(140 days)
Not Found
No
The description focuses on standard in vitro diagnostic technology, sensor measurements, and data calculation/display. There is no mention of AI, ML, or related concepts.
No
The device is described as an "in vitro diagnostic device" used for quantitative testing of blood samples to provide measurements that aid in the diagnosis and treatment of various conditions. It does not directly perform therapy.
Yes
The "Intended Use / Indications for Use" section explicitly states, "The epoc® Blood Analysis System is intended for use by trained medical professionals as an in vitro diagnostic device..." Additionally, the "Device Description" also refers to it as an "in vitro diagnostic device system".
No
The device description clearly outlines three major hardware components: the epoc® BGEM Test Card, the epoc® Reader, and the epoc® Host (a mobile computer). While the epoc Host runs software, the system relies on physical hardware for sample analysis and data acquisition.
Yes, this device is an IVD (In Vitro Diagnostic).
Here's why:
- Intended Use/Indications for Use: The document explicitly states that the epoc® Blood Analysis System is "intended for use by trained medical professionals as an in vitro diagnostic device". It also details the specific analytes measured (blood gases, electrolytes, metabolites) and how these measurements are used in the "diagnosis and treatment" of various medical conditions. This directly aligns with the definition of an IVD, which is used to examine specimens from the human body to provide information for diagnosis, monitoring, or treatment.
- Device Description: The description clearly outlines the components of the system, including the "epoc® Blood Gas Electrolyte Metabolite (BGEM) Test Card" which is a "single-use, device with a port for blood sample introduction which contains the sensor configurations for testing" various analytes. This describes the in vitro nature of the testing, where a sample is taken from the body and analyzed outside of it.
- Predicate Device(s): The listed predicate devices are also identified as "epoc® Blood Analysis System", further indicating that this system falls under the category of IVDs.
The information provided strongly supports the classification of the epoc® Blood Analysis System as an In Vitro Diagnostic device.
N/A
Intended Use / Indications for Use
The epoc® Blood Analysis System is intended for use by trained medical professionals as an in vitro diagnostic device for the quantitative testing of samples of heparinized or un-anticoagulated arterial, venous, or capillary whole blood in the laboratory or at the point of care.
The Blood Gas Electrolyte and Metabolite (BGEM) Test Card panel configuration includes sensors that quantitate pH, pCO2, pO2, Sodium, Potassium, Ionized Calcium, Chloride, Total Carbon Dioxide, Glucose, Lactate, Blood Urea Nitrogen, Creatinine, and Hematocrit.
pH, pCO2, p02 (blood gases) measurements from the epoc® Blood Analysis System are used in the diagnosis and treatment of life-threatening acid-base disturbances.
Sodium and Potassium measurements from the epoc® Blood Analysis System are used in diagnosis and treatment of diseases involving electrolyte imbalance.
lonized Calcium measurements from the epoc® Blood Analysis System is used in diagnosis and treatment of parathyroid disease, a variety of bone diseases, chronic renal disease and tetany.
Chloride measurements from the epoc® Blood Analysis System are used in the diagnosis and treatment of electrolyte and metabolic disorders.
Total Carbon Dioxide measurements from the epoc® Blood Analysis System are used in the diagnosis and treatment of disorders associated with changes in body acid-base balance.
Glucose measurements from the epoc® Blood Analysis System are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, and idiopathic hypoglycemia, and of pancreatic islet cell tumors.
Lactate measurements from the epoc® Blood Analysis System are used to evaluate the acidbase status and are used in the diagnosis and treatment of lactic acidosis (abnormally high acidity of the blood).
Blood Urea Nitrogen measurements from the epoc® Blood Analysis System are used in the diagnosis and treatment of certain renal and metabolic diseases.
Creatinine measurements from the epoc® Blood Analysis System are used in the diagnosis and treatment of certain renal diseases and in monitoring renal dialysis.
Hematocrit measurements from the epoc® Blood Analysis System distinguishes normal from abnormal states of blood volume, such as anemia and erythrocytosis.
Product codes (comma separated list FDA assigned to the subject device)
CHL, JGS, CEM, JFP, JPI, CGA, KHP, CGL, CGZ, CDS, JFL
Device Description
The epoc® Blood Analysis System is an in vitro diagnostic device system for the quantitative testing of blood gases, electrolytes, and metabolites in venous, arterial, and capillary whole blood samples. The epoc® System is comprised of three (3) major subsystems: epoc® Host, epoc® Reader and epoc® BGEM Test Card.
- epoc® Blood Gas Electrolyte Metabolite (BGEM) Test Card: single-use, device with . port for blood sample introduction which contains the sensor configurations for testing Sodium (Na+), Potassium (K+), Calcium (Ca++), Chloride (C)I-, pH, partial pressure of carbon dioxide (pCO2), partial pressure of oxygen (pO2), Glucose (Glu), Lactate (Lact), Creatinine (Crea), Hematocrit (Hct), Blood Urea Nitrogen (BUN) and Total Carbon Dioxide (TCO2).
- . epoc® Reader: portable, battery-powered device component that measures electrical signals from the test card sensors during blood testing and transmits this sensor data wirelessly via Bluetooth to the epoc Host.
- . epoc® Host: mobile computer-based device component for calculating test results from the sensor data sent by the epoc Reader and displaying these results on the graphical user interface. The epoc Host component can be physically connected to the Reader by a cradle component. The epoc Host also incorporates an internal laser barcode scanner for scanning patient and operator IDs. The epoc Host component currently runs on Microsoft® Windows Mobile 6.5 Operating System (OS).
Mentions image processing
Not Found
Mentions AI, DNN, or ML
Not Found
Input Imaging Modality
Not Found
Anatomical Site
Not Found
Indicated Patient Age Range
Not Found
Intended User / Care Setting
Trained medical professionals / laboratory or at the point of care.
Description of the training set, sample size, data source, and annotation protocol
Not Found
Description of the test set, sample size, data source, and annotation protocol
Not Found
Summary of Performance Studies (study type, sample size, AUC, MRMC, standalone performance, key results)
No performance data was required to evaluate the changes introduced with the alternate epoc Host component. Software, hardware and usability validation were adequate to support substantial equivalence.
Key Metrics (Sensitivity, Specificity, PPV, NPV, etc.)
Not Found
Predicate Device(s): If the device was cleared using the 510(k) pathway, identify the Predicate Device(s) K/DEN number used to claim substantial equivalence and list them here in a comma separated list exactly as they appear in the text. List the primary predicate first in the list.
K092849, K093297, K113726, K171247
Reference Device(s): Identify the Reference Device(s) K/DEN number and list them here in a comma separated list exactly as they appear in the text.
Not Found
Predetermined Change Control Plan (PCCP) - All Relevant Information for the subject device only (e.g. presence / absence, what scope was granted / cleared under the PCCP, any restrictions, etc).
Not Found
§ 862.1120 Blood gases (P
CO2 , PO2 ) and blood pH test system.(a)
Identification. A blood gases (PCO2 , PO2 ) and blood pH test system is a device intended to measure certain gases in blood, serum, plasma or pH of blood, serum, and plasma. Measurements of blood gases (PCO2 , PO2 ) and blood pH are used in the diagnosis and treatment of life-threatening acid-base disturbances.(b)
Classification. Class II.
0
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
June 5, 2020
Epocal Inc. Amy Goldberg Sr. Manager, Regulatory Affairs 2060 Walkley Road Ottawa, K1G 3P5 Canada
Re: K200107
Trade/Device Name: epoc® Blood Analysis System Regulation Number: 21 CFR 862.1120 Regulation Name: Blood Gases (PCO2, PO2) and Blood pH Test System Regulatory Class: Class II Product Code: CHL. JGS. CEM. JFP. JPI. CGA. KHP. CGL. CGZ. CDS. JFL Dated: May 1, 2020 Received: May 7, 2020
Dear Amy Goldberg:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
1
801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Marianela Perez-Torres, M.T., Ph.D. Acting Deputy Director Division of Chemistry and Toxicology Devices OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
2
510(k) Number (if known) K200107
Device Name
epoc® Blood Analvsis System
Indications for Use (Describe)
- The epoc Blood Analysis System is intended for use by trained medical professionals as an in vitro diagnostic device for the quantitative testing of samples of heparinized or un-anticoagulated arterial, venous, or capillary whole blood in the laboratory or at the point of care.
- The Blood Gas Electrolyte and Metabolite (BGEM) Test Card panel configuration includes sensors that quantitate pH, pCO2, oO2, Sodium, Potassium, Ionized Calcium, Chloride, Total Carbon Dioxide, Glucose, Lactate, Blood Urea Nitrogen, Creatinine, and Hematocrit.
- pH, pCO2, pO2 (blood gases) measurements from the epoc Blood Analysis System are used in the diagnosis and treatment of lifethreatening acid-base disturbances.
- Sodium and Potassium measurements from the epoc Blood Analysis System are treatment of diseases involving electrolyte imbalance.
- lonized Calcium measurements from the epoc Blood Analysis System are treatment of parathyroid disease, a variety of bone diseases, chronic renal disease and tetany.
- Chloride measurements from the epoc Blood Analysis System are used in the treatment of electrolyte and metabolic disorders.
- Total Carbon Dioxide measurements from the epoc Blood Analysis System are used in the diagnosis and treatment of disorders associated with changes in body acid-base balance.
- Glucose measurements from the epoc Blood Analysis System are used in the diagnosis and treatment of carbolism disorders, including diabetes mellitus and idiopathic hypodycemia, and of pancreatic islet cell tumors.
- Lactate measurements from the epoc Blood Analysis System are used to evaluate the acid-base status and are used in the diagnosis and treatment of lactic acidosis (abnormally high acidity of the blood),
- Blood Urea Nitrogen measurements from the epoc Blood Analysis System are used in the diagnosis and treatment of certain renal and metabolic diseases.
- Creatinine measurements from the epoc Blood Analysis System are used in the diagnosis and treatment diseases and in monitoring renal dialysis.
- Hematocrit measurements from the epoc Blood Analysis System are used to distinguish normal states of blood volume, such as anemia and erythrocytosis.
Type of Use (Select one or both, as applicable) |
---|
------------------------------------------------- |
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the
time to review instructions, search existing data sources, gather and maintain the data needed and complete
and review the collection of information. Send comments regarding this burden estimate or any other aspect
of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services |
---|
Food and Drug Administration |
Office of Chief Information Officer |
Paperwork Reduction Act (PRA) Staff |
PRAStaff@fda.hhs.gov |
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of
information unless it displays a currently valid OMB number."
3
Image /page/3/Picture/1 description: The image contains the logo for Siemens Healthineers. The word "SIEMENS" is written in teal, and the word "Healthineers" is written in orange below it. To the right of the words is a graphic of orange dots arranged in a circular pattern.
Section 5 - 510(k) SUMMARY
This 510(k) Summary is being submitted in accordance with the requirements of the Safe Medical Device Act of 1990 and 21 CFR 807.92.
SUBMITTER INFORMATION 1.0
| Submitter
Name: | Epocal Inc., a Siemens Healthineers Company
2060 Walkley Road
Ottawa, ON K1G 3P5
Canada |
|----------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Company
Contacts: | Amy Goldberg
Sr. Manager, Regulatory Affairs
Email: amy.goldberg@siemens-healthineers.com
Phone: +1 (781) 269-3544
Amanda Barriage
Regulatory Affairs Manager
Email: amanda.barriage@siemens-healthineers.com
Phone: +1 (613) 688 3982 ext. 2232 |
| Date
Prepared: | June 4, 2020 |
2.0 DEVICE INFORMATION
| Trade or
Proprietary
Name: | epoc® Blood Analysis System |
---|---|
Common | |
Name: | Blood Gases (PCO2, PO2) and Blood pH Test System |
3.0 REGULATORY INFORMATION
The following table highlights the classification and regulatory information for all analytes that are measured on the epoc® Blood Analysis System, which is unchanged for this Special 510(k) submission.
4
Image /page/4/Picture/1 description: The image shows the logo for Siemens Healthineers. The word "SIEMENS" is written in teal, and the word "Healthineers" is written in orange below it. To the right of the word "Healthineers" is a graphic of orange dots.
Regulatory Information- Classification | |||||
---|---|---|---|---|---|
Measured | |||||
Analytes | Classification | ||||
Name | Regulation Section | Product | |||
Code | Device | ||||
Class | Classification | ||||
Panel | |||||
pCO2, pO2, pH | Blood gases (pCO2, | ||||
pO2) and blood pH | |||||
test system | 21 CFR § 862.1120 | CHL | II | Clinical | |
Chemistry | |||||
Sodium (Na+) | Sodium test system | 21 CFR § 862.1665 | JGS | II | |
Potassium (K+) | Potassium test | ||||
system | 21 CFR § 862.1600 | CEM | II | ||
Ionized Calcium | |||||
(Ca2+) | Calcium test | ||||
system | 21 CFR § 862.1145 | JFP | II | ||
Chloride (Cl-) | Chloride test | ||||
system | 21 CFR § 862.1170 | CGZ | II | ||
Glucose (Glu) | Glucose test | ||||
system | 21 CFR § 862.1345 | CGA | II | ||
Creatinine (Crea) | Creatinine test | ||||
system | 21 CFR § 862.1225 | CGL | II | ||
Hematocrit (Hct) | Hematocrit | ||||
measuring device | 21 CFR § 862.6400 | JPI | II | ||
Lactate (Lac) | Lactic acid test | ||||
system | 21 CFR § 862.1450 | KHP | I | ||
Blood Urea | |||||
Nitrogen (BUN) | Urea nitrogen test | ||||
system | 21 CFR § 864.6150 | CDS | II | ||
Total Carbon | |||||
Dioxide (TCO2) | Bicarbonate/carbon | ||||
dioxide test system | 21 CFR § 862.1160 | JFL | II | ||
Capillary blood | |||||
collection tube | Capillary blood | ||||
collection tube | 21 CFR § 864.6150 | GIO | I | Hematology |
4.0 DEVICE DESCRIPTION
The epoc® Blood Analysis System is an in vitro diagnostic device system for the quantitative testing of blood gases, electrolytes, and metabolites in venous, arterial, and capillary whole blood samples. The epoc® System is comprised of three (3) major subsystems: epoc® Host, epoc® Reader and epoc® BGEM Test Card.
- epoc® Blood Gas Electrolyte Metabolite (BGEM) Test Card: single-use, device with . port for blood sample introduction which contains the sensor configurations for testing Sodium (Na+), Potassium (K+), Calcium (Ca++), Chloride (C)I-, pH, partial pressure of carbon dioxide (pCO2), partial pressure of oxygen (pO2), Glucose (Glu), Lactate (Lact), Creatinine (Crea), Hematocrit (Hct), Blood Urea Nitrogen (BUN) and Total Carbon Dioxide (TCO2).
5
Image /page/5/Picture/1 description: The image shows the Siemens Healthineers logo. The word "SIEMENS" is written in teal, and the word "Healthineers" is written in orange below it. To the right of the words is a cluster of orange dots.
- . epoc® Reader: portable, battery-powered device component that measures electrical signals from the test card sensors during blood testing and transmits this sensor data wirelessly via Bluetooth to the epoc Host.
- . epoc® Host: mobile computer-based device component for calculating test results from the sensor data sent by the epoc Reader and displaying these results on the graphical user interface. The epoc Host component can be physically connected to the Reader by a cradle component. The epoc Host also incorporates an internal laser barcode scanner for scanning patient and operator IDs. The epoc Host component currently runs on Microsoft® Windows Mobile 6.5 Operating System (OS).
The epoc® Blood Analysis System was previously cleared for prescription use to quantitate Sodium (Na+), Potassium (K+), Calcium (Ca++), Chloride (C)I-, pH, partial pressure of carbon dioxide (pCO2), partial pressure of oxygen (pO2), Glucose (Glu), Lactate (Lact), Creatinine (Crea), Hematocrit (Hct), Blood Urea Nitrogen (BUN) and Total Carbon Dioxide (TCO2) in arterial, venous, and capillary blood samples per K061597, K090109, K092849, K093297, K113726 and K171247.
This Special 510(k) submission has been prepared to introduce hardware and software updates to the current epoc® Host component only.
5.0 INTENDED USE
The epoc® Blood Analysis System is intended for use by trained medical professionals as an in vitro diagnostic device for the quantitative testing of samples of heparinized or unanticoagulated arterial, venous or capillary whole blood in the laboratory or at the point of care.
The Blood Gas Electrolyte and Metabolite (BGEM) Test Card panel configuration includes sensors that quantitate pH, pCO2, pO2, Sodium, Potassium, Ionized Calcium, Chloride, Total Carbon Dioxide, Glucose, Lactate, Blood Urea Nitrogen, Creatinine, and Hematocrit.
6.0 INDICATIONS FOR USE
pH, pCO2, p02 (blood gases) measurements from the epoc® Blood Analysis System are used in the diagnosis and treatment of life-threatening acid-base disturbances.
Sodium and Potassium measurements from the epoc® Blood Analysis System are used in diagnosis and treatment of diseases involving electrolyte imbalance.
lonized Calcium measurements from the epoc® Blood Analysis System is used in diagnosis and treatment of parathyroid disease, a variety of bone diseases, chronic renal disease and tetany.
Chloride measurements from the epoc® Blood Analysis System are used in the diagnosis and treatment of electrolyte and metabolic disorders.
Total Carbon Dioxide measurements from the epoc® Blood Analysis System are used in the diagnosis and treatment of disorders associated with changes in body acid-base balance.
6
Image /page/6/Picture/1 description: The image shows the Siemens Healthineers logo. The word "SIEMENS" is written in teal, and the word "Healthineers" is written in orange below it. To the right of the words is a cluster of orange dots arranged in a circular pattern.
Glucose measurements from the epoc® Blood Analysis System are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, and idiopathic hypoglycemia, and of pancreatic islet cell tumors.
Lactate measurements from the epoc® Blood Analysis System are used to evaluate the acidbase status and are used in the diagnosis and treatment of lactic acidosis (abnormally high acidity of the blood).
Blood Urea Nitrogen measurements from the epoc® Blood Analysis System are used in the diagnosis and treatment of certain renal and metabolic diseases.
Creatinine measurements from the epoc® Blood Analysis System are used in the diagnosis and treatment of certain renal diseases and in monitoring renal dialysis.
Hematocrit measurements from the epoc® Blood Analysis System distinguishes normal from abnormal states of blood volume, such as anemia and erythrocytosis.
7.0 SPECIAL CONDITIONS FOR USE STATEMENT(S):
- For prescription use only. .
- For in vitro diagnostic use only. .
- For point-of-care (POC) or clinical laboratory settings. ●
8.0 REASON FOR SUBMISSION
This Special 510(k) submission has been prepared to introduce hardware and software updates to the current epoc Host component only. The current epoc Host hardware component is a dedicated-use Personal Digital Assistant (PDA) computer-based platform that operates on Microsoft® Windows Mobile 6.5 Operating System (OS). The hardware component is changing to a new tablet-based mobile platform running on Android 9.0 OS. The modified device consists of the existing current epoc Reader and current epoc Test Card along with a new epoc Host component. There is no change to the epoc Reader and Test Card. The epoc Host Application Software has been modified to support the Android-based Operating System. The new Host component is referred to as "epoc NXS Host" throughout this submission.
The epoc® Blood Analysis System with current Host component is the Predicate Device (K061597) for this Special 510(k) submission. Both the current epoc Host and new epoc NXS Host mobile computers are intended to have equivalent functionalities and workflow. A Special 510(k) Premarket Notification is the ideal pathway for this submission based on the following:
- . There is no change to the intended use or indications for use.
- . There is no change to the fundamental scientific technology.
- There is no change to labeled performance claims. ●
- There is no change to principle of operation. ●
- There is no change to cartridge (test card) calibrator formulation and technology. ●
7
Image /page/7/Picture/1 description: The image shows the logo for Siemens Healthineers. The word "SIEMENS" is written in teal, and the word "Healthineers" is written in orange below it. To the right of the text is a cluster of orange dots of varying sizes.
9.0 SUBSTANTIAL EQUIVALENCE INFORMATION
Predicate Device Information | ||||
---|---|---|---|---|
ELEMENT | PREDICATE DEVICE | |||
Predicate Device Name | epoc® Blood Analysis System | |||
Common Name | Blood gases (PCO2, PO2) and blood pH test system | |||
510(k) Numbers | K092849, K093297, K113726, K171247 | |||
FDA Product Codes and | ||||
Analytes per 510(k) | ||||
Clearance | K092849 | K093297 | K113726 | K171247 |
CHL: pH, pCO2, pO2 | ||||
JGS: Sodium | ||||
CEM: Potassium | ||||
JFP: Ionized Calcium | ||||
JPI: Hematocrit | ||||
CGA: Glucose | KHP: Lactate | CGL: Creatinine | ||
CGZ: Chloride | CDS: Blood Urea | |||
Nitrogen | ||||
JFL: Total Carbon | ||||
Dioxide | ||||
Sample Types: | ||||
heparinized or un-anticoagulated arterial, venous or capillary whole blood | ||||
Manufacturer | Epocal Inc., a Siemens Healthineers Company |
The Substantial Equivalency Table in the following pages summarizes the similarities and differences between the epoc® Blood Analysis System with NXS Host (Modified device) and the epoc® Blood Analysis System (Predicate device).
8
Image /page/8/Picture/1 description: The image shows the logo for Siemens Healthineers. The word "SIEMENS" is in teal, and the word "Healthineers" is in orange. To the right of the words is a cluster of orange dots.
Substantial Equivalency | ||
---|---|---|
FEATURE | epoc® Blood Analysis System with NXS Host | |
(MODIFIED DEVICE) | ||
K200107 | epoc® Blood Analysis System | |
(PREDICATE DEVICE) | ||
K092849, K093297, K113726, | ||
K171247 | ||
Intended Use | The epoc® Blood Analysis System is | |
intended for use by trained medical | ||
professionals as an in vitro diagnostic device | ||
for the quantitative testing of samples of | ||
heparinized or un-anticoagulated arterial, | ||
venous or capillary whole blood in the | ||
laboratory or at the point of care. | ||
The Blood Gas Electrolyte and Metabolite | ||
(BGEM) Test Card panel configuration | ||
includes sensors that quantitate pH pCO2 pO2 Sodium Potassium Ionized Calcium Chloride Total Carbon Dioxide Glucose Lactate Blood Urea Nitrogen Creatinine Hematocrit | Same | |
FDA Product | ||
Codes and | ||
Analytes per | ||
510(k) | ||
Clearance | K092849: | |
CHL: pH, pCO2, pO2 | ||
JGS: Sodium | ||
CEM: Potassium | ||
JFP: Ionized Calcium | ||
JPI: Hematocrit | ||
CGA: Glucose |
K093297:
KHP: Lactate
K113726:
CGL: Creatinine
CGZ: Chloride
K171247:
CDS: Blood Urea Nitrogen
JFL: Total Carbon Dioxide | Same |
| Substantial Equivalency | | |
| FEATURE | epoc® Blood Analysis System with NXS Host
(MODIFIED DEVICE)
K200107 | epoc® Blood Analysis System
(PREDICATE DEVICE)
K092849, K093297, K113726,
K171247 |
| Sample Type | Heparinized or un-anticoagulated arterial,
venous or capillary whole blood | Same |
| Sample Volume | At least 92µL | Same |
| Measurement
Technology | An electrochemical multi-sensor array
integrated into a single-use test that is
interpreted by a handheld reader and
associated software | Same |
| Sensors | Contained in Test Cards | Same |
| Calibration | Calibration using in-card calibration fluid. | Same |
| Operating
System
Software | Android 9.0 Operating System | Windows 6.5 Mobile Operating
System |
| Hardware | epoc Host: Dedicated-use tablet-based
mobile platform
Image: epoc NXS Host, epoc Test Card, epoc Reader
epoc Test Card: same
epoc Reader: same | epoc Host: Dedicated-use Personal
Digital Assistant (PDA) mobile
computer-based platform
Image: epoc Host, epoc Test Card, epoc Reader
epoc Test Card: contains an array of
sensors on a sensor module mounted
in a credit-card sized fluidic housing
with a sample entry port, and a sealed
calibrator fluid reservoir.
epoc Reader: Portable, battery-
powered device for reading Test
Cards |
| Connectivity | Wireless
Bluetooth | Same |
| Substantial Equivalency | | |
| FEATURE | epoc® Blood Analysis System with NXS Host
(MODIFIED DEVICE)
K200107 | epoc® Blood Analysis System
(PREDICATE DEVICE)
K092849, K093297, K113726,
K171247 |
| Barcode
Scanner | epoc Reader: Integrated LED (1D) used to
read the Test Card Barcode
epoc Host: Integrated (1D & 2D)
Manufacturer: Zebra
Module model #: SE4710
Sensor resolution: 1280 x 800 pixels | epoc Reader: Same
epoc Host: Integrated (1D & 2D)
Manufacturer: Zebra
Module model #: SE4500DL
Sensor resolution: 752 x 480 pixels |
9
Image /page/9/Picture/1 description: The image shows the logo for Siemens Healthineers. The word "SIEMENS" is in teal, and the word "Healthineers" is in orange. To the right of the text is a cluster of orange dots.
10
Image /page/10/Picture/1 description: The image shows the Siemens Healthineers logo. The word "SIEMENS" is in teal, and the word "Healthineers" is in orange. To the right of the words is a cluster of orange dots arranged in a circular pattern.
10.0 RISK MANAGEMENT
The Risk Management was performed in compliance with EN ISO 14971:2012 and ISO 14971:2007 Medical Devices - Application of Risk Management to Medical Devices. A series of risk assessments were performed in order to identify, or mitigate any potential risks associated to the design changes for the modified device. In addition, a Failure Mode, Effects and Criticality Analysis (FMECA) is performed for any identifiable risks.
The conclusion of the risk management process is that the overall residual risk of the epoc System with the epoc NXS host is acceptable.
11.0 VERIFICATION AND VALIDATION SUMMARY
All software, hardware and usability verification and validation activities were performed in accordance to relevant standards, established plans and protocols and Design Control procedures. Testing verified all acceptance criteria were met.
No performance data was required to evaluate the changes introduced with the alternate epoc Host component. Software, hardware and usability validation were adequate to support substantial equivalence.
12.0 CYBERSECURITY INFORMATION
Cybersecurity design inputs for the epoc® Blood Analysis System with NXS Host were established as part of the Quality Management procedures for software validation and risk analysis activities. Risks were identified, assessed and controls were designed within the software, to mitigate potential security threats, system vulnerabilities, viral attacks and safeguard unauthorized access to patient data.
11
Image /page/11/Picture/1 description: The image contains the logo for Siemens Healthineers. The word "SIEMENS" is written in teal, and the word "Healthineers" is written in orange below it. To the right of the words is a graphic of orange dots in a circular pattern.
13.0 CONCLUSION
Results from verification and validation testing confirmed that the modifications made to epoc® Host hardware, operating system, and system software do not introduce new concerns for safety and effectiveness. The modified device utilizes the same intended use, indications for use, principle of operation, workflow, scientific technology and core algorithm as the predicate device. The epoc Test Card and epoc Reader remains unchanged from the predicate device. Based on these factors, the epoc® Blood Analysis System with NXS Host (Modified device) is substantially equivalent to the previously cleared predicate device, the epoc® Blood Analysis System.