K Number
K171247

Validate with FDA (Live)

Manufacturer
Date Cleared
2018-01-17

(264 days)

Product Code
Regulation Number
862.1770
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Blood Urea Nitrogen and Total Carbon Dioxide tests, as part of the epoc Blood Analysis System, is intended for use by trained medical professionals as an in vitro diagnostic device for the quantitative testing of samples of heparinized or un-anticoagulated arterial, venous or capillary whole blood in the laboratory or at the point of care.

Blood Urea Nitrogen measurements from the epoc Blood Analysis System are used in the diagnosis and treatment of certain renal and metabolic diseases.

Total Carbon Dioxide measurements from the epoc Blood Analysis System are used in the diagnosis and treatment of disorders associated with changes in body acid-base balance.

Device Description

The epoc Blood Analysis System is an in vitro diagnostic device system for the quantitative testing of blood gases, electrolytes, and metabolites in venous, arterial, and capillary whole blood samples. The epoc System is comprised of 3 major subsystems: epoc Host, epoc Reader and epoc BGEM Test Card. The main accessory used with the epoc System includes the epoc Care-Fill Capillary Tubes used to collect and introduce capillary blood samples into the epoc Test Card.

The epoc Blood Analysis System was previously cleared for prescription use to quantitate pH, pCO2, pO2, Na, K, iCa, Cl, Glu, Lact, Crea, and Hct in arterial, venous, and capillary blood samples per K061597, K090109, K092849, K093297, and K113726. This premarket notification submission adds blood urea nitrogen (BUN) and total carbon dioxide (TCO2) quantitation to the epoc BGEM Test Card and Blood Analysis System.

AI/ML Overview

The epoc Blood Urea Nitrogen Test and epoc Total Carbon Dioxide Test, as part of the epoc Blood Analysis System, are intended for use by trained medical professionals as an in vitro diagnostic device for quantitative testing of heparinized or un-anticoagulated arterial, venous or capillary whole blood.

The acceptance criteria and device performance are described in several studies:

Acceptance Criteria and Device Performance:

StudyAcceptance CriteriaReported Device Performance
Analytical Sensitivity (LoB, LoD, LoQ per CLSI EP17-A2)Not explicitly stated as acceptance criteria, but demonstrates detection limits.BUN: LoB 2 mg/dL, LoD 3 mg/dL, LoQ 3 mg/dL
TCO2: LoB 4.0 mM, LoD 4.3 mM, LoQ 4.3 mM
Linearity (per CLSI EP06-A)Not explicitly stated as acceptance criteria, but demonstrates linearity across reportable range.BUN (4-119 mg/dL): Slope 1.020, Intercept 0.4, R 0.9989
TCO2 (4-49 mmol/L): Slope 0.903, Intercept 3.32, R 0.9997
Precision (Aqueous Controls) (CLSI EP05-A3)Not explicitly stated as acceptance criteria, but demonstrates precision.BUN High Level (51.7 mg/dL): SWR 1.01 (2.0% CV), ST 1.16 (2.3% CV)
BUN Low Level (7.1 mg/dL): SWR 0.30 (4.2% CV), ST 0.32 (4.5% CV)
TCO2 High Level (30.7 mmol/L): SWR 0.82 (2.7% CV), ST 0.92 (3.0% CV)
TCO2 Low Level (16.2 mmol/L): SWR 0.88 (5.4% CV), ST 1.02 (6.3% CV)
Interference (CLSI EP07-A2)Unacceptable interference bias defined as producing a significant error more than 5% of the time.Clinically significant interfering substances for BUN and TCO2 are itemized and reported. Various exogenous and endogenous interferences were tested and found to be clinically insignificant below certain thresholds.
Clinical Field Precision (Aqueous Controls) (CLSI EP05-A3)Not explicitly stated as acceptance criteria, but demonstrates precision in a clinical setting.BUN Level 1 (52.1 mg/dL): SWR 1.06 (2.0%), Total Reproducibility 1.54 (3.0%)
BUN Level 2 (17.7 mg/dL): SWR 0.45 (2.5%), Total Reproducibility 1.11 (6.3%)
BUN Level 3 (7.1 mg/dL): SWR 0.24 (3.4%), Total Reproducibility 0.26 (3.7%)
TCO2 Level 1 (15.9 mM): SWR 0.44 (2.8%), Total Reproducibility 0.50 (3.1%)
TCO2 Level 2 (19.7 mM): SWR 0.66 (3.4%), Total Reproducibility 0.78 (3.9%)
TCO2 Level 3 (30.4 mM): SWR 0.58 (1.9%), Total Reproducibility 1.05 (3.4%)
Clinical Field Precision (Whole Blood)Not explicitly stated as acceptance criteria, but demonstrates precision in a clinical setting.BUN Hi-Syringe (57.4 mg/dL): %CV 2.3%
BUN Lo-Cap Tube (7.6 mg/dL): %CV 7.0%
TCO2 Hi-Syringe (36.5 mM): %CV 1.5%
TCO2 Lo-Cap Tube (13.5 mM): %CV 3.5%
Method Comparison (BUN) (CLSI EP09-A3)Not explicitly stated as a numerical acceptance criterion, but implies a high correlation with the reference method.Comparing epoc BUN to Roche Cobas 8000: Slope 0.985, Intercept 0.3, R 0.998, Mean Bias at 26 mg/dL -0.1+0.2
Method Comparison (TCO2)Not explicitly stated as a numerical acceptance criterion, but implies a high correlation with the reference method.Comparing epoc TCO2 to i-STAT-CHEM8+: Slope 1.039, Intercept -0.8, R 0.974, Mean Bias at 20 mM 0.0+0.2
Matrix Comparison: AnticoagulantNo significant difference between results in Li-heparinized, Na-heparinized, and non-anticoagulated blood samplesConcluded no significant difference in BUN and TCO2 results.

Study Information:

  1. Sample sizes used for the test set and the data provenance:

    • Analytical Sensitivity (LoB, LoD, LoQ): Test samples were prepared from dialyzed whole blood. The specific number of samples or runs is not explicitly stated, but the study was conducted according to CLSI EP17-A2.
    • Linearity: Multiple whole blood samples were used, spanning the reportable range. Conducted per CLSI EP06-A. Specific number not provided.
    • Precision (Aqueous Controls): 320 replicates for each level of both BUN and TCO2. These were in-house measurements.
    • Clinical Field Precision (Aqueous Controls): N=170 for BUN Level 1, 171 for Level 2, 168 for Level 3. N=172 for TCO2 Level 1, 170 for Level 2, 169 for Level 3. Data provenance is from "three different clinical sites."
    • Clinical Field Precision (Whole Blood): N=134-136 for BUN samples, N=134-139 for TCO2 samples, depending on the type (syringe/cap tube) and level (high/NB/low). Data provenance is from "three different clinical sites."
    • Precision (Duplicate Epoc Test Results): Over 430 patient tests run in duplicate. "Approximately equal numbers of venous, arterial and capillary samples." Data provenance not explicitly stated (e.g., country of origin), assumed to be from clinical sites in the context of "Clinical Field Precision." This is prospective clinical data.
    • Method Comparison (BUN): N=433 venous, arterial, and capillary blood samples. Performed at "three clinical sites." This is prospective clinical data.
    • Method Comparison (TCO2): N=574 venous, arterial, and capillary patient samples. Performed at "three clinical sites." This is prospective clinical data.
    • Matrix Comparison: Anticoagulant: Over 60 volunteer donors, with samples further aliquoted into 3 vacutainers each. Data provenance not explicitly stated.
  2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable. This device is a quantitative diagnostic test for chemical analytes (BUN, TCO2), not an imaging or qualitative diagnostic device requiring expert interpretation for ground truth. The ground truth for analytical performance studies is typically established using reference methods (e.g., IDMS-traceable laboratory system) or prepared reference materials.

  3. Adjudication method for the test set: Not applicable. The ground truth for quantitative chemical analytes is established by reference methods or gravimetric preparation, not through human adjudication.

  4. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable. This device is a diagnostic testing system for chemical analytes, not an AI-assisted diagnostic imaging or qualitative interpretation tool for human readers.

  5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Yes, the entire performance characterization (analytical sensitivity, linearity, precision, interference, method comparison, and matrix comparison) represents standalone algorithm/device performance. The device provides quantitative results directly. Human-in-the-loop performance is about accuracy of human readers, and the clinical field precision study assesses the precision of the device in the hands of intended users, not the interpretative performance of those users.

  6. The type of ground truth used (expert consensus, pathology, outcomes data, etc):

    • Analytical Sensitivity, Linearity, Precision: Ground truth established via prepared reference materials (dialyzed whole blood, gravimetric mixtures of high/low samples) and aqueous controls with known concentrations.
    • Method Comparison (BUN): Ground truth established by an "IDMS-traceable plasma/serum-based laboratory system" (Roche Cobas 8000).
    • Method Comparison (TCO2): Ground truth established by a "whole blood point-of-care system" (i-STAT-CHEM8+), which is also a predicate device.
    • Interference and Matrix Comparison: Comparisons were made against control samples (e.g., solvent added, or anticoagulant-free) to assess the impact of interfering substances or different matrices.
  7. The sample size for the training set: Not applicable. This document describes the performance of a chemical analyte detection system, not a machine learning or AI model that requires a training set.

  8. How the ground truth for the training set was established: Not applicable, as there is no training set for this device.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo in blue, with the words "U.S. FOOD & DRUG" stacked on top of the word "ADMINISTRATION".

January 17, 2018

Epocal Inc. Jennifer Armstrong Regulatory Affairs Manager 2060 Walkley Road Ottawa, ON K1G 3P5 Canada

Re: K171247

Trade/Device Name: epoc Blood Urea Nitrogen Test, epoc Total Carbon Dioxide Test Regulation Number: 21 CFR 862.1770 Regulation Name: Urea nitrogen test system Regulatory Class: II Product Code: CDS. JFL Dated: December 11, 2017 Received: December 13, 2017

Dear Jennifer Armstrong:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR

{1}------------------------------------------------

Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Kellie B. K

for Courtney H. Lias, Ph.D. Director Division of Chemistry and Toxicology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K171247

Device Name epoc® Blood Urea Nitrogen Test epoc® Total Carbon Dioxide Test

Indications for Use (Describe)

The Blood Urea Nitrogen and Total Carbon Dioxide tests, as part of the epoc Blood Analysis System, is intended for use by trained medical professionals as an in vitro diagnostic device for the quantitative testing of samples of heparinized or un-anticoagulated arterial, venous or capillary whole blood in the laboratory or at the point of care.

Blood Urea Nitrogen measurements from the epoc Blood Analysis System are used in the diagnosis and treatment of certain renal and metabolic diseases.

Total Carbon Dioxide measurements from the epoc Blood Analysis System are used in the diagnosis and treatment of disorders associated with changes in body acid-base balance.

Type of Use (Select one or both, as applicable)
-------------------------------------------------

Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW .

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) SUMMARY K171247

GENERAL INFORMATION

Applicant Name:Epocal Inc.2060 Walkley RoadOttawa, ON K1G 3P5 Canada
Company Contact:Jennifer ArmstrongManager, Regulatory AffairsPhone: (613) 688-3982 x2227Email: jennifer.armstrong@alere.com
Date Prepared:January 15, 2018

DEVICE IDENTIFICATION

Trade or Proprietary Names: epoc® Blood Urea Nitrogen Test epoc® Total Carbon Dioxide Test

REGULATORY INFORMATION

Classification Regulation:21 CFR 862.1770Urea nitrogen test system
21 CFR 862.1160Bicarbonate/carbon dioxide test system
Regulatory Class:Class II
Product Codes:CDSElectrode, Ion Specific, Urea Nitrogen
JFLpH Rate Measurement, Carbon-Dioxide
Predicate Device:i-STAT CHEM8+ Cartridge (K053110; cleared by i-STAT Corporation)

DEVICE DESCRIPTION

The epoc Blood Analysis System is an in vitro diagnostic device system for the quantitative testing of blood gases, electrolytes, and metabolites in venous, arterial, and capillary whole blood samples. The epoc System is comprised of 3 major subsystems: epoc Host, epoc Reader and epoc BGEM Test Card. The main accessory used with the epoc System includes the epoc Care-Fill Capillary Tubes used to collect and introduce capillary blood samples into the epoc Test Card.

{4}------------------------------------------------

The epoc Blood Analysis System was previously cleared for prescription use to quantitate pH, pCO2, pO2, Na, K, iCa, Cl, Glu, Lact, Crea, and Hct in arterial, venous, and capillary blood samples per K061597, K090109, K092849, K093297, and K113726. This premarket notification submission adds blood urea nitrogen (BUN) and total carbon dioxide (TCO2) quantitation to the epoc BGEM Test Card and Blood Analysis System.

INTENDED USE

The Blood Urea Nitrogen and Total Carbon Dioxide tests, as part of the epoc Blood Analysis System, is intended for use by trained medical professionals as an in vitro diagnostic device for the quantitative testing of samples of heparinized or unanticoagulated arterial, venous or capillary whole blood in the laboratory or at the point of care.

Blood Urea Nitrogen measurements from the epoc Blood Analysis System are used in the diagnosis and treatment of certain renal and metabolic diseases.

Total Carbon Dioxide measurements from the epoc Blood Analysis System are used in the diagnosis and treatment of disorders associated with changes in body acid-base balance.

AttributePredicate Devicei-STAT CHEM8+ Cartridge(with i-STAT Portable ClinicalAnalyzer) [K053110]Candidate Deviceepoc BGEM Test Card with epoc BloodAnalysis System
Intended usePortable, prescription use test systemPrescription, point-of-care test system
MeasuredParameterUrea Nitrogen (BUN);Total CO2 (TCO2)Blood Urea Nitrogen (BUN);Total CO2 (TCO2)
CalculatedParameterAnion Gap (AnGap);Anion Gap (AGap, AGapK);BUN/Creatinine ratio (BUN/Crea)
Where usedhospital, point of careSame
Sample typeVenous, arterial and capillary wholebloodSame
TechnologyAn electrochemical multi-sensor arrayintegrated into a single-use test that isinterpreted by a handheld reader andassociated softwareSame
Reportable ranges(BUN and TCO2)BUN 3-140 mg/dLTCO2 5-50 mmol/LBUN 3-120 mg/dLTCO2 same
Sample volume95 µLAt least 92 µL

COMPARISON WITH PREDICATE

{5}------------------------------------------------

PERFORMANCE CHARACTERISTICS

1. Analytical Sensitivity

This study evaluated and verified the performance of the epoc Blood Analysis System for BUN and TCO2 quantitation at the low end of their respective concentration ranges by determining the Limit of Blank (LoB), Limit of Detection (LoD) and Limit of Quantitation (LoQ) according to CLSI EP17-A2. Test samples were prepared from dialyzed whole blood. Results from this study are shown below:

AnalyteLoBLoDLoQ
BUN2 mg/dL3 mg/dL3 mg/dL
TCO24.0 mM4.3 mM4.3 mM

2. Linearity

Linearity was performed in-house on multiple whole blood samples with BUN or TCO2 values spanning the reportable range. Linearity is reported versus theoretical BUN values based on gravimetric mixtures of high and low BUN samples (as measured using an in-house standard whole blood BUN method). Three card lots were used in this study. The study was conducted per CLSI EP06-A.

BUN
Test RangeSlopeInterceptR
4-119 mg/dL1.0200.40.9989
TCO2
Test RangeSlopeInterceptR
4-49 mmol/L0.9033.320.9997

3. Precision (Aqueous Controls)

Analytical precision for BUN and TCO2 measurements was conducted with four card lots using at least 25 epoc Readers where replicate measurements were run in-house twice a day for twenty days for each fluid per CLSI EP05-A3. In the precision data tables below, Swx denotes within-run standard deviation, %CVwr denotes within-run coefficient of variation, Sr denotes total standard deviation, and %CVr denotes total coefficient of variation.

Aqueous ControlUnitsNMeanSWR%CVWRST%CVT
High Level (BUN)mg/dL32051.71.012.0%1.162.3%
Low Level (BUN)mg/dL3207.10.304.2%0.324.5%
High Level (TCO2)mmol/L32030.70.822.7%0.923.0%
Low Level (TCO2)mmol/L32016.20.885.4%1.026.3%

{6}------------------------------------------------

4. Interference

Interferent testing of the BUN and TCO2 measurements on the epoc System was performed as recommended in the CLSI guideline EP07-A2. In each of these tests, human serum specimens were aliquoted into two (2) samples. The test sample was spiked by addition of interferent, while the control sample was spiked by the addition of the solvent of the interferent. The bias between the mean of six (6) replicates on both the control sample and the test sample with added interferent was calculated. Unacceptable interference bias was defined as producing a significant error more than 5% of the time.

Clinically significant interfering substances for BUN measurements are itemized below:

  • . Samples contaminated with benzalkonium salts used as coatings for in-dwelling lines may cause elevated BUN results. For proper line-flushing procedures refer to CLSI H11-A4.
  • Citrate will have no significant effect up to 6.0 mmol/L (176.5 mg/dL) after which it will decrease the BUN concentration by up to 0.26 mg/dL BUN per mmol/L citrate.
  • EDTA will have no significant effect up to 4.5 mmol/L (167 mg/dL) after which it ● will decrease the BUN concentration by up to 0.43 mg/dL BUN per mmol/L EDTA.
  • . Glutathione reduced will have no significant effect up to 1.7 mmol/L (52.2 mg/dL), after which it will increase the BUN concentration by up to 1.91 mg/dL BUN per mmol/L glutathione reduced. Blood glutathione (GSH) in human subjects is ~0.79-1.05 mmol/L. Long term oral glutathione reduced supplementation (250-1,000 mg/day administered for 6 months) increases glutathione plasma levels by ~0.2-8 µmol/L (~0.01-0.25 mg/dL). Short-term, oral intake of glutathione reduced does not affect plasma glutathione levels.
  • ß-Hydroxybutyrate will have no significant effect up to 17.2 mmol/L (216.9 mg/dL), after which it will decrease the BUN concentration by up to 0.11 mg/dL BUN per mmol/L hydroxybutyrate. The reference range for ß-hydroxybutyrate in plasma is <0.4 to 0.5 mmol/L. ß-hydroxybutyrate concentration over 3 mmol/L are indicative of ketoacidosis; in very severe diabetic ketoacidosis the concentration may exceed 25 mmol/L.
  • Hydroxyurea will have no significant effect up to 1.3 mmol/L (9.9 mg/dL), after which it will increase the BUN concentration by up to 1.61 mg/dL BUN per mmol/L hydroxyurea. The recommended dose of hydroxyurea for patients range from 15 mg/kg/day to 30 mg/kg/day. A treatment dose of 2,000 mg/day (~30mg/kg) results in maximum plasma concentration of ~800µmol/L with oral administration and ~1 mmol/L with intravenous method.

{7}------------------------------------------------

ероса I Inc.

  • . N-acetylcysteine will have no significant effect up to 9.2 mmol/L (150.1 mg/dL), after which it will increase the BUN concentration by up to 0.11 mg/dL BUN per mmol/L N- acetylcysteine. It has been reported that 1 mmol/L N-acetyl cysteine is therapeutically unattainable in plasma. The therapeutic level for N-acetyl cysteine is 0.3 mmol/L.
  • Nithiodote will have no significant effect up to 4.1 mmol/L (64.8 mg/dL) after which it will decrease the BUN concentration by up to 0.41 mg/dL BUN per mmol/L Nithiodote. The expected peak sodium thiosulfate plasma concentration following a 12.5 g of Nithiodote is 16.7 mmol/L.

The following levels of exogenous interferences were tested and found to be clinically insignificant for BUN measurements: 1.324 mmol/L (20 mg/dL) acetaminophen, 2 mmol/L (21.6 mg/dL) Li acetoacetic acid, 3.62 mmol/L (65.2 mg/dL) acetyl salicylic acid, 1 mmol/L (5.349 mg/dL) ammonium chloride, 342 µmol/L (6.8 mg/dL) Na ascorbate, 37.5 mmol/L (386 mg/dL) Na bromide, 2.643 mmol/L (125.9 mg/dL) Na cefazolin, 1.46 mmol/L (96.6 mg/dL) Na ceftriaxone, 5.87 µmol/L (0.1 mg/dL) dopamine HCL, 86.8 mmol/L (400 mg/dL) ethanol, 50 umol/L (4.46 mg/dL) (Flaxedil™) gallamine triethiodide, 28 mmol/L (0.5 g/dL) glucose, 2.55 mmol/L (156 mg/dL) oxidized glutathione, 5 mmol/L (38 mg/dL) glycolic acid, 20 U/mL heparin, 2.43 mmol/L (50 mg/dL) ibuprofen, (0.5%) 500 mg/dL intralipid, 1.3 mmol/L (19.4 mg/dL) Na iodide, 1 mmol/L (12 mg/dL) L-cysteine, 25 µmol/L (~0.5 mg/dL) L-Dopa, 3.2 mmol/L (13.5 mg/dL) lithium chloride, 6 mmol/L (210.8 mg/dL) Na metamizole, 2 mmol/L (90 mg/dL) methotrexate, 0.22 mmol/L (4 mg/dL) oxalate (K) monohydrate, 248 µmol/L (6.5 mg/dL) Na pentothal, 1 mmol/L (12.2 mg/dL) Na perchlorate, 4.34 mmol/L (69.5 mg/dL) Na salicylate, 1.72 mmol/L (16.7 mg/dL) K thiocyanate.

The following levels of endogenous interferences were tested and found to be clinically insignificant for BUN measurements: 342 µmol/L (28.8 mg/dL) bilirubin conjugated, 428 umol/L (25 mg/dL) bilirubin unconjugated, 35 mmol/L bicarbonate, Hct 20% to 60% PCV, 6.6 mmol/L (74 mg/dL) lactate, pH 6.8 to 8, 3.5% to 10% total protein, 1.4 mmol/L (23.5 mg/dL) uric acid.

Clinically significant interfering substances for TCO2 measurements are itemized below:

  • Samples contaminated with benzalkonium salts used as coatings for in-dwelling lines may cause significant decrease in TCO2 results. For proper line-flushing procedures refer to CLSI H11-A4.
  • Citrate will have no significant effect up to 11.8 mmol/L (347.0 mg/dL) after ● which it will increase the TCO2 concentration by up to 0.24 mmol/L TCO2 per mmol/L citrate.

{8}------------------------------------------------

  • EDTA will have no significant effect up to 4.8 mmol/L (178.7 mg/dL) after which it will increase the TCO2 concentration by up to 0.57 mmol/L TCO2 per mmol/L EDTA.
  • N-acetyl cysteine will have no significant effect up to 9.6 mmol/L (156.7 mg/dL) ● after which it will increase the TCO2 concentration by up to 0.54 mmol/L TCO2 per mmol/L N-acetyl cysteine. It has been reported that 1 mmol/L N-acetyl cysteine is therapeutically unattainable in plasma. The therapeutic level for Nacetyl cysteine is 0.3 mmol/L.

The following levels of exogenous interferences were tested and found to be clinically insignificant for TCO2measurements: 1.324 mmol/L (20 mg/dL) acetaminophen, 2 mmol/L (21.6 mg/dL) Li acetoacetic acid, 3.62 mmol/L (65.2 mg/dL) acetyl salicylic acid, 1 mmol/L (5.349 mg/dL) ammonium chloride, 342 umol/L (6.8 mg/dL) Na ascorbate, 37.5 mmol/L (386 mg/dL) Na bromide, 2.643 mmol/L (125.9 mg/dL) Na cefazolin, 1.46 mmol/L (96.6 mg/dL) Na ceftriaxone, 5.87 µmol/L (0.1 mg/dL) dopamine HCL, 86.8 mmol/L (400 mg/dL) ethanol, 50 umol/L (4.46 mg/dL) (Flaxedil™) gallamine triethiodide, 28 mmol/L (0.5 g/dL) glucose, 2.55 mmol/L (156 mg/dL) oxidized glutathione, 2.55 mM (78.4 mg/dL) reduced glutathione, 5 mmol/L (38 mg/dL) glycolic acid, 20 U/mL heparin, 2 mmol/L (15 mg/dL) hydroxyurea, 2.43 mmol/L (50 mg/dL) ibuprofen, (0.5%) 500 mg/dL intralipid, 1.3 mmol/L (19.5 mg/dL) Na iodide, 1 mmol/L (12 mg/dL) L-cysteine, 25 µmol/L (~0.5 mg/dL) L-Dopa, 3.2 mmol/L (13.5 mg/dL) lithium chloride, 6 mmol/L (210.8 mg/dL) Na metamizole, 2 mmol/L (90 mg/dL) methotrexate, 16.7 mmol/L (264 mg/dL) Nithiodote, 0.22 mmol/L (4 mg/dL) oxalate (K) monohydrate, 248 µmol/L (6.5 mg/dL) Na pentothal, 1 mmol/L (12.2 mg/dL) Na perchlorate, 4.34 mmol/L (69.5 mg/dL) Na salicylate, 1.72 mmol/L (16.7 mg/dL) K thiocyanate.

The following levels of endogenous interferences were tested and found to be clinically insignificant for TCO2measurements: 342 µmol/L (28.8 mg/dL) bilirubin conjugated, 428 µmol/L (25 mg/dL) bilirubin unconjugated, Hct 20% to 60% PCV, 20 mmol/L (252 mg/dL) ß-hydroxybutyrate, 6.6 mmol/L (74 mg/dL) lactate, pH 6.8 to 8, 3.5% to 10% total protein, 1.4 mmol/L (23.5 mg/dL) uric acid.

5. Clinical Field Precision

The external precision study was conducted to evaluate the precision of the BUN and TCO2 quantitation on the epoc System in the hands of the intended users. The study was evaluated based on CLSI guideline EP05-A3 at three different clinical sites using a different lot of epoc test card at each site. All testing was performed by existing or potential POC operators. Testing was comprised of three parts: 1) aqueous control precision using syringes, 2) whole blood precision using syringes, 3) whole blood precision using capillary tubes.

{9}------------------------------------------------

еросаllnc.

Clinical Field Precision with Aqueous Controls

ParameterAqueous Control Fluid
Level 1Level 2Level 3
BUN [mg/dL]
N170171168
Mean BUN [mg/dL]52.117.77.1
Repeatability (SWR [SD],%CV)1.062.0%0.452.5%0.243.4%
Between-day (SD, %CV)0.941.8%0.482.7%0.030.5%
Between-site (SD, %CV)0.601.2%0.905.1%0.101.4%
Total Reproducibility(ST [SD], %CV)1.543.0%1.116.3%0.263.7%
TCO2 [mM]
N172170169
Mean TCO2 [mM]15.919.730.4
Repeatability (SWR [SD],%CV)0.442.8%0.663.4%0.581.9%
Between-day (SD, %CV)0.161.0%0.201.0%0.762.5%
Between-site (SD, %CV)0.181.1%0.341.7%0.421.4%
Total Reproducibility(ST [SD], %CV)0.503.1%0.783.9%1.053.4%

Clinical Field Precision with Whole Blood

Sample IDNum.RunsNum. ofOperatorsnAvgMinMaxSWR%CV
BUN [mg/dL]
Hi-Syringe121213457.451.872.41.32.3%
Hi-Cap Tube121213655.551.360.31.62.9%
NB-Syringe121213617.312.535.30.74.1%
NB-Cap Tube121213515.611.920.80.63.9%
Lo-Syringe12121367.03.710.60.67.2%
Lo-Cap Tube12121357.65.99.70.57.0%
TCO2 [mM]
Hi-Syringe121213436.533.840.00.61.5%
Hi-Cap Tube121213934.131.736.20.72.1%
NB-Syringe121213627.522.330.90.41.4%
NB-Cap Tube121213725.622.428.30.72.9%
Lo-Syringe121213610.55.015.90.43.7%
Lo-Cap Tube121213413.511.215.00.53.5%

Hi = High Level; NB = Normal Blood Range; Lo = Low Level

{10}------------------------------------------------

Precision was additionally assessed on duplicate epoc test results during the Method Comparison Studies. Over 430 patient tests were run in duplicate with approximately equal numbers of venous, arterial and capillary samples. Pooled pair-wise precision was estimated over three concentration ranges for BUN and two concentration ranges for TCO2.

ParameterBUN [mg/dL]TCO2 [mM]
Range<2222-100>100<40>40
N2531431252423
Average Reading13.144.2111.124.544.8
Pair Precision (SD)0.61.21.60.61.0
%CV4.6%2.7%1.4%2.6%2.2%

6. Method Comparison

Urea method comparison studies were performed at three clinical sites per CLSI EP09-A3. Venous, arterial and capillary blood samples for a total of over 140 results for each blood type were compared an IDMS-traceable plasma/serum-based laboratory system. Pooled results are shown below.

BUN [mg/dL]Roche Cobas 8000
N433
Sxx0.5
Syy0.9
Intercept0.3
Slope0.985
Syx1.8
Xmin3
Xmax118
R0.998
Mean Bias at 26 mg/dL-0.1+0.2

{11}------------------------------------------------

ероса I In с.

TCO2 method comparison studies were performed at three clinical sites. Venous, arterial and capillary patient samples for a total of over 150 results for each blood type were compared with a whole blood point-of-care system. Pooled results are shown below.

TCO2 [mM]i-STAT-CHEM8+
N574
Sxx0.68
Syy0.64
Intercept-0.8
Slope1.039
Syx1.52
Xmin7
Xmax49
R0.974
Mean Bias at 20 mM0.0 + 0.2

7. Matrix Comparison: Anticoagulant

A method comparison approach was used to compare the epoc BUN and TCO2 results in venous blood samples, collected from over 60 volunteer donors into evacuated tubes containing no additive, and further aliquoted into three (3) vacutainers containing noadditive, Li-heparin and Na-heparin to create 3-way matched samples. It was concluded from the analysis that there was no significant difference between BUN and TCO2 results in Li-heparinized, Na-heparinized and non-anticoagulated blood samples on the epoc System.

CONCLUSION

The information provided in this pre-market notification demonstrates that the epoc BGEM Test Card and Blood Analysis System is substantially equivalent to the legally marketed predicate device for its intended use.

§ 862.1770 Urea nitrogen test system.

(a)
Identification. A urea nitrogen test system is a device intended to measure urea nitrogen (an end-product of nitrogen metabolism) in whole blood, serum, plasma, and urine. Measurements obtained by this device are used in the diagnosis and treatment of certain renal and metabolic diseases.(b)
Classification. Class II.