Search Filters

Search Results

Found 45 results

510(k) Data Aggregation

    K Number
    K250250
    Date Cleared
    2025-10-17

    (262 days)

    Product Code
    Regulation Number
    866.5870
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    510k Summary Text (Full-text Search) :

    Trade/Device Name:** ADVIA Centaur Anti-Thyroid Peroxidase II (aTPOII)
    Regulation Number: 21 CFR 866.5870
    ADVIA Centaur aTPOII (K250250) Page 2 of 14**
    October 2025

    | Regulation Number | 21 CFR 866.5870

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The ADVIA Centaur Anti-Thyroid Peroxidase II (aTPOII) assay is for in vitro diagnostic use in the quantitative measurement of autoantibodies against thyroid peroxidase in human serum and plasma (EDTA and lithium heparin) using the ADVIA Centaur XP system.

    Anti-thyroid peroxidase (aTPO) measurements are used, in conjunction with a clinical assessment, as an aid in the diagnosis of autoimmune thyroiditis and/or Graves' disease.

    Device Description

    The ADVIA Centaur Anti-Thyroid Peroxidase II (aTPOII) consists of:

    • aTPOII ReadyPack® primary reagent pack (Lite Reagent, Solid Phase)
    • aTPOII CAL
      Devices sold separately and included in the ADVIA Centaur® Anti-Thyroid Peroxidase II (aTPOII) are:
    • ADVIA Centaur aTPOII MCM (MCM 1, MCM 2–4)
    • ADVIA Centaur aTPOII QC
    • ADVIA Centaur aTPOII DIL ReadyPack ancillary reagent pack
    AI/ML Overview

    N/A

    Ask a Question

    Ask a specific question about this device

    K Number
    K240469
    Date Cleared
    2024-08-09

    (171 days)

    Product Code
    Regulation Number
    866.5870
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    510k Summary Text (Full-text Search) :

    Chaska, Minnesota 55318

    Re: K240469

    Trade/Device Name: Access TPO Antibody Regulation Number: 21 CFR 866.5870
    Access Immunoassay Analyzer Classification Name: Thyroid Autoantibody Classification Requlation: [21 CFR 866.5870

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Access TPO Antibody assay is a paramagnetic particle, chemiluminescent immunoassay for the quantitative determination of thyroperoxidase antibody (TPOAb) levels in human serum and plasma using the Access Immunoassay Systems.

    The detection of TPOAb is an aid in the diagnosis of thyroid autoimmune disorders.

    Device Description

    The Access TPO Antibody assay is a paramagnetic particle, chemiluminescent immunoassay for the quantitative determination of thyroperoxidase antibody (TPO Ab) levels in human serum and plasma using the Access Immunoassay Systems.

    The Access TPO Antibody Calibrators are intended to calibrate the Access TPO Antibody assay for the quantitative determination of TPO Antibody levels in human serum and plasma using the Access Immunoassay Systems.

    The Access TPO Antibody assay is a sequential two-step immunoenzymatic ("sandwich") assay. A sample is added to a reaction vessel with paramagnetic particles coated with thyroperoxidase protein. The serum or plasma TPO Ab binds to the thyroperoxidase. After incubation, materials bound to the solid phase are held in a magnetic field while unbound materials are washed away. Then, the chemiluminescent substrate is added to the vessel and light generated by the reaction is measured with a luminometer. The light production is directly proportional to the concentration of analyte in the sample. Analyte concentration is automatically determined from a stored calibration.

    AI/ML Overview

    The provided document, a 510(k) summary for the Beckman Coulter Access TPO Antibody assay, describes the analytical performance studies conducted to demonstrate substantial equivalence to a predicate device. This is a common regulatory pathway for in vitro diagnostic devices, focusing on demonstrating that the new device is as safe and effective as a legally marketed predicate device.

    Here's an analysis of the acceptance criteria and the study that proves the device meets them, based on the provided text:

    Acceptance Criteria and Reported Device Performance

    The document presents acceptance criteria implicitly through the study designs and results, as it's a submission for an IVD kit rather than an AI/ML algorithm. The performance is assessed by comparing the new device (Access TPO Antibody Assay on Dxl 9000 Access Immunoassay System) to its predicate device (Access TPO Antibody Assay on Access 2 Immunoassay System).

    Table of Acceptance Criteria and Reported Device Performance:

    Performance CharacteristicAcceptance Criteria (Implicit from Study Design/Goal)Reported Device Performance (Access TPO Antibody Assay on Dxl 9000)Study Method/Description
    Method Comparison (Agreement with Predicate)A strong correlation and minimal bias between results from the new device and the predicate device, demonstrated by slope and intercept close to 1 and 0 respectively, and a high correlation coefficient.Slope: 1.06 (95% CI: 1.04 - 1.08) Intercept: -0.26 (95% CI: -0.32 - -0.22) Correlation Coefficient (R): -0.978 (Note: The negative sign for R with a positive slope is unusual and might be a typo in the document; typically, a strong positive correlation is indicated by R closer to +1 for agreement studies).CLSI EP09c, 3rd Edition; Passing-Bablok regression.
    Imprecision (Repeatability and Reproducibility)Within-laboratory imprecision meeting predefined thresholds: - < 0.07 IU/mL SD at concentrations < 0.6 IU/mL - < 12.0% CV at concentrations ≥ 0.6 IU/mL and < 450.0 IU/mL - < 15.0% CV at concentrations ≥ 450.0 IU/mLSample 1 (0.35 IU/mL): Within-Laboratory SD: 0.02, %CV: 6.9 (Meets <0.07 SD) Sample 2 (5.5 IU/mL): Within-Laboratory SD: 0.36, %CV: 6.7 (Meets <12.0% CV) Sample 3 (20 IU/mL): Within-Laboratory SD: 1.23, %CV: 6.2 (Meets <12.0% CV) Sample 4 (318 IU/mL): Within-Laboratory SD: 21.35, %CV: 6.7 (Meets <12.0% CV) Sample 5 (747 IU/mL): Within-Laboratory SD: 96.48, %CV: 12.9 (Meets <15.0% CV)CLSI EP05-A3; multiple samples in duplicate for 2 runs/day over ≥20 days.
    LinearityAssay demonstrates linearity across the claimed measuring interval (0.25 - 1,000 IU/mL)."the assay demonstrated linearity across the measuring interval."CLSI EP06-Ed2
    Detection Capability (LoB, LoD, LoQ)Established quantitative limits relevant for clinical use, demonstrating the ability to detect and quantify low concentrations of the analyte.Limit of Blank (LoB): 0.19 IU/mL Limit of Detection (LoD): 0.23 IU/mL Limit of Quantitation (LoQ): 0.25 IU/mLCLSI EP17-A2

    Study Details:

    This submission focuses on an In Vitro Diagnostic (IVD) assay (Access TPO Antibody), not an AI/ML algorithm. Therefore, many of the questions related to AI/ML development (training set, experts for ground truth, MRMC studies) are not directly applicable in the way they would be for an AI-based medical device. However, I will interpret them in the context of an IVD assay where possible, based on standard IVD validation practices.

    1. Sample Size Used for the Test Set and Data Provenance:

      • Method Comparison: N = 219 patient samples.
      • Imprecision: N = 80 replicates (for each of 5 samples).
      • Linearity & Detection Capability: Sample sizes for these studies are not explicitly stated, but are governed by CLSI guidelines (EP06-Ed2 and EP17-A2 respectively), which typically involve analyzing serially diluted samples.
      • Data Provenance: Not explicitly stated (e.g., country of origin). The studies appear to be analytical performance studies conducted on patient samples, but whether they were retrospective or prospective is not specified. Given the nature of an IVD, these are typically conducted prospectively in a lab setting using banked or newly collected samples.
    2. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of those Experts:

      • Not Applicable in the traditional AI/ML sense. For an IVD assay, the "ground truth" for the test set is either:
        • Clinical Diagnosis: For the diagnostic claim ("aid in the diagnosis of thyroid autoimmune disorders"), the ground truth would be established by independent clinical diagnosis (e.g., by endocrinologists) typically using a variety of clinical data and other diagnostic tests. This information is usually part of clinical utility studies, which are not detailed in this analytical performance summary.
        • Reference Method/Predicate Product: For the analytical performance studies like Method Comparison, the "ground truth" or reference is the result obtained from the predicate device (Access TPO Antibody Assay on Access 2 Immunoassay System). The predicate device itself is validated against established (often clinical) criteria.
        • Known Concentrations/Materials: For linearity, imprecision, and detection capability studies, the "ground truth" is established by using characterized materials with known concentrations, or by measuring the variability around the mean of repeated measurements. This doesn't involve medical experts in the adjudication sense.
    3. Adjudication Method for the Test Set:

      • Not Applicable. Adjudication is typically for subjective interpretations (like image reading). For an IVD, the results are quantitative measurements from an instrument. The "adjudication" is essentially the statistical analysis (e.g., Passing-Bablok regression, CV calculations) of these quantitative results against predefined criteria or a reference method.
    4. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done:

      • No. An MRMC study is relevant for interpreting subjective data (like medical images) where human readers are involved. This document describes the analytical performance of an in vitro diagnostic assay that produces quantitative results, not an image interpretation or decision support AI.
    5. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done:

      • Yes, in the context of an IVD. The entire analytical performance evaluation (Method Comparison, Imprecision, Linearity, Detection Capability) describes the standalone performance of the Access TPO Antibody assay on the Dxl 9000 system. It functions automatically based on its chemiluminescent immunoassay technology to produce a quantitative result. There is no "human-in-the-loop" aspect to the measurement itself, though human lab personnel operate the instrument and interpret the results.
    6. The Type of Ground Truth Used:

      • Known concentrations, reference method results (predicate device), and statistical averages.
        • For Method Comparison, the "ground truth" for comparison is the result obtained from the predicate Access 2 Immunoassay System.
        • For Imprecision, the "ground truth" is the mean concentration of the samples derived from repeated measurements, and the variability around that mean.
        • For Linearity, the ground truth is derived from known dilutions of a high-concentration material.
        • For Detection Capability (LoB, LoD, LoQ), the ground truth involves the statistical analysis of measurements of blank samples and low-concentration samples.
    7. The Sample Size for the Training Set:

      • Not Applicable. This is an IVD assay relying on chemical and immunological reactions, not a machine learning algorithm that requires a "training set" in the AI/ML sense. The device's parameters are determined during its design and optimization phases by Beckman Coulter, based on established immunoassay principles.
    8. How the Ground Truth for the Training Set was Established:

      • Not Applicable. As above, there is no "training set" for this type of device in the AI/ML context. The assay's performance characteristics (e.g., reactivity, linearity, precision) are inherent to its reagent formulation and instrument design, which are optimized by the manufacturer using internal R&D processes guided by scientific principles and regulatory standards.
    Ask a Question

    Ask a specific question about this device

    K Number
    K240996
    Date Cleared
    2024-07-03

    (83 days)

    Product Code
    Regulation Number
    866.5870
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    510k Summary Text (Full-text Search) :

    Minnesota 55318

    Re: K240996

    Trade/Device Name: Access Thyroglobulin Antibody II Regulation Number: 21 CFR 866.5870
    Classification Regulation: 21 CFR 866.5870 Classification Product Code: JNL

    Predicate Device

    Device

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Access Thyroglobulin Antibody II assay is a paramagnetic particle, chemiluminescent immunoassay for the quantitative determination of thyroglobulin antibody levels in human serum and plasma using the Access Immunoassay Systems. The measurement of thyroid autoantibodies may aid in the diagnosis of Hashimoto's disease, nontoxic goiter, and Graves' disease.

    Device Description

    The Access Thyroqlobulin Antibody II assay is a paramagnetic particle, chemiluminescent immunoassay for the quantitative determination of thyroglobulin antibody levels in human serum and plasma using the Access Immunoassay Systems. The measurement of thyroid autoantibodies may aid in the diagnosis of Hashimoto's disease, nontoxic goiter, and Graves' disease.

    The Access Thyroglobulin Antibody II assay is a sequential two-step immunoenzymatic ("sandwich") assay. A sample is added to a reaction vessel with paramagnetic particles coated with the thyroglobulin protein. The TgAb in the sample binds to the thyroglobulin coated on the particles. After incubation, materials bound to the solid phase are held in a magnetic field while unbound materials are washed away. The thyroglobulin-alkaline phosphatase conjugate is added and binds to the TgAb.

    After second incubation, materials bound to the solid phase are held in a magnetic field while unbound materials are washed away. Then, the chemiluminescent substrate is added to the vessel and light generated by the reaction is measured with a luminometer. The light production is directly proportional to the concentration of analyte in the sample. Analyte concentration is automatically determined from a stored calibration.

    AI/ML Overview

    Here's a detailed breakdown of the acceptance criteria and study information for the Beckman Coulter Access Thyroglobulin Antibody II device, extracted from the provided FDA 510(k) summary:

    1. Table of Acceptance Criteria and Reported Device Performance

    ParameterAcceptance Criteria (Predicate Device)Reported Device Performance (Modified Device - Dxl 9000 Access Immunoassay Analyzer)
    Intended UseQuantitative determination of thyroglobulin antibody levels in human serum and plasma to aid in the diagnosis of Hashimoto's disease, nontoxic goiter, and Graves' disease.Same (No change in Intended Use)
    Analyte MeasuredThyroglobulin AntibodySame
    Technology / Format / MethodSandwich immunoassay / Chemiluminescent / AutomatedSame
    Sample TypeHuman serum or plasmaSame
    Sample Volume10 uLSame
    Measuring Range1.5 - 2,500 IU/mLSame
    Blocker ReagentsBiotin and alkaline phosphatase included in reagent pack as blockersSame
    Biotin InterferenceNo significant interference (± 10%) observed in samples containing up to 3,510 ng/mL of biotin.Same (Explicitly stated in the comparison table)
    Imprecision (Repeatability)SD ≤ 1.5 for values < 15 IU/mL; CV ≤ 10.0% for values ≥ 15 IU/mL and < 1000 IU/mL; CV ≤ 15.0% for values ≥ 1000 IU/mLWithin-Laboratory Imprecision: - Sample 1 (2.4 IU/mL): 5.2% CV (meets ≤ 1.5 IU/mL SD, which is equivalent to 62.5% CV. The actual SD is 0.1, making the %CV 4.2% for repeatability) - Sample 2 (188 IU/mL): 4.1% CV (meets ≤ 10.0% CV) - Sample 3 (727 IU/mL): 4.2% CV (meets ≤ 10.0% CV) - (Partial data for Sample ~1000 IU/mL is cut off, but the predicate applies CV ≤ 15.0% for values ≥ 1000 IU/mL)
    ReproducibilityNot explicitly stated as a separate acceptance criterion for the predicate, but implied by the imprecision criteria.Reproducibility (Overall): - Sample 1 (2.6 IU/mL): 5.8% CV (within expected range for low concentration) - Sample 2 (184 IU/mL): 3.8% CV (well within 10.0% CV) - Sample 3 (744 IU/mL): 3.1% CV (well within 10.0% CV) - Sample 4 (1503 IU/mL): 3.6% CV (well within 15.0% CV) - Sample 5 (1966 IU/mL): 6.4% CV (well within 15.0% CV)
    LinearityThe assay demonstrates linearity across the measuring interval.Determined to demonstrate linearity across the measuring interval (no quantitative data given, but implied successful).
    Limit of Blank (LoB)Assumed to be ≤ 0.1 IU/mL (based on reported value)0.1 IU/mL
    Limit of Detection (LoD)Assumed to be ≤ 0.2 IU/mL (based on reported value)0.2 IU/mL
    Limit of Quantitation (LoQ)≤ 1.5 IU/mL for ≤ 20% within-lab CV1.5 IU/mL (at ≤ 20% within-lab CV)
    Method Comparison (Slope)Assumed to be close to 1.0 (for substantial equivalence to predicate)0.97 (95% CI: 0.95 – 0.99)
    Method Comparison (Intercept)Assumed to be close to 0 (for substantial equivalence to predicate)-0.37 (95% CI: -0.99 – 0.047)
    Method Comparison (Correlation Coefficient)Assumed to be close to 1.0 (for substantial equivalence to predicate)1.00

    Note: The primary goal of this submission (K240996) is to demonstrate substantial equivalence of the same device on a new instrument platform (Dxl 9000 Access Immunoassay Analyzer) compared to its predicate on the Access 2 Immunoassay System. Therefore, the "acceptance criteria" are largely derived from the established performance of the predicate device.


    2. Sample Sizes Used for the Test Set and Data Provenance

    • Method Comparison Study (CLSI EP09c):
      • Sample Size: N = 114
      • Data Provenance: Not explicitly stated (e.g., country of origin, retrospective/prospective). However, such studies typically use clinical samples that may be either retrospectively collected or prospectively collected for validation purposes.
    • Imprecision Study (CLSI EP05-A3):
      • Sample Size: 3 distinct "Samples" (concentrations) were tested. Each sample was tested 80 times (duplicate R.L per day for a minimum of 20 days).
      • Data Provenance: Not explicitly stated. Assumed to be laboratory samples or controls prepared for method validation.
    • Reproducibility Study (CLSI EP05-A3):
      • Sample Size: 5 distinct "Samples" (concentrations) were tested. Each sample was tested 75 times (replicates of 5 per day for a minimum of 5 days on 3 instruments).
      • Data Provenance: Not explicitly stated. Assumed to be laboratory samples or controls prepared for method validation.
    • Detection Capability (LoB, LoD, LoQ) Study (CLSI EP17-A2):
      • Sample Size: Not explicitly stated for each determination, but these studies typically involve multiple replicates of blank, low-level, and higher-level samples.
      • Data Provenance: Not explicitly stated. Assumed to be laboratory samples or controls.

    3. Number of Experts Used to Establish Ground Truth and Qualifications

    This document describes a clinical laboratory device (an immunoassay), not an AI/imaging device requiring expert interpretation of results for ground truth. Therefore, the concepts of "experts to establish ground truth" (in the context of clinical interpretation or diagnosis from an image) and their "qualifications" are not applicable here.

    For this type of device, ground truth is established through:

    • Reference Methods / Predicate Devices: The Access 2 Immunoassay System (predicate) served as the comparator for the method comparison study to assess the "truth" of the Dxl 9000 system's measurements.
    • Certified Reference Materials/Standards: Calibrators and controls with known analyte concentrations, often traceable to international standards, are used to establish accuracy and calibration.
    • Clinical Diagnosis: For the "Indications for Use," the device aids in diagnosis, meaning its results are interpreted by clinicians in conjunction with other clinical information. The diagnostic accuracy studies (sensitivity, specificity) in relation to a "gold standard" clinical diagnosis are typically part of a larger clinical trial not detailed in this specific 510(k) summary for a platform change.

    4. Adjudication Method for the Test Set

    Not applicable. As this is an immunoassay device assessing quantitative levels of an antibody, there is no "adjudication method" in the sense of reconciling divergent expert interpretations of qualitative or semi-quantitative data. The "test set" results—the quantitative values—are compared statistically to the reference method (predicate device) and assessed against performance specifications (imprecision, linearity, detection limits).


    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

    No, an MRMC comparative effectiveness study was not done. This type of study is relevant for AI/imaging devices where multiple human readers interpret complex cases (e.g., medical images) and AI assistance might improve their performance. This document is for a medical laboratory immunoassay for quantitative measurement of thyroglobulin antibody, not an AI-assisted diagnostic tool for human interpretation.


    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    While the device operates "standalone" in the sense that the instrument performs the assay and generates a quantitative result without human intervention during the analytical process, this is not an "algorithm-only" study in the context of AI. The performance studies (imprecision, linearity, method comparison, detection capability) represent the standalone analytical performance of the instrument/reagent system. The "human-in-the-loop" would be the clinician interpreting the numerical result in the context of a patient's overall clinical picture, but the device itself functions automatically.


    7. The Type of Ground Truth Used

    • Quantitative Reference Values: For the performance studies, the ground truth is established through various means:
      • Method Comparison: The results obtained from the predicate device (Access Thyroglobulin Antibody II on the Access 2 Immunoassay System) serve as the reference standard.
      • Imprecision & Reproducibility: Derived from repeated measurements of samples (often control materials or pooled patient samples) to assess variability. The "true" value for these samples is either assigned by the manufacturer or determined through extensive testing.
      • Linearity: Determined by creating serially diluted samples from a high-concentration sample, where the expected concentration of each dilution is the "ground truth."
      • Detection Capability (LoB, LoD, LoQ): Established using blank samples and low-concentration spiked samples, with statistical methods determining the lowest detectable/quantifiable levels.

    8. The Sample Size for the Training Set

    This document does not describe a machine learning or AI algorithm development that would typically involve a "training set." The studies described are for analytical validation of an immunoassay on a new instrument platform, focusing on demonstrating equivalent performance to a predicate device. Therefore, a "training set" in the AI sense is not applicable.


    9. How the Ground Truth for the Training Set Was Established

    Not applicable, as there is no "training set" in the context of AI or machine learning for this immunoassay device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K213517
    Date Cleared
    2023-09-26

    (692 days)

    Product Code
    Regulation Number
    866.5870
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    510k Summary Text (Full-text Search) :

    Minnesota 55318

    Re: K213517

    Trade/Device Name: Access Thyroglobulin Antibody II Regulation Number: 21 CFR 866.5870
    Classification Regulation: 21 CFR 866.5870 Classification Product Code: JZO

    {4}---------------------

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Access Thyroglobulin Antibody II assay is a paramagnetic particle, chemiluminescent immunoassay for the quantitative determination of thyroglobulin antibody levels in human serum and plasma using the Access Immunoassay Systems. The measurement of thyroid autoantibodies may aid in the diagnosis of Hashimoto's disease, nontoxic goter, and Graves' disease.

    Device Description

    The Access Thyroqlobulin Antibody II assay is a paramagnetic particle, chemiluminescent immunoassay for the quantitative determination of thyroglobulin antibody levels in human serum and plasma using the Access Immunoassay Systems. The measurement of thyroid autoantibodies may aid in the diagnosis of Hashimoto's disease, nontoxic goiter, and Graves' disease.

    The Access Thyroglobulin Antibody II assay consists of the reagent pack and calibrators. Other items needed to run the assay include substrate and wash buffers. The assay is run on Access 2 Immunoassay Analyzers.

    The device modifications described in this submission impact the Access Thyroqlobulin Antibody II reagent pack only; they do not impact or change the other components that are used with this reagent pack. The modification does not affect the intended use or indications of the device or alter the fundamental scientific technology of the device.

    A description of the reagent pack is provided below.

    WellIngredients
    R1a:Dynabeads* paramagnetic particles coated with streptavidin andcoupled to biotinylated human thyroglobulin, suspended in a TRISbuffer with protein (bovine), < 0.1% sodium azide, and 0.1%ProClin** 300.
    R1b:Human thyroglobulin-alkaline phosphatase (bovine) conjugate in aTRIS buffer with protein (bovine), < 0.1% sodium azide, and 0.1%ProClin 300.
    R1c:TRIS buffer with protein (bovine), < 0.1% sodium azide and 0.1%ProClin 300.
    R1d:TRIS buffer with blocking polymer, < 0.1% sodium azide and 0.1%ProClin 300.
    AI/ML Overview

    Here's an analysis of the provided text, outlining the acceptance criteria and study details for the "Access Thyroglobulin Antibody II" device:

    Device: Access Thyroglobulin Antibody II

    The study in the document focuses on the modified Access Thyroglobulin Antibody II assay and compares it to the previously cleared predicate device (Access Thyroglobulin Antibody II Assay, FDA 510(k) Number K112933). The goal is to demonstrate substantial equivalence of the modified device.


    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not explicitly present acceptance criteria in a dedicated table format. Instead, it describes performance characteristics and the results obtained. I've reconstructed a table based on the implicit criteria derived from the reported performance, especially where a target value or range is given (e.g., for imprecision, biases, or correlation).

    Performance CharacteristicAcceptance Criteria (Implicit/Explicit)Reported Device Performance (Modified Device)
    Measuring Range(Compared to predicate's 0.9-2,500 IU/mL)1.5-2,500 IU/mL
    ImprecisionPredicate: SD < 1.5 for values < 15 IU/mL; CV < 10% for values ≥ 15 IU/mLWithin Laboratory (Total): - SD ≤ 1.5 at concentrations < 15 IU/mL - CV ≤ 10.0 % at concentrations ≥ 15 IU/mL and < 1000 IU/mL - CV ≤ 15.0% for concentrations ≥ 1000 IU/mL
    Reproducibility(Not explicitly defined for predicate, but performance is reported)Reproducibility: - SD ≤ 2.3 at concentrations < 15 IU/mL - CV ≤ 15.0 % at concentrations ≥ 15 IU/mL and < 1000 IU/mL - CV ≤ 20.0% for concentrations ≥ 1000 IU/mL
    High-dose Hook EffectNo significant hook effect expected.No high-dose hook effect at concentrations up to at least 50,000 IU/mL.
    Linearity(Expected to be linear across the measuring range)Demonstrated to be linear across the range of the assay (1.5 to 2,500 IU/mL) in both serum and plasma samples.
    Limit of Blank (LoB)(Lower than LoD/LoQ)0.0 IU/mL
    Limit of Detection (LoD)(Lower than LoQ)0.4 IU/mL
    Limit of Quantitation (LoQ)≤ 20% within-lab CV at LoQ.1.5 IU/mL (with ≤ 20% within-lab CV)
    Analytical Specificity100% agreement with predicate for cross-reactive disease states; No significant interference (≤ ±1.5 IU/mL for <15 IU/mL, ≤ ±10% for ≥15 IU/mL)- Cross-reactive disease states: 100% total agreement with the predicate assay. - Potential interferents (including 3510 ng/mL biotin): No significant interference (defined as change in concentration within ± 1.5 IU/mL for samples < 15 IU/mL and within ± 10% for samples ≥ 15 IU/mL) observed at clinically relevant concentrations (approx. 4 IU/mL and 100 IU/mL).
    Matrix ComparisonSlope of 1.00 ± 0.12 and R² ≥ 0.92 for correlation between serum, lithium heparin, and EDTA plasma samples.Results met the acceptance criteria of slope of 1.00 ± 0.12 and R² ≥ 0.92.
    Method Comparison(Comparison to a commercially available immunoassay, showing good correlation and agreement, implicitly) Reported statistics: Slope 1.03 (1.00-1.06), Y-Intercept -0.13 (-0.68-0.30), R 0.99 (Passing-Bablok)Slope: 1.03 (95% CI: 1.00-1.06) Y-Intercept: -0.13 (95% CI: -0.68-0.30) Correlation Coefficient R: 0.99 (Passing-Bablok regression)

    2. Sample Size Used for the Test Set and Data Provenance

    • Method Comparison: n = 123 samples. Data provenance not specified (country of origin, retrospective/prospective).
    • Imprecision: Not specified, but involved reagent lots and instrument details suggest laboratory testing.
    • Reproducibility: Not specified, but involved reagent lots and instrument details suggest laboratory testing.
    • High-dose Hook Effect: Not specified how many samples were tested, but concentrations up to 50,000 IU/mL were used.
    • Linearity: Not specified how many samples were tested, but both serum and plasma samples were used.
    • Sensitivity (LoB, LoD, LoQ): Not specified how many samples were used, but involved 2 reagent lots and 2 instruments over a minimum of 3 days (LoB) or 5 days (LoD, LoQ).
    • Analytical Specificity:
      • Cross-reactive disease states: "Samples with potential cross-reactive disease states were tested." Number of samples not specified.
      • Potential interferents: "Patient serum samples containing two levels of thyroglobulin antibody at clinically relevant concentrations of approximately 4 IU/mL and 100 IU/mL." Number of samples not specified, but specific interferents (like biotin at 3510 ng/mL) were tested.
    • Matrix Comparison: Fifty (50) matched sets of serum, lithium heparin plasma, and EDTA plasma samples.

    Data Provenance: The document does not specify the country of origin for any of the samples used in these studies, nor does it explicitly state if the studies were retrospective or prospective, though "patient samples" implies retrospective collection or prospective enrollment for the study.


    3. Number of Experts Used to Establish Ground Truth and Qualifications

    This device is an in vitro diagnostic (IVD) immunoassay, not an imaging device typically requiring expert interpretation for ground truth. Therefore, the concept of "experts" establishing ground truth in the traditional sense of clinical diagnosis (e.g., radiologists, pathologists) does not directly apply here.

    For IVDs, "ground truth" is typically established by:

    • Reference methods/devices.
    • Certified reference materials.
    • Pathology or histopathology (for certain tests).
    • Clinical outcomes/diagnosis (for certain tests).
    • Known concentrations in spiked samples or characterized panels.

    In this document:

    • The "Method Comparison" used a "commercially available immunoassay" as a comparator, which serves as a de facto reference for comparison.
    • Analytical specificity testing used "potential cross-reactive disease states" and "potential interferents" for which the expected results (presence/absence of Thyroglobulin Antibody, or known interference levels) would be pre-established or determined by a reference method.
    • Sensitivity studies (LoB, LoD, LoQ) involve statistical calculations based on repeated measurements of samples with very low or zero analyte concentrations, not expert consensus.
    • Concentration ranges for linearity were established using characterized samples.

    Therefore, no information on human experts establishing ground truth is provided, as it's not relevant for this type of device validation.


    4. Adjudication Method for the Test Set

    Adjudication methods (like 2+1, 3+1) are typically used for studies where multiple human readers interpret medical images or complex data, and their interpretations need to be reconciled to establish a ground truth.

    Since this is an in vitro diagnostic device measuring an analyte concentration, and the gold standard for comparison typically involves laboratory methods or reference materials, an adjudication method for a test set is not applicable or described in this document. The "ground truth" for the various performance characteristics is established by analytical methods and comparisons to reference standards or predicate devices.


    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    No, a multi-reader multi-case (MRMC) comparative effectiveness study was not performed. This type of study is relevant for medical imaging devices where the performance of human readers, with and without AI assistance, is evaluated using a set of cases. This document describes an in vitro diagnostic device, an immunoassay, which does not involve human interpretation of images or complex data in a way that MRMC studies would be applicable.


    6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study

    Yes, the studies described are all standalone performance evaluations of the Access Thyroglobulin Antibody II assay. This means the studies assess the device's ability to measure thyroglobulin antibody levels independent of human intervention in the interpretation of the result in the context of the device's analytical performance. The Access Immunoassay Systems are automated, and the assay's performance metrics (imprecision, linearity, sensitivity, specificity, etc.) are solely based on the analytical capabilities of the device itself.


    7. The Type of Ground Truth Used

    The type of "ground truth" used varies depending on the specific study:

    • Method Comparison: A "commercially available immunoassay" served as the comparative standard. The results from this predicate/reference immunoassay formed the basis for comparing the new device's measurements.
    • Imprecision & Reproducibility: Ground truth is implicit in the known characteristics of the control materials or spiked samples used, and the statistical variability observed from repeated measurements defines the performance.
    • High-dose Hook Effect & Linearity: Ground truth is established by using samples with known or precisely characterized concentrations, often prepared by dilution or spiking.
    • Sensitivity (LoB, LoD, LoQ): Statistical models are used based on repeated measurements of blank samples and samples with very low (but known) analyte concentrations.
    • Analytical Specificity:
      • Cross-reactive disease states: Likely involved samples from patients with these conditions, where the presence/absence of TgAb would have been characterized by a reference method or clinical diagnosis. The "ground truth" then is the expected TgAb status based on the sample's origin.
      • Potential interferents: Involved adding known concentrations of interfering substances to samples with known TgAb concentrations. The "ground truth" is the expected TgAb concentration without interference.
    • Matrix Comparison: Used matched samples from different matrices (serum, plasma), where the inherent TgAb concentration within a given patient sample is the ground truth against which each matrix measurement is compared.

    In summary, the ground truth for this IVD device is primarily established through comparisons to established reference methods/predicate devices, known concentrations in characterized samples (spiked or diluted), and statistical analysis of performance with control materials.


    8. The Sample Size for the Training Set

    The document describes studies for validation of a modified device, demonstrating substantial equivalence to a predicate. It does not mention a "training set" in the context of developing an algorithm or AI model. This is an immunoassay, not a machine learning model that typically requires a large training dataset. The studies focus on analytical performance rather than diagnostic accuracy determined by a machine learning pipeline.


    9. How the Ground Truth for the Training Set Was Established

    As noted above, there is no mention of a "training set" or "ground truth" established for training in this document, as the device is an immunoassay not based on a trainable algorithm.

    Ask a Question

    Ask a specific question about this device

    K Number
    K222610
    Device Name
    Elecsys Anti-Tg
    Manufacturer
    Date Cleared
    2023-09-15

    (382 days)

    Product Code
    Regulation Number
    866.5870
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    510k Summary Text (Full-text Search) :

    Indianapolis, Indiana 46250

    Re: K222610

    Trade/Device Name: Elecsys Anti-Tg Regulation Number: 21 CFR 866.5870
    |
    | Product Codes,Regulation Numbers | JZO, 21 CFR 866.5870

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Immunoassay for the in vitro quantitative determination of antibodies to thyroglobulin in human serum and plasma. The anti-Tg autoantibodies determination is used as an aid in the detection of autoimmune thyroid diseases in conjunction with other laboratory and clinical findings.

    The electrochemiluminescence immunoassay "ECLIA" is intended for use on the cobas e 411 immunoassay analyzer.

    Device Description

    The Elecsys Anti-Tg immunoassay makes use of a competitive test principle using biotinylated human antigen and monoclonal human anti-Tg antibodies labeled with a ruthenium complex. The Elecsys Anti-Tg immunoassay is intended for the quantitative determination of antibodies to thyroglobulin in human serum and plasma. It is intended for use on the cobas e immunoassay analyzers.

    Results are determined via a calibration curve which is instrument-specifically generated by 2 point calibration and a master curve provided via the reagent barcode or e barcode.

    AI/ML Overview

    The provided text describes the performance evaluation of the Elecsys Anti-Tg immunoassay, a diagnostic device, and its acceptance criteria. Here's a breakdown of the requested information based on the text:

    1. A table of acceptance criteria and the reported device performance

    The document does not explicitly present a single table labeled "Acceptance Criteria" with corresponding "Reported Device Performance" in a direct side-by-side format. Instead, it describes various performance evaluations and states whether "All predefined acceptance criteria was met" for each. However, we can reconstruct a table based on the provided details for the non-clinical performance evaluation.

    Performance CharacteristicAcceptance Criteria (Implied)Reported Device Performance
    PrecisionAll predefined acceptance criteria met (specific numerical criteria not detailed in this section for repeatability/intermediate precision and lot-to-lot reproducibility)Repeatability (within-run precision) and Intermediate Precision (within-laboratory precision) measured according to CLSI guideline EP05-A3. Accepted for all samples. Lot-to-lot Reproducibility measured according to CLSI guideline EP05-A3 using three reagent lots. Accepted.
    Analytical Sensitivity
    Limit of Blank (LoB)LoB claim in labeling set to 9 IU/mL (implied acceptance)Determined according to CLSI EP17-A2. LoB claim in labeling will be set to 9 IU/mL.
    Limit of Detection (LoD)LoD claim in labeling set to 10 IU/mL (implied acceptance)Determined according to CLSI EP17-A2. LoD claim in labeling will be set to 10 IU/mL.
    Limit of Quantitation (LoQ)LoQ claim in labeling set to 15 IU/mL (implied acceptance)Determined according to CLSI EP17-A2. LoQ claim in labeling will be set to 15 IU/mL.
    LinearityConfirmed to support the measuring range of 15 - 4000 IU/mL (implied acceptance)Assessed according to CLSI EP06-Ed2, study design B, using weighted linear regression analysis. Linearity confirmed to support measuring range of 15 - 4000 IU/mL.
    Endogenous InterferencesAll predefined acceptance criteria met (specific numerical criteria not detailed)Evaluated for Biotin, Lipemia, Hemoglobin, Bilirubin, Rheumatoid Factor, Tg. All predefined acceptance criteria met. Proposed labeling claims: Biotin ≤ 1200 ng/mL, Lipemia ≤ 2000 mg/dL, Hemoglobin ≤ 600 mg/dL (for conc. ≤ 115 IU/mL), Bilirubin ≤ 66 mg/dL, Rheumatoid Factor ≤ 300 IU/mL, Tg ≤ 100 ng/mL.
    Analytical Specificity/Cross-ReactivityNo cross-reaction with Anti-TPO detected (implied acceptance)Cross-reactivity study conducted with anti-TPO. No cross-reaction with Anti-TPO detected.
    Exogenous InterferencesAll predefined acceptance criteria met for all drugs tested (implied acceptance)Evaluated 17 commonly and 14 specially used pharmaceutical compounds. All predefined acceptance criteria met, and no interference observed.
    Sample Matrix ComparisonResults within specification and support use of specified matrices (implied acceptance)Compared values from serum, K2-EDTA, and K3-EDTA plasma. Results were within specification and support the use of Serum (standard or separating gel), K2-EDTA, and K3-EDTA plasma.
    Method Comparison to PredicateNot explicitly stated as pass/fail; presented as regression resultsCompared to predicate device (K053426) using 129 human serum samples. Linear Regression: y = 0.905x + 48.0, r = 0.990 Passing Bablok: y = 0.974x + 1.72, τ = 0.930
    Reagent Stability
    Reagent On-board Stability42 days (6 weeks) on-board (implied acceptance)Tested on one cobas e 411 analyzer. Stable for up to 42 days (6 weeks).
    Reagent Stability After First Opening42 days (6 weeks) after first opening (implied acceptance)Tested on one cobas e 411 analyzer. Stable for up to 42 days (6 weeks) when stored at 2-8°C.
    Calibration Stability
    Lot Calibration StabilityCalibration every 28 days (1 month) (implied acceptance)Tested on one cobas e 411 analyzer. Recommended calibration frequency of every 28 days (1 month).
    On-board Calibration Stability7 days without new calibration (implied acceptance)Tested on one cobas e 411 analyzer. Stable for up to 7 days without a new calibration.

    2. Sample sizes used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    • Test Set Sample Sizes:
      • Precision: Not explicitly stated, but includes "Human serum 1-5" and "PC THYRO1-2" (presumably replicates for each, as per CLSI EP05-A3 which requires sufficient replicates).
      • Lot-to-lot Reproducibility: "three reagent lots" (number of samples per lot not specified).
      • Analytical Sensitivity (LoB, LoD, LoQ): Not explicitly stated, determined according to CLSI EP17-A2 which has sample size recommendations.
      • Linearity: "Six dilution series" using "native human serum samples and sample pools" (number of samples/pools not specified).
      • Endogenous Interferences: Not explicitly stated per substance, but mentions "Six endogenous substances."
      • Analytical Specificity/Cross-Reactivity: Not explicitly stated (for anti-TPO).
      • Exogenous Interferences: "17 commonly and 14 specially used pharmaceutical compounds" (number of samples not stated).
      • Sample Matrix Comparison: "blood from 13 donors" (tested across serum, K2-EDTA, K3-EDTA plasma, and serum separation tubes from 3 manufacturers).
      • Method Comparison to Predicate: "total of 129 human serum samples."
    • Data Provenance: The document does not specify the country of origin for the data or whether the studies were retrospective or prospective. It is a "510(k) Summary" for an FDA submission, reporting on laboratory performance studies. Given they are "non-clinical performance evaluation," these are typically controlled laboratory studies.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    This device is an in-vitro diagnostic (IVD) immunoassay. The ground truth for such devices is established through analytical testing against reference materials, established methods, and clinical samples with known characteristics, not typically by expert consensus in the same way as an imaging AI. The "ground truth" here is the precise concentration or presence/absence of the analyte (thyroglobulin antibodies) as determined by the study's reference method or expected values for standards/controls.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Not applicable for this type of IVD analytical performance study. Adjudication methods like 2+1 or 3+1 are common in clinical trials or imaging studies where expert readers interpret results, but not for direct quantitative measurements from an immunoassay. The acceptance criteria are based on statistical analysis of quantitative results (e.g., precision, linearity, recovery, regression).

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This is an immunoassay for determining antibody levels, not an imaging device or an AI intended to assist human readers. Hence, no MRMC study was performed, and human reader improvement with AI assistance is not relevant to this device's evaluation.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This entire non-clinical performance evaluation section (4. NON-CLINICAL PERFORMANCE EVALUATION) describes the standalone performance of the Elecsys Anti-Tg immunoassay (a device, not an AI algorithm). The measurements are performed by the "cobas e 411 immunoassay analyzer," which acts as the "algorithm" or automated system. There's no human "in the loop" for the direct measurement results themselves, though human operators are involved in running the tests and interpreting the results in a clinical setting.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    The ground truth for the analytical performance studies (precision, linearity, sensitivity, interferences, stability) is based on:

    • CLSI guidelines: The studies adhere to specific Clinical and Laboratory Standards Institute (CLSI) guidelines (e.g., EP05-A3, EP17-A2, EP06-Ed2), which define how such analytical characteristics are determined using reference materials, spiked samples, and statistical methods.
    • Reference Standards/Materials: Implied in sections like "Traceability/Standardization" against the NIBSC 65/93 Standard, and the use of calibrators (Anti-Tg CalSet) and controls (PreciControl ThyroAB).
    • Known Sample Characteristics: For linearity, samples with varying known concentrations are typically used. For interference studies, samples spiked with known interferents are used.
    • Predicate Device Comparison: For method comparison, the predicate device's results serve as a comparative reference.

    8. The sample size for the training set

    Not applicable in the context of an immunoassay. This device is an in-vitro diagnostic test kit (reagents) used on an existing analyzer. It does not involve a "training set" in the machine learning sense. The "development" or "training" of such a diagnostic involves optimizing the chemical and biological components of the assay (reagents, antibodies, detection method) and calibrating the system across a range of known concentrations.

    9. How the ground truth for the training set was established

    As there is no "training set" in the AI/machine learning sense, this question is not applicable. The development process for an immunoassay involves extensive research and development to create reagents that accurately quantify the target analyte. Calibration is done using reference materials with assigned values, and the assay's performance characteristics (as detailed in section 4) are then rigorously validated.

    Ask a Question

    Ask a specific question about this device

    K Number
    K193313
    Manufacturer
    Date Cleared
    2020-02-27

    (90 days)

    Product Code
    Regulation Number
    866.5870
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    510k Summary Text (Full-text Search) :

    Indianapolis, Indiana 46250

    Re: K193313

    Trade/Device Name: Elecsys Anti-TSHR Regulation Number: 21 CFR 866.5870
    |
    | Product Codes,Regulation Numbers | JZO, 21 CFR 866.5870

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Immunoassay for the in vitro quantitation of autoantibodies to thyroid stimulating hormone (TSH) receptor in human serum using a human thyroid stimulating monoclonal antibody. The anti-TSH receptor determination is used in the assessment of patients suspected of Graves' disease (autoimmune hyperthyroidism).

    The electrochemiluminescence immunoassay "ECLIA" is intended for use on cobas e 601 immunoassay analyzers.

    Device Description

    The Elecsys Anti-TSHR is used for the in vitro quantitative determination of autoantibodies to TSHR receptor in human serum using a human thyroid stimulating monoclonal antibody. It is intended for use on the cobas e 601 immunoassay analyzer. The cobas e family of analyzers uses electrochemiluminescence immunoassay "ECLIA" technology.

    AI/ML Overview

    Here's an analysis of the provided text to extract the acceptance criteria and study information for the Elecsys Anti-TSHR device:


    Acceptance Criteria and Device Performance for Elecsys Anti-TSHR

    The Elecsys Anti-TSHR is an immunoassay for the in vitro quantitative determination of autoantibodies to the TSH receptor in human serum, used in the assessment of patients suspected of Graves' disease.

    1. Table of Acceptance Criteria and Reported Device Performance:

    Performance CharacteristicAcceptance Criteria (Implicit from study results meeting "predetermined acceptance criterion")Reported Device Performance
    PrecisionAll samples to meet predetermined acceptance criteria for repeatability and intermediate imprecision.Repeatability: Sample 1: 0.105 IU/L SD (7.5% CV); Sample 2: 0.140 IU/L SD (7.5% CV); Sample 3: 0.114 IU/L SD (5.7% CV); Sample 4: 0.252 IU/L SD (1.1% CV); Sample 5: 0.298 IU/L SD (0.8% CV); PC ThyroAB 1: 0.145 IU/L SD (3.3% CV); PC ThyroAB 2: 0.342 IU/L SD (1.9% CV). Intermediate Precision: Sample 1: 0.129 IU/L SD (9.1% CV); Sample 2: 0.161 IU/L SD (8.6% CV); Sample 3: 0.144 IU/L SD (7.2% CV); Sample 4: 0.347 IU/L SD (1.5% CV); Sample 5: 0.505 IU/L SD (1.3% CV); PC ThyroAB 1: 0.178 IU/L SD (4.0% CV); PC ThyroAB 2: 0.397 IU/L SD (2.2% CV). Lot-to-Lot Reproducibility: "Calculated SD´s and CV´s for the multiple lot (reproducibility) study are comparable to those of the single lot (intermediate) precision study (met acceptance)."
    Analytical SensitivityEach lot to meet the predetermined acceptance criterion.Limit of Blank (LoB): All lots met acceptance. Claim set to 0.5 IU/L. Limit of Detection (LoD): All lots met acceptance. Claim set to 0.8 IU/L. Limit of Quantitation (LoQ): All lots met acceptance. Claim set to 1.1 IU/L.
    Linearity/Reportable RangeDeviations to be within predetermined acceptance criteria across the entire measuring range.Linearity confirmed in the range from 0.8 to 40.0 IU/L (all deviations within predetermined acceptance criteria).
    High Dose Hook EffectNot applicable.Not applicable (device is not susceptible).
    HAMA InterferenceNot susceptible to interference from HAMA.Not susceptible to interference from Human Anti-Mouse Antibodies (HAMA).
    Endogenous Interference:Recovery for each sample to meet "predetermined acceptance criterion" (implicit).Biotin: Claim set to < 600 ng/mL. Hemolysis: Claim set to ≤ 400 mg/dL. Bilirubin: Claim set to ≤ 25 mg/dL. Lipemia: Claim set to ≤ 1500 mg/dL. Rheumatoid Factors (RF): Claim set to ≤ 600 IU/mL.
    Analytical Specificity/Cross-ReactivityAll cross-reactivities to meet predefined acceptance criterion at specified concentration.No influence with human autoantibodies to thyroglobulin (< 4000 IU/mL) or anti-TPO (< 600 IU/mL) detectable. Cross-reactants tested and met criteria: Human LH (< 10000 mIU/ML), Human FSH (< 10000 mIU/ML), hCG (< 50000 mIU/ML).
    Exogenous Interferences (Drugs)Each compound to be non-interfering at the tested drug concentration.For all 29 drugs tested (17 common, 13 special), the specification was met. Example drugs and concentrations: Amiodarone (≤ 200 mg/L), Carbimazole (≤ 30 mg/L), Levothyroxine (≤ 0.250 mg/L), etc.
    Method Comparison to Predicate"Acceptable" agreement and regression results demonstrating substantial equivalence.Agreement: Positive Percent Agreement (PPA) = 97.37% (95% CI: 93.43 - 98.97); Negative Percent Agreement (NPA) = 95.37% (95% CI: 89.62 - 98.01); Overall Percent Agreement (TPA) = 96.54% (95% CI: 93.55 - 98.17). Regression (Passing/Bablok): y = 1.047x - 0.068 (Slope = 1.029 to 1.064; Intercept = -0.188 to 0.032); r = 0.998.
    StabilityStability data to support Roche Diagnostic's claims.Stability studies reviewed and found acceptable; data supports claims in package inserts.

    2. Sample Size Used for the Test Set and Data Provenance:

    • Precision (Repeatability & Intermediate Precision): 5 human serum samples (4 native, 1 spiked) and 2 controls (PC ThyroAB) were tested over 21 days with 2 replicates per run, 2 runs per day. (Total of 84 replicates per sample/control for repeatability and intermediate precision calculations).
    • Lot-to-Lot Reproducibility: 5 human serum samples (4 native, 1 spiked) were tested with 2 replicates per run, 2 runs per day, for 3 reagent lots (n = 28 determinations per lot per sample, totaling 3x7x2x2 measurements over 3 lots for each sample).
    • Analytical Sensitivity (LoB, LoD, LoQ):
      • LoB: Five blank samples with two replicates each per run, for 6 runs on ≥ three days, across three reagent lots. (Total 60 determinations for analyte-free samples).
      • LoD: Five low analyte samples with two replicates each per run, for 6 runs on ≥ three days, across three reagent lots. (60 replicates per sample per reagent lot).
      • LoQ: Samples tested across three reagent lots for 5 days, one run per day (25 replicates per sample per reagent lot).
    • Linearity/Assay Reportable Range: Three high analyte human serum samples (serum pools) were diluted to create 14 concentrations (13 dilutions). Samples were measured in triplicate within a single run.
    • Endogenous Interference (Biotin, Hemolysis, Bilirubin, Lipemia, Rheumatoid Factors): Not explicitly stated, but typically involves a few serum samples spiked with interferents and compared to unspiked controls. Each sample was spiked with the interfering substance, another aliquot was spiked with isotonic NaCl solution (dilution pool), and the interfering pool was diluted into the dilution pool (in 10% increments for some).
    • Analytical Specificity/Cross-Reactivity: One native human serum sample pool with a low concentration of anti-TSHR was used.
    • Exogenous Interferences (Drugs): Two human serum samples (native serum pools) were used.
    • Method Comparison to Predicate:
      • Agreement: 260 clinical samples from the "intended use population".
      • Regression Analysis: A subset of 120 samples, evenly distributed across the measuring range, from the 260 collected samples.

    Data Provenance:

    • For the "Expected values" (Table 2), an external study used samples from 436 apparently healthy individuals, 210 patients with thyroid diseases without Graves' disease, and 102 patients with untreated Graves' disease.
    • For the non-clinical performance evaluation, samples used were primarily "human serum pools" (native and spiked) and controls.
    • For the Method Comparison, "clinical samples from the intended use population" were used.

    The document does not explicitly state the country of origin or whether the data was retrospective or prospective beyond referring to "clinical samples" and an "external study."

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications:

    This document describes an in vitro diagnostic immunoassay. The concept of "experts establishing ground truth for a test set" as typically understood in AI/imaging studies (e.g., radiologists reviewing images) is not directly applicable here. The "ground truth" for an immunoassay is typically established by reference methods, clinical diagnosis, or by defining specific concentrations for spiked samples or control materials.

    For the "Expected Values" in Table 2, an external study was used where an "optimal cutoff of 1.75 IU/L was determined" based on diagnoses of apparently healthy individuals, those with thyroid diseases without Graves', and those with untreated Graves' disease. This implies a clinical diagnostic ground truth, but not direct "expert adjudication" in the sense of multiple experts assigning labels to individual cases for comparison.

    4. Adjudication Method for the Test Set:

    Not applicable in the context of an immunoassay performance study as described. Clinical diagnoses or reference assays serve as the "ground truth" rather than expert adjudication of individual test cases.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done:

    No, an MRMC comparative effectiveness study was not done. This type of study is relevant for diagnostic imaging interpretation devices involving human readers, not for an automated immunoassay such as the Elecsys Anti-TSHR.

    6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) was Done:

    Yes, the studies described are all standalone performance evaluations of the Elecsys Anti-TSHR device itself. This device is an automated immunoassay system that provides quantitative results without human interpretation in the loop as part of its primary function. Its performance characteristics (precision, sensitivity, linearity, interference, method comparison) are evaluated directly.

    7. The Type of Ground Truth Used:

    • Clinical Diagnosis/Disease State: For "Expected Values" (Sensitivity and Specificity), the ground truth was based on patient cohorts: apparently healthy individuals, patients with thyroid diseases (without Graves'), and patients with untreated Graves' disease.
    • Reference Methods/Known Concentrations: For analytical performance studies (Precision, Analytical Sensitivity, Linearity, Interference, Cross-Reactivity), the ground truth was established by:
      • Using predefined control materials with known values.
      • Using native human serum pools.
      • Spiking samples with known concentrations of analyte or interfering substances.
      • Comparison against a predicate device (Elecsys Anti-TSHR Immunoassay K080092) for method comparison.

    8. The Sample Size for the Training Set:

    The document describes performance evaluation studies, not the development or training of an AI algorithm. Therefore, there is no "training set" in the context of artificial intelligence or machine learning. The studies described are for validation and verification of the device's performance characteristics.

    9. How the Ground Truth for the Training Set Was Established:

    As there is no "training set" for an AI model mentioned in the context of this immunoassay, this question is not applicable. The device's underlying principles are based on electrochemiluminescence immunoassay (ECLIA) technology, not machine learning that requires a training set.

    Ask a Question

    Ask a specific question about this device

    K Number
    K151799
    Manufacturer
    Date Cleared
    2016-03-25

    (267 days)

    Product Code
    Regulation Number
    866.5870
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    510k Summary Text (Full-text Search) :

    EliATM Thyroid Positive Control 250 EliATM Thyroid Positive Control 2500/5000 Regulation Number: 21 CFR §866.5870
    Regulation section:
    21 CFR §866.5870, Thyroid autoantibody immunological test system 21 CFR §862.1660

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    EliA anti-TG Immunoassay is intended for the in vitro quantitative measurement of IgG antibodies directed to thyroglobulin (TG) in human serum and plasma (Li-heparin, EDTA) as an aid in the clinical diagnosis of autoimmune thyroiditis and Graves' disease in conjunction with other laboratory and clinical findings. EliA anti-TG uses the EliA IgG method on the instrument Phadia 250.

    EliA anti-TG Immunoassay is intended for the in vitro quantitative measurement of IgG antibodies directed to thyroglobulin (TG) in human serum and plasma (Li-heparin, EDTA) as an aid in the clinical diagnosis of autoimmune thyroiditis and Graves' disease in conjunction with other laboratory and clinical findings. EliA anti-TG uses the EliA IgG method on the instrument Phadia 2500/5000.

    EliA anti-TPO Immunoassay is intended for the in vitro quantitative measurement of IgG antibodies directed to thyroid peroxidase (TPO) in human serum and plasma (Li-heparin, EDTA) as an aid in the clinical diagnosis of autoimmune thyroiditis and Graves' disease in conjunction with other laboratory and clinical findings. EliA anti-TPO uses the EliA IgG method on the instrument Phadia 250.

    EliA anti-TPO Immunoassay is intended for the in vitro quantitative measurement of IgG antibodies directed to thyroid peroxidase (TPO) in human serum and plasma (Li-heparin, EDTA) as an aid in the clinical diagnosis of autoimmune thyroiditis and Graves' disease in conjunction with other laboratory and clinical findings. EliA anti-TPO uses the EliA IgG method on the instrument Phadia 2500/5000.

    EliA Thyroid Positive Control 250 is intended for laboratory use in monitoring the performance of in vitro measurement of antibodies against thyroid peroxidase (TPO) and thyroglobulin (TG) with Phadia 250 using the EliA IgG method.

    EliA Thyroid Positive Control 2500/5000 is intended for laboratory use in monitoring the performance of in vitro measurement of antibodies against thyroid peroxidase (TPO) and thyroglobulin (TG) with Phadia 2500/5000 using the EliA IgG method.

    Device Description

    The method specific reagents on Phadia® 250 and Phadia® 2500/5000 are identical: they are only filled in different containers. Each device consists of:

    • EliA anti-TG wells are coated with a human thyroglobulin antigen 4 carriers (16 wells each), ready to use; or EliA anti-TPO wells are coated with a human recombinant thyroid peroxidase
    • antigen 4 carriers (16 wells each), ready to use; EliA Sample Diluent: PBS containing BSA, detergent and 0.095% sodium azide - 6 bottles, 48 mL each, ready to use; or 6 bottles, 400 mL each, ready to use;
    • -EliA IgG Conjugate 50 or 200: ß-Galactosidase labeled anti-IgG (mouse monoclonal antibodies) in PBS containing BSA and 0.06% sodium azide - 6 wedge shaped bottles, 5 mL each, ready to use; or 6 wedge shaped bottles, 19 mL each, ready to use
    • -EliA Thyroid Positive Control 250 or 2500/5000: Human serum containing IqG antibodies to TG and TPO in PBS containing BSA, detergent and 0.095% sodium azide - 6 single use vials, 0.3 mL each, ready to use;
    • EliA Negative Control 250 or 2500/5000: Human sera from healthy donors in -PBS containing BSA, detergent and 0.095% sodium azide - 6 single-use vials, 0.3 mL each, ready to use;
    • -EliA IgG Calibrator Strips: Human IqG (0, 4, 10, 20, 100, 600 µq/L) in PBS containing BSA, detergent and 0.095% sodium azide - 5 strips, 6 single-use vials per strip, 0.3 mL each, ready to use;
    • -EliA IgG Curve Control Strips: Human IgG (20 µg/L) in PBS containing BSA, detergent and 0.095% sodium azide - 5 strips, 6 single-use vials per strip, 0.3 mL each, ready to use;
    • -EliA IqG Calibrator Well: Coated with mouse monoclonal antibodies - 4 carriers (12 wells each), ready to use.

    The Phadia EliA Immunodiagnostic System is an automated system for immunodiagnostic testing. The EliA reagents are available as modular packages, each purchased separately. All packages except the positive and negative controls are required to carry out an EliA anti-TG or anti-TPO test.

    AI/ML Overview

    Here's an analysis of the acceptance criteria and the study proving the device meets them, based on the provided text.

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria are not explicitly stated as a formal table with pass/fail values. Instead, the document presents performance characteristics and demonstrates that the device meets these characteristics, implying they are the de facto acceptance criteria. I will infer these criteria from the studies conducted.

    EliA anti-TG Immunoassay

    Acceptance Criterion (Inferred)Reported Device Performance (Phadia 250)Reported Device Performance (Phadia 2500/5000)
    Precision (Total Imprecision CV%)
    Low Concentration (30.6 IU/mL)11.9%17.8%
    Mid-range Concentration (184.8 IU/mL)3.1%5.6%
    High Concentration (4147.2 IU/mL)4.7%10.2%
    Linearity/Reportable Range12 – 4794 IU/mL12 – 4794 IU/mL
    Hook EffectNo hook effect observed up to 27.7 times above upper limitNo hook effect observed up to 27.7 times above upper limit
    Limit of Detection (LoD)12.0 IU/mL (single shared LoD)12.0 IU/mL (single shared LoD)
    Interference (Bilirubin F/C, Hemoglobin, Lipemic factor, Rheumatoid factor, Thyroxine, Iodide)No interference observed up to specified concentrations (ratios of blank/spiked 0.94 – 1.09)No interference observed up to specified concentrations (ratios of blank/spiked 0.94 – 1.09)
    Reference Sera Evaluation (CAP, NEQAS)All targets hit ("OK")All targets hit ("OK")
    Carry-overNegligible effectNone (disposable tips)
    Method Comparison (vs. Predicate VarelisA TG)
    Positive Percent Agreement (equivocal as negative)89.7% (95% CI: 83.3% -94.3%)Not separately reported, assumed similar
    Negative Percent Agreement (equivocal as negative)88.7% (95% CI: 84.6% -92.1%)Not separately reported, assumed similar
    Total Percent Agreement (equivocal as negative)89.0% (95% CI: 85.7% -91.8%)Not separately reported, assumed similar
    Positive Percent Agreement (equivocal as positive)86.3% (95% CI: 80.7% - 90.8%)Not separately reported, assumed similar
    Negative Percent Agreement (equivocal as positive)91.3% (95% CI: 87.0% -94.5%)Not separately reported, assumed similar
    Total Percent Agreement (equivocal as positive)89.0% (95% CI: 85.7% - 91.8%)Not separately reported, assumed similar
    Matrix Comparison (Serum vs. EDTA Plasma)Slope: 1.00 (0.97 to 1.03), R2: 1.00Not separately reported, assumed similar
    Matrix Comparison (Serum vs. Li-heparin Plasma)Slope: 1.00 (0.97 - 1.03), R2: 1.00Not separately reported, assumed similar
    Instrument Comparison (Phadia 250 vs. Phadia 2500/5000)Slope: 0.96 (0.93 - 0.97), Intercept: 1.7 (0.7 - 3.8)Not separately reported, assumed compared to Phadia 250
    Clinical Sensitivity (AI Thyroiditis)55.8% (95% CI: 48.9% - 62.6%)Not separately reported, assumed similar
    Clinical Specificity (AI Thyroiditis)88.9% (95% CI: 85.8% - 91.5%)Not separately reported, assumed similar

    EliA anti-TPO Immunoassay

    Acceptance Criterion (Inferred)Reported Device Performance (Phadia 250)Reported Device Performance (Phadia 2500/5000)
    Precision (Total Imprecision CV%)
    Low Concentration (15.7 IU/mL)8.7%13.1%
    Mid-range Concentration (66.7 IU/mL)4.5%7.0%
    High Concentration (1212.6 IU/mL)6.2%9.5%
    Linearity/Reportable Range4 – 1542 IU/mL4 – 1542 IU/mL
    Hook EffectNo hook effect observed up to 13.4 times above upper limitNo hook effect observed up to 13.4 times above upper limit
    Limit of Detection (LoD)4.0 IU/mL (single shared LoD)4.0 IU/mL (single shared LoD)
    Interference (Bilirubin F/C, Hemoglobin, Lipemic factor, Rheumatoid factor, Thyroxine, Iodide)No interference observed up to specified concentrations (ratios of blank/spiked 0.93 – 1.05)No interference observed up to specified concentrations (ratios of blank/spiked 0.93 – 1.05)
    Reference Sera Evaluation (CAP, NEQAS)All targets hit ("OK")All targets hit ("OK")
    Carry-overNegligible effectNone (disposable tips)
    Method Comparison (vs. Predicate VarelisA TPO)
    Positive Percent Agreement (equivocal as negative)100% (95% CI: 97.8% – 100.0%)Not separately reported, assumed similar
    Negative Percent Agreement (equivocal as negative)89.7% (95% CI: 85.9% -92.8%)Not separately reported, assumed similar
    Total Percent Agreement (equivocal as negative)93.2% (95% CI: 90.6% -95.2%)Not separately reported, assumed similar
    Positive Percent Agreement (equivocal as positive)100% (95% CI: 98.3% – 100.0%)Not separately reported, assumed similar
    Negative Percent Agreement (equivocal as positive)73.7% (95% CI: 68.2% -78.7%)Not separately reported, assumed similar
    Total Percent Agreement (equivocal as positive)84.9% (95% CI: 81.5% -88.8%)Not separately reported, assumed similar
    Matrix Comparison (Serum vs. EDTA Plasma)Slope: 1.00 (0.90 to 1.10), R2: 0.99Not separately reported, assumed similar
    Matrix Comparison (Serum vs. Li-heparin Plasma)Slope: 0.99 (0.95 - 1.03), R2: 0.99Not separately reported, assumed similar
    Instrument Comparison (Phadia 250 vs. Phadia 2500/5000)Slope: 0.98 (0.97 – 0.99), Intercept: 0.2 (-0.4 – 0.8)Not separately reported, assumed compared to Phadia 250
    Clinical Sensitivity (AI Thyroiditis)82.3% (95% CI: 76.6% – 87.2%)Not separately reported, assumed similar
    Clinical Specificity (AI Thyroiditis)90.5% (95% CI: 87.5% – 92.9%)Not separately reported, assumed similar

    2. Sample Size Used for the Test Set and Data Provenance

    • Precision/Reproducibility:
      • EliA anti-TG/TPO on Phadia 250/2500/5000: 8 samples, each with 252 replicate determinations (21 runs x 3 instruments x 7 runs each over 7 days).
      • Data Provenance: Not explicitly stated, implied to be internal lab studies.
    • Linearity/Assay Reportable Range:
      • EliA anti-TG/TPO on Phadia 250/2500/5000: 6-7 patient serum samples.
      • Data Provenance: Not explicitly stated, implied to be internal lab studies.
    • Detection Limit (LoB/LoD):
      • EliA anti-TG/TPO on Phadia 250/2500/5000: Four analyte-free samples and four low antibody concentration blood donor samples. Each measured in 36 replicates (6 replicates x 6 runs). Total of 8 samples x 36 replicates = 288 measurements.
      • Data Provenance: Not explicitly stated, implied to be internal lab studies using blood donors.
    • Endogenous Interference:
      • EliA anti-TG/TPO: Three serum samples (one negative, one near cut-off, one high positive). Each spiked with different interfering substances or blanks and analyzed in triplicates. Runs repeated twice.
      • Data Provenance: Not explicitly stated, implied to be internal lab studies.
    • Reference Sera:
      • EliA anti-TG: 8 samples from CAP.
      • EliA anti-TPO: 8 samples from CAP and 12 samples from UK-NEQAS.
      • Data Provenance: External proficiency testing programs (CAP, UK-NEQAS).
    • Carry-over:
      • Phadia 250: A serum sample (diluted 1:2 and 1:20). Tested with EliA Ro (another assay) to demonstrate instrument's general carry-over performance.
      • Data Provenance: Not explicitly stated, implied to be internal lab studies.
    • Assay Cut-off / Expected Values:
      • EliA anti-TG/TPO: 604 apparently healthy blood donor samples.
      • Data Provenance: Caucasian, African American, Hispanic, and Asian individuals, almost equally distributed by sex and age. (Source country not specified but "blood donor samples" are typically collected prospectively for such studies).
    • Method Comparison and Clinical Sensitivity/Specificity:
      • EliA anti-TG/TPO: 718 serum samples from patients with various thyroid and autoimmune conditions (Graves' disease, autoimmune thyroiditis, non-AI thyroid disease, connective tissue disease, Crohn's disease, ulcerative colitis, primary biliary cirrhosis, HIV, HCV, HBV, other infection, cancer, rheumatoid arthritis, hypergamma-globulinemia, systemic lupus erythematosus, Sjögren's syndrome, celiac disease, type 1 diabetes mellitus, type II diabetes mellitus, pregnant women of all trimesters, pre-eclampsia, miscarriage, thyroid cancer, myasthenia gravis, pernicious anemia, chronic lymphocytic thyroiditis, sub-acute thyroiditis, multi-nodular goiter).
      • Data Provenance: Not explicitly stated (e.g., country of origin, retrospective/prospective).
    • Matrix Comparison:
      • EliA anti-TG/TPO: 57 patients from whom serum, lithium heparin plasma, and EDTA plasma were collected.
      • Data Provenance: Not explicitly stated, implied to be patient samples.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    The document does not specify the use of "experts" in the context of diagnostic performance studies. For immunoassay devices like these, ground truth is typically established by:

    • Reference Methods: Comparison to a legally marketed predicate device (VarelisA TG Antibodies, VarelisA TPO Antibodies) as seen in the method comparison study.
    • Clinical Diagnosis: For sensitivity and specificity, samples are categorized based on "diagnosis" (e.g., Graves' disease, autoimmune thyroiditis). This implies a clinical ground truth, but the details of how these diagnoses were established (e.g., by how many clinicians, their specialties, or experience levels) are not provided.
    • External Reference Sera: For reference sera evaluation, targets are established by external institutions like CAP and UK-NEQAS, which rely on peer groups or provider definitions.

    Therefore, there's no mention of a specific number of experts or their qualifications establishing ground truth in the way one might see for image-based diagnostics.

    4. Adjudication Method for the Test Set

    The document does not describe an explicit adjudication method for establishing ground truth, such as 2+1 or 3+1 expert consensus. Instead, it relies on established clinical diagnoses for sensitivity/specificity studies and predicate device results for method comparison.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    No MRMC comparative effectiveness study was done. This device is an in vitro diagnostic (IVD) immunoassay, not an AI-assisted diagnostic tool for human readers. Its performance is evaluated intrinsically and in comparison to a predicate device, not in terms of human reader improvement.

    6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done

    The device is a standalone immunoassay system (EliA anti-TG Immunoassay and EliA anti-TPO Immunoassay run on Phadia 250/2500/5000 instruments). Its performance is inherently "standalone" in the sense of being an automated laboratory test without direct human interpretive input at the point of result generation. Results are quantitative values that clinicians interpret. Therefore, the entire analytical and clinical performance section describes the device's standalone performance.

    7. The Type of Ground Truth Used

    • Clinical Diagnosis: For clinical sensitivity and specificity studies, samples were linked to known diagnoses such as "Graves' Disease" and "Autoimmune Thyroiditis." The basis for these diagnoses (e.g., pathology, clinical outcomes, or expert opinion) is not detailed beyond being "clinically defined."
    • Predicate Device: For method comparison, the reference standard was the legally marketed predicate device (VarelisA TG Antibodies and VarelisA TPO Antibodies).
    • External Program Targets: For reference sera, the ground truth was the targets set by CAP (College of American Pathologists) and UK-NEQAS (United Kingdom National External Quality Assessment Service). These targets are often established by peer consensus or by the program providers.

    8. The Sample Size for the Training Set

    The document describes studies for validation of performance characteristics but does not explicitly mention a "training set" in the context of machine learning or AI development. Since these are immunoassays, they are based on biochemical reactions and calibration rather than a machine learning model that requires training data. The calibration process uses specific calibrator materials.

    9. How the Ground Truth for the Training Set Was Established

    As this is an immunoassay and not an AI/ML device, the concept of a "training set" with established ground truth doesn't apply in the same way. The device is calibrated using:

    • EliA IgG Calibrator Strips: Human IgG at defined concentrations (0, 4, 10, 20, 100, 600 µq/L). These calibrators are traceable to the International Reference Preparation (IRP) 67/86 of Human Serum Immunoglobulins A, G and M from WHO. New batches are compared to a secondary standard or the IRP directly to ensure correct concentration. This traceability constitutes the "ground truth" for calibration.
    • EliA IgG Curve Control Strips: Human IgG (20 µg/L) for monitoring calibration curve performance.

    The "ground truth" for the calibrators is therefore established through established international reference standards and their traceability.

    Ask a Question

    Ask a specific question about this device

    K Number
    K152061
    Date Cleared
    2016-03-03

    (223 days)

    Product Code
    Regulation Number
    866.5870
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    510k Summary Text (Full-text Search) :

    Immulite 200 TSI Assay, Immulite 200 TSI Calibration Verification Material

    Regulation Number: 21 CFR 866.5870
    --------------|---------------------------------------------|----------------------------|
    | 21 CFR 866.5870

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The IMMULITE® 2000 TSI (thyroid-stimulating immunoglobulins) Assay is an in vitro diagnostic immunoassay for the semi-quantitative determination of thyroid stimulating autoantibodies specific to thyroid stimulating hormone receptors (TSHR) in human serum (including Serum Separator tubes) or plasma (K2-EDTA or lithium heparin). The IMMULITE® 2000 TSI Assay is for use on the IMMULITE® 2000 system. The measurement of thyroid stimulating autoantibodies, in conjunction with other clinical and laboratory findings, is used as an aid in the diagnosis of patients suspected of having Graves' disease.

    The IMMULITE® TSI Calibration Verification Material (CVM) is for in vitro diagnostic use in the verification of calibration of the IMMULITE® TSI Assay on the IMMULITE® 2000 Systems.

    Device Description

    The IMMULITE 2000 TSI assay kit consists of the following components:

    • TSI bead pack coated with MAb (3D7) anti-TSHR anchor antibody and hTSHR Capture Chimera
    • TSI reagent wedge containing hTSHR-Chimera alkaline phosphatase conjugate
    • TSI adjustors: low and high, containing TSI negative heat-inactivated bovine serum and thyroid stimulating human MAb (M22)
    • TSI controls: negative, low, and high, containing TSI negative human serum and thyroid stimulating MAb (M22)
    • Multi-Diluent 2
    AI/ML Overview

    Here's a breakdown of the acceptance criteria and the study that proves the device meets them, based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria are generally implied from the "Drift Specifications" for stability and the statistical measures for precision and method comparison. The device performance is the "reported performance".

    Performance CharacteristicAcceptance Criteria (Implied/Stated)Reported Device Performance
    Precision/ReproducibilityNot explicitly stated numerical acceptance criteria for %CV, but generally clinical assays aim for low %CVs.20-Day Imprecision:- Repeatability %CV: Ranged from 3.5% to 7.0%.- Within Lab %CV: Ranged from 5.0% to 8.3%.
    Linearity/Reportable RangeLinearity data for % Difference to be ± 15% or 0.50 IU/L (whichever is greater).Linearity data for % Difference was shown to be ± 15% or 0.50 IU/L (whichever is greater) for most samples; tested within range 0.50 - 40.0 IU/L.
    Assay/Component Stability DriftReagents and Beads: - Control 1 at ≤ -60°C: ≤ 15% from Day 0 mean - Control 2 at ≤ -60°C: ≤ 10% from Day 0 mean - CVM 1 at ≤ -60°C: ≤ 0.15 IU/L from Day 0 mean - MDP 1: ≤ 20% from Day 0 mean - MDP 4: ≤ 10% from Day 0 mean - Cal J: ≤ 10% from Day 0 mean (Also, all Control results must be within QC established range to validate the run).Stability Claims Achieved: - Kit, unopened: 12 Months (2-8 °C) - Bead Pack, open: 90 Days (2-8 °C) - Reagent wedge, open and on-board: 90 Days (2-8 °C) - Sample diluent, open: 30 Days (2-8 °C) - Sample diluent, open frozen aliquotted: 6 Months (-20 °C) - Adjustors open: 90 Days (2-8 °C) - Adjustors frozen aliquotted: 4 Months (-20 °C) - Controls open: 90 Days (2-8 °C) - Controls frozen aliquotted: 6 Months (-20 °C) - CVM, unopened: 12 Months (2-8 °C) - CVM, opened and reconstituted: 30 Days (2-8 °C) Sample Stability: 24 hours at 20-25°C, 7 days at 2-8°C, and 12 months at -20°C for serum and plasma.
    Limit of Blank (LoB)LoB should be low enough for clinical utility.Highest LoB by lot was determined to be 0.03 IU/L.
    Limit of Detection (LoD)LoD should be low enough for clinical utility.Highest LoD by lot was determined to be 0.06 IU/L.
    Analytical Specificity (Interference)No interference (≤ 10% different than control sample).Interferents (Intralipid, Hemoglobin, Bilirubin, K2-EDTA): No interference (≤ 10% difference) except for hemoglobin (≥ 200mg/dL potentially affects recovery) and short draw K2-EDTA (may result in under-recovery). HAMA: Individual sample bias of <4% (at 10ng/mL and 40ng/mL HAMA). Rheumatoid Factor (RF): Individual sample bias of <5% (at 100 IU/mL and 200 IU/mL RF). Cross-Reactivity (TSH, FSH, LH, hCG, aTPO, and aTG): Bias of ≤10% for all interferents and samples.
    Method Comparison (vs. Thyretain TSI Reporter BioAssay)High agreement (generally >90% for positive, negative, and overall).Positive Agreement: 95.8% (95% CI: 93.0 – 97.7) Negative Agreement: 87.7% (95% CI: 84.5 - 90.5) Overall Agreement: 91.0% (95% CI: 88.8 - 92.9)
    Matrix Comparison (Regression Coefficient)Strong correlation (e.g., R-value > 0.95 or 0.98).SST Serum: Slope 1.01, Intercept 0.00, Correlation Coefficient 0.99 K2-EDTA Plasma: Slope 1.03, Intercept -0.01, Correlation Coefficient 0.99 Lithium Heparin Plasma: Slope 0.99, Intercept -0.01, Correlation Coefficient 0.99
    Clinical Sensitivity and Specificity (at 0.55 IU/L cut-off)High clinical sensitivity and specificity (generally >95%).Clinical Sensitivity: 98.6% (95% CI: 96.8 – 99.5) Clinical Specificity: 98.5% (95% CI: 96.8 – 99.5)
    Reference Range (97.5th percentile)Clinically appropriate range.< 0.10 IU/L (using non-parametric 97.5th percentile).

    *Specific Answers to Questions:

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    • Precision/Reproducibility: 6 serum samples and low/high kit controls, tested in duplicate in two runs per day over 20 days (total 80 replicates per sample). Data provenance not specified (likely internal laboratory data, prospective for the study).
    • Linearity: TSI positive and TSI negative samples from "several commercial vendors." Data provenance not specified.
    • LoB: 5 TSI negative samples, tested with 3 reagent lots, 2 runs, 2 replicates per run for 3 days (total >180 measurements). Data provenance not specified.
    • LoD: 5 TSI serum samples, tested in 4 replicates using 3 reagent kit lots for 3 days on 3 systems (total 36 observations per sample). Data provenance not specified.
    • Analytical Specificity (Interference): Three serum pools (for endogenous interferents), HAMA positive serum, RF positive serum, and six potential biological interferent samples. Data provenance not specified.
    • Method Comparison: 811 serum samples from patients with Graves' disease, other thyroid or autoantibody diseases. Specimens tested at two external sites and one internal site for the IMMULITE 2000 TSI assay, and at one external site for the Thyretain device. Data provenance not explicitly stated, but implies mixed sources (internal/external, likely retrospective patient samples).
    • Matrix Comparison: Graves' Disease sets of matched serum and plasma samples (clot tube, lithium heparin, SST, K2-EDTA). Sample size not specified. Data provenance not specified.
    • Clinical Sensitivity and Specificity: 361 treated and untreated hyperthyroid Graves' disease patients and 404 individuals with other thyroid or autoimmune diseases. Data provenance not specified (likely retrospective clinical samples).
    • Reference Range: 842 serum samples from apparently healthy males (n=151), non-pregnant females (n=155), first trimester (n=169), second trimester (n=191), and third trimester (n=176) pregnant donors. Data provenance not specified.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    The document does not specify the number or qualifications of experts used to establish the ground truth for any of the test sets. For example, for clinical sensitivity and specificity, it refers to "Graves' Disease Diagnosis" and "other thyroid or autoimmune diseases" but does not detail how these diagnoses were definitively established or by whom. Similarly, for the method comparison, it refers to patient samples for Graves' disease or other thyroid/autoantibody diseases without detailing expert adjudication of these conditions.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    The document does not describe any specific adjudication method for establishing ground truth diagnoses for the patient samples used in the studies. The diagnoses are simply stated as existing, implying they were pre-determined or established through typical clinical practice, rather than through a separate expert adjudication process for the study.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    No, a multi-reader multi-case (MRMC) comparative effectiveness study was not done. This is an immunoassay device, not an imaging AI device that involves human readers interpreting cases. The comparison study was a "Method Comparison" between the new Immulite 2000 TSI assay and a predicate immunoassay device (Thyretain TSI Reporter BioAssay), analyzing patient samples.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Yes, the studies presented are all standalone validations of the Immulite 2000 TSI assay system. It is an automated immunoassay for in vitro diagnostic use, meaning it provides results directly from patient samples without human interpretation of images or other complex data requiring "human-in-the-loop" decision-making. The performance characteristics (precision, linearity, detection limits, clinical sensitivity/specificity, etc.) reflect the algorithm/device's performance alone.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    The ground truth for the clinical studies (clinical sensitivity/specificity, method comparison) appears to be clinical diagnosis of Graves' disease or other thyroid/autoimmune diseases. For analytical performance studies (precision, linearity, LoB, LoD, interference), the ground truth is based on spiking known concentrations, using negative samples, or established reference standards like the NIBSC standard.

    8. The sample size for the training set

    The document does not explicitly mention a "training set" in the context of machine learning. This is a traditional immunoassay, not an AI device that typically involves distinct training and test sets in the same manner. The studies describe validation sets, not training sets for model development.

    9. How the ground truth for the training set was established

    As there is no distinct "training set" in the context of an AI/machine learning model as understood in typical AI/ML submissions, this question is not directly applicable. For traditional assays, calibrators are used to establish the measurement and are themselves calibrated against reference materials (e.g., NIBSC standard 08/204 for the Immulite 2000 TSI assay).

    Ask a Question

    Ask a specific question about this device

    K Number
    K112933
    Date Cleared
    2011-12-27

    (85 days)

    Product Code
    Regulation Number
    866.5870
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    510k Summary Text (Full-text Search) :

    Re: K112933

    Trade/Device Name: Access Thyroglobulin Antibody II Assay Regulation Number: 21 CFR §866.5870

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Access Thyroglobulin Antibody II assay is a paramagnetic chemiluminescent immunoassay for the quantitative determination of thyroglobulin antibody levels in human serum and plasma using the Access Immunoassay Systems. The measurement of thyroid autoantibodies may aid in the diagnosis of Hashimoto's disease, nontoxic goiter, and Graves' disease.

    Device Description

    The Access Thyroglobulin Antibody II reagents, Thyroglobulin Antibody II calibrators, and the Access Immunoassay analyzers comprise the Access lmmunoassay System for the quantitative determination of thyroglobulin antibody in human serum and plasma.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and study information for the Access Thyroglobulin Antibody II for use on the Access Immunoassay Systems, as extracted from the provided 510(k) summary:

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria are implied by the "Summary of Studies" section, which describes the performance characteristics that were measured and deemed acceptable for demonstrating substantial equivalence. Exact numerical acceptance thresholds for each test (e.g., "imprecision must be less than X%") are not explicitly stated as distinct criteria, but the reported performance values are presented as evidence of meeting acceptable levels.

    Performance MetricAcceptance Criteria (Implied)Reported Device Performance
    Method ComparisonStrong correlation with predicate deviceNew = m(predicate)1.0 ± 0.12, R² ≥ 0.92 (for 397 samples, range <0.9 to 2500 IU/mL)
    ImprecisionLow imprecision across various concentrationsWithin-run: 3.6 to 5.7 %CVBetween-run: 0.0 to 5.2 %CVTotal: 4.8 to 7.7 %CV (at 26.9 to 1889.6 IU/mL).Overall: < 10% CV for ≥ 15 IU/mL; < 1.5 IU/mL SD for < 15 IU/mL.
    High-dose Hook EffectNo hook effect at high concentrationsNo hook to 50,000 IU/mL
    LinearityLinear across the assay's measuring rangeLinear across the range of the assay (0.0 to 2500 IU/mL)
    Limit of Blank (LoB)Lowest measurement with no analyte0.9 IU/mL (n=221)
    Limit of Detection (LoD)Lowest detectable concentration (95% probability)0.9 IU/mL
    Analytical SpecificityNo significant interference; 100% agreement for autoimmune samplesNo significant interference from total protein, bilirubin, hemoglobin, or triglycerides. 100% total agreement with autoimmune disease state samples.

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size: 397 samples were used for the method comparison study. The LoB study used 221 samples. The number of samples for other studies (imprecision, linearity, analytical specificity) is not explicitly stated, although "samples with autoimmune disease state" were "tested" for analytical specificity.
    • Data Provenance: Not explicitly stated (e.g., country of origin). The data is from a prospective study, as it involves running samples on both the new device and the predicate device for comparison and performance characterization.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

    This device is an in vitro diagnostic (IVD) immunoassay that measures an analyte (thyroglobulin antibody) directly. Therefore, there is no "ground truth" established by human experts in the way that would apply to image-based diagnostic AI. The "ground truth" for method comparison and performance comes inherently from the measurements of the predicate device and established analytical methods.

    4. Adjudication Method for the Test Set

    Not applicable, as the device is an immunoassay and does not involve human interpretation or adjudication for its raw output. The "comparison" is between the new device's quantitative output and the predicate device's quantitative output.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and Effect Size of Human Improvement with AI vs. Without AI Assistance

    Not applicable. This device is an automated immunoassay system, not an AI-assisted diagnostic tool that aids human readers in interpretation.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    Yes, the studies described are standalone performance evaluations of the device (immunoassay system), as it functions as an automated system producing quantitative results. There is no "human-in-the-loop" component for the actual measurement process.

    7. The Type of Ground Truth Used

    The "ground truth" used for comparison, particularly in the method comparison study, is the quantitative measurement provided by the predicate device (Access Thyroglobulin Antibody II, K062516). For other performance metrics (imprecision, linearity, LoB, LoD), the "ground truth" is derived from established analytical methods and reference materials (e.g., NIBSC Anti-Thyroglobulin Serum, Human First International Reference Preparation, WHO Coded 65/93 for standardization).

    8. The Sample Size for the Training Set

    Not applicable. This device is an immunoassay, not a machine learning or AI algorithm that requires a "training set" in the conventional sense. The "training" for such devices typically involves calibration using specific calibrators as described ("Utilizes a stored calibration curve").

    9. How the Ground Truth for the Training Set Was Established

    Not applicable. As described above, there isn't a "training set" in the AI/ML context. The calibration of the immunoassay system uses specific calibrators, and its standardization is based on NIBSC Anti-Thyroglobulin Serum, Human First International Reference Preparation, WHO Coded 65/93.

    Ask a Question

    Ask a specific question about this device

    K Number
    K092229
    Date Cleared
    2010-05-18

    (299 days)

    Product Code
    Regulation Number
    866.5870
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    510k Summary Text (Full-text Search) :

    Reporter Classification name: System, Test, Thyroid Autoantibody Product Code: JZO Regulation: 21 CFR § 866.5870
    ------------------|------------------------------------------------|
    | Regulation Number: | 21 CFR §866.5870

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Thyretain ™ TSI Reporter BioAssay is intended for the qualitative detection in serum of thyroid stimulating autoantibodies to the thyroid stimulating hormone (TSH) receptors (TSHRs) on the thyroid. The detection of these stimulating autoantibodies, in conjunction with other clinical and laboratory findings, may be useful as an aid in the differential diagnosis of patients with Graves' disease (GD).

    Device Description

    The number of wells tested per Positive, Reference and Negative control has been reduced from three to two for each. The number of wells tested per patient specimen has been reduced from three to two.

    AI/ML Overview

    The provided submission (K092229) is a special 510(k) for a device modification; therefore, it primarily focuses on demonstrating that the device, post-modification, remains substantially equivalent to its predicate device. This type of submission usually doesn't include new, large-scale clinical studies with specific acceptance criteria as would be found in a De Novo or PMAs. Instead, it relies on bridging data to show the modification doesn't negatively impact performance.

    Here's an analysis based on the provided text:

    Device: Thyretain™ TSI Reporter BioAssay (Modified)

    Predicate Device: Thyretain™ TSI Reporter BioAssay (K083391)

    Modification: Reduction of the number of wells tested per Positive, Reference, and Negative control from three to two. Reduction of wells tested per patient specimen from three to two.


    1. A table of acceptance criteria and the reported device performance

    The document does not explicitly state formal acceptance criteria in a quantitative manner (e.g., "sensitivity must be >90%"). Instead, the performance assessment aims to demonstrate that the modified device remains equivalent to the predicate, particularly in its non-clinical performance.

    Acceptance Criteria (Implied)Reported Device Performance
    No significant negative impact on device performance due to modification."Based on the analysis of the study site data (see Attachment 2) the modification poses little risk."
    Maintain the intended use and diagnostic accuracy.Intent to maintain the same intended use as the predicate device. The FMEA indicated little risk from the modification.

    Note: "Attachment 2" which would contain the detailed study site data is not provided in the input, so specific quantitative performance metrics like sensitivity, specificity, or agreement with the predicate are not available in this summary. The assessment focuses on risk analysis.


    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    The document refers to "study site data" in "Attachment 2". However, the details of this data, including sample size, provenance (country of origin), and whether it was retrospective or prospective, are not provided in the given text extract.


    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    This information is not provided in the given text extract. Given this is a modification to an in vitro diagnostic (IVD) device, the ground truth would likely refer to established diagnostic methods or clinical outcomes, rather than expert interpretation of images.


    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    This information is not provided in the given text extract. It is unlikely to be relevant for this type of IVD modification, which assesses analytical performance rather than interpretation by human readers.


    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    No. A MRMC comparative effectiveness study was not done. This is an in vitro diagnostic device (a bioassay) for detecting autoantibodies, not an imaging device that requires human interpretation or AI assistance.


    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This question is not applicable in the context of this device. The Thyretain™ TSI Reporter BioAssay is a laboratory assay for detecting antibodies, not an algorithm. The "performance" refers to the analytical performance of the assay itself. The submission is not about an algorithm, but a modification to the assay's procedure (number of wells).


    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    The submission is for an immunoassay. The ground truth for such an assay would typically be established by:

    • Reference methods: Comparison to established, clinically validated assays for the same analyte.
    • Clinical diagnosis: Correlation with the clinical diagnosis of Graves' disease (GD), which would involve a combination of clinical symptoms, other laboratory findings, and potentially outcomes data over time.
    • Pathology: Less likely to be the primary ground truth for an antibody assay, but could be part of the broader clinical picture for Graves' disease.

    The provided text does not explicitly state the type of ground truth used for the "study site data" referenced.


    8. The sample size for the training set

    This information is not provided in the given text extract. For an IVD assay modification, there isn't typically a "training set" in the machine learning sense. Any data used to validate the modified procedure would be considered a "test set" or verification/validation data.


    9. How the ground truth for the training set was established

    As there is no "training set" in the context of this IVD device, this question is not applicable. Any ground truth for validation samples would be established by methods similar to those described in point 7.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 5