Search Filters

Search Results

Found 9 results

510(k) Data Aggregation

    K Number
    K230300
    Date Cleared
    2023-10-27

    (266 days)

    Product Code
    Regulation Number
    862.1145
    Reference & Predicate Devices
    Why did this record match?
    Product Code :

    JFP

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The i-STAT CG8+ cartridge with the i-STAT 1 System is intended for use in the in vitro quantification of ionized calcium and hematocrit in arterial or venous whole blood in point of care or clinical laboratory settings.

    The i-STAT CG8+ cartridge with the i-STAT 1 System is intended for use in the in vitro quantification of hematocrit in capillary whole blood in point of care or clinical laboratory settings.

    Ionized calcium measurements are used in the diagnosis, monitoring, and treatment of conditions including, but not limited to, parathyroid disease, a variety of bone disease, chronic renal disease, tetany, and disturbances related to surgical and intensive care.

    Hematocrit measurements can aid in the determination of normal or abnormal total red cell volume status that can be associated with conditions including anemia, erythrocytosis, and blood loss related to trauma and surgery.

    Device Description

    The i-STAT CG8+ cartridge is used with the i-STAT 1 analyzer as part of the i-STAT 1 System and contains test reagents to measure hematocrit (Hct) in arterial, venous or capillary whole blood and to measure ionized calcium (iCa) in arterial and venous whole blood.

    The i-STAT 1 System is an in vitro diagnostic (IVD) medical device intended for the quantitative determination of various clinical chemistry tests contained within i-STAT cartridges using whole blood. The i-STAT 1 System consists of a portable blood analyzer (i-STAT 1 analyzer), single-use disposable test cartridges (i-STAT cartridges), liquid quality control and calibration verification materials, and accessories (i-STAT 1 Downloader/Recharger, i-STAT Electronic Simulator and i-STAT 1 Printer). The i-STAT 1 System, including the i-STAT CG8+ cartridge, is designed for use by trained medical professionals in point of care or clinical laboratory settings and is for prescription use only.

    The i-STAT CG8+ cartridge contains the required sensors, a fluid pack (calibrant pouch), a sample entry well and closure, fluid channels, waste chamber, and the necessary mechanical features for controlled fluid movement within cartridge. The i-STAT cartridge format allows all the tests in the cartridge to be performed simultaneously. All the test steps and fluid movement occur within the i-STAT CG8+ cartridges require two to three drops of whole blood which are applied to the cartridge using a transfer device by the trained user before the cartridge is placed within the analyzer.

    The i-STAT 1 analyzer is a handheld, in vitro diagnostic analytical device designed to run only i-STAT test cartridges. The instrument interacts with the i-STAT CG8+ cartridge to move fluid across the sensors and generate a quantitative result (within approximately 2 minutes).

    AI/ML Overview

    The provided text describes the analytical performance studies for the i-STAT CG8+ cartridge with the i-STAT 1 System, a device for in vitro quantification of ionized calcium (iCa) and hematocrit (Hct). This submission sought to demonstrate substantial equivalence to a predicate device (i-STAT CHEM8+ cartridge with the i-STAT 1 System, K191360).

    Here's an analysis of the acceptance criteria and study details based on the provided information:

    1. A table of acceptance criteria and the reported device performance

    The document does not explicitly state pre-defined acceptance criteria (e.g., target ranges for precision, linearity, or bias). Instead, it presents the results of various analytical performance studies. The conclusion states that the studies demonstrate substantial equivalence to the predicate device, implying that the reported performance metrics met the internal or regulatory thresholds for such a claim.

    However, we can infer performance targets for some aspects by comparing the reported results to general expectations for such devices or by looking at the predicate device's characteristics mentioned (though the predicate's detailed performance is not provided in this document).

    For the purpose of this request, I will present the reported performance values. The "Acceptance Criteria" column will reflect the general expectation of "demonstrating substantial equivalence" or meeting regulatory standards for the tested parameter, as specific numerical acceptance criteria are not provided in the document.

    Test ParameterAcceptance Criteria (Inferred from "Substantial Equivalence")Reported Device Performance (i-STAT CG8+ cartridge with i-STAT 1 System)
    Precision (Aqueous Materials) - 20 DaysPerformance demonstrating substantial equivalence to predicate.iCa (mmol/L): Overall Within-Laboratory SD typically
    Ask a Question

    Ask a specific question about this device

    K Number
    K191360
    Date Cleared
    2020-02-14

    (269 days)

    Product Code
    Regulation Number
    862.1145
    Reference & Predicate Devices
    Why did this record match?
    Product Code :

    JFP

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The i-STAT CHEM8+ cartridge with the i-STAT 1 System is intended for use in the in vitro quantification of ionized calcium in arterial or venous whole blood in point of care or clinical laboratory settings.

    Ionized calcium measurements are used in the diagnosis and treatment of parathyroid disease, a variety of bone diseases, chronic renal disease and tetany.

    Device Description

    The i-STAT CHEM8+ test cartridge contains test reagents to analyze whole blood at the point of care or in the clinical laboratory for ionized calcium (iCa). The test is contained in a single-use, disposable cartridge. Cartridges require two to three drops of whole blood which are typically applied to the cartridge using a transfer device.

    The i-STAT 1 Analyzer is a handheld, in vitro diagnostic analytical device designed to run only i-STAT test cartridges. The instrument interacts with the cartridge to move fluid across the sensors and generate a quantitative result (within approximately 2 minutes).

    The i-STAT 1 System is comprised of the i-STAT 1 analyzer, the i-STAT test cartridges and accessories (i-STAT 1 Downloader/Recharger, electronic simulator and portable printer). The system is designed for use by trained medical professionals at the patient point of care or in the clinical laboratory and is for prescription use only.

    AI/ML Overview

    The provided text is a 510(k) summary for the i-STAT CHEM8+ cartridge with the i-STAT 1 System, specifically for the ionized calcium (iCa) test. It details various performance characteristic studies to demonstrate substantial equivalence to a predicate device.

    Here's an analysis of the acceptance criteria and the study that proves the device meets them, based only on the provided text. Many elements cannot be fully answered as they are not explicitly stated in this specific document, which is a summary.

    Acceptance Criteria and Reported Device Performance

    The acceptance criteria are generally implied by the successful outcomes of the performance studies. For a clearer table, we derive the 'acceptance criteria' from the observed 'performance' meeting regulatory expectations for substantial equivalence.

    Performance CharacteristicAcceptance Criteria (Implied)Reported Device Performance
    Precision (Aqueous Materials)Low %CV values, indicating reproducibility across various levels and within/between runs/days.iCa (mmol/L):
    • CV L1: Total CV 0.6%
    • CV L2/Control L1: Total CV 0.7%
    • CV L3/Control L2: Total CV 0.5%
    • CV L4/Control L3: Total CV 0.5%
    • CV L5: Total CV 1.5% |
      | Precision (Whole Blood - Venous) | Low %CV values across different sample ranges. | Venous Whole Blood:
    • 0.25-0.75 mmol/L: 2.2% CV
    • 0.75-1.2 mmol/L: 1.6% CV
    • 1.2-1.5 mmol/L: 1.0% CV
    • 1.5-2.5 mmol/L: 0.9% CV |
      | Precision (Whole Blood - Arterial) | Low %CV values across different sample ranges. | Arterial Whole Blood:
    • 0.25-0.75 mmol/L: 0.9% CV
    • 0.75-1.2 mmol/L: 3.0% CV
    • 1.2-1.5 mmol/L: 0.8% CV
    • 1.5-2.5 mmol/L: 0.5% CV |
      | Linearity | Demonstrate linearity over the specified reportable range with a good fit. | Linearity demonstrated over 0.25 - 2.50 mmol/L reportable range. Regression: Slope 0.9738, Intercept 0.360, R2 0.999. "Absolute degree of nonlinearity results met the acceptance criteria." |
      | Limit of Quantitation (LoQ) | LoQ should be at or below the lower limit of the reportable range. | LoQ determined to be 0.21 mmol/L, which is below the lower limit of the reportable range (0.25 mmol/L). |
      | Limit of Blank (LoB) | LoB should be very low, indicating minimal signal in the absence of analyte. | LoB determined to be 0.14 mmol/L. |
      | Limit of Detection (LoD) | LoD should be very low, indicating the lowest detectable concentration. | LoD determined to be 0.15 mmol/L. |
      | Interference | No significant interference from common substances or specified interfering substances below allowed error (Ea). | Several substances tested; Interference identified for: Leflunomide, Lithium Lactate, Lithium Thiocyanate, Sodium Thiosulfate, Teriflunomide. For these, details on the specific effect (e.g., "Decreased results ≥ 0.4 mmol/L") are provided. Other substances showed no interference at tested concentrations. |
      | Method Comparison with Predicate Device | Strong correlation (r value close to 1) and a slope close to 1 with an intercept close to 0 when compared to the predicate device. | iCa: N=250, Slope 1.00, Intercept -0.02, r 0.99. |

    Study Details:

    1. A table of acceptance criteria and the reported device performance: See table above.

    2. Sample size used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective):

      • Precision (Aqueous): 80-81 measurements per level (N values in Table 1). Data provenance is not specified regarding country, but it was conducted "at one site." The study design (20-day multi-day precision testing) implies a prospective collection of data for this specific study.
      • Precision (Whole Blood): 241 samples (132 venous, 109 arterial). Data collected across "three point of care sites." Data provenance regarding country or retrospective/prospective nature is not explicitly mentioned, but the context of a validation study typically implies prospective collection for this analysis.
      • Linearity: Not explicitly stated but implies multiple whole blood samples across the range (0.22 - 2.81 mmol/L).
      • LoQ/LoB/LoD: Not explicitly stated how many individual sample measurements were involved, but mentions "whole blood that was altered" and conducted over "four (4) days using two (2) cartridge lots."
      • Interference: Not explicitly stated, but each substance was evaluated by comparing a control sample to a spiked sample.
      • Method Comparison: 250 samples (N value in Table 6). Venous and arterial blood specimens were evaluated. Data provenance regarding country or retrospective/prospective nature is not explicitly mentioned.
    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g., radiologist with 10 years of experience):

      • Not applicable for this type of device. This is an in vitro diagnostic (IVD) device for quantitative measurement. Ground truth is established by reference methods or direct measurement comparisons, not expert subjective interpretation (like radiologists for imaging). The predicate device (Epocal EPOC Blood Analysis System) served as the comparative method for the method comparison study.
    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

      • Not applicable. As an IVD device measuring a quantitative analyte, there is no subjective interpretation requiring adjudication methods.
    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

      • Not applicable. This is not an AI-assisted diagnostic imaging device that involves human reader interpretation. It's an automated quantitative IVD system.
    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

      • Yes, this is a standalone device. The performance characteristics (precision, linearity, LoQ, LoD, LoB, interference) and the method comparison against a predicate device evaluate the device itself without direct human interpretation influencing the measurement result for iCa. The system automatically performs quality checks and suppresses results if specifications are not met.
    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

      • For precision studies, the 'ground truth' is the mean value derived from repeated measurements of stable materials or samples, reflecting the inherent variability.
      • For linearity studies, ground truth is established by accurately prepared samples with known concentrations.
      • For LoQ/LoB/LoD studies, ground truth is based on samples expertly prepared to very low or zero analyte concentrations.
      • For interference studies, ground truth is the expected value of the sample without the interfering substance.
      • For method comparison, the predicate device (Epocal EPOC Blood Analysis System) serves as the "ground truth" or reference for comparison.
    8. The sample size for the training set:

      • Not applicable. This document describes a traditional IVD device validation, not a machine learning or AI model development that requires a "training set" in the conventional sense of AI. The device's underlying technology relies on ion-selective electrodes, not a learned model from a large dataset.
    9. How the ground truth for the training set was established:

      • Not applicable. See point 8.
    Ask a Question

    Ask a specific question about this device

    K Number
    K191626
    Date Cleared
    2019-08-16

    (59 days)

    Product Code
    Regulation Number
    862.1145
    Reference & Predicate Devices
    Why did this record match?
    Product Code :

    JFP

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Stat Profile Prime ES Comp Plus Analyzer System is intended for in vitro diagnostic use by health care professionals in clinical laboratory settings for the quantitative determination of pH, Hematocrit, Ionized Calcium and Ionized Magnesium in heparinized venous whole blood, and pH. Ionized Calcium and Ionized Magnesium in plasma and serum.

    Ionized Calcium (iCa) measurements are used in the diagnosis and treatment of parathyroid disease, a variety ofbone diseases, chronic renal disease and tetany (intermittent muscular contractions or spasms).

    Ionized Magnesium (iMg) measurements are used in the diagnosis and treatment of hypomagnesemia (abnormally low levels of magnesium) and hypermagnesemia (abnormally high levels of magnesium).

    pH measurements are used in the diagnosis and treatment of life-threatening acid-base disturbances.

    Hematocrit measurements of the packed red blood cell volume are used to distinguish normal from abnormal states, such as anemia and erythrocytosis.

    Device Description

    The Stat Profile Prime ES Comp Plus Analyzer is a small, low cost blood electrolyte analyzer. It consists of the analyzer, sensor cartridges, and thermal paper for an onboard printer. Optionally, it provides for reading of barcode labels (such as operator badges and data sheets).

    The Stat Profile Prime ES Comp Plus Analyzer has an enhanced test menu and multiple quality control options. External Control Solutions (ampules) shall be offered, as well as an on-board Quality Management System (QMS), an electronic monitoring approach that insures the analyzer is working properly.

    The Stat Profile Prime ES Comp Plus Analyzer can accommodate either of two sensor cards in the sensor card housing. The analyzer will determine the test configuration of the system by detecting which sensor card is installed.

    The two options for the sensor card are:

    • Sensor Card 1 (Basic Electrolyte Panel plus Hct) shall enable and report the following listed analytes: Hct, Na, K, Cl
    • Sensor Card 2(Full Electrolyte Panel plus pH & Hct) shall enable and report the following listed analytes: pH, Hct, Na, K, Cl, iCa, iMg

    As with the predicates, the Stat Profile Prime ES Comp Plus Analyzer is microprocessor-based and incorporates ion selective electrode technology to measure pH, ionized calcium, ionized magnesium.

    The Prime ES Comp Plus can be configured with an optional sample tray, which allows the user to run up to 10 consecutive samples may be any combination of Serum/Plasma or control solutions. Whole Blood samples may only be run in STAT Mode (not tray mode).

    Calibration standards are provided in sealed pouches within a calibrator pack. Liquid quality control materials are available as external ampules. Sampling and calibration are fully automated.

    The Stat Profile Prime ES Comp Plus Analyzer accepts lithium heparinized whole blood sample from syringes, open tubes, and small cups. The minimum sample sizes for analysis is 100 µL.

    AI/ML Overview

    The provided text describes a 510(k) premarket notification for the "Stat Profile Prime ES Comp Plus Analyzer System," an in vitro diagnostic device for measuring various blood parameters. It does NOT describe an AI/ML-based device. Therefore, the requested information regarding AI/ML acceptance criteria, training/test sets, expert ground truth establishment, MRMC studies, etc., cannot be extracted from this document as it is not relevant to the described device.

    The document focuses on demonstrating substantial equivalence to a predicate device (Nova Stat Profile pHOx Ultra Blood Gas Analyzer) through performance testing typically required for in vitro diagnostic devices, such as method comparison, precision, linearity, specificity/interference, detection limit, and shelf life stability.

    Here's a breakdown of what can be extracted, addressing the closest relevant points, and explicitly stating what cannot be found:

    1. A table of acceptance criteria and the reported device performance

    The document does not explicitly present a table of acceptance criteria paired with the specific numerical reported device performance for each analyte (e.g., pH, iCa, iMg, Hct). Instead, it broadly states that "The results of the testing confirmed that the performance of the Stat Profile Prime ES Comp Plus Analyzer System is substantially equivalent to that of the Nova Stat Profile pHOx Ultra Blood Gas Analyzer (predicate device)." The study types performed are listed as:

    • Method Comparison Studies
    • Precision/Reproducibility Studies
    • Run to Run Precision
    • Linearity Testing
    • Specificity / Interference Testing
    • Detection Limit
    • Shelf Life Stability Testing

    Without the detailed test reports attached to this summary, the specific acceptance criteria thresholds and the measured performance values for each of these studies (e.g., specific bias ranges for method comparison, CV% limits for precision) are not available in this document.

    2. Sample size used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective)

    • Sample Size: The document does not specify the exact sample sizes used for each performance test (method comparison, precision, linearity, etc.).
    • Data Provenance: Not explicitly stated, but clinical laboratory settings are mentioned for intended use, implying laboratory-conducted studies. No information on country of origin or whether the data was retrospective or prospective is provided.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g., radiologist with 10 years of experience)

    This question is not applicable. This device is an in vitro diagnostic analyzer, not an AI/ML system requiring human expert interpretation as ground truth for imaging or similar applications. The "ground truth" for the performance studies would be reference methods (for method comparison), known concentrations/values (for linearity/detection limit), or repeated measurements (for precision), all performed using established laboratory techniques, not expert consensus.

    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set

    Not applicable. This is not a human interpretation study.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This is not an AI-assisted diagnostic device for human readers.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. This is a standalone analytical device, but the term "standalone performance" in the context of your prompt usually refers to the accuracy of an AI algorithm on its own. For this device, all performance listed (method comparison, precision, etc.) is its standalone performance. The device operates without human-in-the-loop interpretation assistance, but rather as an automated analyzer whose results are then used by healthcare professionals.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    The "ground truth" or reference values for this type of device would typically be:

    • For Method Comparison: Results obtained from a well-established, validated reference method or another legally marketed device known for its accuracy.
    • For Linearity: Samples with precisely known concentrations prepared gravimetrically or volumetrically.
    • For Precision/Reproducibility: Statistical analysis of repeated measurements on stable samples.

    The document does not specify the exact reference instruments or methods used, but this is the general approach for IVD devices.

    8. The sample size for the training set

    Not applicable. This is not an AI/ML device that requires a "training set." Its operating principles are based on ion-selective electrodes and impedance sensors, as described.

    9. How the ground truth for the training set was established

    Not applicable. See point 8.

    Ask a Question

    Ask a specific question about this device

    K Number
    K152327
    Manufacturer
    Date Cleared
    2016-12-01

    (471 days)

    Product Code
    Regulation Number
    862.1145
    Reference & Predicate Devices
    N/A
    Why did this record match?
    Product Code :

    JFP

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The EasyLyte Na/K/C/Ca Analyzer is intended for in-vitro diagnostic testing of sodium (Na), Potassium (K), Chloride (CL), and lonized Calcium (Ca++) without pH correction.

    The ionized calcium test on the EasyLyte Na/K/C//Ca analyzer is intended for the quantitative determination of calcium ions (Ca++) in human serum, plasma and whole Blood in clinical laboratories. Calcium measurements are used in the diagnosis and treatment of parathyroid disease, a variety of bone disease and tetany finternittent muscular contractions or spasms).

    Device Description

    Not Found

    AI/ML Overview

    The provided text is a 510(k) premarket notification letter from the FDA to MEDICA CORPORATION for their EasyLyte Na/K/Cl/Ca Analyzer. This document primarily focuses on the regulatory aspects of the device, stating that the FDA has determined it is "substantially equivalent" to legally marketed predicate devices.

    However, the letter does not contain the detailed study information required to answer your specific questions regarding acceptance criteria, device performance, sample sizes, ground truth establishment, expert qualifications, or MRMC studies. This type of information would typically be found in the device's 510(k) submission itself, which is a much larger and more technical document than the FDA's regulatory response letter.

    Therefore,Based on the provided document, I cannot accurately describe the acceptance criteria and the study that proves the device meets the acceptance criteria. The document is an FDA 510(k) clearance letter, which confirms substantial equivalence but does not detail the specific performance studies, acceptance criteria, or their results.

    To answer your questions, I would need access to the actual 510(k) submission for K152327, which contains the detailed technical and clinical performance data.

    Ask a Question

    Ask a specific question about this device

    K Number
    K984299
    Date Cleared
    1998-12-11

    (10 days)

    Product Code
    Regulation Number
    862.1145
    Reference & Predicate Devices
    Why did this record match?
    Product Code :

    JFP

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The AVL OPTI Critical Care Analyzer is intended to be used for the measurement of pH, PCO2, PO2, sodium, potassium, ionized calcium and chloride, total hemoglobin content and oxygen saturation in samples of whole blood, serum, plasma or dialysate in either a traditional blood gas, clinical laboratory setting or point-of-care locations by personnel minimally qualified to perform and report these results.

    The AVL OPTI Critical Care Analyzer intended to be used for the measurement of pH, PCO2, PO2, ctHb, SO2, Na', K', Cl' and Ca* in whole blood, serum, plasma and aqueous dialysate solutions as appropriate by minimally trained personnel qualified to perform and to report these values in either a traditional blood gas, clinical laboratory setting or point-of-care locations by personnel minimally qualified to perform and report these results.

    For Professional Use Only
    For In Vitro Diagnostic Use

    Device Description

    The AVL OPTI Critical Care Analyzer is a small [4.9 x 14.3 x 9.8 in. 10 lbs], instrument using optical fluorescence for the measurement of pH, PCO2, PO2, sodium, potassium, ionized calcium and choride of whole blood, plasma, serum or dialysate as appropriate. In addition, it uses optical reflectance for the measurement of total hemoglobin and oxygen saturation. A disposable, single-use cassette containing six optical fluorescence sensors is packaged in a sealed foil pouch which bears a bar-coded label with calibration and identification information. The OPTI can perform up to 8 tests on a single sample, determined by the type of disposable cassette used.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and study information for the AVL OPTI Critical Care Analyzer, based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not explicitly state numerical acceptance criteria for precision, linearity, or interference. Instead, it describes compliance through comparative studies and general statements of no significant difference. Here's a summary of the reported performance:

    ParameterAcceptance Criteria (explicitly stated)Reported Device Performance
    PrecisionNot explicitly stated"Typical Within-Run (Swr), Between-Day (Sdd) and Total (ST) precision were determined from two runs per day with 2 replicates per run for 20 days on two AVL OPTI instruments using samples: serum, reduced bovine hemoglobin solution and three levels of aqueous quality control solution." (No specific numerical results provided)
    LinearityNot explicitly stated"Wherever possible, linearity for the OPTI measurement has been established against reference materials or methods. Linearity for pH of whole blood is established by measurement of blood specimens which were tonometered to various CO2 values, and measured on an AVL 995 pH/Blood Gas Analyzer standardized to N.I.S.T. traceable pH buffers, and on OPTI Critical Care Analyzers." (No specific numerical results provided)
    InterferencesNot explicitly stated (implied compliance with guidelines)"Representative samples taken the published guidelines for evaluation of interference substances and identified from literature were evaluated." (No specific numerical results or identified interferences provided)
    Clinical CorrelationNo significant difference (P
    Ask a Question

    Ask a specific question about this device

    K Number
    K972673
    Date Cleared
    1997-08-20

    (35 days)

    Product Code
    Regulation Number
    862.1145
    Reference & Predicate Devices
    Why did this record match?
    Product Code :

    JFP

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The AVL 9181 Electrolyte Analyzer is intended to be used for the measurement of sodium, potassium, chloride, ionized calcium and lithium in whole blood, serum or plasma, urine, dialysate solutions, or QC materials as appropriate by minimally trained personnel qualified to perform and to report these values in a clinical laboratory setting. These analytes are commonly used in the diagnosis and management of patients with a broad range of renal, metabolic and cardiovascular disorders and, as such, have come to be among those which are considered by the American Association of Clinical Chemistry to have the potential of being life threatening if left uncontrolled.

    Device Description

    The AVL 9181 Electrolyte Analyzer is a microprocessor-based instrument using ionselective electrodes for the measurement of sodium, chloride, ionized calcium and lithium. The user is able to select any one of the measurement modes: whole blood, serum, urine, standard, OC material, acetate or bicarbonate dialysate, depending on the sample type to be analyzed. The analyzer automatically processes the sample through the necessary steps, then prints and displays the results. The AVL 9181 is an improved design of our existing 9180 Electrolyte Analyzer [K961458] with the addition of an Autosampler to allow automatic sampling of up to 18 sample cups. The 9181 is exactly equivalent to the AVL 9180 with the exception of this single, additional feature. The manual needle mechanism of the 9180 was replaced with the automated needle mechanism from the AVL 988-4. The electronic circuitry to control the needle mechanism are incorporated into the 9180 display board. The sampler is taken from the 988-4 without modification except that the connector is changed to a smaller type. The 9180 main board already incorporated the controller for the sampler. The 9180 software was modified to add the automatic sampling feature. Calibration and measurement sequences are taken without alteration from the 9180 (sample volume, timing and algorithms). Once the sample probe is positioned in the sample cup, aspiration, measurement, wash and recalibration sequences are identical to the 9180.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and study information for the AVL 9181 Electrolyte Analyzer, based on the provided text:

    AVL 9181 Electrolyte Analyzer: Acceptance Criteria and Study Details

    The provided document describes the AVL 9181 Electrolyte Analyzer as an improved design of the existing AVL 9180 Electrolyte Analyzer [K961458], with the primary addition of an Autosampler. The studies presented focus on demonstrating that this new feature, and the overall device, performs equivalently to the predicate devices and existing manual methods.


    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not explicitly present a table of predetermined "acceptance criteria" in the format of specific thresholds for accuracy, precision, etc. Instead, it describes performance characteristics that were evaluated and concluded to be comparable to predicate devices and manual methods. The general acceptance criterion appears to be "no significant difference" (p wr), Between-Day (Sdd), and Total (ST) Precision determined. Tests performed in manual and automated modes. Conclusion: "no significant difference in performance exists between the 9181 in automated and manual measurement modes; and between the 9181 and its predecessor, the AVL 9180." |
    | Linearity (N.I.S.T. SRM 956a) | Strong correlation (R-value close to 1) and low Syx. | Sodium: Slope 1.0134, Intercept -2.5307, Correlation Coefficient 0.99988, Syx 0.2709
    Potassium: Slope 1.0133, Intercept -0.0230, Correlation Coefficient 0.99996, Syx 0.0166
    Lithium: Slope 0.9717, Intercept 0.0482, Correlation Coefficient 0.99998, Sy
    x 0.0067
    (Note: This demonstrates linearity, but a direct comparison to an acceptance threshold isn't stated.) |
    | Linearity (Serum) | Equivalent linearity to predicate methods; no significant difference (p

    Ask a Question

    Ask a specific question about this device

    K Number
    K963694
    Manufacturer
    Date Cleared
    1996-11-27

    (72 days)

    Product Code
    Regulation Number
    862.1145
    Reference & Predicate Devices
    N/A
    Why did this record match?
    Product Code :

    JFP

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use
    Device Description
    AI/ML Overview
    Ask a Question

    Ask a specific question about this device

    K Number
    K961458
    Date Cleared
    1996-06-12

    (58 days)

    Product Code
    Regulation Number
    862.1145
    Reference & Predicate Devices
    Why did this record match?
    Product Code :

    JFP

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The AVL 9180 Electrolyte Analyzer is intended to be used for the measurement of sodium, potassium, chloride, ionized calcium and lithium in whole blood, serum or plasma, urine, dialysate solutions, or QC materials as appropriate by minimally trained personnel qualified to perform and to report these values in a clinical laboratory setting. These analytes are commonly used in the diagnosis and management of patients with a broad range of renal, metabolic and cardiovascular disorders and, as such, have come to be among those which are considered by the American Association of Clinical Chemistry to have the potential of being life threatening if left uncontrolled.

    Device Description

    The AVL 9180 Electrolyte Analyzer is a microprocessor-based instrument using ionselective electrodes for the measurement of sodium, chloride, ionized calcium and lithium. The user is able to select any one of the measurement modes: whole blood, serum, urine, standard, OC material, acetate or bicarbonate dialysate, depending on the sample type to be analyzed. The analyzer automatically processes the sample through the necessary steps, then prints and displays the results.

    In the blood, serum and QC measuring modes, the results for sodium and potassium are reported by default as flame photometry equivalent; chloride, ionized calcium and lithium are reported as ISE direct potentiometric values. The urine mode allows for the measurement of prediluted urine samples for sodium, potassium and chloride. The acetate, bicarbonate and standard modes allow for the measurement of aqueous standards and dialysate solutions and reports as ISE direct potentiometric values.

    AI/ML Overview

    The provided text describes the AVL 9180 Electrolyte Analyzer, an ion-specific electrolyte analyzer for measuring sodium, potassium, chloride, ionized calcium, and lithium. The submission focuses on demonstrating the substantial equivalence and performance of the new device compared to existing predicate devices.

    Here's an analysis of the acceptance criteria and study information:

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria are not explicitly stated as numerical targets in the provided text (e.g., "correlation coefficient > 0.95"). Instead, the document implies that the device meets acceptance criteria if its performance is comparable to predicate devices and falls within "manufacturers claims and expectations." The core of the performance evaluation relies on correlation coefficients, slopes, and intercepts from comparative studies, demonstrating the agreement between the AVL 9180 and established methods.

    The tables presented in the document are the reported device performance. I will present a summary table outlining the key performance metrics from the various comparative studies, grouped by parameter and comparator.

    ParameterPerformance Metric (Type of Study)Reported Value / RangeIndication of Acceptance
    Sodium (Aqueous Solutions)Correlation Coefficient (Linearity)0.99995Excellent linearity
    Slope (Linearity)0.99993Close to ideal slope of 1
    Intercept (Linearity)0.0128Close to ideal intercept of 0
    Potassium (Aqueous Solutions)Correlation Coefficient (Linearity)0.99919Excellent linearity
    Slope (Linearity)0.99838Close to ideal slope of 1
    Intercept (Linearity)0.0119Close to ideal intercept of 0
    Chloride (Aqueous Solutions)Correlation Coefficient (Linearity)0.99994Excellent linearity
    Slope (Linearity)0.97556Close to ideal slope of 1
    Intercept (Linearity)-0.1775Close to ideal intercept of 0
    Ionized Calcium (Aqueous Solutions)Correlation Coefficient (Linearity)0.99980Excellent linearity
    Slope (Linearity)1.01552Close to ideal slope of 1
    Intercept (Linearity)-0.0078Close to ideal intercept of 0
    Lithium (Aqueous Solutions)Correlation Coefficient (Linearity)0.99985Excellent linearity
    Slope (Linearity)0.99850Close to ideal slope of 1
    Intercept (Linearity)0.0087Close to ideal intercept of 0
    Sodium (Serum vs. Flame Absorbance)Correlation Coefficient0.9908Very good correlation
    Slope0.9617Close to ideal slope of 1
    Potassium (Serum vs. Flame Absorbance)Correlation Coefficient0.9991Excellent correlation
    Slope1.0249Close to ideal slope of 1
    Lithium (Serum vs. Flame Absorbance)Correlation Coefficient0.9822Very good correlation
    Slope0.9803Close to ideal slope of 1
    Sodium (Serum vs. ISE Direct Potentiometry - AVL 983)Correlation Coefficient0.9992Excellent correlation
    Slope0.9895Close to ideal slope of 1
    Potassium (Serum vs. ISE Direct Potentiometry - AVL 983)Correlation Coefficient0.9996Excellent correlation
    Slope1.0223Close to ideal slope of 1
    Chloride (Serum vs. ISE Direct Potentiometry - AVL 983)Correlation Coefficient0.9995Excellent correlation
    Slope0.9631Close to ideal slope of 1
    Ionized Calcium (Serum vs. ISE Direct Potentiometry - AVL 984)Correlation Coefficient0.9960Excellent correlation
    Slope0.8898Good correlation, slightly lower slope
    Lithium (Serum vs. ISE Direct Potentiometry - AVL 985)Correlation Coefficient0.9985Excellent correlation
    Slope0.9923Close to ideal slope of 1
    Sodium (Serum vs. Flame Correlation - AVL 9130, 9140)Correlation Coefficient0.9856Very good correlation
    Potassium (Serum vs. Flame Correlation - AVL 9130, 9140)Correlation Coefficient0.9994Excellent correlation
    Chloride (Serum vs. Flame Correlation - AVL 9130, 9140)Correlation Coefficient0.9989Excellent correlation
    Ionized Calcium (Serum vs. Flame Correlation - AVL 9130, 9140)Correlation Coefficient0.9954Excellent correlation
    Sodium (Urine vs. direct ISE - AVL 983)Correlation Coefficient0.9973Excellent correlation
    Slope1.0173Close to ideal slope of 1
    Potassium (Urine vs. direct ISE - AVL 983)Correlation Coefficient0.9976Excellent correlation
    Slope1.0312Close to ideal slope of 1
    Chloride (Urine vs. direct ISE - AVL 983)Correlation Coefficient0.9972Excellent correlation
    Slope0.9817Close to ideal slope of 1
    Sodium (Urine vs. Flame Absorbance - IL 943)Correlation Coefficient0.9901Very good correlation
    Slope0.9173Good correlation, slightly lower slope
    Potassium (Urine vs. Flame Absorbance - IL 943)Correlation Coefficient0.9976Excellent correlation
    Slope1.0312Close to ideal slope of 1
    Chloride (Urine vs. Chloridometry - Labconco Digital)Correlation Coefficient0.9972Excellent correlation
    Slope0.9817Close to ideal slope of 1
    Sodium (Clinical vs. KODAK Ektachem XR700)Correlation Coefficient0.9495Good correlation
    Potassium (Clinical vs. KODAK Ektachem XR700)Correlation Coefficient0.9873Very good correlation
    Chloride (Clinical vs. KODAK Ektachem XR700)Correlation Coefficient0.9762Very good correlation
    Clinical Field Tests (example data presented)No difference in mean valuesP
    Ask a Question

    Ask a specific question about this device

    K Number
    K960120
    Date Cleared
    1996-03-11

    (60 days)

    Product Code
    Regulation Number
    862.1145
    Reference & Predicate Devices
    N/A
    Why did this record match?
    Product Code :

    JFP

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use
    Device Description
    AI/ML Overview
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1