K Number
K161730

Validate with FDA (Live)

Date Cleared
2016-12-08

(168 days)

Product Code
Regulation Number
892.1680
Age Range
All
Reference & Predicate Devices
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Mars1417V-PSI Wireless Digital Flat Panel Detector is indicated for digital imaging solution designed for providing general radiographic diagnosis of human anatomy. It is intended to replace radiographic film/screen systems in all general-purpose diagnostic procedures. This device is not intended for mammography or dental applications.

Device Description

Mars1417V-PS1 Wireless Digital Flat Panel Detector is a kind of wireless digital flat panel detector. It supports the single frame mode, with the key component of TFT/PD image sensor flat panel of active area: 36cm×43cm.

The sensor plate of Mars1417V-PSI Wireless Digital Flat Panel Detector is direct-deposited with Gd2O2S scintillator to achieve the conversion from X-ray to visible photon. The visible photons are transformed to electron signals by diode capacitor array within TFT panel, which are composed and processed by connecting to scanning and readout electronics, consequently to form a panel image by transmitting to PC through the user interface.

The major function of the Mars1417V-PSI Wireless Digital Flat Panel Detector is to convert the X-ray to digital image, with the application of high resolution X-ray imaging. This detector is the key component of DR system, enables to complete the digitalization of the medical Xray imaging with the DR system software.

The iRay DR used for getting Digital X -ray radiography images from the flat panel detectors. iRay DR is used to handle the DICOM protocol (DICOM 3.0), iRay DR is responsible for the DR equipment management, acquisition and processing functions, to provide patient registration, scanning, image processing and forwarding, and other functions ..

AI/ML Overview

The provided document is a 510(k) Summary for the iRay Technology (Shanghai) Ltd. Wireless Digital Flat Panel Detector (Mars1417V-PSI), which is a digital X-ray system. The document focuses on demonstrating substantial equivalence to a predicate device, rather than providing detailed acceptance criteria and a standalone study for the device's clinical performance.

However, based on the information provided, here's a breakdown of the requested elements:

1. Table of Acceptance Criteria and Reported Device Performance

The document does not explicitly present a table of "acceptance criteria" for clinical performance. Instead, it relies on demonstrating substantial equivalence by comparing technological characteristics and an absence of significant difference in clinical images compared to a predicate device (ViZion DR+Wireless, K152279).

Based on the Nonclinical Considerations section, the performance characteristics evaluated are:

Acceptance Criteria (Implicit from Nonclinical Studies)Reported Device Performance (Mars1417V-PSI vs. Predicate)
Detective Quantum Efficiency (DQE)0.27 at 0.5 lp/mm (RQA5, 3.2µGy) - Matches predicate
Quantum limited performanceNot explicitly quantified, implied to be equivalent to predicate
Modulation Transfer Function (MTF)0.75 at 0.5 lp/mm - Matches predicate
Effects of aliasingNot explicitly quantified, implied to be equivalent to predicate
Sensitivity linearityNot explicitly quantified, implied to be equivalent to predicate
LagNot explicitly quantified, implied to be equivalent to predicate
Change in detection sensitivityNot explicitly quantified, implied to be equivalent to predicate
Dose requirement and reciprocity changesNot explicitly quantified, implied to be equivalent to predicate
Stability of device characteristics with timeNot explicitly quantified, implied to be equivalent to predicate
Uniformity of device characteristicNot explicitly quantified, implied to be equivalent to predicate
Noise power spectrum (NPS)Not explicitly quantified, implied to be equivalent to predicate
Spatial resolutionMin. 3.4lp/mm - Matches predicate
Image Acquisition timeNot explicitly quantified, implied to be equivalent to predicate
Black levelNot explicitly quantified, implied to be equivalent to predicate
Software functionality (iRayDR)Passes 83 test cases, complies with intended design specification
Electrical Safety, EMC, Biological EvaluationMeets standard requirements (IEC/ES 60601-1, IEC/EN 60601-1-2, ISO 10993-1)
Clinical Image ComparisonNo significant difference between images of Mars1417V-PSI and predicate device

2. Sample Size Used for the Test Set and Data Provenance

  • Sample Size for Clinical Image Comparison: 30 clinical images.
  • Data Provenance: The document does not specify the country of origin for the clinical images, nor does it explicitly state if the study was retrospective or prospective. Given the context of a 510(k) submission for substantial equivalence based on a comparison, it's highly likely these were retrospective images or images specifically acquired for comparison purposes but not necessarily part of a broader prospective clinical trial.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

The document states: "A concurrence study of 30 clinical images was conducted to compare the performance of the Mars1417V-PSI to that of the predicate device (ViZion DR+Wireless,K152279)."

It does not specify the number of experts used or their qualifications for evaluating these 30 clinical images. It only mentions that "no significant difference" was found.

4. Adjudication Method for the Test Set

The document does not specify any adjudication method (e.g., 2+1, 3+1, none) for the clinical image comparison, nor does it detail how the "no significant difference" conclusion was reached.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

A multi-reader multi-case (MRMC) comparative effectiveness study comparing human readers with and without AI assistance was not done. The study described is a comparison of the device's images to those of a predicate device, not an evaluation of AI assistance for human readers.

6. Standalone Performance Study

A standalone performance study (algorithm only without human-in-the-loop performance) was performed in terms of non-clinical tests evaluating physical characteristics like DQE, MTF, spatial resolution, and others. The software (iRayDR) also underwent standalone testing with 83 test cases. The clinical image comparison, while assessing image quality, doesn't constitute a standalone algorithm performance in the context of an AI device, as this is a digital flat panel detector, not an AI diagnostic algorithm.

7. Type of Ground Truth Used

For the 30 clinical images used in the comparison study, the "ground truth" implicitly referred to the diagnostic quality of the images produced by the predicate device. The study aimed to show "no significant difference" in image quality between the proposed device and the predicate device, implying that the diagnostic utility of the predicate's images served as the reference. There is no mention of pathology or outcomes data as ground truth.

8. Sample Size for the Training Set

The document does not provide any information regarding a training set sample size. This device is a digital flat panel detector, which converts X-rays into digital images, and the software mentioned (iRayDR) handles image processing and DICOM protocols. This is not an AI diagnostic algorithm that requires a specific training set to learn to identify pathology.

9. How the Ground Truth for the Training Set Was Established

As no training set is mentioned in the context of an AI diagnostic algorithm, this question is not applicable. The software components described appear to be for image acquisition, processing, and management, not for automated diagnosis based on
learned patterns.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the seal of the Department of Health & Human Services - USA. The seal features a stylized eagle with three heads facing right, representing the department's mission to protect the health of all Americans and provide essential human services. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the eagle.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

December 8, 2016

iRay Technology (Shanghai) Ltd. % Mr. Wei Pan Registration & Regulatory Affairs Manager Room 202, Building 7, No. 590, Ruiqing Road Shanghai, 201201 CHINA

Re: K161730

Trade/Device Name: Wireless Digital Flat Panel Detector Regulation Number: 21 CFR 892.1680 Regulation Name: Stationary x-ray system Regulatory Class: II Product Code: MOB Dated: November 11, 2016 Received: November 18, 2016

Dear Mr. Pan:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{1}------------------------------------------------

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Michael D'Hara

For

Robert Ochs, Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

Indications for Use

510(k) Number (if known)

K161730

Device Name Wireless Digital Flat Panel Detector

Indications for Use (Describe)

Mars1417V-PSI Wireless Digital Flat Panel Detector is indicated for digital imaging solution designed for providing general radiographic diagnosis of human anatomy. It is intended to replace radiographic film/screen systems in all general-purpose diagnostic procedures. This device is not intended for mammography or dental applications.

Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the
time to review instructions, search existing data sources, gather and maintain the data needed and complete
and review the collection of information. Send comments regarding this burden estimate or any other aspect
of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services

Food and Drug Administration

Office of Chief Information Officer

Paperwork Reduction Act (PRA) Staff

PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

Form Approved: OMB No. 0910-0120

Expiration Date: January 31, 2017

See PRA Statement below.

{3}------------------------------------------------

SECTION 7

510 (k) SUMMARY OF SAFETY AND EFFECTIVENESS

SECTION 7 - 1 of 8

{4}------------------------------------------------

510 (k) SUMMARY OF SAFETY AND EFFECTIVENESS

(As Required by 21 CFR 807.92)

1. Date Prepared [21 CFR 807.92(a)(1)]

October 19, 2015

2. Submitter;s Information [21 CFR 807.92(a)(1)]

Company Name:iRay Technology (Shanghai) Ltd.
Company Address:RM 202, Building 7, No. 590, Ruiqing RD., Zhangjiang East,
Pudong, Shanghai, China 201201
Contact Person:Mr. Wei Pan
Phone:+86 021-50720560
Fax:+86 021-50720561
Email:wei.pan@iraychina.com

3. Trade Name, Common Name, Classification [21 CFR 807.92(a)(2)]

Trade Name:Wireless Digital Flat Panel Detector
Common Name:Flat Panel Detector
Model Name:Mars1417V-PSI
Classification Name:Stationary X-Ray System
Product Code:MQB
Regulation Number:21 CFR 892.1680
Device Class:Class II

4. Identification of Predicate Devices(s) [21 CFR 807.92(a)(3)[

The identification predicates within this submission are as follows:

Manufacturer:Viztek LLC
Trade Name:ViZion DR + Wireless
Model Name:ViZion DR + Wireless
Product Code:MQB

SECTION 7 - 2 of 8

{5}------------------------------------------------

Classification Name: Stationary X-Ray System K152279 FDA 510 (k) #:

5. Description of the Device [21 CFR 807.92(a)(4)]

Mars1417V-PS1 Wireless Digital Flat Panel Detector is a kind of wireless digital flat panel detector. It supports the single frame mode, with the key component of TFT/PD image sensor flat panel of active area: 36cm×43cm.

The sensor plate of Mars1417V-PSI Wireless Digital Flat Panel Detector is direct-deposited with Gd2O2S scintillator to achieve the conversion from X-ray to visible photon. The visible photons are transformed to electron signals by diode capacitor array within TFT panel, which are composed and processed by connecting to scanning and readout electronics, consequently to form a panel image by transmitting to PC through the user interface.

The major function of the Mars1417V-PSI Wireless Digital Flat Panel Detector is to convert the X-ray to digital image, with the application of high resolution X-ray imaging. This detector is the key component of DR system, enables to complete the digitalization of the medical Xray imaging with the DR system software.

The iRay DR used for getting Digital X -ray radiography images from the flat panel detectors. iRay DR is used to handle the DICOM protocol (DICOM 3.0), iRay DR is responsible for the DR equipment management, acquisition and processing functions, to provide patient registration, scanning, image processing and forwarding, and other functions ..

6. Intended Use [21 CFR 807.92(a)(5)]

6.1. Intended Use

Mars1417V-PSI Wireless Digital Flat Panel Detector is indicated for digital imaging solution designed for providing general radiographic diagnosis of human anatomy. It is intended to replace radiographic film/screen systems in all general-purpose diagnostic procedures. This device is not intended for mammography or dental applications.

6.2. Suitable patient

It is suitable for providing digital X-ray imaging for DR system conventional photography but not intended for mammography or dental applications. The remaining notes depend on the DR system.

6.3. Processing of input and output

SECTION 7 - 3 of 8

{6}------------------------------------------------

When FPD works continuously, it can automatically distinguish X-ray and output an imaging for diagnosis of disease, injury, or of any applicable health problem

7. Technological Characteristic [21 CFR 807.92(a)(6)]

ItemProposed Device:Wireless Digital Flat PanelDetectorPredicate Device:ViZion DR + Wireless
510(K)NumberTo be assignedK152279
Intended UseThe Mars1417V-PSI WirelessDigital Flat Panel Detector isindicated for digital imagingsolution designed for providinggeneral radiographic system in allgeneral purpose diagnosticprocedures.ViZion DR + Wireless is intendedfor digital image capture use ingeneral radiographicexaminations, whereverconventional screen-film systemsmay be used, excludingfluoroscopy, angiography andmammography.
ClassificationNameStationary X-ray systemStationary X-ray system
Product CodeMQBMQB
RegulationNumber21 CFR 892.168021 CFR 892.1680
Panel:RadiologyRadiology
Classification:IIII
X-Ray Absorber(Scintillator):Gd2O2SGd2O2S
Installation Type:Wireless, PortableWireless, Portable
ReadoutMechanism:Thin Film TransistorThin Film Transistor
Image Matrix Size:2304 × 2800 pixels2304 × 2800 pixels
Pixel Pitch:150μm150μm
ADC Digitization14 bit14 bit
ItemProposed Device:Wireless Digital Flat PanelDetectorPredicate Device:ViZion DR + Wireless
Effective ImagingArea:355 mm × 434 mm355 mm × 434 mm
Spatial Resolution:Min. 3.4lp/mmMin. 3.4lp/mm
ModulationTransferFunction(MTF)0.75 at 0.5lp/mm0.75 at 0.5 lp/mm
DetectiveQuantumEfficiency(DQE)0.27 at 0.5 lp/mm (RQA5,3.2µGy)0.27 at 0.5 lp/mm (RQA5,3.2µGy)
PowerConsumption:Max. 13WMax. 13W
SoftwareOutputs a DICOM image.Outputs a DICOM image.
DICOM 3YesYes
Communications:Wired: Gigabit Ethernet(1000BASE-T)Wireless: IEEE 802.11a/b/g/n(2.4 GHz / 5 GHz)Wired: Gigabit Ethernet(1000BASE-T)Wireless: IEEE 802.11a/b/g/n (2.4 GHz / 5 GHz)
Imaging protectPlate:Carbon Fiber PlateCarbon Fiber Plate
Cooling:Air coolingAir cooling
Dimensions:384 mm × 460 mm × 15 mm384 mm × 460 mm × 15 mm
Operation:Temperature: +5 ~ +35°CHumidity: 30 ~ 75%(Non-Condensing)Atmospheric pressure: 70 ~ 106 kPaAltitude: Max. 3000 metersTemperature: +5 ~ +35°CHumidity: 30 ~ 75%(Non-Condensing)Atmospheric pressure: 70 ~ 106 kPaAltitude: Max. 3000 meters
ItemProposed Device:Wireless Digital Flat PanelDetectorPredicate Device:ViZion DR + Wireless
Storage andTransportation:Temperature: -20 ~ +55 °CTemperature: -20 ~ +55 °C
Humidity: 10 ~ 90%(Non-Condensing)Humidity: 10 ~ 90%(Non-Condensing)
Atmospheric pressure: 70 ~ 106 kPaAtmospheric pressure: 70 ~ 106 kPa
Altitude: Max. 3000 metersAltitude: Max. 3000 meters
SoftwareiRay DROpal-RAD™ (K063337)
The iRay DR used for gettingOpal - RAID is a software suite of
Digital X -ray radiographyweb based PACS applications that
images from the flat panelwas developed specifically to
detectors. iRay DR is used tohandle the DICOM protocol, for
handle the DICOM protocolboth transmitting and viewing
(DICOM 3.0),DICOM images and data
iRay DR is responsible for theelements. The applications were
DR equipment management,developed so that access to the
acquisition and processingPACS can occur from any
functions, to provide patientMicrosoft Windows computer
registration, scanning, imagewith internet capabilities, and
processing, image forwarding,offer an interface that users find to
image printing and otherbe quite intuitive after some initial
functions.learning. The Opal-RAID
applications deal with all manner
of DICOM images These images
can be viewed, manipulated, annotated, transmitted to other facilities, printed, animated and stored using the Opal-RAID suite. The system does not produce any original medical images.

SECTION 7 - 4 of 8

{7}------------------------------------------------

iRay Technology (Shanghai) Ltd

[510(k)] Application

SECTION 7 - 5 of 8

{8}------------------------------------------------

iRay Technology (Shanghai) Ltd

[510(k)] Application

SECTION 7 - 6 of 8

{9}------------------------------------------------

8. System requirements to operate with other radiographic system components

    1. Recommended Generator Specification:
      Energy range: 40~150kVp

mA range: 10~1000mA (depending on the generator power)

ms range: 106300ms to produce 0.11000mAs (depending on the generator power) Note: To our best knowledge, the detector is compatible with the X-ray generators with the specifications described above. If you have any questions regarding the compatibility issue for other generators, please contact your distributor or iRay's service office.

    1. Application Program Interface (API) for system integration manufacturer Peripheral hardware: Mars1417V-PSI detector connected via wireless or wired communication.
      Operating System: Windows XP/7 32/64bit CPU: Intel Core i73.6G Memory: 4G DDR3 Hard Disk: 640 G LAN Card: Intel Pro EXP9301CT PRO Gigabit Network Adapter with PCIe interface
    1. X-ray exposure mode
      The inner trigger module is a unit can connect X-ray signal in the Mars1417V-PSI. Once there is X-ray generator exposure exist, the inner trigger module will detect the X-ray radiation and output signal to the detector. Until the exposure finished, the detector will receive a signal which represent the end of exposure from the inner trigger module and begin to acquire the image.

9. Substantial Equivalence [21 CFR 807.92(b)(1) and 807.92(b)(2)]

  • Electrical Safety and EMC testing: 1) Electrical, mechanical, environmental safety and performance testing according to IEC/ES 60601-1 was performed, and EMC testing was also conducted in accordance with IEC/EN 60601-1-2. All test results are meet standard requirements.
    1. Biological Evaluation:

SECTION 7 - 7 of 8

{10}------------------------------------------------

The materials of the detector which contact operators' skin have been evaluated with the ISO 10993-1. And the evaluation results and test result assured the safety the same as the predicate device.

    1. Nonclinical Considerations:
      The non-clinical studies have been performed and the results have shown that the Mars1417V-PSI wireless digital X-ray flat panel detector is substantially equivalent to the predicate devices on the Market (ViZion DR+Wireless, K152279 ): Detective quantum efficiency (DQE), Quantum limited performance, Modulation transfer function (MTF), Effects of aliasing, Sensitivity linearity, Lag, Change in detection sensitivity, Dose requirement and reciprocity changes, Stability of device characteristics with time, Uniformity of device characteristic, Noise power spectrum(NPS), Spatial resolution, Image Acquisition time, & Black level.

According to the Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices, the software iRayDR classifies the hazards, defines requirements specification and design specification, all the specification pass the 83 test cases and complies the intended design specification.

    1. Clinical Consideration:
      A concurrence study of 30 clinical images was conducted to compare the performance of the Mars1417V-PSI to that of the predicate device (ViZion DR+Wireless,K152279). Clinical images were provided; these images were not necessary to establish substantial equivalence based on the differences from the predicate but they provide further evidence in addition to the laboratory performance data to show that the complete system works as intended.

There was no significant difference between the images of the Mars1417V-PSI and those of the predicate device.

10. Conclusion [21 CFR 807.92(b)(3)]

In accordance with the Federal Food, Drug and Cosmetic Act, 21 CFR Part 807 and based on the information provided in this premarket notification, iRay Technology (Shanghai) Ltd. Concludes that iRay Mars1417V-PSI Wireless Digital Flat Panel Detectors is substantially equivalent to predicate device with regards to safety and effectiveness.

SECTION 7-8 of 8

§ 892.1680 Stationary x-ray system.

(a)
Identification. A stationary x-ray system is a permanently installed diagnostic system intended to generate and control x-rays for examination of various anatomical regions. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.(b)
Classification. Class II (special controls). A radiographic contrast tray or radiology diagnostic kit intended for use with a stationary x-ray system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.