-12136-00
Obstetrics RN 3L2151.980.063.2%07-12130-00
Obstetrics RN 1L3153.980.133.3%07-12129-00
Phlebotomist 1L3154.220.225.2%07-12132-00
Respiratory Therapist 1L3154.310.225.0%07-12130-00
POC TechnicianL3154.440.184.1%07-12136-00
Obstetrics RN 2L3154.320.153.5%07-12129-00
Phlebotomist 2L3154.170.184.2%07-12129-00
Medical TechnologistL3154.580.255.4%07-12132-00
Obstetrics RN 3L3154.450.173.9%07-12132-00

Figure 5.13 – Table – Creatinine Aqueous Precision

and the comments of the country

:

.

· · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · ·

13

Chloride
Control FluidsQC Level 1QC Level 2QC Level 3
Total Tests
N - Tests165163148
Mean76.598.6123.7
SD0.500.561.06
CV% Total0.7%0.6%0.9%
Run-to-run
N - Runs111110
Mean76.598.6123.7
SD0.210.340.71
CV% R-R0.3%0.3%0.6%
Within-run
N - WR Replicates151515
Avd SD W-R0.440.440.69
CV% W-R0.6%0.4%0.6%
Creatinine
Control FluidsQC Level 1QC Level 2QC Level 3
Total Tests
N - Tests120119120
Mean0.662.044.31
SD0.050.130.27
CV% Total6.8%6.4%6.3%
Run-to-run
N - Runs888
Mean0.662.044.31
SD0.020.010.02
CV% R-R3.4%5.5%4.0%
Within-run
N - WR Replicates151515
Avd SD W-R0.040.080.21
CV% W-R6.1%3.8%4.8%

Figure 5.14 - Tables - Chloride and Creatinine Aqueous Precision Summaries

5.6.4 Matrix Effects

The method comparison studies were performed in field trials at two hospitals on patient samples of whole blood at various locations. Patient specimens were venous, arterial and capillary. The method comparison was against the predicate device.

Chloride epoc vs. i-STAT
VenousArterialCapillaryAll
N494363155
Sxx0.71.40.60.9
Syy0.70.80.90.8
slope1.000.961.020.99
intercept-0.242.35-3.060.20
Syx2.51.91.31.9
X min72697069
X max136136139139
0.970.990.990.98

14

Creatinine epoc vs. Roche Cobas
VenousArterialCapillaryAll
N534249144
Sxx0.090.120.090.10
Syy0.280.320.300.30
slope1.031.041.011.03
intercept-0.12-0.11-0.06-0.10
Syx0.470.480.400.45
X min0.300.300.300.30
X max14.5014.3014.8014.80
R20.990.990.990.99

Figure 5.15 – Table of Method Comparison Summary Against Predicate Device By Sample Matrix Type

Chloride epoc vs. i-STAT, mM
MatrixDecision Level90112
VenousAvg. Bias0.00.0
±95% CI0.80.7
ArterialAvg. Bias-0.9-1.7
±95% CI0.80.4
CapillaryAvg. Bias-1.3-0.9
±95% CI0.40.3
AllAvg. Bias-0.7-1.0
±95% CI0.40.3
Creatinine epoc vs. Roche Cobas, mg/dL
MatrixDecision Level1.251.60
VenousAvg. Bias-0.08-0.07
±95% CI0.050.05
ArterialAvg. Bias-0.06-0.04
±95% CI0.070.07
CapillaryAvg. Bias-0.05-0.04
±95% CI0.030.03
AllAvg. Bias-0.06-0.05
±95% CI0.030.03

Figure 5.16 - Table of Method Comparison Summary Against Predicate Device -Consolidated Bias by Sample Matrix Type :

15

Effect of Anticoagulant 5.6.4.1

The effect of anticoagulant was evaluated on patient samples that were collected using heparinized and non-heparinized collection devices. This study was performed at a hospital (46 samples) and supplemented with in-house studies (29 samples). The data was analyzed using EP9-2A methodology.

Chloride epoc Hep vs. No-HepCreatinine epoc Hep vs. No-Hep
NN
7677
SxxSxx
0.60.26
SyySyy
0.80.33
slopeslope
0.980.99
interceptintercept
1.920.02
SyxSyx
1.20.23
X minX min
990.42
X maxX max
12910.53
R2R2
0.970.99

Figure 5.17 – Table of Heparinized Versus Non-Heparinized Samples

Summary of Conclusions Drawn from Non Clinical and 5.7 Clinical Tests

We conclude from the data presented in section 5.5 that the device performs effectively. We conclude from the data section 5.6 that the clinical performance of the device is substantially equivalent to the predicate devices.

16

DEPARTMENT OF HEALTH & HUMAN SERVICES

Food and Drug Administration

10903 New Hampshire Avenue Silver Spring, MD 20993

EPOCAL Inc. c/o Roy Layer, Vice President, Quality Assurance & Regulatory Affairs 2060 Walkley Road Ottawa, Ontario, Canada K1G 3P5

Re: K113726

Trade Name: epoc Chloride test, epoc Creatinine test Regulation Number: 21 CFR §862.1170 Regulation Name: Chloride test system Regulatory Class: Class II Product Codes: CGZ: CGL Dated: September 12, 2012 Received: September 13, 2012

OCT, 5 2012

Dear Mr. Layer:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820).

17

Page 2

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please 11 you dosire specific as vior Diagnostic Devices and Radiological Health at (301) 796-5450. comact the Office of in Vitro Diaglious "Misbranding by reference to premarket notification" Also, prease note mo regarding postmarket surveillance, please contact CDRH'S (21 CF R Fur 001:37): 1 or questions : 05 St s) Division of Postmarket Surveillance at (301) Office of but vehiclibe and Drolling of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/Medical

Devices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance ...

You may obtain other general information on your responsibilities under the Act from the Tou may ootain other general manational and Consumer Assistance at its toll-free number (800 638-2041 or (301) 796-5680 or at its Internet address http://www.fda/gov/MedicalDevices/ResourcesforYou/Industry/default.htm

Sincerely yours,

$

Courtney H. Lias, Ph.D Director Division of Chemistry and Toxicology Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

18

Indications for Use

510(k): K113726

Device Name: epoc Chloride test

Indication For Use:

The Chloride test, as part of the epoc Blood Analysis System, is intended for use by trained medical professionals as an in vitro diagnostic device for the quantitative testing of samples of heparinized or un-anticoagulated arterial, venous or capillary whole blood in the laboratory or at the point of care.

Chloride measurements from the epoc Blood Analysis System are used in the diagnosis and treatment of electrolyte and metabolic disorders.

Prescription Use _ X (21 CFR Part 801 Subpart D) And/Or

Over the Counter Use _ (21 CFR Part 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE; CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of In Vitro Diagnostic Device Evaluation and Safety (OIVD)

Yungs Chan

Division Sign-Off Office of In Vitro Diagnostic Device Evaluation and Safety

510(k) K113726

19

Indications for Use

510(k): K113726

Device Name: epoc Creatinine test

Indication For Use:

The Creatinine test, as part of the epoc Blood Analysis System, is intended for use by trained medical professionals as an in vitro diagnostic device for the quantitative testing of samples of heparinized or un-anticoagulated arterial, venous or capillary whole blood in the laboratory or at the point of care.

Creatinine measurements from the epoc Blood Analysis System are used in the diagnosis and treatment of certain renal diseases and in monitoring renal dialysis.

Prescription Use X (21 CFR Part 801 Subpart D) And/Or

Over the Counter Use (21 CFR Part 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE; CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of In Vitro Diagnostic Device Evaluation and Safety (OIVD)

Yung Chan

Division Sign-Off Office of In Vitro Diagnostic Device Evaluation and Safety

510(k) K113726