(146 days)
The indications for use of the BeneVision Central Monitoring System include:
- Real time viewing of patient clinical data and alarms .
- . Storage and Historical review of patient clinical data and alarms
- Printing of real time and historical patient data
- Configuration of local settings as well as synchronizing settings across the network to a remote device
- Transfer of patient clinical data and settings between several CentralStations
- Provides a Resting 12 Lead interpretation of previously stored data
The BeneVision Central Monitoring System is a networked patient monitoring system intended for use in a fixed location, installed in professional healthcare facilities to provide clinicians remote patient monitoring. The network connections between the various devices can be any combination of Ethernet (Wireless WIFI (WLAN), and Wireless WMTS. The BeneVision Central Monitoring System supports one or more Mindray compatible physiological monitors and will display, store, print, and transfer information received from the compatible monitors. The Bene Vision Central Monitoring System supports bi-directional configuration of the compatible monitors.
The telemetry monitoring systems are designed to acquire and monitor physiological data for ambulating patients within a defined coverage area. The BeneVision Central Monitoring System supports Telemetry Systems: TMS-6016, Telepack-608, TMS60, TM80 and TM70.
The TMS-6016 transmitter is intended for use on Adult and Pediatric patients to monitor ECG and SpO2 physiological data.
The Panorama Telepack-608 transmitter is intended for use on Adult patients to monitor ECG and SpO2 physiological data.
The TMS60 transmitter is intended for use on Adult and Pediatric patients over three years old to monitor ECG, . SpO2, NIBP and Resp physiological data. The physiological data can be reviewed locally on the display of the transmitter. The CentralStation will support ECG, Heart Rate, SpO2, NIBP, Resp, Pulse Rate, Arrhythmia analysis, QT monitoring, and ST Segment Analysis for the TMS60.
. The TM80/TM70 telemetry monitor is intended for use on Adult and Pediatric patients over three years old to monitor ECG. SpO2. NIBP and Resp physiological data. The physiological data can be analyzed, alarmed, stored, reviewed locally on the display of the monitor, and the CentralStation can config and display the physiological parameters from the TM80/TM70.
The BeneVision Central Monitoring System is intended for use in professional healthcare facilities under the direct supervision of a licensed healthcare practitioner.
The BeneVision Central Monitoring System (CMS) is a networked patient monitoring system intended for use in healthcare settings by, or under the direction of, a physician to provide clinicians remote patient monitoring. The target patient population is adult patients and pediatrics.
The provided text is a 510(k) premarket notification summary for the BeneVision Central Monitoring System. It details the device, its intended use, comparison to predicate devices, and various testing performed to demonstrate substantial equivalence.
However, the document does not contain information about an AI/algorithm where acceptance criteria and specific performance metrics (like sensitivity, specificity, or AUC) or information related to multi-reader multi-case (MRMC) comparative effectiveness studies would typically be found. The changes describe modifications to the software for an existing central monitoring system, primarily related to operating system compatibility, supported monitors, and the addition of a "Resting 12 Lead interpretation of previously stored data" feature.
The document discusses "Software Verification and Validation Testing" and "Bench Testing" to ensure the device meets its specifications, but these tests are for the overall system's functionality and accuracy, not for an AI component that would require the level of detail requested in the prompt.
Therefore, I cannot fulfill the request for a table of acceptance criteria and reported device performance for an AI model, sample sizes, expert qualifications, adjudication methods, MRMC studies, standalone performance, or ground truth establishment relevant to an AI/ML device, as this information is not present in the provided text.
The information that is available in the document regarding testing is general:
- Software Verification and Validation Testing: Conducted "as recommended by FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices."" This ensures the product "works as designed" and validates its "design and performance."
- Bench Testing: Conducted "functional and system level testing to validate the performance of the devices." The results "show that the subject device meets its accuracy specification, and is substantially equivalent to the predicate device."
- Consensus Standards: The device meets relevant consensus standards, specifically mentioning IEC 60601-2-25:2011 for electrocardiographs.
The document explicitly states:
- Animal Testing: "Not applicable. Animal studies are not necessary to establish the substantial equivalence of this device."
- Clinical Data: "Not applicable. Clinical testing is not required to establish substantial equivalence to the predicate device."
Without an AI/ML component described with specific performance criteria and a study demonstrating its meeting those criteria, the detailed information requested in the prompt cannot be extracted from this document.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food & Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the FDA acronym with the agency's name on the right. The symbol on the left is a stylized representation of human figures, while the text on the right reads "FDA U.S. FOOD & DRUG ADMINISTRATION" in blue font.
June 1, 2022
Shenzhen Mindray Bio-Medical Electronics Co., LTD Li Lei Manager Regulatory Affairs, Technical Regulation Department Mindray Building, Keji 12th Road South, Hi-tech Industrial Park, Nanshan Shenzhen, Guangdong 518057 China
Re: K220058
Trade/Device Name: BeneVision Central Monitoring System Regulation Number: 21 CFR 870.2300 Regulation Name: Cardiac Monitor (Including Cardiotachometer And Rate Alarm) Regulatory Class: Class II Product Code: MSX. DRT, DQA, DXN, DSB, MHX, DRO Dated: December 27, 2021 Received: January 6, 2022
Dear Li Lei:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976. the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Jennifer Shih Kozen Assistant Director Division of Cardiac Electrophysiology, Diagnostics and Monitoring Devices Office of Cardiovascular Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Indications for Use
510(k) Number (if known) K220058
Device Name
BeneVision Central Monitoring System
Indications for Use (Describe)
The indications for use of the BeneVision Central Monitoring System include:
- Real time viewing of patient clinical data and alarms .
- . Storage and Historical review of patient clinical data and alarms
- Printing of real time and historical patient data
- Configuration of local settings as well as synchronizing settings across the network to a remote device
- Transfer of patient clinical data and settings between several CentralStations
- Provides a Resting 12 Lead interpretation of previously stored data
The BeneVision Central Monitoring System is a networked patient monitoring system intended for use in a fixed location, installed in professional healthcare facilities to provide clinicians remote patient monitoring. The network connections between the various devices can be any combination of Ethernet (Wireless WIFI (WLAN), and Wireless WMTS. The BeneVision Central Monitoring System supports one or more Mindray compatible physiological monitors and will display, store, print, and transfer information received from the compatible monitors. The Bene Vision Central Monitoring System supports bi-directional configuration of the compatible monitors.
The telemetry monitoring systems are designed to acquire and monitor physiological data for ambulating patients within a defined coverage area. The BeneVision Central Monitoring System supports Telemetry Systems: TMS-6016, Telepack-608, TMS60, TM80 and TM70.
The TMS-6016 transmitter is intended for use on Adult and Pediatric patients to monitor ECG and SpO2 physiological data.
The Panorama Telepack-608 transmitter is intended for use on Adult patients to monitor ECG and SpO2 physiological data.
The TMS60 transmitter is intended for use on Adult and Pediatric patients over three years old to monitor ECG, . SpO2, NIBP and Resp physiological data. The physiological data can be reviewed locally on the display of the transmitter. The CentralStation will support ECG, Heart Rate, SpO2, NIBP, Resp, Pulse Rate, Arrhythmia analysis, QT monitoring, and ST Segment Analysis for the TMS60.
. The TM80/TM70 telemetry monitor is intended for use on Adult and Pediatric patients over three years old to monitor ECG. SpO2. NIBP and Resp physiological data. The physiological data can be analyzed, alarmed, stored, reviewed locally on the display of the monitor, and the CentralStation can config and display the physiological parameters from the TM80/TM70.
The BeneVision Central Monitoring System is intended for use in professional healthcare facilities under the direct supervision of a licensed healthcare practitioner.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
{3}------------------------------------------------
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
In accordance with 21 CFR 807.87(h) and 21 CFR 807.92 the 510(k) Summary for the BeneVision Central Monitoring System is provided below.
SUBMITTER 1.
| Applicant: | SHENZHEN MINDRAY BIO-MEDICALELECTRONICS CO., LTD.Mindray Building, Keji 12th Road SouthHigh-tech Industrial Park, NanshanShenzhen 518057, P.R. ChinaTel: +86 755 81888998Fax: +86 755 26582680 |
|---|---|
| Contact: | Contact Person: Li LeiTitle: Manager Regulatory AffairsPhone: +86 755 81885953Fax: +86 755 26582680E-mail: lilei.js@mindray.com |
| Date Prepared: | December 27, 2021 |
2. DEVICE
| Device Trade Name: | BeneVision Central Monitoring System |
|---|---|
| Device Common Name: | System, network and communication, physiological monitors |
| Classification Name and Regulation | 21 CFR 870.2300, Cardiac monitor (including cardiotachometer and rate alarm) |
| Primary Product Code: | MSX - System, network and communication, physiological monitors |
| Regulatory Class | Class II |
| Panel | Cardiovascular |
{5}------------------------------------------------
| ProductCode | RegulationNumber | Panel | Regulation description | Device Common Name |
|---|---|---|---|---|
| DRT | 21 CFR870.2300 | Cardiovascular | Cardiac Monitor(includingcardiotachometer and ratealarm) | Monitor, cardiac (incl.cardiotachometer & ratealarm) |
| DQA | 21 CFR870.2700 | Anesthesiology | Oximeter | Oximeter |
| DXN | 21 CFR870.1130 | Cardiovascular | Noninvasive bloodpressure measurementsystem | System, measurement,blood-pressure, non-invasive |
| DSB | 21 CFR870.2770 | Cardiovascular | Impedanceplethysmograph | Plethysmograph,impedance |
| MHX | 21 CFR870.1025 | Cardiovascular | Arrhythmia detector andalarm (including ST-segment measurement andalarm). | Monitor, physiological,patient (with arrhythmiadetection or alarms) |
| DRQ | 21 CFR870.2060 | Cardiovascular | Transducer signalamplifier and conditioner | Amplifier and signalconditioner, transducersignal |
| Table 1: | Secondary Product Codes |
|---|---|
| ---------- | ------------------------- |
3. PREDICATE DEVICE
Predicate Device: K193391 - BeneVision Central Monitoring System (SHENZHEN MINDRAY BIO-MEDICAL ELECTRONICS CO., LTD)
K163584 - M3290B Patient Information Center iX Release C.01(Philips Reference Device: Medical Systems)
K202405 - BeneVision N Series Patient Monitors (Including BeneVision N12, BeneVision N15, BeneVision N17, BeneVision N19, BeneVision N22, BeneVision N1) (SHENZHEN MINDRAY BIO-MEDICAL ELECTRONICS CO., LTD)
DEVICE DESCRIPTION 4.
The BeneVision Central Monitoring System (CMS) is a networked patient monitoring system intended for use in healthcare settings by, or under the direction of, a physician to provide clinicians remote patient monitoring. The target patient population is adult patients and pediatrics.
{6}------------------------------------------------
INTENDED USE/INDICATIONS FOR USE 5.
The indications for use of the BeneVision Central Monitoring System include:
- . Real time viewing of patient clinical data and alarms
- . Storage and Historical review of patient clinical data and alarms
- Printing of real time and historical patient data ●
- Configuration of local settings as well as synchronizing settings across the network to a ● remote device
- . Transfer of patient clinical data and settings between several CentralStations
- Provides a Resting 12 Lead interpretation of previously stored data
The BeneVision Central Monitoring System is a networked patient monitoring system intended for use in a fixed location, installed in professional healthcare facilities to provide clinicians remote patient monitoring. The network connections between the various devices can be any combination of Ethernet (Wired), Wireless WIFI (WLAN), and Wireless WMTS.
The BeneVision Central Monitoring System supports one or more Mindray compatible physiological monitors and will display, store, print, and transfer information received from the compatible monitors. The BeneVision Central Monitoring System supports bi-directional configuration of the compatible monitors.
The telemetry monitoring systems are designed to acquire and monitor physiological data for ambulating patients within a defined coverage area. The BeneVision Central Monitoring System supports Telemetry Systems: TMS-6016, Telepack-608, TMS60, TM80 and TM70.
- The TMS-6016 transmitter is intended for use on Adult and Pediatric patients to monitor ECG and SpO2 physiological data.
- The Panorama Telepack-608 transmitter is intended for use on Adult patients to monitor . ECG and SpO2 physiological data.
- . The TMS60 transmitter is intended for use on Adult and Pediatric patients over three years old to monitor ECG, SpO2, NIBP and Resp physiological data. The physiological data can be reviewed locally on the display of the transmitter. The CentralStation will support ECG, Heart Rate, SpO2, NIBP, Resp, Pulse Rate, Arrhythmia analysis, QT monitoring, and ST Segment Analysis for the TMS60.
- . The TM80/TM70 telemetry monitor is intended for use on Adult and Pediatric patients over three years old to monitor ECG, SpO2, NIBP and Resp physiological data. The physiological data can be analyzed, alarmed, stored, reviewed locally on the display of the monitor, and the CentralStation can config and display the physiological parameters from the TM80/TM70.
The BeneVision Central Monitoring System is intended for use in professional healthcare facilities under the direct supervision of a licensed healthcare practitioner.
{7}------------------------------------------------
SUBSTANTIAL EQUIVALENCE 6.
Comparison of Indications
Both the predicate device and the subject device are patient monitoring systems intended to be used in healthcare facilities under the direction of clinical professionals.
The indications for use statement of the subject BeneVision Central Monitoring System has been modified to include support a resting 12 Lead interpretation. Although this feature is not present in the predicate BeneVision Central Monitoring System, it is present in Mindray's N Series Monitors (K202405).
In conclusion, the difference of the indications for use do not change the fundamental intended use of the BeneVision Central Monitoring System.
Technological Comparisons
The table below provides a comparison of the technological features of the BeneVision CentralStation, WorkStation, ViewStation, and CMS Viewer compared to the CentralStation, WorkStation, ViewStation, and CMS Viewer cleared in K193391. The features in gray are features which are different between the predicate device and the subject devices. These changes marked with an asterisks (*) are non-significant.
{8}------------------------------------------------
Feature
| Technological Comparison | |||||
|---|---|---|---|---|---|
| As Cleared in K193391 | Modifications Made toSubjectDevice | Applied To | |||
| CentralStation | WorkStation | ViewStation | CMSViewer | ||
| Microsoft Windows7/8/10/Server 2012/Server2016 | 1.AddedMicrosoftWindowsServer 2019* | A(CentralStation onlysupportsWindows10/Server | A(WorkStation onlysupportsWindows10) | A(ViewStation onlysupportsWindowss 10) | A(CMSVieweronlysupportsWindows |
| 2. Deleted |
Table 2: Technological Com
| ns Made toSubjectDevice | CentralStation | WorkStation | ViewStation | CMSViewer | ||
|---|---|---|---|---|---|---|
| OperationSystem | Microsoft Windows7/8/10/Server 2012/Server2016 | 1.AddedMicrosoftWindowsServer 2019*2. DeletedMicrosoftWindows 73.TheCentralStation supportsinstallationon theVirtualMachinePlatformwhenrunning asservice suchas VMWareand Hyper-V | A(CentralStation onlysupportsWindows10/Server2016/Server 2019) | A(WorkStation onlysupportsWindows10) | A(ViewStation onlysupportsWindows 10) | A(CMSVieweronlysupportsWindows8/10/Server 2012) |
| Display(includingtouchscreencapability) | resolution:1920 x1080,1280*1024 | No change | A | A | A | A |
| Audio | Built-in speakers | No change | A | A | A | NA |
| Recorder | Mindray thermal arraymodule product | No change | A | A | A | NA |
| Network | 100 Mbps, Ethernet 802.3 | No change | A | A | A | A |
| Maxconnectionsaresupportedfor oneCentralStation | Up to 32 WorkStation orViewStation connectionsare supported for oneCentralStation | No change | A | NA | NA | NA |
| Feature | As Cleared in K193391 | Modifications Made to Subject Device | Applied To | |||
| CentralStation | WorkStation | ViewStation | CMS Viewer | |||
| Patient Monitor Numbers – Number Supported | 1.Support 32 monitors running as application.2. Running as serviceUp to 128 montors with no patient display, the display is provided by WorkStations | Support 64 monitors running as application * | A | A | A | A (CMSViewer only support 36 monitors) * |
| Feature | As Cleared in K193391 | Modifications Made to Subject Device | Applied To | |||
| CentralStation | WorkStation | ViewStation | CMS Viewer | |||
| Telemetry Systems and monitors | Supports the following telemetry systems and monitors:– TMS-6016 (K183238)– TMS60 (K183238)– Telepack-608 (K183238)– TM80(K193391)– TM70(K193391) | No change | A | A | A | A |
| Communication protocol (and compatible monitors) | CMS+ protocol:DPM3 (K072235 )DPM4/5 (K070791)DPM6/7 (K092449)Passport 12m/17m (170876)Passport8/12 (K153448)Passport V (K091834)Accutorr 7/VS-900 (K170712)T1 (K152902)ELAN protocol:Spectrum (K062098)Spectrum OR (K062098)Passport II (K020550)V12/21 (K150352)MD2 protocol:CMS Viewer (K193391)ViewStation (K193391)WorkStation (K193391)TM80 (K193391)TM70(K193391)BeneVision N22/N19/N17/N15/N12/N 1(K199391) | 1.Added VS 8/8A/9(K21 1475) *2.Added ePM Series Patient Monitors (Including ePM 10, ePM12, ePM 15, ePM 10M, ePM 12M, ePM 15M) (K200015) * | A | A | A | A |
| Feature | As Cleared in K193391 | Modifications Made to Subject Device | Applied To | |||
| CentralStation | WorkStation | ViewStation | CMS Viewer | |||
| Bi-directionalConfiguration | Patient demographics, alarm settings and parameter settingsFor TM80: patient demographics, alarm settings and parameter setup information can be set by both the CentralStation and TM80.The QRS threshold, ST point/ISO point/J point, and ST and QT template can only be set by the CentralStation. | No change | A | A | NA | NA |
| Calculations | Supports five calculation mode:Drug CalculationHemodynamics CalculationOxygenation CalculationVentilation CalculationRenal Calculation | No change | A | A | A | NA |
| ViewOther Bed | Provides the user the ability to remotely view 32 patient's parameters, waveforms, and alarms from a patient monitor connected to anotherBeneVision Central Monitoring System | No change | A | A | A | NA |
| HL7Output | Provide HL7 interface output | No change | A | NA | NA | NA |
| PagingInterface | Enables transmission of configured alarm notifications to a third-party paging system | No change | A | A | NA | NA |
| Data Review | ||||||
| Dynamicshort trend | 8 hours | No change | A | A | A | A |
| Trendreview | 240 hours | No change | A | A | A | A |
| Feature | As Cleared in K193391 | Modifications Made to Subject Device | CentralStation | WorkStation | ViewStation | CMS Viewer |
| Wave review | 240 hours of full-disclosure waveforms and compressed waveforms | No change | A | A | A | A |
| NIBP review | Most recent 3000 NIBP measurements | No change | A | A | A | A |
| Event review | 3000 events | No change | A | A | A | A |
| 12-lead review | 720 12-lead analysis results, 12 analysis waveforms for each analysis result | No change | A | A | A | A |
| ST review | Most recent 240 hours of ST segments | No change | A | A | A | A |
| Cardiac output review | 720 measurements | No change | A | A | A | A |
| Patient information, real-time waveform, real-time alarm, Alarm Settings, Multi-lead ECG Report, CSA Report, waveform review, Arrhythmia Statistic Result, Trend Review, C.O. measurement, events, 12-lead Review, ST review, QT View Report, drug calculations, hemodynamics calculations, oxygenation calculations, ventilation calculations, renal calculations, ICG hemodynamic parameter, CCO hemodynamic parameter, SvO2/ScvO2 oxygenation parameters | No change | A | A | A | A | |
| Feature | As Cleared in K193391 | Modifications Made toSubjectDevice | Applied To | |||
| CentralStation | WorkStation | ViewStation | CMSViewer | |||
| Records | Patient information, real-time waveform, real-timealarm, waveform review,C.O. measurement, events,12-lead Review, STreview, drug calculation,hemodynamicscalculations, oxygenationcalculations, ventilationcalculations, renalcalculations, ICGhemodynamic parameter,CCO hemodynamicparameter, SvO2/ScvO2oxygenation parameters | No change | A | A | A | NA |
| Datastorage | The patient data will besaved in an encrypted file. | No change | A | NA | NA | NA |
| ECG Features | ||||||
| ECGAlgorithm | Supports Mindray andMortara | No change | A | A | A | A |
| ECGFunctions | 3-lead, 5-lead, 6-leadselectable,Arrhythmia detection, STsegment analysis, QTAnalysis, Heart rate | No change | A | A | A | A |
| HR | Adult:Range:15 | No change | A | A | A | A |
| ST | Range: -2.0~2.0mVAccuracy: ±0.02mV or±10%, whichever isgreater, in the range of -0.8mV to +0.8mV; notspecified in other range | No change | A | A | A | A |
| Feature | As Cleared in K193391 | Modifications Made to Subject Device | CentralStation | WorkStation | ViewStation | CMS Viewer |
| J Point Auto Detection | J-point Auto detection for ST algorithm. Supports automatically detecting the location of the J-point on the ST template. | No change | A | A | A | A |
| ARR | Mindray algorithm:Asystol, V-Fib/V-Tach, V-Tach, Vent Brady, Extreme Tachy, Extreme Brady, PVCs/min, Vent Rhythm, Couplet, Bigeminy Trigeminy, R on T, Run PVCs, PVC, Tachy, Brady, Missed Beats, Pacer Not Pacing, Pacer Not Capture, Multiform PVC, Nonsus V-Tach, Pause, Irr Rhythm, Pauses/min, and A-FibMortara algorithm:Asystol, V-Fib, V-Tach, Vent Rhythm, Couplet, Run PVCs, PVCs/min, Bigeminy Trigeminy, R on T, Multiform PVC, Irr Rhythm, Tachy, Brady, Pacer Not Pacing, Pacer Not Capture, Extreme Tachy, Extreme Brady, Pause and Pauses/min | No change | A | A | A | A (CMS Viewer receives ARR alarms from CentralStation and only displays ARR alarms, therefore there are no modification) |
| Adjustable Leads for Arrhythmia Analysis | Adjustable Leads for Arrhythmia Analysis.Supports selectable ECG leads as primary detection lead, secondary detection lead and beat classification lead for arrhythmia analysis | No change | A | A | NA | NA |
| Feature | As Cleared in K193391 | Modifications Made to Subject Device | Applied To | |||
| CentralStation | WorkStation | ViewStation | CMS Viewer | |||
| QT Analysis | Mindray algorithm:- QT measurement range: [200, 800] ms- QT accuracy [200, 800] ms: ± 30 ms, beyond this range is not specified- QT resolution: [200, 800] ms: 4 ms, beyond this range is not specified- QTc measurement range: [200, 800] ms- QTc resolution [200, 800] ms: 1 ms, beyond this range is not specified- QT-HR measurement range: Adult: [15, 150] bpm, pediatric: [15, 180] bpm | No change | A | A | A | A |
| Mortara algorithm:- QT measurement range: [300, 600] ms- QT accuracy [300, 600] ms: ± 30 ms, beyond this range is not specified- QT resolution: [300, 600] ms: 2 ms, beyond this range is not specified- QTc measurement range: [300, 600] ms- QTc resolution [300, 600] ms: 1 ms, beyond this range is not specified- QT-HR measurement range: Adult: [43, 130] bpm, | ||||||
| Feature | As Cleared in K193391 | Modifications Made to Subject Device | Applied To | |||
| CentralStation | WorkStation | ViewStation | CMS Viewer | |||
| QRS Detection Threshold | Adjustable QRS Detection threshold.QRS threshold range: 0.16-0.48mV. | No change | A | A | NA | NA |
| Pace mark | Detects and marks pace pulse.Amplitude: ±2 to ±700 mVDuration: 0.1 to 2 msRise time: 10 to 100 µs | No change | A | A | A | A |
| Pace pulse rejection | Meets the requirements of IEC60601-2-27 2011:Section 201.12.1.101.13.The following pulses without overshoot will be rejected:Amplitude: ±2 to ±700 mVDuration: 0.1 to 2 msRise time: 10 to 100 με | No change | A | A | A | A |
| New FeaturesThe system provides a resting 12 Lead interpretation of previously stored data | Not provided | Supports the user to select a period of waveform from the historical data storage to analyze, the result will be saved in 12 lead review | A | A | NA | NA |
{9}------------------------------------------------
{10}------------------------------------------------
{11}------------------------------------------------
{12}------------------------------------------------
{13}------------------------------------------------
{14}------------------------------------------------
{15}------------------------------------------------
{16}------------------------------------------------
Substantial Equivalence Conclusion
In conclusion, the differences in respect to the indications for use and technological characteristics do not raise new questions of safety and effectiveness as compared to the predicate device.
To establish the substantial equivalence of the BeneVision Central Monitoring System, Mindray conducted functional and system level testing to validate the performance of the devices. The results of the bench testing show that the subject device meets its specifications and is substantially equivalent to the predicate device.
{17}------------------------------------------------
In addition, Mindray has conducted testing to ensure the subject devices meet relevant consensus standards.
7. PERFORMANCE DATA
Biocompatibility Testing
Not applicable. The changes are just concerned with Software, not relate to Biocompatibility Testing.
Software Verification and Validation Testing
Software verification and validation testing was conducted, and documentation was provided, as recommended by FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices." Verification of the Bene Vision Central Monitoring System was conducted to ensure that the product works as designed. Validation was conducted to check the design and performance of the product.
Electrical safety and electromagnetic compatibility (EMC)
Not applicable. The changes are just concerned with Software, not relate to Electrical safety and electromagnetic compatibility.
Bench Testing
To establish the substantial equivalence of the BeneVision Central Monitoring System, Mindray conducted functional and system level testing to validate the performance of the devices. The results of the bench testing show that the subject device meets its accuracy specification, and is substantially equivalent to the predicate device.
In addition. Mindray has conducted testing to ensure the subject devices meet relevant consensus standards.
- . IEC 60601-2-25:2011 Medical electrical equipment - Part 2-25: Particular requirements for the basic safety and essential performance of electrocardiographs
Animal Testing
Not applicable. Animal studies are not necessary to establish the substantial equivalence of this device.
Clinical Data
Not applicable. Clinical testing is not required to establish substantial equivalence to the predicate device.
{18}------------------------------------------------
8. CONCLUSION
Based on the detailed comparison between the predicate devices and the subject devices, and the functional and system level testing, the BeneVision Central Monitoring System can be found substantially equivalent to the predicate devices.
§ 870.2300 Cardiac monitor (including cardiotachometer and rate alarm).
(a)
Identification. A cardiac monitor (including cardiotachometer and rate alarm) is a device used to measure the heart rate from an analog signal produced by an electrocardiograph, vectorcardiograph, or blood pressure monitor. This device may sound an alarm when the heart rate falls outside preset upper and lower limits.(b)
Classification. Class II (performance standards).