K Number
K201043

Validate with FDA (Live)

Date Cleared
2020-05-12

(22 days)

Product Code
Regulation Number
892.1680
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Wireless Digital Flat Panel Detector (Model:Mars1717V-VSI, Mano4343W) are indicated for digital imaging solutions designed to provide general radiographic diagnosis for human anatomy including both adult and pediatic patients. They are intended to replace film/screen systems in all general-purpose diagnostic procedures. These two devices are not intended for mammography, dental applications.

Device Description

Mars1717V-VSI and Mano4343W Wireless Digital Panel Detectors (Hereinafter referred to as Mars1717V-VSI and Mano4343W) are the kind of wireless digital flat panel detectors. They support the single frame mode, with the key component of TFT/PD image sensor flat panel of active area: 427mm x 427mm. Two models Mars1717V-VSI and Mano4343W are totally same except for label and model name. The sensor plate of Mars1717V-VSI and Mano4343W is direct-deposited with CsI scintillator to achieve the conversion from X-ray to visible photon. The visible photons are transformed to electron signals by diode capacitor array within TFT panel, which are composed and processed by connecting to scanning and readout electronics, consequently to form a panel image by transmitting to PC through the user interface. The major function of the Mars1717V-VSI and Mano4343W is to convert the X-ray to digital image, with the application of high-resolution X-ray imaging. Both kinds of detectors are the key component of DR system, enable to complete the digitalization of the medical X-ray imaging with the DR system software.

AI/ML Overview

The provided text is a 510(k) summary for a medical device (Wireless Digital Flat Panel Detector). It describes the device, its intended use, and comparison to a predicate device to demonstrate substantial equivalence. However, it does not contain information about a study proving the device meets specific acceptance criteria based on diagnostic performance (e.g., sensitivity, specificity, accuracy).

The document focuses on demonstrating substantial equivalence primarily through technical characteristics, safety, and non-clinical performance, rather than clinical diagnostic accuracy or reader performance studies.

Therefore, most of the information requested in your prompt regarding acceptance criteria and studies to prove diagnostic performance (points 1-7, and 9 for the test set) cannot be extracted from this document. The document explicitly states: "clinical consideration may not necessary for changes in the pixel size and resolution with the same x-ray detection material and may not necessary for changes in the wireless functionality if non-clinical information is sufficient to support the substantial equivalence." This indicates that a detailed clinical performance study as you've described was not deemed necessary for this 510(k) submission.

Here's what can be extracted based on the provided text:

Preamble: This 510(k) submission for the Wireless Digital Flat Panel Detector (Models: Mars1717V-VSI, Mano4343W) focuses on demonstrating substantial equivalence to a predicate device (Mars1717XF-CSI, K183713) based on technical, electrical, mechanical, and non-clinical performance characteristics, rather than a clinical study evaluating diagnostic accuracy or human reader performance. Therefore, detailed information on diagnostic performance acceptance criteria, sample sizes for such studies, expert involvement for ground truth, or MRMC studies is not present.


1. A table of acceptance criteria and the reported device performance

  • Acceptance Criteria (Implicit via comparison to predicate and standards adherence): The device is substantially equivalent to the predicate device and meets applicable electrical safety, EMC, and biological evaluation standards. Specific quantitative diagnostic performance metrics are not provided as acceptance criteria in this document.
  • Reported Device Performance (Technical and Safety):
CharacteristicPredicate Device (Mars1717XF-CSI, K183713)Proposed Device (Mars1717V-VSI, Mano4343W)Notes/Performance Statement from Document
Image Matrix Size2832×2836 pixels3072 ×3072 pixels-
Pixel Pitch150μm139μm-
ADC Digitization16 bitSame-
Effective Imaging Area424.8 mm × 425.4 mm427 mm x 427 mm-
Spatial Resolution3.3lp/mm3.6lp/mm-
Modulation Transfer Function (MTF)0.49 at 1 lp/mm0.65 at 1 lp/mm-
Detective Quantum Efficiency (DQE)0.40 at 1 lp/mm (RQA5, 2.5μGy)Same-
Power ConsumptionMax. 20WMax. 30W-
Communication (Wireless Functionality)Wireless: IEEE 802.11a/b/g/n (2.4 GHz / 5 GHz)a) Wired (only for service) : Gigabit Ethernet (1000BASE-T) b) Wireless: IEEE 802.11a/b/g/n/ac (2.4 GHz / 5 GHz)Proposed device has higher wireless transmission speed.
Electrical Safety & EMCStandards met (predicate implied)IEC/ES 60601-1, IEC 60601-1-2 met"All test results are meet the standard requirements."
Biological EvaluationISO 10993-1 met (predicate implied)ISO 10993-1 met"evaluated... with the ISO 10993-1. And the evaluation results and test result assured the safety the same as the predicate device."
Image Quality Perception(Not explicitly stated for predicate)"No significant difference""There was no significant difference between the images of the Mars1717V-VSI/Mano4343W and those of the predicate device." (This is a subjective statement without a defined study or acceptance criteria.)

2. Sample size used for the test set and the data provenance

  • Sample Size: Not applicable/Not specified for diagnostic performance tests. The evaluation was primarily based on technical specifications and non-clinical considerations. The statement "There was no significant difference between the images of the Mars1717V-VSI/Mano4343W and those of the predicate device" implies some form of image comparison, but no formal test set size or study methodology is detailed.
  • Data Provenance: Not applicable for diagnostic test set; the document describes a technical and safety comparison.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

  • Not applicable/Not specified. Ground truth in the context of diagnostic performance studies is not mentioned, as such a study was not performed or submitted.

4. Adjudication method for the test set

  • Not applicable. No diagnostic test set or ground truth establishment method described.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

  • No, an MRMC comparative effectiveness study was not done. This device is a flat panel detector, not an AI-assisted diagnostic tool.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

  • No, this product is hardware (a flat panel detector), not a standalone algorithm.

7. The type of ground truth used

  • Not applicable. No diagnostic ground truth was established as a clinical diagnostic performance study was not performed. The evaluation relied on technical performance metrics, adherence to safety standards, and perceived equivalence of image quality to a predicate device.

8. The sample size for the training set

  • Not applicable. This device is a flat panel detector, not an AI/machine learning algorithm requiring a training set.

9. How the ground truth for the training set was established

  • Not applicable. See point 8.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health and Human Services logo on the left and the FDA logo on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.

May 12, 2020

iRay Technology Taicang Ltd. % Mr. Junjie Qian Registration & Regulation Affairs Engineer No. 33 Xinggang Rd., Taicang Port Economic & Technological Development Zone Taicang, Jiangsu 215434 CHINA

Re: K201043

Trade/Device Name: Wireless Digital Flat Panel Detector (Model: Mars1717V-VSI, Mano4343W) Regulation Number: 21 CFR 892.1680 Regulation Name: Stationary x-ray system Regulatory Class: Class II Product Code: MQB Dated: April 8, 2020 Received: April 20, 2020

Dear Mr. Qian:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part

{1}------------------------------------------------

801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

For

Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K201043

Device Name

Wireless Digital Flat Panel Detector (Model:Mars1717V-VSI, Mano4343W)

Indications for Use (Describe)

Wireless Digital Flat Panel Detector (Model:Mars1717V-VSI, Mano4343W) are indicated for digital imaging solutions designed to provide general radiographic diagnosis for human anatomy including both adult and pediatic patients. They are intended to replace film/screen systems in all general-purpose diagnostic procedures. These two devices are not intended for mammography, dental applications.

There are no differences between the 2 models except the model name and label.

Type of Use (Select one or both, as applicable)

☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510 (k) SUMMARY OF SAFETY AND EFFECTIVENESS

K201043

SECTION 6 - 1 of 12

{4}------------------------------------------------

510 (k) SUMMARY OF SAFETY AND EFFECTIVENESS

(As Required by 21 CFR 807.92)

1. Date Prepared [21 CFR 807.92(a)(1)]

April 8th, 2020

2. Submitter's Information [21 CFR 807.92(a)(1)]

Company Name:iRay Technology Taicang Ltd.
Company Address:No.33 Xinggang Road, Taicang Port Economic andTechnological Development Zone, Jiangsu, China 215434
Contact Person:Junjie Qian
Phone:0512-53690872
Fax:0512-53690872
Email:junjie.qian@iraygroup.com

3. Trade Name, Common Name, Classification [21 CFR 807.92(a)(2)]

Trade Name:Wireless Digital Flat Panel Detector
Common Name:Solid State X-Ray Imager
Model Name:Mars1717V-VSI
Mano4343W
Classification Name:Stationary x-ray system
Product Code:MQB
Regulation Number:21 CFR 892.1680
Device Class:Class II

{5}------------------------------------------------

4. Identification of Predicate Devices(s) [21 CFR 807.92(a)(3)[

The identification predicates within this submission are as follows:

Manufacturer:iRay Technology Taicang Ltd.
Trade Name:Wireless Digital Flat Panel Detector
Model Name:Mars1717XF-CSI
Product Code:MQB
Device Class:Class II
Classification Name:Stationary x-ray system
FDA 510 (k) #:K183713

5. Description of the Device [21 CFR 807.92(a)(4)]

Mars1717V-VSI and Mano4343W Wireless Digital Panel Detectors (Hereinafter referred to as Mars1717V-VSI and Mano4343W) are the kind of wireless digital flat panel detectors. They support the single frame mode, with the key component of TFT/PD image sensor flat panel of active area: 427mm x 427mm. Two models Mars1717V-VSI and Mano4343W are totally same except for label and model name. The sensor plate of Mars1717V-VSI and Mano4343W is direct-deposited with CsI scintillator to achieve the conversion from X-ray to visible photon. The visible photons are transformed to electron signals by diode capacitor array within TFT panel, which are composed and processed by connecting to scanning and readout electronics, consequently to form a panel image by transmitting to PC through the user interface. The major function of the Mars1717V-VSI and Mano4343W is to convert the X-ray to digital image, with the application of high-resolution X-ray imaging. Both kinds of detectors are the key component of DR system, enable to complete the digitalization of the medical X-ray imaging with the DR system software.

iRay SDK(include iDetector) is intend to supply API interface for DR system manufacturers.DR system manufacturer control the detector by SDK interface. SDK is not intended to be used directly by other users beside DR system manufacturers. The iRay SDK is unchanged from the predicate device.

SECTION 6 - 3 of 12

{6}------------------------------------------------

6. Intended Use [21 CFR 807.92(a)(5)]

6.1. Indications for use

Mars1717V-VSI and Mano4343W are indicated for digital imaging solutions designed to provide general radiographic diagnosis for human anatomy including both adult and pediatric patients. They are intended to replace film/screen systems in all general-purpose diagnostic procedures. These two devices are not intended for mammography, dental applications. There are no differences between the 2 models except the model name and label.

6.2. Suitable patient

They are suitable for providing digital X-ray imaging for DR system to provide general radiographic diagnosis for human anatomy including both adult and pediatric patients, but not intended for mammography or dental applications. The remaining notes depend on the DR system.

6.3. Processing of input and output

When flat panel detector works continuously, it can automatically distinguish Xray and output an imaging for diagnosis of disease, injury, or of any applicable health problem.

ItemPredicate Device:Wireless Digital Flat PanelDetectorProposed Device:Wireless Digital Flat PanelDetector
510(K)NumberK183713K201043
Intended UseThe Mars1717XF-CSI WirelessDigital Flat Panel Detector isindicated for digital imagingsame
ItemPredicate Device:Wireless Digital Flat PanelDetectorProposed Device:Wireless Digital Flat PanelDetector
solution designed for providinggeneral radiographic system in allgeneral-purpose diagnosticprocedures.
Indications forUseMars1717XF-CSI WirelessDigital Flat Panel Detector isindicated for digital imagingsolution designed for providinggeneral radiographic diagnosis ofhuman anatomy.It is intended to replaceradiographic film/screen systemsin all general-purpose diagnosticprocedures. This device is notintended for mammography ordental applications.We understand the Agency hasbecome aware of situations wheresolid state detectors inserted intoradiographic systems adverselyimpacted device performance dueto improper integration(reference:http://www.fda.gov/downloads/MedicalDevices/ResourcSame with Mars1717XF-CSI,additionally, in theconsideration of patient size(i.e., height, weight, body partthickness) and usable doserange, the detector could beused for general X raydiagnosis of usual body partfor adults and pediatricpatients..
ItemPredicate Device:Proposed Device:
Wireless Digital Flat PanelDetectorWireless Digital Flat PanelDetector
esforYou/Industry/UCM385149.pdf).
Below is a summary of theinformation from theMars1717XF-CSI user manualscovering key electromechanicaland computer requirementsneeded for X-ray system interfaceand integration.1. Mechanical interfacerequirements.2. Computer requirements3. Data communication interfacerequirements4. Electrical power requirements5. X-ray trigger interfacerequirements
Neither the Mars1717XF-CSIdetector nor its software act as anX-ray generator controller, andtherefore, the device is not subjectto Electronic Product RadiationControl (EPRC) performancestandards and reportingrequirements.
Predicate Device:Proposed Device:
ItemWireless Digital Flat PanelDetectorWireless Digital Flat PanelDetector
ClassificationNameStationary x-ray systemSame
Product CodeMQBSame
RegulationNumber21 CFR 892.1680Same
Panel:RadiologySame
Classification:IISame
X-Ray Absorber(Scintillator):CsISame
InstallationType:Wireless, PortableSame
ReadoutMechanism:Thin Film TransistorSame
Image MatrixSize:2832×2836 pixels3072 ×3072 pixels
Pixel Pitch:150μm139μm
ADCDigitization16 bitSame
EffectiveImaging Area:424.8 mm × 425.4 mm427 mm x 427 mm
SpatialResolution:3.3lp/mm3.6lp/mm
ModulationTransferFunction(MTF)0.49 at 1 lp/mm0.65 at 1 lp/mm
Predicate Device:Proposed Device:
ItemWireless Digital Flat PanelDetectorWireless Digital Flat PanelDetector
DetectiveQuantumEfficiency(DQE)$0.40 \text{ at } 1 \text{ lp/mm (RQA5, } 2.5\mu\text{Gy)}$Same
PowerConsumption:Max. 20WMax. 30W
Communications:(Wirelessfunctionality)Wireless: IEEE 802.11a/b/g/n (2.4 GHz / 5 GHz)a) Wired (only for service) : GigabitEthernet (1000BASE-T)b) Wireless: IEEE802.11a/b/g/n/ac (2.4 GHz / 5 GHz)
Imaging protectPlate:Carbon Fiber PlateSame
Cooling:Air coolingSame
Dimensions:460 mm × 460 mm × 15 mm460 mm × 460 mm × 15.3 mm
Operation:Temperature: +5 ~ +30 °CTemperature: +5 ~ +35 °C
Humidity: 10 ~ 80%(Non-Condensing)Humidity: 5 ~ 95%(Non-Condensing)
Atmospheric pressure: 70 ~ 106 kPaAtmospheric pressure: 55 ~ 106 kPa
Altitude: Max. 3000 metersAltitude: Max. 3000 meters
Storage andTransportation:Temperature: -20 ~ +50 °CTemperature: -10 ~ +55 °C
Humidity: 10 ~ 90%Humidity: 5 ~ 95%
ItemPredicate Device:Wireless Digital Flat PanelDetectorProposed Device:Wireless Digital Flat PanelDetector
( detector )(Non-Condensing)Atmospheric pressure: 70 ~ 106 kPaAltitude: Max. 3000 meters(Non-Condensing)Atmospheric pressure: 55 ~ 106 kPaAltitude: Max. 3000 meters
SoftwareiRay SDK(include iDetector) isintend to supply API interface forDR system manufacturers. DRsystem manufacturer control thedetector by SDK interface. SDKis not intended to use directly byother users beside DR systemmanufacturers.Same
Utilized FDAguidancedocuments1. Guidance for the Submissionof 510(k)'s for Solid State X-ray Imaging Devices;2. The 510(k) Program:Evaluating SubstantialEquivalence in PremarketNotifications[510(k)];3. Content of PremarketSubmissions for Managementof Cybersecurity in MedicalDevices;4. Radio Frequency WirelessTechnology in MedicalDevices.Same with Mars1717XF-CSI,additionally:1. Guidance for "PremarketAssessment of PediatricMedical Devices";2. Guidance for "PediatricInformation for X-rayImaging Device PremarketNotifications".

7. Technological Characteristic [21 CFR 807.92(a)(6)]

SECTION 6 - 4 of 12

{7}------------------------------------------------

SECTION 6 - 5 of 12

{8}------------------------------------------------

SECTION 6 - 6 of 12

{9}------------------------------------------------

iRay Technology Taicang Ltd.

[510(k)] Application

SECTION 6 - 7 of 12

{10}------------------------------------------------

iRay Technology Taicang Ltd.

[510(k)] Application

SECTION 6 - 8 of 12

{11}------------------------------------------------

SECTION 6 - 9 of 12

{12}------------------------------------------------

8. System requirements to operate with other radiographic system components

    1. Recommended Generator Specification:
      Energy range: 40~150keV

mA range: 10~1000mA (depending on the generator power)

ms range: 106300ms to produce 0.11000mAs (depending on the generator power) Note: To our best knowledge, the detector is compatible with the X-ray generators with the specifications described above. If you have any questions regarding the compatibility issue for other generators, please contact your distributor or iRay's service office.

    1. Application Program Interface (API) for system integration manufacturer Peripheral hardware: Mars1717V-VSI and Mano4343W are connected via wireless communication.
      recommended computer hardware requirement is :
  • Windows 7 32/64 bit

  • •Intel Core i7 3.6G

  • •4G DDR3

  • •640 G

  • · Intel Pro EXP9301CT PRO, Gigabit Network Adapter with PCIe interface

    1. X-ray exposure mode

The AED trigger module is a unit can connect X-ray signal in the Mars1717V-VSI and Mano4343W. Once there is X-ray generator exposure exist, the AED trigger module will detect the X-ray radiation and output signal to the detector. Until the exposure finished, the detector will receive a signal which represent the end of exposure from the inner trigger module and begin to acquire the image.

SECTION 6 - 10 of 12

{13}------------------------------------------------

9. Substantial Equivalence [21 CFR 807.92(b)(1) and 807.92(b)(2)]

  • Electrical Safety and EMC testing: 1)
    Electrical, mechanical, environmental safety test according to IEC/ES 60601-1 was performed, and EMC testing was also conducted in accordance with IEC 60601-1-2. All test results are meet the standard requirements.

    1. Biological Evaluation:
      The materials of the detector which contact operators' or patient's skin have been evaluated with the ISO 10993-1. And the evaluation results and test result assured the safety the same as the predicate device.
    1. Non-clinical Considerations:
      One modification from the predicate device to Mars1717V-VSI and Mano4343W is geometric design, related to Amorphous Silicon (A-Si) panel size and structure size design. Another modification is wireless functionality, proposed device has higher wireless transmission speed.

The non-clinical studies have been performed and the results have shown that sections of the non-clinical consideration mentioned in the 'Guidance for the Submission of 510(k)s for Solid State X-ray Imaging Devices' are substantially equivalent to the non-consideration of predicate devices on the Market (Mars1717XF-CSI, K183713).

    1. Clinical Consideration:
      Intended use, fundamental scientific technology, regulatory requirement, non-clinical performance, labeling, quality-assurance program and software keep the same with those of predicate device. Additionally, as mentioned in clinical considerations in 'Guidance for the Submission of 510(k)s for Solid State X-ray Imaging Devices', clinical consideration may not necessary for changes in the pixel size and resolution with the same x-ray detection material and may not necessary for changes in the wireless functionality if non-clinical information is sufficient to support the substantial equivalence.

There was no significant difference between the images of the Mars1717V-VSI/Mano4343W and those of the predicate device.

SECTION 6 - 11 of 12

{14}------------------------------------------------

10. Conclusion [21 CFR 807.92(b)(3)]

In accordance with the Federal Food, Drug and Cosmetic Act, 21 CFR Part 807 and based on the information provided in this premarket notification, iRay Technology Taicang Ltd. concludes that Mars1717V-VSI and Mano4343W are substantially equivalent to predicate device with regards to safety and effectiveness.

SECTION 6 - 12 of 12

§ 892.1680 Stationary x-ray system.

(a)
Identification. A stationary x-ray system is a permanently installed diagnostic system intended to generate and control x-rays for examination of various anatomical regions. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.(b)
Classification. Class II (special controls). A radiographic contrast tray or radiology diagnostic kit intended for use with a stationary x-ray system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.