Search Results
Found 4 results
510(k) Data Aggregation
(90 days)
Drive Chaska, Minnesota 55318
Re: K240182
Trade/Device Name: Access EPO Regulation Number: 21 CFR 864.7250
Access Immunoassay Analyzer Classification Name: Ervthropoietin assay Classification Requlation: [21 CFR 864.7250
The Access EPO assay is a paramagnetic particle, chemiluminescent immunoassay for the quantitative determination of erythropoietin levels in human serum and plasma (heparin) using the Access Immunoassay Systems. This assay is intended as an aid in the diagnosis of anemias and polycythemias.
The Access EPO assay is a paramagnetic particle, chemiluminescent immunoassay for the quantitative determination of erythropoietin levels in human serum and plasma (heparin) using the Access Immunoassay Systems. This assay is intended as an aid in the diagnosis of anemias and polycythemias.
The Access EPO Calibrators are intended to calibrate the Access EPO assay for the quantitative determination of EPO levels in human serum and plasma (heparin) using the Access Immunoassay Systems.
The Access EPO assay is a two-site immunoenzymatic ("sandwich") assay. A sample is added to a reaction vessel along with the paramagnetic particles coated with mouse monoclonal anti-EPO, blocking reagent and the alkaline phosphatase conjugate. After incubation, materials bound to the solid phase are held in a magnetic field while unbound materials are washed away. Then, the chemiluminescent substrate is added to the vessel and light generated by the reaction is measured with a luminometer. The light production is directly proportional to the concentration of analyte in the sample. Analyte concentration is automatically determined from a stored calibration.
The provided text describes the 510(k) premarket notification for the Beckman Coulter "Access EPO" assay when run on the Dxl 9000 Access Immunoassay Analyzer. This document outlines the device's technical characteristics, its intended use, and comparative studies against a predicate device. However, it does not describe an AI-powered diagnostic device or a study involving human readers and AI assistance. Instead, it describes an immunoassay for measuring erythropoietin levels. Therefore, many of the detailed points requested in your prompt (e.g., number of experts for ground truth, MRMC study, effect size of AI, standalone AI performance) are not applicable to this specific submission.
Despite this, I will extract and present the available information that aligns with your request regarding acceptance criteria and study details for this non-AI diagnostic device.
Here's the breakdown based on the provided document:
Device Name: Access EPO
Intended Use: The Access EPO assay is a paramagnetic particle, chemiluminescent immunoassay for the quantitative determination of erythropoietin levels in human serum and plasma (heparin) using the Access Immunoassay Systems. This assay is intended as an aid in the diagnosis of anemias and polycythemias.
1. Table of Acceptance Criteria and Reported Device Performance
The document defines performance characteristics (acceptance criteria implicitly by demonstrating acceptable results) through various analytical studies.
| Study Parameter | Acceptance Criteria (Implicit from demonstrated performance) | Reported Device Performance |
|---|---|---|
| Method Comparison | Slope close to 1, Intercept close to 0, High Correlation Coefficient (comparing new system to predicate) | N = 152 samples (141 native, 11 spiked) - Concentration Range: 0.79 - 697 mIU/mL (Access 2 values) - Slope: 0.99 (95% CI: 0.97 – 1.00) - Intercept: -0.040 (95% CI: -0.19 – 0.22) - Correlation Coefficient (R): 1.00 (Demonstrates substantial equivalence to predicate) |
| Imprecision (Within-Laboratory) | ≤ 0.30 mIU/mL SD at concentrations ≤ 3.0 mIU/mL ≤ 10.0% CV at concentrations > 3.0 mIU/mL | Performed on four Dxl 9000 Access Immunoassay Analyzers, testing four native samples and one spiked sample. All reported within-laboratory %CVs are well below 10% for concentrations > 3.0 mIU/mL, and SDs are below 0.30 mIU/mL for concentrations ≤ 3.0 mIU/mL. - Sample 1 (2.0 mIU/L): Within-Laboratory SD 0.07, %CV 3.8 - Sample 2 (18 mIU/L): Within-Laboratory SD 0.5, %CV 3.0 - Sample 3 (105 mIU/L): Within-Laboratory SD 2.8, %CV 2.7 - Sample 4 (267 mIU/L): Within-Laboratory SD 6.3, %CV 2.4 - Sample 5 (548 mIU/L): Within-Laboratory SD 20.4, %CV 3.7 |
| Linearity | Assay demonstrates linearity across the measuring interval. | "The assay demonstrated linearity across the measuring interval" (Specific range is 0.6-750 mIU/L, as stated in the "Measuring Range" table). |
| Limit of Blank (LoB) | Not explicitly stated as acceptance criteria, but claimed value is provided. | Claimed LoB for Access assay is 0.6 mIU/mL. |
| Limit of Detection (LoD) | Not explicitly stated as acceptance criteria, but claimed value is provided. | Claimed LoD for Access assay is 0.6 mIU/mL. |
| Limit of Quantitation (LoQ) | Not explicitly stated as acceptance criteria, but claimed value is provided. | Claimed LoQ for Access assay is 0.6 mIU/mL. |
2. Sample Size Used for the Test Set and Data Provenance
- Method Comparison Test Set: 152 samples (141 native, 11 spiked).
- Imprecision Test Set: 5 samples (4 native, 1 spiked), tested in duplicate in 2 runs per day for 20 days on each of four analyzers (total 80 replicates per sample).
- Data Provenance: The document does not specify the country of origin for the samples or if they were retrospective or prospective. Given it's a 510(k) submission for a laboratory diagnostic, the samples are typically human biological samples collected for validation purposes.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts
Not applicable. This is not an image-based AI diagnostic device requiring expert adjudication of images. The "ground truth" for an immunoassay is typically established through reference methods or analysis of sample characteristics (e.g., known concentrations for spiked samples, or comparison to a cleared predicate device for native samples).
4. Adjudication Method for the Test Set
Not applicable, as it's an immunoassay for quantitative determination of erythropoietin, not a subjective interpretation task that would require adjudication.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve with AI vs. Without AI Assistance
Not applicable. This device is an automated immunoassay system, not an AI-assisted diagnostic tool that involves human readers.
6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done
Not applicable in the context of an "algorithm only" AI performance. The performance described here is the "standalone" performance of the immunoassay system (Dxl 9000 Access Immunoassay Analyzer with Access EPO assay reagents).
7. The Type of Ground Truth Used
The ground truth for this device's performance evaluation is established through:
- Comparison to a Predicate Device: For the method comparison study, the predicate device (Access EPO Assay on Access 2 Immunoassay System) served as the reference for native samples.
- Known Concentrations: For imprecision and linearity studies, samples of known or targeted concentrations (including spiked samples) are used.
- Reference Standards/Materials: Implied for establishing accurate concentrations for LoB, LoD, and LoQ.
8. The Sample Size for the Training Set
Not applicable. This is an immunoassay, not a machine learning model that requires a "training set" in the conventional sense of AI development. The "training" of the instrument involves calibration using specific calibrators provided with the assay.
9. How the Ground Truth for the Training Set Was Established
Not applicable as there is no "training set" for an AI model. The "calibration" of the immunoassay is done using Access EPO Calibrators, which are intended to calibrate the assay. The ground truth for these calibrators would be established by the manufacturer through rigorous analytical methods to assign their precise concentrations.
Ask a specific question about this device
(269 days)
Re: K183088
Trade/Device Name: ADVIA Centaur Erythropoietin (EPO) assay Regulation Number: 21 CFR 864.7250
Trade Name: ADVIA Centaur® Erythropoietin (EPO) Assay Common Name: Erythropoietin Governing Regulation: 864.7250
The ADVIA Centaur® Erythropoietin (EPO) assay is for in the quantitative measurement of exythropoitin in pediatric and adult human serum or plasma (K2-EDTA, lithium heparin) using the ADVIA Centaur XP system. Measurement of erythropoietin is used as an aid in the diagnosis of anemias and polycythemias.
The ADVIA Centaur EPO assay is a fully automated, one-step sandwich immunoassay using direct chemiluminescent technology. The assay utilizes an acridinium-ester-labeled monoclonal mouse anti-EPO antibody in the Lite Reagent. The Solid Phase consists of mouse anti-EPO monoclonal antibody-coated paramagnetic microparticles.
Here's an analysis of the provided text to extract the acceptance criteria and study details for the ADVIA Centaur Erythropoietin (EPO) assay:
1. Table of Acceptance Criteria and Reported Device Performance
This table compiles information primarily from the "Summary of Non-Clinical Performance" and "Summary of Clinical Performance" sections.
| Acceptance Criterion (Implicit) | Reported Device Performance (ADVIA Centaur EPO assay) |
|---|---|
| Linearity (range over which results are proportional to actual concentration) | Linear from 0.83–750.00 mIU/mL. |
| Dilution Recovery (accuracy after dilution) | Observed percent recovery for individual samples ranged from 76 - 111% when diluted 1:10. |
| Measuring Interval (reportable range) | 0.83 - 750.00 mIU/mL. |
| Limit of Blank (LoB) | 0.46 mIU/mL. |
| Limit of Detection (LoD) (lowest concentration detectable with 95% probability) | 0.75 mIU/mL. |
| Limit of Quantitation (LoQ) (lowest concentration detectable at total error of 30%) | 0.83 mIU/mL. (Results below LoQ should be reported as < 0.83 mIU/mL). |
| High Dose Hook Effect (no paradoxical decrease at high concentrations) | Patient samples with EPO levels as high as 114,500 mIU/mL will assay greater than 750.00 mIU/mL (i.e., no high-dose hook effect within the measuring interval). |
| Cross-reactivity (minimal interference from related substances and normal plasma proteins) | Showed minimal cross-reactivity with normal human alpha-2-macroglobulin, transferrin (iron-saturated and non-saturated), rh Thrombopoietin, alpha-1-antitrypsin, alpha- and beta-globulins, Gamma Globulins, and alpha-1-acid glycoprotein. Cross-reactivity with Epoetin alfa and Dabepoetin alfa was also quantified (e.g., Epoetin alfa at 250 mIU/mL showed 27.13% cross-reactivity, Dabepoetin alfa at 2075 mIU/mL showed 7.14% cross-reactivity). |
| Interference (minimal effect from common endogenous and exogenous interfering substances) | Designed to be ≤ 10% change in EPO values at approximately 4-6 mIU/mL and 25-35 mIU/mL. Insignificant effect from hemolyzed samples (up to 500 mg/dL hemoglobin), icteric (up to 60 mg/dL unconjugated bilirubin, 40 mg/dL conjugated bilirubin), and lipemic (up to 3000 mg/dL Intralipid). No significant interference from Acetaminophen, Acetylsalicylic acid, Biotin, Cholesterol, EPO Soluble Receptor, Heparin, Human Gamma Globulins, Ibuprofen, Multivitamin, Protein Albumin (human), Rheumatoid Factor, Silwet L720, Total Protein, and Triglycerides at specified concentrations (e.g., Acetaminophen >18 mg/dL caused >10% change at 4-6 mIU/mL EPO; Albumin >6.8 g/dL caused >10% change at 4-6 mIU/mL EPO; EPO soluble receptor >31.25 ng/dL caused >10% change at 4-6 mIU/mL EPO; Human gamma globulins (IgG) 6.7 g/dL caused >10% change at 25-35 mIU/mL EPO). |
| Precision (reproducibility and repeatability) | Coefficients of Variation (CV%) for Repeatability (Within-Run) ranged from 1.6% to 4.8%. CV% for Within-Lab (Total) ranged from 2.6% to 8.4% across 7 samples with EPO concentrations from 1.69 to 579.41 mIU/mL. |
| Specimen Collection Comparison (equivalence across different tube types) | Correlation coefficient (r) ≥ 0.95, a slope of 0.90-1.10, and an intercept ± 1.00 mIU/mL for alternate tube types (y) versus human serum (x). Demonstrated r values of 0.99-1.00, slopes of 0.97-1.02 and intercepts of -0.33 to -0.20 for K2-EDTA, Lithium Heparin, Sodium Heparin, Plasma Separator Tube, and Serum Separator Tube compared to human serum. |
| Method Comparison (Agreement with a legally marketed predicate device) | Passing-Bablok regression: ADVIA Centaur EPO (y) = 0.99 (x) + 0.81 mIU/mL (intercept), r = 0.99 (1st study). ADVIA Centaur EPO (y) = 1.07 (x) + 0.00 mIU/mL (intercept), r = 1.00 (2nd study). ADVIA Centaur EPO (y) = 1.01 (x) + 0.36 mIU/mL (intercept), r = 0.99 (3rd multi-site study). |
| Expected Values (establishment of reference ranges for adult and pediatric populations) | Established 95% Reference Range for combined adult male and female: 5.44 - 26.25 mIU/mL. Established pediatric ranges for Male Child (2-12): 4.13-25.52 mIU/mL; Male Adolescent (13-21): 4.15-26.15 mIU/mL; Female Child (2-12): 4.94-24.47 mIU/mL; Female Adolescent (13-21): 4.07-40.30 mIU/mL. |
| Standardization (traceability to international standards) | Traceable to WHO 2nd International Reference Preparation for Erythropoietin (human, urinary derived); NIBSC code: 67/343, and WHO 3rd International Standard for Erythropoietin, recombinant, for bioassay; NIBSC code: 11/170. |
| Substantial Equivalence (Overall conclusion based on studies showing similar performance to predicate) | The ADVIA Centaur EPO assay demonstrated substantially equivalent performance to the Beckman Coulter Access EPO assay. |
2. Sample Size Used for the Test Set and Data Provenance
- Linearity: Not specified, but involved three high EPO samples mixed with low EPO human serum.
- Dilution Recovery: 10 samples (containing high EPO levels: 618.63-986.07 mIU/mL).
- Detection Capability (LoD): 323 determinations using 10 low-level samples.
- Cross-reactivity: Not explicitly stated as a "sample size," but involved numerous cross-reactants (e.g., various plasma proteins, epoetin alfa, darbepoetin alfa).
- Interference: Not explicitly stated as a specific "sample size" for each interferent, but involved various substances tested at different concentrations.
- Precision: 7 pooled serum samples. For each sample, there were 80 observations (replicates of 2, in 2 runs/day, over 20 days).
- Specimen Collection Comparison: 65 samples (serum EPO values ranging from 4.39 - 707.81 mIU/mL).
- Method Comparison:
- Study 1: 216 human serum samples (range: 3.29 – 691.60 mIU/mL).
- Study 2: 100 human serum samples from US population (range: 4.45 - 407.74 mIU/mL).
- Study 3 (Multi-site): 327 human serum samples (range: 3.55 - 596.81 mIU/mL), with ≥ 100 samples per site.
- Expected Values (Adult): 251 apparently healthy subjects (128 males, 123 females), older than 21 years of age.
- Expected Values (Pediatric): 266 apparently healthy children (2 to <13 years) and adolescents (13 to <22 years).
Data Provenance:
The document does not explicitly state the country of origin for all data sets, but mentions:
- Method Comparison Study 2: "100 human serum samples from US population".
- Method Comparison Study 3: "3 sites, 2 within Europe and 1 within the US".
- Most studies imply retrospective (samples collected before the study) or prospective (samples collected during the study for specific evaluation), but this is not consistently specified for each study. For establishing expected values, samples were collected prospectively following specific criteria (e.g., healthy subjects, specific age ranges, exclusion criteria).
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications
- This device is an in vitro diagnostic assay that measures erythropoietin levels. The "ground truth" for such assays typically comes from the actual concentration of the analyte in the biological sample, often confirmed by a reference method or standardized preparations.
- The document describes traceability to World Health Organization (WHO) International Reference Preparations and Standards (NIBSC codes 67/343 and 11/170). These international standards are established through collaborative studies involving multiple expert laboratories and scientists, but the immediate "ground truth" for the test samples in this submission would be their assigned values based on these standards or the results from the predicate device.
- No "experts" in the sense of radiologists or other clinicians interpreting images or assessments are directly involved in establishing the ground truth for individual test samples for this type of device. The ground truth is biochemical measurement.
4. Adjudication Method for the Test Set
- Not applicable. This is a quantitative diagnostic assay. "Adjudication" typically refers to resolving discrepancies between multiple human readers in diagnostic imaging or clinical assessment studies. For an immunoassay, the "ground truth" is a measured concentration.
5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study Was Done, and the Effect Size of How Much Human Readers Improve with AI vs Without AI Assistance
- Not applicable. This is an immunoassay, not an AI-assisted diagnostic imaging or clinical decision support device involving human readers.
6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) Was Done
- Yes, the performance studies described (Linearity, Dilution Recovery, Measuring Interval, Detection Capability, High Dose Hook, Cross-reactivity, Interference, Precision, Specimen Collection Comparison, Method Comparison, Expected Values) represent the standalone performance of the ADVIA Centaur Erythropoietin (EPO) assay itself, meaning the algorithm/instrument's measurement capabilities without direct human interpretation of the final measurement result for diagnostic purposes. The device directly produces a quantitative EPO value.
7. The Type of Ground Truth Used
- Reference materials/standards: The assay is traceable to WHO International Reference Preparations/Standards for Erythropoietin.
- Predicate device comparison: For method comparison studies, the results from the legally marketed predicate device (Beckman Coulter Access EPO Assay) serve as a comparative "ground truth" to demonstrate substantial equivalence.
- Expert consensus (indirectly): The establishment of reference intervals (expected values) involved defining "healthy" populations based on standard medical criteria, which would implicitly rely on shared medical understanding and the consensus of medical professionals. The CLSI protocols referenced (e.g., EP28-A3) also represent a form of expert consensus on how to establish such values.
- Biochemical measurement: The fundamental ground truth is the actual concentration of EPO in the samples, determined through rigorous biochemical methods and validated against international standards.
8. The Sample Size for the Training Set
- For an immunoassay like this, there isn't a "training set" in the machine learning sense. The device is not learning from data in the same way an AI algorithm would. Instead, the assay's reagents, calibration curves, and analytical procedures are developed and optimized through extensive R&D and internal validation studies. The studies described in the 510(k) are primarily for verification and validation (V&V) of the final device's performance characteristics. Therefore, a specific "training set sample size" as might be provided for an AI/ML device is not applicable or provided.
9. How the Ground Truth for the Training Set Was Established
- As above, the concept of a "training set" and its "ground truth" in the context of machine learning is not directly applicable to an immunoassay. The development of the assay involves extensive characterization of reagents, optimization of reaction conditions, and establishment of calibration curves using purified EPO standards of known concentrations, which are themselves traceable to international standards. The "ground truth" during this development phase would be the known concentrations of these standards.
Ask a specific question about this device
(417 days)
Drive Chaska, MN 55318
K052223 Re:
Trade/Device Name: Access® EPO Assay Regulation Number: 21 CFR & 864.7250
The Access EPO assay is a paramagnetic particle, chemiluminescent immunoassay for the quantitative determination of erythropoietin levels in human serum and plasma (heparin) using the Access Immunoassay Systems. This assay is intended as an aid in the diagnosis of anemias and polycythemias. With the advent of the administration of recombinant erythropoietin as a biologic therapy to increase red blood cell mass, an erythropoietin assay may be used also to aid in the prediction and monitoring of response to recombinant erythropoietin treatment in persons with anemias.
The Access EPO calibrators are intended to calibrate the Access EPO assay for the quantitative determination of EPO levels in human serum and plasma (heparin) using the Access Immunoassay Systems.
The Access® EPO assay consists of the reagent pack and calibrators. Other items needed to perform the assay include the Access substrate and wash buffers.
Here's a breakdown of the acceptance criteria and study information for the Access® EPO Assay based on the provided text:
1. Table of Acceptance Criteria and Reported Device Performance:
| Performance Metric | Acceptance Criteria | Reported Device Performance |
|---|---|---|
| Precision | Total precision of ≤ 10% CV at EPO concentrations > 3 mIU/mL | Internal Data: - Precision tested at concentrations from approximately 9 to 475 mIU/mL. - Within-run imprecision ranged from 1.8% CV to 8.7% CV. - Total imprecision ranged from 2.6% CV to 8.7% CV. Conclusion: Meets the criterion; total precision values are all ≤ 10% CV. |
| Analytical Sensitivity | Lowest detectable level of EPO distinguishable from zero (Access EPO Calibrator S0) is ≤0.6 mIU/mL | Internal Data: - Reported as ≤0.6 mIU/mL. |
| Dilution Recovery (Linearity) | Sample mean recovery values for all serum and plasma samples were within the range of 100 ± 15%. | Internal Data: - Dilution recovery studies performed by diluting multiple human serum and plasma (heparin) samples at various levels with Access EPO Calibrator S0. - Sample mean recovery values were within the range of 100 ± 15%. |
| Methods Comparison | Acceptable agreement with predicate device (RDS Quantikine ELISA assays). | Internal Site Study: - Slope of 1.0511, intercept of -1.3595, and correlation coefficient (r) of 0.988. - N=103 with EPO concentration range of approximately 3 to 182 mIU/mL. External Site Study: - Slope of 1.1216, intercept of -2.4168, and correlation coefficient (r) of 0.995. - N=113 with EPO concentration range of approximately 3 to 193 mIU/mL. |
| Analytical Specificity | No significant interference from therapeutic drugs, similar compounds, or potential sample contaminants (total protein, bilirubin, hemoglobin, and triglycerides). | Internal Data: - No significant interference from therapeutic drugs or similar compounds. - No significant interference from potential sample contaminants (total protein, bilirubin, hemoglobin, and triglycerides). |
| Stability | EPO reagents stable for 28 days after opening; calibrators stable for 90 days after opening; calibration curve stable for 28 days. | Internal Data: - EPO reagents are stable for 28 days after opening. - Calibrators are stable for 90 days after opening. - The calibration curve is stable for 28 days. |
2. Sample Sizes Used for the Test Set and the Data Provenance:
- Precision: Not explicitly stated as a separate "test set" in the context of clinical samples, but the study was performed on samples with EPO concentrations ranging from 9 to 475 mIU/mL.
- Dilution Recovery (Linearity): "Multiple human serum and plasma (heparin) samples." The exact number is not provided.
- Methods Comparison:
- Internal Site Study: N=103
- External Site Study: N=113
- Analytical Specificity: Not explicitly stated as a numerical sample size, but indicates testing for interference from "therapeutic drugs or similar compounds" and "potential sample contaminants."
- Data Provenance: The studies are described as "Internal Site Study" and "External Site Study," suggesting the data comes from within Beckman Coulter's testing facilities and at least one other external laboratory. The samples were "human serum and plasma (heparin)." The country of origin is not explicitly stated, but the submission is to the FDA (USA). The studies appear to be prospective in nature, as they involve testing the performance of the new Access® EPO assay.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts:
This information is not provided in the document. For an immunoassay, the "ground truth" for the test set is typically established by comparing the device's results to a well-established, often reference, method. In this case, the predicate device (RDS Quantikine ELISA) served as the reference for method comparison. The document does not describe expert adjudication for these numerical results.
4. Adjudication Method for the Test Set:
This is not applicable in the context of an immunoassay performance study focused on quantitative measurements. Adjudication methods (like 2+1, 3+1) are typically used in imaging or diagnostic accuracy studies where expert consensus is required to establish a qualitative or subjective ground truth. Here, the comparison is against an established quantitative predicate method.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve with AI vs. without AI Assistance:
This information is not applicable to this device. The Access® EPO Assay is an in vitro diagnostic (IVD) immunoassay, not an AI-powered diagnostic imaging device or an AI human-in-the-loop system. Therefore, MRMC studies examining human reader performance with or without AI assistance are not relevant here.
6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done:
Yes, the studies presented represent the standalone performance of the Access® EPO Assay. This is an automated immunoassay system; its performance metrics (precision, sensitivity, linearity, methods comparison, specificity, stability) characterize the algorithm and instrument's ability to quantitatively determine EPO levels without direct human interpretation of the assay's core output.
7. The Type of Ground Truth Used:
For the quantitative performance studies, particularly the "Methods Comparison," the ground truth was effectively the results obtained from established predicate devices, specifically the R&D Systems Quantikine IVD Erythropoietin ELISA Kit. These predicate devices are themselves validated diagnostic assays.
8. The Sample Size for the Training Set:
This information is not provided in the document. For an immunoassay, a "training set" in the context of machine learning isn't directly applicable in the same way it would be for an AI algorithm. Instead, "training" for such a system typically involves:
- Assay development and optimization using numerous samples.
- Establishing calibration curves using specific calibrator materials.
- Method validation experiments that might involve hundreds or thousands of samples over time.
The document refers to "Summary of Performance Studies" as direct validation of the assay's performance attributes rather than detailing a distinct "training set."
9. How the Ground Truth for the Training Set Was Established:
As mentioned above, the concept of a "training set" with an explicitly established ground truth (like expert consensus or pathology for AI) is not directly applicable to this type of IVD immunoassay. The development and calibration of the assay would rely on:
- Reference materials/standards: The assay is standardized against the "WHO 2nd IRP 67/343" (International Reference Preparation). This serves as a fundamental "ground truth" for the quantitative measurement of EPO.
- Known concentrations: During assay development, samples with known EPO concentrations (often characterized by reference methods or gravimetric methods for primary standards) would be used to build and validate the assay's response curve.
- Cross-validation with existing methods: Early in development, comparison with established methods would help refine the assay's performance.
Ask a specific question about this device
(301 days)
and heparinized plasma
LKEPZ (50 tests), LKEP1 (100 tests)
Class III device (pre-amendment) 21CFR 864.7250
IMMULITE EPO is a solid-phase, two-site sequential chemiluminescent enzyme immunometric assay for use with the IMMULITE Automated Analyzer and designed for the quantitative measurement of erythropoietin (EPO) in serum or heparinized plasma. It is intended strictly for in vitro diagnostic use as an aid in the diagnosis of anemias and polycythemias.
IMMULITE® EPO is a solid-phase, two-site chemiluminescent sequential enzyme immunometric assay for use with the IMMULITE® Automated Immunoassay Analyzer
Here's a breakdown of the acceptance criteria and study information for the IMMULITE® EPO device, based on the provided text:
1. Table of Acceptance Criteria and Reported Device Performance
The submission primarily focuses on demonstrating substantial equivalence through a method comparison, rather than pre-defined acceptance criteria with specific thresholds for accuracy, sensitivity, or specificity commonly seen with imaging devices. The acceptance criterion is implicit: the new device's performance must be comparable to the predicate device.
| Acceptance Criteria (Implicit) | Reported Device Performance |
|---|---|
| Correlation with Predicate Device (Incstar's EPO-Trac 125I RIA) | Linear regression analysis yielded: (IMMULITE) = 0.846 (INCSTAR) + 3.818 mU/mL Correlation coefficient (r) = 0.978 Means: 40.96 mU/mL (IMMULITE), 43.90 mU/mL (INCSTAR) |
| EPO Concentration Range tested | 2.5 to 200 mU/mL |
| Agreement in clinical interpretation (aid in diagnosis) | Not explicitly quantified with figures like sensitivity/specificity or positive/negative predictive values. The high correlation coefficient (r = 0.978) suggests strong agreement in quantitative EPO measurements, which would support similar diagnostic aid for anemias and polycythemias. |
2. Sample Size Used for the Test Set and Data Provenance
- Sample Size for Test Set: 130 endogenous serum samples.
- Data Provenance: Not explicitly stated (e.g., country of origin). The data is from "endogenous serum samples," implying human samples. It is retrospective in the sense that these were pre-existing samples used for comparison.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts
- This device measures a quantitative biomarker (EPO levels) in serum/plasma. Its "ground truth" is established by the measurements obtained from a legally marketed predicate device (Incstar's EPO-Trac 125I RIA), which is itself a laboratory assay.
- Therefore, there were no human "experts" in the sense of radiologists reviewing images to establish a ground truth. The predicate device's results served as the reference.
4. Adjudication Method for the Test Set
- Not applicable. This was a quantitative method comparison study against a predicate device, not a study involving human interpretation or adjudication of cases. The comparison was statistical.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve with AI vs. Without AI Assistance
- No. This is an in vitro diagnostic (IVD) device for quantitative measurement of a biomarker, not an imaging device with AI assistance for human readers. Therefore, an MRMC study is not relevant here.
6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done
- Yes, this was effectively a standalone performance study. The IMMULITE® EPO system operates as an automated immunoassay analyzer, and its output (EPO concentration) is compared directly to the output of the predicate device (Incstar's EPO-Trac 125I RIA). There is no "human-in-the-loop" performance in the measurement process itself, beyond the initial sample collection and loading.
7. The Type of Ground Truth Used
- Ground Truth: The quantitative EPO concentration values obtained from the predicate device, Incstar's EPO-Trac 125I RIA.
8. The Sample Size for the Training Set
- This document describes a 510(k) submission, which focuses on demonstrating substantial equivalence of a new device to a predicate device. For such devices, training set information for an "algorithm" as understood in AI/ML is typically not provided or relevant. The IMMULITE® EPO system is a chemiluminescent immunoassay, a chemical and biological measurement system, not a machine learning algorithm that requires a "training set" in the computational sense.
- The assay itself is developed and validated, but the term "training set" doesn't apply in the same way as it would for an AI-powered diagnostic.
9. How the Ground Truth for the Training Set Was Established
- Not applicable for the reasons mentioned in point 8. The device's calibration and assay parameters would be established through a different set of internal validation experiments and reference materials, not a "training set" with ground truth in the AI sense.
Ask a specific question about this device
Page 1 of 1