(260 days)
The Access Toxo IgG assay is a paramagnetic-particle, chemiluminescent immunoassay for the qualitative and quantitative determination of IgG antibodies to Toxoplasma gondii in human serum using the Access Immunoassay Systems. The Access Toxo IgG assay aids in the diagnosis of Toxoplasma gondii infection and may be used to assess the immune status of pregnant women.
This product is not FDA cleared/approved for the screening of blood or plasma donors. Assay performance characteristics have not been established for immunocompromised or immunosuppressed patients, cord blood, neonatal specimens or infants.
The Access Toxo IgG assay is a paramagnetic-particle, chemiluminescent immunoassay for the qualitative and quantitative detection of Toxoplasma gondii-specific IgG antibody in adult human serum using the Access Immunoassay Systems.
The Access Toxo IgG assay consists of the reagent pack, calibrators, and quality controls (OCs), packaged separately. Other items needed to run the assay include substrate and wash buffer.
This document describes the premarket notification (510(k)) for the Beckman Coulter Access Toxo IgG assay, a chemiluminescent immunoassay for detecting IgG antibodies to Toxoplasma gondii in human serum. This product is intended to aid in the diagnosis of Toxoplasma gondii infection and assess the immune status of pregnant women.
The submission claims substantial equivalence to a legally marketed predicate device, the Access Toxo IgG assay (K080869). The primary difference highlighted is the instrument used: the new device runs on the DxI 9000 Access Immunoassay Analyzer, while the predicate runs on the Access 2 Immunoassay System.
Here's an analysis of the provided information, focusing on the study that proves the device meets the acceptance criteria:
1. Table of Acceptance Criteria and Reported Device Performance
Strictly speaking, the document does not present "acceptance criteria" in a separate table with yes/no compliance. Instead, it details specific performance metrics and their measured values. The implicit acceptance criterion for most analytical performance studies (like imprecision and method comparison) is that the new device's performance is acceptable for its intended use and comparable to or better than the predicate. For Linearity, LoB, LoD, and LoQ, the acceptance criterion is that the study supports the claimed values.
| Performance Characteristic | Acceptance Criteria (Implicit from Study Design/Claims) | Reported Device Performance (Access Toxo IgG on DxI 9000) |
|---|---|---|
| Method Comparison (vs. Access 2 Immunoassay System) | High Positive Percent Agreement (PPA) and Negative Percent Agreement (NPA) to demonstrate interchangeability between instruments. | PPA: 100.00% (40/40) with 95% CI = 91.24% to 100% (for Reactive samples) NPA: 100.00% (99/99) with 95% CI = 96.26% to 100.00% (for Non-Reactive samples) |
| Imprecision (Within-Laboratory) | SD < 0.64 IU/mL at concentrations < 3.2 IU/mL; CV < 20.0% at concentrations > 3.2 IU/mL. (These are the design criteria mentioned, implying they are the acceptance threshold.) | Sample 1 (2.7 IU/mL): Overall Precision SD 0.38 (13.9% CV) - Meets SD criterion (0.38 < 0.64) Samples 2-6 (8.3 - 361.0 IU/mL): Overall Precision CVs range from 9.6% to 12.1% - Meets CV criterion (all < 20.0%) |
| Imprecision (Reproducibility / Between-Instrument) | Not explicitly stated as a separate numerical criterion beyond the general imprecision goal, but implicitly to show consistent performance across instruments. | Sample 1 (2.5 IU/mL): Reproducibility SD 0.29 (11.9% CV) Samples 2-6 (7.9 - 377.5 IU/mL): Reproducibility CVs range from 8.9% to 13.4% |
| Linearity | Demonstrated linearity across the claimed measuring interval. | Demonstrated linearity across the measuring interval of 3.2 - 450 IU/mL. |
| Limit of Blank (LoB) | Claimed value supported by study. | 0.8 IU/mL (claimed) |
| Limit of Detection (LoD) | Claimed value supported by study. | 1.5 IU/mL (claimed) |
| Limit of Quantitation (LoQ) | Claimed value supported by study. | 3.2 IU/mL (claimed) |
2. Sample Sizes Used for the Test Set and Data Provenance
-
Method Comparison:
- Test Sample Size: 140 native serum samples.
- Data Provenance: The document does not explicitly state the country of origin or whether the samples were retrospective or prospective. It refers to "native serum samples," which typically implies real-world patient samples.
-
Imprecision (Within-Laboratory):
- Test Sample Size: 6 serum samples (analyzed across 240 replicates each).
- Data Provenance: Not specified for origin or prospective/retrospective.
-
Imprecision (Reproducibility):
- Test Sample Size: 6 serum samples (analyzed across 225 replicates each).
- Data Provenance: Performed at an "internal site." Not specified for origin or prospective/retrospective.
-
Linearity, LoB, LoD, LoQ: The type and number of samples used for these studies are not specified, only that "a study" was performed and values were "claimed."
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts
This type of information (experts, qualifications) is primarily relevant for studies involving human interpretation or clinical diagnosis, such as imaging studies where radiologists establish ground truth.
For this in vitro diagnostic (IVD) device (an immunoassay), the "ground truth" for the test results is established by the measurement itself and comparison to a reference method or validated system (the predicate device). There are no human experts involved in interpreting the raw assay output to establish ground truth for method comparison or analytical performance studies like imprecision. The "truth" is the analytical measurement obtained by the reference method (the predicate device or a highly characterized sample).
4. Adjudication Method for the Test Set
Not applicable. As this is an IVD immunoassay, not a human reader study, adjudication methods (like 2+1 or 3+1 for imaging reads) are not relevant here. The comparison is between two analytical instruments/systems.
5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve with AI vs Without AI Assistance
Not applicable. This device is an in vitro diagnostic immunoassay. It does not involve human readers interpreting images assisted by AI, nor is it an AI-driven device in the sense of image analysis. Its function is to quantitatively and qualitatively measure an analyte in a biological sample.
6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) Was Done
Yes, the primary studies listed (Method Comparison, Imprecision, Linearity, LoB, LoD, LoQ) are all "standalone" in the sense that they evaluate the analytical performance of the device itself (Access Toxo IgG assay on the DxI 9000 Access Immunoassay Analyzer) without human interpretation affecting the measurement outcome. The device's output is a quantitative (IU/mL) or qualitative (Reactive/Non-Reactive/Equivocal) result.
7. The Type of Ground Truth Used (expert consensus, pathology, outcomes data, etc.)
For the method comparison study, the "ground truth" was essentially the results obtained from the predicate device (Access Toxo IgG assay on the Access 2 Immunoassay System). The goal was to show that the new device on the new instrument produced comparable results to the already cleared predicate.
For the analytical performance studies (Imprecision, Linearity, LoB, LoD, LoQ), the "ground truth" refers to the inherent analytical properties of the assay and instrument, which are determined by testing characterized samples (e.g., known concentrations for linearity, low-concentration samples for LoD/LoB).
8. The Sample Size for the Training Set
This document does not provide information about a "training set" in the context of machine learning or AI models. For IVD assays, "training" typically refers to the development and optimization process that precedes validation studies. The analytical performance studies presented here are primarily validation studies.
9. How the Ground Truth for the Training Set Was Established
Not applicable, as a "training set" in the context of AI/ML ground truth establishment is not relevant to this type of IVD device submission. Assay development and optimization for IVDs involve extensive R&D, but the concept of "ground truth for training" as seen in AI is not directly applicable to a chemical immunoassay.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
[3/28/2025]
Beckman Coulter Inc. Audree Demmers Director Quality Assurance 1000 Lake Hazeltine Drive Chaska, Minnesota 55318
Re: K242022
Trade/Device Name: Access Toxo IgG Regulation Number: 21 CFR 866.3780 Regulation Name: Toxoplasma Gondii Serological Reagents Regulatory Class: Class II Product Code: LGD Dated: July 10, 2024 Received: July 11, 2024
Dear Audree Demmers:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safetyreporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatory
{2}------------------------------------------------
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
JORGE L.
MUNOZ-S
Digitally signed by JORGE L. MUNOZ -S Date: 2025.03.28 11:34:01 -04'00'
Jorge Munoz, Ph.D. Deputy Branch Chief Division of Microbiology Devices OHT7: Office of In Vitro Diagnostics Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
Indications for Use
510(k) Number (if known) K242022
Device Name Access Toxo IgG
Indications for Use (Describe)
The Access Toxo IgG assay is a paramagnetic-particle, chemiluminescent immunoassay for the qualitative and quantitative determination of IgG antibodies to Toxoplasma gondii in human serum using the Access Immunoassay Systems. The Access Toxo IgG assay aids in the diagnosis of Toxoplasma gondii infection and may be used to assess the immune status of pregnant women.
This product is not FDA cleared/approved for the screening of blood or plasma donors. Assay performance characteristics have not been established for immunocompromised or immunosuppressed patients, cord blood, neonatal specimens or infants.
| Type of Use (Select one or both, as applicable) |
|---|
| ------------------------------------------------- |
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
510 (k) Summary
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.
| 510(k) Number | K242022 |
|---|---|
| --------------- | --------- |
Date Prepared: ____________ 03/25/2025
Submitted By:
Beckman Coulter, Inc. 1000 Lake Hazeltine Drive Chaska, MN 55318
Primary Contact:
Audree Demmers Director Quality Assurance Phone: (651) 605-5278 Email: aldemmers@beckman.com
Alternate Contact:
Kate Oelberg Senior Staff Ouality and Regulatory Affairs Phone: (612) 431-7315 Email: kmoelberg@beckman.com
Regulatory Information
Regulation Number: 21 CFR 866.3780 Regulation Description: Toxoplasma gondii serological reagents Trade Name: Access Toxo IgG Classification Name: Enzyme Linked Immunoabsorbent Assay, Toxoplasma Gondii Product Code: LGD Device Class: II
Predicate Device
Device Name: Access Toxo IgG 510(k) Numbers: K080869
Device Description
The Access Toxo IgG assay is a paramagnetic-particle, chemiluminescent immunoassay for the qualitative and quantitative detection of Toxoplasma gondii-specific IgG antibody in adult human serum using the Access Immunoassay Systems.
The Access Toxo IgG assay consists of the reagent pack, calibrators, and quality controls (OCs), packaged separately. Other items needed to run the assay include substrate and wash buffer.
{5}------------------------------------------------
Intended Use
The Access Toxo IgG assay is a paramagnetic-particle, chemiluminescent immunoassay for the qualitative and quantitative determination of IgG antibodies to Toxoplasma gondii in human serum, using the Access Immunoassay Systems. The Access Toxo IgG assay aids in the diagnosis of Toxoplasma gondii infection and may be used to assess the immune status of pregnant women.
This product is not FDA cleared/approved for the screening of blood or plasma donors. Assay performance characteristics have not been established for immunocompromised or immunosuppressed patients, cord blood, neonatal specimens or infants.
| Device &PredicateDevice(s): | Candidate Test | Predicate |
|---|---|---|
| K242022 | K080869 | |
| Device TradeName | Access Toxo IgG | Access Toxo IgG |
| General DeviceCharacteristicSimilarities | ||
| IntendedUse/IndicationsFor Use | Same | The Access Toxo IgG assay is aparamagnetic-particle,chemiluminescent immunoassay forthe qualitative and quantitativedetermination of IgG antibodies toToxoplasma gondii in human serumusing the Access ImmunoassaySystems. The Access Toxo IgG assayaids in the diagnosis of Toxoplasmagondii infection and may be used toassess the immune status of pregnantwomen.This product is not FDAcleared/approved for the screening ofblood or plasma donors. Assayperformance characteristics have notbeen established forimmunocompromised orimmunosuppressed patients, cordblood, neonatal specimens or infants. |
| Analyte | Same | IgG antibody to T. gondii |
| Technology | Same | Two-step immunoenzymatic assay |
| Format | Same | Chemiluminescent |
| Method | Same | Automated |
Comparison with Predicate(s)
{6}------------------------------------------------
| Device &PredicateDevice(s): | Candidate Test | Predicate |
|---|---|---|
| K242022 | K080869 | |
| Calibration | Same | Utilizes a stored calibration curve |
| CalibrationFrequency | Same | 28 days |
| Sample Type | Same | Serum |
| AnalyticalMeasuringInterval | 3.2 - 450 IU/mL | 0 - 450 IU/mL |
| ResultsInterpretation | Same | <7.5 IU/mL: Non-Reactive≥7.5 - <10.5 IU/mL: Equivocal≥10.5 IU/mL: Reactive |
| CaptureReagent | Same | Paramagnetic particles coated with T. gondii (strain RH) solubilizedmembrane antigen |
| DetectionAntibody | Same | Mouse monoclonal anti-human IgGantibody- alkaline phosphatase(bovine) conjugate |
| Open KitStability | Same | 28 days after opening, 2-10°C |
| General DeviceCharacteristicDifferences | ||
| Substrate | Lumi-Phos PRO substrate | Access Substrate |
| Instrument | DxI 9000 Access Immunoassay Analyzer | Access 2 Immunoassay System |
Standard/Guidance Document Referenced (if applicable):
CLSI EP05-A3: Evaluation of Precision Performance of Quantitative Measurement Methods; Approved Guideline – Third Edition
CLSI EP06-2nd Edition-: Evaluation of the Linearity of Quantitative Measurement Procedures: A Statistical Approach; Approved Guideline
CLSI EP17-A2: Evaluation of Detection Capability for Clinical Laboratory Measurement Procedures; Approved Guideline – Second Edition
CLSI EP12-Ed3, Evaluation of Qualitative, Binary Output Examination Performance, Third Edition.
{7}------------------------------------------------
Summary of Studies
Method Comparison:
A method comparison of 140 native serum samples using the Access Toxo IgG assay on the DxI 9000 Access Immunoassay Analyzer and the Access 2 Immunoassay System was conducted to compare the performance on both systems. Positive percent agreement (PPA) and negative percent agreement (NPA) between the Access Toxo IgG assay run on the DXI 9000 Immunoassay Analyzer and the Access 2 Immunoassay System was calculated for the Access Toxo IgG assay and are shown in Table 1.
Table 1: Performance Agreement of Access Toxo IgG assay on the Access 2 Immunoassay Analyzer and the DxI 9000 Immunoassay Analyzer (n=140)
| Access Toxo IgG | Access 2 Immunoassay Analyzer | ||||
|---|---|---|---|---|---|
| Reactive | Equivocal | Non-reactive | Total | ||
| DxI9000ImmunoassayAnalyzer | Reactive | 40 | 0 | 0 | 40 |
| Equivocal | 0 | 1 | 0 | 1 | |
| Non-Reactive | 0 | 0 | 99 | 99 | |
| Total | 40 | 1 | 99 | 140 | |
| PPAa | 100.00% (40/40) | 95% CI = 91.24% to 100% | |||
| NPAb | 100.00% (99/99) | 95% CI = 96.26% to 100.00 |
a, b95% CI for PPA and NPA were estimated using the Wilson score method.
Imprecision:
Precision studies were designed to have within-laboratory imprecision as listed below:
- · < 0.64 IU/mL SD at concentrations < 3.2 IU/mL
- · < 20.0% CV at concentrations > 3.2 IU/mL
a) Within-Laboratory Precision: A study based on CLSI EP05-A3 performed on the DxI 9000 Access Immunoassay Analyzer tested multiple samples in 2 runs per day for 20 days. Six serum samples were tested using 3 reagent lot/calibrator lot combinations, where a unique reagent lot and calibrator lot were paired. The data were analyzed for repeatability
{8}------------------------------------------------
(within-run), between-run, between-day, between lot and instrument and overall precision. Within-laboratory precision data summary is shown in Table 2.
| Sample | N | Mean(IU/mL) | Repeatability(Within-Run) | Between-Run | Between-Day | Between LotandInstrumenta | OverallPrecisionb | |||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|
| SD | %CV | SD | %CV | SD | %CV | SD | %CV | SD | %CV | |||
| Sample 1 | 240 | 2.7 | 0.15 | 5.3 | 0.16 | 5.8 | 0.10 | 3.6 | 0.29 | 10.8 | 0.38 | 13.9 |
| Sample 2 | 240 | 8.3 | 0.36 | 4.3 | 0.24 | 2.9 | 0.43 | 5.2 | 0.61 | 7.3 | 0.86 | 10.4 |
| Sample 3 | 240 | 13.1 | 0.56 | 4.3 | 0.54 | 4.2 | 0.35 | 2.7 | 1.34 | 10.2 | 1.59 | 12.1 |
| Sample 4 | 240 | 69.3 | 3.03 | 4.4 | 3.17 | 4.6 | 2.17 | 3.1 | 6.56 | 9.5 | 8.18 | 11.8 |
| Sample 5 | 240 | 174.2 | 7.29 | 4.2 | 6.80 | 3.9 | 6.28 | 3.6 | 14.01 | 8.0 | 18.30 | 10.5 |
| Sample 6 | 240 | 361.0 | 20.45 | 5.7 | 12.55 | 3.5 | 20.38 | 5.6 | 14.07 | 3.9 | 34.48 | 9.6 |
Table 2: Access Toxo IgG assay 20-day Within-Laboratory Precision on the DxI 9000 Access Immunoassay Analyzer
4 Access Toxo IgG reagent lot, Access Toxo IgG calibrator lot and DxI 9000 instruments are confounded, and the confounding effect is represented by between-lot and instrument.
b Overall within-laboratory variability includes within-run, between-day, and between-lot variance components.
- b) Reproducibility (between-Instrument Precision): A reproducibility study based on CLSI EP05-A3 performed on the DxI 9000 Access Immunoassay Analyzer tested multiple samples in replicates of 5 in 1 run per day for a minimum of 5 days. The study tested 6 serum samples with the Access Toxo IgG on 3 DxI 9000 Access Immunoassay Analyzers in an internal site. The samples were tested with 3 lots of Access Toxo IgG reagents, and 1 lot of Access Toxo IgG calibrator on each instrument. Summary of the data is shown in Table 3.
Table 3: Access Toxo IgG assay Reproducibility on the DxI 9000 Access Immunoassay Analyzer
| Sample | N | Mean(IU/mL) | Repeatability(Within-run) | Between-Day/Runª | BetweenReagent Lot | Between-Instrument | Reproducibilityb | |||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|
| SD | %CV | SD | %CV | SD | %CV | SD | %CV | SD | %CV | |||
| Sample 1 | 225 | 2.5 | 0.16 | 6.4 | 0.17 | 7.0 | 0.11 | 4.5 | 0.14 | 5.6 | 0.29 | 11.9 |
| Sample 2 | 225 | 7.9 | 0.40 | 5.0 | 0.42 | 5.3 | 0.25 | 3.1 | 0.35 | 4.5 | 0.72 | 9.1 |
| Sample 3 | 225 | 12.4 | 0.57 | 4.6 | 0.51 | 4.1 | 0.58 | 4.7 | 0.55 | 4.4 | 1.11 | 8.9 |
| Sample 4 | 225 | 67.4 | 3.17 | 4.7 | 2.96 | 4.4 | 3.66 | 5.4 | 2.90 | 4.3 | 6.37 | 9.4 |
| Sample 5 | 225 | 169.7 | 7.83 | 4.6 | 9.04 | 5.3 | 9.64 | 5.7 | 6.24 | 3.7 | 16.58 | 9.8 |
| Sample 6 | 225 | 377.5 | 25.58 | 6.8 | 30.77 | 8.2 | 22.70 | 6.0 | 21.45 | 5.7 | 50.76 | 13.4 |
a Days and runs are confounded.
b Reproducibility includes within-run, between-day/run, between-instrument variance components.
{9}------------------------------------------------
Linearity: A study based on CLSI EP06-ED2 performed with the Access Toxo IgG assay on the DxI 9000 Access Immunoassay Analyzer determined the assay demonstrated linearity across the measuring interval of 3.2 - 450 IU/mL.
Limit of Blank (LoB): The claimed LoB for Access Toxo IgG assay on the DxI 9000 Access Immunoassay Analyzer is 0.8 IU/mL.
Limit of Detection (LoD): The claimed LoD for Access Toxo IgG assay on the DxI 9000 Access Immunoassay Analyzer is 1.5 IU/mL.
Limit of Quantitation (LoQ): The claimed LoQ for Access Toxo IgG assay on the DxI 9000 Access Immunoassay Analyzer is 3.2 IU/mL.
Substantial Equivalence Comparison Conclusion
Beckman Coulter's Access Toxo IgG assay on the DxI 9000 Access Immunoassay Analyzer is substantially equivalent to the Access Toxo IgG assay on the Access 2 Immunoassay System as demonstrated through the information and data provided in this submission. The performance testing presented in this submission provides evidence that the device is safe and effective in its intended use.
§ 866.3780
Toxoplasma gondii serological reagents.(a)
Identification. Toxoplasma gondii serological reagents are devices that consist of antigens and antisera used in serological tests to identify antibodies toToxoplasma gondii in serum. Additionally, some of these reagents consist of antisera conjugated with a fluorescent dye (immunofluorescent reagents) used to identifyToxoplasma gondii from clinical specimens. The identification aids in the diagnosis of toxoplasmosis caused by the parasitic protozoanToxoplasma gondii and provides epidemiological information on this disease. Congenital toxoplasmosis is characterized by lesions of the central nervous system, which if undetected and untreated may lead to brain defects, blindness, and death of an unborn fetus. The disease is characterized in children by inflammation of the brain and spinal cord.(b)
Classification. Class II (performance standards).