(165 days)
No
The summary does not mention AI, ML, deep learning, or any related terms, and the device description focuses on standard physiological parameter monitoring.
No
Explanation: The device is intended for measuring and monitoring physiological parameters, not for treating or providing therapy.
Yes
The device is described as a vital signs monitor used for measuring, storing, and generating prompts for multiple physiological parameters (NIBP, SpO2, PR, TEMP), which are used for monitoring a patient's health status. While it doesn't explicitly state "diagnosis," the collection and presentation of these vital signs are fundamental to diagnostic processes in a hospital environment.
No
The device description explicitly states "iM3s series vital signs monitors including iM3s\iM3As\iM3Bs\ iHM3s are intended to be used for measuring, storing, reviewing of, and generating prompts for multiple physiological parameters". The performance studies also mention electrical safety and EMC testing (IEC 60601-1, IEC 60601-1-2) and performance testing including bench and system level testing, which are indicative of a hardware device.
Based on the provided information, this device is not an IVD (In Vitro Diagnostic).
Here's why:
- Intended Use: The intended use clearly states the device is for measuring physiological parameters (NIBP, SpO2, PR, TEMP) directly from the patient. IVD devices are used to examine specimens (like blood, urine, tissue) taken from the body to provide information about a person's health.
- Device Description: The description reinforces that it's a vital signs monitor, which measures parameters in vivo (within the living body).
- Lack of Specimen Handling: There is no mention of collecting or analyzing biological specimens.
- Parameters Measured: The parameters listed (NIBP, SpO2, PR, TEMP) are all measured directly from the patient's body, not from a specimen.
Therefore, this device falls under the category of a vital signs monitor, which is a type of medical device used for patient monitoring, not an in vitro diagnostic device.
N/A
Intended Use / Indications for Use
The device is intended to be used for measuring, storing, and to generate prompts for, multiple physiological parameters of adults and pediatrics (including neonates). The device is intended for use by trained healthcare professionals in hospital environments.
Parameters include: NIBP, SpO2, PR (pulse rate), TEMP.
The F3000 Quick Temp module is not intended for neonates.
The device is not intended for MRI environments.
Product codes
MWI, DXN, FLL, DQA
Device Description
The iM3s series vital signs monitors including iM3s\iM3As\iM3Bs\ iHM3s are intended to be used for measuring, storing, reviewing of, and generating prompts for multiple physiological parameters of adults and pediatrics.
Mentions image processing
Not Found
Mentions AI, DNN, or ML
Not Found
Input Imaging Modality
Not Found
Anatomical Site
Not Found
Indicated Patient Age Range
adults and pediatrics (including neonates)
Intended User / Care Setting
trained healthcare professionals in hospital environments
Description of the training set, sample size, data source, and annotation protocol
Not Found
Description of the test set, sample size, data source, and annotation protocol
Not Found
Summary of Performance Studies (study type, sample size, AUC, MRMC, standalone performance, key results)
Non-clinical data:
Electrical safety and electromagnetic compatibility (EMC): iM3s Series Vital Signs Monitor were assessed for conformity with the relevant requirements of IEC 60601-1: 2005/A1: 2012/A2: 2020 and IEC 60601-1-2:2014+A1:2020.
Performance testing-Bench: Edan has conducted functional and system level testing to validate the performance of the devices. The results of the bench testing show that the subject device meets its accuracy specification and meet relevant consensus standards (IEC 80601-2-30:2018, ISO 80601-2-56: 2017+A 1:2018, ISO 80601-2-61: 2017).
Software Verification and Validation Testing: Software verification and validation testing were conducted and documentation was provided as recommended by FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices".
Clinical data: Not applicable.
Summary: The non-clinical performance testing showed that the subject devices are as safe and as effective as the predicate device.
Key Metrics (Sensitivity, Specificity, PPV, NPV, etc.)
SpO2 EDAN Module:
Measurement Range: 0% to 100%
Accuracy:
Adult /Pediatric (not including neonate): 70% to100% : ±2%, 0% to 69%: Undefined
Neonate: 70% to100%: ±3%, 0% to 69%: Undefined
PR from SpO2:
Measurement range: 25 to 300 bpm
Accuracy: ±2 bpm
TEMP (Covidien F3000 Quick Temp Module):
Measuring range: 30°C43°C43°C
Prediction measurement range: 35°C
TEMP (Exergen TAT-5000S-RS232-QR Temp Module):
Temperature Range: 61 to 110°F (15.5 to 43°C)
Clinical Accuracy: ± 0.2°F or 0.1°C Per ASTM E1112
Arterial Heat Balance Range for Body Temperature: 94 to 110°F (34.5 to 43°C)
TEMP (HTD8808C Non-contact Infrared Thermometer Temp Module):
Measurement Range: 34 °C to 42.9 °C
Accuracy:
34.034.9°C: ± 0.3°C (93.2-94.8°F: ± 0.5°F);42.0°C: ± 0.2°C (95.0-107.6°F: ± 0.4°F);
35.0
42.1~42.9°C: ± 0.3°C (107.8-109.2°F: ± 0.5°F);
TEMP (T100A Internal Forehead TEMP Module):
Accuracy:
±0.3 °C: 34.0 °C -34.9 °C;
±0.2 °C: 35.0 °C -42.0 °C;
±0.3 °C: 42.1°C -43.0 °C;
EDAN-NIBP(ICFUS):
Measurement Range:
Adult Mode: SYS: 25 mmHg to 290 mmHg, DIA: 10 mmHg to 250 mmHg, MAP: 15 mmHg to 260 mmHg
Pediatric Mode: SYS: 25 mmHg to 240 mmHg, DIA: 10 mmHg to 200 mmHg, MAP: 15 mmHg to 215 mmHg
Neonatal Mode: SYS: 25 mmHg to 140 mmHg, DIA: 10 mmHg to 115 mmHg, MAP: 15 mmHg to 125 mmHg
NIBP PR Measurement range: 40 bpm to 240 bpm
NIBP PR Accuracy: ± 3 bpm or 3.5%, whichever is greater
EDAN-NIBP(IFAST):
Measurement Range:
Adult Mode: SYS: 25 mmHg to 290 mmHg, DIA: 10 mmHg to 250 mmHg, MAP: 15 mmHg to 260 mmHg
Pediatric Mode: SYS: 25 mmHg to 240 mmHg, DIA: 10 mmHg to 200 mmHg, MAP: 15 mmHg to 215 mmHg
Predicate Device(s)
Reference Device(s)
Predetermined Change Control Plan (PCCP) - All Relevant Information
Not Found
§ 870.2300 Cardiac monitor (including cardiotachometer and rate alarm).
(a)
Identification. A cardiac monitor (including cardiotachometer and rate alarm) is a device used to measure the heart rate from an analog signal produced by an electrocardiograph, vectorcardiograph, or blood pressure monitor. This device may sound an alarm when the heart rate falls outside preset upper and lower limits.(b)
Classification. Class II (performance standards).
0
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
March 8, 2024
Edan Instruments, Inc. Tracy Yue Regulatory Affairs Engineer 15 Jinhui Road, Jinsha Community, Kengzi Sub-district Pingshan District, Shenzhen, Shenzhen 518122 China
Re: K233038
Trade/Device Name: Vital Signs monitor, Model: iM3s, iM3As, iHM3s Regulation Number: 21 CFR 870.2300 Regulation Name: Cardiac Monitor (Including Cardiotachometer And Rate Alarm) Regulatory Class: Class II Product Code: MWI Dated: September 25, 2023 Received: September 25, 2023
Dear Tracy Yue:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
1
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
2
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Stephen C. Browning -S
LCDR Stephen Browning Assistant Director Division of Cardiac Electrophysiology, Diagnostics, and Monitoring Devices Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
3
Indications for Use
510(k) Number (if known) K233038
Device Name
Vital Signs monitor, Model: iM3s, iM3As, iM3Bs, iHM3s
Indications for Use (Describe)
The device is intended to be used for measuring, storing, and to generate prompts for, multiple physiological parameters of adults and pediatrics (including neonates). The device is intended for use by trained healthcare professionals in hospital environments.
Parameters include: NIBP, SpO2, PR (pulse rate), TEMP.
The F3000 Quick Temp module is not intended for neonates.
The device is not intended for MRI environments.
Type of Use (Select one or both, as applicable) | |
---|---|
Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
4
510(K) Summary
Prepared in accordance with the content and format regulatory requirements of 21 CFR Part 807.92
| 1. Submitter: | Edan Instruments, Inc.
#15 Jinhui Road, Jinsha Community,
Kengzi Sub-District, Pingshan District,
Shenzhen, 518122 P.R.China.
Tel: +86(0755) 26858736
Fax: +86(0755) 26882223 |
|-------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact person: | Tracy Yue |
| Preparing date: | September 25, 2023 |
| 2. Device name and classification: | Trade Name: Vital Signs Monitor, Model: iM3s, iM3As, iM3Bs, iHM3s
Common/Usual Name: Vital Signs Monitor |
Classification Name | Product code |
---|---|
21 CFR 870.2300 | |
Monitor, Physiological, Patient (Without Arrhythmia Detection Or Alarms) | MWI |
Subsequent Product Code | |
---|---|
21 CFR 870.1130 | |
System, Measurement, Blood-Pressure, Non-Invasive | DXN |
21 CFR 880.2910 | |
Thermometer, Electronic, Clinical | FLL |
21 CFR 870.2700 | |
Oximeter | DQA |
Regulatory Class: Class II |
3. Predicate Device(s):
- Edan Instruments, Inc., Vital Signs Monitor:iM3s, iM3As, iM3Bs, iHM3s, K202892 (Primary predicate device)
-
- Shenzhen Mindray Bio-Meidcal Electronics Co., LTD, VS 9 Vital Signs Monitor, K221267 (Reference device)
-
- HeTaiDa Technology Co., Ltd, Non-contact Infrared Body Thermometer, Model HTD8808C, K203332 (Reference device)
-
- TaiDoc Technology Corporation, FORA IR20B EAR
5
THERMOMETER , K090395 (Reference device)
The iM3s series vital signs monitors including iM3s\iM3As\iM3Bs\ iHM3s are 4. Device Description: intended to be used for measuring, storing, reviewing of, and generating prompts for multiple physiological parameters of adults and pediatrics. 5. Indication for Use The device is intended to be used for measuring, storing, and reviewing of, and to generate prompts for, multiple physiological parameters of adults and pediatrics (including neonates). The device is intended for use by trained healthcare professionals in hospital environments. Parameters include: NIBP, SpO2, PR (pulse rate), TEMP. The F3000 Quick TEMP module is not intended for neonates. The device is not intended for MRI environments.
6. Predicate Device Comparison
The intended use of the subject devices iM3s series is the same as the primary predicate device Vital Signs Monitor: iM3s, iM3As, iHM3s (K202892) which is intended to be used for measuring, storing, and reviewing of, and to generate prompts for, multiple physiological parameters of adults and pediatrics in healthcare environments by trained healthcare professionals.
The table below compares the indication for use and parameter modules of the subject devices to the primary predicate devices.
| Manufacturer/Item |
EDAN Instrument Inc.
iM3s series Vital Signs Monitor |
EDAN Instrument Inc.
iM3s series Vital Signs Monitor | Comparison
Result |
|-----------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------|
| K# | K202892 | Current Submission | -- |
| Intended Use | | | |
| Description | The device is intended to be used for
measuring, storing, and reviewing of,
and to generate prompts for, multiple
physiological parameters of adults and
pediatrics. The device is intended for
use by trained healthcare professionals
in hospital environments.
Parameters include: NIBP, SpO2, PR
(pulse rate), TEMP.
The F3000 Quick TEMP module is not
intended for neonates. | The device is intended to be used
for
measuring, storing, and
reviewing of, and to generate
prompts for, multiple physiological
parameters of adults and pediatrics
(including neonates). The device is
intended for use by trained
healthcare professionals in hospital
environments.
Parameters include: NIBP, SpO2,
PR (pulse rate), TEMP | Different |
| | The device is not intended for MRI | The F3000 Quick TEMP module is | |
| | environments. | not intended for neonates.
The device is not intended for MRI
environments. | |
| Measurement
Parameters | SpO2, PR, NIBP, TEMP | SpO2, PR, NIBP, TEMP | Same |
| Operation mode | Spot-checking, Ward round | Spot-checking, Ward round | |
| SpO2 (EDAN Module) | | | |
| Measurement Range | 0% to 100% | 0% to 100% | |
| | Adult /Pediatric (not including
neonate): | Adult /Pediatric (not including
neonate): | |
| | 70% to100% : ±2% | 70% to100% : ±2% | |
| Accuracy | 0% to 69%: Undefined | 0% to 69%: Undefined | Same |
| | Neonate: | Neonate: | |
| | 70% to100%: ±3% | 70% to100%: ±3% | |
| | 0% to 69%: Undefined | 0% to 69%: Undefined | |
| PR from SpO2 | | | |
| Measurement range | 25 to 300 bpm | 25 to 300 bpm | |
| Accuracy | ±2 bpm | ±2 bpm | Same |
| | TEMP (Covidien F3000 Quick Temp Module) | | |
| Measuring range | 30°C43°C | 30°C43°C | |
| Prediction
measurement range | 35°C43°C | 35°C43°C | Same |
| Sensor type | Oral /axillary /rectal | Oral /axillary /rectal | |
| Measuring Mode | Direct Mode /Adjusted Mode | Direct Mode /Adjusted Mode | |
| | TEMP (Exergen TAT-5000S-RS232-QR Temp Module) | | |
| Temperature Range | 61 to 110°F (15.5 to 43°C) | 61 to 110°F (15.5 to 43°C) | |
| | ± 0.2°F or 0.1°C | ± 0.2°F or 0.1°C | |
| Clinical Accuracy | Per ASTM E1112 | Per ASTM E1112 | |
| Arterial Heat Balance
Range
for Body
Temperature | 94 to 110°F (34.5 to 43°C) | 94 to 110°F (34.5 to 43°C) | Same |
| | TEMP (HTD8808C Non-contact Infrared Thermometer Temp Module) | | |
| Measurement Range | 34 °C to 42.9 °C | 34 °C to 43.0 °C | |
| | 34.034.9°C: ± 0.3°C | 34.034.9°C: ± 0.3°C | |
| | (93.2-94.8°F: ± 0.5°F); | (93.2-94.8°F: ± 0.5°F); | |
| | Accuracy | 35.042.0°C: ± 0.2°C | 35.042.0°C: ± 0.2°C |
| | (95.0-107.6°F: ± 0.4°F); | (95.0-107.6°F: ± 0.4°F); | |
| | 42.142.9°C: ± 0.3°C | 42.143.0°C: ± 0.3°C | |
| (107.8-109.2°F: ± 0.5°F); | | | |
| TEMP (T100A Internal Forehead TEMP Module) | | | |
| Measurement Range | / | | |
| Accuracy | 34°C43°C (93.2°F109.4°F)
±0.3 °C: 34.0 °C -34.9 °C;
±0.2 °C: 35.0 °C -42.0 °C;
±0.3 °C: 42.1°C -43.0 °C; | Different | |
| TEMP (FORA IR20B EAR THERMOMETER TEMP Module) | | | |
| Displayed
Temperature Range | 89.6°F to 109.4°F (32°C to 43°C) | | |
| Accuracy | ±0.4°F (±0.2°C) for the range of
96.8°F to 102.2°F (36.0°C to 39.0°C)
±0.5°F (±0.3°C) from 89.6°F to
96.6°F (32.0°C to 35.9°C) and from
102.4°F to 109.4°F (39.1°C to 43.0°C) | Different | |
| Wi-Fi | | | |
| IEEE | 802.11b/g/n | Same | |
| Frequency Band | 5GHz & 2.4GHz | | |
| Max. Transmit Power
(±2 dBm) | 2.4GHz:
17dBm (802.11b DSSS)
17dBm (802.11b CCK)
17dBm (802.11g/n OFDM)
5GHz:
10dBm (802.11g OFDM)
9dBm (802.11n) | | |
| E-Link | | | |
| Radio Technology | BR, EDR, BLE | Same | |
| Frequency Range | 2402 MHz ~ 2480 MHz | | |
| EDAN-NIBP(ICFUS) | | | |
| Principle | Oscillometry | Same | |
| Measuring parameter | SYS, DIA, MAP, PR | | |
| Measurement Range | Adult Mode: | | |
| | | | |
| | SYS: 25 mmHg to 290 mmHg
DIA: 10 mmHg to 250 mmHg
MAP: 15 mmHg to 260 mmHg | SYS: 25 mmHg to 290 mmHg
DIA: 10 mmHg to 250 mmHg
MAP: 15 mmHg to 260 mmHg | |
| | Pediatric Mode:
SYS: 25 mmHg to 240 mmHg
DIA: 10 mmHg to 200 mmHg
MAP: 15 mmHg to 215 mmHg | Pediatric Mode:
SYS: 25 mmHg to 240 mmHg
DIA: 10 mmHg to 200 mmHg
MAP: 15 mmHg to 215 mmHg | |
| | Neonatal Mode:
SYS: 25 mmHg to 140 mmHg
DIA: 10 mmHg to 115 mmHg
MAP: 15 mmHg to 125 mmHg | Neonatal Mode:
SYS: 25 mmHg to 140 mmHg
DIA: 10 mmHg to 115 mmHg
MAP: 15 mmHg to 125 mmHg | |
| NIBP PR
Measurement range | 40 bpm to 240 bpm | 40 bpm to 240 bpm | |
| NIBP PR Accuracy | $\pm$ 3 bpm or 3.5%, whichever is greater | $\pm$ 3 bpm or 3.5%, whichever is greater | |
| EDAN-NIBP(IFAST) | | | |
| Principle | | Oscillometry | |
| Measuring parameter | | SYS, DIA, MAP | |
| | | Adult Mode:
SYS: 25 mmHg to 290 mmHg
DIA: 10 mmHg to 250 mmHg
MAP: 15 mmHg to 260 mmHg | |
| Measurement Range | / | Pediatric Mode:
SYS: 25 mmHg to 240 mmHg
DIA: 10 mmHg to 200 mmHg
MAP: 15 mmHg to 215 mmHg | Different |
| | | iFAST: 15 s (depend on SYS, arm
circumference and HR/motion
disturbance) (measured with E8
cuff, PR is set as 80 bpm and
systolic pressure within 120/80
mmHg) | |
| Typical
measurement time | | | |
6
7
8
As seen in the comparison tables, the subject and predicate devices have similar design features and performance specifications. The technological differences between the subject and predicate devices do not raise different questions of safety or effectiveness.
7. Performance Data:
9
Non-clinical data:
Electrical safety and electromagnetic compatibility (EMC)
iM3s Series Vital Signs Monitor were assessed for conformity with the relevant requirements of the following standards and found to comply:
- IEC 60601-1: 2005/A1: 2012/A2: 2020 Medical electrical equipment - Part 1: General requirements for basic safety and essential performance.
- IEC 60601-1-2:2014+A1:2020 Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance – Collateral Standard: electromagnetic disturbances – Requirements and tests.
Performance testing-Bench
Edan has conducted functional and system level testing to validate the performance of the devices. The results of the bench testing show that the subject device meets its accuracy specification and meet relevant consensus standards.
- IEC 80601-2-30:2018 Medical electrical equipment - part 2-30: particular requirements for the basic safety and essential performance of automated non-invasive sphygmomanometers
- ISO 80601-2-56: 2017+A 1:2018 Medical electrical equipment - part 2-56: particular requirements for basic safety and essential performance of clinical thermometers for body temperature measurement
- ISO 80601-2-61: 2017 Medical electrical equipment - part 2-61: particular requirements for basic safety and essential performance of pulse oximeter equipment
Software Verification and Validation Testing
Software verification and validation testing were conducted and documentation was provided as recommended by FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices".
Clinical data: Not applicable.
Summary
The non-clinical performance testing showed that the subject devices are as safe and as effective as the predicate device.
8. Conclusion
The bench testing data and software verification and validation demonstrate that iM3s Series Vital Signs Monitor are substantially equivalent to the predicate devices.