Search Filters

Search Results

Found 69 results

510(k) Data Aggregation

    K Number
    K232786
    Date Cleared
    2024-05-16

    (248 days)

    Product Code
    Regulation Number
    890.5850
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    PA&PR series:
    The PA&PR series are a neuromuscular electrical stimulator indical supervision for adjunctive therapy in the treatment of medical diseases and conditions.
    As a powered muscle stimulator, PA&PR series are indicated for the following conditions:

    • · Relaxation of muscle spasms,
    • · Prevention or retardation of disuse atrophy,
    • · Increasing local blood circulation,
    • · Immediate post-surgical stimulation of calf muscles to prevent venous thrombosis,
    • · Maintaining or increasing range of motion,
    • · Muscle re-education.
      Environments of Use: Clinics, hospital. Patient population: Adults
      S&Q series:
      The S&O series are a neuromuscular electrical stimulator indical supervision for adjunctive therapy in the treatment of medical diseases and conditions.
      As a powered muscle stimulator, S&Q series are indicated for the following conditions:
    • · Relaxation of muscle spasms,
    • · Prevention or retardation of disuse atrophy,
    • · Increasing local blood circulation,
    • · Immediate post-surgical stimulation of calf muscles to prevent venous thrombosis,
    • · Maintaining or increasing range of motion,
    • · Muscle re-education.
      As a transcutaneous electrical nerve stimulator for pain relief, S&Q series are indicated for the following conditions:
      · Symptomatic relief and management of chronic (long-term), intractable pain,
    • · Adjunctive treatment in the management of post-surgical pain and post traumatic acute pain.
      Environments of Use: Clinics, hospital and home environments.
      Patient population: Adults
    Device Description

    Subject device have four series, which are PA series, PR series, Q series and S series. These four series are all electrical stimulating devices.

    AI/ML Overview

    This document is an FDA 510(k) clearance letter for the "Stimulation System" manufactured by Edan Instruments, Inc. It asserts substantial equivalence to a predicate device, STIWELL med4 (K080950).

    The provided text does not contain acceptance criteria for an AI-powered device, nor does it describe a study proving such a device meets acceptance criteria.

    Instead, this document is a regulatory submission for a medical device (neuromuscular and transcutaneous electrical nerve stimulators) that appears to be hardware-based, with "software/firmware/microprocessor control" mentioned generally, but not as the primary component being evaluated for AI performance. The document focuses heavily on bench testing (electrical safety, EMC, performance to standards) and comparison to a predicate device's physical and electrical specifications, not on AI performance metrics like accuracy, sensitivity, or specificity.

    Therefore, I cannot fulfill your request for:

    • A table of acceptance criteria and reported device performance for an AI device.
    • Sample sizes, data provenance, ground truth establishment, or expert involvement related to AI performance.
    • MRMC studies or standalone AI performance.

    The document's "Performance Data" section explicitly states: "Clinical data: Not applicable." This further confirms that no clinical studies (which would be necessary to prove performance of an AI system in a clinical context) were conducted or presented for this submission.

    In summary, the provided text does not describe an AI medical device or its performance criteria/study.

    Ask a Question

    Ask a specific question about this device

    K Number
    K233038
    Date Cleared
    2024-03-08

    (165 days)

    Product Code
    Regulation Number
    870.2300
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The device is intended to be used for measuring, storing, and to generate prompts for, multiple physiological parameters of adults and pediatrics (including neonates). The device is intended for use by trained healthcare professionals in hospital environments. Parameters include: NIBP, SpO2, PR (pulse rate), TEMP. The F3000 Quick Temp module is not intended for neonates. The device is not intended for MRI environments.

    Device Description

    The iM3s series vital signs monitors including iM3s\iM3As\iM3Bs\ iHM3s are intended to be used for measuring, storing, reviewing of, and generating prompts for multiple physiological parameters of adults and pediatrics.

    AI/ML Overview

    The provided documentation describes the FDA 510(k) clearance for the Edan Instruments Vital Signs monitor, Model: iM3s, iM3As, iM3Bs, iHM3s. However, the document primarily focuses on demonstrating substantial equivalence to a predicate device through non-clinical performance testing and software verification/validation, rather than detailing a clinical study with specific acceptance criteria for AI/algorithm performance.

    Therefore, many of the requested details regarding acceptance criteria, sample size, expert involvement, and ground truth establishment for a study proving device performance in an AI/algorithm context are not applicable or not provided in this specific FDA clearance document. The device is a vital signs monitor, not explicitly described as incorporating AI for diagnosis or interpretation that would necessitate an MRMC study or detailed ground truth establishment in the context of AI performance.

    The document states "Clinical data: Not applicable," further confirming the absence of a clinical study for this clearance.

    Here's a breakdown based on the information available in the document:

    1. A table of acceptance criteria and the reported device performance:

    The document refers to compliance with consensus standards for device performance, where the "acceptance criteria" are implied by adherence to these standards and the "reported device performance" is that the device "meets its accuracy specification and meet relevant consensus standards." Specific numerical acceptance criteria are mentioned for the measurement accuracy of SpO2, PR, and NIBP, and various temperature modules.

    ParameterAcceptance Criteria (from Predicate Device/Standard)Reported Device Performance (Implied)
    SpO2 (EDAN Module)
    Measurement Range0% to 100%Complies
    Accuracy (Adult/Pediatric, 70-100%)±2%Meets specification
    Accuracy (Adult/Pediatric, 0-69%)UndefinedUndefined (as per spec)
    Accuracy (Neonates, 70-100%)±3%Meets specification
    Accuracy (Neonates, 0-69%)UndefinedUndefined (as per spec)
    PR from SpO2
    Measurement Range25 to 300 bpmComplies
    Accuracy±2 bpmMeets specification
    TEMP (Covidien F3000 Quick Temp Module)
    Measuring Range30°C~43°CComplies
    Prediction Range35°C~43°CComplies
    TEMP (Exergen TAT-5000S-RS232-QR Temp Module)
    Temperature Range61 to 110°F (15.5 to 43°C)Complies
    Accuracy± 0.2°F or 0.1°CMeets specification
    Clinical AccuracyPer ASTM E1112Complies with ASTM E1112
    TEMP (HTD8808C Non-contact Infrared Thermometer Temp Module)
    Measurement Range34 °C to 43.0 °C (Differs slightly from predicate)Complies
    Accuracy (34.0~34.9°C)± 0.3°CMeets specification
    Accuracy (35.0~42.0°C)± 0.2°CMeets specification
    Accuracy (42.1~43.0°C)± 0.3°CMeets specification
    TEMP (T100A Internal Forehead TEMP Module)
    Accuracy (34.0 - 34.9 °C)±0.3 °CMeets specification
    Accuracy (35.0 - 42.0 °C)±0.2 °CMeets specification
    Accuracy (42.1 - 43.0 °C)±0.3 °CMeets specification
    EDAN-NIBP (ICFUS) & EDAN-NIBP (IFAST)
    SYS Measurement RangeAdult: 25-290 mmHg; Pediatric: 25-240 mmHg; Neonatal: 25-140 mmHgComplies
    DIA Measurement RangeAdult: 10-250 mmHg; Pediatric: 10-200 mmHg; Neonatal: 10-115 mmHgComplies
    MAP Measurement RangeAdult: 15-260 mmHg; Pediatric: 15-215 mmHg; Neonatal: 15-125 mmHgComplies
    NIBP PR Measurement Range40 bpm to 240 bpmComplies
    NIBP PR Accuracy± 3 bpm or 3.5%, whichever is greaterMeets specification

    2. Sample size used for the test set and the data provenance:

    • The document explicitly states "Clinical data: Not applicable." The "test set" for performance was conducted through "Non-clinical data" including "Electrical safety and electromagnetic compatibility (EMC)" and "Performance testing-Bench."
    • No specific sample sizes (e.g., number of patients/measurements) for "bench" testing are provided.
    • Data provenance (e.g., country of origin, retrospective/prospective) is not applicable given no clinical study.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • Not applicable as no clinical ground truth or expert-adjudicated test set was used. Device performance was assessed against established engineering and safety standards.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

    • Not applicable as no clinical test set requiring adjudication was used.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • No MRMC study was performed as this device is a vital signs monitor, not an AI-assisted diagnostic device requiring human reader improvement assessment. Clinical data was deemed "Not applicable."

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    • The performance assessment focused on the device's ability to accurately measure vital signs according to specifications and consensus standards (e.g., IEC 80601-2-30 for NIBP, ISO 80601-2-56 for temperature, ISO 80601-2-61 for pulse oximetry). This is essentially a "standalone" performance evaluation of the measurement algorithms within the device against physical standards, but not in the context of an AI-driven diagnostic algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    • For the non-clinical performance testing, the "ground truth" would be the reference measurements from calibrated and traceable sources, as specified by the relevant international standards (e.g., simulated physiological signals/conditions, calibrated temperature sources, etc.).

    8. The sample size for the training set:

    • Not applicable. This document describes the clearance of a vital signs monitor, not an AI/machine learning algorithm requiring a separate training set.

    9. How the ground truth for the training set was established:

    • Not applicable, as no AI training set is described.
    Ask a Question

    Ask a specific question about this device

    K Number
    K232694
    Date Cleared
    2024-02-05

    (153 days)

    Product Code
    Regulation Number
    870.1025
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    MFM-CMS central monitoring system (hereinafter referred to as MFM-CMS) supports centralized management of patients' clinical data provided by EDAN medical devices. Clinicians can obtain patient clinical data via MFM-CMS. The indications for use of the MFM-CMS central monitoring system include:

    • · Viewing patient real-time clinical data and alarms.
    • · Storing and reviewing patient clinical data and alarms.
    • · Printing real-time and history patient data.
    • · Configuring local settings as well as synchronizing settings to a remote device through network.
    • · Accessing patient clinical data between several departments.

    MFM-CMS is intended to be used only in clinical or hospital environment by well-trained healthcare professionals.

    MFM-CMS is indicated for use when monitoring adult and/or pediatric and/or neonate patients as indicated by labeling of the medical device providing the data.

    Device Description

    MFM-CMS is a central monitoring system product, which can connect and manage information from EDAN medical devices. MFM-CMS offers central management for monitoring information from the medical devices. All these collected information can be displayed, printed, alarmed and recorded.

    AI/ML Overview

    The provided text is a 510(k) summary for the Edan Instruments, Inc. MFM-CMS Central Monitoring System. It describes the device, its intended use, and a comparison to predicate devices, but does not contain information related to specific acceptance criteria, reported device performance in those criteria, sample sizes, expert qualifications, or ground truth establishment for a diagnostic AI device.

    The submission is for a "Central Monitoring System" (MFM-CMS), which supports centralized management of patient clinical data from other EDAN medical devices. It is classified as an "Arrhythmia detector and alarm (including ST-segment measurement and alarm)" with product code MHX. However, the summary focuses on the system's ability to display, store, review, and print data, and manage settings, rather than its performance as an arrhythmia detector itself.

    Therefore, many of the requested details cannot be extracted from this document, as they are typically found in the clinical validation studies of algorithms that perform diagnostic or interpretative tasks.

    Here's an analysis based on the information available in the document:

    1. A table of acceptance criteria and the reported device performance

    This information is not provided in the document. The submission focuses on functional changes and comparison to predicates, not specific performance metrics like sensitivity, specificity, or accuracy for an arrhythmia detection algorithm. The "Performance" section within the comparison table refers to features like "Bi-directional Configuration" and "Data Review," not numerical performance criteria.

    2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    This information is not provided. The document states "Clinical testing: Not applicable. Clinical testing is not required to establish substantial equivalence to the predicate device." This indicates that no clinical test set was used to validate the device's performance in a diagnostic capacity.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    This information is not provided, as no clinical test set for diagnostic accuracy was utilized.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    This information is not provided, as no clinical test set for diagnostic accuracy was utilized.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    There is no indication that an MRMC study was done. The device is a "Central Monitoring System" and is not described as an AI-powered diagnostic tool that assists human readers.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This information is not provided, as the document does not describe the MFM-CMS as a standalone diagnostic algorithm. Its primary function is a central data management system. Although it is classified under "Arrhythmia detector and alarm," the detailed description of its updates and comparison to predicates emphasizes data handling and system functionality rather than algorithm performance for arrhythmia detection.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    This information is not provided, as no clinical validation study is described.

    8. The sample size for the training set

    This information is not provided. As the submission focuses on software updates and functional equivalence, details about training sets for an AI algorithm are not relevant or discussed.

    9. How the ground truth for the training set was established

    This information is not provided, as no clinical validation study or AI training is described.

    Summary of what is available in the document:

    The document describes the MFM-CMS as a central monitoring system that connects to other EDAN medical devices to manage patient clinical data. The submission is for an updated version (K232694) of an existing device (K120727), with the primary predicate being the BeneVision Central Monitoring System (K220058).

    Key changes and comparisons:

    The main changes to the software include:

    • Add distributed function.
    • Add license authorization.
    • Support department management, device management, and user management.
    • Support time synchronization function.
    • Support data automatic dump function.
    • Replace the software development platform.
    • Supports simultaneous login of multiple clients.
    • Support domain account to log in to the CMS client.

    The comparison table highlights similarities and differences in intended use, operating system support, data review features, calculations, telemetry support, print capabilities, and network connectivity between the subject device and its predicate.

    Performance Data (as per the document):
    The document states:

    • Non-clinical test: Biocompatibility testing and Electrical safety & electromagnetic compatibility (EMC) are "Not applicable."
    • Software Verification and Validation Testing: Conducted in accordance with FDA guidance for software in medical devices.
    • Bench Testing: Functional and system-level testing was conducted to validate performance, and results "show that the subject device meets relevant consensus standards" (e.g., IEC 60601-1-8:2006 + Am1:2012 for alarm systems).
    • Clinical testing: "Not applicable. Clinical testing is not required to establish substantial equivalence to the predicate device."

    Conclusion: The submission concludes that "The bench testing data and software verification and validation demonstrate that MFM-CMS Central Monitoring System is substantially equivalent to the predicate devices."

    In essence, this FDA 510(k) summary focuses on demonstrating that the updated MFM-CMS system maintains the safety and effectiveness of its predicate devices through non-clinical testing and software verification, rather than providing a detailed performance study of a diagnostic algorithm against specific clinical acceptance criteria.

    Ask a Question

    Ask a specific question about this device

    K Number
    K230184
    Date Cleared
    2023-09-25

    (245 days)

    Product Code
    Regulation Number
    870.2800
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The SA-20 Series recorder combines ambulatory ECG Holter and BP monitor. It is intended to acquire, store and display ambulatory ECG signals from adult and pediatric patient. It is capable of measuring systolic and diastolic blood pressures of adults and pediatrics (> 12 years) over a preprogrammed period of time. These measurements are displayed and stored in the recorder. It is intended to be used by trained personnel under the direction of doctors.

    Device Description

    Holter ECG and ABP System (Model: SA-20, SA-19, SA-18, SA-15, SA-16) is composed of the recorder, NIBP cuff, lead wire. The ambulatory blood pressure adopts the principle of the oscillometric method for non-invasive blood pressure measurement. ECG signals are obtained by placing electrodes on the patient's surface to record the potential change signal of the human heart and then processing the data to form an ECG waveform and store it in the Recorder to monitor the ECG signal.

    AI/ML Overview

    The provided text describes the Holter ECG and ABP System and its performance data to establish substantial equivalence to predicate devices. However, the document does not contain an acceptance criteria table or specific details of a study directly proving the device meets acceptance criteria in the format often associated with AI/ML algorithm validation (e.g., sensitivity, specificity, AUC).

    Instead, the document focuses on non-clinical performance testing to demonstrate the device's adherence to relevant industry standards and functional specifications, ensuring its safety and effectiveness are comparable to the predicate devices.

    Here's a breakdown of the information that can be extracted or inferred based on common medical device submission practices, acknowledging that explicit details on some points are missing:


    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not present a formal "acceptance criteria" table in terms of diagnostic performance metrics (e.g., sensitivity, specificity for disease detection). Instead, the performance evaluations are based on compliance with established medical device standards. The "reported device performance" is the successful compliance with these standards.

    Acceptance Criteria (Compliance with Standard)Reported Device Performance (Status)
    Electrical Safety and Electromagnetic Compatibility (EMC)
    ANSI AAMI ES 60601-1:2005/(R) 2012 and A1:2012, C1:2009(R) 2012 and A2:2010(R) 2012 (Medical electrical equipment - General requirements for basic safety and essential performance)Complies
    IEC 60601-1-2:2014 (Fourth Edition) (Medical electrical equipment – Electromagnetic disturbances - Requirements and tests)Complies
    Performance Testing - Bench
    IEC 60601-2-47 Edition 2.0 2012-02 (Medical electrical equipment - Particular requirements for the basic safety and essential performance of ambulatory electrocardiographic systems)Meets accuracy specifications, complies
    IEC 80601-2-30: Edition 2.0 2018-03 (Medical electrical equipment - Particular requirements for the basic safety and essential performance of automated non-invasive sphygmomanometers)Meets accuracy specifications, complies
    ANSI AAMI EC53:2013/(R) 2020 (ECG trunk cables and patient leadwires)Meets accuracy specifications, complies
    IEC 60601-1-11 Edition 2.1 2020-07 (Medical electrical equipment - General requirements for basic safety and essential performance - Home healthcare environment)Meets accuracy specifications, complies
    Software Verification and ValidationDocumentation provided as recommended by FDA guidance.

    2. Sample size used for the test set and the data provenance

    The document states, "Clinical data: Not applicable," implying that the assessment for this 510(k) submission did not involve a clinical trial or a test set derived from patient data in the typical sense of evaluating diagnostic accuracy. The testing was primarily bench-based and software verification. Therefore, there's no "sample size for a test set" or "data provenance" related to patient data for diagnostic performance evaluation.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    Not applicable, as no clinical test set for diagnostic performance was used. The "ground truth" for the non-clinical testing was defined by the specifications and requirements of the referenced industry standards.

    4. Adjudication method for the test set

    Not applicable, as no clinical test set for diagnostic performance was used.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This device is a Holter ECG and ABP system (hardware for acquisition, storage, and display of physiological signals), not an AI/ML diagnostic algorithm designed to assist human readers or perform automated interpretations beyond basic parameter measurements.

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done

    Not applicable. This is not an AI/ML algorithm in the context of diagnostic interpretation. It is a device for signal acquisition and measurement. The "algorithm" for NIBP (oscillometric method) and ECG signal processing is inherent to the device's function, but it's not described as a standalone diagnostic algorithm requiring performance evaluation against a specific ground truth like an AI model would be.

    7. The type of ground truth used

    For the performance testing, the "ground truth" was the specifications and requirements defined by the referenced international and national consensus standards (e.g., IEC 60601 series, ANSI AAMI standards). The device's measurements (ECG signals, blood pressure) are verified against the expected outputs and accuracy tolerances dictated by these standards.

    8. The sample size for the training set

    Not applicable. This is not an AI/ML algorithm that requires a training set of patient data.

    9. How the ground truth for the training set was established

    Not applicable. This is not an AI/ML algorithm that requires a training set.


    Summary of the Study that Proves the Device Meets Acceptance Criteria (based on the document):

    The "study" proving the device meets its "acceptance criteria" is a series of non-clinical performance tests and software verification and validation activities designed to demonstrate compliance with recognized medical device standards.

    1. Electrical Safety and EMC Testing: The device was tested against ANSI AAMI ES 60601-1 and IEC 60601-1-2 to ensure it is electrically safe and does not cause or succumb to electromagnetic interference.
    2. Performance Testing (Bench): Functional and system-level bench testing was conducted to validate the device's performance against the accuracy specifications and requirements outlined in:
      • IEC 60601-2-47: For ambulatory electrocardiographic systems (ECG Holter functionality).
      • IEC 80601-2-30: For automated non-invasive sphygmomanometers (ABP functionality).
      • ANSI AAMI EC53: For ECG trunk cables and patient leadwires.
      • IEC 60601-1-11: For devices used in the home healthcare environment.
        These tests would involve simulating physiological signals or using calibrated equipment to ensure the device accurately acquires, processes, and displays ECG and ABP data within acceptable error margins defined by these standards.
    3. Software Verification and Validation: This involved rigorous testing of the device's embedded software to ensure it functions as intended, is free from critical bugs, and meets specified requirements, following FDA guidance for software in medical devices.

    The conclusion drawn from these "non-clinical data" is that the "subject devices are as safe and as effective as the predicate device," thus establishing substantial equivalence for market clearance.

    Ask a Question

    Ask a specific question about this device

    K Number
    K222902
    Date Cleared
    2023-07-28

    (308 days)

    Product Code
    Regulation Number
    870.2340
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The SE-1200 Pro&SE-1201 Pro 12-lead electrocardiographs are intended to acquire ECG signals from adult and pediatric patients through body surface ECG electrodes. The electrocardiographs are only intended to be used in hospitals or healthcare facilities by doctors and trained healthcare professionals. The cardiogram recorded by the electrocardiograph can help users to analyze and diagnose heart disease. However, the interpreted ECG with measurements and interpretive statements is offered to clinicians on an advisory basis only.

    Device Description

    SE-1201 Pro&SE-1200 Pro (2 models: SE-1201 Pro and SE-1200 Pro) features a LCD touch screen, an operation panel, user-programmable reports, and the ability to operate on either battery or AC power. It is capable of simultaneous acquisition, display, and print of 12-lead ECG. It uses algorithm to generate measurements, data presentations, graphical presentations and interpretative statements. The record can be saved in flash memory or send to PC.

    AI/ML Overview

    The provided document is a 510(k) premarket notification for Electrocardiograph models SE-1200 Pro and SE-1201 Pro. It aims to demonstrate substantial equivalence to previously cleared predicate devices.

    However, the document makes a clear statement regarding clinical data: "Clinical data: Not applicable." This indicates that no clinical study was conducted to prove the device meets specific acceptance criteria based on human performance or diagnostic accuracy. Instead, the submission relies on non-clinical data, including electrical safety, electromagnetic compatibility (EMC), and performance testing (bench).

    Therefore, I cannot provide the information requested in your prompt regarding acceptance criteria, study details, expert involvement, or MRMC studies, as the submission explicitly states that clinical data was not applicable.

    The device's performance demonstration relies on:

    1. Electrical safety and electromagnetic compatibility (EMC): Conformity to standards like ANSI AAMI ES 606601-1:2005/(R) 2012, IEC 60601-1-2:2014, and IEC 60601-2-25 Edition 2.0 2011-10.
    2. Performance testing (bench): Functional and system-level testing to validate accuracy specifications and compliance with relevant consensus standards (e.g., IEC 60601-2-25).
    3. Software Verification and Validation Testing: Documentation provided as recommended by FDA guidance for software in medical devices.

    The conclusion states that "The bench testing data and software verification and validation demonstrate that Electrocardiograph SE-1201 Pro&SE-1200 Pro is substantially equivalent to the predicate devices." This substantial equivalence is based on the technical and non-clinical performance characteristics, not on a clinical effectiveness study involving human subjects or interpretation.

    Ask a Question

    Ask a specific question about this device

    K Number
    K220308
    Date Cleared
    2022-08-11

    (190 days)

    Product Code
    Regulation Number
    870.1025
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The RespArray™ patient monitor is intended to be used for monitoring, storing, of, and to generate alarms for, multiple physiological parameters of adults, pediatrics and neonates. The monitors are intended for use by trained healthcare professionals in hospital environments.
    The monitor is for prescription use only.
    The monitored physiological parameters include: ECG, respiration (RESP), temperature (TEMP), oxygen saturation of arterial blood (SpO2), pulse rate (PR), non-invasive blood pressure (NIBP), and carbon dioxide (CO2).
    The arrhythmia detection and ST Segment analysis are intended for adult patients.
    The SpO2 (NellcorTM) module is intended to be used for spot-check or continuous non-invasive monitoring of functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate (PR), in motion conditions, and in patients who are well or poorly perfused.
    The MicrostreamTM capnography module is intended for continuous non-invasive monitoring of carbon dioxide concentration of the expired and inspired breath (etCO2) and respiration rate (RR).
    The monitor also provides the clinician with integrated pulmonary index (IPI), apnea per hour (A/hr) and oxygen desaturation index (ODI) values. IPI is not intended for patients up to the age of one year. Allr and ODI are intended for ages 22 and up.
    The monitors are not intended for MRI environments.

    Device Description

    The RespArray patient monitor (hereinafter called RespArray) can perform long-time continuous monitoring of multiple physiological parameters. Also, it is capable of storing, displaying, analyzing and controlling measurements, and it will indicate alarms in case of abnormity so that doctors and nurses can deal with them in time.

    AI/ML Overview

    The provided text is a 510(k) Premarket Notification summary for the "Patient Monitor: RespArray" device. This type of submission focuses on demonstrating substantial equivalence to legally marketed predicate devices, rather than conducting new clinical trials for de novo clearance or PMA approval. Therefore, the details requested about acceptance criteria, specific study design (like MRMC studies, sample sizes, expert ground truth establishment for AI/algorithm performance), and training set information are not typically found in these types of submissions, as the FDA review here centers on comparing the new device's specifications and performance to an existing, already cleared device.

    The document primarily highlights the device's technical specifications and how they compare to a predicate device (Edan Instruments, Inc, Patient Monitor Model X8, X10, X12 - K192514), along with compliance with relevant electrical safety, EMC, and performance standards. It explicitly states "Clinical data: Not applicable."

    Given this, I will extract the information that is present in the document and indicate where the requested information is not applicable or not provided within the scope of a 510(k) submission focused on substantial equivalence.


    Analysis of the Provided Document for Device Acceptance Criteria and Study Proof

    The provided document is a 510(k) premarket notification. For devices cleared via a 510(k), the primary "acceptance criterion" is often substantial equivalence to a legally marketed predicate device, demonstrated through comparative testing and adherence to recognized standards. Direct, explicit "acceptance criteria" presented as quantitative performance targets with a detailed study to prove they are met (as might be seen in AI/ML clearances for algorithms with novel functionalities) are typically not included in this type of submission for a patient monitor.

    The "study" that proves the device meets the acceptance criteria is primarily non-clinical performance testing (bench testing) and software verification/validation to show that the device performs as intended and is as safe and effective as its predicate.

    Here's a breakdown of the requested information based on the provided document:


    1. A table of acceptance criteria and the reported device performance

    The document does not provide a table of explicit acceptance criteria/performance targets with quantitative results in the way one might expect for a new AI/ML algorithm. Instead, it demonstrates performance by stating compliance with recognized consensus standards and by comparing the subject device's specifications to those of its predicate device, showing "similar design features and performance specifications."

    The closest representation of "performance" and "acceptance" is the "Predicate Device Comparison" table (pages 5-6). This table implicitly acts as the performance comparison against the predicate device that serves as the "acceptance" benchmark for substantial equivalence.

    ItemSubject Device: RespArray (Reported Device Performance)Predicate Device: X8 X10 X12 (Implicit Acceptance Criterion/Benchmark)Comparison Result
    Indications for UseMonitoring, storing, reviewing of, and to generate alarms for multiple physiological parameters (ECG, RESP, TEMP, SpO2, PR, NIBP, CO2) for adults, pediatrics, and neonates in hospital environments. Arrhythmia detection and ST Segment analysis for adult patients. SpO2 for spot-check/continuous monitoring in motion/no motion. Microstream™ capnography for etCO2 and RR. Provides IPI, A/hr, ODI values. Not for MRI.Very similar, also monitoring, storing, recording, reviewing of, and to generate alarms for multiple physiological parameters (ECG, RESP, TEMP, SpO2, PR, NIBP, invasive blood pressure (IBP), CO2, cardiac output (C.O.)) for adults, pediatrics, neonates in hospital environments. Arrhythmia detection and ST Segment analysis for adult patients. Not for MRI.Similar (Slight differences, e.g., predicate includes IBP and C.O. vs. subject's more detailed SpO2/CO2 module descriptions and IPI/A/hr/ODI. However, overall intention described as "Similar")
    ECG Monitor Lead Mode3 Electrodes; 5 Electrodes;3 Electrodes; 5 Electrodes; 6 Electrodes ; 10 Electrodes ;Different (Subject supports fewer lead modes)
    Arrhythmia AnalysisASYSTOLE, VFIB/VTAC, COUPLET, VT > 2, BIGEMINY, TRIGEMINY, VENT, R on T, PVC, TACHY, BRADY, MISSED BEATS, IRR, VBRADY, PNC, PNPSame list of arrhythmia types.Same
    RESP Monitor PrincipleThoracic impedanceThoracic impedanceSame
    RESP Measurement Range0 rpm to 200 rpmAdult: 0 to 120 rpm; Pediatric/neonate: 0 rpm to 150rpmDifferent (Subject has a wider stated range, but the intent is likely overall comparable)
    NIBP PrincipleoscillationoscillationSame
    NIBP Measurement RangeSystolic: Adult 25-290, Pediatric 25-240, Neonate 25-140; Diastolic: Adult 10-250, Pediatric 10-200, Neonate 10-115; Mean: Adult 15-260, Pediatric 15-215, Neonate 15-125Same ranges.Same
    PR from NIBP Range40 bpm to 240 bpm40 to 240 bpmSame
    Temperature Range0 °C to 50 °C (32 °F to 122 °F)0 °C to 50 °C (32 °F to 122 °F)Same
    Wireless ConnectionWi-FiWi-FiSame
    Power SupplyAC power: Yes; Rechargeable Battery: YesAC power: Yes; Rechargeable Battery: YesSame
    CO2 ModuleMicrostream™ micorMediCO2 EtCO2 (Substantially equivalent to module cleared by K200594)/ (Predicate doesn't specify module, but supports CO2 monitoring)It is substantial equivalent to the CO2 Module cleared by K200594
    SpO2 ModuleNell-1 (Substantially equivalent to module cleared by K141542)/ (Predicate doesn't specify module, but supports SpO2 monitoring)It is substantial equivalent to the SpO2 Module cleared by K141542

    The document concludes that "the subject and predicate devices have similar design features and performance specifications. The technological differences between the subject and predicate devices do not raise different questions of safety or effectiveness."

    2. Sample size used for the test set and the data provenance

    The document states "Clinical data: Not applicable." Therefore, there isn't a "test set" in the sense of patient data used for clinical validation of, for example, an AI algorithm's performance. The "testing" primarily refers to non-clinical bench testing.

    • Sample size: Not applicable for patient data test set. For bench testing, samples would be physical devices, components, or simulated signals, but a "sample size" in terms of patient numbers is not provided.
    • Data provenance (e.g., country of origin of the data, retrospective or prospective): Not applicable, as no clinical data test set was used/provided.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    Not applicable. Since no clinical data test set was described and "Clinical data: Not applicable" is stated, there was no need for expert ground truth establishment for a test set. This type of information would be relevant for AI/ML device clearances where human expert annotation is part of the ground truth creation.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Not applicable, as no clinical test set requiring adjudication of ground truth was described.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    No. The document explicitly states "Clinical data: Not applicable." MRMC studies are typically for evaluating the impact of AI algorithms on human reader performance, which is not the scope of this 510(k) submission for a patient monitor.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    The device is a patient monitor with various physiological parameter measurements and alarms. Its "performance" is inherent in its ability to accurately measure these parameters and detect events like arrhythmias. The non-clinical bench testing demonstrated its standalone performance by showing compliance with relevant standards (e.g., IEC 60601-2-27 for ECG, IEC 80601-2-30 for NIBP, ISO 80601-2-61 for pulse oximeter).

    While not explicitly called "standalone algorithm performance" in the AI/ML sense, the "Performance testing-Bench" section (page 8) confirms that "Edan has conducted functional and system level testing to validate the performance of the results of the bench testing show that the subject device meets its accuracy specification and meet relevant consensus standards." This demonstrates the device's functional performance in isolation.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    For the non-clinical bench testing, the "ground truth" would be established by:

    • Reference instruments or calibrated signals (e.g., precise electrical signals for ECG, known pressure values for NIBP, calibrated gas mixtures for CO2).
    • Standardized measurement protocols defined by the cited IEC/ISO standards.
    • Accuracy specifications found within those standards or the device's own specifications.

    There's no mention of expert consensus, pathology, or outcomes data as "ground truth" because this is a measurement and alarm device, not a diagnostic imaging AI algorithm, and no clinical data was used for validation in this submission.

    8. The sample size for the training set

    Not applicable. This device is a patient monitor, not an AI/ML algorithm that undergoes a distinct training phase on a dataset. The underlying algorithms for parameter measurement (e.g., NIBP oscillometric algorithm, arrhythmia detection) are established engineering designs, not typically "trained" in the machine learning sense with large datasets.

    9. How the ground truth for the training set was established

    Not applicable, as there is no specific "training set" for an AI/ML algorithm described. The "ground truth" for the development and calibration of the monitor's measurement algorithms would have been established through engineering principles, laboratory testing with calibrated instruments, and referencing physiological models and data, but this is part of the device's fundamental design and not a separate "training set" as understood in current AI/ML contexts.

    Ask a Question

    Ask a specific question about this device

    K Number
    K212278
    Date Cleared
    2022-04-07

    (261 days)

    Product Code
    Regulation Number
    870.2340
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The iSE series electrocardiograph is intended to acquire ECG signals from adult and pediatric patients through body surface ECG electrodes. The electrocardiograph is only intended to be used in hospitals or healthcare facilities by doctors and trained healthcare professionals. The cardiogram recorded by the electrocardiograph can help users to analyze and diagnose heart disease. However, the interpreted ECG with measurements and interpretive statements is offered to clinicians on an advisory basis only. The electrocardiograph is capable of network communications and supports the informatized management of workflows in hospital and healthcare facilities.

    Device Description

    iSE-1210 and iSE-1810 Electrocardiograph can support ECG Data Sampling of maximum 12 leads and maximum 18 leads separately. iSE series electrocardiograph uses algorithm to generate measurements, data presentations, graphical presentations and interpretative statements. The electrocardiograph is only intended to be used in hospitals or healthcare facilities by doctors and trained healthcare professionals.

    AI/ML Overview

    The provided text is a 510(k) Premarket Notification summary for the Edan Instruments, Inc. Electrocardiograph (iSE series). It focuses on establishing substantial equivalence to previously cleared devices.

    Based on the provided document, here's a breakdown of the acceptance criteria and the study that proves the device meets them:

    Key Takeaway: This 510(k) summary does not include a clinical study with real patient data to prove the device's diagnostic performance against a ground truth for conditions like heart disease. Instead, it relies heavily on non-clinical bench testing, electrical safety, EMC, human factors, and software verification/validation to demonstrate that the new device (iSE series) is as safe and effective as its predicate devices, primarily in its ability to acquire and process ECG signals according to established industry standards. The interpretation of ECGs with measurements and interpretive statements is explicitly stated as advisory only.

    Therefore, the acceptance criteria and study described will pertain to the technical performance and safety of the ECG acquisition and processing rather than the diagnostic accuracy of an AI algorithm in interpreting ECGs for specific heart conditions.


    Acceptance Criteria and Device Performance

    The document does not present a formal table of acceptance criteria with corresponding performance results in the format requested for diagnostic performance. Instead, it states that the device meets "accuracy specification" and "relevant consensus standards" through non-clinical bench testing.

    Based on the "Performance Specifications" table (Table 1) and the "Performance testing-Bench" section, we can infer some performance aspects that were likely tested against standards.

    Inferred Acceptance Criteria and Reported Device Performance (based on technical specifications and implied testing):

    Acceptance Criteria CategorySpecific Criterion (Inferred from standards/specs)Reported Device Performance
    Electrical SafetyCompliance with ANSI AAMI ES 60601-1:2005/(R) 2012 and A1:2012, C1:2009/(R) 2012 and A2:2010/(R) 2012"assessed for conformity with the relevant requirements" (Pass)
    Electromagnetic Compatibility (EMC)Compliance with IEC 60601-1-2:2014 (Fourth Edition)"assessed for conformity with the relevant requirements" (Pass)
    ECG Signal Accuracy / PerformanceCompliance with IEC 60601-2-25 Edition 2.0 2011-10 (Particular requirements for electrocardiographs)"meets its accuracy specification and meet relevant consensus standards" (Pass)
    Heart Rate (HR) Recognition RangeHR Range for accurate recognition: 30 BPM - 300 BPM30 BPM ~ 300 BPM (Same as predicate)
    Analog-to-Digital ConversionA/D Resolution: 24 bits24 bits (Same as predicate)
    Sampling Frequency (ECG Unit)Sampling Frequency: 64,000 Hz (for general ECG unit)64,000 Hz (Higher than predicate's 16,000 Hz, implying improved capability)
    Frequency ResponseFrequency Response: 0.01Hz ~ 350Hz (-3 dB)0.01 Hz ~ 350 Hz (-3 dB) (Wider than predicate's 0.01Hz~300Hz, implying improved capability)
    Filter PerformanceFunctionality of AC, DFT, EMG, and LOWPASS FiltersAC Filter: Same; DFT Filter: Same; EMG Filter: Same; LOWPASS Filter: Different (350Hz/300Hz/270Hz/150Hz/100Hz/75 Hz vs. 300Hz/270Hz/150Hz/100Hz/75Hz for predicate) (Implied Pass)
    Pacemaker Detection Sampling FrequencyPacemaker Detection Sampling Frequency (for specific models)iSE-1810: 16,000Hz, Rhythm Lead; iSE-1210: 80,000Hz, Rhythm Lead (Different from predicate) (Implied Pass)
    Human Factors/UsabilityCompliance with "Applying Human Factors and Usability Engineering to Medical Devices" guidance and IEC 62366-1:2020"safe and effective for the intended users, uses, and use environments" (Pass)
    Software Verification & ValidationCompliance with FDA's "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices""conducted and documentation was provided" (Implied Pass)

    Study Details (Based on the provided document)

    1. Sample size used for the test set and the data provenance:

      • Test Set Sample Size: Not specified for any performance testing beyond "Functional and system level testing" and "usability testing." No specific number of patient ECGs are mentioned for any diagnostic validation.
      • Data Provenance: The document does not specify the country of origin for any data (if patient data was used implicitly in some "accuracy specification" bench testing, which is unlikely given the context). The studies mentioned (Electrical safety, EMC, performance bench testing, human factors, software verification) are non-clinical bench and lab-based studies. There is no mention of prospective or retrospective patient data collection.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

      • Not Applicable. For non-clinical bench testing, the "ground truth" is typically defined by engineering specifications, calibration standards, and validated test equipment rather than expert human interpretation of medical data.
      • The interpretative statement for the ECGs is offered "on an advisory basis only," indicating that the device's interpretation is not intended to replace a physician's diagnosis.
    3. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

      • Not Applicable. The studies described are technical performance tests against standards, not human reader studies requiring adjudication of medical diagnoses.
    4. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

      • No. An MRMC study was not conducted or reported. The device's interpreted ECGs are advisory, and there's no claim of AI assistance for human readers in the diagnostic process. The "algorithm" mentioned (SEMIP V1.92) generates measurements, data presentations, graphical presentations, and interpretative statements, but its performance is not assessed in a comparative effectiveness study with human readers.
    5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

      • Partially Yes, for technical performance. The bench testing (e.g., meeting accuracy specifications, sampling frequency, frequency response) can be considered a standalone performance assessment of the device's ability to acquire and process ECG signals according to engineering and industry standards. However, this is distinct from a standalone diagnostic performance study typically seen for AI/CADe devices. The document explicitly states "Clinical data: Not applicable." This confirms that no standalone clinical diagnostic performance study was performed or needed for this 510(k) submission, as the device's interpretations are advisory.
    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

      • Technical Specifications/Industry Standards. For the non-clinical tests, the "ground truth" for the device's performance is compliance with established electrical safety, EMC, and ECG technical performance standards (e.g., IEC 60601 series). There is no "ground truth" of patient diagnosis (e.g., pathology, outcomes, expert consensus on disease states) used as this device is broadly an ECG acquisition and processing device, with its interpretive statements being advisory.
    7. The sample size for the training set:

      • Not applicable / Not specified. The document makes no mention of AI model training or a training set. The "algorithm" (SEMIP V1.92) is mentioned as generating interpretive statements, but its development process (e.g., use of a training set, machine learning) is not described or evaluated in this 510(k) summary. Given the nature of a foundational ECG device and the "advisory basis only" statement for interpretations, it's highly likely this algorithm is rule-based or statistical rather than a deep learning model requiring a large training set of annotated ECGs for diagnostic purposes.
    8. How the ground truth for the training set was established:

      • Not applicable. As no training set is mentioned or evaluated in the provided document, the process for establishing its ground truth is not described.
    Ask a Question

    Ask a specific question about this device

    K Number
    K210139
    Date Cleared
    2021-10-01

    (255 days)

    Product Code
    Regulation Number
    870.1130
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Ambulatory Blood Pressure Monitor is capable of measuring systolic and diastolic blood pressures, and pulse of adults and pediatrics (> 12 years) over a preprogrammed period of time. These measurements are stored in the monitor and can be transferred to analysis system via wire or wireless transmission. It is intended for use as an aid or adjunct to diagnosis and treatment.

    Device Description

    The SA Series ABPM (Ambulatory Blood Pressure Monitor), primarily composed of Ambulatory Blood Pressure Monitor (ABPM) and ABPM Analysis software (Smart ABPM View), is designed to measure systolic and diastolic blood pressures, and pulse of adults and pediatrics (> 12 years) over a preprogrammed period of time. The monitor inflates and deflates the cuff on the upper arm to measure blood pressures and pulse by the oscillometric method and stores the measurement. The stored measurements are transferred into analysis software installed in generally used PC via wire transfer or Bluetooth. The monitor is carried around by patients. Patients are requested to come back to medical treatment site after recording time. The systolic and diastolic blood pressures, and pulse data recorded in monitor will be transferred into PCs and then analyzed, displayed and edited by Analysis software.

    AI/ML Overview

    The provided text describes the 510(k) premarket notification for the Edan Instruments, Inc. Ambulatory Blood Pressure Monitor (Models: SA-10, SA-05, SA-06, SA-08 and SA-09). The device measures systolic and diastolic blood pressure and pulse using the oscillometric method.

    Here's an analysis of the acceptance criteria and the study that proves the device meets them:

    Acceptance Criteria and Reported Device Performance

    The device's performance was evaluated against the international standard for non-invasive sphygmomanometers, ISO 81060-2.

    Acceptance Criteria (ISO 81060-2:2018+A1:2020)Reported Device Performance (SA series ABPM)
    Mean error: ±5 mmHgMean error: ±5 mmHg
    Standard deviation: Max. 8 mmHgStandard deviation: Max. 8 mmHg
    Heart Rate Acuracy: Not explicitly stated but the predicate device criteria was used
    Heart Rate accurate within ±3 bpm or ±3.5%, whichever is greater.

    Note: The document explicitly states that the "Blood Pressure results meet or exceed ANSI/AAMI/ISO 81060-2:2013 standards for non-invasive accuracy: ±5 mmHg mean error & 8 mmHg standard deviation" in the predicate comparison table, and later confirms compliance with the updated ISO 81060-2:2018+A1:2020 standard. For heart rate accuracy, the subject device's performance criterion is listed as "Heart Rate accurate within ±3 bpm or ±3.5%, whichever is greater," which is also the reported performance.

    Study Information

    Here's a breakdown of the study details based on the provided text:

    1. Sample size used for the test set and the data provenance:

      • Sample Size: The document mentions "investigation with qualified subject distribution" for the clinical test but does not specify the exact sample size.
      • Data Provenance: The document implies the data is prospective, as it describes a "clinical investigation" undertaken specifically to validate the device. The country of origin of the data is not specified.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

      • The document does not provide details on the number or qualifications of experts used to establish ground truth. For blood pressure monitors, ground truth typically involves simultaneous measurements by a trained professional using a reference standard (e.g., auscultation with a mercury sphygmomanometer) as part of the ISO 81060-2 standard protocol.
    3. Adjudication method for the test set:

      • The document does not describe a specific adjudication method. For validation against ISO 81060-2, the comparison is typically direct against the reference standard.
    4. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

      • No MRMC comparative effectiveness study was done. This device is a standalone blood pressure monitor, not an AI-assisted diagnostic tool for interpretation by human readers.
    5. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

      • Yes, a standalone performance study was done for the device. The "Clinical test" section states, "The subject devices have been undertaken clinical investigation in accordance with ISO 81060-2:2018+A1:2020 to validate the clinical safety and effectiveness within its intended use." This standard is for the performance of automated non-invasive sphygmomanometers themselves.
    6. The type of ground truth used:

      • The ground truth would be established through simultaneous or closely timed measurements by a "reference standard" as defined by ISO 81060-2, typically involving trained observers using a mercury or other validated sphygmomanometer performing auscultatory measurements. This is inherent to the ISO 81060-2 clinical validation protocol.
    7. The sample size for the training set:

      • The document does not mention a training set sample size. Ambulatory Blood Pressure Monitors typically use proprietary algorithms for oscillometric measurements, which are developed and validated during the product's R&D phase; however, details of an algorithm's training data are not usually part of a 510(k) summary for this type of device. The ISO 81060-2 study is a validation, not a training activity.
    8. How the ground truth for the training set was established:

      • The document does not provide information about how ground truth was established for any training set, as it does not discuss a training set for the NIBP algorithm.
    Ask a Question

    Ask a specific question about this device

    K Number
    K202892
    Date Cleared
    2021-01-28

    (122 days)

    Product Code
    Regulation Number
    870.2300
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The device is intended to be used for measuring, storing, and to generate prompts for, multiple physiological parameters of adults and pediatrics. The device is intended for use by trained healthcare professionals in hospital environments.

    Parameters include: NIBP, SpO2, PR (pulse rate), TEMP.

    The F3000 Quick TEMP module is not intended for neonates.

    The device is not intended for MRI environments.

    Device Description

    The iM3s series vital signs monitors including iM3s\iM3As\iM3Bs\ iHM3s are intended to be used for measuring, storing, reviewing of, and generating prompts for multiple physiological parameters of adults, pediatrics and neonates.

    AI/ML Overview

    Here's an analysis of the provided text regarding the acceptance criteria and study data for the Vital Signs Monitor (iM3s, iM3As, iM3Bs, iHM3s).

    Please note: The provided document is an FDA 510(k) summary, which focuses on demonstrating substantial equivalence to a predicate device. It primarily presents performance specifications and shows that the new device meets those, often by referencing existing standards or previous clearances. It does not typically include detailed experimental setups for clinical or performance studies in the way a research paper would. Therefore, some information requested (like sample size for training sets, specific adjudication methods, MRMC studies, or detailed ground truth establishment for training data) is not present in this type of regulatory document.


    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria are generally implied by the predicate device's specifications and compliance with relevant ISO/IEC standards. The "Reported Device Performance" for the subject device is stated as meeting these same specifications.

    ParameterAcceptance Criteria (from Predicate/Standards)Reported Device Performance (Subject Device)
    SpO2 AccuracyAdult/Pediatric:Adult/Pediatric:
    70 to 100%: ±2 %±2% (70% to 100% SpO2)
    0-69% unspecifiedUndefined (0% to 69% SpO2)
    Neonate:Neonate:
    70 to 100%: ±3%±3% (70% to 100% SpO2)
    0-69% unspecifiedUndefined (0% to 69% SpO2)
    PR from SpO2 Accuracy±2 bpm±2 bpm
    TEMP Measurement Range30°C~43°C30°C~43°C
    TEMP Prediction Measurement Range35°C~43°C35°C~43°C
    NIBP PR Accuracy(Referenced to NIBP module of X series cleared by K192514, which would presumably have its own stated accuracy. The document states NIBP PR accuracy as)±3 bpm or 3.5%, whichever is greater
    Electrical SafetyCompliance with ANSI AAMI ES 60601-1:2005/(R) 2012 and A1:2012, C1:2009(R) 2012 and A2:2010/(R) 2012Complies (assessed for conformity)
    EMCCompliance with IEC 60601-1-2:2014 (Fourth Edition)Complies (assessed for conformity)
    NIBP PerformanceCompliance with IEC 80601-2-30:2009+A1:2013Meets accuracy specification and relevant consensus standards
    TEMP PerformanceCompliance with ISO 80601-2-56: 2017+A 1:2018Meets accuracy specification and relevant consensus standards
    Pulse Oximeter PerformanceCompliance with ISO 80601-2-61: 2017Meets accuracy specification and relevant consensus standards
    Software Verification & ValidationAdherence to FDA Guidance "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices"Documentation provided and testing conducted

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size for Test Set: Not specified in the document. The performance is validated through "functional and system level testing" and compliance with standards, not a clinical trial with a specified patient sample size for device comparison.
    • Data Provenance: The studies are described as "non-clinical data" and "bench testing." No country of origin is mentioned for patient data, as no clinical studies with human subjects are detailed. The nature of the testing implies it's retrospective, based on existing standards and validation procedures, rather than prospective clinical data.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Their Qualifications

    • This information is not provided in the document. The "ground truth" for these performance tests would typically be established by highly calibrated reference instruments or simulated physiological signals according to the requirements of the cited ISO/IEC standards, rather than expert human interpretation in a clinical context.

    4. Adjudication Method for the Test Set

    • This information is not applicable and not provided. As no clinical studies with human subjects or interpretation tasks are described, there is no need for an adjudication method. The assessment is against technical specifications and standards.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

    • No, an MRMC comparative effectiveness study was not done. The document explicitly states "Clinical data: Not applicable." The focus is on the device's standalone performance against engineering and international standards.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    • Yes, the performance reported is essentially standalone. The "bench testing" and "software verification and validation testing" assess the device's accuracy and functionality in isolation, without an explicit human-in-the-loop component for interpreting outputs in the context of the device's core physiological measurements (NIBP, SpO2, PR, TEMP).

    7. The Type of Ground Truth Used

    • The ground truth for the performance testing is based on:
      • Reference measurements from calibrated equipment: For parameters like SpO2, NIBP, and TEMP, the "ground truth" would be simulated physiological signals or measurements from highly accurate, traceable reference devices as mandated by the respective ISO/IEC standards.
      • Compliance with international standards: The device is tested against the specific requirements and accuracy tolerances defined in standards such as IEC 80601-2-30 (NIBP), ISO 80601-2-56 (thermometers), and ISO 80601-2-61 (pulse oximeters).

    8. The Sample Size for the Training Set

    • This information is not provided and is not applicable in the context of this device and document. This device is a vital signs monitor, not typically an AI/machine learning diagnostic device that relies on "training sets" in the conventional sense of machine learning algorithms. Its parameters are measured directly through hardware and firmware, validated against physical standards.

    9. How the Ground Truth for the Training Set Was Established

    • This information is not provided and is not applicable as there is no "training set" for an AI algorithm described for this device. The device's fundamental operational principles are based on established physiological measurement techniques, not on learning from a dataset.
    Ask a Question

    Ask a specific question about this device

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Acclarix AX3 series/Acclarix LX3 series Diagnostic Ultrasound System is intended for use by a qualified physician or allied health professional for ultrasound evaluations in hospitals and clinical applications include:Abdominal. Gvnecology, Obstetric. Cardiac, Small parts. Urology, Musculosketal . Peripheral vascular. Intraoperative, Pediatric, Neonatal , Adult Cephalic.

    Device Description

    Acclarix AX3 Series/ Acclarix LX3 Series is a software controlled Diagnostic Ultrasound System, which consists of a main unit along with associated transducers. It is intended for use by a qualified physician or allied health professional for ultrasound evaluations. This system is a Track 3 device to acquire and display ultrasound data in various imaging modes.

    AI/ML Overview

    The provided text describes a 510(k) premarket notification for a Diagnostic Ultrasound System. This document focuses on demonstrating substantial equivalence to previously cleared predicate devices, rather than establishing acceptance criteria and conducting a new study to prove a device meets those criteria for novel technology.

    Therefore, many of the requested elements for a study proving device performance against acceptance criteria are not available in this document. The document primarily focuses on explaining how the new device is similar to already approved ultrasound systems and transducers.

    Here's a breakdown of the available information:

    1. Table of Acceptance Criteria and Reported Device Performance:

    This information is not explicitly stated in terms of specific performance metrics with acceptance criteria and their corresponding reported values. Instead, the document claims substantial equivalence to predicate devices based on:

    • Intended Use/Indications for Use
    • Mode of Operations
    • Design principles
    • Hardware and software platforms
    • Transducer types
    • Acoustic Output (stated to be below FDA limits)
    • Materials of transducers and needle guide brackets

    2. Sample size used for the test set and data provenance:

    • Test Set Sample Size: Not provided. The document states "Clinical testing is not required" and primarily relies on non-clinical tests and comparison to predicate devices.
    • Data Provenance: The document implies that the "clinical applications" and "imaging modes" added to the new device are based on what was already cleared for the Acclarix LX9 series (K192879) and other predicate devices (K200232 and K201693). This suggests a reliance on retrospective data and established performance of predicate devices, rather than new prospective clinical trials for this specific submission. The origin of the data for the predicate devices is not specified in this document.

    3. Number of experts used to establish the ground truth for the test set and qualifications of those experts:

    • Not applicable. Since clinical testing was not required for this submission, there was no separate test set requiring new ground truth establishment by experts for specific AI algorithm performance.

    4. Adjudication method for the test set:

    • Not applicable. No specific adjudication method is mentioned as there was no dedicated clinical test set described for new performance evaluation.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, and the effect size:

    • No, an MRMC study was NOT done. The document explicitly states "Clinical testing is not required" and focuses on substantial equivalence based on technical specifications and previously cleared indications. There is no mention of human readers or AI assistance.

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

    • Not applicable. This document is for a diagnostic ultrasound system, not an AI algorithm. While the system may employ internal processing (which could include algorithmic components), the submission does not detail testing of standalone AI performance against a ground truth.

    7. The type of ground truth used:

    • Not applicable for novel performance claims. The "ground truth" implicitly relies on the established and cleared performance of the predicate diagnostic ultrasound systems and their associated transducers. The "Acoustic output testing" was done "as per the guideline 'Marketing Clearance of Diagnostic Ultrasound Systems and Transducers' dated June 27, 2019," which suggests adherence to industry standards and regulatory guidance for ultrasound device performance, rather than clinical ground truth for a specific medical condition.

    8. The sample size for the training set:

    • Not applicable. This document does not describe the development or training of a new AI algorithm. It refers to a diagnostic ultrasound system and its transducers.

    9. How the ground truth for the training set was established:

    • Not applicable. As no new AI algorithm development is described, there is no training set or ground truth specific to an algorithm's training.

    Summary of what the document does provide:

    The document serves as a 510(k) submission arguing for the substantial equivalence of the new Edan Instruments Diagnostic Ultrasound Systems (Acclarix AX3 series and Acclarix LX3 series) to existing cleared predicate devices. The arguments are based on:

    • Non-clinical performance tests demonstrating compliance with relevant electrical safety, EMC, and acoustic output standards (ANSVAAMI ES 60601-1, IEC 60601-1-2, IEC 60601-2-37, ISO 10993-1, ISO 10993-5, ISO 10993-10).
    • Comparison of intended use, indications for use, mode of operations, design principles, hardware/software platforms, and transducer types to previously cleared Edan Instruments ultrasound systems (K192791, K192879) and other predicate devices (K200232, K201693).
    • The argument is that the differences are minor and do not raise new questions of safety or effectiveness. For example, new clinical applications, transducers, and imaging/measurement modes have already been cleared on other predicate devices.
    Ask a Question

    Ask a specific question about this device

    Page 1 of 7