Search Results
Found 19 results
510(k) Data Aggregation
(267 days)
The Access anti-HAV assay is a paramagnetic particle, chemiluminescent immunoassay for the in vitro qualitative detection of total antibodies (anti-HAV IgG and IgM) to hepatitis A virus (HAV) in human pediatric (2 through 21 years) and adult serum and serum separator tubes or plasma [lithium heparin, lithium heparin separator tubes, sodium citrate, acid-citrate-dextrose (ACD), and citrate phosphate-dextrose (CPD)] using the Dxl 9000 Access Immunoassay Analyzer.
The Access anti-HAV assay is indicated as an aid in the diagnosis of current or past HAV infection in persons with risk factors and/or signs or symptoms of hepatitis A, when used in conjunction with other serological and clinical information. The assay may also be used in the identification of HAV susceptible individuals and to determine the presence of an antibody response to HAV in vaccine recipients.
This assay is not intended for use for screening donors of blood or blood products or human cells, tissues, or cellular or tissue-based products (HCT/Ps).
The Access anti-HAV assay requires Access anti-HAV (reagent packs), Access anti-HAV Calibrator (C1), and Access anti-HAV QC (QC1-QC2). The Access anti-HAV assay is a two-step competitive immunoassay. During incubation, the anti-HAV antibodies present in the patient sample bind to the coated antigen. After incubation in a reaction vessel, materials bound to the solid phase are held in a magnetic field while unbound materials are washed away. A monoclonal anti-HAV antibody alkaline phosphatase conjugate is added to the reaction vessel and the conjugate competes with the bound patient antibodies to affix the HAV antigen coated on the particles. After a second incubation and wash step, the chemiluminescent substrate is added to the vessel and light generated by the reaction is measured with a luminometer. The light production is compared to the cutoff value defined during calibration. The Qualitative assessment is automatically determined from a stored calibration.
Quality control (QC) materials simulate the characteristics of patient samples and are essential for monitoring the system performance of the Access anti-HAV immunoassay. In addition, they are an integral part of good laboratory practices. When performing assays with Access reagents for anti-HAV, include quality control materials to validate the integrity of the assay. The assayed values should fall within the acceptable range if the test system is working properly.
The Access anti-HAV reagents are provided in liquid ready-to-use format designed for optimal performance on the Beckman Coulter DxI 9000 Access Immunoassay Analyzer only. Each reagent kit contains two reagent packs. The Access anti-HAV Calibrator kit contains one vial, and the Access anti-HAV QC kit contains three vials, each of anti-HAV positive control and anti-HAV negative control. Other items needed to run the assay include Lumi-Phos PRO (chemiluminescent substrate) and UniCel DxI Wash Buffer II.
N/A
Ask a specific question about this device
(266 days)
Access hsTnI is a paramagnetic particle, chemiluminescent immunoassay for the quantitative determination of cardiac troponin I (cTnI) levels in human serum and plasma using the Unicel DxI Immunoassay Systems to aid in the diagnosis of myocardial infarction (MI).
The Access hsTnI is a two–site immunoenzymatic ("sandwich") assay. Monoclonal anti–cTnI antibody conjugated to alkaline phosphatase is added to a reaction vessel along with a surfactant–containing buffer and sample. After a short incubation, paramagnetic particles coated with monoclonal anti–cTnI antibody are added. The human cTnI binds to the anti–cTnI antibody on the solid phase, while the anti–cTnI antibody–alkaline phosphatase conjugate reacts with different antigenic sites on the cTnI molecules. After incubation, materials bound to the solid phase are held in a magnetic field while unbound materials are washed away. Then, the chemiluminescent substrate is added to the vessel and light generated by the reaction is measured with a luminometer. The light production is directly proportional to the concentration of analyte in the sample. Analyte concentration is automatically determined from a stored calibration.
Here's a breakdown of the acceptance criteria and study information for the Access hsTnI device, based on the provided FDA 510(k) clearance letter:
Acceptance Criteria and Reported Device Performance
| Acceptance Criteria Category | Acceptance Criteria | Reported Device Performance (Access hsTnI Candidate on UniCel DxI 800) |
|---|---|---|
| Method Comparison | Slope of Passing-Bablok linear regression model: 1.00 ± 0.10 | Met acceptance criteria (exact slope not provided, but stated that the study "met the acceptance criteria of slope 1.00 ± 0.10"). Bias data supported that reference intervals have not changed appreciably from the commercialized product. |
| Imprecision | Within-laboratory (total) CV: ≤ 10% for concentrations ≥ 11.5 pg/mLWithin-laboratory (total) SD: ≤ 1.15 pg/mL for concentrations < 11.5 pg/mL | Within-laboratory (total) % CV ranged from 3% to 4% for Access hsTnI concentrations ≥ 11.5 pg/mL. The within-laboratory (total) SD was 0.48 pg/mL for Access hsTnI concentration < 11.5 pg/mL. |
| Linearity | For each sample concentration range, the higher order (2nd or 3rd) term of the polynomial fit should be non-significant (p > 0.05). If significant, the fit of the polynomial regression demonstrating significance should have ≤ 10% bias across the analytical measuring range. | The analysis of the data found that across the UniCel DxI 800 instruments, and for each sample concentration range, the higher order (2nd or 3rd) term of the polynomial fit is non-significant (p > 0.05), and if significant, the fit of the polynomial regression demonstrating significance had ≤ 10% bias across the analytical measuring range. |
| LoB/LoD | Not explicitly stated as an "acceptance criteria" but limits are reported for the predicate. | LoB estimate of the Access hsTnI is 1.5 (serum and plasma).LoD estimate of the Access hsTnI is 1.8 (serum and plasma). |
| LoQ | Not explicitly stated as an "acceptance criteria" but limits are reported for the predicate. | The LoQ for Access hsTnI at ≤20% within-lab CV was determined to be 1.3 pg/mL (serum) and 1.2 pg/mL (plasma). |
| Carryover | Not explicitly stated as an "acceptance criteria" for numeric limits, but the sponsor performed studies and included a limitation in labeling acknowledging observed carryover. | When a sample with cTnI > 150,000 pg/mL (ng/L) was tested, intra-assay carryover was observed if an Access hsTnI was tested after a high cTnI sample. Estimated carryover was 3-5 pg/mL (ng/L) from a high sample at 270,000 pg/mL (ng/L) and 5-8 pg/mL (ng/L) from a high sample at 500,000 pg/mL (ng/L). Limitation statements related to carryover are to be added. |
| Analytical Measuring Range | 2.3 pg/mL to 27,027 pg/mL (Predicate) | Similar (Candidate) |
Study Details
The provided document describes a study primarily focused on demonstrating the substantial equivalence of the Access hsTnI assay when run on the UniCel DxI 800 Immunoassay System compared to its predicate device (Access hsTnI on the Access 2 Immunoassay System). This is achieved through performance testing of various analytical aspects.
-
Sample Size Used for the Test Set and Data Provenance:
- Method Comparison: 239 samples (119 Lithium Heparin Plasma and 120 Serum).
- Data Provenance: Not specified (e.g., country of origin). The document indicates it's a retrospective comparison between the "IVD Access hsTnI (Current Assay Protocol File (APF))" and the "proposed Access hsTnI (Proposed APF)" on the UniCel DxI 800 instruments, implying existing samples or previously collected data.
-
Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts:
- Not Applicable. This device is an in-vitro diagnostic (IVD) immunoassay for quantitative determination of cTnI levels. The "ground truth" for the test set in this context refers to the measured cTnI values, which are inherently quantitative and determined by the predicate device's method and the proposed device's method, not by expert consensus or interpretation of images/clinical findings.
-
Adjudication Method for the Test Set:
- Not Applicable. As noted above, this is a quantitative analytical method comparison, not a diagnostic interpretation or clinical outcome study that would require adjudication.
-
If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve with AI vs Without AI Assistance:
- Not Applicable. This is an in-vitro diagnostic device, not an AI-based image interpretation or diagnostic aid system involving human readers.
-
If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done:
- Yes, implicitly. The studies described (Method Comparison, Imprecision, Linearity, LoB/LoD, LoQ, Carryover) evaluate the performance of the analytical instrument and assay without human intervention in the measurement process. The device itself is an automated immunoassay system that produces quantitative results.
-
The Type of Ground Truth Used:
- The "ground truth" in this context refers to the quantitative measurements of cTnI levels themselves. For the method comparison, the predicate device (Access hsTnI on the Access 2 Immunoassay System, or the "Current Assay Protocol File (APF)" on the DxI 800) essentially serves as the reference for comparison against the "Proposed APF" on the UniCel DxI 800. Therefore, it's a comparison against an established, legally marketed reference measurement method.
-
The Sample Size for the Training Set:
- Not specified. This documentation primarily focuses on the validation of the device's analytical performance. While there would have been internal development and optimization (which could be considered "training"), the document does not distinguish a formal "training set" (as might be seen with AI/ML models) from "internal validation" data. The tested datasets described are for analytical validation.
-
How the Ground Truth for the Training Set Was Established:
- Not specified / Not applicable in the traditional sense. As an IVD assay, the "ground truth" for developing such a test is the accurate quantitative measurement of the analyte (cTnI) in biological samples, requiring highly controlled reference methods and materials. The document indicates the device's principle is a "two-site immunoenzymatic ('sandwich') assay," which is a well-established biochemical technique. The development process would involve extensive characterization against reference standards and known concentrations, rather than establishing ground truth through, for example, expert labeling of clinical data.
Ask a specific question about this device
(260 days)
The Access Toxo IgG assay is a paramagnetic-particle, chemiluminescent immunoassay for the qualitative and quantitative determination of IgG antibodies to Toxoplasma gondii in human serum using the Access Immunoassay Systems. The Access Toxo IgG assay aids in the diagnosis of Toxoplasma gondii infection and may be used to assess the immune status of pregnant women.
This product is not FDA cleared/approved for the screening of blood or plasma donors. Assay performance characteristics have not been established for immunocompromised or immunosuppressed patients, cord blood, neonatal specimens or infants.
The Access Toxo IgG assay is a paramagnetic-particle, chemiluminescent immunoassay for the qualitative and quantitative detection of Toxoplasma gondii-specific IgG antibody in adult human serum using the Access Immunoassay Systems.
The Access Toxo IgG assay consists of the reagent pack, calibrators, and quality controls (OCs), packaged separately. Other items needed to run the assay include substrate and wash buffer.
This document describes the premarket notification (510(k)) for the Beckman Coulter Access Toxo IgG assay, a chemiluminescent immunoassay for detecting IgG antibodies to Toxoplasma gondii in human serum. This product is intended to aid in the diagnosis of Toxoplasma gondii infection and assess the immune status of pregnant women.
The submission claims substantial equivalence to a legally marketed predicate device, the Access Toxo IgG assay (K080869). The primary difference highlighted is the instrument used: the new device runs on the DxI 9000 Access Immunoassay Analyzer, while the predicate runs on the Access 2 Immunoassay System.
Here's an analysis of the provided information, focusing on the study that proves the device meets the acceptance criteria:
1. Table of Acceptance Criteria and Reported Device Performance
Strictly speaking, the document does not present "acceptance criteria" in a separate table with yes/no compliance. Instead, it details specific performance metrics and their measured values. The implicit acceptance criterion for most analytical performance studies (like imprecision and method comparison) is that the new device's performance is acceptable for its intended use and comparable to or better than the predicate. For Linearity, LoB, LoD, and LoQ, the acceptance criterion is that the study supports the claimed values.
| Performance Characteristic | Acceptance Criteria (Implicit from Study Design/Claims) | Reported Device Performance (Access Toxo IgG on DxI 9000) |
|---|---|---|
| Method Comparison (vs. Access 2 Immunoassay System) | High Positive Percent Agreement (PPA) and Negative Percent Agreement (NPA) to demonstrate interchangeability between instruments. | PPA: 100.00% (40/40) with 95% CI = 91.24% to 100% (for Reactive samples) NPA: 100.00% (99/99) with 95% CI = 96.26% to 100.00% (for Non-Reactive samples) |
| Imprecision (Within-Laboratory) | SD < 0.64 IU/mL at concentrations < 3.2 IU/mL; CV < 20.0% at concentrations > 3.2 IU/mL. (These are the design criteria mentioned, implying they are the acceptance threshold.) | Sample 1 (2.7 IU/mL): Overall Precision SD 0.38 (13.9% CV) - Meets SD criterion (0.38 < 0.64) Samples 2-6 (8.3 - 361.0 IU/mL): Overall Precision CVs range from 9.6% to 12.1% - Meets CV criterion (all < 20.0%) |
| Imprecision (Reproducibility / Between-Instrument) | Not explicitly stated as a separate numerical criterion beyond the general imprecision goal, but implicitly to show consistent performance across instruments. | Sample 1 (2.5 IU/mL): Reproducibility SD 0.29 (11.9% CV) Samples 2-6 (7.9 - 377.5 IU/mL): Reproducibility CVs range from 8.9% to 13.4% |
| Linearity | Demonstrated linearity across the claimed measuring interval. | Demonstrated linearity across the measuring interval of 3.2 - 450 IU/mL. |
| Limit of Blank (LoB) | Claimed value supported by study. | 0.8 IU/mL (claimed) |
| Limit of Detection (LoD) | Claimed value supported by study. | 1.5 IU/mL (claimed) |
| Limit of Quantitation (LoQ) | Claimed value supported by study. | 3.2 IU/mL (claimed) |
2. Sample Sizes Used for the Test Set and Data Provenance
-
Method Comparison:
- Test Sample Size: 140 native serum samples.
- Data Provenance: The document does not explicitly state the country of origin or whether the samples were retrospective or prospective. It refers to "native serum samples," which typically implies real-world patient samples.
-
Imprecision (Within-Laboratory):
- Test Sample Size: 6 serum samples (analyzed across 240 replicates each).
- Data Provenance: Not specified for origin or prospective/retrospective.
-
Imprecision (Reproducibility):
- Test Sample Size: 6 serum samples (analyzed across 225 replicates each).
- Data Provenance: Performed at an "internal site." Not specified for origin or prospective/retrospective.
-
Linearity, LoB, LoD, LoQ: The type and number of samples used for these studies are not specified, only that "a study" was performed and values were "claimed."
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts
This type of information (experts, qualifications) is primarily relevant for studies involving human interpretation or clinical diagnosis, such as imaging studies where radiologists establish ground truth.
For this in vitro diagnostic (IVD) device (an immunoassay), the "ground truth" for the test results is established by the measurement itself and comparison to a reference method or validated system (the predicate device). There are no human experts involved in interpreting the raw assay output to establish ground truth for method comparison or analytical performance studies like imprecision. The "truth" is the analytical measurement obtained by the reference method (the predicate device or a highly characterized sample).
4. Adjudication Method for the Test Set
Not applicable. As this is an IVD immunoassay, not a human reader study, adjudication methods (like 2+1 or 3+1 for imaging reads) are not relevant here. The comparison is between two analytical instruments/systems.
5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve with AI vs Without AI Assistance
Not applicable. This device is an in vitro diagnostic immunoassay. It does not involve human readers interpreting images assisted by AI, nor is it an AI-driven device in the sense of image analysis. Its function is to quantitatively and qualitatively measure an analyte in a biological sample.
6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) Was Done
Yes, the primary studies listed (Method Comparison, Imprecision, Linearity, LoB, LoD, LoQ) are all "standalone" in the sense that they evaluate the analytical performance of the device itself (Access Toxo IgG assay on the DxI 9000 Access Immunoassay Analyzer) without human interpretation affecting the measurement outcome. The device's output is a quantitative (IU/mL) or qualitative (Reactive/Non-Reactive/Equivocal) result.
7. The Type of Ground Truth Used (expert consensus, pathology, outcomes data, etc.)
For the method comparison study, the "ground truth" was essentially the results obtained from the predicate device (Access Toxo IgG assay on the Access 2 Immunoassay System). The goal was to show that the new device on the new instrument produced comparable results to the already cleared predicate.
For the analytical performance studies (Imprecision, Linearity, LoB, LoD, LoQ), the "ground truth" refers to the inherent analytical properties of the assay and instrument, which are determined by testing characterized samples (e.g., known concentrations for linearity, low-concentration samples for LoD/LoB).
8. The Sample Size for the Training Set
This document does not provide information about a "training set" in the context of machine learning or AI models. For IVD assays, "training" typically refers to the development and optimization process that precedes validation studies. The analytical performance studies presented here are primarily validation studies.
9. How the Ground Truth for the Training Set Was Established
Not applicable, as a "training set" in the context of AI/ML ground truth establishment is not relevant to this type of IVD device submission. Assay development and optimization for IVDs involve extensive R&D, but the concept of "ground truth for training" as seen in AI is not directly applicable to a chemical immunoassay.
Ask a specific question about this device
(86 days)
The Access Toxo IgM II assay is a paramagnetic-particle chemiluminescent immunoassay for the qualitative detection of Toxoplasma gondii-specific IgM antibody in adult human serum and plasma using the Access Immunoassay Systems.
The Access Toxo IgM II assay is presumptive for the diagnosis of acute, recent, or reactivated Toxoplasma gondii infection in males and pregnant females. It is recommended this assay be performed in conjunction with a Toxoplasma gondii-specific IgG antibody assay.
Note: This assay has not been cleared/approved by the FDA for the screening of blood or plasma donors in the United States.
The Access Toxo IgM II assay is a paramagnetic-particle chemiluminescent immunoassay for the qualitative detection of Toxoplasma gondii-specific IgM antibody in human serum and plasma using the Access Immunoassay Systems. The Access Toxo IgM II Calibrators are intended for use with the Access Toxo IgM II assay for the qualitative detection of Toxoplasma gondii-specific IgM antibody in adult human serum and plasma using the Access Immunoassay Systems. The Access Toxo IgM II QC is intended for monitoring system performance of the Access Toxo IgM II assay. The Access assay consists of the reagent pack, calibrators and QCs. Other items needed to run the assay include substrate and wash buffer. The Access assay reagent pack, Access assay callorators, Access QCs, along with the UniCel DxI Wash Buffer II are designed for use with the DxI 9000 Access Immunoassay Analyzer in a clinical laboratory setting.
The provided text is a 510(k) premarket notification summary for the Access Toxo IgM II assay, a diagnostic immunoassay, not an AI/ML-driven device. Therefore, many of the requested criteria (e.g., sample size for training set, number of experts for ground truth, MRMC study, AI assistance effect size) are not applicable to this type of medical device submission.
However, I can extract and present the relevant information regarding the device's acceptance criteria and the study proving it meets these criteria based on the provided text.
Acceptance Criteria and Device Performance for Access Toxo IgM II Assay
This document describes the validation of the Access Toxo IgM II assay on the DxI 9000 Access Immunoassay Analyzer, demonstrating its substantial equivalence to the previously cleared Access Toxo IgM II assay on the Access 2 Immunoassay System. The primary performance metrics reported are Positive Percent Agreement (PPA), Negative Percent Agreement (NPA), and Imprecision (CV%).
1. Table of Acceptance Criteria and Reported Device Performance
The acceptance criteria are implied by the results presented, which showed 100% agreement for both positive and negative samples, and the imprecision results were well within the design specification.
| Performance Metric | Acceptance Criteria (Implied/Design Goal) | Reported Device Performance | Study Type |
|---|---|---|---|
| Method Comparison/Accuracy | High agreement with predicate device | Method Comparison | |
| Positive Percent Agreement (PPA) | N/A (demonstrated 100% agreement) | 100% (95% CI: 91.43% to 100.00%) | Method Comparison (Access 2 vs. DxI 9000) |
| Negative Percent Agreement (NPA) | N/A (demonstrated 100% agreement) | 100% (95% CI: 96.53% to 100.00%) | Method Comparison (Access 2 vs. DxI 9000) |
| Imprecision (Within-Laboratory) | ≤ 20.0% CV (Design Goal) | Precision (CLSI EP05-A3) | |
| Sample 1 (Non-Reactive) Overall CV% | ≤ 20.0% | 6.8% | Within-Laboratory Precision |
| Sample 2 (Reactive, Low) Overall CV% | ≤ 20.0% | 5.9% | Within-Laboratory Precision |
| Sample 3 (Reactive, Mid) Overall CV% | ≤ 20.0% | 5.9% | Within-Laboratory Precision |
| Sample 4 (Reactive, High) Overall CV% | ≤ 20.0% | 5.7% | Within-Laboratory Precision |
| Imprecision (Reproducibility) | N/A (Overall CV% for precision) | Reproducibility (CLSI EP05-A3) | |
| Sample 1 (Non-Reactive) Overall CV% | N/A (demonstrated acceptable precision) | 6.8% | Reproducibility |
| Sample 2 (Reactive, Low) Overall CV% | N/A (demonstrated acceptable precision) | 5.3% | Reproducibility |
| Sample 3 (Reactive, Mid) Overall CV% | N/A (demonstrated acceptable precision) | 4.1% | Reproducibility |
| Sample 4 (Reactive, High) Overall CV% | N/A (demonstrated acceptable precision) | 4.6% | Reproducibility |
2. Sample Size Used for the Test Set and Data Provenance
- Method Comparison Study: 152 samples.
- Data Provenance: Samples were "collected from the intended use population." The study was "performed at an internal site." No specific country of origin or whether the data was retrospective or prospective is mentioned, but "intended use population" generally implies clinical samples.
- Imprecision Studies (Within-Laboratory & Reproducibility):
- For each of the 4 samples tested: N = 240 (for within-laboratory precision) and N = 225 (for reproducibility). These represent the number of individual measurements.
- The study involved testing multiple samples in duplicate (for precision) or in replicates of 5 (for reproducibility) over multiple days, across three reagent/calibrator lots and three analyzers.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications
N/A. This is a laboratory diagnostic immunoassay, not an image-based AI/ML device where expert consensus for ground truth is typically required. The "ground truth" for the method comparison study was the result from the FDA-cleared predicate device (Access Toxo IgM II on the Access 2 Immunoassay System).
4. Adjudication Method for the Test Set
N/A. The comparison was directly between the candidate device (DxI 9000) and the predicate device (Access 2). There was no human adjudication process involved in settling discrepancies between results from different analyzers beyond standard laboratory procedures for confirming unexpected results.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done
No, an MRMC study was not done. This type of study is relevant for imaging devices or AI-assisted diagnostic tools where human reader performance is a key metric. This submission is for an automated immunoassay.
6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done
Yes, the studies evaluate the standalone performance of the device/assay system (Access Toxo IgM II on the DxI 9000) against a reference standard (predicate device or established precision metrics). The system itself performs the measurement and provides a qualitative (Reactive, Equivocal, Non-Reactive) result. There isn't an "algorithm only" in the AI/ML sense, but the device operates autonomously to produce results.
7. The Type of Ground Truth Used
- Method Comparison: The results obtained from the FDA-cleared predicate device (Access Toxo IgM II on the Access 2 Immunoassay System) served as the reference for determining agreement.
- Imprecision: The consistency of the device's own measurements provided the "ground truth" for precision relative to established statistical methods (CLSI EP05-A3).
8. The Sample Size for the Training Set
N/A. This is not an AI/ML device that requires a distinct "training set." The device is a chemical immunoassay system. The development of such assays involves extensive R&D, reagent formulation, and analytical validation, but not "training data" in the machine learning sense.
9. How the Ground Truth for the Training Set was Established
N/A. As stated above, there is no "training set" or corresponding ground truth establishment in the context of an AI/ML model. The assay's analytical performance relies on validated biochemical reactions and instrument calibration.
Ask a specific question about this device
(171 days)
The Access TPO Antibody assay is a paramagnetic particle, chemiluminescent immunoassay for the quantitative determination of thyroperoxidase antibody (TPOAb) levels in human serum and plasma using the Access Immunoassay Systems.
The detection of TPOAb is an aid in the diagnosis of thyroid autoimmune disorders.
The Access TPO Antibody assay is a paramagnetic particle, chemiluminescent immunoassay for the quantitative determination of thyroperoxidase antibody (TPO Ab) levels in human serum and plasma using the Access Immunoassay Systems.
The Access TPO Antibody Calibrators are intended to calibrate the Access TPO Antibody assay for the quantitative determination of TPO Antibody levels in human serum and plasma using the Access Immunoassay Systems.
The Access TPO Antibody assay is a sequential two-step immunoenzymatic ("sandwich") assay. A sample is added to a reaction vessel with paramagnetic particles coated with thyroperoxidase protein. The serum or plasma TPO Ab binds to the thyroperoxidase. After incubation, materials bound to the solid phase are held in a magnetic field while unbound materials are washed away. Then, the chemiluminescent substrate is added to the vessel and light generated by the reaction is measured with a luminometer. The light production is directly proportional to the concentration of analyte in the sample. Analyte concentration is automatically determined from a stored calibration.
The provided document, a 510(k) summary for the Beckman Coulter Access TPO Antibody assay, describes the analytical performance studies conducted to demonstrate substantial equivalence to a predicate device. This is a common regulatory pathway for in vitro diagnostic devices, focusing on demonstrating that the new device is as safe and effective as a legally marketed predicate device.
Here's an analysis of the acceptance criteria and the study that proves the device meets them, based on the provided text:
Acceptance Criteria and Reported Device Performance
The document presents acceptance criteria implicitly through the study designs and results, as it's a submission for an IVD kit rather than an AI/ML algorithm. The performance is assessed by comparing the new device (Access TPO Antibody Assay on Dxl 9000 Access Immunoassay System) to its predicate device (Access TPO Antibody Assay on Access 2 Immunoassay System).
Table of Acceptance Criteria and Reported Device Performance:
| Performance Characteristic | Acceptance Criteria (Implicit from Study Design/Goal) | Reported Device Performance (Access TPO Antibody Assay on Dxl 9000) | Study Method/Description |
|---|---|---|---|
| Method Comparison (Agreement with Predicate) | A strong correlation and minimal bias between results from the new device and the predicate device, demonstrated by slope and intercept close to 1 and 0 respectively, and a high correlation coefficient. | Slope: 1.06 (95% CI: 1.04 - 1.08) Intercept: -0.26 (95% CI: -0.32 - -0.22) Correlation Coefficient (R): -0.978 (Note: The negative sign for R with a positive slope is unusual and might be a typo in the document; typically, a strong positive correlation is indicated by R closer to +1 for agreement studies). | CLSI EP09c, 3rd Edition; Passing-Bablok regression. |
| Imprecision (Repeatability and Reproducibility) | Within-laboratory imprecision meeting predefined thresholds: - < 0.07 IU/mL SD at concentrations < 0.6 IU/mL - < 12.0% CV at concentrations ≥ 0.6 IU/mL and < 450.0 IU/mL - < 15.0% CV at concentrations ≥ 450.0 IU/mL | Sample 1 (0.35 IU/mL): Within-Laboratory SD: 0.02, %CV: 6.9 (Meets <0.07 SD) Sample 2 (5.5 IU/mL): Within-Laboratory SD: 0.36, %CV: 6.7 (Meets <12.0% CV) Sample 3 (20 IU/mL): Within-Laboratory SD: 1.23, %CV: 6.2 (Meets <12.0% CV) Sample 4 (318 IU/mL): Within-Laboratory SD: 21.35, %CV: 6.7 (Meets <12.0% CV) Sample 5 (747 IU/mL): Within-Laboratory SD: 96.48, %CV: 12.9 (Meets <15.0% CV) | CLSI EP05-A3; multiple samples in duplicate for 2 runs/day over ≥20 days. |
| Linearity | Assay demonstrates linearity across the claimed measuring interval (0.25 - 1,000 IU/mL). | "the assay demonstrated linearity across the measuring interval." | CLSI EP06-Ed2 |
| Detection Capability (LoB, LoD, LoQ) | Established quantitative limits relevant for clinical use, demonstrating the ability to detect and quantify low concentrations of the analyte. | Limit of Blank (LoB): 0.19 IU/mL Limit of Detection (LoD): 0.23 IU/mL Limit of Quantitation (LoQ): 0.25 IU/mL | CLSI EP17-A2 |
Study Details:
This submission focuses on an In Vitro Diagnostic (IVD) assay (Access TPO Antibody), not an AI/ML algorithm. Therefore, many of the questions related to AI/ML development (training set, experts for ground truth, MRMC studies) are not directly applicable in the way they would be for an AI-based medical device. However, I will interpret them in the context of an IVD assay where possible, based on standard IVD validation practices.
-
Sample Size Used for the Test Set and Data Provenance:
- Method Comparison: N = 219 patient samples.
- Imprecision: N = 80 replicates (for each of 5 samples).
- Linearity & Detection Capability: Sample sizes for these studies are not explicitly stated, but are governed by CLSI guidelines (EP06-Ed2 and EP17-A2 respectively), which typically involve analyzing serially diluted samples.
- Data Provenance: Not explicitly stated (e.g., country of origin). The studies appear to be analytical performance studies conducted on patient samples, but whether they were retrospective or prospective is not specified. Given the nature of an IVD, these are typically conducted prospectively in a lab setting using banked or newly collected samples.
-
Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of those Experts:
- Not Applicable in the traditional AI/ML sense. For an IVD assay, the "ground truth" for the test set is either:
- Clinical Diagnosis: For the diagnostic claim ("aid in the diagnosis of thyroid autoimmune disorders"), the ground truth would be established by independent clinical diagnosis (e.g., by endocrinologists) typically using a variety of clinical data and other diagnostic tests. This information is usually part of clinical utility studies, which are not detailed in this analytical performance summary.
- Reference Method/Predicate Product: For the analytical performance studies like Method Comparison, the "ground truth" or reference is the result obtained from the predicate device (Access TPO Antibody Assay on Access 2 Immunoassay System). The predicate device itself is validated against established (often clinical) criteria.
- Known Concentrations/Materials: For linearity, imprecision, and detection capability studies, the "ground truth" is established by using characterized materials with known concentrations, or by measuring the variability around the mean of repeated measurements. This doesn't involve medical experts in the adjudication sense.
- Not Applicable in the traditional AI/ML sense. For an IVD assay, the "ground truth" for the test set is either:
-
Adjudication Method for the Test Set:
- Not Applicable. Adjudication is typically for subjective interpretations (like image reading). For an IVD, the results are quantitative measurements from an instrument. The "adjudication" is essentially the statistical analysis (e.g., Passing-Bablok regression, CV calculations) of these quantitative results against predefined criteria or a reference method.
-
If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done:
- No. An MRMC study is relevant for interpreting subjective data (like medical images) where human readers are involved. This document describes the analytical performance of an in vitro diagnostic assay that produces quantitative results, not an image interpretation or decision support AI.
-
If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done:
- Yes, in the context of an IVD. The entire analytical performance evaluation (Method Comparison, Imprecision, Linearity, Detection Capability) describes the standalone performance of the Access TPO Antibody assay on the Dxl 9000 system. It functions automatically based on its chemiluminescent immunoassay technology to produce a quantitative result. There is no "human-in-the-loop" aspect to the measurement itself, though human lab personnel operate the instrument and interpret the results.
-
The Type of Ground Truth Used:
- Known concentrations, reference method results (predicate device), and statistical averages.
- For Method Comparison, the "ground truth" for comparison is the result obtained from the predicate Access 2 Immunoassay System.
- For Imprecision, the "ground truth" is the mean concentration of the samples derived from repeated measurements, and the variability around that mean.
- For Linearity, the ground truth is derived from known dilutions of a high-concentration material.
- For Detection Capability (LoB, LoD, LoQ), the ground truth involves the statistical analysis of measurements of blank samples and low-concentration samples.
- Known concentrations, reference method results (predicate device), and statistical averages.
-
The Sample Size for the Training Set:
- Not Applicable. This is an IVD assay relying on chemical and immunological reactions, not a machine learning algorithm that requires a "training set" in the AI/ML sense. The device's parameters are determined during its design and optimization phases by Beckman Coulter, based on established immunoassay principles.
-
How the Ground Truth for the Training Set was Established:
- Not Applicable. As above, there is no "training set" for this type of device in the AI/ML context. The assay's performance characteristics (e.g., reactivity, linearity, precision) are inherent to its reagent formulation and instrument design, which are optimized by the manufacturer using internal R&D processes guided by scientific principles and regulatory standards.
Ask a specific question about this device
(84 days)
The Access sTfR assay is a paramagnetic particle, chemiluminescent immunoassay for the quantitative determination of soluble transferrin receptor (sTR) levels in human serum and plasma (heparin) using the Access Immunoassay Systems. This assay is intended as an aid in the diagnosis of Iron Deficiency Anemia (IDA), and for the differential diagnosis of IDA and Anemia of Chronic Disease (ACD).
This assay may also be used in conunction with an Access Ferritin measurement to provide a calculated sTR log ferritin index. This index is intended as an aid in the diagnosis of IDA, and for the differential diagnosis of IDA and ACD.
Access sTfR: The sTfR assay reagent pack consists of two specific reagents: (R1a) paramagnetic particles coated with streptavidin:biotinylated soluble transferrin receptor monoclonal antibody, proteins (mouse, goat, bovine), bovine serum albumin (BSA), 0.1% sodium azide, and 0.17% ProClin 300; and (R1b) Monoclonal mouse anti-human soluble transferrin receptor alkaline phosphatase (bovine) conjugate, BSA, 0.1% sodium azide and 0.17% ProClin 300. Two assay packs containing 50 tests per pack are provided for a total of 100 assay determinations.
The Access sTfR assay is a sequential two-step immunoenzymatic ("sandwich") assay. A sample is added to a reaction vessel along with paramagnetic particles coated with anti-sTfR antibody. During incubation, the sTfR antigen in the sample binds to the immobilized anti-sTfR antibody on the solid phase. Alkaline phosphatase conjugated anti-sTfR antibody is then added and reacts with a different antigenic site on the sTfR molecule.
After incubation, materials bound to the solid phase are held in a magnetic field while unbound materials are washed away. Then, the chemiluminescent substrate is added to the vessel and light generated by the reaction is measured with a luminometer. The light production is directly proportional to the concentration of analyte in the sample. Analyte concentration is automatically determined from a stored calibration.
Here's a breakdown of the acceptance criteria and study information for the Access sTfR device, based on the provided FDA 510(k) summary:
Acceptance Criteria and Reported Device Performance
| Acceptance Criteria Category | Specific Acceptance Criteria | Reported Device Performance | Meets Criteria? |
|---|---|---|---|
| Method Comparison | Slope: 0.99 – 1.01 (95% CI) | Slope: 1.00 (95% CI: 0.99 – 1.01) | Yes |
| Intercept: -0.14 - 0.15 (95% CI) | Intercept: -0.016 (95% CI: -0.14 - 0.15) | Yes | |
| Correlation Coefficient (R): Not explicitly stated as acceptance, but comparison shows R = 1.00 | Correlation Coefficient (R): 1.00 | Yes (Excellent) | |
| Imprecision | ≤ 0.72 nmol/L SD at concentrations ≤ 9 nmol/L | Sample 1 (9.1 nmol/L): 0.24 nmol/L SD | Yes |
| ≤ 8.0% CV at concentrations > 9 nmol/L | Sample 2 (17 nmol/L): 2.6% CV < 8% | Yes | |
| Sample 3 (20 nmol/L): 2.4% CV < 8% | Yes | ||
| Sample 4 (33 nmol/L): 1.8% CV < 8% | Yes | ||
| Sample 5 (67 nmol/L): 2.0% CV < 8% | Yes | ||
| Sample 6 (118 nmol/L): 2.2% CV < 8% | Yes | ||
| Linearity | "The assay demonstrated linearity across the measuring interval." | "The assay demonstrated linearity across the measuring interval." | Yes (Direct statement) |
| Detection Capability | Not explicitly stated as acceptance criteria, but listed as targets to be determined. | LoB: 0.04 nmol/L | Achieved targets |
| LoD: 0.05 nmol/L | Achieved targets | ||
| LoQ: 0.05 nmol/L (≤ 20% within-lab CV) | Achieved targets |
Study Details
-
Sample size used for the test set and the data provenance:
- Method Comparison: N = 200 patient samples.
- Imprecision: N = 6 samples (Sample 1-6) tested in duplicate, in 2 runs/day for a minimum of 20 days (total N for each sample is 80 measurements).
- Linearity & Detection Capability: Sample sizes for these specific studies are not explicitly stated, but the studies were performed on the Dxl 9000 Access Immunoassay Analyzer.
- Data Provenance: Not specified in the provided document (e.g., country of origin, retrospective or prospective).
-
Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
- This device is an in vitro diagnostic (IVD) immunoassay for the quantitative determination of soluble transferrin receptor (sTfR) levels. The performance studies (method comparison, imprecision, linearity, detection capability) do not involve human experts establishing a "ground truth" in the way an imaging AI might. Instead, the "ground truth" (or reference standard) for method comparison is the measurement from the predicate device (Access sTfR Assay on Access 2 Immunoassay System). For imprecision, linearity, and detection capability, the ground truth is established by the analytical measurement procedures themselves against defined statistical targets.
-
Adjudication method for the test set:
- Not applicable. As noted above, this is an IVD immunoassay, not a system requiring human adjudication of results in the traditional sense. The reference method (predicate device) serves as the comparator for the method comparison study.
-
If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
- Not applicable. This is an IVD immunoassay, not an imaging AI or a device that assists human readers.
-
If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:
- Yes, the performance studies (method comparison, imprecision, linearity, detection capability) represent the standalone performance of the Access sTfR assay on the Dxl 9000 Access Immunoassay Analyzer. Its output is a quantitative sTfR value.
-
The type of ground truth used:
- Method Comparison: Measurements from the predicate device (Access sTfR Assay on Access 2 Immunoassay System). This serves as the reference for comparison against the new device.
- Imprecision, Linearity, Detection Capability: Analytical measurements are compared against pre-defined statistical performance targets (e.g., SD, CV, LoB, LoD, LoQ) established according to CLSI guidelines.
-
The sample size for the training set:
- Not applicable. This is a conventional immunoassay, not a machine learning/AI device requiring a separate "training set" in that context. The device uses established biochemical reactions and a stored calibration curve.
-
How the ground truth for the training set was established:
- Not applicable (see above). The device establishes its "calibration" using internal calibrators and controls to create a stored calibration curve, as is typical for immunoassays. This is distinct from machine learning model training with labeled ground truth data.
Ask a specific question about this device
(83 days)
The Access Thyroglobulin Antibody II assay is a paramagnetic particle, chemiluminescent immunoassay for the quantitative determination of thyroglobulin antibody levels in human serum and plasma using the Access Immunoassay Systems. The measurement of thyroid autoantibodies may aid in the diagnosis of Hashimoto's disease, nontoxic goiter, and Graves' disease.
The Access Thyroqlobulin Antibody II assay is a paramagnetic particle, chemiluminescent immunoassay for the quantitative determination of thyroglobulin antibody levels in human serum and plasma using the Access Immunoassay Systems. The measurement of thyroid autoantibodies may aid in the diagnosis of Hashimoto's disease, nontoxic goiter, and Graves' disease.
The Access Thyroglobulin Antibody II assay is a sequential two-step immunoenzymatic ("sandwich") assay. A sample is added to a reaction vessel with paramagnetic particles coated with the thyroglobulin protein. The TgAb in the sample binds to the thyroglobulin coated on the particles. After incubation, materials bound to the solid phase are held in a magnetic field while unbound materials are washed away. The thyroglobulin-alkaline phosphatase conjugate is added and binds to the TgAb.
After second incubation, materials bound to the solid phase are held in a magnetic field while unbound materials are washed away. Then, the chemiluminescent substrate is added to the vessel and light generated by the reaction is measured with a luminometer. The light production is directly proportional to the concentration of analyte in the sample. Analyte concentration is automatically determined from a stored calibration.
Here's a detailed breakdown of the acceptance criteria and study information for the Beckman Coulter Access Thyroglobulin Antibody II device, extracted from the provided FDA 510(k) summary:
1. Table of Acceptance Criteria and Reported Device Performance
| Parameter | Acceptance Criteria (Predicate Device) | Reported Device Performance (Modified Device - Dxl 9000 Access Immunoassay Analyzer) |
|---|---|---|
| Intended Use | Quantitative determination of thyroglobulin antibody levels in human serum and plasma to aid in the diagnosis of Hashimoto's disease, nontoxic goiter, and Graves' disease. | Same (No change in Intended Use) |
| Analyte Measured | Thyroglobulin Antibody | Same |
| Technology / Format / Method | Sandwich immunoassay / Chemiluminescent / Automated | Same |
| Sample Type | Human serum or plasma | Same |
| Sample Volume | 10 uL | Same |
| Measuring Range | 1.5 - 2,500 IU/mL | Same |
| Blocker Reagents | Biotin and alkaline phosphatase included in reagent pack as blockers | Same |
| Biotin Interference | No significant interference (± 10%) observed in samples containing up to 3,510 ng/mL of biotin. | Same (Explicitly stated in the comparison table) |
| Imprecision (Repeatability) | SD ≤ 1.5 for values < 15 IU/mL; CV ≤ 10.0% for values ≥ 15 IU/mL and < 1000 IU/mL; CV ≤ 15.0% for values ≥ 1000 IU/mL | Within-Laboratory Imprecision: - Sample 1 (2.4 IU/mL): 5.2% CV (meets ≤ 1.5 IU/mL SD, which is equivalent to 62.5% CV. The actual SD is 0.1, making the %CV 4.2% for repeatability) - Sample 2 (188 IU/mL): 4.1% CV (meets ≤ 10.0% CV) - Sample 3 (727 IU/mL): 4.2% CV (meets ≤ 10.0% CV) - (Partial data for Sample ~1000 IU/mL is cut off, but the predicate applies CV ≤ 15.0% for values ≥ 1000 IU/mL) |
| Reproducibility | Not explicitly stated as a separate acceptance criterion for the predicate, but implied by the imprecision criteria. | Reproducibility (Overall): - Sample 1 (2.6 IU/mL): 5.8% CV (within expected range for low concentration) - Sample 2 (184 IU/mL): 3.8% CV (well within 10.0% CV) - Sample 3 (744 IU/mL): 3.1% CV (well within 10.0% CV) - Sample 4 (1503 IU/mL): 3.6% CV (well within 15.0% CV) - Sample 5 (1966 IU/mL): 6.4% CV (well within 15.0% CV) |
| Linearity | The assay demonstrates linearity across the measuring interval. | Determined to demonstrate linearity across the measuring interval (no quantitative data given, but implied successful). |
| Limit of Blank (LoB) | Assumed to be ≤ 0.1 IU/mL (based on reported value) | 0.1 IU/mL |
| Limit of Detection (LoD) | Assumed to be ≤ 0.2 IU/mL (based on reported value) | 0.2 IU/mL |
| Limit of Quantitation (LoQ) | ≤ 1.5 IU/mL for ≤ 20% within-lab CV | 1.5 IU/mL (at ≤ 20% within-lab CV) |
| Method Comparison (Slope) | Assumed to be close to 1.0 (for substantial equivalence to predicate) | 0.97 (95% CI: 0.95 – 0.99) |
| Method Comparison (Intercept) | Assumed to be close to 0 (for substantial equivalence to predicate) | -0.37 (95% CI: -0.99 – 0.047) |
| Method Comparison (Correlation Coefficient) | Assumed to be close to 1.0 (for substantial equivalence to predicate) | 1.00 |
Note: The primary goal of this submission (K240996) is to demonstrate substantial equivalence of the same device on a new instrument platform (Dxl 9000 Access Immunoassay Analyzer) compared to its predicate on the Access 2 Immunoassay System. Therefore, the "acceptance criteria" are largely derived from the established performance of the predicate device.
2. Sample Sizes Used for the Test Set and Data Provenance
- Method Comparison Study (CLSI EP09c):
- Sample Size: N = 114
- Data Provenance: Not explicitly stated (e.g., country of origin, retrospective/prospective). However, such studies typically use clinical samples that may be either retrospectively collected or prospectively collected for validation purposes.
- Imprecision Study (CLSI EP05-A3):
- Sample Size: 3 distinct "Samples" (concentrations) were tested. Each sample was tested 80 times (duplicate R.L per day for a minimum of 20 days).
- Data Provenance: Not explicitly stated. Assumed to be laboratory samples or controls prepared for method validation.
- Reproducibility Study (CLSI EP05-A3):
- Sample Size: 5 distinct "Samples" (concentrations) were tested. Each sample was tested 75 times (replicates of 5 per day for a minimum of 5 days on 3 instruments).
- Data Provenance: Not explicitly stated. Assumed to be laboratory samples or controls prepared for method validation.
- Detection Capability (LoB, LoD, LoQ) Study (CLSI EP17-A2):
- Sample Size: Not explicitly stated for each determination, but these studies typically involve multiple replicates of blank, low-level, and higher-level samples.
- Data Provenance: Not explicitly stated. Assumed to be laboratory samples or controls.
3. Number of Experts Used to Establish Ground Truth and Qualifications
This document describes a clinical laboratory device (an immunoassay), not an AI/imaging device requiring expert interpretation of results for ground truth. Therefore, the concepts of "experts to establish ground truth" (in the context of clinical interpretation or diagnosis from an image) and their "qualifications" are not applicable here.
For this type of device, ground truth is established through:
- Reference Methods / Predicate Devices: The Access 2 Immunoassay System (predicate) served as the comparator for the method comparison study to assess the "truth" of the Dxl 9000 system's measurements.
- Certified Reference Materials/Standards: Calibrators and controls with known analyte concentrations, often traceable to international standards, are used to establish accuracy and calibration.
- Clinical Diagnosis: For the "Indications for Use," the device aids in diagnosis, meaning its results are interpreted by clinicians in conjunction with other clinical information. The diagnostic accuracy studies (sensitivity, specificity) in relation to a "gold standard" clinical diagnosis are typically part of a larger clinical trial not detailed in this specific 510(k) summary for a platform change.
4. Adjudication Method for the Test Set
Not applicable. As this is an immunoassay device assessing quantitative levels of an antibody, there is no "adjudication method" in the sense of reconciling divergent expert interpretations of qualitative or semi-quantitative data. The "test set" results—the quantitative values—are compared statistically to the reference method (predicate device) and assessed against performance specifications (imprecision, linearity, detection limits).
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done
No, an MRMC comparative effectiveness study was not done. This type of study is relevant for AI/imaging devices where multiple human readers interpret complex cases (e.g., medical images) and AI assistance might improve their performance. This document is for a medical laboratory immunoassay for quantitative measurement of thyroglobulin antibody, not an AI-assisted diagnostic tool for human interpretation.
6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done
While the device operates "standalone" in the sense that the instrument performs the assay and generates a quantitative result without human intervention during the analytical process, this is not an "algorithm-only" study in the context of AI. The performance studies (imprecision, linearity, method comparison, detection capability) represent the standalone analytical performance of the instrument/reagent system. The "human-in-the-loop" would be the clinician interpreting the numerical result in the context of a patient's overall clinical picture, but the device itself functions automatically.
7. The Type of Ground Truth Used
- Quantitative Reference Values: For the performance studies, the ground truth is established through various means:
- Method Comparison: The results obtained from the predicate device (Access Thyroglobulin Antibody II on the Access 2 Immunoassay System) serve as the reference standard.
- Imprecision & Reproducibility: Derived from repeated measurements of samples (often control materials or pooled patient samples) to assess variability. The "true" value for these samples is either assigned by the manufacturer or determined through extensive testing.
- Linearity: Determined by creating serially diluted samples from a high-concentration sample, where the expected concentration of each dilution is the "ground truth."
- Detection Capability (LoB, LoD, LoQ): Established using blank samples and low-concentration spiked samples, with statistical methods determining the lowest detectable/quantifiable levels.
8. The Sample Size for the Training Set
This document does not describe a machine learning or AI algorithm development that would typically involve a "training set." The studies described are for analytical validation of an immunoassay on a new instrument platform, focusing on demonstrating equivalent performance to a predicate device. Therefore, a "training set" in the AI sense is not applicable.
9. How the Ground Truth for the Training Set Was Established
Not applicable, as there is no "training set" in the context of AI or machine learning for this immunoassay device.
Ask a specific question about this device
(39 days)
The Access Intrinsic Factor Ab assay is a paramagnetic particle, chemiluminescent immunoassay for the detection of intrinsic factor antibody in human serum and plasma using the Access Immunoassay Systems.
It is intended for in vitro diagnostic use as an aid in the diagnosis of pernicious anemia.
The Access Intrinsic Factor Ab assay is a competitive binding immunoenzymatic assay. The Access assay consists of the reagent pack, calibrators and QCs. Other items needed to run the assay include substrate and wash buffer. The Access assay reagent pack, Access assay calibrators, Access QCs, along with the UniCel Dxl Wash Buffer II are designed for use with the Dxl 9000 Access Immunoassay Analyzer in a clinical laboratory setting.
The acceptance criteria and device performance information provided in the document focuses on the Access Intrinsic Factor Ab assay's performance on the Dxl 9000 Access Immunoassay Analyzer compared to the Access 2 Immunoassay System (predicate device).
Here's a breakdown of the requested information based on the provided text:
1. Table of Acceptance Criteria and Reported Device Performance
The document doesn't explicitly state "acceptance criteria" as a separate table, but it provides performance metrics that would serve as the basis for such criteria, particularly for method comparison and imprecision. The comparison is against an existing predicate device, implying that performance should be comparable or non-inferior.
| Performance Metric | Acceptance Criteria (Implied/Derived from context) | Reported Device Performance (Access Intrinsic Factor Ab on Dxl 9000) |
|---|---|---|
| Method Comparison | Agreement with predicate device (Access 2 Immunoassay System) for Negative, Positive, and Total Agreement. (Specific thresholds are not explicitly stated, but high agreement is expected for substantial equivalence). | Negative Agreement: 100% (39/39)Positive Agreement: 97.5% (78/80)Total Agreement: 96.9% ((39 + 78 + 7) / 128) |
| Within-Laboratory Imprecision (Total CV) | CV ≤ 10.0% | Sample 1 (Mean 1.0 AU/mL): 4.2%Sample 2 (Mean 1.3 AU/mL): 4.3%Sample 3 (Mean 1.5 AU/mL): 3.9%Sample 4 (Mean 1.6 AU/mL): 4.4%Sample 5 (Mean 3.4 AU/mL): 3.7%Sample 6 (Mean 14 AU/mL): 3.3%(All reported Total CVs are ≤ 10.0%) |
2. Sample Size and Data Provenance
- Test Set Sample Size:
- Method Comparison: 128 serum samples.
- Imprecision Study: For each of the 6 samples tested, N=80 measurements were performed (implying 80 duplicate measurements over 20 days or similar).
- Data Provenance: Not explicitly stated (e.g., country of origin). The studies appear to be prospective analytical validation studies conducted by the manufacturer, Beckman Coulter, Inc.
3. Number and Qualifications of Experts for Ground Truth
This information is not provided in the document. The studies described are analytical performance validations comparing a new instrument platform to an existing one, and assessing precision. They do not involve human expert interpretation of images or other subjective data for ground truth establishment. The ground truth for the method comparison is the result obtained from the predicate device (Access 2 Immunoassay System), which itself is a quantitative assay.
4. Adjudication Method for the Test Set
This information is not applicable/provided. The studies focus on objective quantitative measurements (concentration levels, agreement with a predicate assay). There is no mention of subjective interpretation requiring adjudication by multiple readers or experts.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study
This information is not applicable/provided. The device is an in vitro diagnostic immunoassay, not an imaging AI device that would typically involve human reader evaluation or assistance.
6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study
This information is not applicable. The device is an immunoassay system, not an AI algorithm. Its performance is inherent to the assay and instrument mechanics. The "standalone" performance is effectively what is reported in the method comparison and imprecision studies.
7. Type of Ground Truth Used
- For Method Comparison: The ground truth for evaluating the Access Intrinsic Factor Ab assay on the Dxl 9000 was the results obtained from the predicate device (Access Intrinsic Factor Ab assay on the Access 2 Immunoassay System). This is a comparative analytical performance study.
- For Imprecision: The ground truth is the measured concentration of the intrinsic factor antibody within the samples themselves, with variability around that measurement being assessed.
8. Sample Size for the Training Set
This information is not applicable/provided. As an immunoassay, the device does not employ machine learning or AI algorithms that would require a "training set" in the conventional sense. The "training" for such a system would involve optimizing the assay reagents and instrument parameters during development, not a data-driven training set for an algorithm.
9. How the Ground Truth for the Training Set was Established
This information is not applicable/provided for the same reasons as #8.
Ask a specific question about this device
(90 days)
The Access EPO assay is a paramagnetic particle, chemiluminescent immunoassay for the quantitative determination of erythropoietin levels in human serum and plasma (heparin) using the Access Immunoassay Systems. This assay is intended as an aid in the diagnosis of anemias and polycythemias.
The Access EPO assay is a paramagnetic particle, chemiluminescent immunoassay for the quantitative determination of erythropoietin levels in human serum and plasma (heparin) using the Access Immunoassay Systems. This assay is intended as an aid in the diagnosis of anemias and polycythemias.
The Access EPO Calibrators are intended to calibrate the Access EPO assay for the quantitative determination of EPO levels in human serum and plasma (heparin) using the Access Immunoassay Systems.
The Access EPO assay is a two-site immunoenzymatic ("sandwich") assay. A sample is added to a reaction vessel along with the paramagnetic particles coated with mouse monoclonal anti-EPO, blocking reagent and the alkaline phosphatase conjugate. After incubation, materials bound to the solid phase are held in a magnetic field while unbound materials are washed away. Then, the chemiluminescent substrate is added to the vessel and light generated by the reaction is measured with a luminometer. The light production is directly proportional to the concentration of analyte in the sample. Analyte concentration is automatically determined from a stored calibration.
The provided text describes the 510(k) premarket notification for the Beckman Coulter "Access EPO" assay when run on the Dxl 9000 Access Immunoassay Analyzer. This document outlines the device's technical characteristics, its intended use, and comparative studies against a predicate device. However, it does not describe an AI-powered diagnostic device or a study involving human readers and AI assistance. Instead, it describes an immunoassay for measuring erythropoietin levels. Therefore, many of the detailed points requested in your prompt (e.g., number of experts for ground truth, MRMC study, effect size of AI, standalone AI performance) are not applicable to this specific submission.
Despite this, I will extract and present the available information that aligns with your request regarding acceptance criteria and study details for this non-AI diagnostic device.
Here's the breakdown based on the provided document:
Device Name: Access EPO
Intended Use: The Access EPO assay is a paramagnetic particle, chemiluminescent immunoassay for the quantitative determination of erythropoietin levels in human serum and plasma (heparin) using the Access Immunoassay Systems. This assay is intended as an aid in the diagnosis of anemias and polycythemias.
1. Table of Acceptance Criteria and Reported Device Performance
The document defines performance characteristics (acceptance criteria implicitly by demonstrating acceptable results) through various analytical studies.
| Study Parameter | Acceptance Criteria (Implicit from demonstrated performance) | Reported Device Performance |
|---|---|---|
| Method Comparison | Slope close to 1, Intercept close to 0, High Correlation Coefficient (comparing new system to predicate) | N = 152 samples (141 native, 11 spiked) - Concentration Range: 0.79 - 697 mIU/mL (Access 2 values) - Slope: 0.99 (95% CI: 0.97 – 1.00) - Intercept: -0.040 (95% CI: -0.19 – 0.22) - Correlation Coefficient (R): 1.00 (Demonstrates substantial equivalence to predicate) |
| Imprecision (Within-Laboratory) | ≤ 0.30 mIU/mL SD at concentrations ≤ 3.0 mIU/mL ≤ 10.0% CV at concentrations > 3.0 mIU/mL | Performed on four Dxl 9000 Access Immunoassay Analyzers, testing four native samples and one spiked sample. All reported within-laboratory %CVs are well below 10% for concentrations > 3.0 mIU/mL, and SDs are below 0.30 mIU/mL for concentrations ≤ 3.0 mIU/mL. - Sample 1 (2.0 mIU/L): Within-Laboratory SD 0.07, %CV 3.8 - Sample 2 (18 mIU/L): Within-Laboratory SD 0.5, %CV 3.0 - Sample 3 (105 mIU/L): Within-Laboratory SD 2.8, %CV 2.7 - Sample 4 (267 mIU/L): Within-Laboratory SD 6.3, %CV 2.4 - Sample 5 (548 mIU/L): Within-Laboratory SD 20.4, %CV 3.7 |
| Linearity | Assay demonstrates linearity across the measuring interval. | "The assay demonstrated linearity across the measuring interval" (Specific range is 0.6-750 mIU/L, as stated in the "Measuring Range" table). |
| Limit of Blank (LoB) | Not explicitly stated as acceptance criteria, but claimed value is provided. | Claimed LoB for Access assay is 0.6 mIU/mL. |
| Limit of Detection (LoD) | Not explicitly stated as acceptance criteria, but claimed value is provided. | Claimed LoD for Access assay is 0.6 mIU/mL. |
| Limit of Quantitation (LoQ) | Not explicitly stated as acceptance criteria, but claimed value is provided. | Claimed LoQ for Access assay is 0.6 mIU/mL. |
2. Sample Size Used for the Test Set and Data Provenance
- Method Comparison Test Set: 152 samples (141 native, 11 spiked).
- Imprecision Test Set: 5 samples (4 native, 1 spiked), tested in duplicate in 2 runs per day for 20 days on each of four analyzers (total 80 replicates per sample).
- Data Provenance: The document does not specify the country of origin for the samples or if they were retrospective or prospective. Given it's a 510(k) submission for a laboratory diagnostic, the samples are typically human biological samples collected for validation purposes.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts
Not applicable. This is not an image-based AI diagnostic device requiring expert adjudication of images. The "ground truth" for an immunoassay is typically established through reference methods or analysis of sample characteristics (e.g., known concentrations for spiked samples, or comparison to a cleared predicate device for native samples).
4. Adjudication Method for the Test Set
Not applicable, as it's an immunoassay for quantitative determination of erythropoietin, not a subjective interpretation task that would require adjudication.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve with AI vs. Without AI Assistance
Not applicable. This device is an automated immunoassay system, not an AI-assisted diagnostic tool that involves human readers.
6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done
Not applicable in the context of an "algorithm only" AI performance. The performance described here is the "standalone" performance of the immunoassay system (Dxl 9000 Access Immunoassay Analyzer with Access EPO assay reagents).
7. The Type of Ground Truth Used
The ground truth for this device's performance evaluation is established through:
- Comparison to a Predicate Device: For the method comparison study, the predicate device (Access EPO Assay on Access 2 Immunoassay System) served as the reference for native samples.
- Known Concentrations: For imprecision and linearity studies, samples of known or targeted concentrations (including spiked samples) are used.
- Reference Standards/Materials: Implied for establishing accurate concentrations for LoB, LoD, and LoQ.
8. The Sample Size for the Training Set
Not applicable. This is an immunoassay, not a machine learning model that requires a "training set" in the conventional sense of AI development. The "training" of the instrument involves calibration using specific calibrators provided with the assay.
9. How the Ground Truth for the Training Set Was Established
Not applicable as there is no "training set" for an AI model. The "calibration" of the immunoassay is done using Access EPO Calibrators, which are intended to calibrate the assay. The ground truth for these calibrators would be established by the manufacturer through rigorous analytical methods to assign their precise concentrations.
Ask a specific question about this device
(266 days)
The Access NT-proBNP assay is a paramagnetic particle, chemiluminescent immunoassay for the quantitative determination of N-terminal pro B-type natriuretic peptide levels in human serum and plasma using the automated DxI Access Immunoassay Analyzers to aid in the following:
- diagnosis of patients suspected of having acute heart failure in the Emergency Department
- assessment of heart failure severity
- risk stratification of patients with heart failure
- risk stratification of patients with acute coronary syndrome
The Access NT-proBNP assay is a paramagnetic particle, chemiluminescent immunoassay for the quantitative determination of N-terminal pro B-type natriuretic peptide levels in human serum and plasma using the automated Dxl 9000 Access Immunoassay Analyzers to aid in the following: 1) diagnosis of patients suspected of acute heart failure in the Emergency Department, 2) assessment of heart failure severity, 3) risk stratification of patients with heart failure, 4) risk stratification of patients with acute coronary syndrome.
The Access NT-proBNP is a two-site immunoenzymatic (sandwich) assay. Paramagnetic particles coated with monoclonal anti-NT-proBNP antibody and monoclonal anti-NTproBNP antibody conjugated to alkaline phosphatase are added to a reaction vessel along with a surfactant-containing buffer and serum or plasma sample. The human NTproBNP binds to the anti-NT-proBNP antibody on the solid phase, while the anti-NTproBNP antibody-alkaline phosphatase conjugate reacts with a different antigenic site on the NT-proBNP molecule. After incubation, materials bound to the solid phase are held in a magnetic field while unbound materials are washed away. Then the chemiluminescent substrate is added to the vessel and light generated by the reaction is measured with a luminometer. The light production is directly proportional to the concentration of analyte in the sample. Analyte concentration is automatically determined from a stored calibration.
Other items required to use the assay include calibrators, Lumi-Phos PRO, and wash buffer. The Access NT-proBNP reagent packs. Access NT-proBNP calibrators, along with the Access wash buffer, and Lumi-Phos PRO are designed for use on the Dxl 9000 Access Immunoassay Analyzers in a clinical laboratory setting.
Here's a breakdown of the acceptance criteria and study details for the Beckman Coulter Access NT-proBNP device, based on the provided text:
Acceptance Criteria and Reported Device Performance
The acceptance criteria are implied by the clinical and non-clinical studies conducted, with the device performance needing to meet expected ranges and exhibit substantial equivalence to predicate devices. Specific quantitative acceptance criteria are given for imprecision, LoB/LoD/LoQ, and linearity.
| Acceptance Criteria Category | Specific Criteria (Implied or Stated) | Reported Device Performance |
|---|---|---|
| Imprecision | - ≤ 4.0 ng/L SD at concentrations ≤ 50 ng/L - ≤ 8.0% CV at concentrations > 50 ng/L | Sample 1 (Mean 31 ng/L): SD 1.1 ng/L (3.5% CV). Meets criterion. Sample 2 (Mean 129 ng/L): SD 7.8 ng/L (6.0% CV). Meets criterion. Sample 3 (Mean 266 ng/L): SD 16.7 ng/L (6.3% CV). Meets criterion. Sample 4 (Mean 377 ng/L): SD 26.4 ng/L (7.0% CV). Meets criterion. Sample 5 (Mean 1,777 ng/L): SD 89.5 ng/L (5.0% CV). Meets criterion. Sample 6 (Mean 12,076 ng/L): SD 676.7 ng/L (5.6% CV). Meets criterion. Sample 7 (Mean 26,126 ng/L): SD 1516.1 ng/L (5.8% CV). Meets criterion. |
| High Dose Hook Effect | No observed high dose hook effect within a specified high concentration range (e.g., up to 300,000 pg/mL, similar to predicate) | No high dose hook effect observed up to 400,000 ng/L. Meets implied criterion (exceeds predicate device's demonstrated hook effect range). |
| Limit of Blank (LoB) | < 10.0 ng/L (pg/mL) | Maximum Observed Result: 1.1 ng/L. Meets criterion. |
| Limit of Detection (LoD) | ≤ 10.0 ng/L (pg/mL) | Maximum Observed Result: 4.8 ng/L. Meets criterion. |
| Limit of Quantitation (LoQ) | ≤ 10.0 ng/L (pg/mL) with ≤ 20% within-lab CV | Maximum Observed Result: 4.8 ng/L. Meets criterion (implied that CV was ≤20%). |
| Linearity | - Within ± 10% for values > 50 ng/L - Within ± 5.0 ng/L for values ≤ 50 ng/L | Demonstrated acceptable non-linearity across the analytical measuring range (10.0 - 35,000 ng/L) and meets the specified ±10% and ±5.0 ng/L criteria. |
| Matrix Comparison | All indicated sample types (serum, lithium heparin plasma, EDTA plasma) are suitable for use. | Study with 68 matched samples showed all sample types are suitable. Meets criterion. |
| Interfering Substances | No significant interference (defined as > 10% shift in dose) by common substances at specified concentrations. | None of the tested compounds caused significant interference (>10% shift). Meets criterion. |
| Cross Reactivity | No significant cross-reactivity (>10%) with structurally similar substances. | No significant cross-reactivity (>10%) observed. Meets criterion. |
| Clinical Performance (AHF Diagnosis) | Aid in diagnosis of acute heart failure with comparable diagnostic accuracy to predicate device (Elecsys proBNP II), as evidenced by ROC AUC. | AUC for Access NT-proBNP was 0.8536 (95% CI: 0.8362 - 0.8710), comparable to Elecsys proBNP II at 0.8562 (95% CI: 0.8361 - 0.8762). Meets criterion of comparability. |
| Clinical Performance (NYHA Correlation) | Significant trend relationship between NT-proBNP values and NYHA classification for all subjects, females, and males. | JT test of trending resulted in statistically significant p-values (<0.0001 for all subjects, 0.0005 for females, 0.0033 for males), indicating a significant trend relationship. Meets criterion. |
| Clinical Performance (Risk Stratification) | Demonstrated prognostic utility for risk stratification in patients with heart failure and acute coronary syndrome (supported by peer-reviewed literature). | Supported by analysis of peer-reviewed literature and guidelines for the same analyte, confirming correlation between NT-proBNP and cardiovascular events/mortality. Meets criterion. |
Study Details
-
Sample Size and Data Provenance:
- Test Set (Clinical Study for AHF Diagnosis): 2,384 patients presenting to the Emergency Department with clinical suspicion of acute heart failure.
- Provenance: This was a "multicenter prospective study." While specific countries are not mentioned, regulatory submissions to the FDA typically involve studies conducted in the US or other regions following ICH GCP guidelines. The study type is explicitly prospective.
- Test Set (Reference Interval Study): 675 apparently healthy adults for the overall ULN, broken down into 306 males and 369 females.
- Provenance: This was a "multicenter prospective reference interval study." Again, specific countries are not mentioned but it was prospective.
- Test Set (Imprecision): Number of runs (minimum 20 days) and replicates (duplicate) are specified for samples 1-7, with N ranging from 83 to 86 tests per sample.
- Test Set (LoB, LoD, LoQ): Multiple reagents lots and 3 instruments over 3-5 days.
- Test Set (Linearity): Native patient samples.
- Test Set (Matrix Comparison): Sixty-eight (68) matched serum, lithium heparin, and EDTA plasma samples.
- Test Set (Interfering Substances/Cross Reactivity): Lithium heparin plasma samples (concentrations approx. 125 ng/L and 1,800 ng/L) spiked with various substances. Number of individual samples not specified, but the number of substances tested is large.
- Test Set (Clinical Study for AHF Diagnosis): 2,384 patients presenting to the Emergency Department with clinical suspicion of acute heart failure.
-
Number of Experts and Qualifications for Ground Truth:
- For the clinical sensitivity study (AHF diagnosis), "Final diagnoses were adjudicated by an independent committee of medical doctors."
- Number of Experts: Not explicitly stated (e.g., "3 independent physicians"). It mentions "an independent committee," which implies more than one.
- Qualifications: "Medical doctors." No specific specialties (e.g., cardiologists, emergency physicians) or years of experience are listed in the provided text.
- For risk stratification, the basis for effectiveness comes from "an objective and systematic analysis of recent peer-reviewed literature and clinical practice guidelines," implying consensus from the broader medical community rather than a specific set of experts for this particular study.
-
Adjudication Method for the Test Set:
- For the AHF diagnosis study, the text states: "Final diagnoses were adjudicated by an independent committee of medical doctors who decided on presence of acute heart failure. Adjudicators were blinded to the Access NT-proBNP assay results."
- The specific method (e.g., 2+1, 3+1) is not explicitly detailed. It simply states "adjudicated by an independent committee," which could imply consensus, majority vote, or a specific tie-breaking rule, but the exact mechanism is not provided.
-
Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:
- No, this device is an in-vitro diagnostic (IVD) assay, not an AI/imaging device. Therefore, an MRMC study and human reader improvement with AI assistance are not applicable. The comparison made is against another IVD assay (Elecsys proBNP II) as a standalone algorithm/test.
-
Standalone Performance:
- Yes, the primary performance shown is standalone (algorithm only/test only) performance. The Access NT-proBNP assay is a quantitative determination of NT-proBNP levels. Its performance (e.g., analytical performance, diagnostic accuracy via ROC curves) is measured and reported as the output of the assay itself. The clinical performance section directly presents the assay's ability to diagnose AHF and correlates its results with NYHA classification and risk stratification, without human interpretation of the assay's output as an input to a further AI system.
-
Type of Ground Truth Used:
- Clinical Diagnosis (Expert Consensus/Adjudication): For the acute heart failure diagnosis study, the ground truth was established by an "independent committee of medical doctors" adjudicating the final diagnoses. This falls under expert consensus based on clinical information.
- Clinical Outcomes/Literature Review: For risk stratification, the ground truth is based on the correlation of NT-proBNP with "increased incidence of cardiovascular events, mortality and composite outcomes" as established in existing peer-reviewed literature and clinical practice guidelines.
- Reference Intervals for Healthy Population: Established by recruiting "apparently healthy adults" based on defined exclusion criteria and screening with additional biomarkers (eGFR, hsTnl) to ensure the health status.
-
Sample Size for the Training Set:
- The provided document describes the validation studies (test set) for the Access NT-proBNP device. It does not mention a separate "training set" in the context of an AI/machine learning model. This is an IVD assay, not an AI device trained on data. The development process for such an assay involves R&D, optimization, and internal verification stages, but the term "training set" doesn't directly apply in the same way it would for AI products.
-
How Ground Truth for the Training Set Was Established:
- As this is not an AI/machine learning device, the concept of a "training set" with established ground truth as per AI/ML typically doesn't apply. The development of the assay itself would involve internal studies and optimization based on known reference materials and clinical samples, but these are part of the assay's design and analytical verification, not a training phase for a learning algorithm.
Ask a specific question about this device
Page 1 of 2