K Number
K203703

Validate with FDA (Live)

Device Name
I-Q View
Manufacturer
Date Cleared
2021-03-24

(96 days)

Product Code
Regulation Number
892.1680
Age Range
All
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

This software is intended to generate digital radiographic images of the skull, spinal column, extremities, and other body parts in patients of all ages. Applications can be performed with the patient sitting, or lying in the prone or supine position and is intended for use in all routine radiography exams. The product is not intended for mammographic applications.

This software is not meant for mammography, fluoroscopy, or angiography.

Device Description

The I-Q View is a software package to be used with FDA cleared solid-state imaging receptors. It functions as a diagnostic x-ray image acquisition platform and allows these images to be transferred to hard copy, softcopy, and archive devices via DICOM protocol. The flat panel detector is not part of this submission. In the I-Q View software, the Digital Radiography Operator Console (DROC) software allows the following functions:

    1. Add new patients to the system; enter information about the patient and physician that will be associated with the digital radiographic images.
    1. Edit existing patient information.
    1. Emergency registration and edit Emergency settings.
    1. Pick from a selection of procedures, which defines the series of images to be acquired.
    1. Adiust technique settings before capturing the x-ray image.
    1. Preview the image, accept or reject the image entering comments or rejection reasons to the image. Accepted images will be sent to the selected output destinations.
    1. Save an incomplete procedure, for which the rest of the exposures will be made at a later time.
    1. Close a procedure when all images have been captured.
    1. Review History images, resend and reprint images.
    1. Re-exam a completed patient.
    1. Protect patient records from being deleted by the system.
    1. Delete an examined Study with all images being captured.
    1. Edit User accounts.
    1. Check statistical information.
    1. Image QC.
    1. Image stitching.
    1. Provides electronic transfer of medical image data between medical devices.
AI/ML Overview

The provided document is a 510(k) summary for the I-Q View software. It focuses on demonstrating substantial equivalence to a predicate device through bench testing and comparison of technical characteristics. It explicitly states that clinical testing was not required or performed.

Therefore, I cannot provide details on clinical acceptance criteria or a study proving the device meets them, as such a study was not conducted for this submission. The document relies on bench testing and comparison to a predicate device to establish substantial equivalence.

Here's a breakdown of what can be extracted from the provided text regarding acceptance criteria and the "study" (bench testing) that supports the device:

1. Table of Acceptance Criteria and Reported Device Performance

Since no clinical acceptance criteria or performance metrics are provided, this table will reflect the general statements made about the device performing to specifications.

Acceptance Criteria (Implied)Reported Device Performance
Device functions as intended for image acquisition.Demonstrated intended functions.
Device performs to specification.Performed to specification.
Integration with compatible solid-state detectors performs within specification.Verified integration performance within specification.
Software is as safe and functionally effective as the predicate.Bench testing confirmed as safe and functionally effective as predicate.

2. Sample size used for the test set and the data provenance

  • Test Set Sample Size: Not applicable/not reported. The document describes bench testing, not a test set of patient data.
  • Data Provenance: Not applicable. Bench testing generally involves internal testing environments rather than patient data from specific countries or retrospective/prospective studies.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

  • Not applicable. As no clinical test set was used, no experts were needed to establish ground truth for patient data. Bench testing typically relies on engineering specifications and verification.

4. Adjudication method for the test set

  • Not applicable. No clinical test set or human interpretation was involved.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

  • No, an MRMC comparative effectiveness study was not done. The document explicitly states: "Clinical Testing: The bench testing is significant enough to demonstrate that the I-Q View software is as good as the predicate software. All features and functionality have been tested and all specifications have been met. Therefore, it is our conclusion that clinical testing is not required to show substantial equivalence." The device is software for image acquisition, not an AI-assisted diagnostic tool.

6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

  • Yes, in a sense. The "study" described is bench testing of the software's functionality and its integration with solid-state detectors. This is an evaluation of the algorithm/software itself.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

  • For bench testing, the "ground truth" would be the engineering specifications and expected functional behavior of the software and its interaction with hardware components. It's about verifying that the software performs according to its design requirements.

8. The sample size for the training set

  • Not applicable. The I-Q View is described as an image acquisition and processing software, not an AI/machine learning model that typically requires a training set of data.

9. How the ground truth for the training set was established

  • Not applicable, as there is no mention of a training set or AI/machine learning component.

Summary of the "Study" (Bench Testing) for K203703:

The "study" conducted for the I-Q View software was bench testing. This involved:

  • Verification and validation of the software.
  • Demonstrating the intended functions and relative performance of the software.
  • Integration testing to verify that compatible solid-state detectors performed within specification as intended when used with the I-Q View software.

The conclusion drawn from this bench testing was that the software performs to specification and is "as safe and as functionally effective as the predicate software." This was deemed sufficient to demonstrate substantial equivalence, and clinical testing was explicitly stated as not required.

{0}------------------------------------------------

March 24, 2021

Image /page/0/Picture/1 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the seal of the Department of Health & Human Services. To the right of the seal is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, and the word "ADMINISTRATION" in a smaller font below.

First Source, Inc. % Ms. Susan Ryan-Kron Quality & Regulatory Manager 3495 Winton Place, Bldg. E, Ste 1 ROCHESTER NY 14623

Re: K203703

Trade/Device Name: I-O View Regulation Number: 21 CFR 892.1680 Regulation Name: Stationary x-ray system Regulatory Class: Class II Product Code: MQB, IZL, LLZ Dated: February 5, 2021 Received: February 8, 2021

Dear Ms. Ryan-Kron:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmp/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see

{1}------------------------------------------------

https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

For

Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K203703

Device Name I-Q View

Indications for Use (Describe)

This software is intended to generate digital radiographic images of the skull, spinal column, extremities, and other body parts in patients of all ages. Applications can be performed with the patient sitting, or lying in the prone or supine position and is intended for use in all routine radiography exams. The product is not intended for mammographic applications.

This software is not meant for mammography, fluoroscopy, or angiography.

Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510 (k) Summary I-Q View (K203703)

Company Name:First Source, Inc.
Address:3495 Winton PlaceBldg E, Ste 1Rochester, NY 14623
Telephone No:585.272.1690
Registration No.:3004063527
Contact person:Susan Ryan-Kron
Date Prepared:February 5, 2021
Trade/Device Name:I-Q View
Classification Name:Stationary X-ray System
DR Regulation No.:21 CFR 892.1680
Regulatory Name:Stationary x-ray system
Classification Panel:Radiology
Regulation Class:Class II
Product Code:MQB, IZL, LLZ

Predicate:

K number: K193644 Company Name: LiverMoreTech, Inc. Trade/Device Name: E-COM DR-2000 DR DR Regulation Number: 21 CFR 892.1680 Requlation Name: Stationary x-ray system Regulatory Class: Class II Product Code: MQB

Device description:

The I-Q View is a software package to be used with FDA cleared solid-state imaging receptors. It functions as a diagnostic x-ray image acquisition platform and allows these images to be transferred to hard copy, softcopy, and archive devices via DICOM protocol. The flat panel detector is not part of this submission. In the I-Q View software, the Digital Radiography Operator Console (DROC) software allows the following functions:

    1. Add new patients to the system; enter information about the patient and physician that will be associated with the digital radiographic images.
    1. Edit existing patient information.
    1. Emergency registration and edit Emergency settings.

{4}------------------------------------------------

    1. Pick from a selection of procedures, which defines the series of images to be acquired.
    1. Adiust technique settings before capturing the x-ray image.
    1. Preview the image, accept or reject the image entering comments or rejection reasons to the image. Accepted images will be sent to the selected output destinations.
    1. Save an incomplete procedure, for which the rest of the exposures will be made at a later time.
    1. Close a procedure when all images have been captured.
    1. Review History images, resend and reprint images.
    1. Re-exam a completed patient.
    1. Protect patient records from being deleted by the system.
    1. Delete an examined Study with all images being captured.
    1. Edit User accounts.
    1. Check statistical information.
    1. Image QC.
    1. Image stitching.
    1. Provides electronic transfer of medical image data between medical devices.

Indications for Use (intended use):

This software is intended to generate digital radiographic images of the skull, spinal column, chest, abdomen, extremities, and other body parts in patients of all ages. Applications can be performed with the patient sitting, standing, or lying in the prone or supine position and is intended for use in all routine radiography exams. The product is not intended for mammographic applications.

This software is not meant for mammography, fluoroscopy, or angiography.

Technological Characteristics:

The I-Q View has the same technological characteristics (e.g. design) as the predicate software. Specifications are essentially the same or similar and there is no impact to the safety nor effectiveness of the diagnostic image quality as a result of this I-Q View software.

Comparison Table:

A comparison chart provides the similarities and differences of technical features of the predicate software and the software for this submission.

CharacteristicsPredicate: LiverMoreTechE-COM DR-2000 (K193644)I-Q ViewThis submission
Indications forUseIntended for digital image capture usein general radiographic examinations,wherever convention screen-filmsystems may be used, excludingThis software is intended to generatedigital radiographic images of theskull, spinal column, chest, abdomen,extremities, and other body parts in
fluoroscopy, angiography andmammography. The kit allows imagingof the skull, chest, shoulders, spine,abdomen, pelvis, and extremities.patients of all ages. Applications canbe performed with the patient sitting,standing, or lying in the prone orsupine position and is intended foruse in all routine radiography exams.This software is not meant formammography, fluoroscopy, orangiography.
PlatformWindows 10Windows 10
Compatiblesolid-statereceptorsToshiba FDX4343R Detector(K130883)Toshiba FDX3543RP Detector(K130883)Toshiba FDX3543RPW Detector(K171353)Toshiba FDX2530RPW Detector(K162687)Toshiba FDX4343RPW Detector(K162687)Thales Pixium RAD 4143 Detector(K133139)Thales Pixium RAD 4343 Detector(K133139)Thales Pixium RAD 3543 Detector(K141440)Thales Pixium RAD 3543EZ Detector(K141440)Thales Pixium RAD 2430EZ Detector(K141440)Thales Pixium RAD 3543DR Detector(K191813)Varian PaxScan 4336R Detector(K172007)Varian PaxScan 4343R Detector(K172007)Varian PaxScan 4336X Detector(K172007)Varian PaxScan 4336W Detector(K172007)Toshiba FDX4343R Detector(K130883)Toshiba FDX3543RP Detector(K130883)Toshiba FDX3543RPW Detector(K171353)Toshiba FDX2530RPW Detector(K162687)Toshiba FDX4343RPW Detector(K162687)Thales Pixium RAD 4143 Detector(K133139)Thales Pixium RAD 4343 Detector(K133139)Thales Pixium RAD 3543 Detector(K141440)Thales Pixium RAD 3543EZ Detector(K141440)Thales Pixium RAD 2430EZ Detector(K141440)Thales Pixium RAD 3543DR Detector(K191813)Varian PaxScan 4336R Detector(K172007)Varian PaxScan 4343R Detector(K172007)Varian PaxScan 4336X Detector(K172007)Varian PaxScan 4336W Detector(K172007)Careray - 1500Cw (K150929)DRTech - ExprimerEVS2430(K171137),EVS3643 (K162552),EVS4343W (K193017),EVS4343A (K192400)iRay - Mars 1417V (K161730),1717V (K201043)Konica Minolta - AeroDR 1210HQ(K130936)

{5}------------------------------------------------

{6}------------------------------------------------

DICOM 3.0compatibilityYes1417HQ/HD (K151465),1717HQ (K172793)LG – DXD 1417 (K182348),17HK700G (K180332)Pixxgen - PIXX1212 (K182533),PIXX1417 (K182533),PIXX1717 (K182533)Thales - Pixium RAD 4143RC/RG(K131483),4343RC/RG (K131483),Pixium RAD 4343RC-E/RG-E(K170113),Pixium RAD 3543DR-X (K170113)Varex - XRpad 4336 (K140551),4343F (K142698),XRpad2 3025 (K161942),4336 (K161966)
Yes

Bench Testing:

Bench Testing was conducted via verification and validation. Testing was done to demonstrate the intended functions and the relative performance of the I-Q View software. Integration testing was done to verify that the compatible solid-state detectors listed in the Comparison Table above, performed within specification and as intended. Non-clinical test results have confirmed that the software/device performs to specification and that the submitted I-Q View software is as safe and as functionally effective as the predicate software.

Clinical Testing:

The bench testing is significant enough to demonstrate that the I-Q View software is as good as the predicate software. All features and functionality have been tested and all specifications have been met. Therefore, it is our conclusion that clinical testing is not required to show substantial equivalence.

Conclusion:

The First Source, Inc. I-Q View does not introduce any new indications for use, nor does the use of the software result in any new potential hazards. First Source, Inc. considers the I-Q View digital imaging acquisition software to be substantially equivalent with the predicate with respect to safety and effectiveness.

§ 892.1680 Stationary x-ray system.

(a)
Identification. A stationary x-ray system is a permanently installed diagnostic system intended to generate and control x-rays for examination of various anatomical regions. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.(b)
Classification. Class II (special controls). A radiographic contrast tray or radiology diagnostic kit intended for use with a stationary x-ray system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.