Search Results
Found 249 results
510(k) Data Aggregation
(156 days)
IPF
Ask a specific question about this device
(379 days)
IPF
EMS is used for:
- Relaxation of muscle spasms
- Prevention or retardation of disuse atrophy
- Increasing local blood circulation
- Muscle re-education
- Maintaining or increasing range of motion
- Immediate postsurgical stimulation of calf muscles to prevent venous thrombosis
TENS is used for:
- Symptomatic relief and management of chronic, intractable pain
- Post-surgical acute pain
- Post-trauma acute pain
EVE Synergy has Synergy handpiece (1EA), Pro handpiece (2EA), Micro cable, Tip (Ball type, T type, COMB type) and Power adaptor & cable.
EVE Synergy has Synergy mode and Pro mode.
Synergy mode has radio frequency and low frequency functions using a synergy handpiece. The Synergy handpiece has a temperature sensor, so the output is automatically turned off when the skin surface temperature exceeds 43℃.
Pro mode has low frequency function that uses the Pro handpiece with tip.
It should be noted that the device has radio frequency (RF Mode) indicated for relief of minor muscle aches and pain, relief of muscle spasm, temporary improvement of local blood circulation. This RF mode was previously reviewed by the FDA under K212253 and was cleared under regulation 21 CFR 878.4400 (Product code PBX). In the subject submission, there are no changes being proposed related to the RF Mode, and, hence, the review of the RF Mode is not within the scope of the subject 510(k) submission.
The provided text is a 510(k) clearance letter and an accompanying 510(k) summary for the EVE Synergy (EVE-20M) device. This document focuses on demonstrating substantial equivalence to previously cleared predicate devices, rather than proving the device meets specific acceptance criteria through a comparative effectiveness study involving AI assistance or human-in-the-loop performance.
Therefore, many of the requested details regarding acceptance criteria for AI performance, specific study designs (MRMC, standalone), ground truth establishment for AI training/test sets, and sample sizes for these, are not present in this type of regulatory submission. This document primarily describes the physical and electrical characteristics of the device and compares them to predicate devices, along with standard electrical safety, EMC, and software validation testing.
Here's a breakdown of the requested information based on the provided text, highlighting what is present and what is not:
1. A table of acceptance criteria and the reported device performance
The document does not present "acceptance criteria" in the traditional sense of a specific performance metric (like accuracy, sensitivity, specificity) for a clinical outcome or an AI model's performance. Instead, it demonstrates compliance with safety and performance standards and similarity to predicate devices.
The "acceptance criteria" are implicitly met by showing:
- Compliance with various IEC standards (IEC 60601-1, IEC 60601-1-2, IEC 60601-2-10, IEC 62304) for electrical safety, EMC, and software.
- The device's technical specifications (e.g., output voltage, current, pulse width, frequency, power density) are comparable to or within the safe limits established by the predicate devices.
Table of Performance (extracted from comparison tables):
Characteristic | Acceptance Criteria (Implied by Predicate & Standards) | EVE Synergy (EVE-20M) Reported Performance |
---|---|---|
General Compliance | Compliance with IEC 60601-1, IEC 60601-1-2, IEC 60601-2-10, ISO 14971, IEC 60601-1-6, IEC 62304 | Passed all testing in accordance with internal requirements, national standards, and international standards. |
Device Configuration | Similar design, components, and method of current isolation to predicates. | Consists of AC/DC power supply, controller, LCD touch screen; Supports Synergy and Pro handpieces; Independent transformer isolation; Type BF Electrical Type. |
Electrical Safety | Patient Leakage Current – Normal Condition: |
Ask a specific question about this device
(257 days)
IPF
For Prescription and Home Use by prescription from a medical professional:
The TrainFES advanced is a neuromuscular electrical stimulator indicated for use under medical supervision for adjunctive therapy in the treatment of medical diseases and conditions.
As a powered muscle stimulator, TrainFES advanced is indicated for the following conditions:
- Relaxation of muscle spasms
- Prevention or retardation of disuse atrophy
- increasing local blood circulation
- immediate post-surgical stimulation of calf muscles to prevent venous thrombosis
- Maintaining or increasing range of motion
- Muscle re-education
As an external functional neuromuscular stimulator (FES), TrainFES Advanced is indicated for the following conditions:
- Helps to relearn voluntary motor functions of the extremities.
TrainFES' Advanced intended population is anybody aged 22 or over.
Environments of use: TrainFES Advanced devices can be used by both therapists and patients, in the clinic, hospitals or at home.
Platform: TrainFES is a battery-powered, wireless device, accessible through software.
TrainFES Advanced is a portable functional electrostimulator with 6 channels designed for use in the clinics and hospitals by the medical professionals as well as in the home environment by the patient. This device generates electrical impulses to stimulate the musculature of paralyzed seqments and facilitate both the relearning of movement and neuromodulation of tone.
TrainFES Advanced is a battery-powered, wireless device, configurable from the TRAINFES App, available for Smartphone and Tablets, which allows you to adjust different parameters and follow a training plan from your smartphone. Session settings can be retrieved from the PC or Cloud.
The provided document describes the FDA 510(k) premarket notification for the "TrainFES Advanced" device, a neuromuscular electrical stimulator. The purpose of this submission is to demonstrate substantial equivalence to a predicate device (Stella BIO, K210002).
Here's an analysis of the acceptance criteria and study information provided:
1. Table of Acceptance Criteria and Reported Device Performance:
The document doesn't explicitly state "acceptance criteria" for the device's performance in a quantitative manner (e.g., target accuracy, sensitivity, specificity). Instead, the substantial equivalence justification relies on demonstrating that the TrainFES Advanced device performs similarly to or meets the safety and effectiveness standards of the predicate device, K210002.
The table below summarizes the comparison of key technical characteristics between the TrainFES Advanced and its predicate, Stella BIO, highlighting where "performance" is discussed in terms of meeting relevant standards or being considered equivalent. The reported "performance" for TrainFES Advanced is intrinsically linked to its compliance with these standards and the assertion that differences do not raise new safety or effectiveness concerns.
Characteristic / Specification | Acceptance Criteria (implied by predicate comparison) | Reported Device Performance (TrainFES Advanced) |
---|---|---|
Basic Device Characteristics | ||
Classification | Class II | Class II |
Prescription/OTC Use | Prescription and Home Use | Prescription and Home Use |
Environment of Use | Clinics, hospitals, and home | Clinics, hospitals, and home |
Indications for Use | Similar to predicate (specific conditions) | Similar to predicate, with specific conditions listed, and the functions of powered muscle stimulator and external functional neuromuscular stimulator are exactly the same |
Power Source | Battery-powered, compliant with IEC 62133 | Battery: Li-Ion 3.7V (4000mAh), compliant with IEC 62133 |
Method of Line Current Isolation | N/A (Battery) | N/A (Battery) |
Patient leakage current (Normal) |
Ask a specific question about this device
(318 days)
IPF
Quantum Mitohormesis Medical Device is intended for:
- Stimulating healthy limb muscles to improve and facilitate muscle performance.
- Preventing or retardation of disuse atrophy of muscles.
Quantum Mitohormesis (QMT) is a medical device that uses unique pulsed electromagnetic fields for muscle therapy. Users (adults ≥ 22 years old) place one group of muscles in the lower limbs into the field applicator of QMT device, which for a defined duration generates specific electromagnetic fields that stimulates healthy muscles to improve and facilitate muscle performance, and prevents and retards muscle disuse atrophy.
QMT is applied directly to muscles (without direct contact) to stimulate muscle activity. In order to achieve optimized treatment outcome, QMT is applied to muscles on a weekly basis, once or twice a week, with each treatment application lasting a brief 10-minutes.
The device consists of the following high-level components:
- Base unit, which includes signal generator, power amplifier, and micro-controller unit.
- Field applicator, which includes electro-magnetic coils generating specific magnetic pulses.
- User Interface Panel, which allows users to operate the device and display necessary device information.
- Signal cord.
- Height and tilt adjustment knobs, which allow users to adjust the height and tilt of field applicator; and
The provided text does not contain information about the acceptance criteria and the study that proves the device meets the acceptance criteria.
The document is a 510(k) premarket notification letter from the FDA to QuantumTX Pte. Ltd. regarding their Quantum Mitohormesis (QMT) device. It confirms the device's substantial equivalence to legally marketed predicate devices, allowing its marketing.
The document discusses:
- Regulatory information: Device classification, regulation numbers, and general controls.
- Device description: How the QMT device works (pulsed electromagnetic fields for muscle therapy) and its components.
- Intended Use/Indications for Use: Stimulating healthy limb muscles to improve and facilitate muscle performance, and preventing or retarding disuse atrophy of muscles.
- Comparison to Predicate Devices: A detailed table comparing the QMT device with two predicate devices (Storz Medical MAGNETOLITH Muscle Stimulator and Bemer Classic & Pro) across various parameters (device name, manufacturer, intended use, principle of action, technical specifications, etc.). This comparison aims to demonstrate substantial equivalence, not to prove performance against specific acceptance criteria.
- Summary of Non-Clinical Testing: States that the QMT device passed various non-clinical performance tests (biocompatibility, electrical safety, electromagnetic disturbance/immunity, and software verification/validation) in accordance with relevant ISO and IEC standards. These are safety and engineering compliance tests, not performance studies against specific clinical acceptance criteria for the intended use.
- Summary of Clinical Testing: Explicitly states, "There is no clinical data submitted with this submission."
Therefore, I cannot provide the requested information, particularly the table of acceptance criteria and reported device performance, the sample sizes, expert details, adjudication methods, MRMC studies, standalone performance, ground truth types, or training set details because this information is not present in the provided text. The submission relies solely on non-clinical testing and substantial equivalence to predicate devices, not on a clinical performance study.
Ask a specific question about this device
(182 days)
IPF
EMS is used for:
- · Relaxation of muscle spasms
- · Prevention or retardation of disuse atrophy
- · Increasing local blood circulation
- · Muscle re-education
- · Maintaining or increasing range of motion
- · Immediate postsurgical stimulation of calf muscles to prevent venous thrombosis
eMVFit has MULTI handpiece, EMS handpiece, patient switch, cradle, distilled water injection hose & funnel, key switch and power cable.
eMVFit has a user interface including an LCD touch screen and has MULTI mode and EMS mode. MULTI mode has RF and EMS functions that use the MULTI handpiece.
In the case of RF, the target temperature can be set and a vacuum can be used together, and a cooling function can be used as a safety system to prevent burns on the skin surface.
In addition, the MULTI handpiece has a temperature sensor, so the output is automatically turned off when the skin surface temperature exceeds 43 ℃.
EMS mode has an EMS function that uses the EMS handpiece.
The provided text is a 510(k) Summary for the eMVFit (MVF-10M) device, a powered muscle stimulator. It focuses on demonstrating substantial equivalence to predicate devices through various comparisons and non-clinical performance data.
However, the document does not contain information on acceptance criteria for a study demonstrating device performance in terms of clinical outcomes, nor does it describe a study involving human subjects or expert readers to prove the device meets specific acceptance criteria for diagnostic or clinical effectiveness.
The document primarily details:
- Device Description and Indications for Use: What the device is and what it's used for (muscle spasms, disuse atrophy, circulation, re-education, range of motion, DVT prevention).
- Predicate Device Comparison: A detailed table comparing the eMVFit's technical specifications (electrical parameters, dimensions, operation modes) with three predicate devices (INMODE LTD.'s Evolve System with T3 and Tone Applicators, and TimeWaver Production GmbH's TimeWaver Frequency).
- Substantial Equivalence Discussion: Explains that differences between the subject device and predicates do not raise new questions of safety or effectiveness, often citing compliance with IEC standards.
- Non-Clinical (Bench) Performance Data: Lists a series of standards (IEC 60601-1, IEC 60601-1-2, IEC 60601-2-10, ISO 14971, IEC 60601-1-6, ISO 10993-10, ISO 10993-23, IEC 62304) to which the device was tested to demonstrate basic safety, essential performance, EMC, risk management, usability, biocompatibility, and software functions. It states the device passed all testing.
Therefore, I cannot fulfill your request for: a table of acceptance criteria and reported device performance (in the context of clinical outcomes), sample size, data provenance, number/qualifications of experts, adjudication methods, MRMC studies, standalone performance, or ground truth establishment based on the provided text.
The document describes technical compliance and safety testing against engineering standards, not direct clinical performance or effectiveness studies with human data or expert evaluations. The "performance data" referred to is "Non-Clinical (Bench) Performance Data," meaning lab tests, not clinical trials.
The FDA's 510(k) pathway often relies on demonstrating substantial equivalence to a legally marketed predicate device, rather than requiring new clinical efficacy studies, especially for devices with well-understood mechanisms of action like muscle stimulators. The information provided aligns with this regulatory pathway.
Ask a specific question about this device
(201 days)
IPF
BTL-899MS is indicated to be used for:
· Improvement of abdominal tone, strengthening of the abdominal muscles, development of firmer abdomen.
- · Strenathening, toning and firming of buttocks, thighs and calves.
- · Improvement of muscle tone and firmness, for strengthening muscles in arms.
The BTL-899MS device is intended to be used under medical supervision for adjunctive therapy for the treatment of medical diseases and conditions.
The BTL-899MS device is indicated for use in stimulating neuromuscular tissue for bulk muscle excitation in the legs or arms for rehabilitative purposes.
Indications for Use for Muscle Stimulators:
- · Relaxation of muscle spasms
- · Prevention or retardation of disuse atrophy
- Increasing local blood circulation
- · Muscle re-education
- · Immediate post-surgical stimulation of calf muscles to prevent venous thrombosis
- · Maintaining or increasing range of motion
The BTL-899MS is a non-invasive therapeutic device. The device is comprised of a main unit and applicators that deliver electromagnetic energy to the targeted tissue. The device's two outputs enable hands-free simultaneous treatment by two applicators.
The application of electromagnetic energy to the patient's feeling of magnetic stimulation of underlying tissue accompanied by a pleasant feeling of warming in the application area, which may contribute to increased local blood circulation and help the patient to relax and make the treatment more comfortable.
The BTL-899MS is equipped with a large color touch-screen that facilitates the use of the device. The on-screen information guides the user step-by-step through the entire therapy procedure. The therapeutic parameters are set using the touch screen of the device. During the therapy, the device displays information about the applied therapy type, remaining therapy time and main therapy parameters on the screen.
Based on the provided FDA 510(k) summary for the BTL-899MS device (K240234), here is an analysis of its acceptance criteria and the study proving it meets them:
Important Note: The provided document is a 510(k) summary, which focuses on demonstrating substantial equivalence to legally marketed predicate devices rather than providing detailed clinical study results for a novel device. Therefore, the information regarding acceptance criteria, specific performance metrics, and detailed study methodologies (like sample size for test sets, number of experts, adjudication methods, MRMC studies, standalone performance, and ground truth establishment for novel AI/software devices) is limited or absent. This document primarily outlines compliance with safety and performance standards relevant to a medical device's physical and electrical characteristics.
1. Table of Acceptance Criteria and Reported Device Performance
For this specific device, the "acceptance criteria" are primarily related to meeting established electrical safety, electromagnetic compatibility, usability, risk management, and biocompatibility standards, and demonstrating substantial equivalence to predicate devices in terms of technological characteristics and intended use. There isn't a table of quantitative performance metrics for disease detection or diagnostic accuracy in this submission because it's a physical therapy device, not an AI diagnostic tool.
Acceptance Criteria Category | Specific Standard/Requirement | Reported Device Performance/Compliance |
---|---|---|
Electrical Safety | IEC 60601-1: General requirements for basic safety and essential performance | The BTL-899MS device has been "thoroughly evaluated for electrical safety" and "found to comply with the applicable medical device safety standards." |
Electromagnetic Compatibility (EMC) | IEC 60601-1-2: EMC – Requirements and tests | Complies with the standard. |
Usability | IEC 60601-1-6: General requirements for basic safety and essential performance – Collateral standard: Usability | Complies with the standard. |
Software Life Cycle Processes | IEC 62304: Medical device software – Software life cycle processes | Complies with the standard. |
Risk Management | ISO 14971: Medical devices – Application of risk management to medical devices | Complies with the standard. |
Biocompatibility | ISO 10993-1: Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process | Complies with the standard. (Specific tests mentioned: ISO 10993-5 for in vitro cytotoxicity and ISO 10993-10 for irritation and skin sensitization). |
Substantial Equivalence | Comparison to predicate devices (K203488, K200382, K203710) in terms of intended use, technological characteristics, and safety/effectiveness questions. | "The BTL-899MS device has the same intended use and identical technological characteristics and principles of operation to its predicate devices." Differences "do not raise any new types of safety or effectiveness questions." Parameters (magnetic field intensities, pulse durations, magnetic energy densities) are "within the range of values" of primary/third predicates or "equivalent" to the second predicate. Therapy duration differences are addressed by repeated use for comparable energy delivery. Added warming feature is already cleared in a predicate. |
2. Sample Size Used for the Test Set and Data Provenance
The document explicitly states "Clinical Testing: Not applicable" on page 6. This indicates that a clinical study with real patient data (test set) for performance evaluation, as would be expected for an AI/software device, was not conducted or deemed necessary for this 510(k) submission. The evaluation was based on non-clinical testing (bench testing) and comparison to predicate devices, focusing on the device's physical and electrical characteristics and mechanism of action.
Therefore, questions regarding sample size, data provenance (country, retrospective/prospective), experts, and adjudication methods for a test set are not applicable in this context.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts
Not applicable, as no clinical test set for performance evaluation was conducted.
4. Adjudication Method (e.g., 2+1, 3+1, none) for the Test Set
Not applicable, as no clinical test set for performance evaluation was conducted.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and What Was the Effect Size of How Much Human Readers Improve with AI vs Without AI Assistance
Not applicable. This is a powered muscle stimulator, not an AI diagnostic device intended to assist human readers (e.g., radiologists interpreting images).
6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done
Not applicable. This is a physical device, not an algorithm, and its functionality is not standalone in the sense of an AI model's performance.
7. The Type of Ground Truth Used
The "ground truth" for this medical device's acceptance is based on:
- Compliance with established international and national technical standards (e.g., IEC 60601-1, ISO 14971) for safety, EMC, usability, and biocompatibility.
- Demonstrated substantial equivalence to previously cleared predicate devices, meaning its intended use and technological characteristics are similar enough that it does not raise new questions of safety or effectiveness. The "ground truth" for this equivalence often relies on the established safety and efficacy of the predicate devices themselves, backed by their own clearances or long-standing market presence.
8. The Sample Size for the Training Set
Not applicable. This device is not an AI/ML model that requires a training set of data.
9. How the Ground Truth for the Training Set Was Established
Not applicable, as there is no training set for this device.
Ask a specific question about this device
(311 days)
IPF
The MYOTouch Muscle Stimulator is indicated for use as an adjunct in the treatment of high and low anorectic malformations by helping with the identification of striated muscles to be used in anal reconstructions. The MYOTouch device is intended for use by trained personnel who are competent to apply the appropriate stimuli.
MYOTouch is a hand-held device designed to be used for surgical correction of neonatal anorectal malformations. MYOTouch is intended for prescription use. The device is a 9V battery-powered, one-channel muscle stimulator. The device is supplied with a Sterile Stylus, which connects to the control unit by cable and plug. The MYOTouch muscle stimulator has been specially designed to meet the requirements of pediatric surgeons who need a reliable and easy to use muscle locating stimulator when carrying out the Pena, "Pull-through", or posterior sagittal anorectoplasty (PSARP) anorectal reconstruction technique. A trained professional or pediatric surgeon must use the MYOTouch device. The pediatric surgeon uses manual, push-button controls to control the electrical stimulation level. The unit is intended for prescription use and is designed with simplicity and ease of use. It has two preset modes available.
I am sorry, but the provided text is a 510(k) premarket notification for a medical device called the MYOTouch Muscle Stimulator. It describes the device, its intended use, and compares it to a predicate device to establish substantial equivalence.
However, the document explicitly states in Section VIII, "No clinical tests were submitted to support this 510(k)."
Therefore, based on the information provided, I cannot answer your request regarding the acceptance criteria and the study that proves the device meets those criteria, as no such study is presented in this document. The submission relies on demonstrating substantial equivalence to a legally marketed predicate device through technical comparisons and compliance with recognized standards, rather than direct performance studies against acceptance criteria.
Ask a specific question about this device
(248 days)
IPF
PA&PR series:
The PA&PR series are a neuromuscular electrical stimulator indical supervision for adjunctive therapy in the treatment of medical diseases and conditions.
As a powered muscle stimulator, PA&PR series are indicated for the following conditions:
- · Relaxation of muscle spasms,
- · Prevention or retardation of disuse atrophy,
- · Increasing local blood circulation,
- · Immediate post-surgical stimulation of calf muscles to prevent venous thrombosis,
- · Maintaining or increasing range of motion,
- · Muscle re-education.
Environments of Use: Clinics, hospital. Patient population: Adults
S&Q series:
The S&O series are a neuromuscular electrical stimulator indical supervision for adjunctive therapy in the treatment of medical diseases and conditions.
As a powered muscle stimulator, S&Q series are indicated for the following conditions: - · Relaxation of muscle spasms,
- · Prevention or retardation of disuse atrophy,
- · Increasing local blood circulation,
- · Immediate post-surgical stimulation of calf muscles to prevent venous thrombosis,
- · Maintaining or increasing range of motion,
- · Muscle re-education.
As a transcutaneous electrical nerve stimulator for pain relief, S&Q series are indicated for the following conditions:
· Symptomatic relief and management of chronic (long-term), intractable pain, - · Adjunctive treatment in the management of post-surgical pain and post traumatic acute pain.
Environments of Use: Clinics, hospital and home environments.
Patient population: Adults
Subject device have four series, which are PA series, PR series, Q series and S series. These four series are all electrical stimulating devices.
This document is an FDA 510(k) clearance letter for the "Stimulation System" manufactured by Edan Instruments, Inc. It asserts substantial equivalence to a predicate device, STIWELL med4 (K080950).
The provided text does not contain acceptance criteria for an AI-powered device, nor does it describe a study proving such a device meets acceptance criteria.
Instead, this document is a regulatory submission for a medical device (neuromuscular and transcutaneous electrical nerve stimulators) that appears to be hardware-based, with "software/firmware/microprocessor control" mentioned generally, but not as the primary component being evaluated for AI performance. The document focuses heavily on bench testing (electrical safety, EMC, performance to standards) and comparison to a predicate device's physical and electrical specifications, not on AI performance metrics like accuracy, sensitivity, or specificity.
Therefore, I cannot fulfill your request for:
- A table of acceptance criteria and reported device performance for an AI device.
- Sample sizes, data provenance, ground truth establishment, or expert involvement related to AI performance.
- MRMC studies or standalone AI performance.
The document's "Performance Data" section explicitly states: "Clinical data: Not applicable." This further confirms that no clinical studies (which would be necessary to prove performance of an AI system in a clinical context) were conducted or presented for this submission.
In summary, the provided text does not describe an AI medical device or its performance criteria/study.
Ask a specific question about this device
(99 days)
IPF
The accufit is indicated for use for Relaxation of muscle spasms, Prevention or retardation of disuse atrophy, Increase local blood circulation, Muscle re-education, Maintaining or increasing range of motion, Immediate post-surgical stimulation of calf muscles to prevent venous thrombosis, Improvement of abdominal tone, for strengthening of the abdominal muscles, for development of firmer abdomen, and Improvement of muscle tone and firmness, for strengthening muscles in arms, thighs and buttocks areas.
The principle of operation of the accufit device is to provide direct electrical stimulation of muscles which produces muscle contractions in and around the area of the treatment electrodes. The accufit generates electrical stimulation to contract the muscles to achieve the desired intended use using unique IntelliPhase waveforms and IntelliSTIM electrodes. It offers two unique IntelliPhase waveforms, Biphasic and Interferential (4P), that are designed to optimize muscle re-education and activation for patients. The four IntelliPhase waveform protocols are twist, hold, grip, and tap. Each protocol may be used for 15, 30, or 45 minutes. Not only are there four IntelliPhase protocols, but there is also a fully automated protocol (IntelliCycle) that integrates all the above listed waveforms in a comprehensive, interlaced series. This engages all the targeted muscles in a continuous program to maximize muscle effects with the touch of a button. accufit is both the proprietary-trade name and generic name. accufit is a supplement to existing treatments and does not replace any traditional treatment or therapy. The system console is the heart of the accufit and contains the Touch LCD/GUI, electrodes, and power supply module. The main console also includes a key switch used to turn the power on and off, and an emergency stop push button that quickly de-energizes the system in emergency situations. There are 4 casters in the console base that can be used when moving the system.
The accufit device is a Powered Muscle Stimulator (Product Codes IPF, NGX) indicated for various muscle-related therapies, including:
- Relaxation of muscle spasms
- Prevention or retardation of disuse atrophy
- Increase local blood circulation
- Muscle re-education
- Maintaining or increasing range of motion
- Immediate post-surgical stimulation of calf muscles to prevent venous thrombosis
- Improvement of abdominal tone, for strengthening of the abdominal muscles, for development of firmer abdomen.
- Improvement of muscle tone and firmness, for strengthening muscles in arms, thighs and buttocks areas.
The device operates by providing direct electrical stimulation to muscles, causing contractions using unique IntelliPhase waveforms (Biphasic and Interferential (4P)) and IntelliSTIM electrodes. It offers four IntelliPhase waveform protocols (twist, hold, grip, and tap), each usable for 15, 30, or 45 minutes, as well as a fully automated IntelliCycle protocol.
1. Table of Acceptance Criteria and Reported Device Performance
The provided document details a clinical study primarily focused on the improvement of muscle tone and firmness, specifically for strengthening muscles in the arms. While acceptance criteria in a quantitative sense (e.g., a specific percentage increase in strength) are not explicitly stated, the study's overall finding that "most patients saw improvement in muscle strength" serves as the reported device performance against an implied criterion of demonstrating clinical effectiveness for its new indication.
Given the available information, here is a summary related to the clinical study and its findings:
Acceptance Criteria (Implied) | Reported Device Performance |
---|---|
Device effectively strengthens arm muscles. | "most patients saw improvement in muscle strength from the initial to final treatment, as measured by the dynamometer." |
Device is well-tolerated with no significant adverse events. | "There were no significant adverse events noted in the duration of the study." |
2. Sample Size and Data Provenance for Test Set
- Sample Size: 45 subjects were enrolled in the study.
- Data Provenance: Not explicitly stated regarding the country of origin. The study appears to be prospective as subjects were "enrolled" and "treated" with the device, and follow-up assessments were conducted.
3. Number and Qualifications of Experts for Ground Truth
Not applicable. This was a clinical study assessing the physiological effect of the device on muscle strength, not an AI/image-based diagnostic device requiring expert interpretation for ground truth. Muscle strength was measured objectively using a dynamometer.
4. Adjudication Method for Test Set
Not applicable. Muscle strength was measured using a dynamometer, which provides objective, quantitative data, thus negating the need for an adjudication method typically used for subjective assessments or interpretations.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study
No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not done. The study described is a clinical trial assessing the direct effect of the device on muscle strength in human subjects.
6. Standalone (Algorithm Only) Performance
Not applicable. This device is a powered muscle stimulator for direct application to the human body, not an algorithm-only or AI-driven diagnostic tool. Its performance is intrinsically linked to human-device interaction.
7. Type of Ground Truth Used
The ground truth used in the clinical study was objective physiological measurement: muscle strength measured by a dynamometer at specified time points (baseline, final treatment, 30-day post-treatment, and 90-day post-treatment).
8. Sample Size for Training Set
Not applicable. This device is a physical therapeutic device, not an AI/machine learning model that requires a "training set" in the conventional sense of data-driven algorithm development. The "training" for the device would be its engineering and design based on established principles of electrical muscle stimulation.
9. How Ground Truth for Training Set Was Established
Not applicable, as there is no specific "training set" for the type of device described. The device's design and operational parameters would be based on scientific and engineering principles for powered muscle stimulators. Its effectiveness for specific indications is then demonstrated through clinical studies, as detailed above.
Ask a specific question about this device
(91 days)
IPF
This Zynex device has been designed for muscle re-education, prevention of disuse atrophy, increase local blood circulation, maintain, or increase range of motion, and relaxation of muscle spasms.
Not Found
I am sorry, but the provided text does not contain any information about acceptance criteria or a study that proves a device meets acceptance criteria. The document is an FDA 510(k) clearance letter for a device called "M-Wave," a powered muscle stimulator. It outlines the regulatory approval process and requirements but does not detail any performance studies, acceptance criteria tables, or specifics on how the device's efficacy was verified. Therefore, I cannot fulfill your request for that specific information based on the given input.
Ask a specific question about this device
Page 1 of 25