K Number
K200232
Date Cleared
2020-06-23

(145 days)

Product Code
Regulation Number
892.2050
Reference & Predicate Devices
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

L Vivo platform is intended for non-invasive processing of ultrasound images to detect, measure, and calculate relevant medical parameters of structures and function of patients with suspected disease.

Device Description

The LVivo System analyzes echocardiographic patient examination DICOM movies for Global ejection fraction (EF) evaluation. EF is evaluated using two orthogonal planes, four-chamber (4CH) and two-chamber (2CH) views, to provide fully automated analyses of LV function from the echo examination movies. It also has the ability to measure strain and to evaluate the Right Ventricle and well as to measure the bladder.

AI/ML Overview

This document describes the acceptance criteria and study results for DiA Imaging Analysis Ltd's LVivo Software Application, specifically focusing on its LVivoRV (Right Ventricular) and LVivo Bladder modules.

1. Table of Acceptance Criteria and Reported Device Performance

ModuleMetricAcceptance CriteriaReported Device Performance
LVivoRVFAC Correlation75% correlation (r ≥ 0.75) between LVivoRV's FAC measurement and manual FAC measurements by sonographers. This value is based on statistical data reported by FDA cleared systems for semi-automated RV function evaluation (e.g., EchoInsight by Epsilon).Primary Endpoint Met: Excellent correlation (r=0.79, p<0.0001) between LVivoRV FAC and manual measurements.
EDA Correlation(Secondary Endpoint) Implied expectation of good correlation with manual measurements.Excellent correlation (r=0.92) between LVivoRV EDA and the average of two sonographers' manual measurements (p<0.0001).
ESA Correlation(Secondary Endpoint) Implied expectation of good correlation with manual measurements.Excellent correlation (r=0.93) between LVivoRV ESA and the average of two sonographers' manual measurements (p<0.0001).
TAPSE Correlation(Secondary Endpoint) Implied expectation of correlation similar to "real life" performance of VVI compared to M-mode.Correlation of 0.62 with manual M-Mode measurements (similar to VVI vs. M-Mode correlation of r=0.66). When 5 outlier cases were omitted, r=0.72.
Free Wall Strain Correlation(Secondary Endpoint) Implied expectation of good correlation with manual measurements.Positive correlation of R=0.6 with 2D VVI. When 6 outlier cases were omitted, r=0.78.
LVivo BladderBladder Volume Agreement (200mL threshold)Kappa of at least 0.61 (substantial agreement) between the automated and manual method when differentiating between post-voiding volumes above or below 200mL. This threshold is considered clinically important for catheter placement decisions.Primary Endpoint Met: Excellent Kappa of 0.84, indicating excellent agreement between methods for the 200mL threshold. High overall agreement (0.93) and high sensitivity (100%) and specificity (80%).
Bladder Volume Correlation(Secondary / Implied) Implied expectation of good correlation between automated bladder volume calculation and manual tracing.Very high correlation (r=0.94) between automated bladder volume calculation by LVivo Bladder and volume calculated by manual tracing (routinely used method).

2. Sample Size Used for the Test Set and Data Provenance

  • LVivoRV: The exact sample size for the test set is not explicitly stated. The study mentions that examinations were collected over a period of 22 months retrospectively. It includes RV clips from 4CH and 4CH modified views that had 2-3 stable recorded beats.

    • Data Provenance: Retrospective, single-center study. The data was retrieved from PACS systems available on-site at the study center.
  • LVivo Bladder: 226 bladder images (113 pairs of transverse and longitudinal views) were included.

    • Data Provenance: Retrospective, single-center study. Examinations were retrieved from PACS systems available in Terem's clinic. The data was collected as part of routine abdominal examinations.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

  • LVivoRV:

    • FAC, EDA, ESA: Ground truth was established by two sonographers who manually traced the ED and ES frames. Their qualifications are not explicitly detailed beyond being "sonographers."
    • TAPSE: Ground truth was established by a sonographer (using M-Mode) and an echo cardiologist (using VVI).
    • S': Ground truth was established by a sonographer (using M-Mode) and an echo cardiologist (using VVI).
    • RV Strain: Ground truth was established by an echo cardiologist (using VVI).
    • Overall Qualification Level: Sonographers and echo cardiologists are qualified medical professionals routinely performing these measurements.
  • LVivo Bladder:

    • Ground truth was established by one expert sonographer who performed manual measurements of bladder volume (D1, D2, D3) from the trans and long views. The manual measurements were considered the reference/ground truth.

4. Adjudication Method for the Test Set

  • LVivoRV:

    • For EDA, ESA, and FAC, the LVivoRV's automated values were compared to the average of the values obtained manually by the two sonographers. This implies a form of consensus or averaging method.
    • For TAPSE and S', measurements from a sonographer (M-Mode) and an echo cardiologist (VVI) were used for comparison, but it's not explicitly stated if there was an adjudication for a single ground truth value when both measurements were available. However, the report compares LVivoRV to "manual measurement using M-Mode" and also notes correlation between VVI and M-Mode, suggesting separate comparisons rather than a combined adjudication.
  • LVivo Bladder:

    • The manual measurements performed by the single expert sonographer served as the ground truth. There was no explicit adjudication among multiple experts since only one expert provided the manual measurements for the validation.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

  • No MRMC comparative effectiveness study, evaluating human readers with and without AI assistance, was reported in this document. The studies were focused on the standalone performance of the LVivoRV and LVivo Bladder modules against manual measurements or established methods.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

  • Yes, standalone performance studies were done for both LVivoRV and LVivo Bladder.
    • LVivoRV: The algorithm's output was generated "by an automated batch processing after all data was ready and considered locked." This indicates the algorithm processed the data without human intervention to influence its initial measurements. Manual adjustments are available in the device, but the reported study focuses on the automated output.
    • LVivo Bladder: The algorithm was applied "by an automated batch processing on all pairs of trans and long views," also indicating standalone performance for the reported results. The device allows for manual caliper adjustments, but the validation appears to have used the automated result.

7. The Type of Ground Truth Used

  • LVivoRV: Expert consensus/manual measurements from qualified medical professionals (sonographers and echo cardiologists) based on conventional echocardiography methods (2D manual measurements, M-Mode, VVI).
  • LVivo Bladder: Expert manual measurements by an expert sonographer, considered the routinely used method for bladder volume measurement.

8. The Sample Size for the Training Set

  • The document does not explicitly state the sample size used for the training set for either LVivoRV or LVivo Bladder. It only describes the algorithms and their application to the test sets.

9. How the Ground Truth for the Training Set Was Established

  • The document does not provide details on how the ground truth for the training set was established. It describes the technology for LVivoRV as combining "image processing and Deep Learning Neural Network (NN)" and for LVivo Bladder as using "a combination of machine learning and active contour," implying the use of training data, but no specifics are given regarding its ground truth establishment.

{0}------------------------------------------------

June 23, 2020

Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

DiA Imaging Analysis Ltd % Mr. George J. Hattub Senior Staff Consultant Medicsense USA 291 Hillside Avenue SOMERSET MA 02726

Re: K200232

Trade/Device Name: LVivo Software Application Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: Class II Product Code: QIH Dated: May 13, 2020 Received: May 18, 2020

Dear Mr. Hattub:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for

{1}------------------------------------------------

devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

For

Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K200232

Device Name LVivo Software Application

Indications for Use (Describe)

L Vivo platform is intended for non-invasive processing of ultrasound images to detect, measure, and calculate relevant medical parameters of structures and function of patients with suspected disease.

Type of Use (Select one or both, as applicable)

X Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

K200232 510(k) Summary

Pursuant to CFR 807.92, the following 510(k) Summary is provided:

    1. (a) Submitter George J. Hattub Address: MedicSense, USA 291 Hillside Avenue Somerset, MA 02726 www.medicsense.com 1. (b) Manufacturer DiA Imaging Analysis Ltd Address: HaEnergia Street 77 Beer-Sheva, Israel 8470912 Mfg. Phone: Tel.: +972 77 7648318 Contact Person: Mrs. Michal Yaacobi Date: June 21, 2020 Automated Radiological Image Processing Software- classified as Class 2 2. Device & QIH, Regulation Number 21 CFR 892.2050 Classification Name: LVivo Software Application Predicate Device: K132544 TomTec-Arena for RV Evaluation 3. Reference K161382 & K130779- DiaCardio's LVivo Software Application Devices: K180995 GE Viscan for Bladder and Android 4. Description: The LVivo System analyzes echocardiographic patient examination DICOM movies for Global ejection fraction (EF) evaluation. EF is evaluated using two orthogonal planes, four-chamber (4CH) and two-chamber (2CH) views, to provide fully automated analyses of LV function from the echo examination movies. It also has the ability to measure strain and to evaluate the Right Ventricle and well as to measure the bladder. 5. DiA's LVivo platform is intended for non-invasive processing of ultrasound Intended Use: images to detect, measure, and calculate relevant medical parameters of structures and function of patients with suspected disease 6. Comparison of With respect to technology and intended use, DiA's LVivo Software Application is substantially equivalent to its predicate devices. Based upon Technological Characteristics: the outcomes from clinical trials, DiA believes that their device does not raise additional safety of efficacy concerns. At the end of this summary, a comparison table is provided. 7. A summary of the clinical evaluation is provided on the proceeding pages. Clinical Tests:

{4}------------------------------------------------

Submitted DevicePredicate Device
Features/CharacteristicsLVivoTomTec-Arena 1.0 K132544
Product CodeQIHLLZ
Intended UseCalculateand measurementof LV, RV andBladderCalculation andmeasurement LV, RV,fetal and abdomen
Indication for UseDiA's LVivo platform isintended for non-invasive processing ofultrasound images todetect, measure, andcalculate relevantmedical parameters ofstructures and functionof patients withsuspected disease.Indication for use of Tom-Tec- Arena software arediagnostic review,quantification and reporting ofcardiovascular, fetal, andabdominal structures andfunctions of patients withsuspected disease.
Automationyesyes
Manual Adjustmentyesyes
RV Calculation from2d3d
RV ED Volumenoyes
FACyesyes
ED Areayesno
EDVinoyes
ES Volnoyes
ES areayesno
EFnoyes
SVnoyes
TAPSEyesyes
Strain values free wallyesyes
Strain values septumnoyes
S primeyesno
View4ch viewMulti 2d view
510(k) #K200232K132544
Submitted DeviceReference Devices
Features/CharacteristicsLVivoLVivo K130779 & K161382
Product CodeQIHLLZ
Indications For UseDiA's LVivo platform isintended for non-invasive processing ofultrasound images todetect, measure, andcalculate relevantmedical parameters ofstructures and functionof patients withsuspected disease.DiaCardio's LVivo SoftwareApplication is intended fornon-invasive processing ofalready acquiredechocardiographic images inorder to detect, measure, andcalculate the left ventricularwall for left ventricularfunction evaluation. Thismeasurement can be used toassist the clinician in acardiac evaluation.
ModulesLVivo EF, LVivo SG,LVivo SAX, LVivo RV &LVivo BladderLVivo EF, LVivo SG, LVivoSAX
AutomationSameFully Automated
Bi plane EF evaluationYesYes
Simultaneous 2CHand 4CH evaluationYesYes
Off line EFevaluation usingDICOM clips of anyvondarYesYes
Automated ED andES frames selectionYesYes
Dynamic left ventricularYes. Frame byframe trackingYes. Frame by frametracking
Manual editing by usercapabilityMinimum of 7 borderpoints manipulation(dragging) and onlinecontour presentation.Possible to apply toany frame in the clip.Border detection isrecalculation isapplied to the entireclip.Yes. 7 border pointsmanipulation (dragging) andonline contour presentation.Possible to apply to anyframe in the clip.Border detection isrecalculation is applied tothe entire clip.
Visually confirm EFYesYes
Automated rejectionof false resultsYesYes
Volume calculation bySimson's method ofdiscsYesYes
Volume curvePresentationYesYes
EF results presentationDisplaying full clip withborder tracking. Andtable with results foreach cycle for selectedED & ES frames foreach beat.Displaying full clip with bordertracking. And table withresults for each cycle forselected ED & ES frames foreach beat.
Enables presentation EFresults for different cycleYesYes
AlgorithmImage segmentation forborder detectionFor the RV- DeepLearning TechnologyImage segmentation forborder detectionFrom image processing
Calculation speedLess than 1s per cyclefor biplane evaluationLess than 1s per cycle forbiplane evaluation
Capability or a part of abigger package (device)for LV functionevaluationYesYes
Segmental LongitudinalStrain MeasureYesYes
Global LongitudinalStrain MeasureYesYes
Segmental wall motionevaluationYesYes
RV EvaluationYesNo
Bladder MeasurementYesNo
Operating SystemWindows/Linux (withAndroid MobileOption for LVivo EF)Windows
510(k) #K200232K130779 & K161382

{5}------------------------------------------------

{6}------------------------------------------------

{7}------------------------------------------------

Submitted DeviceReference Device
Features/CharacteristicsLVivoViscan K180995
Indication for UseDiA's LVivo platform isintended for non-invasive processing ofultrasound images todetect, measure, andcalculate relevantmedical parameters ofstructures and functionof patients withsuspected disease.Vscan Extend is ageneral-purposediagnostic ultrasoundimaging system for useby qualified and trainedhealthcare professionalsenabling visualization andmeasurement ofanatomical structures andfluid. The specific clinicalapplications and examtimes include Cardiac,Abdominal and Urology]
Automationyesyes
Calculate From Borderyesyes
Display Calipersyesyes
Allow User to AdjustCalibrationyesyes
Present Volumeyesyes
Present Diameteryesyes
Image processingalgorithmyesyes
FrameTrans and LongViewTrans and Long View
Operating SystemWindows,Android Option(With LVivo EF)Android
510(k) #K200232K180995

{8}------------------------------------------------

Clinical Summary -

LVivoRV

Technology

The LVivo RV, a part of the LVivo platform, is a decision support system that uses 2D echocardiographic examinations to automatically evaluate the Right Ventricular (RV) function from 4 chamber apical views (focused or modified). The LVivoRV utilizes clips in DICOM format from the apical 4CH view, focused or modified, as an input without any additional user input (such as selection of a starting frame or manual starting points as required by Epsilon). The Algorithm combines image processing and Deep Learning Neural Network (NN) for the RV analysis. The endocardial boundaries and the location of the anulus of the tricuspid valve are identified by the NN model. These boundaries are further enhanced and tracked using image processing methods that are already established in other approved modules of the LVivo Platform. The algorithm provides measurements of RV size and function: ESA, EDA, FAC, TAPSE S' and Free Wall Strain. There are 3D technologies available for RV evaluation, the most known is Image-Arena Platform by TomTec

Protocol:

In this presented clinical validation, the LVivoRV output for RV function was compared with conventional methods that are used in echocardiography for RV function evaluation (2D manual measurements by technicians and when relevant to M-Mode, doppler and VVI).

Study: Retrospective, single center study.

    1. All examinations were retrospectively retrieved from the PACS systems available on site. Examinations of patients referred to the echo unit who have had a standard echo examination were collected over a period of 22 months retrospectively according to the inclusion / exclusion criteria
    1. All examinations were performed as part of the routine and according to the ASE guidelines, using available ultrasound systems (EPIQ, Affinity, IE33, & Philips)
    1. Only RV Clips from 4CH and 4CH modified views that have 2-3 stable recorded beats were included. Also, the dataset included only examinations in which measurements of TAPSE and S' were reported
    1. The selected RV clips, were anonymized and saved separately without the patient's details according to the patient number
    1. conventional methods were used to evaluate RV function:
    • EDA and ESA were measured by 2 sonographers by selection and manually tracing the ED and ES frames, and FAC results were calculated from the EDA and the ESA.
    • TAPSE was measured by a sonographer using M-Mode and by an echo cardiologist using VVI.
    • S' was measured by a sonographer using M-Mode and by an echo cardiologist using VVI.
    • -RV Strain was measured by echo cardiologist using VVI
    1. LVivoRV evaluation was done by an automated batch processing after all data was ready and considered locked
    1. Manual measurements were compared to automated measurements according to statistical plan

{9}------------------------------------------------

Study Objectives

  • Primary endpoint was to compare LVivoRV measurement of FAC to manual FAC measurement
  • Secondary endpoints were: -
    • to compare LVivoRV measurements to the manual measurements of EDA, ESA, TAPSE, S' and FREE WALL STRAIN
    • To compare RV function by visual assessment to the categorized result from FAC, TAPSE, S' and STRAIN and
    • To evaluate inter and intra observer variability.

Acceptance of the statistical data was 75% correlation between FAC by LVivoRV and the same measurements performed manually by sonographers. This value is based on statistical data reported by FDA cleared system for semi-automated RV function evaluation (Echolnsight by Epsilon)

Results and Conclusions

The results showed that the primary end point was successfully met with a very good correlation of FAC between LVivoRV and manual measurements (r=0.79, p<0.0001). RV Global function is evaluated by measuring EDA and calculating FAC accordingly. The values obtained automatically by the LVivoRV were compared to the average of the values obtained manually by the 2 sonographers.

The correlations between the sonographer's average for EDA and ESA and the LVivoRV EDA and ESA were excellent with r=0.92 and r=0.93 respectively (p<0.0001).

For the TAPSE, we saw that the correlation of the manual measurement using M-Mode was 0.62, similarly to the correlation found between VVI and M-Mode that was r=0.66. With these correlations, the LVivoRV performance is close to "real life" performance of VVI in compare to M-mode. When looking at the TAPSE correlation without 5 outlier cases, we got r=0.72. The Free Wall Strain correlation between 2D VVI and LVivoRV was evaluated and found to be a positive correlation of R=0.6. When omitting 6 outlier cases from the correlation analysis, we obtained r= 0.78.

The inter-observer reliability between readers within the same group was also tested. Interobserver reliability between sonographers for EDA. ESA and FAC was 0.85, 0.9 and 0.77 respectively (p<0.0001).

10 subjects were randomly chosen for the evaluation of the intra-observer reliability. The intraobserver reliability was evaluated for the 2 sonographers that reported measurable values using 2D and one physician that reported measurable values with VVI. The intra-observer correlation for FAC for sonographer1 was 0.95 [95%Cl: 0.80-0.98] and for sonographer 2 it was 0.94 [95%Cl: 0.72-0.98]. The intra-observer correlation for TAPSE for the physician used the VVI was 0.85 [95%CI: 0.40-0.96]

It is known that the conventional RV assessment is not perfect and has its own intermeasurement variabilityF, thus we don't expect a perfect agreement between the LVivoRV and currently used techniques. However, the differences between RV parameters obtained using the LVivoRV software and the conventional measurements were similar to the variability of the current conventional methods.

Overall, the correlation of the LVivoRV to acceptable used methods for RV evaluation was very good. The LVivoRV can use as a decision support system.

{10}------------------------------------------------

LVivoBladder

Technology

The LVivo Bladder, a part of the LVivo platform, is a decision support system for automated Bladder Volume evaluation from ultrasound examinations.

The LVivo Bladder utilizes images in DICOM format. It automatically Identifies the bladder borders on the ultrasound scan and locates the calipers for measuring the bladder's 3 dimensions on the longitudinal and sagittal views, with a corresponding value for the 3 dimensions and for the bladder volume. The calipers can be changed by the sonographer to make adjustments if required.

This application automatically measures bladder volume by segmenting bladder contours from sagittal and transverse ultrasound views using a combination of machine learning and active contour

Protocol:

In this clinical validation study the automated bladder volume measurements obtained by DiA's automated tool were compared to manual measurements obtained by an expert.

Study: Retrospective, single center study.

  • 226 bladder images (113 pairs) were included 1.
    1. All examinations were retrospectively retrieved from the PACS systems available in Terem's clinic. Since the data is retrospectively collected, all examinations are acquired as part of routine abdominal examinations, by qualified sonographers using available ultrasound systems (Siemens, Mindray). Collected data will be in DICOM format.
    1. Since Abdominal tests do not necessarily include bladder volume measurements, the first step of data collection was filtering of the abdominal tests to include only cases where bladder volume is reported.
    1. Assuming ~30% of tests with bladder volume examinations will not include full sets of images (as described in item 2 under 'exclusion criteria), 80 consecutive bladder ultrasound examinations with bladder volume higher than 200ml and 80 consecutive bladder ultrasound examinations with bladder volume less than 200ml were extracted from the database
    1. An expert sonographer reviewed extracted cases to include only cases with full sets of images (as described in the inclusion criteria)
  • All images were anonymized by the study coordinator and were sent to the sponsor via 6. Dropbox. Patient details including demographic details were collected anonymously and reported as well in the excel sheet.
    1. Since some of the tests included more than one long view image and more than one trans view image and since these images do not represent the images that were used to measure the volume in practice, the pairs of images that were selected for the validation were the ones that were most similar to the ones that appear in the biplane view from which the original reported measurements were obtained. The rest of the pairs were used for verification
    1. In tests that included post voiding good quality interpretable images in addition to prevoiding images, both were used for the validation
    1. After obtaining the required data set for the validation test, all pairs (trans and long views) were sent to an expert sonographer for manual measurements of the bladder volume, to acquire the 3 dimensions of the bladder in long and trans views (i.e. D1, D2 and D3). For standardization purposes the dimension measured by the expert in the long view was always marked from top right to bottom left of the image.
    1. The LVivo Bladder algorithm was applied by an automated batch processing on all pairs of trans and long views. Here as well the algorithm was configured to draw the dimension in the long view from top right to bottom left of the image, for standardization purposes

{11}------------------------------------------------

    1. The calculated volume by the manual measurements was compared to the automated calculated volume by LVivo Bladder.

Study Objectives

The clinical validation objective was to compare LVivo Bladder measurement of bladder volume to bladder volume by manual tracing.

Post Voiding Residual volume of 200mL indicates inadequate emptying and may be associated with catheter insertion. In Point of care settings, it is specifically important to evaluate whether the residual volume is lower or higher than this 200ml threshold, thus the method that was used to compare the automated and the manual methods were based on categorial evaluation of more/less than 200mL.

We expected to have a good agreement between the automated and the manual method (which is an accepted method for BV measurement) based on 200ml threshold, with kappa of at least 0.61 which is considered substantial agreement according to accepted Kappa interoperations.

Results and Conclusions

The results showed that the primary end point was successfully met- excellent agreement between methods was obtained by differentiating between post voiding volume, which is considered to be large (200ml) and is indicative for catheter placement, with excellent Kappa of 0.84, as well as very high agreement (0.93) and high sensitivity and specificity (100 and 80 respectively). Excellent results were also obtained comparing automated bladder volume calculation by LVivo Bladder to volume calculated by manual tracing which is the routinely used method with very high correlation (r=0.94).

The LVivo Bladder can easily, accurately and automatically measure the bladder volume from 2D ultrasound. Such a tool addresses the need in rapid and accurate evaluation in the POC environment. The LVivo Bladder's algorithm was found to be very robust and accurate in automatically measure bladder volume.

§ 892.2050 Medical image management and processing system.

(a)
Identification. A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.(b)
Classification. Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).